Technical Publication SMS-M-002 (Rev. 1.0)

Service Manual

Model: PX-300HF

Important! Please read this manual thoroughly and use the system correctly before using the system. This manual desires to keep in promised place to use always again after reading.

EcoRay Co., Ltd. 621-14, Docheon-Dong, Gwangsan-Gu, Gwangju, Korea 506-301 Tel: +82-62-952-3251 / Fax: +82-62-951-5261 Seoul Office: #402, Cygnus B/D, 115-12, Nonhyun-Dong, Kangnam-Ku, Seoul, Korea 135-821 Tel) +82 70 7510 3402

Fax) +82 70 8630 3420 E-mail: [email protected]

EcoRay

PX-300HF

MPORTANT!

X-RAY PROTECTIONS!

X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS SAFETY INSTRUCTIONS ARE STRICTLY OBSERVED IN USE! Accordingly, when use this product should be thoroughly read the "Radiation Safety Control Regulation" and then compliance with it in use. This product is serious injury's danger if do not keep use guideline or regulation therefore, read and understood this manual before in use. This product was designed by safety level (SL) presenting in Radiation Safety Control Regulation. However, if do not keep using method and Radiation Safety Control Regulation, will not be able to avoid serious injury's danger from radiation. Therefore, to prevent accident from radiation, everyone who work connected with radioactive emission must understand radiation safety control regulation and understand product manual fulfill thorough education and formality. Our company recognizes enough about danger of excessive exposure to operator, installation people, and maintenance people. Therefore, our company strongly advises that operator, installation people, maintenance people must be educated about radiation safety control regulation and should be used this product when understand fluently about all knowledge of operation this product. Also, our company informs that operator, installation people, and maintenance people have a responsibility to be educated fluently about radiation safety control regulation and understood about product use method. Also, our company has no responsibility about injury or calamity that occurs does not keep use method of this product and radiation safety control regulation. Therefore, our company strongly advises that to be used this product after that person in charge of installation, maintenance and operator should understood radiation safety control regulation of domestic or foreign. Thanks for purchasing product of EcoRay. Before using the system, please read this manual thoroughly and use the system correctly. The precautions and prohibitions are seen though the manual is classified as follows.

(Rev. 1.0)

1

Service Manual

EcoRay

PX-300HF

NOTICE AND CONTRAINDICATION!

States a direct danger that may cause death or serious injury if it is not avoided.

States an indirect or potential danger that may cause death or serious injury if it is not avoided.

State a danger that may cause slight or medium injury or may cause damage in equipment or fire if it is not avoided. NOTE ☞ States the information, which helps to use the system correctly. CONTRAINDICATION! This x-ray system is not suitable for therapeutic applications.

(Rev. 1.0)

2

Service Manual

EcoRay

PX-300HF

Precautions in the use of X-ray system (For the Safety and Prevention of Danger) 1. Only one experienced technician should operate the equipment. 2. When installing the equipment, pay attention to the following items. 2.1 Do not install it near water faucet or similar equipment. 2.2 Install it away from potential sources of problems such as abnormal pressure, temperature or humidity, drafts, direct sunlight, dust, chlorine or sulphur gas. 2.3 During transportation and operation of the equipment, avoid tilting, vibration and sharp impact against it. 2.4 Keep the equipment away from the areas where chemicals or gases are stored. 2.5 Use only the correct electrical power source with matching frequency, voltage and current (or wattage). 2.6 Check the condition of the battery power source (Power Supply Unit). (Power and polarity before operating the equipment.) 3. Before operating the equipment, pay attention to the following items: 3.1 Check the conditions of switch contacts, polarity, and make sure the equipment performs correctly. 3.2 Confirm that the ground is connected properly. 3.3 Check all wiring for proper and correct connections. 3.4 Check the condition of the external electric circuit breaker. 3.5 Circuit breaker will be directly connected to the system and do not connect with another system. 3.6 When the equipment has been leaved for a long time (more than 10days), You must warm-up equipment and X-ray tube sufficiently before use. 4. While operating the equipment, pay attention to the following items: 4.1 Do not ever-exceed time or the amount of equipment use needed for diagnosis. 4.2 Observe the equipment and the patient continuously for detection of problems. 4.3 When a problem is detected with the equipment or the patient, take proper action to stop the equipment without harming the patient. 4.4 Do not let the equipment touched by the patient. 5. After operating the equipment, pay attention to the following items: 5.1 Turn of the switch and then, turn off the main power switch. 5.2 Clean the part that contacted with body. 6. When the equipment is out of order, call a certificated engineer. (Do not try to repair it.) 7. Do not modify any part of the equipment. 8. The equipment must be checked periodically for good operation condition without troubles. 9. Do not use x-ray system connecting with other system.

(Rev. 1.0)

3

Service Manual

EcoRay

PX-300HF

1. The owner/user has a responsibility for management and maintenance of this equipment. 2. This equipment must be operated by only authorized personnel. 3. Contact authorized service engineer when repair or inspection of the inside of the equipment. 4. In general, modifications are strictly prohibited by the Regulations. Please contact our agency if it is needed to modify the device. 5. Do not use this system in conjunction with the other machines because seriously danger could be happened unexpectedly.

Periodic Inspection 1. Maintenance is required to maintain safety and performance of this system for a long time. 2. Maintenance must be done periodically with maintenance plan. 3. This system must be inspected from the inspection office per every 3 years.

Warranty 1. This warranty shall apply only to defects that arise within 12months of purchase. 2. This product could have a free of charge service if troubles are occurred during the warranty period. 3. The warranty does not cover defects caused by abuse, mishandling, accidents or improper installation. 4. The following will not be covered under the warranty. 4-1. Modifications to the product, or repairs made by unauthorized person but manufacturer or authorized person by manufacturer. 4-2. Acts of God. Including fire, flood, storm, power or voltage surges. 4-3. Improper use or installation. 4-4. Defects that arise in conjunction with the use of other machines. 4-5. Use of improper parts or modification the circuit. 4-6. Improper use 5. In case of repaired during the warranted period, User shall advise the result of test to manufacture.

(Rev. 1.0)

4

Service Manual

EcoRay

PX-300HF

Precautions in Use When using this equipment, please observe the following precautions for the safety of the operator and the patient.

1. Improper use of the X-ray equipment might cause the operator or the patient to be accidentally exposed to X-ray radiation. 2. During X-ray radiation, any person other than the subject the patient should not stay in the room. 3. X-ray room must have radiation protection facility by "Radiation Safety Control Regulation"

1. Only qualified personnel should handle the equipment 2. Electric shock or leakage of electricity can happen unless ground is not installed necessarily. 3. Keep the equipment dry 4. Otherwise it may cause an electric shock to the operator of the patient and if any liquid or water flew into electric device turn off the power and contact company or service agency.

NOTE ☞ For the safety of operator and patient, opening the X-ray generator cover, X-ray tube, Collimator and other equipment cover are strictly prohibited. Only qualified engineer could open the cover.

(Rev. 1.0)

5

Service Manual

EcoRay

PX-300HF

Scope 1. This manual contains information of system operation, components, control keys, switches and hazards and technical reference information. 2. This manual may not be reproduced in whole or in part without the written permission of EcoRay Co., Ltd. 3. The text of this manual is a translation based on the Korean alphabet (Hangŭl) original.

EcoRay Co., Ltd. 621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea

Revision History

Revision No. Rev. 1.0

(Rev. 1.0)

Revision Date 2008-03-10

6

Change Description Release version

Service Manual

EcoRay

PX-300HF

Table of Contents Chapter 1 1. Pre-installation ……………………………………………………………….……….………………………….……….…. 10 1.1 An Overview ……………………………………………………………………………………………………….…………………….……..… 10 1.2 Responsibility of Customers ……………………………….………………………………………….………….…………….….……. 10 1.3 Constructions and Function .…………………………………………….………….…...………………………………………..….…. 10 1.4 General details …………..……………….………………………….……………………………..…………………………………….….…. 11 1.5 Service Numbers …………..……………………………………………...…………………………….…………………………….………. 11 1.6 Product Pre-installation Data ..……….…………………………………………………………………………………………………. 12 1.7 X-ray room Requirements ……...……………………………..………………………………………………………………….………. 13

Chapter 2 2. Installation …………………….…………………………………..…………………………………………………………… 18 2.1 An Overview ……………………………………………………………………………………………………………………………….………. 18 2.2 Pre-installation checks points and Tools and Testers ………………………………………..….…………………………... 18 2.3 Pre-installation Check ………………………………..……………..…………………………………………..…….……………………… 19 2.4 Installation of Generator ……….………………………………………………………..……………………….….……………………... 20 2.5 Installation of Tube Stand ………………………………………………………………………………….……….……………………... 27 2.6 System Cable Wiring ……………………………………………………………….………………………………………..……………... 35

Chapter 3 3. Technical Information …………….………………………………..………………….……………….……………..……. 37 3.1 An Overview …………………………………………………………………………………….……………………………..…………..……… 37 3.2 Installation information ……………………………………………………….…………………………..…………….………....…..…… 37 3.3. Exposure condition data ………………………………………………………………..……………………………..………...………… 37 3.4 Maintenance History ……………………………………………………………………………..…………………………….…..………… 38

Chapter 4 4. Calibration …………….…………………………………………….………………………….……………………………… 40 4.1 An Overview ..……………………………………………………………………………………………….…….……………………………… 40 4.2 Introduction ………………………………………….……………………………………………….…………………………………...……… 40 4.3 Needed tool and testers when Calibration …..……………………………………………………………………………..…… 40 4.4 Check point before calibration …..………………………………….………………………………………………….………...…… 40 4.5 Consist of Calibration Program …..…………………………………………………………………………..……………….……..… 41 4.6 How to use Calibration program ………………………………………………………………………………….….…….…….…… 41

