Mammography X-ray System

MX - 600 Service Manual Document No.: RD0701001 Issued Date: 30-January-2001 Revision: 1.0 The original version of MX-600 is being written in English

RD0309001

Table of Contents

Section 1.

Specification Manual

Section 2.

Technical Document Manual

Section 3

Calibration Manual

Section 4

Maintenance Manual

Section 5

Schematic Manual

Series

Mammography X-ray System

1

Service Manual

Section 1. Specification Manual of the MX-600

Diagnostic Mammography X-ray System

RADIEQ CO., LTD.

MX-600 1

Mammography X-ray System

Series

1

Service Manual

1. System Power Requirements

………………...………………………………………..

3

2. Environment ………...………………………………………………...............................

3

3. X-ray Tube(Insert)

3

………...…………………………………………………………….

4. X-ray Tube(Housing)

……………...…………………………………………………….

4

5. Bucky Device Assembly ………...……………………………………………………….

4

6. X-ray Source Assembly ………...……………………….……………………………….

4

7. x-ray Generator

4

RADIEQ CO., LTD.

………...……………………….……………………………………...

MX-600 2

Mammography X-ray System

Series

1

Service Manual

1. SYSTEM POWER REQUIREMENTS Input power :

3500W (6500VA) max

Output power :

2500W max

Output Voltage :

22~39kV

Output Current :

10~85mA

Frequency :

20KHz

Line Voltage :

220~240 ± 10%

IRMS(rest) :

1A

IRMS(exposure):

25A

IPEAK:

55A

50/60 Hz

2. ENVIRONMENT Operational :

10˚C~40˚C , 30~75% rH , 70~106kPa

Transport & Storge

-40˚C~70˚C , 10~85% rH , 50~106kPa

3. X-RAY TUBE(INSERT) Manufacturer :

VARIAN

Model :

RAD 85S

Target Materials :

Molybdenum

Nominal X-ray Tube Voltage :

22~39kV

Inherent Filtration :

0.63mm Be (IEC 60522)

Nominal Focal Spot Value :

0.1/0.3 mm (IEC 60613)

Anode Heat Storage :

300kHU

Anode Heat Cooling :

60 kHU / min

Filament Current

Large Focus : 2.0~4.0 A Small Focus : 2.0~4.0 A

Filament Voltage

Large Focus : 6.0~8.5 V Small Focus : 1.0~1.5

Serial NO. : RADIEQ CO., LTD.

MX-600 3

Series

Mammography X-ray System

1

Service Manual

4. X-RAY TUBE ASSEMBLY(HOUSING) Assembler :

VARIAN

Type of X-ray Tube :

Rotating Anode

Model :

B110

Target Angle :

16°

Maximum Housing cooling rate :

135HU

Housing Heat Storage Capacity :

500kHU

Weight :

22.5lbs (10.2kg)

Serial NO. :

5. Bucky Device Assembly Manufacturer:

Radieq Medical Systems

Model :

RB-600

Electric Energy Input :

24VDC

Grid :

Ratio 4:1 , Line 36 L/cm

6. X-ray Source Assembly Assembler :

Radieq Medical Systems

Position of The Focal Spot :

600mm From Sticking Surface of The Head cover

Serial No.

7. X-ray Generator Generator Type :

High Frequency Inverter

Input Power :

Single Phase 220VAC , 60Hz

Radiographic Ratings :

Large Focal Point 22-39kV / 1-600mA Small Focal Point 22-35kV / 1-100mA

Power Rating :

RADIEQ CO., LTD.

3.5kW

MX-600 4

Series

Mammography X-ray System

2

Service Manual

Section 2. Technical Document of the MX-600

Diagnostic Mammography X-ray System

RADIEQ CO., LTD.

MX-600 1

Series

Mammography X-ray System

2

Service Manual

Table of Contents

1. Single Load Ratings IEC 60613……....………………………………………………….

3

2. Focus and tolerance between the focus and datum axis……………………………...

5

3. The attenuation of X-ray beam..………………………………………………………….

6

4. The location of focus mark .………...…………………………………………………….

7

5. X-ray tube teat data …… ...……………………………………………………………… 6. The limit of X-ray tube temperature …………………………………………………….

9 10

RADIEQ CO., LTD.

MX-600 2

Series

Mammography X-ray System

2

Service Manual

1. Single Load Ratings IEC 60613

RADIEQ CO., LTD.

MX-600 3

Series

Mammography X-ray System

2

Service Manual

RADIEQ CO., LTD.

MX-600 4

Series

Mammography X-ray System

2

Service Manual

2. Focus and tolerance between the focus and datum axis.

RADIEQ CO., LTD.

MX-600 5

Series

Mammography X-ray System

2

Service Manual

3. The attenuation of X-ray beam. 1. Measurement at Non-attenuation

X-Ray

xxx R/min

2. Measurement at 0.3mmAl

O.3mmAl X-Ray xxx R/min

Measurement equipment :Radiation Detector

3. Measurement at Bucky(Grid + FRP)

Grid X-Ray xxx R/min

Bucky

RADIEQ CO., LTD.

MX-600 6

Series

Mammography X-ray System

2

Service Manual

4. Measurement data ( At 30kV, 20mAs, Large focus) Item 1st mR Non-attenuation 15.28 0.3mmAl 8.29 Bucky(with Grid) 8.94

2nd mR 15.45 8.30 8.83

3rd mR 15.33 8.30 8.96

Avr. mR 15.35 8.30 8.91

% 100 54 58

Standard Pass

- Blank -

RADIEQ CO., LTD.

MX-600 7

Series

Mammography X-ray System

2

Service Manual

4. The location of focus mark

RADIEQ CO., LTD.

MX-600 8

Series

Mammography X-ray System

2

Service Manual

5. X-ray tube Test data PURPOSE : To verify housing/unit integrity under simulated catastrophic conditions. The T/U will be ran at a power level that will bring the housing to thermal equilibrium @ 80°C for 10 minutes, following this the T/U power will be increased to obtain 100% heat storage for addition 2 min. < Over pressure test>

RADIEQ CO., LTD.

MX-600 9

Series

Mammography X-ray System

2

Service Manual

6. The limit of X-ray tube temperature. - Max. Temperature shutdown Test PURPOSE : To verify that safety devices on the unit are functioning properly during an over temperature / pressure condition. Temperature rating of thermal cutoff : 80°C Thermocouple location : Center of housing : The thermocouple is actually rated at 79.4°C ±3.9°C. The thermocouple is inside the housing and connected in series with the common leg of the SRATOR. See attached and data.

