41 1 11MB
Confidential and Proprietary
SM-30-4054 Part No. 1836-541, Revision A February 2010
Diagrams
Parts Removal / Replacement
=
Parts
Adjustment and Calibration
sÉêëáçå=RKR
Appendices
Troubleshooting Error codes
flij~ëíÉê=
pÉêîáÅÉ=j~åì~ä= Checkout and Preventive Maintenance
General Information
Table of Contents
Service Manual Part No. 1836-541
IOLMaster 5.5
Confidential and Proprietary
ii
SM-30-4054-en Revision A, Feb. 2010
IOLMaster 5.5
Table of Contents
PROPRIETARY NOTICE Information contained in this document is copyright Carl Zeiss Meditec AG and may not be reproduced in full or in part by any person without prior written approval of Carl Zeiss Meditec. Its purpose is to provide the User with adequately detailed information so as to repair, maintain, and order spare parts for the instrument supplied. Every effort has been made to keep the information contained in this document current and accurate as of the date of publication or revision. However, no guarantee is given or implied that the document is error-free or that it is accurate with regard to any specification.
All trademarks, registered trademarks, and trade names that appear in this manual are the property of their respective holders. Copyright © 2010 by Carl Zeiss Meditec AG All rights reserved. Carl Zeiss Meditec AG Goeschwitzer Strasse 51-52 07745 Jena GERMANY
SM-30-4054-en Revision A, Feb. 2010
Confidential and Proprietary
iii
Service Manual Part No. 1836-541
Table of Contents
IOLMaster 5.5
REVISION CONTROL LIST Document:
IOLMaster 5.5
Service Manual Part No.: Issued Date:
SM-30-4054 1836-541 February 2010
Listed at the bottom of each page is the part number of the manual, along with the Revision letter and date for that page (e.g. Part No. 1836-541, Revision A, Feb. 2010) and the document number (e.g. SM-30-4054). Subsequent revisions to a page will be noted by a corresponding change to the Revision letter and date.
Service Manual Part No. 1836-541
Confidential and Proprietary
iv
SM-30-4054-en Revision A, Feb. 2010
IOLMaster 5.5
Table of Contents
Table of Contents Section 1 – General Information
Section 2 – Checkout and Preventive Maintenance
Section 3 – Parts Removal / Replacement
Section 4 – Adjustment and Calibration
Section 5 – Troubleshooting / Error codes
Section 6 – Diagrams
Section 7 – Parts
Section 8 – Appendices
SM-30-4054-en Revision A, Feb. 2010
Confidential and Proprietary
v
Service Manual Part No. 1836-541
Section 1 – General Information
1.1 About This Manual.........................................................................................................3 1.1.1 General Information ........................................................................................................ 3 1.1.2 Conventions .................................................................................................................... 4 1.2 Service Bulletins .............................................................................................................4 1.3 IOLMaster Service Strategy ............................................................................................5 1.4 General Safety Instruction ..............................................................................................5 1.5 Disposal of the product within the EU............................................................................5
SM-30-4054-en Section1, Revision A, Feb. 2010
Confidential and Proprietary
1-1
Service Manual Part No. 1836-541
General Information
IOLMaster 5.5
Notes:
Service Manual Part No. 1836-541
Confidential and Proprietary
1-2
SM-30-4054-en Section 1, Revision A, Feb. 2010
IOLMaster 5.5
1.1
General Information
About This Manual
1.1.1 General Information This manual is the field service reference for troubleshooting, repair, adjustment, and calibration of the IOLMaster. The manual is intended for use by field service engineers who have completed Carl Zeiss Meditec service training on the IOLMaster. The service manual is designed to support Level 1 of a two-level service strategy. Level 1 (on-site) field service employs modular replacement of printed circuit boards and other assemblies that are most effectively repaired at a central repair facility. This is the service strategy used in field service, and presented in Carl Zeiss Meditec (CZM) training classes. Level 2 (workshop repair) include procedures which are not possible on-side or where on-side repair was not successful. See Section 1.3 for additional information regarding the IOLMaster service strategy. The procedures in this manual assume that the reader is familiar with operation of the instrument. Complete operating instructions are contained in the IOLMaster User Manual. Information presented in the user manual is not repeated in this service manual. The user manual can be ordered separately by standard Zeiss parts order. Update revisions to this service manual will be issued by Service Bulletin, as required. The general layout of the service manual is shown below. For greater detail, please refer to the Table of Contents. Level 1 Service Manual Layout Section 1 - General Information Section 2 - System Checkout and Preventive Maintenance Section 3 - Parts Removal / Replacement Section 4 - Adjustment and Calibration Section 5 - Troubleshooting Section 6 - Diagrams Section 7 - Spare Parts Section 8 - Appendices A: Service Bulletins B: Instrument Checklist
SM-30-4054-en Section 1, Revision A, Feb. 2010
Confidential and Proprietary
1-3
Service Manual Part No. 1836-541
General Information
IOLMaster 5.5
1.1.2 Conventions The following conventions apply in this manual:
Front and Back sides of the instrument is as viewed from the front (patient side) of the instrument, unless noted otherwise. All tool sizes for screws and nuts mentioned in the instructions are metric unless noted otherwise.
1.2 Service Bulletins Service Bulletins are a vital element of service support. Bulletins are used to quickly convey technical information on a variety of field service topics, including:
instrument design changes technical problems and corrections software updates
new troubleshooting procedures problem alerts
service manual revisions new calibration or adjustment procedures upgrade announcements/ procedures system checkout – checklist
service disclaimer form
Your service bulletins should be filed where easily accessible for quick reference.
NOTICE Service Bulletins are Confidential and Proprietary, for the sole use of personnel employed by Carl Zeiss Meditec, Carl Zeiss Meditec affiliates, and authorized Carl Zeiss Meditec distributors. Carl Zeiss Meditec has a well-deserved reputation for high quality, reliable instruments, unsurpassed in the industry. As a Carl Zeiss Meditec employee, affiliate, or distributor you are required to handle your service bulletins as appropriate for proprietary and confidential information.
Service Manual Part No. 1836-541
Confidential and Proprietary
1-4
SM-30-4054-en Section 1, Revision A, Feb. 2010
IOLMaster 5.5
General Information
1.3 IOLMaster Service Strategy The service strategy for IOLMaster envisages that field service must only be performed by appropriately trained technicians authorized by CZM AG. These activities are described in this service manual and are dealt with in detail during the service training for this product. Only original spare parts must be used. For maintenance and repair, proper ESD protection measures, the tools and devices specified in the tools list must be used.
1.4 General Safety Instruction If you need to gain access to the interior electronics of the IOLMaster, especially the computer base, always disconnect the instrument from any and all power sources first. For maintenance and repair, proper ESD protection measures, the tools and devices specified in the tools list must be used (see Section 5). Only original spare parts listed in Section 7 and the official CZM spare part price list must be used. Do not use or install software which is not approved and tested by CZM. Other procedures than described in this manual are not allowed to be performed by the field service engineer. After any kind of manipulation or changes perform a System Checkout in accordance to the instrument checklist and initiate a technical safety check according to Section 2, especially after the service of electronic parts directly connected to the main power, e.g. power supply, main switch and power entry module.
1.5 Disposal of the product within the EU EU directives and national regulations currently in force at the time of marketing prohibit the disposal of the product specified on the delivery note in domestic waste or by municipal waste disposal companies. For more information on disposal of the product, please contact your local dealer, the manufacturer or their legal successors. Please also take note of the manufacturer’s current online information. If the product or its components are resold, the seller has the duty to notify the buyer that the product must be disposed of in accordance with currently valid national regulations.
SM-30-4054-en Section 1, Revision A, Feb. 2010
Confidential and Proprietary
1-5
Service Manual Part No. 1836-541
Section 2 – Checkout and Preventive Maintenance 2.1 Checkout and Maintenance ............................................................................................3 2.2 Technical and Electrical Safety Check .............................................................................3 2.2.1 General Information ........................................................................................................ 3 2.2.2 Procedure of the technical safety check ........................................................................... 3 2.2.3 Reading points ................................................................................................................ 5
SM-30-4054-en Section 2, Revision A, Feb. 2010
Confidential and Proprietary
2-1
Service Manual Part No. 1836-541
Checkout and Preventive Maintenance
IOLMaster 5.5
Notes:
Service Manual Part No. 1836-541
Confidential and Proprietary
2-2
SM-30-4054-en Section 2, Revision A, Feb. 2010
IOLMaster 5.5
2.1
Checkout and Preventive Maintenance
Checkout and Maintenance
Procedure: Please follow the instructions of the Checkout and Maintenance form (see Section 8 Appendices) as accurate as possible. For any nonconformity please perform the necessary adjustments and alignments.
2.2
Technical and Electrical Safety Check
2.2.1 General Information The detailed description of the technical safety check procedure can be found in Service Bulletin 400. This Bulletin is for all Carl Zeiss Meditec AG medical products applicable and equates the latest international rules for electrical safety. Additionally, this service manual also describes the most important procedures inclusive the instrument-depending varieties, e.g. the reading points for the different tests. Note If you have national rules for electrical safety, SB 400 is in force only analogous. In this case you have to fulfil the values of your national standards (e.g. NFPA 99, AAMI ES-1). Only qualified technicians shall perform these tests. Qualification shall include training on the subject, knowledge, experience, and acquaintance with the relevant technologies and regulations. The technician shall be able to assess safety and shall be able to recognize possible consequences and hazards arising from non-conforming equipment.
2.2.2 Procedure of the technical safety check Complete the following points in line with the test procedures of the Instrument Checklist: • • •
Check that the housing (in particular the ventilation slots) is undamaged and free from contamination. Check all cables (in particular the insulation) for damage. Check that all notes and labels are complete and correctly positioned using the user manual.
Checking the electrical safety (This test must be performed as part of the start-up procedure, the annual maintenance service and technical safety check, or after repairs affecting safety-relevant components). During the technical safety check, the device must be tested for electrical safety. Perform the measurements in accordance with IEC 62353:2007 and DIN EN 62353:2008. − The measuring results must be documented in the technical safety check report (See SB 400)
SM-30-4054-en Section 1, Revision A, Feb. 2010
Confidential and Proprietary
2-3
Service Manual Part No. 1836-541
Checkout and Preventive Maintenance
IOLMaster 5.5
Procedure: 1. Measuring protective ground resistance • A maximum test voltage of 24V and a test current of 10A AC are recommended for measurement. • Connect the device with its connecting cable to the power socket of the safety tester (e.g. RIGEL 288). • Firmly press the test probe against a screw head on the housing of the IOLMaster (see Figure 2 – 1) • Start the measurement and note down the measured value. • If the IOLMaster is powered via the power socket of the IT table, repeat the measurement for the table. The same applies if the system is operated via a multiple socket. Include the multiple socket into the measurement in this case. • Note down the highest value measured. • The limit value that must not be exceeded is 0.3 Ohms for devices with a power cord and 0.5 Ohms for systems with a multiple socket. 2. Measuring the leakage current • We recommend using the differential current method, where the device is measured during operation and both phase directions are taken into account. The IOLMaster connected and operated via the safety tester. • When using the RIGEL 288 tester, select IEC 62353, differential measurement and semi automatic mode (to boot and shut down the computer manually during reversal of polarity). • Please note the risk of a computer damage using safety testers with fully automatic reversal of polarity! • Switch on the instrument at the main switch. • Firmly press the test probe against a screw head on the housing of the IOLMaster (see Figure 2 – 1 and Figure 2 – 2) and start measurement. The tester automatically performs a reversal of polarity. • Note down the highest value measured. • The limit value of 0.5 mA must not be exceeded in any of the measurements. 3. Measuring the insulation resistance • Perform this measurement at 500V DC. • Connect the power cord of the IOLMaster to the power socket of the safety tester. • Switch on the instrument at the main switch. • Start the measurement. • Note down the measured value. • The measured value must not be less than the limit value of 1 MOhm. • If the IOLMaster is operated via the power socket of the instrument table and/or via a multiple socket, repeat the measurements accordingly.
Service Manual Part No. 1836-541
Confidential and Proprietary
2-4
SM-30-4054-en Section 2, Revision A, Feb. 2010
IOLMaster 5.5
Checkout and Preventive Maintenance
2.2.3 Reading points According to Service Bulletin 400 following tests are necessary: • • •
Measuring of protective earth impedance Measuring of leakage currents Measuring of insulation resistance
For all measurements and tests during the electrical safety check please use the two points according to Figure 2 – 1 and Figure 2 – 2.
Fig. 2 - 1: Computer base
SM-30-4054-en Section 1, Revision A, Feb. 2010
Fig. 2 - 2: Head rest
Confidential and Proprietary
2-5
Service Manual Part No. 1836-541
Section 3 – Parts Removal / Replacement 3.1 Instrument head.............................................................................................................4 3.1.1 Opening the instrument head....................................................................................... 4 3.1.2 Head covers / spacers................................................................................................... 4 3.1.3 Complete head exchange ............................................................................................. 5 3.1.4 Control cable ............................................................................................................... 6 3.1.5 Display complete.......................................................................................................... 7 3.1.6 TFT Display and LED-Converter ..................................................................................... 8 3.1.7 Keratometer complete.................................................................................................. 8 3.1.8 Slit module................................................................................................................... 9 3.1.9 CCD Camera ................................................................................................................ 9 3.1.10 Digital IOL PCB ........................................................................................................... 14 3.1.11 APD PCB .................................................................................................................... 15 3.1.12 Aperture / diaphragm module .................................................................................... 15 3.1.13 Aperture / diaphragm (Sensor).................................................................................... 15 3.1.14 Magnification changer (Motor)................................................................................... 16 3.1.15 Magnification changer (Sensor) .................................................................................. 17 3.1.16 Interferometer Motor ................................................................................................. 17 3.1.17 Interferometer edge detector (Sensor) ........................................................................ 18 3.2 Computer Base .............................................................................................................19 3.2.1 Opening the Computer Base ...................................................................................... 19 3.2.2 Complete computer base ........................................................................................... 20 3.2.3 Hard drive (HDD) ........................................................................................................ 20 3.2.4 Power supply ............................................................................................................. 21 3.2.5 Fan ............................................................................................................................ 22 3.2.6 Power ON/OFF PCB .................................................................................................... 22 3.2.7 Motherboard ............................................................................................................. 23 3.2.8 Battery ....................................................................................................................... 24 3.2.9 Inductive Sensor (left/right eye switch)........................................................................ 24 3.2.10 Master switch and power entry module ..................................................................... 25 3.3 Instrument Base / Cross Stage ......................................................................................26 3.3.1 Opening the instrument base ..................................................................................... 26 3.3.2 Complete instrument base.......................................................................................... 27 3.3.3 Joystick with release button ....................................................................................... 27 3.3.4 Spur gear ................................................................................................................... 28 3.3.5 Spindle nut / spindle complete.................................................................................... 28 3.3.6 Cross table locking knob ............................................................................................ 29 3.4 Head Rest.....................................................................................................................29 3.4.1 3.4.2 3.4.3 3.4.4
Complete head rest .................................................................................................... 30 Chin rest unit ............................................................................................................. 30 Chin rest .................................................................................................................... 30 Forehead strap ........................................................................................................... 31
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3-1
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
Notes:
Service Manual Part No. 1836-541
Confidential and Proprietary
3-2
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
Adjustment / Calibration Requirements Following Replacements This table identifies checks/adjustments that must be made following removal/replacement procedures. The numbers listed under Follow-ups Required are keyed to the Follow-up Actions listed in the second column. For each Assembly Replaced, the follow-up numbers are listed in the order in which they should be performed, not necessarily in numerical sequence.
Assembly/Part Replaced Instrument head Complete head exchange............ Control cable……………………. Display complete......................... TFT Display and LED-Converter…. Keratometer complete………...... Slit module…………………….. CCD Camera…………………….. Digital IOL PCB…………………... Aperture / diaphragm module... Aperture / diaphragm (Sensor)…. Magnification changer (Motor)…………………………... Magnification changer (Sensor)………………………….. Interferometer Motor...…………. Interferometer edge detector…... Computer Base Opening the computer base……. Complete computer base……….. Hard drive (HDD)……….............. Power supply……………………. Fan……………………………….. Power on/off PCB……………….. Motherboard…………………..... Battery…………………………… Master switch and power entry module…………………………... Induction sensor (left/right eye switch)…………………………… Instrument Base / Cross Table Joystick with release button…….
Follow-ups Required 1 1 1 1 11, 13, 12, 7, 8, 18, 17, 1 13, 17, 1 6, 17, 18, 19, 20, 1 4, 3, 5, 6, 7, 8, 9, 21, 1 12, 1 12, 1 12, 14, 1
Follow-up Actions 12345678-
14, 1 9, 1 10, 1
910 -
1 15, 1 15, 1 22, 1, 23 1 1, 23 1 1
11 12 13 14 15 16 17 18 -
1, 23 1
19 20 -
1
21 22 23 -
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 3
System Checkout (Section 8 B) Optic cleaning (4.2.1) Laser power “Adjustment power” (4.2.3) Laser power "100% power" (4.2.4) Laser Monitoring (4.2.5) CCD camera sensitivity (4.2.6) Surrounding field illumination (4.2.7) Keratometer LED illumination (4.2.8) Interferometer-motor (4.2.9) ALM Measurement range alignment (4.2.10) Keratometer alignment (4.2.11) Aperture alignment (4.2.12) Slit module alignment (4.2.13) Magnification changer alignment (4.2.15) Re-imaging (4.3.1) ALM calibration (4.4.1) ACD calibration (4.4.2) Keratometer calibration (4.4.3) WTW calibration (4.4.4) Spot size calibration (4.4.5) Read / Write EEPROM (4.1.8) Power supply Voltages (4.3.6) Technical/Electrical Safety Check (Section 2)
Service Manual Part No. 1836-541
Parts Removal / Replacement
3.1
IOLMaster 5.5
Instrument head
3.1.1 Opening the instrument head Removal: 1) Loosen six hex socket screws (2.5 mm hex key) on the left cover (viewed from user side). Note that the screws are of different sizes when reassembling. 2) Remove cover. 3) Loosen only 3 screws (2.5 mm hex key) as shown in red in figure 3 - 1. 4) Remove right cover together with the spacers.
Reassembly: Reassemble in reverse order. Please do not tighten the screws as far as it will go.
Fig. 3 – 1: Head covers
3.1.2 Head covers / spacers Spare parts:
1836-516
Set of spare covers for IOLMaster V5.5
Removal: 1) Remove both head covers – see 3.1.1 2) Unscrew the 3 spacers (TORX TX10) from the right cover. 3) Replace covers or spacers as necessary. Reassembly: Reassemble in reverse order. Please do not tighten the screws as far as it will go.
Service Manual Part No. 1836-541
Confidential and Proprietary
3-4
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.1.3 Complete head exchange Spare part:
1772-810
IOLMaster head incl. packaging
Removal: 1) Remove both head covers – see 3.1.1 2) Disconnect control cable from the base. 3) Disconnect joystick cable from X209 of Digital IOL PCB and remove any cable fastener if present 4) Remove three fastening screws (4 mm hex key) of the instrument head and take off the instrument head.
