US-751
Service Manual
ITO CO., LTD. 1st Edition
US-751 Repair Information Menu Clicking the blue colored character in this page can enter the page where detail information is written.
1. Specification & Circuit Concept
Index Product outline
Specification(General)
Specification(Mechanism) Trouble Shooting
Function
Inspection Procedure
Specification (Ultrasound) Block Diagram
Spare Parts
Error Messages
FAQ
Adjustments
2. Assembly / Disassembly
LCD & Upper Case
Lower Case
US-751 Table of Contents Contents
Page
1
Product Outline
1
4
Specification
3
5
Used material
7
6
General specification
8
7
Notes for operation and usage
9
8
Dimension and parts name
10
9
Overall Block Diagram
12
10
Block Diagram (+B line No1,card)
13
11
Block Diagram (+B line No2,card)
14
12
Block Diagram (+B line No4,card)
15
13
Block Diagram (Signal Path No1 card)
16
14
Block Diagram (Signal Path No5 card)
19
15
Spare parts
20
16
Trouble shooting
21
17
Error messages
25
18
Disassembly procedure
26
19
Ultrasound Power Measurement
35
US-751 Product Outline 1. Outline US-751 is a dual frequency ultrasound therapy device designed for overseas use only. The standard probe is the L-type ultrasound probe which delivers the two frequency outputs of 1MHz and 3MHz. Gel and the L-type probe are also supplied as the standard accessories and the the S-type probe and the EST cable are provided as the option. 2. Purpose of use / expected efficacy or effect. Microscopic massage, pain relief of muscles and joint.
US-751 is composed of 8 PCBs as shown in the below table. The contingent measure for failure is basically changing the defective PCB and module. REF 1 2 3 4 5 6 7 8 9
PCB name No.1 PCB (DCN) No.2 PCB (SW) No.3 PCB (Encoder) No.4 PCB (US) No.5 PCB (US probe (L)) No.5 PCB (US probe (S)) No.7 PCB (AC line filter) No.8 PCB (EMC filter) No.9 PCB (AC filter)
Description Controls LCD display and touch panel. Controls ultrasound output Encoder mounted board US output board Main board inside L probe Main board inside S probe Power filter US output filter board Power filter board
1
US-751 Specification (1) No. 1 2 3 4 5 6 7 8 9 10
Main unit and accessories Name Classification Quantity Main unit Standard 1 Power cord Standard 1 Ultrasound probe (L) Standard 1 Ultrasound probe (S) Optional 1 Probe holder Standard 1 Therapeutic gel Standard 1 E.S.T cable (Type A L) Optional 1 E.S.T cable (Type B S) Optional 1 User manual Standard 1 ferrite core Standard 1
Remarks New specification Type differs by country
New specification
Electric specification Power supply voltage
AC100 - 240v +/-10%
Power supply frequency
50 / 60 Hz
Power input
Less than 85vA +10%
Type of medical equipment
Class II , MDD Class IIa IEC, JIS category Class I BF
Display
7 inch color LCD with touch panel Open error: Generated when probe is disconnected during output.
Error function
Probe error: Generated when abnormal connection of probe is detected. System error: Generated when the main CPU found abnormal communication with boards.
2
US-751 Specification (1) Two types of frequency of 1 MHz and 3 MHz are provided. ・Ultrasound output: The maximum frequency according to sales regions is shown below. The values in ( ) are in pulse mode.
IEC FDA
Ultrasound probe (L) 1 MHz 3MHz 10.0W 10.0W (15.0W) (15.0W) 11.0W 12.0W (16.5W) (18.0W)
Ultrasound probe (S) 1 MHz 3MHz 1.4W 1.0W (2.1W) (1.5W) 1.8W 1.8W (2.7W) (2.7W)
(2.0W/cm2 MAX in CW mode) *Deviation: within +/-20% when output is above 0.2W/cm2 (3.0W/cm2 peak-peak MAX in Pulse mode) *Deviation: within +/-20% when output is above 0.3W/cm2 The output adjustment is available through encoder and menu by 0.05W/cm2 step. ・Vibration frequency:
a) Ultrasound probe (L) 1MHz +/-10% 3MHz +/-10% b) Ultrasound probe (S) 1MHz +/-10%
3MHz +/-10%
・Transducer: Installed inside of the probe head. The ultrasound probe (L) and (S) is selected according to the parts of therapy. In addition, the auto contact function enables output, only when the probe head is in contact with tissue. When output, the LCD window, the front panel LED and the probe LED indicates the output. ・ERA/BNR: ERA, BNR and emission characteristic are shown below. ERA tolerance is +/-20% and BNR tolerance is +/-30%.