RADIEQ CO., LTD.

MX-600 10

Series

Mammography X-ray System

2

Service Manual

RADIEQ CO., LTD.

MX-600 11

Series

Mammography X-ray System

3

Service Manual

Section 3. Calibration Manual of the MX-600

Diagnostic Mammography X-ray System

RADIEQ CO., LTD.

MX-600 1

Series

Mammography X-ray System

3

Service Manual

Table of Contents

1. Calibration Overview

………………...………………………………………………….

3

2. Required Staff and Used Tools ………...………………………………………………...

5

3. Pre-Check list

………...………………………………………………………………….

6

4. Calibration and Adjustment ………...…………………………………………………….

7

Block Diagram ………...……………………………………………………………….

7

4.1. Cover Removal of Control Box ………...……………………………………….

8

4.2. Mode Switch of CPU Board

9

………...………………………………………...

4.3. Selection Method in Adjustment Mode

………...…………………………….

10

4.4. Selection Method in Adjustment Sub Mode 1.x ………...…………………….

10

4.5. Initial Data Transmission of CPU Board ………...……………………………..

11

4.6. Connection of Measuring Instrument ...………………………………………...

13

4.7. Setting Adjustment Mode ...……………………………………………………...

14

4.8. Adjustment of X-ray Tube Current ...……………………………………………

15

4.9. Adjustment of AEC ...……………………………………………………………..

17

4.10 Mode Setting ..................……………………………………………………….

23

4.11 Data Sheet ...…………..…………………………………………………………

24

. 5. Mechanical

Adjustment ...………………………………………………………………

28

5.1. Adjusting the Light Field ...………………………………………………………

28

5.2. Adjusting X-ray exposure field ...………………………………………………..

29

RADIEQ CO., LTD.

MX-600 2

Series

Mammography X-ray System

3

Service Manual

1. Calibration Overview All new equipment installations will require that M101-5401 MAIN CPU board be adjusted. If you are comfortable with your evaluation of kV/mA output findings, proceed to AEC & mAs procedures. ALL Radieq Co.,Ltd. EQUIPMENT IS PRE-CALIBRATED AT THE FACTORY. THE FOLLOWING PROCEDURES ARE PROVIDE ONLY IN EVENT THAT THE UNIT FAILS TO MEET SPECIFICATIONS. It is recommended that all electrical and mechanical adjustments be checked annually. When performing the following procedures, follow the sequence at which the procedure is given, as well as its content. It is suggested that you calibrate the equipment annually, first verify the performance to determine whether or not a complete calibration is required. Always begin by checking the kV/mA/mAs DIGITAL REGERENCE SIGNAL that were recorded on the "DATA SHEET". You can either spot check or verify all selected ranges for output comparison for what was recorded on the "DATA SHEET" on the initial installation or previous calibration. Always start with a new copy of the "DATA SHEET" and add it to the equipment records. In the event the CPU(IC1,Main CPU Board) is replaced or the Non-volatile RAM should fail, the information on the "DATA SHEET" will reduce the number of X-ray's required to re-calibrate the mammography unit. By following the procedures, we are actually reprogramming the information fund in the RAM memory. This allows us to fine tune the broad exposure information supplied to the buffer from Non-volatile RAM chip. The information in the Non-volatile RAM will be maintained even with power off by inner the battery back-up system. Should this system fail for whatever reason, the RAM memory will be lost and no exposure can take place. In this event, the programming must be re-supplied to the Non-volatile RAM from the ROM memory and the kV and mA calibration performed again. Having recorded all of the data on the "DATA SHEET" and within this section of the manual, you should be able to quickly restore the conditions within a very short time.

RADIEQ CO., LTD.

MX-600 3

Series

Mammography X-ray System

3

Service Manual

Caution:

Remember we are dealing with small focal spot size. Take your time to prevent tube over loading, and you will gain faster result in calibration. Allow 30 seconds between exposures.

RADIEQ CO., LTD.

MX-600 4

Series

Mammography X-ray System

3

Service Manual

2. Required Staff and Used Tools The following personnel and tools are required to Calibration MX-600, and they must be organized beforehand. (1) Tools and instruments required (2) One set of general purpose tools „ „ „

Screwdriver (small and medium size): Flat screwdriver for adjustments: M6 and M10 wrenches:

Each 1 1 Each 1

(3) Unpacking tools:

1 set

(4) AC voltmeters to measure the power supply voltage:

1

(5) One X-ray tube voltmeter set: 1 set „ X-ray tube voltmeter/mAs meter as Dynalizer III or equivalent „ X-ray tube voltmeter cable (6) X-ray tube current meter: 1 set „ X-ray tube voltmeter/mAs meter as Dynalizer III or equivalent (7) Acrylic phantoms for AEC adjustment „ 10mm thickness (120mm x 120mm):

8

(8) Density meter:

1

RADIEQ CO., LTD.

MX-600 5

Series

Mammography X-ray System

3

Service Manual

3. Pre-Check List Equipment S/N:

Date Installed : Caution:

Do not apply power without first checking that all boards are properly seated and all connectors and wires are connected exactly, damaged could occur to the unit.

Changing the Power Regulator (PSU1) supply voltages can result in a change in the calibration of the mammography unit. Record below the initial PWR supply voltage setting, if a PWR supply is replaced adjust it to the recorded value below to avoid extensive re-calibration. 1) Turn on the line breaker on the Control Box side. 2) Check that all power supply voltages present, do not change outputs if they are within the tolerance of 2%: +5VDC

+12VDC

-12VDC

+24VDC

3) Turn off the line breaker on the Control Box side.

RADIEQ CO., LTD.

MX-600 6

PSU1

RADIEQ CO., LTD.

ROTOR & IF BOARD

MAIN CONTROL BOARD

MAIN OP BOARD

PSU2

SUB OP1 BOARD

M_INV BOARD1

INVERTER BOARD

COM. MOTOR DRIVER BOARD

MOTOR CONTROL BOARD

SUB OP1 BOARD

MX-600 BLOCK DIAGRAM M_INV BOARD2

HFG

TUBE

Series

Mammography X-ray System

Service Manual

3

4. Calibration and Adjustment

Fig.1 MX-600 Block Diagram.

MX-600 7

Mammography X-ray System

Series

3

Service Manual

4.1. Cover Removal of Control Box Before starting calibration and adjustment, please remove sideboards of control box as shown in Fig. 2.

