Manual de Servicio Iol Master [PDF]

  • 0 0 0
  • Gefällt Ihnen dieses papier und der download? Sie können Ihre eigene PDF-Datei in wenigen Minuten kostenlos online veröffentlichen! Anmelden
Datei wird geladen, bitte warten...
Zitiervorschau

Confidential and Proprietary

SM-30-4054 Part No. 1836-541, Revision A February 2010

Diagrams

Parts Removal / Replacement

=

Parts

Adjustment and Calibration

sÉêëáçå=RKR

Appendices

Troubleshooting Error codes

flij~ëíÉê=

pÉêîáÅÉ=j~åì~ä= Checkout and Preventive Maintenance

General Information

Table of Contents

Service Manual Part No. 1836-541

IOLMaster 5.5

Confidential and Proprietary

ii

SM-30-4054-en Revision A, Feb. 2010

IOLMaster 5.5

Table of Contents

PROPRIETARY NOTICE Information contained in this document is copyright Carl Zeiss Meditec AG and may not be reproduced in full or in part by any person without prior written approval of Carl Zeiss Meditec. Its purpose is to provide the User with adequately detailed information so as to repair, maintain, and order spare parts for the instrument supplied. Every effort has been made to keep the information contained in this document current and accurate as of the date of publication or revision. However, no guarantee is given or implied that the document is error-free or that it is accurate with regard to any specification.

All trademarks, registered trademarks, and trade names that appear in this manual are the property of their respective holders. Copyright © 2010 by Carl Zeiss Meditec AG All rights reserved. Carl Zeiss Meditec AG Goeschwitzer Strasse 51-52 07745 Jena GERMANY

SM-30-4054-en Revision A, Feb. 2010

Confidential and Proprietary

iii

Service Manual Part No. 1836-541

Table of Contents

IOLMaster 5.5

REVISION CONTROL LIST Document:

IOLMaster 5.5

Service Manual Part No.: Issued Date:

SM-30-4054 1836-541 February 2010

Listed at the bottom of each page is the part number of the manual, along with the Revision letter and date for that page (e.g. Part No. 1836-541, Revision A, Feb. 2010) and the document number (e.g. SM-30-4054). Subsequent revisions to a page will be noted by a corresponding change to the Revision letter and date.

Service Manual Part No. 1836-541

Confidential and Proprietary

iv

SM-30-4054-en Revision A, Feb. 2010

IOLMaster 5.5

Table of Contents

Table of Contents Section 1 – General Information

Section 2 – Checkout and Preventive Maintenance

Section 3 – Parts Removal / Replacement

Section 4 – Adjustment and Calibration

Section 5 – Troubleshooting / Error codes

Section 6 – Diagrams

Section 7 – Parts

Section 8 – Appendices

SM-30-4054-en Revision A, Feb. 2010

Confidential and Proprietary

v

Service Manual Part No. 1836-541

Section 1 – General Information

1.1 About This Manual.........................................................................................................3 1.1.1 General Information ........................................................................................................ 3 1.1.2 Conventions .................................................................................................................... 4 1.2 Service Bulletins .............................................................................................................4 1.3 IOLMaster Service Strategy ............................................................................................5 1.4 General Safety Instruction ..............................................................................................5 1.5 Disposal of the product within the EU............................................................................5

SM-30-4054-en Section1, Revision A, Feb. 2010

Confidential and Proprietary

1-1

Service Manual Part No. 1836-541

General Information

IOLMaster 5.5

Notes:

Service Manual Part No. 1836-541

Confidential and Proprietary

1-2

SM-30-4054-en Section 1, Revision A, Feb. 2010

IOLMaster 5.5

1.1

General Information

About This Manual

1.1.1 General Information This manual is the field service reference for troubleshooting, repair, adjustment, and calibration of the IOLMaster. The manual is intended for use by field service engineers who have completed Carl Zeiss Meditec service training on the IOLMaster. The service manual is designed to support Level 1 of a two-level service strategy. Level 1 (on-site) field service employs modular replacement of printed circuit boards and other assemblies that are most effectively repaired at a central repair facility. This is the service strategy used in field service, and presented in Carl Zeiss Meditec (CZM) training classes. Level 2 (workshop repair) include procedures which are not possible on-side or where on-side repair was not successful. See Section 1.3 for additional information regarding the IOLMaster service strategy. The procedures in this manual assume that the reader is familiar with operation of the instrument. Complete operating instructions are contained in the IOLMaster User Manual. Information presented in the user manual is not repeated in this service manual. The user manual can be ordered separately by standard Zeiss parts order. Update revisions to this service manual will be issued by Service Bulletin, as required. The general layout of the service manual is shown below. For greater detail, please refer to the Table of Contents. Level 1 Service Manual Layout Section 1 - General Information Section 2 - System Checkout and Preventive Maintenance Section 3 - Parts Removal / Replacement Section 4 - Adjustment and Calibration Section 5 - Troubleshooting Section 6 - Diagrams Section 7 - Spare Parts Section 8 - Appendices A: Service Bulletins B: Instrument Checklist

SM-30-4054-en Section 1, Revision A, Feb. 2010

Confidential and Proprietary

1-3

Service Manual Part No. 1836-541

General Information

IOLMaster 5.5

1.1.2 Conventions The following conventions apply in this manual:

ƒ ƒ

Front and Back sides of the instrument is as viewed from the front (patient side) of the instrument, unless noted otherwise. All tool sizes for screws and nuts mentioned in the instructions are metric unless noted otherwise.

