SP-8800 SYRINGE PUMP
SERVICE MANUAL
CONTENT
1.
STRUCTURE
--------------------------------------
2
1-1.
Block Diagram
-----------------------------------------
2
DISASSEMBLY AND ASSEMBLY Replace Main Board / LED Board Replace Power Board Replace a Sub Board Replace a Handle Replace a Slide Replace a Clamp Replace a Battery Board Replace a Battery Handle Lever Replace Key pad
--------------------------------------
3~12
-----------------------------------------
3
-----------------------------------------
--------------------------------------
4 5
-----------------------------------------
6
-----------------------------------------
7
-----------------------------------------
8
-----------------------------------------
9
-----------------------------------------
10
-----------------------------------------
11
-----------------------------------------
12
Trouble Shooting
--------------------------------------
13~17
3-1.
Power on malfunction
-----------------------------------------
13
3-2.
“Slider Hook Open LED” is on.
-----------------------------------------
14
3-3.
“Clutch Open LED” is on.
-----------------------------------------
15
3-4.
“Clamp Open LED” is on.
-----------------------------------------
16
3-5.
“Occlusion Alarm LED” is on.
-----------------------------------------
17
3-6.
“ERROR” message with Error code
-----------------------------------------
18~19
3-7.
TROUBLE SHOOTING Table
-----------------------------------------
20~22
2. 2-1. 2-2. 2-3. 2-4. 2-5. 2-6. 2-7 2-8. 2-9.
2-10 3.
SPECIFICATION
23
*DO NOT EXPOSE THIS MANUAL TO OTHER PARTY WITHOUT PERMISSION FROM MANUFACTURER, AMPALL. 1
SP-8800 / SME-03-REV.0-08B27
1. STRUCTURE
1-1. Block Diagram
2
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY
2-1. Replace Main Board
1
6 Nuts(3mm)
2
fox(5mm)
3
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY
2-2. Replace Power Board
1
2
3
4
5
1. 2. 3. 4.
Detach a bottom case. Remove nuts on upper case and bottom case. Remove a cable tie fixing on a trance. Detach a Power Board.
4
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-3. Replace a Sub Board
1
2
3
4
5
6
1. 2. 3. 4.
Detach a bottom case. Remove bolts on upper case and bottom case. Detach a cable connecting Sub Board. Detach a Sub Board.
5
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-4. Replace a Handle
1
2
3
4
Cable
Cable Tie
5
1. 2. 3. 4.
Remove a bolt on Handle Out Cover and detach a Handle Cover. Remove a cable tie and diconnect cable. Hit a handle part and change handle part. Assembley Out Cover.
6
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-5. Replace a Slide 1
2
3
4
5
6
7
8
9
1. 2. 3. 4. 5. 6.
Detach a bottom case. Remove bolts on upper case and bottom case. Remove nuts on Slide Rod. Detach a cable on Motor Sensor Board. Remove bolts on Slide. Detach a slide.
7
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY
2-6. Replace a Clamp
1
2
3
4
5
6
7
Detach after remove a Snap Ring 1. 2. 3. 4. 5.
Detach a bottom case. Remove bolts on upper case and bottom case. Remove a bolts on Slide. Detach a Slide. Detach a Snap Ring fixing on a Clamp with pincette.
8
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-7. Replace a Battery Board 1
2
3
4
5
6
1. 2. 3. 4.
Detach a bottom case. Remove bolts on upper case and bottom case. Detach a cable connecting to Board. Detach a Battery Board.
9
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-8. Replace a Battery
1
2
3
Connected by connector cable
4
1. Detach a Battery Cover. 2. Detach a Batery cable. 3. Detach a Battery.
10
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-9. Handle Lever
1
2
3
1. Remove bolt on Handle, and detach a Handle Out Cover. 2. Change a Handle Lever. 3. Assembly Out Cover.
11
SP-8800 / SME-03-REV.0-08B27
2. DISASSEMBLY & ASSEMBLY 2-10. Replace Key pad
1
2
3
4
5
6
7
8
9
1. 2. 3. 4.
Detach a bottom case. Remove bolts on upper case and bottom case. Disconnect Key button cable connecting to Main Board. Detach a Key pad.
12
SP-8800 / SME-03-REV.0-08B27
3. TROUBLE SHOOTING 3-1. Power on malfunction
1)
Is the cord properly plugged in?
No
Plug the supplied cord and check the supplied voltage.
No
Change the power switch with new one.
No
Change anther.
No
Change the power board with new one.
No
Contact our A/S team.
Yes 2)
Is the contact of the power switch good?
Yes 3)
Is the contact of the power button keypad good?
the
keypad
with
Yes 4)
Is the input and output of the power board normal?
Yes
5)
Replace the main board and test. Does it work normal?