4

2

3 1

Fig.2 Cover Removal of Control Box. 1) 2) 3) 4)

At first loosen screws and remove the front cover 1. Next loosen screws and remove the side cover 2. Next loosen screws and remove the side cover 3. Finally loosen screws and remove the rear cover 4.

RADIEQ CO., LTD.

MX-600 8

Mammography X-ray System

Series

3

Service Manual

The Boards are showed in Fig. 3 as below.

MAIN CPU BOARD

I/F&ROTOR BOARD

C1

PSU1

T1

S2 (MODE S/W)

TB10

HIGHTENSION TRANSFORMER

INVERTER

HIGHTENSION TRANSFORMER

FUSE BOARD

TB2 INVERTER

TB1 Control Box [ FRONT ]

Control Box [ BACK ]

Fig. 3 Position of Boards

4.2 Mode Switch of CPU Board Each following mode is chosen by selection of the number of the MODE switch(S2) on Main CPU board. 0 - - - - - - - - - - - - - - - Normal mode(Operation mode) 1 - - - - - - - - - - - - - - - Adjustment mode (Calibration mode)

CE F012

3 RADIEQ CO., LTD.

Mode switch : S2

MX-600 9

Series

Mammography X-ray System

3

Service Manual

4.3 Selection Method in Adjustment Mode The contents of Display in the adjustment mode are displayed by 7 segment Light Emitting Diode on the left-hand side of kgf. The adjustment mode is chosen by Density switch “▼“ 0‐‐‐‐‐‐‐‐ Adjusting mode of mA

0.

kgf

1‐‐‐‐‐‐‐‐Adjusting mode of AEC

1.

kgf

2 3 4

‐‐‐‐‐‐‐ Setting for a various Mode

2.

kgf

‐‐‐‐‐‐‐Selection mode of the transmission data

3.

kgf

4.4 Selection Method in Adjustment Sub Mode 1.x The contents of Display in the adjustment mode are 7 segment LED on the right-hand side of Density. It is displayed numerically. The switch “▲” of Density chooses the adjustment sub mode. In the case of AEC adjustment: 0 - - - - - - - - - - - - - - Semi AEC adjustment 1-------------

RADIEQ CO., LTD.

Density Data adjustment

1. 0

kgf

1. 1

kgf

MX-600 10

Mammography X-ray System

Series

3

Service Manual

4.5 Initial Data Transmission of CPU Board Initial data other than individual data of the following clause is transmitted to Non-volatile RAM. (1)Disconnect the connection cable between Inverter and MAIN CPU board(M101-5401) at MAINCPU board side. (2)The MODE switch S2 of CPU board is set as No. 1. (3)Turn on the circuit breaker

kV

mAs

DENSITY

The display “—“ of kV blinks. NOTE:



“ is showed a blank mark.

4.5.1 (1)

CPU Board Individual Data Transmission Set up the adjustment mode Push Switch"▼" of DENSITY. Then a display is made into 3. by 7 segment LED on the left side of Density. 3.

0 kV

mAs

DENSITY

kgf mm DEG.

(2)

Push Switch “▲” “▼“ of kV, A display of 7 segment LED of kV is made into the following selection number. 8…………………………...CPU RAM to Non-volatile RAM 3.

8 kV

mAs

DENSITY

kgf mm DEG.

(3)

RADIEQ CO., LTD.

Pushing AEC switch is continued for 1 second. 7 segment LED of mAs blinks, about 10 seconds after, a buzzer sounds for 1 second and completion is told.

MX-600 11

Series

Mammography X-ray System

3

Service Manual

(4)

Push Switch “▲” “▼“ of kV, A display of 7 segment LED of kV is made into the following selection number. 9 ………………………….Non-volatile RAM to Buffer 3.

9 kV

mAs

DENSITY

kgf mm DEG.

(5)

Pushing AEC switch is continued for 1 second. 7 segment LED of mAs blinks, about 10 seconds after, a buzzer sounds for 1 second and completion is told.

NOTE: 4.5.1 clauses needs to be operated, only when the Non-volatile RAM for backup the case where MAIN CPU board is changed into a new , or CPU board is exchanged Here, since the data transmission from RAM / Buffer finished to Non-volatile RAM, Line Breaker is turned off, and the cable between an inverter and CPU board is inserted in a connector.

RADIEQ CO., LTD.

MX-600 12

Mammography X-ray System

Series

3

Service Manual

4.6 Connection of Measuring Instrument Connect tube voltmeter and mA or mAs meters as shown in Fig. 4. High-voltage transformer High-voltage cable for measurement

Control box

Unit high-voltage cable Anode

E

TB 2 MA+

X-ray tube

MA-

Cathode

mA meter

kV meter

or

mAs meter

Fig. 3 Connection for kV and mA/mAs meter (1) X-ray tube voltmeter Between a high-voltage generator and an X-ray tube, the breeders for X-ray high-voltage measurement (DYNALYZER, etc.) are connected using the high-voltage cable adapter for the mammography device. (2) X-ray tube current The jumper line between terminal MA+ to MA- of X-ray tube current measurement terminal stand TB2 is removed, and an mA meter(or mAs meter) is connected between them.

Caution:

If between terminal MA+ to MA- of the terminal stand TB2 become open, the unusual high voltage is impressed by the X-ray tube, and since it may be damaged, you be careful of connection of an ammeter enough.

RADIEQ CO., LTD.

MX-600 13

Mammography X-ray System

Series

3

Service Manual

4.7 Setting Adjustment Mode (1) The MODE switch S2 of CPU board is set as position 1. (2) Turn on Line breaker.

kV

mAs

DENSITY

Display becomes as it is shown in a figure. NOTE: The setting method of the MODE switch S2 unites a number with the portion with a cut of a knob.

Caution:

X-ray exposure of continuation is not made. Since there is a possibility that equipment may be damaged, please take the pause time of 30sec or more in X-ray exposure.

(3)Push On-(internal) switch of Control Panel . Switch “ I” ON(

)

The control circuit will be activating. Display becomes as it is shown in a figure.

Lamp LARGE F.S. lights up.

(4) The switch ” ▼ “ of DENSITY is pushed and the adjustment mode is chosen. Display becomes as it is shown in a figure. 0.

2 2 X X. X kV

mAs

DENSITY

kgf mm DEG.

RADIEQ CO., LTD.

MX-600 14

Series

Mammography X-ray System

3

Service Manual

4.8 Adjustment of X-ray Tube Current NOTE:

There is no adjustment of X-ray tube voltage.