(Rev. 1.0)

7

Service Manual

EcoRay

PX-300HF

4.7 Time of Calibration …………………………………………………………………………………………………….………...…….…… 42 4.8 Must calibrated when Main Parts are repaired or replaced ………………………….…………….…………..….…… 43

Chapter 5 5. Trouble Shooting ………………………………..……………………………..………….….………………….…………. 45 5.1 An Overview ………………………………………………….……………………………………………….……..……………..……….…… 45 5.2 Error type ……………………………………………………………………………………………………………...………………..……… 45 5.3 Trouble Shooting ……………………………………………………………………………………….………....……………………...…… 51 5.4 Handle Bar (SMS-TS-340 Board) …………………………….…..…………………………………..……………………...……..… 73

Chapter 6 6. Circuit Diagram ……………………………………………………………………………….…………………………...…. 75

(Rev. 1.0)

8

Service Manual

EcoRay

PX-300HF

Chapter 1

Pre-installation High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0)

9

Service Manual

EcoRay

PX-300HF

1. Pre-installation

1.1 An Overview 1.1.1 This manual contains operation instructions for radiological technicians, radiologists and other medical personnel who are trained to operate radiation equipment. 1.1.2 These instructions are not suitable as a replacement for components or exchange of circuit or medical training. 1.1.3 This system was made for trained personnel of medical examination, diagnosis and before surgical operation and after. 1.1.4 This manual contains of PX-300HF product information, its associated components, needed environment and electrical requirements. 1.2 Responsibility of Customers 1.2.1 Customer has a responsibility for planning and preparation of installation. 1.2.2 The following items ((1.2.3) are must be checked in pre-installation site by customer and additional work may be needed by depending on the specific site circumstances. 1.2.3 Install required material before delivery of system components. 1. Complete room floor, ceiling, wall and cleaning. 2. Electric power facility (Circuit Breaker, ground, duct, raceway, etc.) 3. Provide current room dimension, hallway and entry door sizes. 4. Cost of alteration and modifications when not specifically provided in sales contract.

NOTE ☞ Complete and proper pre-installation will avoid delays and confusion.

1.3 Constructions and Function This X-ray system is consist of Tube Stand which support X-ray tube and moveable, Console, X-ray generator, Patient table, and X-ray tube unit. 1.3.1 Tube Stand is consist of main power supply unit, electric circuit, handle bar and mechanical devices and could move Tube stand with floor rail. 1.3.2 Console is consist of each switch and could adjust exposure values and display it. 1.3.3 X-ray generator is consists of control unit and high voltage transformer unit and control x-ray exposure values and make high voltages by console. 1.3.4. X-ray tube unit consist of X-ray tube, Collimator and Handle bar and could move Tube Stand

(Rev. 1.0)

10

Service Manual

EcoRay

PX-300HF

and adjust x-ray exposure area. 1.3.5 X-ray radiation will be made by pushing of X-ray switch of console or pushing of hand switch. 1.3.6 X-ray radiation will permeate patient from x-ray tube to bucky device and after permeation, could archive x-ray images by x-ray film or digital detector. 1.4 General details 1.4.1 Manufacturer EcoRay Co., Ltd. 621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea 1.4.2 Place of Manufacture of the PX-300HF EcoRay Co., Ltd. 621-14, Docheon-dong, Gwangsan-gu, Gwang-ju, 506-301, Korea 1.4.3 Representative in European Union 1.5 Service Numbers 1.5.1 If the system does not work reliably, or if errors occur in terms of control functions which are described in the manual, the owner should contact the nearest EcoRay Service Center. 1.5.2 The EcoRay Service Center can be reached at the following Phone and Fax numbers: 1. Service for Europe 2. Service for Outside of Europe, Worldwide Tel) +82 70 7510 3402

(Rev. 1.0)

Fax) +82 70 8630 3420 E-mail: [email protected]

11

Service Manual

EcoRay

PX-300HF

1.6 Product Pre-installation Data 1.6.1 Generator 1. Physical characteristics Item

Dimension (mm). Approx. deviation ±10%

Weight (Kg)

Length

Width

Height

Approx.

1300

640

1980

300

590

600

440

105

Mobile Cart (Including PSU, Generator, Tube Stand) Generator (Including H/V Transformer) (Installed inside of Mobile Cart)

2. Dimension

Note! Console and X-ray generator, PSU was already installed inside of Mobile Cart!

1.6.2 Tube Stand (SMS-TS-3) 1. Physical characteristics Configurations Tube Stand

(Rev. 1.0)

width

length

height

Weight(Kg) Deviation less than ±5 Kg

180

150

1,690

Approx .120

Size (mm) deviation ±10%

12

Service Manual

EcoRay

PX-300HF

2. Dimension

1.7 X-ray room Requirements 1.7.1 Transportation and Storage Requirements 1. An ambient temperature: +10°C ~ +60°C 2. A relative humidity: 10% ~ 90% (Without dew condensation) 3. An atmospheric pressure: 700hPa ~ 1,060hPa 1.7.2 Electrical requirements 1. An ambient temperature: +10°C ~ +35°C 2. A relative humidity: 30% ~ 65% (Without dew condensation) 3. An atmospheric pressure: 700hPa ~ 1,060hPa 4. This system must be used in the without corrosion and explosion gas.

Troubles are could be occurred in the use or storage under unsuitable environment condition.

(Rev. 1.0)

13

Service Manual

EcoRay

PX-300HF

1.7.3 Electrical requirements 1. This generator contains advanced circuitry which will maintain the selected X-ray values during irregular input voltage condition. If input voltage condition is over or less, then could not get a good X-ray radiation and have a good electric facility to be supplied with good regulation condition. 2. Supply specific electric power to the system. (1) Do not under size distribution transformer. (2) Insure the suitable electric facility for the system. 3. Generator Power line requirements. (1) Power line voltage variation must not exceed ± 10%. (2) Maximum line regulation for the maximum kVA demand is 10%. Model

PX-300HF

Input Power

Max. 1 kVA

Input Voltage

Single phase, 230V

4. Recommended wire size (1) Correct sizing of the feeder wire is critical to proper generator operation. (2) Wire size is depend on the kW rating of the generator, line voltage and the distance from the generator and electric circuit breaker. (3) Minimum wire size from generator and electric circuit breaker.

(Rev. 1.0)

Generator

Line Voltage

1 kVA

230 VAC

Wire size 9m 2 mm2

14

Service Manual

EcoRay

PX-300HF

1.7.4 Wiring connection of main devices. Cable 4

X-ray tube

Tube Stand

Handle bar Cable 3

Cable 5

Main Power

Cable 1

Generator

H/V Transformer

(Mobile Cart)

Cable 2

Console

1.7.5 Main wiring connection (Refer to System Cable Wiring. P43) Cable No

Q’ty

Cable length

Function

Cable 1

1

9,000mm

Cable 2

1

Assembled

Signal line

Connect generator with console

Cable 3

1

4,000mm

Power, Tube Stator line

Connect generator with handle bar

Cable 4

1

650mm

Tube Stator

Including temperature relay signal

Cable 5

1 pair

2.,500mm

High Voltage Cable

Connect H/V transformer with X-ray tube

Single phase power (120VAC)

Remarks

Connect line with indicated.

NOTE ☞ Use recommended ground wire size by local regulation. And ground terminal is located in beside of generator main fuse.

(Rev. 1.0)

15

Service Manual

EcoRay

PX-300HF

1.7.6 Fuse capacity of main parts. No

Fuse Name

Position

Capacity

Quantity

1

F1,F2

Main Fuse

500V, 63A

2

2

F3,F4

Trans Fuse

250V, 15A

2

3

F701

ECO-XR5-MiO

250V, 10A

1

4

F901

ECO-XR5-MiO

250V, 3.15A

1

5

F801

ECO-XR5-CnD

250V, 10A

1

6

F501

ECO-XR5-FnD

250V, 3.15A

1

7

F701

ECO-XR5-GC

250V, 20A

1

8

F700,F704,F705,F708

ECO-XR5-GC

250V, 5A

4

9

F702,F703,F706

ECO-XR5-GC

250V, 3.15A

3

10

F707

ECO-XR5-GC

250V, 10A

1

1.7.7 Interconnection and grounding requirements. 1. Main line of the generator must be connecting with main electric circuit breaker. (1) 415VAC 30A circuit breaker must be used. 2. Connected external device with generator must have electric circuit breaker or fuse. 3. Ground terminal of the generator must be grounded with ground line of the main power line. 4. Use recommended ground wire size by local regulation.

(Rev. 1.0)

16

Service Manual

EcoRay

PX-300HF

Chapter 2

Installation High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0)

17

Service Manual

EcoRay

PX-300HF

2. Installation 2.1 An Overview 2.1.1 This installation depends on generator’s type and configuration and must be installed by instruction of this manual. 2.2 Pre-installation checks points and Tools and Testers. 2.2.1 Pre-installation checks points 1. Before installation, engineer must check and notice as follows. (1) Check input voltage, temperature, humidity of installation place. (2) Read and understand installation manual before installation. (3) Install X-ray system with patient safety first. 2.2.2 Tools and Testers. 1. Described installation tools and test equipments are necessary and could be changeable installation method by optional items and then engineer should prepare tools and tester in needed. (1) Standard and generally tools and tester set used by engineer. (Multi meter, Screwdriver, Nipper, Long nose, Wrench set, Solder wire, Soldering iron, Hex Wrench) (2) Electric Drill and assorted bits set. (3) Grinding machine for cutting steel. (4) Cleaning things for parts. (5) kV meter (6) mA meter (7) Radiation Dose meter (8) Operation & Installation Manual

Only qualified (must trained by EcoRay) engineer could install this X-ray system!