13
SP-8800 / SME-03-REV.0-08B27
3. TROUBLE SHOOTING 3-2. “Slider Hook Open”
1)
Check if you have properly inserted the slit to the grip, the plunger to the clutch
No
and the body to inside the clamp.
Insert the parts again and check visually if they have been properly inserted.
Yes 2)
Check if the plunger is correctly
No
Don’t press the handle lever by hand, but fit it in a normal way.
inserted into the clutch and the handle lever is pressed down to the bottom.
Yes 3)
When you press the handle lever, do you feel something rough or blocking?
No
Open the handle assembly and check if there are no impurities inside.
No
Contact our A/S team.
Yes
4)
Replace the handle assembly and test. Does it work normal?
14
SP-8800 / SME-03-REV.0-08B27
3. TROUBLE SHOOTING 3-3. “Clutch Open”
1)
Check if you have correctly inserted the syringe inside the clamp and the
No
plunger to the clutch.
Insert the part again and check visually if it has been properly inserted. (Fig. 2-1)
Yes 2)
Check if the plunger is correctly inserted into the clutch and the handle
No
lever is lowered to the bottom.
Don’t press the handle lever by hand, but fit it in a normal way. (Fig 2-2)
Yes 3)
When you press the handle lever, do you feel something rough or blocking?
No
Does the slide shake much up and down, when you inject medicines?
No
Open the handle assembly and check if there are no impurities inside. (Fig. 2-3)
Yes 4)
Check if the slide fixing bolts have been tightened up. (Fig 2-3)
Yes
5)
Replace the handle assembly and test.
No
Contact our A/S team.
15
SP-8800 / SME-03-REV.0-08B27
Does it work normal?
3. TROUBLE SHOOTING 3-4. “Clamp Open” “Red alarm lamp lights”
1)
No
Insert the part again and check visually if it has been properly inserted. (Fig. 2-1)
No
Choose the maker of the inserted syringe from User’s Menu and test.
Check if you have correctly inserted the syringe inside the clamp and the plunger to the clutch.
Yes 2)
Is the maker of the inserted syringe the same maker as shown on the left upper of LCD display? Yes
3)
Does the value of SIZADC change at the Factory Mode No. 8?
Check if the value changes from 24.0
No
before inserting the syringe to 75.0 when the clamp slides back. If not, contact our A/S team.
Yes 4)
Set the external diameter of the syringe again and test. Does it work normal?
Find the syringe maker at Factory
No
Mode No. 5 and set the external diameter of the syringe again.
Yes
5)
Replace the main board and test. Does it work normal?
No
16
Contact our A/S team.
SP-8800 / SME-03-REV.0-08B27
3. TROUBLE SHOOTING 3-5. “Occlusion alarm ”
1)
Check if the injecting set is twisted.
No
Twisted injecting set may cause blocking. Be sure not to twist the injecting set while inserting.
Yes
2)
When you push the plunger by hand
If it is hard to push and pull the
No
plunger by hand, you should not
before mounting the syringe, do
mount the syringe to the device.
you feel it stiff?
(Medicines with high viscosity can’t be used in this device.)
Yes 3)
Check if there are no impurities stuck to the piston stopper (silver
Impurities stuck to the piston
No
stopper (silver disc) may raise the value of OCCADC.
disc) that pushes the syringe.
Yes
4)
Check if the value of OCCADC at the factory mode is normal.
The value should not change from the
No
highest to the lowest. When the value varies excessively, contact our A/S team.
Yes
5)
When you push the piston stopper (projected part of the silver disc) by
Replace the handle assembly and test
No
the working. If the trouble remains unsolved, contact our A/S team.
hand, does the value rise?
17
SP-8800 / SME-03-REV.0-08B27
3. TROUBLE SHOOTING 3-6. “ERROR” message with error code come up during injection
1) Type of Errors Error Type
Relation
Code of error part (A)
Code of error type (B)
Description
Communication Errors
MainSlave
01
001~005
Communication-concerned errors
Memory Errors
MainMemory
03
001
Read Error
002
Write Error
Reset Errors
MainReset
04
002~005
Reset-concerned errors
001
Abnormality at the motor encoder pulse (When the motor does not work or when it does not coincide with injection quantity.)
Motor Errors
SlaveMotor
11
2) How to cope with “ 01 001 ~ 01 005 ” errors Have you changed the main board?
No
Change the main board and test the working, as defective main board may cause errors.
Yes Have you changed the slave board?
No
Change the slave board and test the working, as defective slave board may cause errors.
Yes Contact our A/S team.
18
SP-8800 / SME-03-REV.0-08B27
“ERROR” message with error code come up during injection. 3) How to cope with “ 03 001 ~ 03 002 “ and “ 04 002 ~ 04 005 “ errors Have you changed the main board and tested the working?