An adjusting point is shown below. „ Large Focus:22kV, 25kV, 30kV, 35kV, 39kV . „ Small Focus:22kV, 25kV, 30kV, 35kV 4.8.1Adjustment of mA NOTE: Record this adjustment data to DATA SHEET which are indicated by 4.12.1 clause. These data is used as future adjustment and reference data at the time of maintenance. 1)

A setup in the adjustment mode Display becomes as it is shown in a figure. Lamp LARGE F.S. lights up.

0.

2 2 X. X X kV

mAs

kgf

DENSITY

2) The switch "▲" of mAs and "▼" are pushed and mA data is inputted. 0.

2 2 7 5. 8 kV

mAs

kgf

DENSITY

A switch "AEC" is pushed and data is saved.

3) An X-ray exposure switch is pushed and expose, X-ray tube voltage and current are measured.

The tube current is displayed on mAs display position during about 5 seconds after an X-ray exposure end. (At the time of the adjustment mode).

RADIEQ CO., LTD.

MX-600 15

Mammography X-ray System

Series

3

Service Manual

When measured value does not fulfill specified value, a switch "▲" or "▼" of mAs is pushed, mA setting value is changed, and it measures again. If standard value is satisfied, switch "AEC" is pushed and data is saved.

4) The setting value of kV is changed by switch "▲" or "▼" of kV, and measurement and adjustment of 3) are performed about each kV. LARGE Focus : ① ② ③ ④ ⑤

22KV Æ 70mA 25KV Æ 85mA 30KV Æ 72mA 35KV Æ 60mA 49KV Æ 60mA

5) A switch "HELP" is pushed and a focus is changed to a small focus.

It is made 22kV. Display becomes as it is shown in a figure. SMALL F.S. lamp turns on.

NOTE:If "HELP" switch is pushed again, it will change to a large focus. 6)Adjustment to 1 to 5 clause performed on the large focus is similarly performed with a small focus. 0.

2 2 X. X X kV

mAs

kgf

DENSITY

SMALL Focus : ① ② ③ ④

RADIEQ CO., LTD.

22KV Æ 13mA 25KV Æ 15mA 30KV Æ 12mA 35KV Æ 10mA

MX-600 16

Mammography X-ray System

Series

3

Service Manual

4.9 Adjustment of AEC The tube voltage set as adjustment is 24kV, 27kV, 30kV, 32kV and 35kV, and a radiograph subject thick is 20mm, 40 mm, 60mm, and 80mm. Adjustment of MAIN CPU board of the clause 4.9.1 is done at the time of factory shipment. When the time of exchange of MAIN CPU board or AEC Detector About density adjustment of the clause 4.9.2 and 4.9.3 when it is installed or changed to other combination with film and a screen, please carry out. 4.9.1 AEC Tilt Adjustment of CPU Board 1) The switch "▼" and "▲" of DENSITY is pushed and the adjustment mode is set to 1.1 Display becomes as it is shown in a figure.

2)

24

2. 5

kV

mAs

1. 1

kgf

DENSITY

Conditions are set up. Tube Voltage: 24 kV(Fixed) Radiography conditions: ① BUCKY device ② Compression paddle ③ 50mm acrylics

3)The switch "▼" or "▲" of mAs is pushed, and mAs display is set to 2.5. Display becomes as it is shown in a figure.

24

2. 5

kV

mAs

1. 1

kgf

DENSITY

4)A switch “AEC” is pushed and data is saved.

RADIEQ CO., LTD.

MX-600 17

Series

Mammography X-ray System

3

Service Manual

5 ) An X-ray exposure switch is pushed and X-rays is exposed, then Voltage/Time is measured at the test pin TP1 on CPU board.

6)The adjustment of the relation between the voltage waveform of TP1 and time is carried out by VR1(at Main CUP board) at approximately 2.5V/1sec as shown in the following figure. 2.5V

*Adjustment of VR1

1.0sec

(Time will be extended if it rotates to clockwise)

4.9.2 100mAs Adjustment of CPU Board 1) Same condition 4.9.1.(AEC Tilt Adjustment of CPU Board) Conditions are set up. Tube Voltage:24 kV(Fixed) Radiography conditions: ① BUCKY device ② Compression paddle ③ 50mm acrylics 2) The switch "▲" of DENSITY is pushed and the adjustment mode is set to 1.1.

3) The adjustment of the relation between the voltage waveform of mAs meter or Dynalizer III is carried out by VR2(at Main CPU board) at exactly 100mAs.

*Adjustment of VR2

RADIEQ CO., LTD.

(Time will be extended if it rotates to clockwise)

MX-600 18

Series

Mammography X-ray System

3

Service Manual

4.9.3 Density Adjustment in Bucky Device 1) Bucky device is attached in a radiography stand. Radiography conditions: ① BUCKY device ② Compression paddle ③ 20mm acrylics : 4pcs ④ Cassette(with a non-exposed film) 2)The switch "▲" of DENSITY is pushed and the adjustment mode is set to 1.0. 1.01 1.

2 4 0. 1 4 kV

mAs

kgf

DENSITY

3) The switch of “FILM & SCREEN” is pushed and "1" is chosen. “1” Factory Setting, “2,3” User Setting.

1: KORDARK FLIM(MIN R) + KORDARK CASSETTE(MIN R)

4) It is set as 24kV by the switch "▼" or "▲" of kV. (acrylics 20mm) It is set as density data by the switch "▼" or "▲" of mAs Display becomes as it is shown in a figure. 1.01 1.

2 4 0. 1 4 kV

5)

mAs

kgf

DENSITY

A switch "AEC" is pushed and data is saved.

6)An X-ray exposure switch is pushed and X-ray is exposed.

The exposed film is developed and density is measured. NOTE 1.30~ 1.40 of density is set when the unit released from the factory.

RADIEQ CO., LTD.

MX-600 19

Series

Mammography X-ray System

3

Service Manual

7) When density is darker, it is smaller in setting value, and setting value is enlarged when density is lighter. 8)The operation from 6) clause to 7) clause is repeated until density suits.

9)It is set as 27kV by the switch "▼" or "▲" of kV. (acrylics 20mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

2 7 0. 1 4 kV

mAs

kgf

DENSITY

10) The operation from 6) clause to 7) clause is repeated until density suits. 11) It exchanges for an acrylics 40mm board, and the tube voltage is set as 24kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

2 4 0. 2 4 kV

mAs

kgf

DENSITY

12)The operation from 6) clause to 7) clause is repeated until density suits. 13)It is set as 27kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

2 7 0. 2 8 kV

mAs

kgf

DENSITY

14)The operation from 6) clause to 7) clause is repeated until density suits. 15)

It is set as 32kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

3 2 0. 3 8 kV

mAs

kgf

DENSITY

16) The operation from 6) clause to 7) clause is repeated until density suits.