Reassembly:
Fig. 3 – 2: Instrument head
Reassemble in the reverse order by following the instructions below. 1) Attach the new instrument head and secure it with three (4 mm hex key) fastening screws. 2) Measure resistance (grounding) between base plate and head frame with an appropriate meter. Ensure that both spots (especially on the head frame) are bare metal (screw heads should NOT be anodized). The resistance must be less than or equal to 0.5 Ω. Ensure that you zero out the meter first, to allow for any lead resistance. Note: Proper grounding is essential for a good axial-length SNR reading. Without good grounding the instrument may be able to read the test-eye (good SNR), but will be unable to read human eyes. 3) Reconnect joystick cable to X209 of the Digital IOL PCB. 4) Reconnect the control cable to the base. 5) Attach both covers and secure them with screws – see 3.1.1 6) Perform a System Checkout in accordance with the instrument checklist.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 5
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.1.4 Control cable Spare part: 1761-131
Cable IOLMaster (X601) - head
Removal: 1) Disconnect control cable from the computer-base. 2) Remove both head covers – see 3.1.1 Note: Please be extremely careful with the tiny cables and do NOT rip them out! 3) 4) 5) 6) 7) 8)
Disconnect connector X302 from the APD module. Disconnect connectors X201 and X231 from the Digital IOL PCB. Cut all cable ties Loosen the two cable clamps (2.5 mm hex key) Remove cable connections from the LCD display – see details at 3.1.5 Completely remove the control cable.
Reassembly: Reassemble in reverse order observing the following instructions. 1) Reconnect the LCD cable connections - see details at 3.1.5 2) Install the two cable clamps (2.5 mm hex key) 3) Reconnect the connectors X201 and X231 to the Digital IOL PCB. 4) Reconnect X302 to the APD module. 5) Attach the control cable clip. 6) Measure resistance (grounding) between base plate and head frame with an appropriate meter. Ensure that both spots (especially on the head frame) are bare metal (screw heads should NOT be anodized). The resistance must be less than or equal to 0.5 Ω. Ensure that you zero out the meter first, to allow for any lead resistance. Note: Proper grounding is essential for a good axial-length SNR reading. Without good grounding the instrument may be able to read the test-eye (good SNR), but will be unable to read human eyes. 7) Mount both head-covers – see 3.1.1 8) Perform a System Checkout in accordance with the instrument checklist
Service Manual Part No. 1836-541
Confidential and Proprietary
3-6
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.1.5 Display complete Spare part:
1770-344
LCD, without cable
Removal: 1) Remove both head covers - see 3.1.1 2) Disconnect the control cable from the base and disconnect the connectors from the Digital IOL PCB and APD- see 3.1.4 3) Cut the cable tie 4) Loosen the cable clamp close to the LC-display. 5) Loosen but do not remove the two M5 screws (4 mm hex key) on the LCD support, remove the display and place it together with the cable on a clean, padded surface (screen facing downward). 6) Open the LCD housing, carefully (TORX TX10) 7) Carefully disconnect cables from the LCD (CN1 and CN2 on TFT Color Display first) –see Fig. 3 – 3, and place down the display. 8) Disconnect display cable connections from the LEDconverter unit (CN1) Note: Taping the connectors together (in a row, see Fig. 3 – 4) will make the removal of the display cable easier. 9) Remove the strain relief device and pull the cables with connectors out of the housing.
Fig. 3 – 3: Display (opened)
Reassembly: Reassemble in reverse order observing the following instructions. 1) Exercise caution when tightening the screws for the cable Fig. 3 – 4: Display connectors clamp. The mount is plastic, and over tightening of the metal screws can cause the threads of the mount to break. 2) Carefully remove the adhesive tape or heat shrink tubing on the two connectors, plug the connectors onto LEDconverter unit (CN1 and CN2) and TFT panel. 3) Close and secure the display housing (do not bruise the tiny cables) with the removed screws. Mount the display on the IOLMaster head and press it till the screw heads, routing the cable to the left side of the housing. 4) Carefully reconnect the plugs of the control cable. 5) Adjust the friction of the swivel with a 3 mm hex key, if Fig. 3 – 5: adjustment of the friction necessary (Fig. 3 – 5) 6) Perform a System Checkout in accordance with the instrument checklist. SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 7
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.1.6 TFT Display and LED-Converter Spare part:
0510-510 0510-511
TFT-Color-Display 6.5-inch LED-Converter Unit for Backlight
Removal: 1) Remove both head covers – see 3.1.1 2) Remove the display and open the enclosure – see 3.1.5 3) To exchange a damaged LED-converter disconnect CN1 and CN2 first (see Fig. 3 – 3) and unscrew the two fixations screws with a crosstip screwdriver 4) For an exchange of the TFT-display just remove it from the front cover. Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount the LED-converter with the fixation screws and connect the cables to CN1 and CN2 (see Fig. 3 – 3) with great caution. 2) Be very careful and do not bruise the tiny cables during the reassembly. 3) Perform a System Checkout in accordance with the instrument checklist.
3.1.7 Keratometer complete Spare part:
1757-077
Keratometer, complete
Removal: 1) Remove both head covers – see 3.1.1 2) Disconnect the two connectors at the slit assemblies and those of the Keratometer-LEDs at the Digital IOL PCB (X217, X218 and X214). 3) Remove three fixation-screws (2.5 mm hex key) and carefully remove the whole Keratometer unit (see Fig. 3 – 6 item A). Fig. 3 – 6: Keratometer
Reassembly:
Reassemble in reverse order observing the following instructions. 1) Assemble Keratometer unit and loosely secure it with the three screws (2.5 mm hex key). 2) Plug-in connectors. 3) Perform the mechanical alignment – see 4.2.11 4) Check and if necessary align slit groups – see 4.2.13 5) Check and if necessary align aperture group – see 4.2.12 6) Check and if necessary adjust surrounding field illumination – see 4.2.7 7) Check and if necessary adjust KER-illumination – see 4.2.8 8) Check and if necessary calibrate KER – see 4.4.3 9) Check and if necessary calibrate ACD – see 4.4.2 10) Perform a System Checkout in accordance with the instrument checklist. Service Manual Part No. 1836-541
Confidential and Proprietary
3-8
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.1.8 Slit module Spare part:
1757-148
ACD slit module (identical for left and right slit module)
Removal: 1) Remove both head covers – see 3.1.1 Note: Observe ESD protection measures when working on electronic components. 2) Disconnect connector on the relevant slit module. 3) Remove two screws (1.5 mm hex key) and remove slit module. Note! Fig. 3 – 7: Slit module Do NOT remove both slit modules at the same time, unless you are replacing the entire Keratometry assembly. One slit module is the reference for the other. In the event that both modules must be replaced, either replace the entire Keratometry/Slit assembly, or replace the slit PCBs one at a time, aligning the first before attempting to replace the second. Reassembly: Reassemble in reverse order observing the following instructions. 1) Loosely screw new slit module on. 2) Reconnect cable to Digital IOL PCB. 3) Align the slit-module – see 4.2.13 4) Check and if necessary calibrate ACD – see 4.4.2 5) Perform a System Checkout in accordance with the instrument checklist.
3.1.9 CCD Camera Spare part:
1770-340
CCD camera set, 9V
Note: The (redesigned) alignment device for IOLMaster is absolutely vital for replacing the CCD camera. Without this tool it is impossible to verify the required focus level of the camera. This may lead to incorrect measurement, especially with regard to ACD measurements. Alignment issues between the overview and Keratometry positions may also present themselves. The exchange of the CCD camera without using the (redesigned) alignment device is not permitted. Old alignment devices (bought before July 2008) can be retrofitted. Please send the related tools to our Jena (Germany) repair center.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 9
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
Identifying a “worn out” CCD camera Note: Before determining whether a camera is “worn out”, several preconditions must be checked and / or set on an IOLMaster to ensure that the camera parameters are properly adjusted. 1) Preconditions: Proper connections Frame grabber settings (see 4.1.3) Clean optics (see 4.2.1) Camera sensitivity voltage settings (see 4.2.6) Surrounding field illumination settings (see 4.2.7) Aperture alignment (see 4.2.12) 2) Signs for a worn out CCD Camera Most easily noticed in the Overview mode, a black strip will be seen along the left edge of the Fig. 3 – 8: Frame Grabber settings camera image. This black strip will encompass the full height of the camera image, and will widen with time. Fig. 3 - 9 shows a typical black strip. When in the Overview mode or/and Keratometer mode the illumination and fixation markers will have a black shadow to the side. In extreme cases the shadows will appear as bars that trail to the side as indicated in Fig. 3 - 9.
Checklist before replacement Black stripe at the left edge
Note: The following points are vital to be checked as “YES” to ensure a successful camera replacement.
Black shadows/trails beside the illumination and fixation lights
Fig. 3 – 9: Worn out camera
1) Camera alignment tool (See 5.3 – tool list) available? 2) Live-image visible? Note: Heating up a CCD camera which does not show any image can easily re-activate it. Use a heat gun for this purpose. Be extremely carefully while blowing hot air into the instrument. There is a high risk of damaging other components. 3) CCD camera is still in the original position (compare crosshair position between Overview and Keratometer mode with the test-eye) and measurements (ALM) show good results in relation to the SNR value? Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 10
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
Note: If the CCD camera is no longer in its original position and / or the SNR of the ALM measurements is not within specification, the replacement of the camera must be aborted, and the measurement head must be sent in for repair in house. Removal: Note: The following procedure describes only the replacement of the CCD camera. The positioning will only be as good as the old / worn-out camera was. It is not possible to re-align the camera if once the original position has been lost or if the present camera is misaligned. This requires either a head-replacement or in-house repair. 1) Remove both head covers – see 3.1.1
1
2) Remove the test-eye from the chin rest, if present. 3) Attach the alignment tool to the Keratometer. Position the tool 30° tilted from the axis and center the ring to the hole of the Keratometer. Rotate the tool until the hooks are behind the Keratometer frame and tighten the screws carefully. Check that the tool is immovable. 4) If not already done, switch on the IOLMaster, create a test patient and change to the Keratometer mode. The sphere side of the rotatable head must be in the direction to the instrument.
2
5) Focus the spots (using the Z-drive screw of the tool) and align the X and Y axis in relation to the crosshair. Loosen the knurledhead screw to position the sphere. Tighten the screw when done. Note: The image in Keratometer mode must be exactly centered as shown below before proceeding further.
Fig. 3 – 10: Aligning the tool
6) Cut off all cable ties (zip ties) from the monitor and control cable. Remove screws from any cable clamps which may be present. Loosen the two screws (4 mm hex key) which hold the LCD in place, and carefully lay the LCD next to the instrument. Exercize caution so as not to stretch the cable. Note: Observe ESD protection measures when working on electronic components. 7) Shut down the instrument and disconnect the camera cable from Digital IOL PCB (X219) 8) Remove the four screws (2 mm hex key) and replace the CCD camera SM-30-4054-en Section 3, Revision A, Feb. 2010
Fig. 3 – 11: Tool positioning
Confidential and Proprietary
3- 11
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
9) Loosely secure it with the four screws (2 mm hex key) and plug-in the camera cable (X219) 10) Adjust the camera sensitivity voltage as necessary (2.9 ± 0.3V at R75, next to X219) and align the camera to the tool (in Keratometer mode). Remember that the image on the screen must be exactly centered. Be as accurate as possible while aligning the camera. 11) Check overview and Keratometer position by moving the magnification changer manually in PIOTest (motor-tab). If the CCD camera was well positioned, both crosshairs should match with the sphere (no horizontal / vertical gap). Focus shift determination / compensation 1) Rotate the tool-head 180° (“black-finished” back-side must face to the instrument). 2) Change to Keratometer mode and run PIO-Test (service-mode). Go to “Chanel 3” of the PIO-Ports tab and switch on the slit groups alternately.
Fig. 3 – 12: Focus shift
3) By turning the Z-axis screw of the tool, the background of the tool-head can be focused. If the focus level of the CCD camera matches to the optical alignment of the instruments, both slits are projected to the middle of the crosshair and do not “virtually” move while switching on the left and right hand side of the slit groups alternately. As a tolerance range, the inner circle of the crosshair can bee seen. While focused, the slit projections must lie within this (inner) circle. 4) If there is a focus shift present, the slits are not projected as described in Fig 3 - 12 - they are out of the middle position while the background is focused – see Fig. 3 - 13
Fig. 3 – 13: Focus shift Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 12
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
5) See the Fig. 3 - 14 and 3 - 15 to determine if the focus level of the CCD camera is too far in the front or too far in the back. 6) If the left slit is projected too far left and the right slit projected too far right, the focus level of the CCD chip is too far back – add shims between the white camera bases and the metal plate.
Fig. 3 – 14: Focus shift – CCD chip too far back
7) If the left slit is projected too far right and the right slit projected too far left, the focus level of the CCD chip is too far in the front – add shims between the casting and the metal plate of the camera.
Fig. 3 – 15: Focus shift – CCD chip too far in the front
Note: Adding shims will require you to take the camera out again. Follow the related sections of the replacement procedure. Afterwards continue to determine the focus shift again. Note: Always add equal numbers of shims to all four fixation points – either one shim (0.2 mm) or two shims (0.4 mm) at each point. 8) Continue as long as the focus shift meets the tolerance range (While the background is focused, the slit projections must lie within the inner circle of the crosshair). Reassembly: Reassemble in reverse order observing the following instructions. 1) Reassemble the instrument in reverse order. Secure the LCD cable loop safely using zip ties. Ensure free movement for the changer. Remove the tool and reattach the head covers. Quit PIOtest and service mode 2) Take measurements with the test-eye and verify the calibration. Re-calibrate KER and ACD if needed.-see 4.4.3 and 4.4.2 3) Compare SNR values (ALM measurements) of “before” and “after” the camera replacement, values should be equal. 4) Perform a System Checkout in accordance with the instrument checklist.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 13
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.1.10 Digital IOL PCB Spare part:
1534-730
Digital IOL PCB (DIGIT)
Removal: 1) Enter Service mode, System tab, Read/Write EEPROM and write a new ini file as a backup of the calibration data – see 4.1.8 2) Enter Service mode, System-tab and run Windows® Explorer copy and paste the created ini file on a reliable storage – see 4.1 3) Shut down and switch off the instrument. Note: Observe ESD protection measures when working on electronic components. 4) Remove both head covers – see 3.1.1 5) Disconnect all connectors. It is recommended to mark the connectors which look similar. Please be extremely careful with the tiny cables and do NOT rip them out! Caution: Immediately connect short-circuit plug (X221) Fig. 3 – 16: IOL PCB into the connector (X216) of the removed laser diode cable. 6) Remove three screws (2.5 mm hex key) on the fixation points (see Fig. 3 – 16 item A) and remove PCB. Note: Watch that the spacers between the PCB and frame do not get lost. Re-secure them with adhesive if necessary. Reassembly: Reassemble in reverse order observing the following instructions. 1) Adjust potentiometer R7, R9 to minimum (several turns anticlockwise) to ensure that the laser diode will not be overloaded and damaged. 2) Re-plug all connections. 3) Switch the instrument on (errors related to the laser power and/or the changer may appear). 4) Adjust laser power (100%) - see 4.2.4 5) Adjust laser power (adjustment power) - see 4.2.3 6) Adjust laser monitoring - see 4.2.5 Fig. 3 – 17: Digital IOL PCB with 7) Adjust camera-sensitivity - see 4.2.6 marked pots 8) Adjust surrounding field illumination - see 4.2.7 9) Adjust Keratometer illumination - see 4.2.8 10) Adjust motor speed - see 4.2.9 11) Compare calibration data and if necessary reload them from the iol.ini file which has been made before the removal - see 4.1.8 12) Perform a function test in accordance with the checklist. Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 14
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.1.11 APD PCB Æ The APD PCB is NOT a field replaceable item. Replacement of the APD PCB MUST be performed at a Carl Zeiss Meditec In house repair center.
3.1.12 Aperture / diaphragm module Spare part:
1756-302
Keratometer diaphragm module
Fixation screws
Aperture assembly
Removal: 1) 2) 3) 4)
Shut down and switch off the instrument. Remove both head covers – see 3.1.1 Disconnect connectors X205, X212 on the Digital IOL PCB. Remove two screws (2.5 mm hex key) (see Fig. 3 - 18) and remove aperture assembly (diaphragm module).
Reassembly: Reassemble in reverse order observing the following instructions. 1) Insert new aperture assembly and secure it loosely. 2) Insert and secure connectors X205, X212 on the Digital IOL PCB. 3) Align aperture assembly – see 4.2.12 4) Perform a System Checkout in accordance with the instrument checklist.
3.1.13 Aperture / diaphragm (Sensor) Spare part:
1693-881 Optical sensor diaphragm X205
Removal:
Fig. 3 – 18: Diaphragm / Aperture
1) Remove the aperture assembly- see 3.1.11 2) Mark the position of the sensor as exactly as possible 3) Remove two screws (2.5 mm hex key) of the sensor (see Fig. 3 – 19) Reassembly: Reassemble in reverse order observing the following instructions. 1) Insert new sensor according to the marking 2) Attach the diaphragm module- see 3.1.11 3) Perform a System Checkout in accordance with the instrument checklist. SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 15
Fig. 3 – 19: aperture sensor
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.1.14 Magnification changer (Motor) Spare part:
1693-816
Changer motor
Removal: 1) 2) 3) 4) 5) 6)
Shut down and switch off the instrument. Remove both head covers – see 3.1.1. Remove diaphragm assembly – see 3.1.11. Disconnect connector X211 on the Digital IOL PCB. Remove the spindle nut. Remove two screws (2 mm hex key) and remove motor with spindle.
Motor
Screws
Spindle
Spindle nut
Reassembly:
Fig 3 – 20: Changer motor
Reassemble in reverse order observing the following instructions. 1) Install new motor with spindle and tighten two screws (2 mm hex key). 2) Attach spindle nut and align spindle in such a way that the changer can be easily moved and the spindle goes through the center of the motor. Note: Should a vertical deviation of the measuring image become noticeable after installation of the motor, this can be aligned by slightly turning the spindle (fine positioning between 2 full steps of the step motor). 3) Tighten spindle nut and plug connection X211. 4) Mount and align diaphragm module –see 3.1.11 and 4.2.12 5) Perform a System Checkout in accordance with the instrument checklist.
Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 16
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.1.15 Magnification changer (Sensor) Spare part:
1693-867
Optical sensor changer X207
Removal: 1) Shut down and switch off the instrument. 2) Remove both head covers – see 3.1.1 3) Disconnect connector X207 on the Digital IOL PCB. 4) Remove 2 securing screws (2.5 mm hex key) and take out the sensor. Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount new sensor and loosely secure it with screws. 2) Re-connector X207 on the Digital IOL PCB. 3) Align the sensor position – see 4.2.15, section “Sensor”. 4) Perform a System Checkout in accordance with the instrument checklist.
Sensor Fig. 3 – 21: Changer sensor
3.1.16 Interferometer Motor Spare part:
1755-837
Motor interferometer, complete
Removal: 1) Shut down and switch off the instrument. 2) Remove both head covers – see 3.1.1 3) Disconnect motor cable from X208 on the Digital IOL PCB. 4) Move the prism slide to the bottom 5) Remove the two screws (2.5 mm hex key) and remove the motor.
Reassembly:
Fig 3 – 22: Interferometer motor
Reassemble in reverse order observing the following instructions. 1) Install motor ensuring that smooth drive motion is observed. 2) Adjust the interferometer motor speed - see 4.2.9 3) Perform a System Checkout in accordance with the instrument checklist.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 17
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.1.17 Interferometer edge detector (Sensor) Spare part:
1693-879 1694-348
Optical sensor interferometer X203 Optical sensor interferometer X204
Removal: 1) Shut down and switch off the instrument. 2) Remove both head covers – see 3.1.1. 3) Disconnect the cable of the damaged sensor, either X203 (blue/upper) or X204 (red/lower), see Fig. 3 - 23 4) Move the prism slide in a good position, where the screws are reachable 5) Remove the two fixation screws (2.5 mm hey key) and remove the sensor
Fig 3 – 23: Interferometer sensor
Reassembly: Reassemble in reverse order observing the following instructions. 1) Install the new sensor. Ensure that the flag of the interferometer sled can easily pass through the sensor. 2) Reconnect the sensor cable at the appropriate place 3) Align the ALM measurement range- see 4.2.10 4) Perform a System Checkout in accordance with the instrument checklist.
Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 18
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
3.2
Parts Removal / Replacement
Computer Base
Fig. 3 – 24: IOLMaster Computer base
3.2.1 Opening the Computer Base Removal: 1) Shut down, switch off the instrument and unplug the main power cord. 2) Disconnect control cable from the base. 3) Remove four screws (2 mm hex key) on the connector panel. 4) Remove eight M5 hex socket head screws (3 mm hex key). Note: It is recommended to position the instrument over the edge of its table in order to access the screws under the base. After removing the screws from one side, then position the instrument over the other side of the table to access the remaining screws. 5) Place the computer base with the head rest upright again. 6) Lift the base cover with the head rest and measuring head off the base plate and put it aside. Fig. 3 – 25: Opening the base
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 19
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
Reassembly: Reassemble in reverse order observing the following instructions. 1) When reattaching the computer base, take care not to bend the left/right switch (sensor). 2) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 3) Perform a System Checkout in accordance with the instrument checklist.
3.2.2 Complete computer base Spare part:
1773-042
Computer IOLMaster with cover
Removal: 1) If possible, back up the existing lens and user database before exchanging the computer base - see user manual for details. 2) Shut down and switch off the instrument. 3) Open the computer base – see 3.2.1. Reassembly: Reassemble in reverse order observing the following instructions. 1) When reattaching the new computer base, take care not to bend the left/right switch (sensor). 2) Tighten the four screws (2 mm hex key) on the Fig. 3 – 26: Computer connector panel before the screws on the bottom of the base. 3) Restore regional settings – see 4.3.1 4) Restore program settings – see 4.3.1 5) Install the backed up lens- and user database – see 4.3.1 6) Re-install printer drivers – see 4.3.1 7) Re-install software-options– see 4.3.1 8) Restore network settings – see 4.3.1 9) Create a local ghost image – see 4.3.1 10) Perform a System Checkout in accordance with the instrument checklist.
3.2.3 Hard drive (HDD) Spare part:
0509-551
Fig 3 – 27: computer base (bottom view)
Hard disk 2.5” SATA
Removal: 1) If possible, back up the existing lens and user database before exchanging the HDD - see user manual for details. 2) Shut down and switch off the instrument. 3) Open the computer base – see 3.2.1 4) Turn the computer base around 180 degrees Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 20
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
5) Remove the four mounting screws (2.5 mm hex key), see Fig. 3 - 27 (item A). Exercise caution so that the screws do not drop out (Place tape over the holes). 6) Turn again the computer base and carefully unplug the HDD in direction of the control cable connector Reassembly: Reassemble in reverse order observing the following instructions. 1) Connect the new HDD and secure it with the four screws (2.5 mm hex key) 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) Image the HDD from the related Service-USB flash drive – see 4.3.1 5) Restore regional settings – see 4.3.1 6) Restore program settings – see 4.3.1 7) Install the backed up lens and user database – see 4.3.1 8) Re-install printer drivers – see 4.3.1 9) Re-install software-options– see 4.3.1 10) Restore network settings – see 4.3.1 11) Create a local ghost image – see 4.3.1 12) Perform a System Checkout in accordance with the instrument checklist.
3.2.4 Power supply Spare part:
0508-752
Power supply 90 W 12 V (DC)/ 7.5 A
Removal: 1) 2) 3) 4) 5) 6)
Shut down and switch off the instrument. Disconnect the main power cord. Open the computer base – see 3.2.1 Cut all of the zip ties from the power supply. Disconnect all cables. Remove the four screws (2 mm hex key) from the bottom side (see Fig. 3 – 27, item B). Exercise caution as the power supply may drop down.
Reassembly: Reassemble in reverse order observing the following instructions. 1) Do not forget to reconnect any and all ground connections! The latch of the blade connector should be set in the direction to the screw head. 2) Check and if necessary adjust the output voltage – see 4.3.6 3) When reattaching the computer base, take care not to bend the left/right switch (sensor). 4) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 5) Perform a System Checkout inclusive the technical safety check in accordance with the instrument checklist and Section 2. SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 21
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.2.5 Fan Spare part:
1693-872
Fan complete
Removal: 1) 2) 3) 4)
Shut down and switch off the instrument. Open the computer base – see 3.2.1 Disconnect the cable connection of the fan and remove the cable clamp. Remove two screws (2 mm hex key) from the bottom side (see Fig. 3 – 27, item C) and replace the fan.
Reassembly: Reassemble in reverse order observing the following instructions. 1) Tighten the screws loosely. 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) If the computer base is closed, slide the fan (after loosening the screws from the bottom side) as far as possible in the direction of the cover (avoid any vibration or noise), then tighten all screws. 5) Perform a System Checkout in accordance with the instrument checklist.
3.2.6 Power ON/OFF PCB Spare part:
1671-563 1694-129 1694-354
Power ON/OFF PCB Cable for power ON/OFF (X853 to X602) Power ON/OFF - mains cable
Removal: 1) 2) 3) 4) 5)
Shut down and switch off the instrument. Disconnect the main power cord. Open the computer base – see 3.2.1 Disconnect the X851, X885 and X853 connections Remove the two M3 screws with 2.5 mm hex key and one M3 screw with an 2.0 mm hex key and remove the Power ON/OFF PCB
Reassembly: Reassemble in reverse order observing the following instructions. 1) Re-connect all connection. 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) Perform a System Checkout inclusive the technical safety check in accordance with the instrument checklist and Section 2. Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 22
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.2.7 Motherboard Spare part:
1528-161
IOLMaster MED-COM PCB
Removal: 1) Shut down and switch off the instrument. 2) Open the computer base – see 3.2.1
1
Note: Observe ESD protection measures when working on electronic components. 3) Disconnect all cables shown in the illustration (Fig. 3 – 28). 4) Remove the mounting bolts for the VGA and COM port (4.5 mm hex socket wrench) 5) Remove the other screws (2.5 mm hex key) and remove the board 6) Remove the HDD – see 3.2.3 Reassembly: Reassemble in reverse order observing the Fig 3 – 28: MED-COM Motherboard following instructions. 1) Assemble the used HDD and the new motherboard 2) Insert new motherboard and secure it in position. 3) Re-connect all cables. 4) When reattaching the computer base, take care not to bend the left/right switch (sensor). 5) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 6) Perform a System Checkout in accordance with the instrument checklist.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 23
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.2.8 Battery Spare part:
0509-839
Lithium coin cell 3V/230mAh 3,2x20mmØ
Removal: 1) Shut down and switch off the instrument. 2) Open the computer base – see 3.2.1 3) Remove the battery with a tool, e.g. a screw driver (see Fig. 3 – 29) Note: Place the defective battery in an approved depleted battery receptacle for proper disposal.
Reassembly:
Fig 3 – 29: BIOS CMOS backup Battery
1) Mount the new battery 2) Verify the proper battery voltage of 3.0 V DC or higher using a calibrated digital volt meter. 3) When reattaching the computer base, take care not to bend the left/right switch (sensor). 4) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 5) Perform a System Checkout in accordance with the instrument checklist.
3.2.9 Inductive Sensor (left/right eye switch) Spare part:
1693-874
Inductive sensor switch
Removal: 1) Shut down and switch off the instrument. 2) Open the computer base – see 3.2.1 3) Disconnect the sensor (X603) and remove two screws (3 mm hex key) Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount the new sensor at the bar 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). It is recommended to open the white plate at the base cover to check the position of the sensor. 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) Perform a System Checkout in accordance with the instrument checklist.
Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 24
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.2.10 Master switch and power entry module Spare part:
1693-874 0505-635
Master switch, 2 pins, green power entry module medical version 3A
Removal: 1) Shut down and switch off the instrument. 2) Disconnect the main power cord 3) Open the computer base – see 3.2.1 Master switch: 1) Disconnect the four cables 2) Compress the snap latches and remove the switch Power entry module: 1) 2) 3) 4)
Remove the fan – see 3.2.5 Remove the master switch Disconnect the three cables Carefully compress the lower and upper snap latch with an appropriate screw driver and remove the power entry module
Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount the new part Note: Take care of the grounding measures and connect all grounding cables at the right place!
Fig 3 – 30: master switch and power entry module
2) Reinstall the fan– see 3.2.5 3) Perform a System Checkout inclusive the technical safety check in accordance with the instrument checklist and Section 2.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 25
Service Manual Part No. 1836-541
Parts Removal / Replacement
3.3
IOLMaster 5.5
Instrument Base / Cross Stage
Fig. 3 – 31: Instrument base
3.3.1 Opening the instrument base 1) Shut down and switch off the instrument. 2) Move the instrument to the highest position (by using the joystick). Note: Always use the right screw driver size and be very careful. Do not scratch the paint. 3) 4) 5) 6)
Remove the two screws on the side of the housing and remove the housing. Remove both head covers – see 3.1.1 Remove the two screws on the side of the cover and remove the cover. Remove the front plate (four screws)
Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 26
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.3.2 Complete instrument base Spare part:
266101-4501-927
Cross Stage for IOLMaster
Removal: 1) Shut down and switch off the instrument. 2) Remove the complete measurement head – see 3.1.3 3) Remove the four screws of both gear racks (2.5 mm hex key), see Fig. 3 - 32 4) Remove the instrument base Reassembly: Reassemble in reverse order observing the following instructions. 1) Install the gear rack onto the computer base, insert the instrument base and install the second toothed rack. Fig. 3 – 32: gear rack Note: Make sure that the gears are identically positioned on the left and right. The shaft must have an endplay clearance of approx. 1/10 mm in each position. 2) Check upper and lower stop of the horizontal movement range and readjust if necessary.
3.3.3 Joystick with release button Spare part:
266101-4560-514
IOLMaster joystick with release button
Removal: 1) 2) 3) 4)
Shut down and switch off the instrument. Open the instrument base – see 3.3.1 Move instrument head to the lowest position (by using the joystick) Mark the position of the joysticks stop-pin and the stop at the spur gear. Note: Before exchanging gears, mark their precise position and insert the new part accordingly. The subsequent adjustment of the upper and lower stops may take some time.
5) Disconnect joystick cable on X209 of Digital IOL PCB and remove cable fixation if present. 6) Remove the four fixation screws (2.5 mm hex key) and remove joystick assembly. Reassembly: Reassemble in reverse order observing the following instructions. 1) Re-position the joystick as the markings show. 2) Check upper and lower stop of the horizontal movement range and readjust if necessary.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 27
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.3.4 Spur gear Spare part:
266101-4533-010
Spur gear
Removal: 1) 2) 3) 4)
Shut down and switch off the instrument. Open the instrument base – see 3.3.1 Move instrument head to the lowest position (by using the joystick). Mark the position of the joysticks stop-pin and the stop at the spur gear. Note: Before exchanging gears, mark their precise position and insert the new part accordingly. The subsequent adjustment of the upper and lower stops may take some time. 5) Mark the position of the defective gear, and do not move the joystick afterwards. 6) Remove the spur gear (loosen the screw with 3 mm hex key). Reassembly:
Reassemble in reverse order observing the following instructions. 1) Apply the marking to the new spur gear. 2) Insert the new spur gear in accordance with the position marking and tighten the fixation screw. 3) Test the upper and lower stops of the movement range, and readjust if necessary.
3.3.5 Spindle nut / spindle complete Spare part:
266101-4502-514 266101-4510-010
Spindle, complete Spindle nut
Removal: 1) Shut down and switch off the instrument. 2) Open the instrument base – see 3.3.1 3) Move instrument head to the lowest position (by using the joystick). Note: Before exchanging gears, mark their precise position and insert the new part accordingly. The subsequent adjustment of the upper and lower stops may take some time. 4) Remove the two screws (2.5 mm hex key) and then remove the spindle nut. 5) Remove the screws (2.5 mm hex key) and then remove the spindle.
Reassembly: Reassemble in reverse order observing the following instructions. 1) When positioning the new spindle nut, take care that the same horizontal position in relation to the spindle will be chosen. 2) Test the upper and lower stops of the movement range, and readjust if necessary. Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 28
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.3.6 Cross table locking knob Spare part:
1775-095
Service Set cross table locking knob IOL
Removal: 1) Remove the cap with an appropriate tiny screw driver 2) Loosen the screw and remove the turning knob. Reassembly: Reassemble in reverse order observing the following instructions. 1) Be careful to replace the parts in their proper order. 2) Press the knob down as far as it will go and fasten the fixation screw. 3) Test the proper function. Fig. 3 – 33: Service Set locking knob
3.4
Head Rest
Fig. 3 – 34: Head rest SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 29
Service Manual Part No. 1836-541
Parts Removal / Replacement
IOLMaster 5.5
3.4.1 Complete head rest Spare part:
1025-670
Head rest
Removal: 1) Shut down and switch off the instrument. 2) Remove head rest from the fixtures, loosen the two locking screws with a 4 mm hex key (Fig. 3 - 34 – item 1) and pull the head rest out.
Reassembly: Reassemble in reverse order observing the following instructions. 1) When putting the head rest back into the fixtures, take care that the tubes are completely in before tightening the fixation screws.
3.4.2 Chin rest unit Spare part:
1294-613
Chin rest unit, assembled
Removal: 1) Shut down and switch off the instrument. 2) Remove head rest - see 3.4.1 3) Slide the chin rest unit off from the head rest tube.
Reassembly: Reassemble in reverse order observing the following instructions. 1) When putting the head rest back into the fixtures, take care that the tubes are completely in before tightening the fixation screws.
3.4.3 Chin rest Spare part:
319752-8016-000
Chin rest
Removal: 1) Shut down and switch off the instrument. 2) Remove the fixation screw (3 mm hex key), see Fig. 3 - 34 item 3 3) Remove the chin rest.
Reassembly: Reassemble in reverse order. Service Manual Part No. 1836-541
Confidential and Proprietary
3 - 30
SM-30-4054-en Section 3, Revision A, Feb. 2010
IOLMaster 5.5
Parts Removal / Replacement
3.4.4 Forehead strap Spare part:
319752-0024-000
Forehead strap
Removal: 1) Shut down and switch off the instrument. (2.5 mm hex key) 2) Remove the two fixation screws (2.5 mm hex key), see Fig. 3 - 34 item 2 3) Remove the forehead strap
Reassembly: Reassemble in reverse order observing the following instructions. 1) Fasten the two screws with great care, do not break the plastic of the strap. Note: Do not use any adhesives for locking the two screws. Chemical reactions may weaken the forehead strap and lead to breakage under load.
SM-30-4054-en Section 3, Revision A, Feb. 2010
Confidential and Proprietary
3- 31
Service Manual Part No. 1836-541
Section 4 – Adjustment and Calibration 4.1 Service Mode..................................................................................................................3 4.1.1 Entering the Service Mode............................................................................................ 3 4.1.2 Calibration Tabs (ALM, ACD, KER, WTW)...................................................................... 4 4.1.3 Grabber Tab................................................................................................................. 4 4.1.4 Prog Tab ...................................................................................................................... 5 4.1.5 System Tab .................................................................................................................. 5 4.1.6 Delete Data .................................................................................................................. 6 4.1.7 Speed Test ................................................................................................................... 6 4.1.8 Read / Write EEPROM................................................................................................... 7 4.1.9 PIO-Test ....................................................................................................................... 7 4.2 Adjustments / Alignments of the Measuring Head .........................................................9 4.2.1 Optics cleaning ............................................................................................................ 9 4.2.2 Power supply voltage Digital IOL board ...................................................................... 14 4.2.3 Laser power “Adjustment power” .............................................................................. 15 4.2.4 Laser power "100% power" ....................................................................................... 16 4.2.5 Laser Monitoring........................................................................................................ 17 4.2.6 CCD camera sensitivity ............................................................................................... 18 4.2.7 Surrounding field illumination..................................................................................... 18 4.2.8 Keratometer LED illumination ..................................................................................... 19 4.2.9 Interferometer-motor ................................................................................................. 20 4.2.10 ALM Measurement range alignment .......................................................................... 22 4.2.11 Keratometer alignment .............................................................................................. 24 4.2.12 Aperture alignment .................................................................................................... 24 4.2.13 Slit group alignment................................................................................................... 25 4.2.14 APD alignment ........................................................................................................... 26 4.2.15 Magnification changer alignment ............................................................................... 26 Sensor........................................................................................................................ 27 Flag............................................................................................................................ 29 Changer Steps............................................................................................................ 30 4.3 Adjustments / Alignments of the computer..................................................................31 4.3.1 Re-imaging................................................................................................................. 31 Re-imaging options .................................................................................................... 31 Re-imaging procedure ................................................................................................ 32 Installation of Software Version 5.5............................................................................ 33 Settings after re-imaging ............................................................................................ 34 4.3.2 HDD Structure (Databases and Logfiles)...................................................................... 36 4.3.3 Database Exchange .................................................................................................... 36 4.3.4 Deleting of corrupted Databases ................................................................................ 36 4.3.5 Deleting the administrator-password .......................................................................... 37 4.3.6 Update the Software .................................................................................................. 37 4.3.7 Power supply Voltages ............................................................................................... 37 4.4 Calibration....................................................................................................................38 4.4.1 ALM calibration.......................................................................................................... 38 4.4.2 ACD calibration .......................................................................................................... 39 4.4.3 Keratometer calibration.............................................................................................. 40 4.4.4 WTW calibration ........................................................................................................ 41 4.4.5 Spot size calibration ................................................................................................... 41 SM-30-4054-en Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4-1
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Notes:
Service Manual Part No. 1836-541
Confidential and Proprietary
4-2
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
4.1 Service Mode The IOLMaster has an embedded service mode which can be entered with special passwords. The passwords are case sensitive and are not permitted to distribute to any non-authorized person. The usage of the service mode is restricted to certified service engineers, trained from an authorized Carl Zeiss Meditec trainer.
4.1.1 Entering the Service Mode 1) The service mode can be activated via Æ “Options” Fig. 4 - 1: Entering Options Æ “Service” 2) A password prompt appears. Enter the related password and confirm. 3) Depending on whether the “Service” password or “Calibration” password is used, the Service Menu will open differently. While all calibration pages are able to be viewed with the service password, the calibration password also allows changes within these pages and so the sections are highlighted.
Fig. 4 - 2: Entering Service mode
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 3
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Fig. 4 - 4: Calibration mode
Fig. 4 - 3: Service mode
Note: Usage of the calibration password is only recommended for the purpose of calibration. Exiting the calibration mode by clicking “OK” will cause the EEPROM to be re-written. Because of the delays which the rewriting causes, the rewriting of the EEPROM should be minimized to extend its lifetime.
4.1.2 Calibration Tabs (ALM, ACD, KER, WTW) Æ See “4.4 Calibrations” for detailed information.
4.1.3 Grabber Tab The frame grabber settings are of vital importance in relation to all image analyzing measurements of the IOLMaster (ACD, KER and WTW). These settings are to remain at the factory default values and must not be changed at any time.