1.2 Service Bulletins Service Bulletins are a vital element of service support. Bulletins are used to quickly convey technical information on a variety of field service topics, including:

ƒ ƒ ƒ ƒ ƒ

instrument design changes technical problems and corrections software updates

ƒ ƒ

new troubleshooting procedures problem alerts

ƒ

service manual revisions new calibration or adjustment procedures upgrade announcements/ procedures system checkout – checklist

ƒ

service disclaimer form

ƒ

Your service bulletins should be filed where easily accessible for quick reference.

NOTICE Service Bulletins are Confidential and Proprietary, for the sole use of personnel employed by Carl Zeiss Meditec, Carl Zeiss Meditec affiliates, and authorized Carl Zeiss Meditec distributors. Carl Zeiss Meditec has a well-deserved reputation for high quality, reliable instruments, unsurpassed in the industry. As a Carl Zeiss Meditec employee, affiliate, or distributor you are required to handle your service bulletins as appropriate for proprietary and confidential information.

Service Manual Part No. 1836-541

Confidential and Proprietary

1-4

SM-30-4054-en Section 1, Revision A, Feb. 2010

IOLMaster 5.5

General Information

1.3 IOLMaster Service Strategy The service strategy for IOLMaster envisages that field service must only be performed by appropriately trained technicians authorized by CZM AG. These activities are described in this service manual and are dealt with in detail during the service training for this product. Only original spare parts must be used. For maintenance and repair, proper ESD protection measures, the tools and devices specified in the tools list must be used.

1.4 General Safety Instruction If you need to gain access to the interior electronics of the IOLMaster, especially the computer base, always disconnect the instrument from any and all power sources first. For maintenance and repair, proper ESD protection measures, the tools and devices specified in the tools list must be used (see Section 5). Only original spare parts listed in Section 7 and the official CZM spare part price list must be used. Do not use or install software which is not approved and tested by CZM. Other procedures than described in this manual are not allowed to be performed by the field service engineer. After any kind of manipulation or changes perform a System Checkout in accordance to the instrument checklist and initiate a technical safety check according to Section 2, especially after the service of electronic parts directly connected to the main power, e.g. power supply, main switch and power entry module.

1.5 Disposal of the product within the EU EU directives and national regulations currently in force at the time of marketing prohibit the disposal of the product specified on the delivery note in domestic waste or by municipal waste disposal companies. For more information on disposal of the product, please contact your local dealer, the manufacturer or their legal successors. Please also take note of the manufacturer’s current online information. If the product or its components are resold, the seller has the duty to notify the buyer that the product must be disposed of in accordance with currently valid national regulations.

SM-30-4054-en Section 1, Revision A, Feb. 2010

Confidential and Proprietary

1-5

Service Manual Part No. 1836-541

Section 2 – Checkout and Preventive Maintenance 2.1 Checkout and Maintenance ............................................................................................3 2.2 Technical and Electrical Safety Check .............................................................................3 2.2.1 General Information ........................................................................................................ 3 2.2.2 Procedure of the technical safety check ........................................................................... 3 2.2.3 Reading points ................................................................................................................ 5

SM-30-4054-en Section 2, Revision A, Feb. 2010

Confidential and Proprietary

2-1

Service Manual Part No. 1836-541

Checkout and Preventive Maintenance

IOLMaster 5.5

Notes:

Service Manual Part No. 1836-541

Confidential and Proprietary

2-2

SM-30-4054-en Section 2, Revision A, Feb. 2010

IOLMaster 5.5

2.1

Checkout and Preventive Maintenance

Checkout and Maintenance

Procedure: Please follow the instructions of the Checkout and Maintenance form (see Section 8 Appendices) as accurate as possible. For any nonconformity please perform the necessary adjustments and alignments.

2.2

Technical and Electrical Safety Check

2.2.1 General Information The detailed description of the technical safety check procedure can be found in Service Bulletin 400. This Bulletin is for all Carl Zeiss Meditec AG medical products applicable and equates the latest international rules for electrical safety. Additionally, this service manual also describes the most important procedures inclusive the instrument-depending varieties, e.g. the reading points for the different tests. Note If you have national rules for electrical safety, SB 400 is in force only analogous. In this case you have to fulfil the values of your national standards (e.g. NFPA 99, AAMI ES-1). Only qualified technicians shall perform these tests. Qualification shall include training on the subject, knowledge, experience, and acquaintance with the relevant technologies and regulations. The technician shall be able to assess safety and shall be able to recognize possible consequences and hazards arising from non-conforming equipment.