17)

It is set as 35kV by the switch "▼" or "▲" of kV. (acrylics 40mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

3 5 0. 5 0 kV

RADIEQ CO., LTD.

mAs

kgf

DENSITY

MX-600 20

Series

Mammography X-ray System

3

Service Manual

18) The operation from 6) clause to 7) clause is repeated until density suits. 19) It exchanges for an acrylics 60mm board, and the tube voltage is set as 27kV by the switch "▼" or "▲" of kV. It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

2 7 0. 4 8 kV

mAs

kgf

DENSITY

20) The operation from 6) clause to 7) clause is repeated until density suits. 21)It is set as 30kV by the switch "▼" or "▲" of kV. (acrylics 60mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

3 0 0. 6 0 kV

mAs

kgf

DENSITY

22) The operation from 6) clause to 7) clause is repeated until density suits. 23)It is set as 32kV by the switch "▼" or "▲" of kV. (acrylics 60mm) It is set as density data by the switch "▼" or "▲" of mAs

1. 10

3 2 0. 6 4 kV

mAs

kgf kgf

DENSITY

24) The operation from 6) clause to 7) clause is repeated until density suits. 25) It is set as 35kV by the switch "▼" or "▲" of kV. (acrylics 60mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

3 5 0. 7 4 kV

mAs

kgf

DENSITY

26) It exchanges for an acrylics 80mm board, and the tube voltage is set as 32kV by the switch "▼" or "▲" of kV. It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

3 2 0. 9 4 kV

mAs

kgf

DENSITY

27) The operation from 6) clause to 7) clause is repeated until density suits.

RADIEQ CO., LTD.

MX-600 21

Mammography X-ray System

Series

3

Service Manual

28)It is set as 35kV by the switch "▼" or "▲" of kV. (acrylics 80mm) It is set as density data by the switch "▼" or "▲" of mAs 1.01 1.

3 5 1. 0 2 kV

mAs

kgf

DENSITY

29) The operation from 6) clause to 7) clause is repeated until density suits.

RADIEQ CO., LTD.

MX-600 22

Series

Mammography X-ray System

3

Service Manual

4.10 Mode Setting (1) Set up the adjustment mode Push Switch"▼" of DENSITY. Then a display is made into 2. by 7 segment LED on the left side of Density. Display becomes as it is shown in a figure.

0

0

kV

mAs

2.

kgf

DENSITY

(2) Push Switch “▲” “▼“ of kV, select modes. And Push Switch “▲” “▼“ of mAs Select data. 0‐‐‐Focal spot 1‐‐‐Filter conversion 2‐‐‐Autofilmchanger 3‐‐‐Rh Filter 4‐‐‐REL conversion

1:Manual 0:Auto 1:Manual 0:Auto 1:Install 0: Not Install 1:Rh 0:Al 1:Manual 0:Auto

(3) A switch "AEC" is pushed and data is saved.

4.10.1 Fiducial loading condition

kV 22 25 30 35 36 37 38 39

RADIEQ CO., LTD.

Large focus mA mAs(max) 70 600 86 600 72 600 60 600 60 500 60 400 60 320 60 320

kV 22 25 30 35

Small focus mA mAs(max) 13 100 15 100 12 100 10 100

MX-600 23

Series

Mammography X-ray System

3

Service Manual

4.11 Data sheet 4.11.1

Adjustment and measured value of X-ray tube voltage and current

Performed date : Condition: Large Focus Set Value

Measured Value by mA adjusted the instruments Ref. Digital Code kV mA

Specification Deviation 4% of Set Value

KV

mA

KV

mA

22

70

21.2



22.8

67.2



72.8

25

85

24.0



26.0

81.6



88.4

30

72

28.8



31.2

69.1



74.9

35

60

33.6



36.4

57.6



62.4

39

60

37.4



40.5

57.6



62.4

Small Focus Set Value

Measured Value by mA adjusted the instruments Ref. Digital Code kV mA

Specification Deviation 4% of Set Value

KV

mA

22

13

21.2



22.8

12.5



13.5

25

15

24.0



26.0

14.4



15.6

30

12

28.8



31.2

11.5



12.5

35

10

33.6



36.4

9.6



10.4

RADIEQ CO., LTD.

KV

mA

MX-600 24

Mammography X-ray System

Series

3

Service Manual

4.11.2 mAs measurement table Performed date: Condition: Radiography method:

Set Value

Measured Value by the instruments L.F.

mAs

BUCKY + compression paddle + 50mm acrylic

22kV / 70mA

S.F.

mAs

22kV / 13mA mAs

Specification Deviation 4% of Set Value

1.00

0.90



1.10

1.30

1.17



1.43

1.60

1.44



1.76

2.00

1.80



2.20

3.20

2.88



3.52

4.00

3.60



4.40

5.00

4.50



5.50

6.40

5.76



7.04

8.00

7.20



8.80

10.0

9.00



11.00

20.0

18.00



22.00

50.0

45.00



55.00

100

90.00



110.00

540.00



660.00

600

RADIEQ CO., LTD.







MX-600 25

Mammography X-ray System

Series

3

Service Manual

4.11.3 The data sheet for mAs of AEC mode Performed date: Condition: Radiography method

KV

24

Thick. of acrylic (mm)

1 Ref. of mAs

Actual mAs

BUCKY (L.F.)

Magnification(S.F.)

2 Density

Ref. Actual of mAs mA

3 Density

Re f. of

Actual mAs

Density

20 40 20

27

40 60

30

40 60 40

32

60 80 40

35

60 80

RADIEQ CO., LTD.

MX-600 26

Mammography X-ray System

Series

3

Service Manual

4.11.4

The data sheet for Density of AEC mode

Performed date: Condition: Radiography

Density Position

%

-9

-45

-8

-40

-7

-35

-6

-30

-5

-25

-4

-20

-3

-15

-2

-10

-1

-5

0

0

1

+5

2

+10

3

+15

4

+20

5

25

6

+30

7

+35

8

+40

9

+45

RADIEQ CO., LTD.