ERA (cm2)
IEC FDA IEC BNR FDA Emission characteristic
1MHz 5.0 5.5 3.0 2.9
Probe (L)
3MHz 5.0 6.0 2.4 2.9
1MHz 0.7 0.9 2.9 2.9
Probe (L)
COLLIMATED COLLIMATED COLLIMATED
3MHz 0.5 0.9 2.4 2.9
COLLIMATED
・Timer: 1-30 minutes (set by minutes step. Deviation: 30 minutes +/-1 minutes) *If timer setting is changed during operation, counting will restart from the beginning.
3
05W/cm2 step.
L) and (S) is
4
US-751 Specification (2) ・Pulse frequency: Duty cycle at 16Hz/48H/100Hz Duty 16Hz 48Hz 100Hz
100% Continuous Continuous Continuous
50% 31ms 10ms 5ms
40% 25ms 8ms 4ms
30% 20% 19ms 13ma 6ms 4ms 3ms 2ms * Deviation +/-5%
10% 6ms 2ms 1ms
5% 3ms 1ms 0.5ms
・Output stability: +/-20% ・Other functions A) Auto contact function During ultrasound therapy, the impedance change of the transducer is used to judge the contact/no contact (or in water/in air) of the probe head with the tissue. In the contact (in water) state, the probe performs output as is set. In the no contact (in air) state, output goes into standby mode and weakens the output slightly. When output, the LCD window, the front panel LED and the probe LED indicate output. . Also, therapy stops if output standby status lasts for 3 minutes. 2.4 Output waveform
・Vibration frequency: 1.9MHz / 3.0MHz +/-10% ・Pulse frequency:
16Hz / 48Hz / 100Hz +/-5%
・Duty:
5, 10, 20, 30, 40, 50, 100% Variable in +/-5%
2.5 Envirnmental condition ・Usage environment: 10℃ to 40℃、30% to 70%, 700 to 1060hPa ・Storage environment: -10℃ to 60℃、30% to 95%, 800 to 1060hPa ・Transport environment: -10℃ to 60℃、30% to 95%, 800 to 1060hPa
5
US-751 Specification (3)
ITEM
SPECIFICATION
Dimension
90mm(W) x 23mm(D) x 93mm(H)
(Including leg: 96mm)
Weight
About 3.0kg
Intensity
In accordance with the design standard of ultrasound therapy device for business purposes "84E-0019" Device ……… 5years
Mechanical life
Accessary……3 years (except consumable items such as therapeutic gel) Consumable.. therapeutic gel
6
US-751 Used material 1. Raw material (1) Main body Name Case LCD Output control VR Stop switch Ultrasound probe socket Power switch AC power cable inlet E.S.T cable socket LED AC power filter
Used material Synthetic resin General electric parts General electric parts General electric parts General electric parts General electric parts General electric parts General electric parts General electric parts Synthetic resin
(2) Power cable
Synthetic resin
Ref 1 2 3 4 5 6 7
(3) Ultrasound probe Name Transducer Case Cable Plug Code protector Output indicator Core filter
Used material Titanium alloy or Aluminum alloy Synthetic resin Synthetic resin Synthetic resin or elastomer Polyamide resin General electric parts Ferrite, Nylon cover
Ref 1 2
(4) Probe holder Name Probe holder Lid
Used material Synthetic resin or Silicon Synthetic resin or Silicon
(5) Treatment gel
Aquasonic 100 gel 250ml
Ref 1 2 3 4 5 6
(6) E.S.T cable (Type A: Large) Name Output connector Code Core filter Core filter Rectangular connector Φ2 plug
Used material General electric parts Synthetic resin Ferrite, Nylon cover Ferrite, Nylon cover General electric parts Brass Nickel plating pin
Ref 1 2 3 4 5
(7) E.S.T cable (Type B: Small) Name Output connector Code Core filter Rectangular connector Φ2 plug
Used material General electric parts Synthetic resin Ferrite, Nylon cover General electric parts Brass Nickel plating pin
(8) Ferrite core
General electric parts
Ref. 1 2 3 4 5 6 7 8 9 10
7