Fig. 4 - 5: Grabber Tab
Service Manual Part No. 1836-541
Confidential and Proprietary
4-4
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
4.1.4 Prog Tab The Prog Tab provides the “Save” section, the “Output” section and the “Database” section. The defaults are shown in Fig. 4 - 6. “Save” section The “Grab Images” checkbox can be activated to highlight the checkboxes next to it. These checkboxes can be individual activated to save the images during the measurement (KER, ACD or WTW). The related images will be saved in the IOLMaster folder and may be used later for detailed troubleshooting
Fig. 4 - 6: Prog Tab
Note: After the images have been made, the feature must be deactivated manually again. Otherwise, the speed of the measurements will slow down and the HDD will be overloaded with images. “Output” section Activating the “Detailed Output” checkbox will show the detailed measurement of the KER. Instead of an average measurement, the five individual results will be displayed. “Manual R/L Switch” allows switching via touchpad between the left and right eye side, without moving the instruments head. The magnetic sensor which normally detects the right and left side is temporary disabled. To switch between both sides, just click in the related column (OD or OS). “Save and show Raw Signal” is only used at the factory. It shows the raw signal in ALM mode. “Database” section With both of these checkboxes checked, older patient files will be deleted from the HDD after the time specified in “Program Settings” menu has elapsed. Unchecking these boxes should not be done, as this would allow the HDD to fill up.
4.1.5 System Tab The system tab page provides links to the Windows® Explorer, the Control Panel, the PIO-Test, the Speed Test and the EEPROM. It is also possible to erase some data via the “Delete Data” link
Fig. 4 - 7: System tab SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 5
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
4.1.6 Delete Data The Delete Data submenu enables several options. – see Fig. 4 - 8 If the related checkboxes are checked, the corresponding data will be erased by clicking “OK”. The following window shows the selection again and needs to be confirmed to finally delete the data.
Fig. 4 - 8: Delete data
4.1.7 Speed Test The speed test allows direct access and control to the interferometer motor. Æ See 4.2.9 Interferometer motor for details. The menu buttons also allows for opening any of the other speed files, which are stored at C:\ (root directory) as “vmess_hsp3.dat”. Speed test files can also be printed.
Service Manual Part No. 1836-541
Confidential and Proprietary
4-6
Fig. 4 - 9: Speed-Test Navigation
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
4.1.8 Read / Write EEPROM The EEPROM is the main location of all calibration data. There is also a backup file for this data, which can be found in the IOLMaster folder. The name of the file is “iol.ini”. “Write INI File” This option will manually create a backup file (ini-file). The destination can be chosen within a popup-window. Please note that the destination must be entered by hand (in the “File name” field) and can not be selected by browsing folders. Note: The changer steps will also be written into the *.ini-file. Fig. 4 - 10: EEPROM
“Read INI File” This option reads an existing ini-file. The destination can be chosen within a popup-window. Please note that the destination must be entered by hand and can not be selected by browsing folders. “Write to EEPROM” This option re-writes the EEPROM with the data currently displayed in the sections of the EEPROMscreen (see Fig. 4 - 10). The instrument must be restarted after the EEPROM has been rewritten. It is highly recommended to compare the data of the ini-file and the EEPROM afterwards to verify the success of the writing procedure. Open the related ini-file via “Notepad” and compare the sections with those of the EEPROM screen. If they are not the same, repeat the procedure.
4.1.9 PIO-Test The PIO Test contains three tabs which can be used for field-service. The “Inbetriebnahmetest” tab is only for the manufacturing. For efficient use of the PIO-Test, the service-mode must be closed. Therefore click outside of the PIOTest window to get back to the main service mode. Exit the service mode and get back to the PIOTest window by using the + key-combination on the keyboard.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 7
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
PIO-Ports-tab The PIO-Ports tab contains several sections for manual control of the instruments functions. Channel 0: Channel 0 provides information about laser and APD. By checking the available radio buttons, the related information will be displayed as increments (numbers) within the numeric section within this channel. Channel 1: Clicking the check box switches the overview illumination LED’s on. Channel 2: The check boxes and radio buttons within this channel control the laser diode. Also the fixation LED can be controlled within this channel.
Fig. 4 - 11: PIO-ports
Channel 3: The LED’s of the Keratometer and those of the slitgroups (ACD = VKT) can be controlled with the radio buttons of this channel. Channel 4: The interferometer motor movement (up and down) can be controlled within this channel. Channel 5: Several radio buttons allow a manual frame grabber exposure control.
Motors tab Section Step motors Both of the step motors of the aperture and changer can be controlled manually within this section. Furthermore, the changer steps can be set here.
Fig. 4 - 12: Motors-tab
Test Hardware Pressing this button will start a complete hardware test of the system. Any errors which may occur will be displayed in the window next to it. All error messages are also logged in a file named as “czgif.log” and is located in D:\IOLMaster folder. Permanent Test This button will run a permanent test of the systems hardware. The test can be stopped by clicking the button again. At the end of the test a log window will popup containing information about start and end time and errors.
Service Manual Part No. 1836-541
Confidential and Proprietary
4-8
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
EEPROM tab The EEPROM tab contains a section for the “Serial Number” of the instrument, which accepts 7 digits. In this example, the serial number which is saved will be printed on the speed test printouts. (See 4.1.7).
4.2
Adjustments / Alignments of the Measuring Head Fig. 4 - 13: EEPROM
4.2.1 Optics cleaning Tools and Material required o
Alpha Swab with Long Handle TX 761
part-no.: 000000-0433-973
Fig. 4 - 14: Micro fiber swab
OR o
Cotton Swab with wooden handle
part-no.: 000000-0500-895
Fig. 4 - 15: Cotton swab
o
Optics cleaning mixture “L” (50ml Bottle)
part-no.: 000000-1526-414
Fig. 4 - 16: Optic cleaning mixture SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 9
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Optics Cleaning Procedure Note: The instrument must be switched off for cleaning – do not switch the instrument on before the cleaning mixture has been dried off. Use Carl Zeiss cleaning mixture “L” for cleaning only. This cleaning agent is approved. Be carefully while cleaning components near painted surfaces. Use small quantities at one time – do not “flood” the instrument. Keep the cleaning mixture clean (do not dip the dirty wiper into the mixture bottle). The cleaning should be made in the following sequence: -
Cleaning of the test eye Cleaning of the optical surfaces of the Keratometer assembly (outside and inside) Cleaning of the optical surfaces (inside the head).
1) Cleaning the test eye
Fig. 4 - 17: Cleaning the Test eye
2) Cleaning the optical surfaces of the Keratometer assembly
Fig. 4 - 18: Cleaning Keratometer Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 10
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
3) Cleaning the optical surfaces (inside the head). Note: It is recommended to perform inside the head cleaning with compressed-air, as the inside components normally are only contaminated with dust (no grease contaminants). If using canned compressed air, do not allow liquid propellant to strike optical surfaces, as spotting may result on the surfaces. If liquid cleaning is necessary (e.g. fingerprints), proceed as showed in the following sections.
Beam splitter
Fig. 4 - 19: Beam splitter
a) Top side
b) Bottom side
Interferometer triple prism
Fig. 4 - 20: Triple prism
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 11
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Prism 26
Fig. 4 - 21: Prism 26
a) Backside
b) Bottom side
a) Frontside
b) Backside
Prism 34
Fig. 4 - 22: Prism34
Lens (behind prism 34)
Fig. 4 - 23: lens
Service Manual Part No. 1836-541
a) Frontside
b) Backside
Confidential and Proprietary
4 - 12
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
Magnification changer lenses
Fig. 4 - 24: Changer lenses
a) upper lens
b) lower lens
Camera mirror (small)
Fig. 4 - 25: Camera mirror
Caution: After successfully cleaning the optics, the laser output power must be checked and adjusted. See 4.2.3 and 4.2.4 for detailed procedure. Laser power adjustment also requires adjustment of the laser monitoring.
+++ Perform a function test in accordance with the checklist. +++
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 13
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
4.2.2 Power supply voltage Digital IOL board Procedure: 1) 2) 3) 4)
Remove the six hex socket screws (2.5 mm hex key) from the left cover only. Remove the cover. Switch on the instrument. Measure the supply voltages as Fig. 4 - 26 shows. Correct values will ensure a proper functionality of the measurement head. Note: If the voltage at pin 1 (see Fig. 4 - 26) is too low, an adjustment can be made at the power supply (computer base). See 4.3.6 for details.
´
Service connector X220: X220: X220: X220:
1 +12V 2 + 9V 3 + 5V 4 GND
Fig. 4 - 26: Power supply voltage / digital IOL board
+++ Perform a function test in accordance with the checklist. +++
Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 14
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
4.2.3 Laser power “Adjustment power” Note: After adjustment of the laser power, the laser monitoring must also be re-adjusted. Special Tools: Power meter (such as Coherent Fieldmaster or Ophir Nova) Adequate probe for the wavelength of 780 nm (such as LM2 or PD300) Optional recommended: Measuring adapter 1094-546 for the LM2 measuring head. Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. 3) Enter service mode and start the PIO Test – see 4.1. 4) Set power meter to 780 nm as well as in an adequate range for the expected power of 40 μW. 5) Remove any present filters from the probes (e.g. PD300) and offset the probe for ambient light. 6) Switch off the fixation LED. It is further recommended to switch off also the overview LEDs and KER LEDs and slit groups if the laser measuring head is positioned not directly on the front lens. 7) Select "Adjustment power" in Channel 2 and check the "Laser On/Off" checkbox for “On”. 8) Position the probe close to the front lens and move slightly for the best measurement result. 9) Use “Motor forward” / “Motor backward” in Channel 4 to determine the position with the highest measured power.
Fig. 4 - 27: Laser power adjustment
Caution: If the laser power is determined as too low, ensure that the optical path is clean before adjusting the laser power – see 4.2.1. Note: If the laser power is unstable, the most likely cause is that the laser diode is defective. Caution: The highest reading of the both positions must be adjusted according the default (40 μW). It is absolutely normal that the lower reading differ from the higher. This difference can be up to 5 μW. It is not required that the lower reading must lie in the tolerance range. 10) Use R7 potentiometer on the Digital IOL PCB (see Fig. 4 - 27) to set the adjustment laser power to (40 ± 2) μW.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 15
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Note: If once the laser power has been adjusted, the laser monitoring must be adjusted as well. See 4.2.5. +++ Perform a function test in accordance with the checklist. +++
4.2.4 Laser power "100% power" Note: After adjustment of the laser power, the laser monitoring must also be re-adjusted. Special Tools: Power meter (such as Coherent Fieldmaster or Ophir Nova) Adequate probe for the wavelength of 780 nm (such as LM2 or PD300) Optional recommended: Measuring adapter 1094-546 for the LM2 measuring head. Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. 3) Enter service mode and start the PIO Test – see 4.1. 4) Set the power meter to 780 nm as well as in an adequate range for the expected 100% power of 400 μW. 5) Remove any present filters from the probes (e.g. PD300) and offset the probe for straight light. 6) Select "100% power" in Channel 2 and check the "Laser On/Off" box for “On”. 7) Switch off the fixation LED. It is further recommended to switch off also the overview LEDs and KER LEDs and slit groups if the laser measuring head is positioned not directly on the front lens. Caution: Do not operate the laser diode at "100% Power" for more than 5 minutes.
8) Position the probe close to the front lens and move slightly for the best measurement result. 9) Use “Motor forward” / “Motor backward” in Channel 4 to determine the highest measurement position. Caution: If the laser power is determined as too low, ensure that the optical path is clean before adjusting the laser power – see 4.2.1 Note: If the laser power is unstable, the most likely cause is that the laser diode is defective.
Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 16
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
Caution: The highest reading of the both positions must be adjusted according the default (400 μW). It is absolutely normal that the lower reading differ from the higher. This difference can be up to 50 μW. It is not required that the lower reading must lie in the tolerance range. 10) Use R9 potentiometer on the Digital IOL PCB (see Fig. 4 - 27) to set the adjustment laser power to (400 ± 5) μW. Note: If once the laser power has been adjusted, the laser monitoring must be adjusted as well. See 4.2.5 +++ Perform a function test in accordance with the checklist. +++
4.2.5 Laser Monitoring Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. 3) Enter service mode and start the PIO Test – see 4.1. 4) Switch off the fixation LED in Channel 2. 5) Click on "Laser monitoring" (Channel 0) and "100% Power" (Channel 2); if necessary, click "Laser on/off" to switch the box inactive. 6) Note down displayed value as "Laser-off reading", depends on ambient conditions (typical reading approx. 128) – see Fig. 4 - 28 part a) 7) Switch on the laser ("Laser on/off" checkbox active) and note down measurement result (Channel 0) as "Value 100% Power" (typical reading approx. 48). – see Fig. 4 - 28 part b) Caution: Do not operate the laser diode at "100% Power" for more than 5 minutes!
a)
b)
c)
8) The difference between the measured values ("Value dark signal" minus "Value 100% Power") must be 80 ± 3 increments. 9) In the case of deviations, adjust the R26 potentiometer (at the APD –PCB, see Fig. 4 - 29) accordingly.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 17
Fig. 4 - 28: Laser monitoring
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Caution: The APD board has more than one potentiometer. Ensure that you adjust only R26. A wrong adjustment of any of the other pots can not be corrected in the field. Do not change the settings of the two other potentiometers on the APD PCB. 10) After completion of the potentiometer adjustment, switch to the "Adjustment Power" function and note down the measurement result as "Adjustment power reading (typical reading approx. 120). – see Fig. 4 - 28 part c) 11) The difference between the measured values ("Value dark signal" minus "Value adjustment power") must be 8 ± 2 increments.
Fig. 4 - 29: Monitoring adjustment
+++ Perform a function test in accordance with the checklist. +++
4.2.6 CCD camera sensitivity Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument and measure the sensitivity voltage between any of the groundpins and the pin (P1) below R75 on the Digital IOL PCB. - see Fig. 4 - 30 3) Adjust the sensitivity voltage at DC (2.8 ± 0.3) V
+++ Perform a function test in accordance with the checklist. +++ Fig. 4 - 30: reading points for camera sensitivity
4.2.7 Surrounding field illumination Preconditions: Frame grabber settings according the defaults – see 4.1.3. CCD camera sensitivity adjusted according the specifications (and according the camera generation). Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 18
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only. 2) Switch on the instrument and select overview mode. Note: It is of vital importance, that the “Check, Device” patient has been selected and the Test Eye is in front of the instrument. 3) Focus the Test Eye (surface!) in overview mode, so that the grainy structure appears in focus. 4) Increase the illumination with R3 on the Digital IOL PCB (see Fig. 4 - 31) until the image seems to be too bright (white areas are visible at the Test Eye’s surface). 5) Now reduce the illumination until the surface of the Test Eye is well illuminated and the image is without glare (no white areas). Note: If it is impossible to increase/decrease the surrounding field illumination in a way to get a well illuminated Test Eye surface, the camera sensitivity voltage can be increased/decreased slightly within the tolerance range.
Fig. 4 - 31: Potentiometer for adjustment of Field and KER illumination
+++ Perform a function test in accordance with the checklist. +++
4.2.8 Keratometer LED illumination Preconditions: • • •
Frame grabber settings according the defaults – see 4.1.3. CCD camera sensitivity adjusted according the specs (and according the camera generation). Aperture alignment OK – see 4.2.12.
Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. Fig. 4 - 32: Calibration setting 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 3) Position the Test Eye in front of the instrument and focus the spots, so that also the halo ring will be visible. Note: It is of vital importance, that no external ambient light strikes the instrument or Test Eye while adjusting the illumination.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 19
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
4) Use potentiometer R40 on the Digital IOL PCB (see Fig. 4 - 31) to adjust the brightness of the Keratometer spots in such a way that they appear without glare - with a bright center and a clearly visible, but weaker halo ring. 5) Enter Service Mode, KER-Tab (see Fig. 4 - 32) and uncheck the box for “adjustment aid calibrated” – see 4.1 6) After closing the “Program Setting” dialog with “Save” and change to the Dual mode. You will be prompted to calibrate the Keratometer aid. 7) Position the measure-head so that the spots appear blurry. Ensure that the horizontal and vertical position is OK – the six spots must be visible and appear equal in the cross-hair. 8) Focus the spots by moving the joystick in the direction where the image becomes sharper. Move further in the same direction until the spots appear blurry again before pressing the joystick-button. 9) Now you will be prompted to repeat the process once again for comparison. Please do so. 10) If the calibration process was successful, a message to this effect will be displayed. 11) If a message appears, stating that variance between the calibration values of the two calibrations was too wide, the process (starting at item 8) must be repeated. Fig. 4 - 33: KER-aid calibration 12) After the calibration process, a prompt shows the spot size of the Keratometer-LED’s in focus. The value must lie in a range of 50 – 55. If the value is in range then confirm and calibration procedure would be successful 13) Use R 40 potentiometer to adjust the brightness of the Keratometer spots and repeat the procedure, starting with item 6 until the spot size lies within 50-55. +++ Perform a function test in accordance with the checklist. +++
4.2.9 Interferometer-motor Special Tools: Thermometer (min. range 10°C - 35°C (50°F – 95°F) / accuracy ± 0,5°C / resolution 0,2°C). Recommended thermometer: TESTO “testo 608-H1“ – See 5.3 (Toollist).
Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Enter Service-Mode and run Speed-Test (System – Tab). Run motor about 10 times for „warm up“ and stabilisation of sledge speed – use Æ
Fig. 4 - 34: Speed adjustment Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 20
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
3) Measure the surrounding temperature and determine the corresponding sledge speed which must be adjusted – see diagram below (Fig. 4 - 35). 4) Enter Service-Mode and run Speed-Test (System – Tab) 5) If not already present, change to “Speed-View“, by pressing the “Speed” button. use Æ Adjust first the determined sledge speed in forward direction (up) using potentiometer R64 on the Digital IOL PCB (see Fig. 4 - 34). Adjusting tolerance is ± 0,2 mm/s. 6) Ensure that the displayed speed is reached after 80 ms (first dotted vertical line – black). 74,0
sledge speed [mm/s]
73,0 72,0
71,6 71,0 70,0 69,0
35°C (95°F)
34°C (93,2°F)
33°C (91,4°F)
32°C (89,6°F)
31°C (87,8°F)
30°C (86°F)
29°C (84,2°F)
28°C (82,4°F)
27°C (80,6°F)
25°C (77°F)
26°C (78,8°F)
24°C (75,2°F)
23°C (73,4°F)
22°C (71,6°F)
20°C (68°F)
21°C (69,8°F)
19°C (66,2°F)
18°C (64,4°F)
17°C (62,6°F)
16°C (60,8°F)
15°C (59°F)
14°C (57,2°F)
13°C (55,4°F)
12°C (53,6°F)
11°C (51,8°F)
10°C (50°F)
68,0
temperature [ °C (°F) ]
Fig. 4 - 35: Temperature graph with example
Note: If the determined slide speed can’t be reached after 80ms, the present motor might be hard-steering. In that case the motor must be exchanged! If the problem still persists, the IOLMaster must be sent for in-house repair! 7) Change to “Oscillation-View”, by pressing the Osillation button (see Fig. 4 - 37). Check the maximum and minimum value of the speed: Max value < 81 mm/s Fig. 4 - 36: Speed graph Min value > 61 mm/s Following that, check the speed-oscillation of the motor. If any peak is in the red zones, the slide motor must be re-adjusted / repositioned. While doing the re-positioning, also the motor gear can be manually turned (90°…180°) before placing the motor back in the rail. This may eliminate the peaks in the red area. SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 21
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Note: If the motor cannot be adjusted so that all peaks lie within the white zone, the motor must be replaced! If the problem still persists, the entire measurement head must be replaced, or the instrument must be sent to a CZM repair center for in– house repair. 8) If not already present, change back to „Speed-View“, by pressing the “Speed” button. Afterwards use Æ Adjust the determined sledge speed in Fig. 4 - 37: Speed oszillation backward direction (down) using potentiometer R77 on the Digital IOL PCB (see Fig. 4 - 34). Adjusting tolerance is again ± 0,2 mm/s. 9) Perform the same check as in item 7 described. +++ Perform a function test in accordance with the checklist. +++
4.2.10 ALM Measurement range alignment Preconditions: Interferometer motor speed adjusted – see 4.2.9 Caution: The ALM measurement range alignment must only be performed with IOLMaster Software Version 7 or higher. Software Version 7 has to be installed before the range can be aligned! Please use the Service USB flash drive (Option 2) and perform all necessary steps before. – see 4.3.1 Procedure: 1) Check the IOLMaster software version and install the Version 7 or higher, if necessary. 2) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 3) Position the Test Eye in front of the instrument (ALM sphere), choose “Check, Device” patient and select Dual mode press the key until the ALM mode is reached. Fig. 4 - 38: Measurement range (lower) Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 22
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
4) Mark an individual measurement (not the composite measurement) from the active column of measurements and move the white spot manually over the left end of the red graph using the right mouse button. 5) Mark another individual measurement (not the composite measurement) from the active column of measurements and move the white spot manually over the right end of the red graph using the right mouse button. Note: Do not use the left mouse button because the left button will automatically determine any peak in the area of positioning the white spot. In that case you may not be at the end of the measurement range!