27kV

BUCKY + Comp. Paddle + 40mm acrylic

Adjustment coefficient

Actual mAs

Density

Actual %

MX-600 27

Series

Mammography X-ray System

3

Service Manual

5. Mechanical Adjustment 5.1. Adjusting the Light Field 1) Do not attach any units to the mammography table (Bucky unit, etc). 2) Mount the 24 x 30 cm mask. 3) Turn on the manual switch for field lamp, then measure dimensions a, b, and c as Fig.

c

b

a Fig. 5

4)

Check the size of a, b, and c. If the following conditions are met, adjustment is not necessary. | a – b| ≤ 6mm, c = 230 ±6mm.

RADIEQ CO., LTD.

MX-600 28

Series

Mammography X-ray System

3

Service Manual

5.2.

Adjusting X-ray exposure field Preparation Prepare the cassette shown by Fig. 6. z The edge marker of an aluminum board with a thickness of 0.5mm and a width of 10mm is stuck on a cassette from edge at about 20mm place. z A marker is given with an aluminum board with a thickness of 0.5mm so that right and left of a film may be known.

„

about 20mm

Radiography table

Cassette

Marker Edge marker Fig. 6 Test cassette (a) (b) (c) (d)

Set the cassette as shown in Fig. 6. Mount the 24 x 30 mask Exposure X-ray to the film ( large focus, 30kV – 20mAs ) Develop the film, and measure any exposure field deviation.

Center of the film X-ray exposure field

c a

b

Fig. 7 If | a – b| ≤ 3mm and 1 ≤ c ≤ 4, adjustment is not required. RADIEQ CO., LTD.

MX-600 29

Series

Mammography X-ray System

3

Service Manual

If the conditions above are not satisfied, then proceed with (f) below. (e) Place the developed film on the cassette, then the cassette is set by the preposition. (f) Align the lamp exposure field with the X-ray exposure field shown on the film. (g) Using the adjusted lamp field as a guide, move the mask holder by loosening the four M3 nuts fixing it, so that the conditions above are met.

Four M3 nuts

Mask holder Fig. 8 (h) After the lamp exposure field has been adjusted to within the conditions above, tighten the four M3 nuts, exposure X-ray and check the results on the film. (i) After checking the developed film, make sure that all nuts are securely tightened, then remount the covers.

RADIEQ CO., LTD.

MX-600 30

Series

Mammography X-ray System

4

Service Manual

Section 4. Maintenance Manual of the MX-600

Diagnostic Mammography X-ray System

RADIEQ CO., LTD.

MX-600 - 1

Mammography X-ray System

Series

4

Service Manual

Table of Contents 1. Maintenance

……………………………………………………………………………..

1.1. Everyday Check and Inspection 1.1.1. Check Display

3

………………………………………………

3

……………….……………………………………………

3

1.1.2. Check Radiography Stand

…………………………..………….....……

5

1.1.3. Evaluation Film Quality ( Density ) ………………………………….……

6

1.2. Cleaning and Disinfection 1.2.1. Cleaning

…………………………………..……….…...……

7

…………………………………………..………………..…..…

7

1.2.2. Disinfection

……………………………………………….……….………

1.3. Periodical Inspection

…………………...……………………..……………

7 8

2.1. REPLACEMENT OF LAMP …………………………………………………….. 12 2.2. REPLACEMENT OF LAMP …………………………………………………….. 13 2.3. Messages and Corrective Actions ……………………………………………… 15 2.3.1 Error & Check Massages ……………………………………………….… 15

RADIEQ CO., LTD.

MX-600 - 2

Series

Mammography X-ray System

4

Service Manual

1.

Maintenance 1.1 Everyday Check and Inspection A user needs to check that all safe related equipment operate pertinently, before using equipment for diagnosis. Attention also needs to be paid to check of whether a display and a pilot lamp operate correctly. An X-ray exposure indicator is turned on only within while operating the X-ray radiography button. While not operating these buttons, when this display lights up, the X-ray equipment must be stopped immediately and must be connected with our company service part. 1.1.1 Check Display (1) Engage the circuit breaker located on the Control Box side. The mark “-“ in the kV display on the panel flashes JUDGEMENT

.

(2) Then push the ON (internal) switch located on the Control Panel The display of the last radiographic condition is lighted. JUDGEMENT

.

(3) Press the kV switch +, and kV value on the display increases. JUDGEMENT

.

(4) Press the kV switch - , and kV value on the display decreases. JUDGEMENT

.

(5) Press the mAs switch + , and mAs value on the display increases. JUDGEMENT

.

(6) Press the mAs switch - , and mAs value on the display decreases. JUDGEMENT

.

(7) Remove the BUKY DEVICE. The BUKY lamp should extinguish. Next attach the Magnification box. MAGNIFY and SMALL F.S. lamp should illuminate. . JUDGEMENT

RADIEQ CO., LTD.

MX-600 - 3

Series

Mammography X-ray System

4

Service Manual

(8) Depress the AEC switch. Each time the display will sequence between MANUAL and Full AEC, Semi AEC. . JUDGEMENT (9) In Full AEC mode the kV and the mAs display should be blanked. JUDGEMENT

.

(10) In Semi AEC mode the mAs display should be blanked. JUDGEMENT

.

(11) In AEC mode the DENSITY display should illuminate. JUDGEMENT

.

When depress the DENSITY switch, DENSITY display should be variable between +9 to –9. . JUDGEMENT (12) Depress the FILM SPEED switch. Each time the display will sequence between 1, 2, 3 to 1. . JUDGEMENT

RADIEQ CO., LTD.

MX-600 - 4

Series

Mammography X-ray System

4

Service Manual

1.1.2. Check Radiography Stand (1) Vertical and Rotation movement of C-arm C-arm rotation Switch

This switch can make C-arm’s rotation. :Counterclockwise

:Clockwise. C-arm Up / Down switch

Keep pressing to make C-arm’s continuous rotation :Up movement

:Down movement

JUDGEMENT

.

(2) Attachment and detachment of a pressure paddle Each pressure paddle of accessories can attach and detach easily among a pressure paddle support part. . JUDGEMENT (3) Vertical movement of a Compression Paddle • Where a pressure paddle is attached, turn a Manual compression knob by hand and carry out vertical movement smoothly. JUDGEMENT . • Where a pressure paddle is attached, operate a Compression foot switch and carry out vertical movement smoothly. JUDGEMENT .

RADIEQ CO., LTD.