No
Change the main board and test the working.
Yes Contact our A/S team.
4) How to cope with “ 11 001 “ errors Have you changed bracket assembly?
the
slide
No
Change the slide bracket assembly and test the working, as defective slide bracket assembly may cause errors.
Have you changed the slave board?
No
Change the slave board and test the working, as defective slave board may cause errors.
Yes
Yes Contact our A/S team.
19
SP-8800 / SME-03-REV.0-08B27
Trouble Shooting Take the following actions if any trouble occurs. When the troubles could not be solved with the following actions, Please contact your local authorized dealer immediately. NOTE
• Whenever
alarm sounds, the pump stops infusion. That is, alarm sounds only in error situation during infusion.
Symptom Pump
Cause
cannot
be
switched on.
Action
• AC or DC power cord is not
▶Check the AC or DC power cord connection.
inserted properly
• Never connect both AC and DC power to the pump at the same time
• Internal battery has deteriorated
▶Stop the operation of the pump and replace with a new battery through your local authorized dealer.
• The voltage of the internal battery is low [OCCLUSION] flashes
and
LED alarm
▶Recharge the battery fully for more than 24 hours by connecting the pump to an AC power outlet.
• The infusion line is kinked or
▶1. Turn the alarm off by pressing [SILENCE] button. ▶2. Check the infusion line and take a corrective action like untwisting or
twisted
sounds.
replacing with a new one to solve the problem of occlusion ▶3. Press [STOP] button to show “STAND-BY” on the LCD. ▶4. Restart infusion by pressing [START] button. • When infusion is restarted, the
[FINISH] LCD display
▶The system is in trouble.
alarm sounds again, and the
▶Turn the alarm off by pressing [SILENCE] button.
pump stops
▶Please contact your local authorized dealer.
• The syringe is completely empty
▶1. Turn the alarm off by pressing [SILENCE] button. ▶2. To continue the infusion, replace the syringe with a new one.
and alarm sounds
▶3. Press [STOP] button to show “STAND-BY” on the LCD. ▶4. Restart infusion by pressing [START] button. [NEAR EMPTY] LED
• The syringe is nearly empty
▶1. Press [STOP] button to stop infusion.
flashes
(Finishing injection in 5 minutes
▶2. To continue the infusion, replace with the new syringe filled with
sounds.
and
alarm
before,
the
flash
light
warning sound occurring)
and
solution. ▶3. Restart infusion by pressing [START] button.
20
SP-8800 / SME-03-REV.0-08B27
Trouble Shooting
Symptom
Cause
[SLIDER HOOK] LED
• Slider Hook does not fix Push-
flashes
and
alarm
Button of syringe
Action ▶1. Turn the alarm off by pressing [SILENCE] button. ▶2. Make Slider Hook fix Push-Button of syringe. ▶3. Press [STOP] button to show “STAND-BY” on the LCD.
sounds
▶4. If re-infusion is needed, restart infusion by pressing [START] button. [CLAMP] LED flashes
• Clamp does not fix Barrel of syringe
▶1. Turn the alarm off by pressing [SILENCE] button. ▶2. Pull the clamp, turn it to the left up to 90° and put the finger grips of
and alarm sounds
syringe into the slit. ▶3. Press [STOP] button to show “STAND-BY” on the LCD. ▶4. If re-infusion is needed, restart infusion by pressing [START] button. [CLUTCH] LED flashes
• Clutch is not properly engaged
▶1. Turn the alarm off by pressing [SILENCE] button. ▶2. Make the clutch engaged properly.
and alarm sounds
▶3. Press [STOP] button to show “STAND-BY” on the LCD. ▶4. If re-infusion is needed, restart infusion by pressing [START] button. “BATT-LOW”
on
the
• The voltage of internal battery is low
▶1. Press [STOP] button to stop infusion. ▶2. Recharge the battery fully for 24 hours by connecting the pump to
LCD and alarm sounds
an AC power outlet. ▶3. Restart infusion by pressing [START] button. “KEEP-VEIN”
on
the
LCD and alarm sounds
• In the operation, the total volume
▶ Stop K.V.O. rate by pressing [STOP] button
delivered reaches the delivery limit • The total volume delivered exceeds the delivery limit
▶ Stop K.V.O. rate by pressing [STOP] button ▶ When infusion is newly needed, set the delivery volume again or clear the total volume delivered.
When [PURGE] button
• The syringe is not set properly
▶ Set the syringe correctly.
is pressed, the alarm
• Even when the syringe is set again,
▶ The system is in trouble.
sounds. “ERROR” is on the LCD
purging does not occur
Please contact your local authorized dealer
• The system is in trouble
▶ Please contact your local authorized dealer
21
SP-8800 / SME-03-REV.0-08B27
Trouble Shooting Symptom
Cause
[START]
• The value of “flow rate, delivery
button is pressed, the
limit, dose rate, body weight, drug
alarm sounds, and the
volume, or solution volume” is not
pump does not start.
set correctly
Even
when
Action ▶ Confirm the values that were set.