(Rev. 1.0)

18

Service Manual

EcoRay

PX-300HF

2.3 Pre-installation Check

2.3.1 Prior to beginning installation, it is recommended to inspect the site and verify that the X-Ray room complies with Pre-installation requirements, such as: 1. Incoming line of X-ray room (Voltage etc.) 2. Main circuit breaker 2.3.2 Precaution when connecting electric power. 1. First, check the input voltage with tester and then connect the power cable with circuit breaker. 2. The input power must be single phase 120V AC and could check voltage with tester.

Power Cable Connected with 230V AC

Power cable must be connected with 230V AC wall outlet! If connected with low or high voltage wall outlet, it cause of troubles!

(Rev. 1.0)

19

Service Manual

EcoRay

PX-300HF

2.4 Installation of Generator (Actually, X-ray Generator was installed inside of Mobile Cart!) 2.4.1 Appearance 1. Console

2. Generator

3. High Voltage Transformer

2.4.2 Consist of Main Parts and Functions 1. Consist of Main Parts (1) Console Membrane Switches

External terminal

Console PCB

Note! X-ray Generator, High Voltage Transformer, PSU, Console was already installed inside of Mobile Cart!

(Rev. 1.0)

20

Service Manual

EcoRay

PX-300HF

(2) Generator (Assembled inside of Mobile Cart)

ECO-XR5-FnD Board

Communication Terminal

Socket Receptacle High Voltage Transformer

IGBT Block ECO-XR5-Mio Board

ECO-XR5-CnD Board

2. Function of Main Parts (1) Console ● Membrane Switches: Select and control functions. ● Console PCB: Control each function after receive signal from Membrane switches. ● External terminal: Connect Console PCB with Generator Main CPU. (2) Generator ● Communication Cable: Signal cable which connects Console PCB with Generator Main CPU. ● ECO-XR5-FnD Board: Control PCB for driver circuit of IGBT and Tube Filament. ● ECO-XR5-CnD Board: Control PCB for charging and discharging condition of Condenser. ● ECO-XR5- Mio Board: Main PCB assembled Microprocessor. ● Emergency Stop switch: Power off the electric power quickly in emergency situation. ● IGBT Block: Semiconductor switch unit for high voltage to control input and output power to be supplied High Voltage Transformer. ● High Voltage Transformer: It is a kind of transformer that supply high voltage to the X-ray tube after received voltage from generator and boost voltage with maximum 150,000V. ● Socket Receptacle: Connecting socket for High Voltage Cable

(Rev. 1.0)

21

Service Manual

EcoRay

PX-300HF

2.4.3. Unpacking 1. X-ray generator was assembled with mobile cart to be installed easily and easy move. 2. Check X-ray generator external condition (damaged or not) after unpacking. 3. If generator was damaged, stop installation and notify this fact to carrier and distributor immediately. 4. After unpacking, have to check all parts are exactly same as like order and do not dump any of packing parts such as envelope, box, documents until confirmation is completely.

2.4.4 Start Installation 1. Installation procedure of X-ray generator (1) Check serial no on the attached label with enveloped certificate’s serial no. (2) Remove the generator cabinet. (Remove bolts beside of generator cover) (3) Check external condition and internal cable wiring condition. If damage parts are found, order that parts immediately to distributor to prevent delayed installation. (4) If released connector is found, should connect it in the installation site. (5) Connect cables by wiring diagram. (Refer to Wiring diagram) (6) Check cable wiring condition again after wiring. (7) Connect main electric power after checking wiring. (8) Must check input voltage with tester before connect it with generator. (9) Power ON the circuit breaker after checking input voltage. (10) Check input voltage of each part. (11) Check functions after Power ON of Console. (12) Engineer never leave installation site until installation is completely finished. (Rev. 1.0)

22

Service Manual

EcoRay

PX-300HF

X-ray Generator was assembled inside of Mobile cart for easy installation!

Never use re-connected communication cable! If cut communication cable, it cause of trouble! 2 Generator Terminal for External Devices (1) Generator Terminal for External Devices

High Voltage Sockets

Connect Console

Connect Tube Stand

Generator (2) Consist of External Devices

Console

Picture 1

Tube Stand

NOTE ☞ X-ray Generator, Console, Tube Stand was already assembled with Mobile Cart!

(Rev. 1.0)

23

Service Manual

EcoRay

PX-300HF

(3) Images after installation

3. Connect High Voltage Transformer with X-ray Tube (1) Wiring High Voltage Cable (Connect High Voltage Transformer and X-ray tube) Socket Receptacle

Connected Image

X-ray Tube

(2) Connect High Voltage Cable Cable Plug

Insulation silicon oil for Receptacle Socket

reinforcement

The terminal pins of the high voltage cable are extremely delicate and easily damaged. And then, handle them with very carefully.

(Rev. 1.0)

24

Service Manual

EcoRay

PX-300HF

(3) Manufacturer’s Instructions 1) Prepare receptacle socket of high voltage transformer that will be installed and high voltage cable. (High voltage cable must be handling with always clean condition) 2) Apply enough insulation silicon to the entire of surface of Cable plug (High voltage transformer) and near of the pin. (Put approximately 1cm of insulation silicon into the High Voltage Transformer receptacle socket.) 3) Fix cable nut with firmly after connecting of high voltage cable with high voltage transformer. 4) Apply enough insulation silicon to the entire of surface of Cable plug (X-ray tube) and near of the pin. 5) Connect opposite side of high voltage cable to the X-ray tube receptacle socket. 6) Fix cable nut with firmly after connecting of high voltage cable with X-ray tube.

High voltage cable wiring must be done in dryness place without humidity. 4. Connect Tube Stator (1) Wiring of X-ray tube Anode Stator

External terminal of Generator (ECO-XR5-GCboard)

Back side of Handle

Anode side of X-ray Tube

NOTE ☞ When connecting of Tube Stator, never use “Temperature Relay” point. SC: Stator Common (Terminal 1) SM: Main Winding of the Stator (Terminal 2) SS: Shift of the Stator (Terminal 3) NC: Not use (Terminal 4) ET: Connect Earth (Terminal E) Tube 1 Temp: Temp. Relay (Terminal 5) Tube 1 Temp: Temp. Relay (Terminal 6)

(Rev. 1.0)

25

Service Manual

EcoRay

PX-300HF

(2) Specification of X-Ray Tube Anode Stator

X-ray tube Anode Stator wiring is must correctly with recommended by X-ray tube manufacturer. Never connect terminal No.1 to No. 5 and No. 6.

The terminal pins of the high voltage cable are extremely delicate and easily damaged. And then, handle them with very carefully. 5. Power Cable Connection (1) The power cable cord was assembled with X-ray generator and located bottom side of Mobile cart. (2) The power cable of this generator must be connected correctly with wall outlet (120V AC). (3) Before power cable connecting, please check input power voltage. (4) Connect power cord when input voltage for generator is correct, and push the Power On switch if want to run X-ray generator. (5) Main power will be supplied to the generator if power cord is connected with wall outlet. Remove the power cord from wall outlet, if break electric power completely.

After wiring, adjust voltage level sensor circuit of CnD Board. How to adjust: Refer to Charge & Discharge circuit error (Check ECO-XR5-CnD Board).

(Rev. 1.0)

26

Service Manual

EcoRay

PX-300HF

6. Specification Item

Functions

PX-300HF

Generator Type

High Frequency Inverter

KVp range

40 ~ 125kVp ±5%

mA range

50 ~ 300mA ±5%

Exposure Time Generator

High Voltage Transformer

1mSec ~ 6Sec (Below 100mSec: inside ±1mSec. Over 100mSec: inside ±10%)

APR

288 (Standard)

AEC

Optional item

Output Frequency

40kHz

Maximum Power

32kW

Line Voltage

Single phase, 230V ±10% , 50/60Hz

Rating

125kV / 200mA,

100kV / 300mA at 100mSec

Maximum kVp

125kVp

Maximum mA

300mA

2.5 Installation of Tube Stand 2.5.1 Appearance of Tube Stand (SMS-TS-3)

Front view

(Rev. 1.0)

Side view

27

Service Manual

EcoRay

PX-300HF

2.5.1 Consist of Main parts 7. Electromagnetic lockup device for vertical travel

6. Electromagnetic lockup device for tube rotation (Tube Arm)

1. Stand

5. X-ray tube 2. Vertical move device 4. Handle Bar

3. Stand base

3. Functions of main components (1) Stand: Device helps the X-ray tube travelling up & down (2) Vertical move device: holding X-ray tube and to be connected to Tube Stand to support up & down travel. (3) Stand base: holding Tube Stand and to be connected to Floor rail for longitudinal travel. (4) Handle Bar: The device has a Angle display panel, emergency PWR-off button, electromagnetic lockup buttons and user holding handles for X-ray tube position control. (5) X-ray tube device: where it has X-ray tube, Handle Bar and collimator assembled. (6) Tube Arm: Helping X-ray tube rotation from Handle Bar. (7) Electromagnetic lockup device for vertical travel: Helping X-ray tube up/down travel from Handle Bar. 2.5.2 Unpacking 1. For easy installation, Mobile X-ray system almost assembled. And after unpacking, engineer just connects X-ray tube, Handle, Collimator. 2. Check if all the devices are well condition after unpacked. 3. In case any of the devices are broken or missed, please contact your local agent or manufacturer quickly. 4. They are packed as a set of Stand, a set of Handle Bar, a set of balance weight. 5. Do not scrap the packing boxes, envelops or inside papers before you make sure all about the contents. 6. Sometime, Tube Stand separated with Mobile cart by buyer’s asking.