US-751 General information Operation theory US-751 is a ultrasound pain relief device utilizing thermal and non thermal physiologic reaction of the tissue which is caused by the delivered ultrasound of the device. US-751 generates the two frequency of 1MHz and 3MHz of which their level are controlled by the level controller circuit and drives the transducer to inject the ultrasound vibration to treat the affected tissue. The power of each circuit are supplied from commercial AC which is converted to DC in the power regulator. CPU controls output level, timer, error detection, display control, and display functions. LCD and LED display the status of manipulation of touch panel, switches, and encoder. User can select the ultrasound probe from S-type or L-type according to the treatment part. Function of each part 1. Main unit (1) Cover Cover is made of synthetic resin which is insulator. (2) Power supply Converts commercial AC to DC and supply it to the circuit. (3) Display control Displays the status of manipulated switch and touch panel. Communicates with the CPU in the ultrasound probe (L & S) to control the ultrasound output level, display error, and identify probe type. Generates 1MHz or 3MHz, drives the output power circuit to deliver the ultrasound to the probe. Detects the phase to control the frequency of the ultrasound while detecting voltage and current of the ultrasound to control the output power source. Communicates with the display controller and the CPU of the ultrasound probe to exchange error information and others. The output VR increases or decreases the ultrasound output. (4) Ultrasound probe (L) / (S) Converts the ultrasound signal to the ultrasound vibration for treatment which is irradiated to the treatment part.
8
US-751 Notes of Operation and Usage 1. Usage (1) Parameter setting. LCD touch panel is used to select, change, or set the output parameters. (2) Intensity setting of the LCD LCD touch panel is used to set the LCD intensity. (3) Starting output Rotation of the output controller VR or manipulation of the LCD touch panel starts output. LED of the ultrasound probe (L,S) will light up while the probe head touches on the skin properly. (4) Stop of the output Pressing the STOP switch stops the output. 2. Low frequency combination treatment (1) US-751 is able to perform the combination treatment with low frequency treatment equipment and low frequency treatment/interference current type low frequency combination physio therapy equipment. For the use of combination treatment, connect the E.S.T connector of US-751 to low frequency treatment equipment with the dedicated E.S.T cable. Connect low frequency electrode pad to the ƒ³2 plug, connects the square connector to the E.S.T cable outlet of US-751 and connect the output connector to the low frequency therapy equipment. Note: The low frequency electrode pad is electrically plus (+) and the ultrasound probe is minus (-). Please maintain the low frequency electrode pad distance from the ultrasound probe head. 3. After use (1) Clean the probe head and keep it in a dry place.
9
US-751 Dimension and Parts name (1)
Ref. No.
Name
1
Main body case
2
LCD with touch panel
3
Output control VR
4
Stop button
5
Ultrasound probe socket
6
Power SW
7
AC power cable inlet
8
EST cable inlet
9
LED
10
Power filter
10
US-751 Dimension and Parts name (2) (1) AC power cable No.
mm
Name
1
Plug
2
Code
3
Connector
(2) Ultrasound probe (L) No.
Name
1
Transducer
2
Case
3
Cable
4
Plug
5
Code protector
6
Output indicator
7
Core filter
mm
(3) Ultrasound probe (S) No.
mm
11
Name
1
Transducer
2
Case
3
Cable
4
Plug
5
Code protector
6
Output indicator
7
Core filter
US-751 Dimension and Parts name (3) (4) E.S.T. cable TypeA (Large) mm
(2) Ultrasound probe (L)
No.
Name
1
Output connector
2
Code
3
Core filter
4
Core filter
5
Rectangular connector
6
Φ2 plug
(5) E.S.T. cable TypeB (Small) mm
No.