MX-600 - 5

Series

Mammography X-ray System

4

Service Manual

1.1.3.Evaluation Film Quality ( Density ) The quality control test of a film development machine is performed, and after checking that the parameter of a film development machine is in control within the limits, the function of automatic exposure control (AEC) is checked by performing phantom photography and checking film density. The MX-600 is designed to provide consistent densities on film across the full spectrum of tissue densities、regardless of breast size. Normally adjusted to provide an optical density of 1.4 at AEC mode, it can be adjusted to any density specified. This can be done to suit particular tastes or to compensate for processing irregularities etc..

RADIEQ CO., LTD.

MX-600 - 6

Series

Mammography X-ray System

4

Service Manual

1.2. Cleaning and disinfection Caution:

When performing cleaning and disinfection, it is required to turn off a power supply input.

1.2.1 Cleaning The parts that contact a patient among the parts of equipment must be cleaned before each use. Please make cloth become wet with neutral detergent thinned with water, and wipe equipment. . Please do not use a cleanser containing abrasive and cleaning liquid containing organic solvents or solvents (for example, benzene, a stain-removal solvent).

Caution:

Please do not hang a spray on equipment! cleaning equipment cleaning liquid.

In any case, don’t permeate

1.2.2.Disinfection Immerse the compression paddle, the magnification kit and the Bucky devices into a cleaning disinfection solution of minimum 2.4% glutaldehyde (ex. CIDEX, Omnicide., etc.) in a separate container for 45 minutes in every 14 days. Make sure the surface is completely moistened. After that take it off and rinse it with the clean water.

Caution:

1. Switch off the device before disinfecting . Disconnect it from the mains. 2. Take care that no liquid gets into the equipment. 3. If you put into the disinfection fluid too long, the life time of the compression paddles, the magnification kit and the Bucky device are reduced. 4. Do not perform the disinfection with E.O. Gas. 5. Do not use the degenerated solvent.

RADIEQ CO., LTD.

MX-600 - 7

Series

Mammography X-ray System

4

Service Manual

1.3. Periodical Inspection In order to secure the safety of a patient, an operator, and a third person, equipment needs to be checked every 12 months and reliability and operatively need to be maintained. The parts which may cause danger in respect of safety if it exhausts need to undergo the inspection by the qualified person periodically every at least 12 months. Moreover, parts need to be exchanged if needed. When the local rule and regulation define still frequent check and maintenance, please follow the regulation. 1. Stand (Inspecting the C-arm support) Item

1

2

3

Contents

1) Remove all cable lubricant from the cable with organic solvent. 2) Make sure there are no broken strands at the Cable cable ends or sections inspection that contact pulleys. Apply new cable lubricant uniformly to sections which contact the pulley over the entire C-arm vertical stroke. Vertical Press movement switches movement on both sides of the C-arm inspection to check operation Press movement switches Vertical on both sides of the C-arm movement belt to check operation. Make sure that the belt inspection fixing nut is not loose.

RADIEQ CO., LTD.

Standard

No-broken strands

Insp. jig

Inter-val 6M 1Y

Processing

Replace the cable.



Must operate



Should operate smoothly. Activation force should be 5kgf or less.



Replace belt.

MX-600 - 8

Series

Mammography X-ray System

4

Service Manual

Item 4

5

Insp. jig

Interval 6M 1Y

Contents

Standard

Rotational movement inspection

Press movement switches on both side of the C-arm to check operation.

Must operate



Cable operation confirmation

Move C-arm up and down; make sure that the cable moves smoothly inside the stand.

Make sure that the cable does not interfere with support section vertical movement.



Processing

Adjust cable

2. C-arm unit Item

1

2

3

Rotational shaft bracket screw tightness check Screw tightness check for the aluminum casting section of the mammography table mount Screw tightness check for the aluminum frame of the tube mount

Standard

Remove the rotational shaft section cover, then make sure that the six M8 screws are tight.

Screws must be tight



Tighten

Remove the plastic cover, and make sure that the four M6 screws are tight.

The screw must be tight.



Tighten

Remove the plastic cover, and make sure that the four M6 screws are tight.

The screws must be tight.



Tighten

RADIEQ CO., LTD.

Insp. jig

Interval 6M 1Y

Contents

Processing

MX-600 - 9

Series

Mammography X-ray System

4

Service Manual

3. Inspection the compression unit Insp. jig

Interval 6M 1Y

Item

Contents

Standard

1

Chain tension and damage check

1) Because tension cannot be measured, check it by touch 2) Check for damage.

When the belt is pressed by hand. There should not be loosened too.



2

Positioning motor drive inspection

It checks that pressure power is below 5kgves.

Less than 5kgf



Processing

4. X-ray high voltage unit Item

Tube voltage, Tube current 1 and mAs check

Contents

Insp. jig

Interval 6M 1Y

kV: 24.0 to 26.0 Generate X-ray using the mA: conditions below, 81.6 to 88.4 checking tube current mAs: and mAs. 18.0 to 22.0 Large focus: 25kV, 20mAs Small focus: 25kV, 10mAs

RADIEQ CO., LTD.

10

Standard



kV: 24.0 to 26.0 mA: 14.4 to 15.6 mAs: 9.00 to 11.0

Processing

If it is not satisfied, perform adjustment Refer Service Manual

MX-600 -

Series

Mammography X-ray System

4

Service Manual

5. AEC (Film density check) Item

1

2

Film density check

Actual mAs display check

Contents

Insp. jig

Perform AEC radiography using a 40mm thick acrylic *40mm phantom using the tube Make sure that thickness voltage setting below and the density is acrylic with DENSITY at “0”, then within the phantom check density. reference density set at Conditions the time of * Tube voltage: 23kV *Density installation. * Using the film system meter and the screen. * Bucky radiography Connect a mAs meter according to the tube voltage/current measurement methods The difference given section 4.6, then between the perform AEC compression mAs meter radiography using the and the *mAs conditions below. Make operation sure that the operation meter panel mAs panel mAs display is display should accurate. be within ±10%. - 40mm thickness acrylic phantom - Tube voltage: 25kV - DENSITY: 0%

RADIEQ CO., LTD.

11

Standard

Interval 6M 1Y

Processing



If it is not satisfied, perform adjustment Refer Service Manual



If it is not satisfied, perform adjustment Refer Service Manual

MX-600 -

Series

Mammography X-ray System

4

Service Manual

2.1. REPLACEMENT OF LAMP Replacement of halogen lamp (1)Tool required Screw driver Glove to replace the lamp

1 sec 1pair

(2)Lamp specification

Model : HLWT-J / 12V, 50W Maker :

NARVA SPEZIALLAMPEN Main in GERMANY

(3)Replacement procedure ① Remove the cover plate.