(Rev. 1.0)

28

Service Manual

EcoRay

PX-300HF

It is necessary to have 2 or more people to move and install the Tube Stand devices. Box 1 Tube Stand

Front view

Side view

2.5.3 Installation procedure of Tube Stand (Actually Tube Stand already installed with Mobile Cart!) 1. Check if the serial No. on the label is match with the serial No. on the quality certificate. 2. During the unpacking, check if all the devices are in well condition, especially at cables. In case any broken or damaged parts, please contact your local agent for replacements. 3. Bring the devices to the installation place. (Do not assemble the devices in advanced) 2 or more people are needed to move them because the devices are very heavy. 4. Fix (install) the Floor rail on a right position. 5. Connect Tube Stand into the Stand base. 6. Press down the tube head (tube arm) to the level where you can put the holding pin. It is necessary to pull it down without weight balance. 7. Put the holding pins around the balance weight box position. 8. Then install X-ray tube, Handle Bar and collimator. 9. And put balance weights into the tube stand (in the empty space of balance weights). The balance weights are very heavy, so please be careful when you move them into the space. 10. Check the cable connections and make them tight if necessary. 11. Make all the cable connections according to the cable diagram. 12. Check if all the connections are correct. 13. If the cable & wire connections are O.K, put main AC input power to the Generator. (Check AC input power level before you make Power ON) 14. One person holds the X-ray tube and the other removes the holding pin at behind of Tube Stand. 15. Check if the weight balance is O.K, if the balance is not perfect, add or remove the weight

(Rev. 1.0)

29

Service Manual

EcoRay

PX-300HF

balances. 16. PWR ON and check if electromagnetic lockup functions are all O.K – test it from Handle Bar. 17. The engineers should stay in the room until the installation works are all over.

Holding pin position 1. Set up the stand on Mobile base. 2. Fix the stand with bolts. 3. Press the tube arm down to the position where you can put the holding pin.

Balance weights Holding pin position 4. Check the holding pin position and put it in. 5. Insert balance weights into the stand. 6. Holding the Arm well and remove the holding pin then check if the weight balance is O.K

(Rev. 1.0)

30

Service Manual

EcoRay

PX-300HF

2.5.4 Starting of installation. 1. Understanding Wiring (1) Understanding of ECO-XR5-GC Board. (Upper side of Generator)

PIN NO

Tube Support part of ECO-XR5-GC Board (Upper side of Generator)

1

Stator Com (SC1)

2

Stator Shift (SS1)

3

Stator Main (SM1)

4

AC Out

5

AC 0V

6

NC

7

DC 0V

8

DC24V

9

EMG

10

EMG

11

NC

12

Tube1 Temp.

13

Tube1 Temp.

14

Door

15

Door

16

GND

Tube Support Unit

(2) Understanding of SMS-TS-340 Board (Handle bar of Tube Stand)

PIN NO.

P6 Connector

(Rev. 1.0)

31

P6 Connector of SMS-TS-340 Board (Handle bar of Tube Stand)

1

Stator Com (SC1)

2

Stator Main (SM1)

3

Stator Shift (SS1)

4

AC 0V

5

AC 24V

6

EMG

7

EMG

8

Tube1 Temp.

9

Tube1 Temp.

10

GND

Service Manual

EcoRay

PX-300HF

(2) Connect ECO-XR5-GCof Tube Support Unit and P6 Connector of SMS-TS-340

Tube support unit

Tube Support unit of

P6 Connector

P6 Connector of

ECO-XR5-GCBoard

SMS-TS-340

(Upper side of Generator)

(Handle bar)

1

Stator Com (SC1)

1

Stator Com (SC1)

2

Stator Shift (SS1)

2

Stator Main(SM1)

3

Stator Main (SM1)

3

Stator Shift (SS1)

4

AC Out

4

AC 0V

5

AC 0V

5

AC 24V

6

NC

NC

NC

7

DC 0V

NC

NC

8

DC24V

NC

NC

9

EMG

6

EMG

10

EMG

7

EMG

11

NC

NC

12

Tube1 Temp.

8

Tube1 Temp.

13

Tube1 Temp.

9

Tube1 Temp.

14

Door

NC

NC

15

Door

NC

NC

16

GND

10

GND

(4) Connect P7 of SMS-TS-340 with X-ray Tube

SMS-TS-340 SMS-TS-340의

Handle bar

Anode side of X-ray tube

(Rev. 1.0)

32

P7 Connector

Anode side of

(Inside of Handle bar)

X-ray Tube

1

Stator Com (SC1)

1

Stator Com (SC1)

2

Stator Main(SM1)

2

Stator Main(SM1)

3

Stator Shift (SS1)

3

Stator Shift (SS1)

4

Tube1 Temp.

5

Tube1 Temp.

5

Tube1 Temp.

6

Tube1 Temp.

6

GND

E

GND

Service Manual

EcoRay

PX-300HF

(5) Connect P8 connector of SMS-TS-340 with Emergency Switch (There is no electric polarity) (6) Connect P4 connector of SMS-TS-340 with Collimator (There is no electric polarity) (7) Connect P9 connector of SMS-TS-340 with X-ray Tube rotation electromagnetic rock. (Take notice of electric polarity) (8) Connect P12 connector of SMS-TS-340 with X-ray Tube rotation electromagnetic rock. (Take notice of electric polarity) (9) Connect P5 connector of SMS-TS-340 with SID display unit (Optional) 2. Assemble of X-ray tube device (1) X-ray tube device consists of X-ray tube, tube arm (rotation), Handle Bar and collimator.

X-ray tube

Tube arm

Handle Bar

Collimator

X-ray tube device

3. Verification of cables. (1) Compare the cable configuration is matching to the circuit diagram or cable diagram. (2) Connect PWR input to Tube Stand (3) Main PWR ON (4) Test the electromagnetic locker functions by controlling Handle Bar.

(Rev. 1.0)

33

Service Manual

EcoRay

PX-300HF

4. Cabling between Tube Stand and HFG (1) Power line connection → The PWR cables are coming from HFG and it is necessary to measure input PWR level before the actual PWR engages to Tube Stand. → The PWR level is 24VDC and should be measure by a multi-meter to make it sure.

Tube stand (side view)

(Rev. 1.0)

SMS-GC-430 (Upper side of Generator

34

Service Manual

EcoRay

PX-300HF

2.6 System Cable Wiring

SMS-TS-340 Board X-ray Tube (Handle bar of Tube Support)

Electromagnetic lock for Tube rotation ECO-XR5-GCBoard (Upper side of Generator)

NOTE! If do not use Door and EMG functions, then never remove the attached pin of tube support connector pin (Door and EMG). If needs these functions, use these pin connectors

(Rev. 1.0)

35

Service Manual

EcoRay

PX-300HF

Chapter 3

Technical Information High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0)

36

Service Manual

EcoRay

PX-300HF

3. Technical Information 3.1 An Overview This technical information contains that maintenance history and calibration data. 3.2 Installation information After installation, enter the installation information as below form. Hospital Name Install Engineer

Install Date

3.3. Exposure condition data 3.3.1 Main Parts. Install Date

Hospital Name

Model

X-Ray

Generator

Tube

S/N. H/V Cable Length

6m

8m

Model S/N. 10m

12m

15m

3.3.2 Calibration Data Small Focal 50mA

75mA

Large Focal 100mA

150mA

40kV

40kV

60kV

60kV

80kV

80kV

100kV

100kV

125kV

125kV

(Rev. 1.0)

37

200mA

300mA

Service Manual

EcoRay

PX-300HF

3.4 Maintenance History

3.4.1 Enter the following information after each data modification in this book or Periodic Maintenance Service Date

(Rev. 1.0)

Engineer Name

Service Note

38

Service Manual

EcoRay

PX-300HF

Chapter 4

Calibration High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0)

39

Service Manual

EcoRay

PX-300HF

4. Calibration 4.1 An Overview This manual contains that adjustment method of exposure condition and how to calibrate exposure data. Calibration data will be memorized at memory chip at the ECO-XR5-Mio Board with digital value. 4.2 Introduction 4.2.1 This calibration tool is specified program supplied by EcoRay and it will be operated at the personal computer. 4.2.2 Control exposure value (Ex. kV, mA, etc) 4.2.3 This calibration tool must be operated by qualified engineer by EcoRay. 4.2.4 This calibration tool was designed by many experience data of manufacturer and this tool will be operated by specified program supplied manufacturer. 4.3 Needed tool and testers when Calibration Following tools and testers are needed in generally and another tools and testers must be prepared if needed. 4.3.1 Calibration program 4.3.2 Generally testers and tools. 4.3.3 Insulation silicon for high voltage cable reinforcement. 4.3.4 Digital voltage meter. 4.3.5 kVp meter 4.3.6 mA meter 4.3.7 Time meter 4.3.8 Radiation Dose meter 4.3.9 Al filter set for half-valuelayer test. 4.3.10 Phantoms set for Fluoroscopy test. 4.3.11 Another tester for calibration. 4.4 Check point before calibration 4.4.1 Input voltage 4.4.2 Ground condition 4.4.3 Prepare calibration tools and tester set. 4.4.4 X-ray tube model (technical data) and external equipment.