2.2.2 Procedure of the technical safety check Complete the following points in line with the test procedures of the Instrument Checklist: • • •

Check that the housing (in particular the ventilation slots) is undamaged and free from contamination. Check all cables (in particular the insulation) for damage. Check that all notes and labels are complete and correctly positioned using the user manual.

Checking the electrical safety (This test must be performed as part of the start-up procedure, the annual maintenance service and technical safety check, or after repairs affecting safety-relevant components). During the technical safety check, the device must be tested for electrical safety. Perform the measurements in accordance with IEC 62353:2007 and DIN EN 62353:2008. − The measuring results must be documented in the technical safety check report (See SB 400)

SM-30-4054-en Section 1, Revision A, Feb. 2010

Confidential and Proprietary

2-3

Service Manual Part No. 1836-541

Checkout and Preventive Maintenance

IOLMaster 5.5

Procedure: 1. Measuring protective ground resistance • A maximum test voltage of 24V and a test current of 10A AC are recommended for measurement. • Connect the device with its connecting cable to the power socket of the safety tester (e.g. RIGEL 288). • Firmly press the test probe against a screw head on the housing of the IOLMaster (see Figure 2 – 1) • Start the measurement and note down the measured value. • If the IOLMaster is powered via the power socket of the IT table, repeat the measurement for the table. The same applies if the system is operated via a multiple socket. Include the multiple socket into the measurement in this case. • Note down the highest value measured. • The limit value that must not be exceeded is 0.3 Ohms for devices with a power cord and 0.5 Ohms for systems with a multiple socket. 2. Measuring the leakage current • We recommend using the differential current method, where the device is measured during operation and both phase directions are taken into account. The IOLMaster connected and operated via the safety tester. • When using the RIGEL 288 tester, select IEC 62353, differential measurement and semi automatic mode (to boot and shut down the computer manually during reversal of polarity). • Please note the risk of a computer damage using safety testers with fully automatic reversal of polarity! • Switch on the instrument at the main switch. • Firmly press the test probe against a screw head on the housing of the IOLMaster (see Figure 2 – 1 and Figure 2 – 2) and start measurement. The tester automatically performs a reversal of polarity. • Note down the highest value measured. • The limit value of 0.5 mA must not be exceeded in any of the measurements. 3. Measuring the insulation resistance • Perform this measurement at 500V DC. • Connect the power cord of the IOLMaster to the power socket of the safety tester. • Switch on the instrument at the main switch. • Start the measurement. • Note down the measured value. • The measured value must not be less than the limit value of 1 MOhm. • If the IOLMaster is operated via the power socket of the instrument table and/or via a multiple socket, repeat the measurements accordingly.

Service Manual Part No. 1836-541

Confidential and Proprietary

2-4

SM-30-4054-en Section 2, Revision A, Feb. 2010

IOLMaster 5.5

Checkout and Preventive Maintenance

2.2.3 Reading points According to Service Bulletin 400 following tests are necessary: • • •

Measuring of protective earth impedance Measuring of leakage currents Measuring of insulation resistance

For all measurements and tests during the electrical safety check please use the two points according to Figure 2 – 1 and Figure 2 – 2.