• Clutch is not properly engaged
▶ Engage the clutch correctly.
• The syringe is not set properly
▶ Set the syringe correctly.
• The low voltage of the internal
▶ Recharge the battery fully for more than 24 hours by connecting the
battery
pump to an AC power outlet. ▶ Make sure that the syringe, which is used, is the same as the one
[SYRINGE SIZE] LED
• The syringe, which is used, may be
and [CLAMP] LED flash
different from the one specified in
specified in the pump in terms of manufacturer.
the pump in terms of manufacturer
To use the syringe of different manufacturer, please contact your
• When infusion is restarted, the
local authorized dealer.
alarm sounds again, and the pump
▶ The system is in trouble.
stops
▶1. Turn the alarm off by pressing [SILENCE] button. ▶2. Turn off the power. ▶3. Please contact your local authorized dealer for repair
The accuracy of flow
• The syringe, which is used, may be
rate is questionable.
▶ Make sure that the syringe, which is used, is the same as the one
different from the one specified in
specified in the pump in terms of manufacturer.
the pump in terms of manufacturer
To use the syringe of different manufacturer, please contact your local authorized dealer.
• Otherwise, the system is suspected
▶ Please contact your local authorized dealer for repair.
to be in trouble The operation hour by
•
Internal battery has deteriorated
through your local authorized dealer.
battery is too short after • Otherwise, the system is suspected
recharging the battery
▶ Stop the operation of the pump and replace with a new battery
▶ Please contact your local authorized dealer for repair.
to be in trouble Battery is not recharged even
when
it
is
• AC power cord is not inserted
▶ Check the AC power cord connection.
properly
connected to AC power
• AC power cord has deteriorated
▶ Replace with a new AC power cord.
outlet.
• Otherwise, the system is suspected
▶ Please contact your local authorized dealer for repair.
to be in trouble Battery is not recharged even
when
it
is
• DC power cord is not inserted
▶ Check the DC power cord connection.
properly
connected to DC power
• DC power cord has deteriorated
▶ Replace with a new DC power cord.
outlet.
• Otherwise, the system is suspected
▶ Please contact your local authorized dealer for repair.
to be in trouble
22
SP-8800 / SME-03-REV.0-08B27
SPECIFICATIONS GENERAL SPECIFICATIONS OF SP-8800 SYRINGE PUMP
DELIVERY PERFORMANCE Syringe 10ml: 0~300ml/h Syringe 20ml: 0~400ml/h FLOW RATE RANGE Syringe 30ml: 0~500ml/h Syringe 50ml: 0~1200ml/h 0.1~999.9ml ( 0.1ml steps)
VOLUME LIMIT
Syringe 10ml: 300ml/h Syringe 20ml: 400ml/h PURGE RATE Syringe 30ml: 500ml/h Syringe 50ml: 1200ml/h 0.1ml/h
K.V.O. RATE
Mechanical accuracy: ±1% ACCURACY WITH APPROVED SYRINGES Accuracy including syringe: ±3%
ALARM OCCLUSION, OPEN,
NEAR EMPTY,
LOW BATTERY,
K.V.O. RATE,
CLUTCH OPEN,
SLIDER HOOK OPEN, CLAMP
REPEAT ALARM
FEATURES CLASSIFICATIONS
ClassⅠ, CF, IPX2
OCCLUSION DETECTION PRESSURE
1.3kgf /㎠
NURSE CALL
DC12V, 1A
TEMPORARY STOPPING,
CLEAR VOLUME DELIVERED,
SETTING DOSAGE MODE,
SETTING MENU
ALARM VOLUME / OCCLUSION PRESSURE CONTROL
OTHER PARAMETERS POWER REQUIREMENTS / CONSUMPTION
AC110 / 230V, 50 / 60Hz or DC9.6V, DC24V 1A
/
40VA
DIMENSIONS (W × D × H)
220×130×108 (mm)
WEIGHT
Approximately 2kg
BATTERY / OPERATION / CHARGING
Ni-Cd, Ni-MH9.6V/2 hours( 50mLsyringe, Flow rate 5mL/h) / 24 hours
BATTERY LIFE
1.5~2 years
OPERATION CONDITIONS
5~40℃, 30~90% RH (no condensation)
STORAGE CONDITIONS
-20~45℃, 10~95% RH (no condensation)
WARRANTY PERIOD
1 year
※Specifications and design are subject to change for improvement without prior notice
23
SP-8800 / SME-03-REV.0-08B27