(Rev. 1.0)

40

Service Manual

EcoRay

PX-300HF

4.5 Consist of Calibration Program 4.5.1 Consist of Calibration Program

2. Calibration mode 1. Controller mode

5. Communication state display & Memory display window

3. X-ray Tube selection window

4. Bucky selection window

4.5.2 Main Functions. 1. Controller mode: Select radiographic, fluoroscopic and AEC function and display. 2. Calibration mode: After running of calibration, adjust calibration data and save it. 3. X-ray Tube selection window: Select radiographic x-ray tub and fluoroscopic x-ray tube and set tube stator working condition. 4. Bucky selection window: Select Bucky position, Bucky type and change data. 5. Communication state display & Memory display window: Communicate generator with PC and read or save setting data. 4.6 How to use Calibration program 4.6.1 Function of Icons 1. Execution Icon. If click this icon on the PC screen, program will be running automatically. 2. Power Icon. If click this icon, program will be activated. (Running: Green lamp, Stop: Yellow lamp condition) 3. Communication port: Only COM1 port supplied. 4. Select Radiographic, Fluoroscopy, AEC function 5. Display window for kV, mA, Time 6. Microprocessor information window: memory mA data, kV data, exposure count, product serial No., microprocessor program version.

7. Calibration mode ON/OFF switch.

8. Select calibration area and display input data.

(Rev. 1.0)

41

Service Manual

EcoRay

PX-300HF

9. Calibration data input window

10. Compensation mode of mA data

11. Used compensate mA data if mA data tolerance is too high. Click “Advanced..” icon if want to compensate mA data. 12.

After running, could see mA graphic and could adjust it.

13. Initialize the compensated mA data. 14. Save the compensated mA data. 15. Read the compensated mA data. 16. Read the compensated mA data from PRAM of Microprocessor. 17. Save the compensated mA data from PRAM of Microprocessor. 18. Save the compensated mA data to another computer. 19. Read the compensated mA data from another computer.

4.6.2 Start Calibration 1. Click Execution Icon on the PC screen. 2. Select Radiographic, Fluoroscopic, AEC functions on the controller mode. 3. Select X-ray tube position and Tube stator on the Tube selection window. 4. Select Bucky position and type on the Bucky selection window. 5. Click ON icon of Calibration mode and then program will be activated. 6. Adjust calibration data after running of calibration program. (Calibration program must be operated by only qualified engineer by manufacturer.) 4.7 Time of Calibration 4.7.1 In installation: After installation, this x-ray system must be inspected by radiation inspection office even though this x-ray system was calibrated at the factory before delivery. And must be calibrated when needs calibration. 4.7.2 In periodic inspection: This x-ray system must be inspected by radiation inspection office per every 3 years. And must be calibrated when needs calibration. 4.7.3 In repaired or replaced: If main parts are repaired or replaced, x-ray system must be inspected by radiation inspection office. And must be calibrated when needs calibration. 4.7.4. In re-installation or move: If re-install or install it another place then x-ray system must be inspected by radiation inspection office. And must be calibrated when needs calibration.

(Rev. 1.0)

42

Service Manual

EcoRay

PX-300HF

4.8 Must calibrated when Main Parts are repaired or replaced 4.8.1 Repaired or Replaced of X-ray Tube 4.8.2 Repaired or Replaced of ECO-XR5-MiO Board 4.8.3 Repaired or Replaced of ECO-XR5-CnD Board 4.8.4 Repaired or Replaced of ECO-XR5-FnD Board 4.8.5 Repaired or Replaced of High Voltage Transformer 4.8.6 Replaced of High Voltage Cable or replaced with another length of high voltage cable.

(Rev. 1.0)

43

Service Manual

EcoRay

PX-300HF

Chapter 5

Trouble Shooting High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0)

44

Service Manual

EcoRay

PX-300HF

5. Trouble Shooting 5.1 An Overview 5.1.1 This generator has a self-diagnosing function and display error code on the console display if system has an error. 5.2 Error type 5.2.1 Error type Error type

No

1

2

3

Charge

Circuit

problem

Cause 1. Charging to Main Capacity is not normal. 2. Input voltage is out of boundary. (Too high or too low) (Standard input voltage: Single phase, 220VAC ±10%)

Tube stator

1. Cable connection to Tube stator is wrong.

Circuit problem

2.

Tube Filament Circuit problem

ECO-XR5-MiO in Stator circuit is out of order.

1. Tube Filament circuit connection is wrong. 2. Filament circuit is out of order. 3. Tube Filament is short. HV-DRV-OLP (ECO-XR5-FnD) signal or circuit is wrong.

the error.

(See circuit diagram) Generator

4

Protection

Circuit

problem

IGBT-OCP (ECO-XR5-FnD) signal or

Over current to IGBT Module may occur the

circuit is wrong. (See circuit diagram)

error

DRV-PWR-UVP (ECO-XR5-FnD) signal or circuit is wrong. FIL-OCP

(ECO-XR5-FnD)

Voltage from the IGBT Driver IC is lower than standard.

(See circuit diagram) signal

or

circuit is wrong. (See circuit diagram)

5

Over current to IGBT Driver IC may occur

Over current on the Tube Filament

Communication

1. Wrong connection of communication cable.

program

2. Communication circuit in the Main board or Console board is out of order. 1. Wrong connection of communication cable.

6

AEC problem

2. Wrong cable connection between AEC Board and Ion Chamber. 3. AEC Board or Ion Chamber is out of order.

7

Irregular Exposure

1. Actual exposure time is shorter than set time.

Time problem

2. Exposure switch is bad condition.

(Rev. 1.0)

45

Service Manual

EcoRay

PX-300HF

5.2.2 Error type and display Error code will be displayed on the console when occurred error.

Error displayed as like image.

1. Error type 1: Displayed E r r C r F P at kVp, mA, Time section Charge Circuit error

Tube

Refer to

Filament

Stator

Circuit

Circuit

error

error

HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Commu

AEC

Filament

nication

Circuit

time

- OLP

error

error

error

Normal

Normal

Normal

To be checked

To be checked

(Any one of 4 is possibly faulty)

Exposure

2. Error type 2: Displayed E r r C r F at kVp, mA, Time section

Charge Circuit error

Refer to

Tube

Circuit

Circuit

error

error

HV-DRV

Circuit error

Tube Circuit error

To be checked

AEC

DRV-

Filament

nication

Circuit

OCP

PER

- OLP

error

error

Normal

Normal

Normal

3. Error type 3: Displayed E r r C r

Stator

Commu IGBT-

Filament

Stator

To be checked

Charge

ECO-XR5-CnD Circuit

Circuit error Normal

HV-DRV

Normal

P at kVp, mA, Time section

Refer to

Filament

Exposure time error

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

To be checked (Any one of 4 is possibly faulty)

Exposure time error

Normal

4. Error type 4: Displayed E r r C r at kVp, mA, Time section Charge Circuit error

Tube Stator Circuit error

To be checked

(Rev. 1.0)

Filament Circuit error Normal

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Normal

46

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Exposure time error

Normal

Service Manual

EcoRay

PX-300HF

5. Error type 5: Displayed E r r C FP at kVp, mA, Time section Charge Circuit error To be checked

Tube Stator Circuit error Normal

Filament Circuit error

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

To be

To be checked

checked

(Any one of 4 is possibly faulty)

Exposure time error

Normal

6. Error type 6: Displayed E r r C F at kVp, mA, Time section Charge Circuit error To be checked

Tube Stator Circuit error Normal

Filament Circuit error

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

To be

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

checked

Exposure time error

Normal

7. Error type 7: Displayed E r r C P at kVp, mA, Time section Charge Circuit error To be checked

Tube Stator Circuit error Normal

Filament Circuit error Normal

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

To be checked (Any one of 4 is possibly faulty)

Exposure time error

Normal

8. Error type 8: Displayed E r r CHA at kVp, mA, Time section Charge Circuit error To be checked

Tube Stator Circuit error Normal

Filament Circuit error

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

Normal

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

Exposure time error

Normal

9. Error type 9: Displayed E r r r FP at kVp, mA, Time section Charge Circuit error Normal

(Rev. 1.0)

Tube Stator Circuit error

Filament Circuit error

To be checked

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

To be checked (Any one of 4 is possibly faulty)

47

Exposure time error

Normal

Service Manual

EcoRay

PX-300HF

10. Error type 10: Displayed E r r r F at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error

Filament Circuit error

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

To be checked

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

Exposure time error

Normal

11. Error type 11: Displayed E r r r P at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error To be checked

Refer

Filament Circuit error Normal

HV-DRV

Circuit error Normal

Tube Stator Circuit error To be checked

Filament Circuit error

IGBT-OCP

DRV-PER

To be checked (Any one of 4 is possibly faulty)

12. Error type 12: Displayed E r r Charge

ECO-XR5-CnD Circuit

Normal

roto at kVp, mA, Time section

Refer to HV-DRV

Exposure time error

ECO-XR5-CnD Circuit

IGBT-OCP

Normal

DRV-PER

Normal

Exposure time error

Normal

13. Error type 13: Displayed E r r FP at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error Normal

Filament Circuit error

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

To be

To be checked

checked

(Any one of 4 is possibly faulty)

Exposure time error

Normal

14. Error type 14: Displayed E r r FILA at kVp, mA, Time section Charge Circuit error Normal

(Rev. 1.0)