Name
1
Output connector
2
Code
3
Core filter
4
Rectangular connector
5
Φ2 plug
12
US-751 Overall Block Diagram
No.9 board
No.6 board
US-751 Block Diagram (+B Line No.1 card)
CN3 1 2
TP2: +3V
TP36: +5V
TP25: +15V
U1 DC Reg
U6 DC Reg
Touch panel TP26: +3.3VPLL
TP38: +3.8V
U10 DC Reg
LCD Backlight TP28: +3.3V
LCD Controller LRC_VccPLL LCD Controller ULC_VccPLL TP29: +2.5V
U8 DC Reg
U7 DC Reg
LCD Controller Vcc
TP42: +1.2V U11 DC Reg DIPSW U2 Vcc LCD Controller Vcc LCD Controller IO Vcc LCD Controller Aux Vcc U9: RES LCD
14
LCD Controller Vcc
US-751 Block Diagram (+B Line No.2 Card)
from Power Supply
CN1 1 2 3 4
TP51: +3.3V
TP54: +5.5V
TP52: +15V
U5 DC / DC
U4 DC / DC
Inv. +3.3v
TP55: +3.3V U6 DC / DC CN7
1 2
U3 SP Drive
Sound SW
No.4 Card (US unit) No.1 Card (DCN)
15
DC Fan Encoder 1, 2 EST STOP SW CPU OSC CPU TX CPU RX CPU Vdd
US-751 Block Diagram (+B Line No.4Card) TP61
from No.2
1 2 3 4
TP62: +15v
TP59: +5.7v U33 DC / DC
TP55: -5.7v U36 DC / DC
U35 DC / DC
TP56: -5v U34 DC / DC
Pulse shaping U4,5,10,11
DC Fan TP57: 3.3v
TP60: +5v FB SW Reg U36 DC / DC Under voltage
+V
Pulse Shaping: U4, U5, U10,
U32
CPU
Phase Cont.: U1, U2, U3, U6, U9, U12, U13,
Pulse Gen.: U1, U4, U5, U7, U10, U11 RY1,2,3,4 Coil I DET: Q27,, Q29, Q30
CPU CPU Vref
16
US-751 Block Diagram (Signal Path No.1 Card)
+3.3v
+3.3v
U4 RAM U2 Flash ROM +3.0v
+3.3v PLL
Character Memory
TP3 X1
+2.5v
TP5: Clock U3 LCD Controller
+3.3v
TP4 Touc h Panel
+3.3v
+3.3v TP5
U5 CPU
TP6
1 2 3 4
+3.3v
U9
+3.3v
LCD
14 15 17 18
17
US-751 Block Diagram (Signal Path No.1 Card) +3.3v
TP4
STOP
TP5
1
CMOMBI_DE
2
TP19
+15v LED2 TR1:B TR2:R
COMB U1 CPU
+15v
LED1
TR3:G
+3.3v
3
ES-520 IN
TP20 TP24
4
ES-320 IN
TP81
5
SU520SW-A SU520SW-B
EST
U2 U3 Audio
U7 RESET
U8
18
US-751 Block Diagram (Signal Path No.1 Card)
Switching Regulator Power Source I DATA Q20, Q28,vQ25, Q28, U27 Q29, Q30 Switching Regulator
V Data
CPU
U26, U28, U29, U21, U31 1MHz /
1MHz, 3MHz OSC
Pulse Amp
U16,U19, U39, U17 U18, U14, U15, U6,
U8, U4, U5, U10, U11, U7
P DET U22, U20 U12, U13, U9
19
RY4 Frequency Select 1MHz /
RY1 Chanel Select CH1/ CH2
US-751 Block Diagram (Signal Path No.5 Card) J From No.4 card
1
+15V
2
COM_PR CN_DET
TP7:+15v
TP2
3
TP10:+11v
TP11:+3.3v TP5: 3M SEL
U2 V Reg
D6
Green LED Red LED Blue LED
TP3
4
Q1
Drive+ 5
Drive-
Q2
6 Q3
7 8
D1
TP1
D3
9 L probe
10
TP4 TP15
S probe
20
U1 CPU
Q5
US-751 Spare Parts
Name
No.
Part No.
1
No.1 Card (DNC)
210687
2
No.2 Card (SW)
210668
3
No.4 Card (US)
210670
4
No.9 Card(AC filter PCB)
210688
5
Power unit
080601
6
Upper case cover
200755
7
LCD
070280
8
LCD cushion 2
9
Power switch
140154
10
Motor fan
091534
11
Speaker harness sub-assy
091546
12
Harness 10 (CH1 output connector assy)
091565
13
Harness 13 (CH2 output connector assy)
091567
220506
21
US-751 Troubleshooting (No Power) No Power (1)
Step 1
Check +15 at TP52 of No.2 card.