Fig. 4 - 39: Measurement range (upper)
Note: Sometimes, if the white spot extends past the end of the range, instead a logical value (e.g. 13.92 mm), a dash “-“ which symbolizes an “error” or a high numeric value appears. In this case the spot is out of the available range. Slightly move the spot (using the right mouse button) in direction to the inside of the measurement range until an mm-value appears. 6) If needed, adjust the measurement range to ≤ 13.5 mm and ≥ 38.5 mm (lower than 13.5 mm and greater than 38.5 mm). It is not needed to widen the range much more than that. In the case of the lower end (13.5 mm) can’t be reached, you have to adjust the lower sensor (edge detector) of the interferometer - you have to move it upwards. Note: After each adjustment of the sensor, at least two new measurements must be taken before verifying the range-values again. Caution: If the alignment has been completed successfully, please install the Software Version 5.5 again, if it was the previous installed Software Version. – see 4.3.1
Fig. 4 - 40: Sensor alignments
+++ Perform a function test in accordance with the checklist. +++ SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 23
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
4.2.11 Keratometer alignment Preconditions: Frame grabber settings according the defaults – see 4.1.3 CCD camera sensitivity adjusted according the specs (and according the camera generation). Procedure: 1) Remove both head covers – see 3.1.1 2) Cover the optical path, to avoid any light interference. 3) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 4) Position the Test Eye is in front of the instrument and focus the spots, so that also the halo rings will be visible. 5) Loosen the three fixation-screws of the Keratometer unit – see 3.1.7 6) Position the Keratometer is such a way, that the
Fig. 4 - 41: Keratometer alignment
spots appear centred in their surrounding halo rings. The two middle spots must also be parallel to and bisected by the centre line of the crosshair. 7) Recalibrate, if necessary, the KER and ACD – see 4.4.3 and 4.4.2 +++ Perform a function test in accordance with the checklist. +++
4.2.12 Aperture alignment Preconditions: Frame grabber settings according the defaults – see 4.1.3 CCD camera sensitivity adjusted according the specs (and according the camera generation).
Fig. 4 - 42: Bad Aperture alignment
Procedure: 1) Remove both head covers – see 3.1.1. 2) Cover the optical path, to avoid any light interference. 3) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 4) Position the Test Eye is in front of the instrument and focus the spots, so that the halo rings will also be visible. 5) Loosen the two screws of the aperture-assembly.
Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 24
Fig. 4 - 43: Good Aperture alignment
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
6) Position the aperture in a way that no flare (see Fig. 4 - 42 and Fig. 4 - 43) is shown around the Keratometer spots. The halo rings must be clearly visible around the spots. +++ Perform a function test in accordance with the checklist. +++
4.2.13 Slit group alignment Preconditions: Frame grabber settings according the defaults – see 4.1.3. CCD camera sensitivity adjusted according the specs. No CCD camera focus shift present – see 3.1.9 section “Focus shift determination / compensation”. Procedure: 1) Remove both head covers – see 3.1.1. 2) Cover the optical path, to avoid any light interference. 3) Position the Test Eye (ACD Sphere) in front of the instrument. 4) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 5) Run PIO-Test (service-mode). Go to “Chanel 3” of the PIOPorts tab and switch on the slit group which needs to be adjusted. Ensure that the slit image will be projected to the black surface of the Test Eye (left from the ACD sphere). 6) Focus the black surface of the Test eye by moving the instruments head forward/backward.
Fig. 4 - 44: Slit group alignment
Note: Please align only one slit group at same time, otherwise the second group as a reference get lost. 7) Now align the related slit group in a way, that the projected slit becomes as thin as possible. The position of the slit must be in the middle of the crosshair. The inner circle can be seen as the tolerance range for the adjustment. 8) Double check both slit groups. The projected slit images should virtually “stay” in the same position while switching between both slit groups. If necessary also adjust the other slit group, too. SM-30-4054-de Section 4, Revision A, Feb. 2010
Fig. 4 - 45: Slit group adjustment
Confidential and Proprietary
4- 25
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Note: If one or both slit groups have been adjusted, a re-calibration of the ACD might be necessary. – see 4.4.2 +++ Perform a function test in accordance with the checklist. +++
4.2.14 APD alignment Æ The APD alignment must be performed in our in-house repair center.
4.2.15 Magnification changer alignment
Note: There are three possible alignments of the magnification changer – the sensor, the flag and the changer steps (software setting). The alignments, if necessary, should be done in order. Overview: Magnification optics Direction of movement towards sensor Signal of sensor
Stepp cycle of 2-phase stepper motor
011001100 001100110 Signal change of sensor
Magnification optics
Adjusting and measurement position of magnification changer
Magnification optics
Overview position
Measurement position Overview optics Adjustable actuating flag
Fixed actuating flag
Overview optics Fixed actuating flag
Direction of movement towards sensor
Movement direction 290-300 steps
Adjustable actuating flag
Sensor
Sensor
Fig. 4 - 46: Changer – schematic overview
Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 26
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
Sensor Note: The Flag alignment primarily changes the measurement-position. The overview-position will also be affected by this alignment and must be re-aligned afterwards. The sensor must be adjusted according to the following criteria: the position according the optical path (rough positioning) and the position in relation to the changer motor phase (fine positioning) Procedure: 1) Remove both head covers – see 3.1.1. 2) Cover the optical path, to avoid any light interference. 3) Position the Test Eye (ALM Sphere) in front of the instrument. 4) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. Position the measurement head according the sphere. Run PIO-Test (service-mode). Go to “Chanel 2” of the PIO-Ports tab and switch on the laser (100%). Caution: Do not operate the laser diode at "100% Power" for more than 5 minutes.
Fig. 4 - 47: Laser Reflections
5) Change back to ALM mode (click outside the PIO-Test) and position the measurement head in a way, that both, the bright laser spot (primary reflection) and the secondary reflection (see Fig. 4 - 47) are over each other and fix the position of the measurement head. Note: If this position is not roughly aligned with the crosshair, this can be taken as prove that the sensor of the magnification changer must be aligned. 6) Switch between measurement-position and overview-position back and forward. At the moment check only the measurement position. If the position where both laser reflexes lie over each other don’t match with the crosshair position, the sensor must be re-aligned. 7) Loosen the sensor (two screws) and change the position slightly. Depending if the crosshair stays left or right from the laser spot, the sensor must be aligned to the left or right hand side. Note: Whatever will be done at the sensor, do not move the position of the measurement head. It is of vital importance the measurement head keeps positioned as the laser reflections stay over each other.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 27
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
8) Switch again between measurement-position and overview-position back and forward to see the effect of the alignment. Repeat the procedure as long as the crosshair will stay over the laser spot in measuring mode (not overview-mode!). 9) Go to KER-Mode and bring back to the PIO-Test ( + ) and switch the laser back to adjustment power. Change to the Motor-tab and slide the window to bottom of the screen, so that you are able to see the KER spots as well as the control of the changer motor.
Fig. 4 - 48: Changer Steps
Note: You are now in KER mode and should see the spots and the crosshair perfectly aligned. If not, manually move the changer once again (click the checkbox next to “Changer”) to overview- and back to measurement position. If necessary also re-position the head. 10) Click “Step +” several times (at least 8 times). The motor will move the optics (image) now to the left hand side (out of the crosshair). 11) Click “Step –“ until the “End” position switches from 1 to 0 (indicates that the sensor has detected the flag). Continue pressing “Step –“until the “phase” shows 00 and count the number of motor-steps in between of the switch of the end-position and the 00 (The switch itself from 1 to 0 does not count). 12) The accepted range for this “switch-gap” is 2 or 3 steps. Any other number requires a fine adjustment of the sensor position. Loosen the sensor again and slightly move it. Be carefully, the necessary alignment requires a minor positioning only. Note: The fine adjustment may take some time and some attempts. Be accurate with this alignment, because it will ensure the stable positioning of the measurement lens. Remember: All measurements taken by the instrument are related to this positioning! 13) Use the manual control of the changer (PIO-Test / Motor tab) to switch between the measurement-position and overview-position and back to see the effect of the sensor-alignment. Always check the alignment according the laser spots before checking the fine-alignment according the steps (switch-gap). 14) If the sensor is successfully aligned, finally tighten the screws. The sensor doesn’t need to be aligned for the overview position of the magnification changer. +++ Perform a function test in accordance with the checklist. +++ Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 28
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
Flag Note: The Flag alignment will only change the overview-position. The measurement position will not affected by this alignment. Procedure: 1) Remove both head covers – see 3.1.1 2) Cover the optical path, to avoid any light interference. 3) Position the Test Eye (ALM Sphere) in front of the instrument. 4) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. Position the measurement head according the sphere. Run PIO-Test (service-mode). – see 4.1 Change to the Motor-tab and slide the window to bottom of the screen, so that you are able to see the KER spots as well as the control of the changer motor. See Fig. 4 - 48. Note: You are now in KER mode and should see the spots and the crosshair perfectly aligned. If not, manually move the changer once again (click the checkbox next to changer) to overview and back to measurement position. If necessary also re-position the instruments head. 5) Manually move the changer now (click the checkbox next to “Changer”) to the overview-position. Note: Depending upon the step-information (setting in the “Steps” window), the position of the crosshair according the image might be correct or not. Due to this “offset” will be adjusted later, it’s not necessary to be correct for the flag-alignment. Anyhow, do not move the position of the instruments head. It is of vital importance the instruments head keeps positioned according the previously aligned measurement position. 6) Use “Step +“/ “Step –“to manual position the crosshair according the image. 7) If in this position the “End” position indicates 1, the flag must be roughly adjusted first. Loosen the screws and move the flag to the left. If the “end position” indicates 0, skip to section 9. 8) Move the changer manually to the measurement-position and back to overview to see the effect of the alignment. Use again “Step +“/ “Step –“to manual position the crosshair according the image. If still the “end position” shows 1, the flag must be adjusted again. 9) Now click “Step –” several times (at least 8 times). The motor will move the optics (image) now to the right hand side (out of the crosshair). The “end position” should switch from 0 to 1 somewhere in between. 10) Now, click “Step +“as long as the “end position” switches from 1 to 0 (indicates that the sensor has detected the [adjustable] flag). Continue pressing “Step +“until the crosshair matches with the image and count the number of motor-steps in between of the switch of the end-position and the position when the crosshair matches (The switch itself from 1 to 0 does not count). 11) The accepted range for this “switch-gap” is 2 or 3 steps. Any other number requires a fine adjustment of the flag position. Loosen the flag and slightly move it. Be carefully, the necessary alignment requires a minor positioning only.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 29
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Note: The fine adjustment may take some time and some attempts. Be accurate with this alignment, because it will ensure the stable positioning of the overview lens. +++ Perform a function test in accordance with the checklist. +++
Changer Steps Note: The changer steps will ensure a good overview-position only in combination with a well aligned sensor and flag. It is vital to check (or if necessary align) the sensor and flag before setting the changer steps. Procedure: 1) Position the Test Eye (ALM Sphere) is in front of the instrument. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. Position the measurement head according the sphere. Run PIO-Test (service-mode). – see 4.1 Change to the Motor-tab and slide the window to bottom of the screen, so that you are able to see the KER spots as well as the control of the changer motor. See Fig. 4 - 48 Note: You are now in KER mode and should see the spots and the crosshair perfectly aligned. If not, manually move the changer once again (click the checkbox next to changer) to overview and back to measurement position. If necessary also re-position the instruments head. 3) Manually move the changer now (click the checkbox next to “Changer”) to the overview-position. Listen to the sound of the changer motor. Note: If the motor stops in between very shortly but continues straight away in direction (or exactly) to the overview-position, the number of steps is too less. Most of the instruments have their setting around the 295. It is recommended to set this as a default, before continuing with the adjustment. 4) If, in overview, with this default the crosshair stands too far right of the image, the number of steps must be increased. If it stands too far left of the image, the number of steps must be decreased. Note: Setting (changing) of the changer steps is only possible in overview-mode. The Set button won’t work in measurement mode.
+++ Perform a function test in accordance with the checklist. +++
Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 30
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
4.3
Adjustment and Calibration
Adjustments / Alignments of the computer
4.3.1 Re-imaging Re-imaging options The service USB flash drive menu provides the following options: 1) “Recover Windows XP on system partition from USB-Stick” Note: This option will delete all data on the C:\ partition. Customized data, such as printer drivers, regional settings, computer name, network settings and installed software options will be lost after the re-imaging. 2) “Recover Windows XP including all partitions from USB-Stick” Note: This option will delete all data on the C:\ and D:\ partition. Customized data, such as printer drivers, regional settings, computer name, network settings and installed software options will be lost after the re-imaging. Furthermore all lens, user and patient data will be lost as well as the IOLMaster program settings, such as refractive index, clinic name, language, export settings and so on. Please be prepared for redoing all customized data (C:\ partition) after the re-imaging process. Lens and user data should be available from the customer’s backup If not, these data can be backed up before the re-imaging (as long as the HDD and the related database are not corrupt). Patient data (visits) can NOT be restored after re-imaging. 3) “Recover system backup from local drive” Note: With this option the C:\ partition will be restored from an image normally existing at the D:\ partition (D:\IOLMSYS.GHO). All data on the C:\ partition will be replaced by this from the image file. Customized data, such as printer drivers, regional settings, computer name, network settings and installed software options will be restored as at the time as the image was created. Please be prepared for redoing some customized data if they were not correctly restored (in case, they have not been included once in the image-backup). 4) “Create system backup onto local drive” Note: By choosing this option, the C:\ partition will be backed up into an image file at the D:\ partition (D:\IOLMSYS.GHO). This backup should always been made after changes into the C:\ partition (system) such as printer drivers, regional settings, computer name, network settings, installed software options and so on.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 31
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Re-imaging procedure Note: Before performing the HDD re-imaging, use the following checklist to ensure the completion after the re-imaging. Printer drivers (Option 1, 2, maybe also 3) Software option installation media (Option 1, 2, maybe also 3) Backup of lens and user data (Option 2) Program setting information (Option 2)
Customer property – only approved printers are allowed to be installed. Customer property (Option N, Option T) Customers backup USB flash drive According customers requests
1) Switch off the instrument and switch on again while pressing F2 alternately (do NOT keep pressing) until the BIOS program is started. 2) Enter the BIOS password and change the boot-sequence for “USB HDD” as first priority. Connect the right service USB flash drive (see Tool list 5.3), save settings and quit BIOS (F10). 3) The instrument automatically reboots now from the service USB flash drive. Please make your selection according the needs. While imaging an empty HDD, “Option 2” must be chosen. 4) If option 2 has been chosen, a prompt will appear, asking for the right partition sizes. Check carefully that the full space of the HDD will be used depending upon the size. 5) HDD will be re-imaged now. When finished, please disconnect the USB flash drive and reboot twice. If the re-imaging has been made from the well-kept local system-backup (Option 3) the following steps might not be necessary. Please continue with item 9).
Fig. 4 - 49: Imaging menu
Fig. 4 - 50: Imaging process
Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 32
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
Installation of Software Version 5.5 Note: The installation of Software Version 5.5 is only mandatory, if the reimaging procedure Option 1 or Option 2 has been performed. 1) After either Option 1 or Option 2 always restart the instrument 2) Connect the IOLMaster Software Version 5.5 USB flash drive (see Tool list 5.3) with the IOLMaster 3) Enter the Settings menu, click on the wrench symbol ¼ 4) Click on the Program tab 5) Click on the Update button (see Fig. 4 - 51) 6) Choose the appropriate language 7) Follow the instructions on the screen 8) If a password prompt appears, enter the IOLMaster XP service password and continue with OK (see Fig. 4 - 53) 9) After a successful installation the system will be restarted.
Fig. 4 - 51: Update button with SW Vers. 7
Fig. 4 - 52: Update button with SW Vers. 5.5
Fig. 4 - 53: Password SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 33
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Settings after re-imaging 1) Set “Date & Time” as well as “Time Zone” according to the customers needs. Æ “Options” Æ “Setup” “Date/Time”
Fig. 4 - 54: Date and Time
2) Set regional settings, such as language and keyboard according to the customer and type of the keyboard used with the instrument. Æ “Options“ “Setup” Æ “Regional Settings” Æ “Languages” tab Æ “Details”
Fig. 4 - 55: Input language
3) Re-install all options (N / T) which are the customer’s property for the related instrument. The option USB flash drive should be provided by the customer. Æ “Options“ “Setup” Æ “SW Option” “Install” Follow the instructions on the screen. The installed options are displayed after installation (at the system information page).
Fig. 4 - 56: Options
4) Append the serial number to the existing computer name “iolm”. Æ “Options“ Æ “Service“ (password required) Æ “System“ tab Æ “Control Panel“ Æ “System“ Æ “Computer Name“ tab Æ “Change“ Æ Format “IOLMSerial-Number“ (e.g. IOLM987654) Æ “OK“ Fig. 4 - 57: Computer name Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 34
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
5) Re-install the printer. Use only printer according the approved specifications! Unapproved printers are not supported by Carl Zeiss Meditec and may lead into serious problems using the instrument. Æ “Options“ “Setup” Æ “Printer” Note: See user manual for detailed installation of network-printers.