Tube Stator Circuit error Normal

Filament Circuit error To be checked

Refer to HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Normal

48

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Exposure time error

Normal

Service Manual

EcoRay

PX-300HF

15. Error type 15: Displayed E r r 232 at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error Normal

Refer to

Filament Circuit error

HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

Normal

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

To be

Normal

checked

Normal

Exposure time error

Normal

16. Error type 16: Displayed E r r AEC at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error Normal

Refer to

Filament Circuit error

HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

Normal

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

To be checked

Exposure time error

Normal

17. Error type 17: Displayed E r r SEC at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error Normal

Refer to

Filament Circuit error

HV-DRV

ECO-XR5-CnD Circuit

IGBT-OCP

Normal

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

Exposure time error To be checked

18. Error type 18: Displayed E r r Pro----O at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error Normal

Refer to

Filament Circuit error Normal

HV-DRV To be checked

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

Normal

Normal

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

Exposure time error

Normal

19. Error type 19: Displayed E r r Pro--I-- at kVp, mA, Time section Charge Circuit error Normal

(Rev. 1.0)

Tube Stator Circuit error Normal

Filament Circuit error Normal

Refer to HV-DRV

Normal

ECO-XR5-CnD Circuit

IGBT-OCP

DRV-PER

To be

Normal

checked

49

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

Exposure time error

Normal

Service Manual

EcoRay

PX-300HF

20. Error type 20: Displayed E r r Pro---P- at kVp, mA, Time section Charge Circuit error Normal

Tube Stator Circuit error Normal

Filament Circuit error Normal

Refer to

ECO-XR5-CnD Circuit

HV-DRV

IGBT-OCP

Normal

Normal

DRV-PER

Commu

AEC

Filament

nication

Circuit

- OLP

error

error

Normal

Normal

Normal

Normal

Commu nication error

AEC Circuit error

Exposure time error

Normal

Normal

Normal

To be checked

Exposure time error

21. Error type 21: Displayed E r r Pro----F at kVp, mA, Time section Charge Circuit error

Tube Stator Circuit error

Filament Circuit error

Normal

Normal

Normal

Refer to

ECO-XR5-CnD Circuit

HV-DRV

IGBT-OCP

DRV-PER

Filament - OLP

Normal

Normal

Normal

To be checked

22. Error type 22: Displayed E r r

Inte at kVp, mA, Time section

1. Inter Lock signal trouble and circuit trouble 2. Inter Lock is Door Open, Tube Temperature, Bucky start signal, Emergency Stop switch.

(Rev. 1.0)

50

Service Manual

EcoRay

PX-300HF

5.3 Trouble Shooting 5.3.1 Charge & Discharge circuit error (Check ECO-XR5-CnD Board) 1. Consist of part

(Rev. 1.0)

51

Service Manual

EcoRay

PX-300HF

2. Check Procedures 1 (1) Board Name: ECO-XR5-CnD (2) Functions: This board controls charge/discharge of Main Condenser. (3) Followings are the check procedures for ECO-XR5-CnD board when charge/discharge trouble. Under normal condition Provide main input power to generator and check if LED (804) is ON. *Before the power ON at the console.

Under defective

How to maintain 1. Check Connector (P803) and

No light at LED

measure the voltage (Normal: 220VAC)

(D804)

2. Check the Fuse F801. 3. Measure the voltage TP801 and U801 Heat sink (Gnd) (Normal: 21VDC ±0.5V)

Console Power ON. Standby lamp flickering then Standby lamp is flickering of Console.

turns yellow and error code displayed on the FND

Standby

lamp

turned

1. Check LED (D816) lightening. (Normal: Red light) 2. Check the Connector (P805) and measure the voltage on it. (Normal: 310VDC between pin 1 & 4) 3. Check the Connector (P804) and measure the voltage on it.

to

(Normal: 220VAC between pin 1 & 2)

green and LCD display goes

4. Check the Connector (P802).

enabled.

5. Measure the voltage between Test point TP803 & TP804 (Normal: +5VDC ±0.2V) In case of Relay (RL801, 802 do

Operating is ready.

the frequency at Test point (TP802,

how to adjust the frequency.

(Rev. 1.0)

point TP803 & TP805 (Normal: +12VDC ± 0.2V)

not work correctly, check

Chapter 4.1.3.1 is describing

6. Measure the voltage between Test

803) and make

7. Check the function of Relay (RL801) Normal: it works when OP console PWR is ON and stops when main capacitor charges about 90% 8. Check the function of Relay (RL802)

sure if it is 75kHz

Normal: RL802 starts to work

or not.

immediately when RL801 stops.

52

Service Manual

EcoRay

PX-300HF

(4) Board Name: ECO-XR5-CnD (5) Functions: This board controls charge/discharge of Main Condenser. (5) Adjust frequency between Test point TP803 and TP802. Under normal condition

Under defective

How to maintain In order to readjust input power freq....

OP console Power On.

1. Short circuit of P801 pin1 & pin2. 2. Tester probe connection to TP803 (Gnd) and TP802.

Study the function of P801 at ECO-XR5-CnD Board.

Freq. readjustment

3. Read the detected value. 4. Adjust VR801 to meet 75KHz (deviation allowed: 75KHz ±0.5KHz)

P801 function of Connector P801 pin 1: Voltage input freq. selective terminal (L. volt) P801 pin 2: Common ground P801 pin 3: freq. Selective terminal for charge (C. volt)

5. with PSU: 76-78KHz (deviation:±1KHz )

Charging freq. check

In order to adjust charging power freq..

Measuring input power freq.

1. Short circuit of P801 pin2 & pin3.

1. Connect pin 1 & 2 of

2. Tester probe connection to TP803

P801. 2. Tester probe connection to TP803 (GND) and TP802.

(Gnd) and TP802. Freq. readjustment

3. Read the detected value. 4. Adjust VR801 to meet 76KHz

3. Read the detected value.

(deviation allowed: 76KHz ±0.5KHz)

4. Normal value: 75KHz

5. with PSU: 76-78KHz

(with PSU: 76-78KHz)

Measuring charging power freq. 1. Connect pin 2 & 3 of P801. 2. Tester probe connection to TP803 (Gnd) and TP802 3. Read the detected value. 4. Normal value: 76KHz (with PSU: 76-78KHz)

Measure the charging power frequency.

(deviation:±1KHz)

Check charge frequency after adjustment of input power frequency.

1. Check input power and charging power freq after adjustment of charging frequency. 2. System turn off and on again, and measure the frequency. 3. Open P801 pin 1, 2, 3 if input power frequency is normal.

(Rev. 1.0)

53

Service Manual

EcoRay

PX-300HF

5.3.2 Tube Stator Circuit error (Check ECO-XR5-Mio Board) 1. Consist of part

2. Consist of Tube Stator Temperature Relay Tube Anode Socket Receptacle

Anode Stator wiring terminal

Anode Stator Coil terminal

M: Main wiring of X-ray tube stator. A: Assistance wiring of X-ray tube stator. NC: Not use ET: Earth

Note!

(Rev. 1.0)

Never connect terminal 1 to terminal 5 or 6.

54

Service Manual

EcoRay

PX-300HF

3. Check procedure (1) Board Name: ECO-XR5-Mio Board (2) Functions: Controlling generator including Tube Stator movement. (3) Under the tube stator trouble, check tube stator wiring condition and then check ECO-XR5Mio Board and check points are as follows; Under normal condition

Under defective

How to maintain 1. Check the cable connection between

Provide main power to generator

HFG and X-ray tube (Tube Stator lines)

and Power ON at console.

(7-5. Tube Stator circuit error sub 2, 3) 2. Check the Connector (P901) and voltage on it -- 220VAC is normal 3. Check the Fuse (F901, F701). 4. Measure the voltage output between

Press ‘Ready’ button on OP

Test point (TP109) and Test point

console.

(TP105); 3VDC ±0.2V is normal. 5. If above voltage is not normal, adjust VR701 to fix it at 3VDC. 6. Measure the voltage between Test

Ready lamp lightening of console.

No Ready lamp light on of console

point (TP109) and Test point (TP106) ; 3VDC ±0.2V is normal. 7. In case the voltage output is not good, adjust VR702 to fix it at 3VDC. 8. Check the output voltage from Diode (D701, 702) -- 7~8 VDC is normal. 9. In case the output voltage is not O.K,

Functioning well.

check the coil winding status on transformer T701, 702 and measure the

Contact service dept.

If you still have

coil value. (see the circuit diagram for

the same problem

check points)

(No LED light ON

Normal: Input pin 1 & 3 = 1 output pin 4 & 5 = 1.5

with Ready)

(Rev. 1.0)

55

Service Manual

EcoRay

PX-300HF

5.3.3 Tube Filament Circuit error (Check ECO-XR5-FnD Board) 1. Consist of part

2. High Voltage Cable

Measuring method and Procedure - Must power off before checking X-ray tube filament short condition. - Remove Cathode ( - ) from the high voltage transformer. - Check resistance value C with S or C with L of cable plug. - If short condition, resistance value could not check.

(Rev. 1.0)

56

Service Manual

EcoRay

PX-300HF

3. Check procedure (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling Tube Filament (3) Under Tube Filament circuit trouble, check high voltage and wiring condition of circuit and then check ECO-XR5-FnD board. Under normal condition

Under defective

1. Check Connector (P501) and measure

Turn on the Main power and

the voltage on it ; 220VAC is normal

then turn on the Console Power

2. Check Fuse (F501). LED (D516, 517, 518)

ECO-XR5-FnD

(Driver

Board)

LED (D516, 517, 518)

How to maintain

Not lighting Normal: All LED lighted with green

3. Fix the Relay (RL501) position and check its functioning. Normal: Relay working and Ready lamp ON 4. Measure the voltage between Heat sink (U406) and Test point (M5V) Nomal: 5VDC ±0.1V 5. Measure the voltage between Heat sink

Set exposure values

(U406) and KA324 (U518 pin 4) ; 12VDC ±0.1V is normal 6. Measure the voltage between Heat sink

Push the “Ready” button

Error when Ready

(U406) and KA324 (U518 pin 11) ; -12VDC ±0.1V is normal 7. In case above voltages are not good, check input PWR of ECO-XR5-Mio (Main Board). 8. Fix the Connector (P403) well.