Fig. 2 - 1: Computer base

SM-30-4054-en Section 1, Revision A, Feb. 2010

Fig. 2 - 2: Head rest

Confidential and Proprietary

2-5

Service Manual Part No. 1836-541

Section 3 – Parts Removal / Replacement 3.1 Instrument head.............................................................................................................4 3.1.1 Opening the instrument head....................................................................................... 4 3.1.2 Head covers / spacers................................................................................................... 4 3.1.3 Complete head exchange ............................................................................................. 5 3.1.4 Control cable ............................................................................................................... 6 3.1.5 Display complete.......................................................................................................... 7 3.1.6 TFT Display and LED-Converter ..................................................................................... 8 3.1.7 Keratometer complete.................................................................................................. 8 3.1.8 Slit module................................................................................................................... 9 3.1.9 CCD Camera ................................................................................................................ 9 3.1.10 Digital IOL PCB ........................................................................................................... 14 3.1.11 APD PCB .................................................................................................................... 15 3.1.12 Aperture / diaphragm module .................................................................................... 15 3.1.13 Aperture / diaphragm (Sensor).................................................................................... 15 3.1.14 Magnification changer (Motor)................................................................................... 16 3.1.15 Magnification changer (Sensor) .................................................................................. 17 3.1.16 Interferometer Motor ................................................................................................. 17 3.1.17 Interferometer edge detector (Sensor) ........................................................................ 18 3.2 Computer Base .............................................................................................................19 3.2.1 Opening the Computer Base ...................................................................................... 19 3.2.2 Complete computer base ........................................................................................... 20 3.2.3 Hard drive (HDD) ........................................................................................................ 20 3.2.4 Power supply ............................................................................................................. 21 3.2.5 Fan ............................................................................................................................ 22 3.2.6 Power ON/OFF PCB .................................................................................................... 22 3.2.7 Motherboard ............................................................................................................. 23 3.2.8 Battery ....................................................................................................................... 24 3.2.9 Inductive Sensor (left/right eye switch)........................................................................ 24 3.2.10 Master switch and power entry module ..................................................................... 25 3.3 Instrument Base / Cross Stage ......................................................................................26 3.3.1 Opening the instrument base ..................................................................................... 26 3.3.2 Complete instrument base.......................................................................................... 27 3.3.3 Joystick with release button ....................................................................................... 27 3.3.4 Spur gear ................................................................................................................... 28 3.3.5 Spindle nut / spindle complete.................................................................................... 28 3.3.6 Cross table locking knob ............................................................................................ 29 3.4 Head Rest.....................................................................................................................29 3.4.1 3.4.2 3.4.3 3.4.4

Complete head rest .................................................................................................... 30 Chin rest unit ............................................................................................................. 30 Chin rest .................................................................................................................... 30 Forehead strap ........................................................................................................... 31

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3-1

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

Notes:

Service Manual Part No. 1836-541

Confidential and Proprietary

3-2

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

Adjustment / Calibration Requirements Following Replacements This table identifies checks/adjustments that must be made following removal/replacement procedures. The numbers listed under Follow-ups Required are keyed to the Follow-up Actions listed in the second column. For each Assembly Replaced, the follow-up numbers are listed in the order in which they should be performed, not necessarily in numerical sequence.

Assembly/Part Replaced Instrument head Complete head exchange............ Control cable……………………. Display complete......................... TFT Display and LED-Converter…. Keratometer complete………...... Slit module…………………….. CCD Camera…………………….. Digital IOL PCB…………………... Aperture / diaphragm module... Aperture / diaphragm (Sensor)…. Magnification changer (Motor)…………………………... Magnification changer (Sensor)………………………….. Interferometer Motor...…………. Interferometer edge detector…... Computer Base Opening the computer base……. Complete computer base……….. Hard drive (HDD)……….............. Power supply……………………. Fan……………………………….. Power on/off PCB……………….. Motherboard…………………..... Battery…………………………… Master switch and power entry module…………………………... Induction sensor (left/right eye switch)…………………………… Instrument Base / Cross Table Joystick with release button…….

Follow-ups Required 1 1 1 1 11, 13, 12, 7, 8, 18, 17, 1 13, 17, 1 6, 17, 18, 19, 20, 1 4, 3, 5, 6, 7, 8, 9, 21, 1 12, 1 12, 1 12, 14, 1

Follow-up Actions 12345678-

14, 1 9, 1 10, 1

910 -

1 15, 1 15, 1 22, 1, 23 1 1, 23 1 1

11 12 13 14 15 16 17 18 -

1, 23 1

19 20 -

1

21 22 23 -

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 3

System Checkout (Section 8 B) Optic cleaning (4.2.1) Laser power “Adjustment power” (4.2.3) Laser power "100% power" (4.2.4) Laser Monitoring (4.2.5) CCD camera sensitivity (4.2.6) Surrounding field illumination (4.2.7) Keratometer LED illumination (4.2.8) Interferometer-motor (4.2.9) ALM Measurement range alignment (4.2.10) Keratometer alignment (4.2.11) Aperture alignment (4.2.12) Slit module alignment (4.2.13) Magnification changer alignment (4.2.15) Re-imaging (4.3.1) ALM calibration (4.4.1) ACD calibration (4.4.2) Keratometer calibration (4.4.3) WTW calibration (4.4.4) Spot size calibration (4.4.5) Read / Write EEPROM (4.1.8) Power supply Voltages (4.3.6) Technical/Electrical Safety Check (Section 2)

Service Manual Part No. 1836-541

Parts Removal / Replacement

3.1

IOLMaster 5.5

Instrument head

3.1.1 Opening the instrument head Removal: 1) Loosen six hex socket screws (2.5 mm hex key) on the left cover (viewed from user side). Note that the screws are of different sizes when reassembling. 2) Remove cover. 3) Loosen only 3 screws (2.5 mm hex key) as shown in red in figure 3 - 1. 4) Remove right cover together with the spacers.

Reassembly: Reassemble in reverse order. Please do not tighten the screws as far as it will go.