6) Restore lens and user data (Option 2 only) Æ “Options“ Æ “Lens database” Æ Administrator logon required Æ “Restore”
Fig. 4 - 58: Restore lens data
7) Program settings (Option 2 only). Configure the application settings according the customers needs. Æ “Options“ Æ “Setup” Æ “Program Settings” Customize the settings of “Program”, “Export”, “Network” and “User management” tabs as necessary. 8) Create / update local backup-image. Reboot the instrument with the service USB flash drive. The menu shown in Fig. 4 - 49 will appear again. Select option 4 for creating a new image from the C:\ partition. This image will be saved at the D:\ partition (D:\IOLMSYS.GHO)
Fig. 4 - 59: Program settings
9) IMPORTANT: After the image has been successfully created, please enter the BIOS again to change back the boot sequence. Restart the instrument while pressing F2 alternately (do NOT keep pressing) until the BIOS program is started. Enter the password and change boot-sequence for “Hard Drive” as first priority. Save settings and quit BIOS (F10). 10) Remove the service USB flash drive. +++ Perform a function test in accordance with the checklist. +++ SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 35
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
4.3.2 HDD Structure (Databases and Logfiles) IOLMaster : Patient Database: Patient visits User and lens Database Keratometer logfile ALM logfile Error message logfile
D:\DATA\CZM_BaseDB.MDB D:\DATA\ *.iol D:\DATA\ IolmXP.MDB D:\IOLMaster\Keratometer.log D:\IOLMaster\Alm.log D:\IOLMaster\czgif.log
4.3.3 Database Exchange Procedure: 1) Enter service mode / system-tab and run Windows® Explorer. 2) Browse to the D:\ partition. 3) Copy the whole DATA folder to a media (USB flash drive, USB HDD). 4) Now, enter service mode / system-tab and run Windows® Explorer of that instrument where the copied data folder should be replaced. 5) Rename the existing DATA folder (“D:\” partition) of the related instrument to DATAOLD. Note: Without renaming the existing DATA folder, the paste of the backed up DATA folder won’t work. This is because opened files can’t be replaced while they are open. 6) Now copy the backed up DATA folder to the “D:\” partition of the instrument. 7) Reboot the instrument and check the patient data, the user data and lens data. 8) If the exchange was successfully, the DATAOLD folder can be deleted. +++ Perform a function test in accordance with the checklist. +++
4.3.4 Deleting of corrupted Databases 1) Enter service mode / system-tab and run Windows® Explorer – see 4.1 2) Browse to the D:\Tools folder and run “taskmgr”. 3) End the IOLMaster task. 4) Browse back to the D:\Data folder and delete the related database (CZM_BaseDB.MDB and/or IolmXP.MDB). Note: All data will be finally lost if deleting a database. The automatically recreated database will be empty. 5) Also delete the related database (CZM_BaseDB.MDB and/or IolmXP.MDB) in the D:\Data\Save folder. 6) After rebooting the instrument, the deleted database will be automatically re-created (empty!) +++ Perform a function test in accordance with the checklist. +++ Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 36
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
4.3.5 Deleting the administrator-password Procedure: 1) Enter Service mode (system tab). Click “Erase Data” and check the “administrator password” checkbox (see 4.1.6). Confirm with OK. +++ Perform a function test in accordance with the checklist. +++
4.3.6 Update the Software Procedure: 1) Switch on the instrument. 2) Connect the USB flash drive containing the adequate update software 3) Click on “Options” “Setup” “Update” and follow the instruction on the screen – See Fig. 4 - 60. +++ Perform a function test in accordance with the checklist. +++
4.3.7 Power supply Voltages
Fig. 4 - 60: Software Update
Procedure: 4) Open the computer base – See 3.2.1 5) Measure the voltage output at the recommended points for the power supply – See Fig. 4 - 61. 6) The output voltage range must be DC (+12 ± 0.1) V. 7) Adjust the voltage with a long and appropriate screw driver at the related potentiometer Vadj – See Fig. 4 - 61. Caution: Be careful with the screw driver and do not damage parts closeby the pot. There is no high voltage within this region, but there is still the risk of electric shock. SM-30-4054-de Section 4, Revision A, Feb. 2010
GND
Vadj +12V
Fig. 4 - 61: Power supply IOLMaster
Confidential and Proprietary
4- 37
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
Note: It is recommended to check the supply voltage at the Digital IOL board, too. This will ensure a proper functionality of the measurement head. See 4.2.2 for details. +++ Perform a function test in accordance with the checklist. +++
4.4
Calibration Note: No other changes as described are permitted to be performed. Changes of other values may lead into wrong measurements.
4.4.1 ALM calibration Procedure: Note: The re-calibration must be done only if the readings of the ALM are out of the tolerance range (± 0.05 mm) of the test-eyes imprinted value. 1) 2) 3) 4)
Install Test Eye (ALM sphere). Take some ALM measurements. Enter Service mode (calibration password), ALM tab – see 4.1 Run the calibration spread-sheet at an external PC (Microsoft® Excel required) and perform the recalibration as Fig. 4 - 62 shows.
Fig. 4 - 62: ALM calibration Service Manual Part No. 1836-541
Confidential and Proprietary
4 - 38
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
5) After determining the new Origin Offset, enter it as the Origin Offset value on the ALM tab. The new ALM calibration Origin Offset value will be saved in the EEPROM and iol.ini-file by clicking OK to leaving the service-mode. 6) Verify the calibration by measuring the test-eye again (ALM measurement). Note: A new patient (visit) must be created to take new readings. If the same patient (as with the old calibration values) would be used, the average reading or composite signal would not give the correct output according the new calibration values. Note: If the deviation between the Test Eye value and the measurement result was quite big, the calibration process (Excel calculation) must be repeated several times. +++ Perform a function test in accordance with the checklist. +++
4.4.2 ACD calibration Preconditions:
Clean Test eye sphere without scratches. Optical path of ACD clean. Frame grabber settings OK (see 4.1.3). Ensure that no ambient light strikes the test eye or enters the exit lens of the IOLMaster head. Well adjusted slit modules (see 4.2.13). CCD camera sensitivity voltage well adjusted (see 4.2.6).
Procedure: 1) Install Test Eye (ACD sphere). 2) Take ACD measurements (with “Check, Device” patient). 3) Enter Service mode (calibration password), ACD tab - see 4.1 4) Run the calibration spread-sheet at an external PC (Microsoft® Excel required). Note: The individual conversion factors for right and left side are always viewed from the patient and not from the user. Do not mix the values up.
Fig. 4 - 63: ACD calibration
5) Enter the values as requested: Test Eye Value is the value of the test eye which is printed on the test eye label as VKT Measurement is the value that the instrument reports for the measured ACD. Conversion Old is taken from the Service Menu ACD tab (before calibration). Conversion New is the value to correct the ACD calibration.
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 39
Service Manual Part No. 1836-541
Adjustment and Calibration
IOLMaster 5.5
6) Enter the calculated conversion factor into the conversion section of the ACD tab. 7) Verify the calibration by taking new measurements. If necessary repeat the calibration process (Excel calculation). +++ Perform a function test in accordance with the checklist. +++
4.4.3 Keratometer calibration Preconditions:
Clean Test eye sphere without scratches. Optical path of KER clean. Frame grabber settings OK (see 4.1.3). Ensure that no ambient light strikes the test eye or enters the exit lens of the IOLMaster head. Well adjusted Keratometer assembly (see 4.2.11). Well adjusted Aperture assembly (see 4.2.11). CCD camera sensitivity voltage well adjusted (see 4.2.6). Keratometer LEDs illumination well adjusted (see 4.2.8).
Procedure: 1) Enter Service mode (calibration password), KER tab – see 4.1 2) Enter the default values in the conversion section and the section for the factors (0°, 60°, and 120°). They are 0.035 / 1 / 1 / 1 (see Fig. 4 - 64) 3) Save the values by leaving the service mode with “OK” 4) Install Test Eye (KER sphere). 5) Take KER measurements. 6) Enter the service mode again, run Microsoft® Explorer (System-Tab) – see 4.1
Fig. 4 - 64: Keratometer calibration
7) Browse to the IOLMaster folder. 8) Open “Keratometer.log” and scroll down to the last block of the log file. 9) Run the calibration spread-sheet at an external PC (Microsoft® Excel required). 10) Enter 5 values each for Kappa 1 / 2 / 3 (from the log file) into the calibration spread-sheet fields highlighted in gray. 11) Enter now the two average values for radius R1 and R2 (from the log file) into the calibration spread-sheet fields highlighted also in gray.
Service Manual Part No. 1836-541
Fig. 4 - 65: Keratometer calibration spread-sheet
Confidential and Proprietary
4 - 40
SM-30-4054-en Section 4, Revision A, Feb. 2010
IOLMaster 5.5
Adjustment and Calibration
12) Enter the imprinted radius value from the Test Eye as the “Radius Test Eye” in the calibration spread-sheet. 13) Open service mode, KER tab and replace the default values with those calculated from the calibration spread-sheet (highlighted in green). 14) Verify the calibration by taking new measurements. If necessary repeat the calibration process (Excel calculation). +++ Perform a function test in accordance with the checklist. +++
4.4.4 WTW calibration Preconditions: White-to-white measurement activated (“Settings” “Measurement” tab “Activate WTW” checkbox) Clean Test eye sphere without scratches, WTW calibration scale attached. Optical path of WTW clean Frame grabber settings OK (see 4.1.3) Ensure that no ambient light strikes the test eye or enters the exit lens of the IOLMaster head. CCD camera sensitivity voltage well adjusted (see 4.2.6). Surrounding field illumination well adjusted (see 4.2.7).
Procedure: 1) Enter Service mode (calibration password), WTW tab. 2) Uncheck the “calibrated” checkbox. 3) Install Test Eye (WTW scale must face to the instrument). 4) Save setting by leaving the service mode with “OK”. 5) Chose WTW – the calibration prompt appears. 6) Position the head in a way to see the full calibration scale in the live image. Focus the scale and press joystick button.
Fig. 4 - 66: WTW calibration
7) Repeat the focusing as prompted once more and press joystick button again. Note: If the calibration has been accepted, the “calibration” checkbox is automatically checked from the system. This procedure must not be made manually. +++ Perform a function test in accordance with the checklist. +++
4.4.5 Spot size calibration Æ See 4.2.8 Keratometer LED illumination
SM-30-4054-de Section 4, Revision A, Feb. 2010
Confidential and Proprietary
4- 41
Service Manual Part No. 1836-541
Section 5 – Troubleshooting 5.1 Troubleshooting Image ..................................................................................................3 5.1.1 General image troubleshooting ....................................................................................... 3 5.1.2 Detailed troubleshooting I (KER – image) ......................................................................... 4 5.1.3 Detailed troubleshooting II (ACD – image) ....................................................................... 5 5.1.4 Detailed troubleshooting III (WTW – image)..................................................................... 6 5.2 Error messages...............................................................................................................7 5.2.1 Error messages during powering up ................................................................................ 7 5.2.2 Error messages during operation ................................................................................... 11 5.2.3 Typical error messages and causes therefore ................................................................. 13 5.3 Tool list IOLMaster .......................................................................................................14
SM-30-4054-en Section 5, Revision A, Feb. 2010
Confidential and Proprietary
5-1
Service Manual Part No. 1836-541
Troubleshooting
IOLMaster 5.5
Notes:
Service Manual Part No. 1836-541
Confidential and Proprietary
5-2
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
5.1
Troubleshooting
Troubleshooting Image
5.1.1 General image troubleshooting Preconditions: - Instrument switched on - Boot procedure complete successfully - No error messages - Test Eye installed and “Test Eye“ Mode active or “Check,Device“ patient choosen
OK
Is Patient Manager visible?
No
Control cable maybe improperly connected. A general troubleshoot for the computerbase would be neccessary.
No
Check the connections of the related cables. If necessary, a slightly warm up of the camera may help. (See 3.1.8)
No
Frame grabber settings must be adjusted. (See 4.1.3)
Yes
Is the live camera image in Overview mode present?
Yes
Are the frame-grabber settings according the defaults?
Yes
Is the CCD camera sensitivity voltage according the defaults (2.9 V DC)?
No
Adjust the sensitivity voltage. (See 4.2.6)
No
Adjust the brightness of the surrounding field illumination LEDs. (See 4.2.7)
Yes
Is the surrounding field illumination well adjusted? (“Test Eye“ Mode active or “Check,Device“ patient choosen)
Yes
Are all optical components clean and free of dust?
No
Clean the optics. (See 4.2.1)
Yes
Replace the camera. (See 3.1.8)
No
Instrument head exchange or in-house repair necessary (prism 26 must be replaced)
Yes
Are there signs of a “worn out“ CCD camera?
No
Coating of prism 26 OK?
Continue with detailed troubleshoot of KER image
SM-30-4054-en Section 5, Revision A, Feb. 2010
Continue with detailed troubleshoot of ACD image
Confidential and Proprietary
5- 3
Continue with detailed troubleshoot of WZW image
Service Manual Part No. 1836-541
Troubleshooting
IOLMaster 5.5
5.1.2 Detailed troubleshooting I (KER – image) Detailed troubleshoot of KER image
Are ambient lights too bright?
Yes
Avoid direct or indirect light interference. Dim room lights.
No
Replace Keratometer assembly. (See 3.1.6)
No
Adjustment of Keratometer LEDs intensity. (See 4.2.8)
No
Adjustment of the aperture. (See 4.2.12)
No
Instrument head exchange or in-house repair necessary (prism 26 must be replaced)
No
Are all KER LEDs working?
Yes
Are KER LEDs intensity (spot size) adjusted?
Yes
Does the aperture move in and out of position correctly?
Yes
Coating of prism 26 OK?
Yes
Further troubleshoot by the next Carl Zeiss Meditec in-house repair center. Service Manual Part No. 1836-541
Confidential and Proprietary
5-4
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
Troubleshooting
5.1.3 Detailed troubleshooting II (ACD – image)
SM-30-4054-en Section 5, Revision A, Feb. 2010
Confidential and Proprietary
5- 5
Service Manual Part No. 1836-541
Troubleshooting
IOLMaster 5.5
5.1.4 Detailed troubleshooting III (WTW – image)
Service Manual Part No. 1836-541
Confidential and Proprietary
5-6
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
5.2
Troubleshooting
Error messages
5.2.1 Error messages during powering up Error message (plain text) Changer: coupler always on.
Possible Causes
Error Correction
The signal of the changer light barrier always shows "0" (on). - Defective light barrier. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective motherboard MED-COM. - Defective control cable connection. Changer: coupler The signal of the changer light not found. barrier always shows "1" (off). - Defective light barrier or connection interrupted. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective control cable connection. - Defective motherboard MED-COM. Changer: - Faulty alignment of changer. step error in forward - Defective stepper motor. direction - Stiff motion of changer. Changer: - Faulty alignment of changer. step error in - Defective stepper motor. backward direction - Stiff motion of changer. Aperture: coupler The signal of the aperture light always on. barrier always shows "0" (on). - Defective light barrier. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective motherboard MED-COM. - Defective control cable connection. Aperture: step error
- Defective stepper motor. - Stiff motion of aperture.
SM-30-4054-en Section 5, Revision A, Feb. 2010
- Change light barrier, see 3.1.15 - Change Digital IOL PCB or stepper motor - Change Digital IOL PCB, see 3.1.10 - Change motherboard MED-COM, see 3.2.7 - Repair or change control cable, see 3.1.4
- Change light barrier, repair connection, see 3.1.15 - Change Digital IOL PCB or stepper motor - Change Digital IOL PCB, see 3.1.10 - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7. -
Align changer, see 4.2.15 Change stepper motor, see 3.1.14 Repair changer, see 3.1.15 Align changer, see 4.2.15 Change stepper motor, see 3.1.14 Repair changer, see 3.1.15
- Change light barrier, see 3.1.15 - Change Digital IOL PCB or stepper motor. - Change Digital IOL PCB, see 3.1.10 - Change motherboard MED-COM, see 3.2.7. - Repair or change control cable, see 3.1.4 - Change stepper motor, see 3.1.12 - Repair aperture, see 3.1.12
Confidential and Proprietary
5- 7
Service Manual Part No. 1836-541
Troubleshooting
Error message (plain text) Aperture: coupler not found.
IOLMaster 5.5
Possible Causes
Error Correction
The signal of the aperture light barrier always is "1" (off). - Defective light barrier or connection interrupted. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective control cable connection. - Defective motherboard MED-COM. Slider: coupler does The signal of the lower light barrier not switch. on the prism slider always is "1" (off). No AL measurement - Defective light barrier or possible, prism slider connection interrupted. does not move - Defective Digital IOL PCB. Interferometer does not run up. - Defective motor or connection.
- Change light barrier or repair connection, see 3.1.13 - Change Digital IOL PCB or stepper motor. - Change Digital IOL PCB, see 3.1.10 - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7.
- Change light barrier or repair connection, see 3.1.17 - Change Digital IOL PCB, see 3.1.10 - Change motor, repair connection, see 3.1.16 - Change Digital IOL PCB, see 3.1.10
- Defective Digital IOL PCB. The signal of the lower light barrier always is "1". - Defective light barrier. - Change lower light barrier (right) , see 3.1.17 - Defective Digital IOL PCB. - Change Digital IOL PCB, see 3.1.10 Slider: upper coupler The signal of the upper light barrier not found. always is "1". - Defective light barrier. - Change upper light barrier (left) , see 3.1.17 - Defective Digital IOL PCB. - Change Digital IOL PCB, see 3.1.10 LIE4: Motor of prism slider does not run. no zero pulse - Defective motor or connection. - Change interferometer motor, see 3.1.16 - Change Digital IOL PCB, see 3.1.10 backward - Defective Digital IOL PCB. Slider: bottom coupler not found.
Motor of prism slider runs. - Defective Digital IOL PCB. - Defective encoder or connection. - Defective control cable connection. - Defective motherboard MEDCOM.
Service Manual Part No. 1836-541
- Change Digital IOL PCB, see 3.1.10 - Change measuring head or repair connection. - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7.
Confidential and Proprietary
5-8
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
Troubleshooting
Error message Possible Causes (plain text) LIE4: - Incorrect motor speed. measurement length - Defective interferometer. backward too short - Stiff motion of prism slider/motor - Defective Digital IOL PCB. - Upper or lower light barrier out of alignment. LIE4: no zero pulse forward
Laser: adjustment power too high.
Laser: 100% power too low.
- Adjust speed, see 4.2.9 - Change measuring head, see 3.1.3 - Change measuring head or change interferometer motor - Change Digital IOL PCB, see 3.1.10 - Move light barrier concerned as far upward or downward as possible, increasing the path for the slider as far as possible.
Motor of prism slider does not run. - Defective motor or connection. - Change motor, see 3.1.16 - Defective Digital IOL PCB. - Change Digital IOL PCB, see 3.1.10
Motor of prism slider runs. - Defective Digital IOL PCB. - Defective encoder or connection. - Defective control cable connection. - Defective motherboard MED-COM. LIE4: - Defective interferometer. measurement length - Stiff motion of prism forward too short slider/motor - Incorrect motor speed. - Defective Digital IOL PCB. - Upper or lower light barrier out of alignment. Laser: adjustment power too low.
Error Correction
- Faulty adjustment of laser power - Defective reflecting diode. - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode. - Faulty adjustment of laser power - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode. - Faulty adjustment of laser power - Defective reflecting diode. - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode.
SM-30-4054-en Section 5, Revision A, Feb. 2010
- Change Digital IOL PCB, see 3.1.10 - Change measuring head or repair connection. - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7. - Change measuring head, see 3.1.3 - Change measuring head or change interferometer motor. - Adjust speed, see 4.2.9 - Change Digital IOL PCB, see 3.1.10 - Move light barrier concerned as far upward or downward as possible, increasing the path for the slider as far as possible. - Adjust laser power, see 4.2.3 - Change measuring head, see 3.1.3 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3 - Adjust laser power, see 4.2.3 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3 - Adjust laser power, see 4.2.4 - Change measuring head, see 3.1.3 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3
Confidential and Proprietary
5- 9
Service Manual Part No. 1836-541
Troubleshooting
Error message (plain text) Laser: 100% power too high.
No laser switch off. + 12 V too low
+ 12 V too high
+ 5 V too low. + 5 V too high - 5 V too low. - 5 V too high. - 12 V too low. - 12 V too high. +12V: too low +5V: too low -5V: too high -12V: too high
IOLMaster 5.5
Possible Causes
Error Correction
- Faulty adjustment of laser power - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode. - Defective Digital IOL PCB. - Defective APD PCB. - Defective power supply unit. - Defective APD PCB. - Defective motherboard MED-COM. - Defective control cable connection. - Defective Digital IOL PCB. - Defective power supply unit. - Defective APD PCB. - Defective measuring head/PC connection - Defective APD PCB. - Defective motherboard MED-COM. - Defective measuring head/PC connection - Defective power supply unit. - Defective APD PCB.