② Change the halogen lamp. (Don’t touch directly glass)

③ Get the cover plate back to its original position. ④ Press the lamp switch and make sure that lamp is lit up.

CAUTION Be sure not to touch the lamp with your bear hands as it is very hot, use gloves to protect from heat.

RADIEQ CO., LTD.

12

MX-600 -

Series

Mammography X-ray System

4

Service Manual

2.2. Bucky Device Maintenance for cassette size (for Kodak MIN-R II and Standard cassette) If the cassette thickness is not about 19mm(Kodak MIN-R type) that is 14.5mm(Standard cassette type) then please replace stopper and attach guide rail. * If the guide rail installed in the bucky then that is standard cassette type(14.5mm). (See No.9 Picture) (Standard cassette thickness : about 14.5mm -Kodak MIN-R II , Standard cassette type , CR Mammo cassette) Exchange of Part in order to use a standard cassette

1. Exchange of Cassette Stopper

2. Removal of Old Stopper

3. Exchange to New Stopper

4. Exchange of Both side Stopper

5. Setting of a Cassette Guide Rail

RADIEQ CO., LTD.

13

MX-600 -

Series

Mammography X-ray System

4

Service Manual

6. Tightening Bolts at Two points of Front Guide Rail

8. Tightening Bolts at Two points of Rear Guide Rail

RADIEQ CO., LTD.

14

7. Setting of a Rear Cassette Guide Rail

9. Completion of exchanging parts

MX-600 -

Mammography X-ray System

Series

4

Service Manual

2.3 Messages and corrective actions 2.3.1. Error & Check messages This unit has self-diagnostic function. Check when INHIBIT lamp is on.



Code

Corrective action

C01 – 00

Cassette is not inserted in Insert cassette into bucky device. bucky device

C01 – 01

Exposed cassette is in bucky.

Insert cassette with new film.

C02 – 01

Sensor can not sense filter.

Filter might be stuck at somewhere out of sensing range. If operation is unavailable, call engineer.

C04 – 00

No beam limiting mask is set.

Insert beam limiting mask.

C04 – 01

Beam limiting mask is 18×24 Exchange cassette with same size one. type but cassette is not for it.

C04 – 02

Beam limiting mask is 24×30 Exchange cassette with same size one. type but cassette is not for it.

C06 – 00

BUCKY error

RADIEQ CO., LTD.

15

Meanings of error

Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.

MX-600 -

Series

Mammography X-ray System

4

Service Manual

Error code

Corrective action

C06 – 01

Inverter / alarm 1 error

Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.

C06 – 02

Inverter / alarm 2 error

Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.

C06 – 03

Inverter / filament error

Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.

C06 – 04

Turn off circuit breaker then ON again. X-ray tube anode rotation error Turn on “ I “ switch. / Tube over heat If error repeats, call engineer.

C08 – 00

Turn off circuit breaker then ON again. Main Board and Motor Turn on “ I “ switch. Controller connection error If error repeats, call engineer.

C08 – 01

Emergency switch pushed

Emergency switch turn off. Turn off circuit breaker then ON again. Turn on “ I “ switch.

C08 – 02

Motor not moving error

Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer.

C09 – 00

X-ray exposure setting mAs data.

small

then

C09 – 01

X-ray exposure setting mAs data.

large

then

C09 – 02

X-ray exposure mistake. CHECK lamp on error.

RADIEQ CO., LTD.

16

Meanings of error

Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer. Turn off circuit breaker then ON again. Turn on “ I “ switch. If error repeats, call engineer. X-ray can not be made during check lamp is on. Press CLR switch to restart.

MX-600 -

Series

Mammography X-ray System

5

Service Manual

Section 5. Schematics of the MX-600

Diagnostic Mammography X-ray System

RADIEQ CO., LTD.

MX-600 1

Series

Mammography X-ray System

5

Service Manual

Table of Contents 1

BLOCK DIAGRAM

2

FUSE BOARD : M101-5403

………………………………………………..………

4

3

COMPRESS MOTOR DRIVE(HEADER BOARD) : M101-5101……….…………

5

4

INTERFACE & ROTOR BOARD : M101-5402……………………………......……

6

5

MAIN CONTROL BOARD : M101-5401………………………………...……...……

7

6

MAIN OP BOARD : M101-5501………………………………..……….…...…….…

8

7

MOTOR INVERTER BOARD : M101-5405..…………………………..……..…..…

9

8

MOTOR CONTROL BOARD : M101-5103………………………….………………

10

9

SIDE OP BOARD : M101-5201...………………………...……………..……………

11

10

HFG PART BLOCKDIAGRAM

12

RADIEQ CO., LTD.

…………………………………………………………..………

.………………………...……………..……………

3

MX-600 2

Series

Mammography X-ray System

5

Service Manual

1.

BLOCK DIAGRAM

RADIEQ CO., LTD.

MX-600 3

Series

Mammography X-ray System

5

Service Manual

2.

FUSE BOARD : M101-5403

RADIEQ CO., LTD.

MX-600 4

Series

Mammography X-ray System

5

Service Manual

3.

COMPRESS MOTOR DRIVE(HEADER BOARD) : M101-5101

RADIEQ CO., LTD.

MX-600 5

Series

Mammography X-ray System

5

Service Manual

4.

INTERFACE & ROTOR BOARD : M101-5402

RADIEQ CO., LTD.

MX-600 6

Series

Mammography X-ray System

5

Service Manual

5.

MAIN CONTROL BOARD : M101-5401

RADIEQ CO., LTD.

MX-600 7

Series

Mammography X-ray System

5

Service Manual

6.

MAIN OP BOARD : M101-5501

RADIEQ CO., LTD.

MX-600 8

Series

Mammography X-ray System

5

Service Manual

7.

MOTOR INVERTER BOARD : M101-5405

RADIEQ CO., LTD.

MX-600 9

Series

Mammography X-ray System

5

Service Manual

8.

MOTOR CONTROL BOARD : M101-5103

RADIEQ CO., LTD.

10

MX-600

Series

Mammography X-ray System

5

Service Manual

RADIEQ CO., LTD.

11

MX-600

Series

Mammography X-ray System

5

Service Manual

9.

SIDE OP BOARD : M101-5201

RADIEQ CO., LTD.

12

MX-600

Series

Mammography X-ray System

5

Service Manual

10.

HFG PART BLOCKDIAGRAM

RADIEQ CO., LTD.

13

MX-600

Series

Mammography X-ray System

5

Service Manual

RADIEQ CO., LTD.

14

MX-600