Fig. 3 – 1: Head covers

3.1.2 Head covers / spacers Spare parts:

1836-516

Set of spare covers for IOLMaster V5.5

Removal: 1) Remove both head covers – see 3.1.1 2) Unscrew the 3 spacers (TORX TX10) from the right cover. 3) Replace covers or spacers as necessary. Reassembly: Reassemble in reverse order. Please do not tighten the screws as far as it will go.

Service Manual Part No. 1836-541

Confidential and Proprietary

3-4

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.1.3 Complete head exchange Spare part:

1772-810

IOLMaster head incl. packaging

Removal: 1) Remove both head covers – see 3.1.1 2) Disconnect control cable from the base. 3) Disconnect joystick cable from X209 of Digital IOL PCB and remove any cable fastener if present 4) Remove three fastening screws (4 mm hex key) of the instrument head and take off the instrument head.

Reassembly:

Fig. 3 – 2: Instrument head

Reassemble in the reverse order by following the instructions below. 1) Attach the new instrument head and secure it with three (4 mm hex key) fastening screws. 2) Measure resistance (grounding) between base plate and head frame with an appropriate meter. Ensure that both spots (especially on the head frame) are bare metal (screw heads should NOT be anodized). The resistance must be less than or equal to 0.5 Ω. Ensure that you zero out the meter first, to allow for any lead resistance. Note: Proper grounding is essential for a good axial-length SNR reading. Without good grounding the instrument may be able to read the test-eye (good SNR), but will be unable to read human eyes. 3) Reconnect joystick cable to X209 of the Digital IOL PCB. 4) Reconnect the control cable to the base. 5) Attach both covers and secure them with screws – see 3.1.1 6) Perform a System Checkout in accordance with the instrument checklist.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 5

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.1.4 Control cable Spare part: 1761-131

Cable IOLMaster (X601) - head

Removal: 1) Disconnect control cable from the computer-base. 2) Remove both head covers – see 3.1.1 Note: Please be extremely careful with the tiny cables and do NOT rip them out! 3) 4) 5) 6) 7) 8)

Disconnect connector X302 from the APD module. Disconnect connectors X201 and X231 from the Digital IOL PCB. Cut all cable ties Loosen the two cable clamps (2.5 mm hex key) Remove cable connections from the LCD display – see details at 3.1.5 Completely remove the control cable.

Reassembly: Reassemble in reverse order observing the following instructions. 1) Reconnect the LCD cable connections - see details at 3.1.5 2) Install the two cable clamps (2.5 mm hex key) 3) Reconnect the connectors X201 and X231 to the Digital IOL PCB. 4) Reconnect X302 to the APD module. 5) Attach the control cable clip. 6) Measure resistance (grounding) between base plate and head frame with an appropriate meter. Ensure that both spots (especially on the head frame) are bare metal (screw heads should NOT be anodized). The resistance must be less than or equal to 0.5 Ω. Ensure that you zero out the meter first, to allow for any lead resistance. Note: Proper grounding is essential for a good axial-length SNR reading. Without good grounding the instrument may be able to read the test-eye (good SNR), but will be unable to read human eyes. 7) Mount both head-covers – see 3.1.1 8) Perform a System Checkout in accordance with the instrument checklist

Service Manual Part No. 1836-541

Confidential and Proprietary

3-6

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.1.5 Display complete Spare part:

1770-344

LCD, without cable

Removal: 1) Remove both head covers - see 3.1.1 2) Disconnect the control cable from the base and disconnect the connectors from the Digital IOL PCB and APD- see 3.1.4 3) Cut the cable tie 4) Loosen the cable clamp close to the LC-display. 5) Loosen but do not remove the two M5 screws (4 mm hex key) on the LCD support, remove the display and place it together with the cable on a clean, padded surface (screen facing downward). 6) Open the LCD housing, carefully (TORX TX10) 7) Carefully disconnect cables from the LCD (CN1 and CN2 on TFT Color Display first) –see Fig. 3 – 3, and place down the display. 8) Disconnect display cable connections from the LEDconverter unit (CN1) Note: Taping the connectors together (in a row, see Fig. 3 – 4) will make the removal of the display cable easier. 9) Remove the strain relief device and pull the cables with connectors out of the housing.

Fig. 3 – 3: Display (opened)

Reassembly: Reassemble in reverse order observing the following instructions. 1) Exercise caution when tightening the screws for the cable Fig. 3 – 4: Display connectors clamp. The mount is plastic, and over tightening of the metal screws can cause the threads of the mount to break. 2) Carefully remove the adhesive tape or heat shrink tubing on the two connectors, plug the connectors onto LEDconverter unit (CN1 and CN2) and TFT panel. 3) Close and secure the display housing (do not bruise the tiny cables) with the removed screws. Mount the display on the IOLMaster head and press it till the screw heads, routing the cable to the left side of the housing. 4) Carefully reconnect the plugs of the control cable. 5) Adjust the friction of the swivel with a 3 mm hex key, if Fig. 3 – 5: adjustment of the friction necessary (Fig. 3 – 5) 6) Perform a System Checkout in accordance with the instrument checklist. SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 7