- Defective motherboard MED-COM. - Defective measuring head/PC connection Power limit reached. Laser power too high Please call service. - Incorrect adjustment of laser power - Defective Digital IOL PCB. - Defective APD PCB.
Changer: step error in forward direction +12 V: too low + 5 V: too low - 5 V too high. - 12 V too high.
Service Manual Part No. 1836-541
- Defective laser diode. - Incorrect adjustment of laser monitoring - Defective contol cable - Defective control cable connector - Defective motherboard MED-COM
- Adjust laser power, see 4.2.4 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3 - Change Digital IOL PCB, see 3.1.10 - Change head or send instrument to next in-house repair to change the APD PCB. - Change power supply unit, see 3.2.4 - Change head or send instrument to next in-house repair to change the APD PCB. - Change motherboard MED-COM, see 3.2.7. - Repair or change control cable, see 3.1.4 - Change Digital IOL PCB, see 3.1.10 - Change power supply unit, see 3.2.4 - Change head or send instrument to next in-house repair to change the APD PCB. - Repair connection. - Change head or send instrument to next in-house repair to change the APD PCB. - Change motherboard MED-COM, see 3.2.7. - Repair connection. - Change power supply unit, see 3.2.4 - Change head or send instrument to next in-house repair to change the APD PCB. - Change motherboard MED-COM, see 3.2.7. - Check all connections - Perform adjustment. - Change Digital IOL PCB, see 3.1.10 - Change head or send instrument to next in-house repair to change the APD PCB. - Change measuring head - Adjusting laser monitoring - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7 - Change motherboard MED-COM, see 3.2.7
Confidential and Proprietary
5 - 10
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
Troubleshooting
5.2.2 Error messages during operation Error message (plain text) ERR: no image possible
ERR: 0x%x laser adjustment power too high.
Possible Causes
Error Correction
Image recording could not be started - Defective motherboard MEDCOM (frame grabber). - Problems in IOL Master Software. Laser power too high - Incorrectly set laser power of 40 μW - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.
-
Change motherboard MED-COM, see 3.2.7. Perform IOL Master Recovery/Re-imaging.
-
Set laser power
-
Set laser monitoring
-
-
Defective laser diode. Laser power too low. - Incorrectly set laser power of 40 μW. - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.
-
Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3
-
Set laser power
-
Set laser monitoring
-
Defective laser diode. Laser power too high during measurement. - Incorrectly set laser power of 400 μW. - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.
-
Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3
-
Set laser power
-
Set laser monitoring
-
Defective laser diode. ERR: Laser power too low during 0x%x laser measurement. measurement power - Incorrectly set laser power of 400 μW. too low. - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.
-
Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3
-
Set laser power
-
Set laser monitoring
-
Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3
-
ERR: 0x%x laser adjustment power too low.
-
ERR: 0x%x laser adjustment power too high, measurement interrupted.
-
-
Defective laser diode.
SM-30-4054-en Section 5, Revision A, Feb. 2010
-
-
-
-
Confidential and Proprietary
5- 11
Service Manual Part No. 1836-541
Troubleshooting
IOLMaster 5.5
Error message Possible Causes (plain text) ERR: Laser power much too high Power limit reached. (end of AD conversion range Please call service. reached) - Defective Digital IOL PCB. - Defective APD PCB.
Error Correction
- Change Digital IOL PCB, see 3.1.10 - Change head or send instrument to next inhouse repair to exchange the APD PCB. - Change measuring head, see 3.1.3
- Defective laser diode. Zero pulse of LIE4 path0x00000xxx , no measuring encoder is missing. zero pulse - Defective path-measuring - Change measuring head, see 3.1.3 encoder. - Change Digital IOL PCB, see 3.1.10 (xxx stands for the - Defective Digital IOL PCB. - Repair or change control cable, see 3.1.4 relevant error code) - Connection in control cable interrupted. - Defective motherboard MED- - Change motherboard MED-COM, see 3.2.7 COM. 0x00000180 // images could not be grabbed 0x00000181 // No zero pulse during AL measurement detected 0x00000182 // Data-file could not be read 0x00000183 // Incorrect distance between zero-pulses
Service Manual Part No. 1836-541
Confidential and Proprietary
5 - 12
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
Troubleshooting
5.2.3 Typical error messages and causes therefore Error message (plain text) changer: step error in forward direction +12V: too low +5V: too low -5V: too high -12V: too high
Possible Causes - Incorrect Control-cable (connection!) - Defect motherboard MED-COM
additional "krrrrrrr" sound from the changer - changer-sensor defect / not well plugged without "krrrrrr" sound changer: optocoupler always on from the changer changer: step error in forward - changer-motor defect / direction not well plugged - sensor aperture defect / not well plugged aperture: optocoupler not found - motor aperture defect / not well plugged - lower sensor Interferometer defect / not well plugged sledge: optocoupler don't switch - motor Interferometer defect / not well plugged System error! No measurement AL measurement possible – possible. but no result laser: adjustment power too low - APD defect / not well laser: 100% power too low plugged +12V: too low +5V: too low -5V: too high -12V: too high AL measurement shows LIE4: no zero pulse backward laser: adjustment power too low "ERR: 0x181, no zero pulse. : 0" laser: 100% power too low - motherboard MED-COM +12V: too low defect +5V: too low -5V: too low -12V: too low Difference during determination of the increments (Interferometer-scale). ERR: 0x183, measurement system "99" is the offset as a nu: 99 meric value. - damaged scale or scale reader AL-graph looks like a sine wave or - defective laser diode otherwise abnormal Laser-Power (400 μW) unstable or - defective laser diode unable to adjust changer: optocoupler not found changer: step error in forward direction
SM-30-4054-en Section 5, Revision A, Feb. 2010
Confidential and Proprietary
5- 13
Error Correction - Check the control cable connection - Change motherboard MED-COM
- Check connection or change sensor, see 3.1.15 - Check connection or change the changer motor, see 3.1.14 - Check connection or change sensor, see 3.1.13 - Check connection or change aperture module, see 3.1.12 - Change sensor or repair connection, see 3.1.17 - Change motor, repair connection, see 3.1.16 - Check connection - Change head or send instrument to next in-house repair to exchange the APD PCB.
- Change the motherboard MEDCOM
- Change head or send instrument to next in-house repair. - Change head or send instrument to next in-house repair. - Change head or send instrument to next in-house repair. Service Manual Part No. 1836-541
Troubleshooting
5.3
IOLMaster 5.5
Tool list IOLMaster Tool Designation
Order Number
Fieldmaster power meter with power adapter
Service Manual Part No. 1836-541
Use of Tool Indicating device of laser power
115V 230V
0115-811 0115-810
Detector head LM 2
0113-503 Sensor of metering laser power
Measuring Adapter for LM2
1094-546 To use with the LM2 detector head for exact positioning.
Ophthalmometer Headrest Clamp
302086-0000-000 Holder for detector head on head rest
Ophir NOVA power meter
320866-0209-022 Indicating device of laser power
Confidential and Proprietary
5 - 14
SM-30-4054-en Section 5, Revision A, Feb. 2010
IOLMaster 5.5
Troubleshooting
Tool Designation
Order Number
Ophir power head PD 300 – 3W
SM-30-4054-en Section 5, Revision A, Feb. 2010
Use of Tool
On request Sensor of metering of laser power
Hygrometer TESTO 608-H1
0492-815 Speed adjustment
Service USB flash drive and IOLMaster Software Version 5.5 USB flash drive
1778-631 Re-imaging the HDD of the IOLMaster 1790-501
Alignment device for IOLMaster
1676-641 CCD camera alignment
Laser short circuit plug
1078-821 For protecting the laser diode (ESD) during service activities.
3/32" hex key
0118-751 For unscrewing the power supply
Torx tool kit T5-T50
1209-366 For service on the LC display
Confidential and Proprietary
5- 15
Service Manual Part No. 1836-541
Section 6 – Diagrams 6.1 Diagrams........................................................................................................................3 6.1.1 Overview of functional modules ...................................................................................... 3 6.1.2 Measuring voltages and signals ....................................................................................... 4 Voltages on the power supply unit .................................................................................. 4 Voltages and signals on the Power ON/OFF PCB .............................................................. 4 Voltages and signals on APD 3 PCB ................................................................................. 5 Voltages and signals on Digital IOL PCB........................................................................... 6
SM-30-4054-en Section 6, Revision A, Feb. 2010
Confidential and Proprietary
6-1
Service Manual Part No. 1836-541
Diagrams
IOLMaster 5.5
Notes:
Service Manual Part No. 1836-541
Confidential and Proprietary
6-2
SM-30-4054-en Section 6, Revision A, Feb. 2010
IOLMaster 5.5
6.1
Diagrams
Diagrams
6.1.1 Overview of functional modules Measuring head
Interferometer
Control Digital IOL
X202 X202 X160
EEPROM
Instrument base
X208 X208
speed down
speed up
Fire button
Light detector end
X203 X203
X209
X1
Light detector beginning M Motor IOL
X204 X204
R77
R64
R40
6 IR LED cycled(KER)
Aperturte diaphragm X205 X205
X401 X214
6 IR LED overview illum.
Measuring system IOLMaster
Light detector M Motor aperture
X212 X212
R3
5 LED ACD left X402
X402
X217 X217
5 LED ACD right X402
X402
X218 X218
Changer X207 X207
Light detector M Motor changer
X211 X211
R7 40μW R75
CCD camera
X150
X150
X219
Gain CCD
X219
X216 X170
R9 400μW
X222 X502
+9V +5V
LED converter unit for backlight
LC-Display 640x480
+12V
+12V
CN1 CN1
X231 X201
X213
X581 X580
X231 X201
X213
IR receiver X501
X304 Power
X304
Isolating transformer 115/230 VAC 115VA
Monitor Laser Diode
Reflex Fixier Diode
+12 V
+12V X301 +5V1/2/3 -5V1/2/3 -12V R26
+5V -5V -12V
MUX
X302 X302
Laser Monitoring
APD Diode
APD-Module IOLMaster
Power
Power INPUT 100...240 VAC Assym. table
S1
Power
1a 1 2 2a
Keyboard
X6 X5X2 X3
DISPLAY optional
Touch pad
USB flash drive optional
X855 X851 Power on/off X853
PS2 PS2 X611 Mouse Keyboard
PSU 100-240V AC
X601
VGA
USB (4x)
X601
X609
X608
+12V +5V A
Power IN +12V
X600
LAN optional
Network isolator (optional)
Ethernet X607
Computer MED-COM D
X602
X604
X603
Fan
Inductive position switch
X7
Printer optional
SATA X606
COM X610
HDD
Computer base IOLMaster SM-30-4054-en Section 6, Revision A, Feb. 2010
Confidential and Proprietary
6- 3
Service Manual Part No. 1836-541
Diagrams
IOLMaster 5.5
6.1.2 Measuring voltages and signals Voltages on the power supply unit +12V +12V GND GND GND
DC output voltage of the PSU
Power Supply 100 -240 V (AC)
Input voltage of the PSU 100 –240V AC Line input voltage of the Power ON/OFF PCB
Line Input 100 -240 V (AC)
Voltages and signals on the Power ON/OFF PCB S1
Power Supply 100-240V AC S1
Line input 100–240V AC
F1/F2 T 3.15A X853:1 ◄ +5V of MED-COM X853:2 ► ON/OFF X853:3 ◄ Delete X853:4 ◄ GND of MED-COM
risk of electric shock
X602 Motherboard MED-COM
Service Manual Part No. 1836-541
Confidential and Proprietary
6-4
SM-30-4054-en Section 6, Revision A, Feb. 2010
IOLMaster 5.5
Diagrams
Voltages and signals on APD 3 PCB
R26 Laser Monitoring
◄ X304:1 selection MUX ◄ X304:2 selection MUX ◄ X304:3 selection MUX ◄ X304:4 0V ◄ X304:5 +12V ◄ X304:6 0V X213 Digital IOL
► X302:1 ► X302:2 ► X302:3 ► X302:4
0V Analog signal MUX -5V2 +5V2
X601 MED-COM ◄ X301:1 anode ◄ X301:2 cathode X501 IR Laser Monitoring Diode
SM-30-4054-en Section 6, Revision A, Feb. 2010
Confidential and Proprietary
6- 5
Service Manual Part No. 1836-541
Diagrams
IOLMaster 5.5
Voltages and signals on Digital IOL PCB
40μW laser power R7 Service connector
400μW laser power R9
X220: 1 +12V X220: 2 + 9V X220: 3 + 5V X220: 4 GND
Brightness Keratometer R40 Surrounding field illumination R3 CCD Camera sensitivity R75 Interf. motor speed (down) R77 Interf. motor speed (up) R64
CCD Camera Module ► X219:4 +9V ► X219:5 GND X201:12 +12V ◄MED-COM X601 X201: 9 GND ◄MED-COM X601 X231: 9 +5V
◄MED-COM X601
APD Module X304 ► X213:5 +12V ► X213:6 GND
X209:1 NC X209:2 GND X209:3 NO FIRE
Service Manual Part No. 1836-541
Confidential and Proprietary
6-6
SM-30-4054-en Section 6, Revision A, Feb. 2010
Section 7 - Parts IOLMaster V5.5 IOLMaster V5.5
4
Zubehör IOLMaster Accessories IOLMaster
SM-30-4054 Section 7, Revision A, Feb. 2010
50
Confidential and Proprietary
7-1
Service Manual Order No. 1836-541
SM-30-4054 IOLMaster V5.5 000000-1836-541 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1
IOLMaster V5.5
IOLMaster V5.5
2
Zubehör IOLMaster
Accessories IOLMaster
~ 000000-1790-501
SW-Paket IOLMaster V5.5
SW-Paket IOLMaster V5.5
~ 000000-1778-631
SW-Paket IOLMaster Service Vs 7.1.2
SW-Packet IOLMaster Service 7.1.2
Service Manual Order No. 1836-541
Confidential and Proprietary
7-2
4 50
SM-30-4054 Section 7, Revision A, Feb. 2010
SM-30-4054 IOLMaster V5.5 000000-1836-541
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-3
Service Manual Order No. 1836-541
IOLMaster V5.5 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1 000000-1836-516
Set Ersatzhauben IOLMaster V5.5
Set of spare covers for IOLMaster V5.5
2
ISO 4762-M 3 x 10-A2-70
Screw M3x10 (screws for scanner)
3
ISO 4762-M 3X 20-A2-70
ISO 4762-M 3X 20-A2-70
6
Untersatz IOLMaster
Overview IOLMaster base
12
7
Gerätekopf IOLMaster m. Verpackung
IOLMaster head incl. packaging
32
Service Manual Order No. 1836-541
Confidential and Proprietary
7-4
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-5
Service Manual Order No. 1836-541
IOLMaster V5.5 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
50 000000-1692-640
Kabel MS (X160) - DIGIT (X202)
Cable MS (X160) - DIGIT (X202)
51 000000-1078-821
Kurzschlussstecker
Short Circuit Connector IOL Master
52 000000-1534-730
FBG DIGITAL IOL (DIGIT)
Digital IOL (DIGIT)
53 000000-1693-879
Koppler Interferometer X203
Interferometer Coupler X203
54 000000-1694-348
Koppler Interferometer X204
Optical sensor interferometer X204
55 000000-1761-131
Kabel IOLMaster (X601) - Gerätekopf
Cable IOLMaster (X601) - head
44
57 000000-1692-644
Kabel DIGIT (X213) - APD (X304)
Cable DIGIT (X213) - APD (X304)
46
58 000000-1693-867
Koppler Wechsler X207
Optical sensor changer X207
59 000000-1692-642
Kabel CCD (X150) - DIGIT (X219)
Cable CCD (X150) - DIGIT (X219)
Service Manual Order No. 1836-541
Confidential and Proprietary
7-6
42
40
48
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-7
Service Manual Order No. 1836-541
IOLMaster V5.5 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
60 000000-1693-816
Motor Wechsler
Changer motor
61 000000-1693-881
Koppler Blende X205
Optical sensor diaphragm X205
62 000000-0510-510
TFT-Color-Display 6,5-inch
TFT-Color-Display 6,5-inch
63 000000-0510-511
LED-Converter Unit für Backlight
LED-Converter Unit for Backlight
Service Manual Order No. 1836-541
Confidential and Proprietary
7-8
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-9
Service Manual Order No. 1836-541
IOLMaster V5.5 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
6
Untersatz IOLMaster
Overview IOLMaster base
12
7
Gerätekopf IOLMaster m. Verpackung
IOLMaster head incl. packaging
32
50 000000-1692-640
Kabel MS (X160) - DIGIT (X202)
Cable MS (X160) - DIGIT (X202)
42
52 000000-1534-730
FBG DIGITAL IOL (DIGIT)
Digital IOL (DIGIT)
40
53 000000-1693-879
Koppler Interferometer X203
Interferometer Coupler X203
54 000000-1694-348
Koppler Interferometer X204
Optical sensor interferometer X204
55 000000-1761-131
Kabel IOLMaster (X601) - Gerätekopf
Cable IOLMaster (X601) - head
44
57 000000-1692-644
Kabel DIGIT (X213) - APD (X304)
Cable DIGIT (X213) - APD (X304)
46
58 000000-1693-867
Koppler Wechsler X207
Optical sensor changer X207
59 000000-1692-642
Kabel CCD (X150) - DIGIT (X219)
Cable CCD (X150) - DIGIT (X219)
60 000000-1693-816
Motor Wechsler
Changer motor
61 000000-1693-881
Koppler Blende X205
Optical sensor diaphragm X205
62 000000-0510-510
TFT-Color-Display 6,5-inch
TFT-Color-Display 6,5-inch
63 000000-0510-511
LED-Converter Unit für Backlight
LED-Converter Unit for Backlight
Service Manual Order No. 1836-541
Confidential and Proprietary
7-10
48
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-11
Service Manual Order No. 1836-541
Overview IOLMaster base Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1 000000-1025-670
Kopfstütze IOL
Head Rest
14
2 266101-4501-927
Kreuztisch IOL
Cross Stage for IOL Master
16
3
DIN 6912-M5X 16-A2-70
DIN 6912-M5X 16-A2-70
4
Computer IOLMaster mit Haube
Computer IOLMaster with cover
5
AT Computer IOLMaster mit Haube
EC IOLMaster computer with cover
6
ISO 7380-M 5x 8 A2-70
ISO 7380-M 5x 8 A2-70
7 266101-4510-310
Gleitplatte
Sliding plate for Cross stage
8 000000-1210-002
Zahnradpaar mit Rollen
Pair of gear wheels with rollers
9
ISO 7380-M 4x12 A2-70
Screw M4x12 (screws for cover)
Zahnstange RAL U713
Gear rack RAL U713
10 266101-4500-562
Service Manual Order No. 1836-541
Confidential and Proprietary
7-12
20
SM-30-4054 Section 7, Revision A, Feb. 2010
Overview IOLMaster base
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-13
Service Manual Order No. 1836-541
Head Rest 000000-1025-670 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1 319752-0024-000
Stirnband
Forehead strap
2
ISO 7380-M 4x 6-A2-70
ISO 7380-M 4x 6-A2-70
3 319752-8016-000
Kinnschale
Chin rest
4 266181-7000-551
Papierhalter
Paper holder
5
ISO 4762-M 5 x 20-A2-70
ISO 4762-M 5 x 20-A2-70
6
ISO 4762-M 4 x 8-A2-70
Screw M4x8
7 000000-1294-613
Kinntraverse montiert IOL
Chin rest unit, assembled
Service Manual Order No. 1836-541
Confidential and Proprietary
7-14
SM-30-4054 Section 7, Revision A, Feb. 2010
Head Rest 000000-1025-670
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-15
Service Manual Order No. 1836-541
Cross Stage for IOL Master 266101-4501-927 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1
ISO 1580-M3 x 5-A2-70
ISO 1580-M3 x 5-A2-70
2
DIN 84-M 2,5x 4-A2-70
DIN 84-M 2,5x 4-A2-70
3 000000-1775-095
Set Kleinteile KT
Service Set cross table locking knob IOL
4
DIN 920-M3 x8 -A1-70
DIN 920-M3 x8 -A1-70
5
ISO 4762-M 3 x 10-A2-70
Screw M3x10 (screws for scanner)
6
DIN 125-A 3,7-A2
Disk DIN 125-A 3,7-A2
7 266101-4510-010
Spindelmutter
Spindle nut
8 266101-4502-514
SPINDEL. KOMPL.