X-ray switch could use after Ready lamp lightening.

1. Check HV cables. (Anode, Cathode directions and contact condition) 2. Tube Focal LED lightening check. 3. Check cable connections between

Could use system after repair

Connector (P401) and HV Transformer. 4. Fix the RL401 Relay of focal change, and check its functioning

7. Voltage checks before Ready between Heat sink (U406) and

(Normal: Relay On -> Large, Relay off ->

Resistor (R426).

Small)

(Normal: approx.RMS 0.8VDC ±0.1V and voltage is difference at

5. Voltage check between Connector (P401)

each mA with Ready condition)

pin 3 and pin 4 before Ready at Small focus

8. Before pressing Ready, measure the voltage between Heat sink

(Normal: 35VDC ±2V)

(U406) and KA39 3(U519 pin3)

6. Voltage check between Connector (P401)

(Normal: 1VDC ±0.1V and this is different at each mA under Ready

pin 2 & 4 before Ready at Large focus.

condition)

(Normal: 35VDC ±2V)

(Rev. 1.0)

57

Service Manual

EcoRay

PX-300HF

5.3.4 IGBT Device signal and Protection circuit trouble (Check ECO-XR5-FnD Board) 1. Consist of part

(Rev. 1.0)

58

Service Manual

EcoRay

PX-300HF

2. Check procedure (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check high voltage cable, filament short condition of X-ray tube, wiring condition of circuit when IGBT Device signal and protection circuit trouble and then check ECO-XR5-FnD Board condition as bellows. Under normal condition

Under defective

How to maintain

1. Power OFF before the circuit check.

Set exposure value

2. Check resistance value between P502 pin 1 & 2, and check Diode voltage on Driver board. (Detection point is same and measure it at Power OFF status only) (Normal: resistance -> 1 ㏀ ± 50Ω,

Push the Ready button

Diode voltage -> 0.7VDC ± 0.1V) 3. Measure the resistance between P502 pin 3 & 4 and voltage Error comes

Push the X-ray button

when Xray exposure

at Diode in Power OFF condition. (Normal: Resistance -> 1㏀ ± 50Ω, Diode voltage -> 0.7VDC ± 0.1V) 4. Measure resistance value between P503 pin 1 & 2 and Diode voltage under Power OFF status. (Normal: Resistance -> 1㏀ ± 50Ω, Diode voltage -> 0.7VDC ± 0.1V) 5. Measure resistance between P503 pin 3 & 4 and Diode voltage at Power OFF status.

X-ray exposure is OK

(Rev. 1.0)

(Normal: Resistance Diode voltage -> 0.7VDC ± 0.1V)

59

Service Manual

EcoRay

PX-300HF

5.3.5 HV-DRV-OLP trouble or Error Pro O- - - error (ECO-XR5-FnD Board) 1. Consist of PCB

2. Check procedure about Board (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check wiring condition of ECO-XR5-FnD Board and check waveforms at the HV-DRV-OLP check point with oscilloscope when IGBT Device signal and protection circuit trouble and checked waveforms are as follows. Error code display

Normal waveform

Abnormal waveform

Displayed on console

(Error Pro o---)

(Rev. 1.0)

60

Service Manual

EcoRay

PX-300HF

5.3.6 IGBT_OCP trouble or Error Pro - I- - error (ECO-XR5-FnD Board) 1. Consist of PCB

2. Check procedure about Board (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check wiring condition of ECO-XR5-FnD Board and then check IGBT_OCP trouble when IGBT Device signal and protection circuit trouble. Check procedure is as follows. Normal Operation

Under defective

How to check and repair 1. Fix the Connector (P504)

Set exposure value

(connected to IGBT current sensor) 2. Detect waveform between U519 pin 7 and

Push Ready button

Error

comes

when

X-Ray

Heat sink (U406). 3. If the waveform is nor normal, adjust VR702 to make proper waveform.

exposure

4. This setting is always completed in factory, so do not adjust VR702 except educated engineers. Push X-ray button

5. In case IGBT_OCP circuit is out of order, please change IGBT or call Service center.

X-ray exposure is done

(Rev. 1.0)

61

Service Manual

EcoRay

PX-300HF

5.3.7 IGBT_OCP trouble or Error Pro - I- - error (ECO-XR5-FnD Board) 1. Consist of PCB

2. Check procedure (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check wiring condition of ECO-XR5-FnD Board and check IGBT_OCP trouble when IGBT Device signal and protection circuit troubles. And check procedures are as follows. Error code display

Normal waveform

Abnormal waveform

Displayed on console

(Error Pro -I--)

(Rev. 1.0)

62

Service Manual

EcoRay

PX-300HF

5.3.8 DRV _ PWR _ UVP trouble or Error Pro - - P – error (ECO-XR5-FnD Board) 1. Consist of PCB

2. Check procedure (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check wiring condition of ECO-XR5-FnD Board and the check IGBT_OCP trouble when IGBT Device signal and protection circuit trouble. And check procedures are as follows. Normal Operation

Under defective

How to check and repair 1. LED light ON check at D516, 517, 518

Set exposure value

(Normal: All LEDs are light ON) 2. Fix the Connector (P501) and check voltage (normal : 220VAC)

Error comes Push Ready button

3. Check the Fuse (F501).

when X-ray

4. Fix the Relay (RL501) and check its

exposure

functioning (normal: Relay working with lamp ON) 5. Measure the voltage between Test point Push X-ray button

(TP500, 501) (normal: 18VDC ±0.2V) 6. If above voltage is not O.K, adjust VR501 to meet the proper range at 18VDC. 7. In case there is no voltage out, check the Resistor (R502).

X-ray exposure is done

(Rev. 1.0)

63

Service Manual

EcoRay

PX-300HF

5.3.9 DRV _ PWR _ UVP trouble or Error Pro --P- error (ECO-XR5-FnD Board) 1. Consist of PCB.

2. Check procedure (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check wiring condition of ECO-XR5-FnD Board and then check IGBT _ OCP troubles when IGBT Device signal and protection circuit troubles. And check procedures are as follows. Normal Operation

Set exposure value

(Rev. 1.0)

Under defective

Error comes when X-ray exposure

64

How to check and repair

8. See next page about repair

Service Manual

EcoRay

Normal Operation

PX-300HF

Under defective

Set exposure value

How to check and repair

9. Check voltage between Heat sink (U504) and Diode Cathode (D504) (Normal: 18VDC ±0.3V) 10. Check voltage between Heat sink (U504) and Error comes

Push Ready button

Diode Anode (D505) (Normal: -13VDC ±0.3V)

when X-ray

11. Check voltage between Heat sink (U504) and

exposure

Diode Anode (D524) (Normal: -10VDC ±0.3V) 12. Check voltage between Heat sink (U504) and Resister (R526) (Normal: 15VDC ±0.3V) 13. Check voltage between Heat sink (U506) and

Push X-ray button

Diode Cathode (D506) (Normal: 18VDC ±0.3V) 14. Check voltage between Heat sink (U506) and Diode Anode (D507) (Normal: -13VDC ±0.3V) 15. Check voltage between Heat sink (U506) and

X-ray exposure is done

Diode Anode (D519) (Normal: -10VDC ±0.3V) 16. Check voltage between Heat sink (U505) and Resister (R521) (Normal: 15VDC ±0.3V) 17. Check voltage between Heat sink (U508) and Diode Cathode (D508) (Normal: 18VDC ±0.3V) 18. Check voltage between Heat sink (U508) and Diode Anode (D509) (Normal: -13VDC ±0.3V) 19. Check voltage between Heat sink (U508) and Diode Anode (D521) (Normal: -10VDC ±0.3V) 20. Check voltage between Heat sink (U508) and Resister (R523) (Normal: 15VDC ±0.3V)

(Rev. 1.0)

65

Service Manual

EcoRay

PX-300HF

5.3.10 FIL _ OCP trouble or Error Pro - - F - error (ECO-XR5-FnD Board) 1. Consist of PCB

2. Check procedure (1) Board Name: ECO-XR5-FnD Board (2) Functions: Controlling IGBT and Tube Filament (3) Check wiring condition of ECO-XR5-FnD Board and then check FIL _ OCP trouble when IGBT Device signals and protection circuit trouble. And check procedures are as follows.

Normal Operation

Under defective

How to check and repair

1. The function of FIL_OCP is to protect tube

Set exposure value

focal spots. 2. Check voltage between Test point (TP402) and Heat sink (U406)

Error comes Push Ready button

(Normal: 3.6VDC ±0.1V)

when X-ray

3. If the voltage is not O.K, adjust VR402 to

exposure

fix the correct voltage. 4. It is strongly recommended that VR402 is strictly prohibited from uneducated persons. (Keeping factory setting is very Push X-ray button

important) 5. In case FIL_OCP circuit is defective, X-ray tube or Driver board should be replaced. Call to Service center for the further actions.