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.1.6 TFT Display and LED-Converter Spare part:

0510-510 0510-511

TFT-Color-Display 6.5-inch LED-Converter Unit for Backlight

Removal: 1) Remove both head covers – see 3.1.1 2) Remove the display and open the enclosure – see 3.1.5 3) To exchange a damaged LED-converter disconnect CN1 and CN2 first (see Fig. 3 – 3) and unscrew the two fixations screws with a crosstip screwdriver 4) For an exchange of the TFT-display just remove it from the front cover. Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount the LED-converter with the fixation screws and connect the cables to CN1 and CN2 (see Fig. 3 – 3) with great caution. 2) Be very careful and do not bruise the tiny cables during the reassembly. 3) Perform a System Checkout in accordance with the instrument checklist.

3.1.7 Keratometer complete Spare part:

1757-077

Keratometer, complete

Removal: 1) Remove both head covers – see 3.1.1 2) Disconnect the two connectors at the slit assemblies and those of the Keratometer-LEDs at the Digital IOL PCB (X217, X218 and X214). 3) Remove three fixation-screws (2.5 mm hex key) and carefully remove the whole Keratometer unit (see Fig. 3 – 6 item A). Fig. 3 – 6: Keratometer

Reassembly:

Reassemble in reverse order observing the following instructions. 1) Assemble Keratometer unit and loosely secure it with the three screws (2.5 mm hex key). 2) Plug-in connectors. 3) Perform the mechanical alignment – see 4.2.11 4) Check and if necessary align slit groups – see 4.2.13 5) Check and if necessary align aperture group – see 4.2.12 6) Check and if necessary adjust surrounding field illumination – see 4.2.7 7) Check and if necessary adjust KER-illumination – see 4.2.8 8) Check and if necessary calibrate KER – see 4.4.3 9) Check and if necessary calibrate ACD – see 4.4.2 10) Perform a System Checkout in accordance with the instrument checklist. Service Manual Part No. 1836-541

Confidential and Proprietary

3-8

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.1.8 Slit module Spare part:

1757-148

ACD slit module (identical for left and right slit module)

Removal: 1) Remove both head covers – see 3.1.1 Note: Observe ESD protection measures when working on electronic components. 2) Disconnect connector on the relevant slit module. 3) Remove two screws (1.5 mm hex key) and remove slit module. Note! Fig. 3 – 7: Slit module Do NOT remove both slit modules at the same time, unless you are replacing the entire Keratometry assembly. One slit module is the reference for the other. In the event that both modules must be replaced, either replace the entire Keratometry/Slit assembly, or replace the slit PCBs one at a time, aligning the first before attempting to replace the second. Reassembly: Reassemble in reverse order observing the following instructions. 1) Loosely screw new slit module on. 2) Reconnect cable to Digital IOL PCB. 3) Align the slit-module – see 4.2.13 4) Check and if necessary calibrate ACD – see 4.4.2 5) Perform a System Checkout in accordance with the instrument checklist.

3.1.9 CCD Camera Spare part:

1770-340

CCD camera set, 9V

Note: The (redesigned) alignment device for IOLMaster is absolutely vital for replacing the CCD camera. Without this tool it is impossible to verify the required focus level of the camera. This may lead to incorrect measurement, especially with regard to ACD measurements. Alignment issues between the overview and Keratometry positions may also present themselves. The exchange of the CCD camera without using the (redesigned) alignment device is not permitted. Old alignment devices (bought before July 2008) can be retrofitted. Please send the related tools to our Jena (Germany) repair center.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 9

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

Identifying a “worn out” CCD camera Note: Before determining whether a camera is “worn out”, several preconditions must be checked and / or set on an IOLMaster to ensure that the camera parameters are properly adjusted. 1) Preconditions: ƒ Proper connections ƒ Frame grabber settings (see 4.1.3) ƒ Clean optics (see 4.2.1) ƒ Camera sensitivity voltage settings (see 4.2.6) ƒ Surrounding field illumination settings (see 4.2.7) ƒ Aperture alignment (see 4.2.12) 2) Signs for a worn out CCD Camera ƒ Most easily noticed in the Overview mode, a black strip will be seen along the left edge of the Fig. 3 – 8: Frame Grabber settings camera image. This black strip will encompass the full height of the camera image, and will widen with time. Fig. 3 - 9 shows a typical black strip. ƒ When in the Overview mode or/and Keratometer mode the illumination and fixation markers will have a black shadow to the side. In extreme cases the shadows will appear as bars that trail to the side as indicated in Fig. 3 - 9.

Checklist before replacement Black stripe at the left edge

Note: The following points are vital to be checked as “YES” to ensure a successful camera replacement.