Spindle, complete
9
ISO 4762-M 4 x 20-A2-70
ISO 4762-M 4 x 20-A2-70
10
4,2X12X 0,8 CZN790 X12CRNI17.7K
4,2X12X 0,8 CZN790 X12CRNI17.7K
11 266101-4533-010
Stirnrad
Spur gear
12 266101-4560-514
Steuergriff
IOL joystick with release button
13
AT Steuergriff IOLMaster
EC IOLMaster Joystick w. release button
14
ISO 4762-M 3 x 8-A2-70
ISO 4762-M 3 x 8-A2-70
15
DIN 125-A 3,2-A2
DIN 125-A 3.2-A2
16
DIN 125-A 5,3-A2
DIN 125-A 5,3-A2
17
DIN 6912-M 5x 12-A2-70
DIN 6912-M 5x 12-A2-70
Service Manual Order No. 1836-541
Confidential and Proprietary
7-16
18
SM-30-4054 Section 7, Revision A, Feb. 2010
Cross Stage for IOL Master 266101-4501-927
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-17
Service Manual Order No. 1836-541
IOL joystick with release button 266101-4560-514 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1
AT Steuergriff IOLMaster
EC IOLMaster Joystick w. release button
2 266101-4515-110
Kappe
Cap
Service Manual Order No. 1836-541
Confidential and Proprietary
7-18
SM-30-4054 Section 7, Revision A, Feb. 2010
IOL joystick with release button 266101-4560-514
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-19
Service Manual Order No. 1836-541
Computer IOLMaster with cover Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1
AT Computer IOLMaster mit Haube
EC IOLMaster computer with cover
2 000000-1693-874
induktiver Näherungsschalter
Induct. Sensor Switch
3
ISO 4762-M 4X 10-A2-70
ISO 4762-M 4X 10-A2-70
4
ISO 4762-M 3 x 5-A2-70
ISO 4762-M 3 x 5-A2-70
5
DIN 125-A 3,2-EP-HP S2000E/FR3NATUR
DIN 125-A 3,2-EP-HP S2000E/FR3NATUR
6 000000-0509-551
Festplatte 2,5" SATA 80GB
Hard disk 2,5" SATA 80GB
7 000000-1528-161
FBG MED-COM IOLMaster (MED-COM)
IOLMaster MED-COM PCB
8
AT FBG MED-COM IOLMaster (MED-COM)
EC PCB MED-COM IOLMaster (MED-COM)
9
ISO 4762-M 3 x 8-A2-70
ISO 4762-M 3 x 8-A2-70
FBG POWER ON/OFF 2 (POWER)
ASSEMBLY POWER ON/OFF 2
24
10A 000000-1694-354
Kabel Power - On/Off Netz
Power On/Off - mains cable
26
11 000000-0508-752
Netzteil 90W 12Vdc/7,5A
Power supply 90W 12VDC/7,5A
12
ISO 7380-M 3x 5-A2-70
ISO 7380-M 3x 5-A2-70
13
DIN 125-A 3,2-A2
DIN 125-A 3.2-A2
14 000000-1693-872
Lüfter kpl.
Fan complete
15 000000-1531-678
Schalter RSM 2pol. grün Sonderdruck
Master switch, 2 pins, green
16 000000-0505-635
Netzmodul/MED 3A Fi
Power entry modul medical version 3A
10 000000-1671-563
Service Manual Order No. 1836-541
steck6308
Confidential and Proprietary
7-20
SM-30-4054 Section 7, Revision A, Feb. 2010
Computer IOLMaster with cover
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-21
Service Manual Order No. 1836-541
Computer IOLMaster with cover Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
2 000000-1693-874
induktiver Näherungsschalter
Induct. Sensor Switch
6 000000-0509-551
Festplatte 2,5" SATA 80GB
Hard disk 2,5" SATA 80GB
7 000000-1528-161
FBG MED-COM IOLMaster (MED-COM)
IOLMaster MED-COM PCB
8
AT FBG MED-COM IOLMaster (MED-COM)
EC PCB MED-COM IOLMaster (MED-COM)
FBG POWER ON/OFF 2 (POWER)
ASSEMBLY POWER ON/OFF 2
24
10A 000000-1694-354
Kabel Power - On/Off Netz
Power On/Off - mains cable
26
11 000000-0508-752
Netzteil 90W 12Vdc/7,5A
Power supply 90W 12VDC/7,5A
14 000000-1693-872
Lüfter kpl.
Fan complete
15 000000-1531-678
Schalter RSM 2pol. grün Sonderdruck
Master switch, 2 pins, green
16 000000-0505-635
Netzmodul/MED 3A Fi
Power entry modul medical version 3A
17 000000-1694-129
Kabel POWER(X853)-MED-COM(X602)
Cable POWER(X853)-MED-COM(X602)
28
18 000000-1694-203
Kabel MED-COM (X600) - NT (X7)
Cable MED-COM (X600) - NT (X7)
30
19 000000-0509-839
Knopfzelle Lithium 3V/230mAh 3,2x20mmØ
Lithium coin cell 3V/230mAh 3,2x20mmØ
10 000000-1671-563
Service Manual Order No. 1836-541
steck6308
Confidential and Proprietary
7-22
SM-30-4054 Section 7, Revision A, Feb. 2010
Computer IOLMaster with cover
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-23
Service Manual Order No. 1836-541
ASSEMBLY POWER ON/OFF 2 000000-1671-563
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-24
SM-30-4054 Section 7, Revision A, Feb. 2010
ASSEMBLY POWER ON/OFF 2 000000-1671-563
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-25
Service Manual Order No. 1836-541
Power On/Off - mains cable 000000-1694-354
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-26
SM-30-4054 Section 7, Revision A, Feb. 2010
Power On/Off - mains cable 000000-1694-354
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-27
Service Manual Order No. 1836-541
Cable POWER(X853)-MED-COM(X602) 000000-1694-129
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-28
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Cable POWER(X853)-MED-COM(X602) 000000-1694-129
Confidential and Proprietary
7-29
Service Manual Order No. 1836-541
Cable MED-COM (X600) - NT (X7) 000000-1694-203
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-30
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Cable MED-COM (X600) - NT (X7) 000000-1694-203
Confidential and Proprietary
7-31
Service Manual Order No. 1836-541
IOLMaster head incl. packaging Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1
ISO 4762-M 5 X16-A2-70
ISO 4762-M 5 X16-A2-70
2
DIN 125-A 5,3-A2
DIN 125-A 5,3-A2
3 000000-1770-344
LCD, ohne Kabel
LCD, without cable
4 000000-1757-148
Spaltgruppe VKT
ACD slit module
5
DIN 125-A 2,2-A2
DIN 125-A 2,2-A2
6
ISO 4762-M 2 X 8-A2-70
ISO 4762-M 2 X 8-A2-70
7 000000-1757-077
Keratometer,kpl.
Keratometer, compl.
8
ISO 4762-M 3 x 6-A2-70
ISO 4762-M 3 x 6-A2-70
9 000000-1756-302
Blendengruppe Keratometer
Keratometer diaphragm module
10
DIN 433- 3,2-A1
Washer DIN 433-3,2-A1
11
ISO 4762-M 3 x 8-A2-70
ISO 4762-M 3 x 8-A2-70
12 000000-1693-816
Motor Wechsler
Changer motor
13
ISO 4762-M 2,5x 4-A2-70
ISO 4762-M 2,5x 4-A2-70
14
DIN 125-A 2,7-A2
DIN 125-A 2,7-A2 disc
15
APD-LP montiert
APD PCB, mounted
16
DIN 125-A 3,2-A2
DIN 125-A 3.2-A2
17
ISO 4762-M 3 x 10-A2-70
Screw M3x10 (screws for scanner)
18 000000-1770-340
Set, CCD Kamera 9V
CCD camera set, 9V
19
ISO 4762-M 2,5x 6-A2-70
ISO 4762-M 2,5x 6-A2-70
20
DIN 9021- 2,7-140 HV-A2
DIN 9021- 2,7-140 HV-A2
21
Prismengruppe 26-er N-SF2
Prism module 26, N-SF2
22
DIN 137-A 3 -A2
DIN 137-A 3 -A2
Service Manual Order No. 1836-541
Confidential and Proprietary
7-32
36
38
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster head incl. packaging
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-33
Service Manual Order No. 1836-541
IOLMaster head incl. packaging Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
ISO 4762-M 3 x 6-A2-70
ISO 4762-M 3 x 6-A2-70
10
DIN 433- 3,2-A1
Washer DIN 433-3,2-A1
11
ISO 4762-M 3 x 8-A2-70
ISO 4762-M 3 x 8-A2-70
23 000000-1693-867
Koppler Wechsler X207
Optical sensor changer X207
24
ISO 4762-M 3 x 5-A2-70
ISO 4762-M 3 x 5-A2-70
25
Distanzhülse Pa/sw 3,2x5,0x3,0
distance bush Pa/bk 3,2x5,0x3,0
26
DIN 125-A 3,2-EP-HP S2000E/FR3NATUR
DIN 125-A 3,2-EP-HP S2000E/FR3NATUR
27 000000-1078-821
Kurzschlussstecker
Short Circuit Connector IOL Master
28 000000-1534-730
FBG DIGITAL IOL (DIGIT)
Digital IOL (DIGIT)
29 000000-1755-837
Motor Interferometer, kpl.
Motor Interferometer, complete
30 000000-1693-879
Koppler Interferometer X203
Interferometer Coupler X203
31 000000-1694-348
Koppler Interferometer X204
Optical sensor interferometer X204
32
Interferometer mit Verpackung
Interferometer incl. packaging
33
ALM-Wechsler
ALM converter
8
Service Manual Order No. 1836-541
Confidential and Proprietary
7-34
40
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster head incl. packaging
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-35
Service Manual Order No. 1836-541
ACD slit module 000000-1757-148
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-36
SM-30-4054 Section 7, Revision A, Feb. 2010
ACD slit module 000000-1757-148
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-37
Service Manual Order No. 1836-541
Keratometer diaphragm module 000000-1756-302 Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1 000000-1693-881
Koppler Blende X205
Optical sensor diaphragm X205
2
ISO 4762-M 3 x 5-A2-70
ISO 4762-M 3 x 5-A2-70
3
DIN 433- 3,2-A1
Washer DIN 433-3,2-A1
Service Manual Order No. 1836-541
Confidential and Proprietary
7-38
SM-30-4054 Section 7, Revision A, Feb. 2010
Keratometer diaphragm module 000000-1756-302
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-39
Service Manual Order No. 1836-541
Digital IOL (DIGIT) 000000-1534-730
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-40
SM-30-4054 Section 7, Revision A, Feb. 2010
Digital IOL (DIGIT) 000000-1534-730
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-41
Service Manual Order No. 1836-541
Cable MS (X160) - DIGIT (X202) 000000-1692-640
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-42
SM-30-4054 Section 7, Revision A, Feb. 2010
Cable MS (X160) - DIGIT (X202) 000000-1692-640
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-43
Service Manual Order No. 1836-541
Cable IOLMaster (X601) - head 000000-1761-131
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-44
SM-30-4054 Section 7, Revision A, Feb. 2010
Cable IOLMaster (X601) - head 000000-1761-131
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-45
Service Manual Order No. 1836-541
Cable DIGIT (X213) - APD (X304) 000000-1692-644
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-46
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Cable DIGIT (X213) - APD (X304) 000000-1692-644
Confidential and Proprietary
7-47
Service Manual Order No. 1836-541
Cable CCD (X150) - DIGIT (X219) 000000-1692-642
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-48
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Cable CCD (X150) - DIGIT (X219) 000000-1692-642
Confidential and Proprietary
7-49
Service Manual Order No. 1836-541
Accessories IOLMaster Pos. Item
Bestell-Nr. Order no.
IOLMaster V5.5 Seite Page
Benennung
Designation
1 000000-0433-973
Polyester-Tupfer TX 761 -Basan-
Alpha Swab with Long Handle TX 761
2 000000-0500-895
neo-Lab Wattestäbchen, 150mm, 1Pck=100St Cotton swab 150mm (100 pcs)
3 000000-1046-387
Halterung (Prüfkugel)
Holder Test Eye Compl for IOL Master
52
4 000000-1313-258
Tastatur ACK 540, deutsch
Keyboard ACK 540, German
54
5 000000-1313-260
Tastatur ACK-540, Englisch
Keyboard ACK-540, English
56
6 000000-1526-414
Optikputzmischung L , Flasche 50ml
Optics cleaning mixture L, 50ml bottle
7 000000-1676-641
Prüfvorrichtung für Kamera
Alignment device for IOL master
Service Manual Order No. 1836-541
Confidential and Proprietary
7-50
SM-30-4054 Section 7, Revision A, Feb. 2010
Accessories IOLMaster
IOLMaster V5.5
KEINE ZEICHNUNG VORHANDEN NO DRAWING AVAILABLE
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-51
Service Manual Order No. 1836-541
Holder Test Eye Compl for IOL Master 000000-1046-387
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-52
SM-30-4054 Section 7, Revision A, Feb. 2010
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Holder Test Eye Compl for IOL Master 000000-1046-387
Confidential and Proprietary
7-53
Service Manual Order No. 1836-541
Keyboard ACK 540, German 000000-1313-258
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-54
SM-30-4054 Section 7, Revision A, Feb. 2010
Keyboard ACK 540, German 000000-1313-258
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-55
Service Manual Order No. 1836-541
Keyboard ACK-540, English 000000-1313-260
IOLMaster V5.5
KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE
Service Manual Order No. 1836-541
Confidential and Proprietary
7-56
SM-30-4054 Section 7, Revision A, Feb. 2010
Keyboard ACK-540, English 000000-1313-260
IOLMaster V5.5
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-57
Service Manual Order No. 1836-541
pé~êÉ=m~êíë=iáëí Order No.
Designation
000000-0433-973
Alpha Swab with Long Handle TX 761
000000-0500-895
Cotton swab 150mm (100 pcs)
000000-0505-635
Power entry modul medical version 3A
000000-0505-635
Power entry modul medical version 3A
000000-0508-752
Power supply 90W 12VDC/7,5A
000000-0508-752
Power supply 90W 12VDC/7,5A
000000-0509-551
Hard disk 2,5" SATA 80GB
000000-0509-551
Hard disk 2,5" SATA 80GB
000000-0509-839
Lithium coin cell 3V/230mAh 3,2x20mmØ
000000-0510-510
TFT-Color-Display 6,5-inch
000000-0510-510
TFT-Color-Display 6,5-inch
000000-0510-511
LED-Converter Unit for Backlight
000000-0510-511
LED-Converter Unit for Backlight
000000-1025-670
Head Rest
000000-1046-387
Holder Test Eye Compl for IOL Master
000000-1078-821
Short Circuit Connector IOL Master
000000-1078-821
Short Circuit Connector IOL Master
000000-1210-002
Pair of gear wheels with rollers
000000-1294-613
Chin rest unit, assembled
000000-1313-258
Keyboard ACK 540, German
000000-1313-260
Keyboard ACK-540, English
000000-1526-414
Optics cleaning mixture L, 50ml bottle
000000-1528-161
IOLMaster MED-COM PCB
000000-1528-161
IOLMaster MED-COM PCB
000000-1531-678
Master switch, 2 pins, green
000000-1531-678
Master switch, 2 pins, green
000000-1534-730
Digital IOL (DIGIT)
000000-1534-730
Digital IOL (DIGIT)
000000-1534-730
Digital IOL (DIGIT)
000000-1671-563
ASSEMBLY POWER ON/OFF 2
000000-1671-563
ASSEMBLY POWER ON/OFF 2
000000-1676-641
Alignment device for IOL master
000000-1692-640
Cable MS (X160) - DIGIT (X202)
000000-1692-640
Cable MS (X160) - DIGIT (X202)
000000-1692-642
Cable CCD (X150) - DIGIT (X219)
000000-1692-642
Cable CCD (X150) - DIGIT (X219)
000000-1692-644
Cable DIGIT (X213) - APD (X304)
000000-1692-644
Cable DIGIT (X213) - APD (X304)
000000-1693-816
Changer motor
000000-1693-816
Changer motor
000000-1693-816
Changer motor
000000-1693-867
Optical sensor changer X207
000000-1693-867
Optical sensor changer X207
Service Manual Order No. 1836-541
Confidential and Proprietary
7-58
SM-30-4054 Section 7, Revision A, Feb. 2010
pé~êÉ=m~êíë=iáëí Order No.
Designation
000000-1693-867
Optical sensor changer X207
000000-1693-872
Fan complete
000000-1693-872
Fan complete
000000-1693-874
Induct. Sensor Switch
000000-1693-874
Induct. Sensor Switch
000000-1693-879
Interferometer Coupler X203
000000-1693-879
Interferometer Coupler X203
000000-1693-879
Interferometer Coupler X203
000000-1693-881
Optical sensor diaphragm X205
000000-1693-881
Optical sensor diaphragm X205
000000-1693-881
Optical sensor diaphragm X205
000000-1694-129
Cable POWER(X853)-MED-COM(X602)
000000-1694-203
Cable MED-COM (X600) - NT (X7)
000000-1694-348
Optical sensor interferometer X204
000000-1694-348
Optical sensor interferometer X204
000000-1694-348
Optical sensor interferometer X204
000000-1694-354
Power On/Off - mains cable
000000-1694-354
Power On/Off - mains cable
000000-1755-837
Motor Interferometer, complete
000000-1756-302
Keratometer diaphragm module
000000-1757-077
Keratometer, compl.
000000-1757-148
ACD slit module
000000-1761-131
Cable IOLMaster (X601) - head
000000-1761-131
Cable IOLMaster (X601) - head
000000-1770-340
CCD camera set, 9V
000000-1770-344
LCD, without cable
000000-1775-095
Service Set cross table locking knob IOL
000000-1778-631
SW-Packet IOLMaster Service 7.1.2
000000-1790-501
SW-Paket IOLMaster V5.5
000000-1836-516
Set of spare covers for IOLMaster V5.5
266101-4500-562
Gear rack RAL U713
266101-4501-927
Cross Stage for IOL Master
266101-4502-514
Spindle, complete
266101-4510-010
Spindle nut
266101-4510-310
Sliding plate for Cross stage
266101-4515-110
Cap
266101-4533-010
Spur gear
266101-4560-514
IOL joystick with release button
266181-7000-551
Paper holder
319752-0024-000
Forehead strap
319752-8016-000
Chin rest
SM-30-4054 Section 7, Revision A, Feb. 2010
Confidential and Proprietary
7-59
Service Manual Order No. 1836-541
Section 8 – Appendices
IOLMaster Instrument Checklist and Maintenance Report Service Bulletins
SM-30-4054-en Section 8, Revision A, Feb. 2010
Confidential and Proprietary
8-1
Service Manual Part No. 1836-541