+15v exists.
No
OK Step 2
DC motor rotates.
Step 3
Step 4
OK Check TP25 of No.1 card.
+15v exists.
Step 5
Change No.9 card or change FU1 and FU2.
Check DC fan drive
No
Change No.2 card.
No
OK Check TP-28 of No.1 card.
Check harness connection between No.2 and No.1 card.
OK +3.3v exists.
No
OK Step 6
Change No.1 ard.
Check TP-28 of No.1 card.
No +1.2v exists. OK Change LCD
22
US-751 Troubleshooting
LCD does not light
Step 1
Check TP25 of No.1 card.
No
+15v exists.
Step 2
OK Check TP-28 of No.1 card.
Check harness connection between No.2 and No.1 card.
OK +3.3v exists.
No
OK Step 3
Change No.1 ard.
Check TP-28 of No.1 card.
No +1.2v exists. OK Change LCD
23
US-751 (No output) No output Change Probe with new one.
Output is relived.
Yes
No
Change probe.
Check TP-61
+15v exists.
No
Ye
Check harness connection between No.4 and No.2 card.
Check TP12 or TP22.
+15v exists No
Yes
Change probe
Change No.4 card.
24
US-751
Auto contact does Confirm that output is normal.
Output is normal. Yes
No
Go to Troubleshooting (No Output)
Change the probe..
25
US-751 Error messages
Open error. Indicates that probe was disconnected from the output terminal during treatment.
Probe error. Indicates that the connected probe has problem. Probe change is required.
System error. Indicates 15v DET abnormality, fan rotation abnormality or circuit misalignment of control board. Indicates abnormality of main fan rotation. Indicates communication error between DCN and US board or between DCN and SW board. 26
US-751 Disassembly procedure (1) Disassembling Procedure Separate the upper case and the lower case
Remove the five screws from the bottom case to separate the upper case and the lower case.
No.2 card
27
US-751 (Lower case)
Disassembly of the Lower Case (1)
EMC insulation
DC Fan Speaker
No.8 (EMC) card
No.4 (US) card
Power supply unit US Fan
No.2 card
28
US-751 (Disassembly No2 & No.8 card) Disassembly of the Lower Case (2) No.2 Card
NC1
NC10 NC2
NC12 NC3
NC7
Remove the six screws, CN1,2,3, 7,9,10,
No.8 Card
CN3
NC9
CN4
CN1 CN2
Remove the four screws , disconnect CN1,2,3,4 to take out the No.8 (EMC) card.
29
US-751 (Disassembly No.4 card) Disassembly of the Lower Case (3) No.4 Card
Remove the harness connecting the No.2 and the No.4 cards.
Remove the four screws to take out the No.4 card.
Disconnect the harnesses from CN2 & CN12 to take out the upper shield case.
Remove the four screws.
Remove the two screws to take out the No.4 card from the shield case.
30
US-750 (Disassembly Power supply unit) Disassembly of the Lower Case (4) Power supply unit
Remove the four screws , disconnect TB1 and TB2 to take out the power supply unit.
Grounding point of the power supply unit.
31
US-751 (Disassembly upper case 1) Disassembly of the Upper Case (1)
Remove the screws.
Remove the screws.
Unlock the fixing
32
US-751 (Disassembly upper case 2) Disassembly of the Upper Case (2)
Remove the screw, pull out the Pull up the upper case.
Disconnect the harnesses, pull out the No.1 card plate.
Peel off the connector by using a cutter.
33
US-751 (Disassembly upper case 3) Disassembly of the Upper Case (3)
Past a new connector on the new LCD board.
Install the new LCD card on the upper case.
34
Ultrasound Power Measurement
Probe holder (Differs by probe)
Probe mounting plate (Differs by probe) Jigs Provided by Ito: Probe holding jigs (Taisei Technical) The tank is filled with de-aired water. Simple method of making de-aired water. (1) Boil tap water for 15 minutes. (2) Cool the water down in the same container used for boiling. (3) Keep the water cool until it becomes room temperature. (4) Pour the water into the tank slowly to avoid catching the air in the water.
Soaks the probe head
Ultrasound power meter Ultrasound Power Meter : Ohmic Instruments Co. ・UPM-DT-1AV (Locally purchase)
35