Black shadows/trails beside the illumination and fixation lights

Fig. 3 – 9: Worn out camera

1) Camera alignment tool (See 5.3 – tool list) available? 2) Live-image visible? Note: Heating up a CCD camera which does not show any image can easily re-activate it. Use a heat gun for this purpose. Be extremely carefully while blowing hot air into the instrument. There is a high risk of damaging other components. 3) CCD camera is still in the original position (compare crosshair position between Overview and Keratometer mode with the test-eye) and measurements (ALM) show good results in relation to the SNR value? Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 10

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

Note: If the CCD camera is no longer in its original position and / or the SNR of the ALM measurements is not within specification, the replacement of the camera must be aborted, and the measurement head must be sent in for repair in house. Removal: Note: The following procedure describes only the replacement of the CCD camera. The positioning will only be as good as the old / worn-out camera was. It is not possible to re-align the camera if once the original position has been lost or if the present camera is misaligned. This requires either a head-replacement or in-house repair. 1) Remove both head covers – see 3.1.1

1

2) Remove the test-eye from the chin rest, if present. 3) Attach the alignment tool to the Keratometer. Position the tool 30° tilted from the axis and center the ring to the hole of the Keratometer. Rotate the tool until the hooks are behind the Keratometer frame and tighten the screws carefully. Check that the tool is immovable. 4) If not already done, switch on the IOLMaster, create a test patient and change to the Keratometer mode. The sphere side of the rotatable head must be in the direction to the instrument.

2

5) Focus the spots (using the Z-drive screw of the tool) and align the X and Y axis in relation to the crosshair. Loosen the knurledhead screw to position the sphere. Tighten the screw when done. Note: The image in Keratometer mode must be exactly centered as shown below before proceeding further.

Fig. 3 – 10: Aligning the tool

6) Cut off all cable ties (zip ties) from the monitor and control cable. Remove screws from any cable clamps which may be present. Loosen the two screws (4 mm hex key) which hold the LCD in place, and carefully lay the LCD next to the instrument. Exercize caution so as not to stretch the cable. Note: Observe ESD protection measures when working on electronic components. 7) Shut down the instrument and disconnect the camera cable from Digital IOL PCB (X219) 8) Remove the four screws (2 mm hex key) and replace the CCD camera SM-30-4054-en Section 3, Revision A, Feb. 2010

Fig. 3 – 11: Tool positioning

Confidential and Proprietary

3- 11

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

9) Loosely secure it with the four screws (2 mm hex key) and plug-in the camera cable (X219) 10) Adjust the camera sensitivity voltage as necessary (2.9 ± 0.3V at R75, next to X219) and align the camera to the tool (in Keratometer mode). Remember that the image on the screen must be exactly centered. Be as accurate as possible while aligning the camera. 11) Check overview and Keratometer position by moving the magnification changer manually in PIOTest (motor-tab). If the CCD camera was well positioned, both crosshairs should match with the sphere (no horizontal / vertical gap). Focus shift determination / compensation 1) Rotate the tool-head 180° (“black-finished” back-side must face to the instrument). 2) Change to Keratometer mode and run PIO-Test (service-mode). Go to “Chanel 3” of the PIO-Ports tab and switch on the slit groups alternately.

Fig. 3 – 12: Focus shift

3) By turning the Z-axis screw of the tool, the background of the tool-head can be focused. If the focus level of the CCD camera matches to the optical alignment of the instruments, both slits are projected to the middle of the crosshair and do not “virtually” move while switching on the left and right hand side of the slit groups alternately. As a tolerance range, the inner circle of the crosshair can bee seen. While focused, the slit projections must lie within this (inner) circle. 4) If there is a focus shift present, the slits are not projected as described in Fig 3 - 12 - they are out of the middle position while the background is focused – see Fig. 3 - 13

Fig. 3 – 13: Focus shift Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 12

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

5) See the Fig. 3 - 14 and 3 - 15 to determine if the focus level of the CCD camera is too far in the front or too far in the back. 6) If the left slit is projected too far left and the right slit projected too far right, the focus level of the CCD chip is too far back – add shims between the white camera bases and the metal plate.

Fig. 3 – 14: Focus shift – CCD chip too far back

7) If the left slit is projected too far right and the right slit projected too far left, the focus level of the CCD chip is too far in the front – add shims between the casting and the metal plate of the camera.