X-ray exposure is done

(Rev. 1.0)

66

Service Manual

EcoRay

PX-300HF

5.3.11 Communication error between generator and console or Err

232 error

(ECO-XR5-Mio Boar & SMS-CONB-340 Board) 1. Consist of PCB

2. Check procedure (1) Board Name: ECO-XR5-Mio Boar and SMS-CONB-340 Board (2) Functions: Communication transfer between generator and console. (3) Check communication cable condition when communication problem between console and ECO-XR5-Mio Board. And check procedures are as follows. Normal Operation

Under defective

1. Communication cable connection check. 2. LED (D107, 108) light check. Normal: D107 flickering when parameters are being set (green light) Abnormal: D108 light ON (Red color) 3. LED (D906, 911, 912) check. Normal: D906 light on before Power On .D911, 912 light ON after Power ON. Abnormal: No LED light ON after Power ON. 4. Connector (P901) checks and measures the voltage. (Normal: 220VAC) 5. Check the Fuse (F901) 6. Check voltage between Test point (TP109) and Test point (TP110) (Normal: 7.5VDC ± 0.1V) 7. If the voltage (TP110) is not correct, adjust VR901 to fix it at 7.5VDC. 8. Check voltage between Test point (TP109) and Test point (TP111) (Normal: 15VDC ± 0.1V) 9. In case the voltage from TP111 is wrong, adjust VR902 to fix it at 15VDC. 10. Check RX signal between MAX232 (U101) pin 13 and Test point (TP109). Normal: see the next page (Oscilloscope waveform) 11. Check TX signal between MAX232 (U101) pin 14 and Test point (TP109). Normal: see the next page (Oscilloscope waveform)

Turn ON power of console

Self-diagnosing automatically by program

Err 232 appeared on the console

System is ready to work

Set exposure value

How to check and repair

Err 232 appeared on the console

X-ray exposure

(Rev. 1.0)

67

Service Manual

EcoRay

PX-300HF

5.3.12 Error AEC error (ECO-XR5-Mio Boar and SMS-CONB-340 Board) 1. Connect part

2. Check procedure (1) Check Ion Chamber connector condition. (2) Chamber position and Bucky position must be same. (Same position) (3) Check Number flickering or not on the Time display of console after X-ray exposure with ‘beep’ sound (4 times). (Normal: Changed normal display condition after Number flickering on the Time display)

(Rev. 1.0)

68

Service Manual

EcoRay

PX-300HF

3. Consist of part 1

4. Check procedure 1 (1) Board Name: SMS-CONB-340 Board (2) Functions: Controlling Ion Chamber (3) Check communication cable condition and then check circuit when communication troubles occurred between console and ECO-XR5-Mio Board and Ion Chamber. And check procedures are as follows. Normal Operation

Under defective

How to check and repair

Turn ON power of console

1. Fix the Communication cable between OP console and HFG 2. Fix the Connector (CCON7).

Self-diagnosing automatically by program

Err AEC comes

3. Fix the Connector (CCON6).

on the display

4. Fix the Connector (PCON1). 5. Check LED (PD1, 2, 3) Normal: LED (PD1, 2, 3) light ON.

System is ready to be used.

6. Check LED (MD2) Normal: MD2 is flickering with data communication. 7. Check the voltage between Connector (PCON1) pin 1 & 4 Normal: 7.5VDC ±0.2V

Select AEC function

8. Check the voltage of Capacitor (PC6)-both end Normal: 5VDC ±0.1V Set exposure value

9. Check the voltage of Capacitor ( PC17)-both end Normal: 3.3VDC ±0.1V 10. Check the voltage of Capacitor (PC19)-both end

X-ray exposure

(Rev. 1.0)

Normal: 1.9VDC ±0.1V

69

Service Manual

EcoRay

PX-300HF

5. Consist of part 2

6. Check procedure 2 (1) Board Name: SMS-AEC Board (2) Functions: Controlling Ion Chamber (3) Check communication condition when communication trouble occurred between input/output signal of Ion Chamber and circuit error. And check procedures are as follows. Normal Operation

Under defective

How to check and repair 1. Fix the communication cable between OP console and

Turn ON power of console

AEC board. Self-diagnosing automatically by program

2. Check the Connector (CCON1).

Err AEC comes

3. Check Connector (CCON3, 4, 5).

on the display

4. Check the console and Connector (CON1) cable connections. 5. Check LED (LED1, 2, 3)

System is ready to be used.

Normal: LED (LED1, 2, 3) light ON 6. Check LED (LED4) Normal: LED4 flickers during data communication 7. Measure the voltage of Capacitor (PC1) -both end

Select AEC function

Normal: 5VDC ±0.1V 8. Measure the voltage of Capacitor (PC2)-both end Normal: 1.9VDC ±0.1V

Set exposure value

9. Measure the voltage of Capacitor (PC4)-both end Normal: 3.3VDC ±0.1V 10. Measure the voltage of Capacitor (PC17)-both end

X-ray exposure

Normal: 3.3VDC ±0.1V

(Rev. 1.0)

70

Service Manual

EcoRay

PX-300HF

7. Consist of part 3

8. Check procedure 3 (1) Board Name: ECO-XR5-Mio Board (2) Functions: Controlling Ion Chamber Normal Operation

Under defective

How to check and repair

1. Check communication cable between OP console and

Turn ON power of console

Connector (P101). 2. Check cable connection between SMS-AEC Self-diagnosing automatically by program

(AEC Board) and Connector (P102).

Err AEC comes

3. Check LED (D107, 108)

on the display

Normal: D107 flickers during data setting (green) Abnormal: D108 light ON (Red) 4. Check LED (D906, 911, 912) Normal: D906 lights ON before Power On

System is ready to be used.

D911, 912 light ON after Power On Abnormal: No LED light ON after Power On 5. Check the Connector (P901) and voltage. (Normal: 220VAC)

Select AEC function

6. Check the Fuse (F901) 7. Measure the voltage Test point (TP109) and Test point (TP110) (Normal: 7.5VDC ±0.1V)

Set exposure value

8. In case the TP110 voltage is not normal, adjust VR901 to fix 7.5VDC. 9. Measure the voltage Test point (TP109) and Test point (TP111)

X-ray exposure

(Normal: 15VDC ±0.1V) 10. In case Test point (TP111) voltage is not normal, adjust VR902 to meet 15VDC.

(Rev. 1.0)

71

Service Manual

EcoRay

PX-300HF

9. Consist of working signal

10. Followings are standard sensitivity value by program. Sensitivity range -8 ~ +8

Value from Test point(DC Out) & Test point(CCON2 pin3)-GND Sensitivity switch in OP console (Density) is operated from -8 to +8, and the voltage between Test point(DC Out) & Test point (CCON2 pin3)-GND will move from -1.5V to -9.5V (deviation: ±0.1V)

5.3.13 Error SEC error (SMS-CONB-340 Board) 1. Error type: Caused by SMS-CONB-340 (Console Board) trouble, Membrane switch trouble, Hand switch trouble 1. Refer to clause 4 of 4.1.3.12 about SMS-CONB-340 Board checking. 2. Check hand switch working condition. 3. Check membrane switch condition. 4. The error comes when you release the exposure switch earlier than set exposure time, which it is not a failure.

(Rev. 1.0)

72

Service Manual

EcoRay

PX-300HF

5.4 Handle Bar (SMS-TS-340 Board) 5.4.1 Compositions Rectification diode D1

Fuse F1, F2

5.4.2 How to check and verify

P6 Connector (ECO-XR5-GCBoard in HFG)

(1) Board name: ECO-XR5-GCBoard (2) Function: Transmit and receive the signal & voltage of external devices to HFG. (3) Followings are check sequence when ECO-XR5-GCBoard is out of order

Normal Function

Faulty status

Handle Bar functioning at

Maintain Handle Bar

1. SMS-TS-340 Board F1, F2 fuse check

PCB

2. SMS-TS-340 Board D1 rectification diode

Tube stand

(SMS-TS-340)

Failure details and trouble shooting

check 3. Input power check on SMS-TS-340 Board P6 Connector 4. SMS-TS-340 Board의 P8 wire connection check (disconnected or not) 5. ECO-XR5-GCBoard function check

Electromagnetic functioning

Check HFG external interface board

1. SMS-TS-340 Board F1, F2, F3, F4, F5 Fuse check 2. SMS-TS-340 Board D1, D2 rectification

(ECO-XR5-

diode check

GCBoard)

3. SMS-TS-340 Board’s P6 Connector input power check 4. SMS-TS-340 Board의 P8 wire connection

Normal functioning

(Rev. 1.0)

Repair completed

73

check (if disconnected or not) 5. ECO-XR5-GCBoard function check

Service Manual

EcoRay

PX-300HF

Chapter 6

Circuit Diagram High Frequency Inverter Mobile X-ray System

Model: PX-300HF

(Rev. 1.0)

74

Service Manual

EcoRay

PX-300HF

6. Circuit Diagram 6.1 Block Diagram

(Rev. 1.0)

75

Service Manual

EcoRay

PX-300HF

6.2 Console Board

(Rev. 1.0)

76

Service Manual

EcoRay

(Rev. 1.0)

PX-300HF

77

Service Manual

EcoRay

(Rev. 1.0)

PX-300HF

78

Service Manual

EcoRay

(Rev. 1.0)

PX-300HF

79

Service Manual

EcoRay

PX-300HF

6.3 ECO-XR5-Mio Board

(Rev. 1.0)

80

Service Manual

EcoRay

(Rev. 1.0)

PX-300HF

81

Service Manual

EcoRay

PX-300HF

6.4 ECO-XR5-CnD Board

(Rev. 1.0)

82

Service Manual

EcoRay

PX-300HF

6.5 ECO-XR5-FnD Board

(Rev. 1.0)

83

Service Manual

EcoRay

PX-300HF

6.6 Handle Bar (ECO-XR5-HANDLE)

(Rev. 1.0)

84