Fig. 3 – 15: Focus shift – CCD chip too far in the front

Note: Adding shims will require you to take the camera out again. Follow the related sections of the replacement procedure. Afterwards continue to determine the focus shift again. Note: Always add equal numbers of shims to all four fixation points – either one shim (0.2 mm) or two shims (0.4 mm) at each point. 8) Continue as long as the focus shift meets the tolerance range (While the background is focused, the slit projections must lie within the inner circle of the crosshair). Reassembly: Reassemble in reverse order observing the following instructions. 1) Reassemble the instrument in reverse order. Secure the LCD cable loop safely using zip ties. Ensure free movement for the changer. Remove the tool and reattach the head covers. Quit PIOtest and service mode 2) Take measurements with the test-eye and verify the calibration. Re-calibrate KER and ACD if needed.-see 4.4.3 and 4.4.2 3) Compare SNR values (ALM measurements) of “before” and “after” the camera replacement, values should be equal. 4) Perform a System Checkout in accordance with the instrument checklist.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 13

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.1.10 Digital IOL PCB Spare part:

1534-730

Digital IOL PCB (DIGIT)

Removal: 1) Enter Service mode, System tab, Read/Write EEPROM and write a new ini file as a backup of the calibration data – see 4.1.8 2) Enter Service mode, System-tab and run Windows® Explorer copy and paste the created ini file on a reliable storage – see 4.1 3) Shut down and switch off the instrument. Note: Observe ESD protection measures when working on electronic components. 4) Remove both head covers – see 3.1.1 5) Disconnect all connectors. It is recommended to mark the connectors which look similar. Please be extremely careful with the tiny cables and do NOT rip them out! Caution: Immediately connect short-circuit plug (X221) Fig. 3 – 16: IOL PCB into the connector (X216) of the removed laser diode cable. 6) Remove three screws (2.5 mm hex key) on the fixation points (see Fig. 3 – 16 item A) and remove PCB. Note: Watch that the spacers between the PCB and frame do not get lost. Re-secure them with adhesive if necessary. Reassembly: Reassemble in reverse order observing the following instructions. 1) Adjust potentiometer R7, R9 to minimum (several turns anticlockwise) to ensure that the laser diode will not be overloaded and damaged. 2) Re-plug all connections. 3) Switch the instrument on (errors related to the laser power and/or the changer may appear). 4) Adjust laser power (100%) - see 4.2.4 5) Adjust laser power (adjustment power) - see 4.2.3 6) Adjust laser monitoring - see 4.2.5 Fig. 3 – 17: Digital IOL PCB with 7) Adjust camera-sensitivity - see 4.2.6 marked pots 8) Adjust surrounding field illumination - see 4.2.7 9) Adjust Keratometer illumination - see 4.2.8 10) Adjust motor speed - see 4.2.9 11) Compare calibration data and if necessary reload them from the iol.ini file which has been made before the removal - see 4.1.8 12) Perform a function test in accordance with the checklist. Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 14

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.1.11 APD PCB Æ The APD PCB is NOT a field replaceable item. Replacement of the APD PCB MUST be performed at a Carl Zeiss Meditec In house repair center.

3.1.12 Aperture / diaphragm module Spare part:

1756-302

Keratometer diaphragm module

Fixation screws

Aperture assembly

Removal: 1) 2) 3) 4)

Shut down and switch off the instrument. Remove both head covers – see 3.1.1 Disconnect connectors X205, X212 on the Digital IOL PCB. Remove two screws (2.5 mm hex key) (see Fig. 3 - 18) and remove aperture assembly (diaphragm module).

Reassembly: Reassemble in reverse order observing the following instructions. 1) Insert new aperture assembly and secure it loosely. 2) Insert and secure connectors X205, X212 on the Digital IOL PCB. 3) Align aperture assembly – see 4.2.12 4) Perform a System Checkout in accordance with the instrument checklist.

3.1.13 Aperture / diaphragm (Sensor) Spare part:

1693-881 Optical sensor diaphragm X205

Removal:

Fig. 3 – 18: Diaphragm / Aperture

1) Remove the aperture assembly- see 3.1.11 2) Mark the position of the sensor as exactly as possible 3) Remove two screws (2.5 mm hex key) of the sensor (see Fig. 3 – 19) Reassembly: Reassemble in reverse order observing the following instructions. 1) Insert new sensor according to the marking 2) Attach the diaphragm module- see 3.1.11 3) Perform a System Checkout in accordance with the instrument checklist. SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 15

Fig. 3 – 19: aperture sensor

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.1.14 Magnification changer (Motor) Spare part:

1693-816

Changer motor

Removal: 1) 2) 3) 4) 5) 6)

Shut down and switch off the instrument. Remove both head covers – see 3.1.1. Remove diaphragm assembly – see 3.1.11. Disconnect connector X211 on the Digital IOL PCB. Remove the spindle nut. Remove two screws (2 mm hex key) and remove motor with spindle.

Motor

Screws

Spindle

Spindle nut

Reassembly:

Fig 3 – 20: Changer motor

Reassemble in reverse order observing the following instructions. 1) Install new motor with spindle and tighten two screws (2 mm hex key). 2) Attach spindle nut and align spindle in such a way that the changer can be easily moved and the spindle goes through the center of the motor. Note: Should a vertical deviation of the measuring image become noticeable after installation of the motor, this can be aligned by slightly turning the spindle (fine positioning between 2 full steps of the step motor). 3) Tighten spindle nut and plug connection X211. 4) Mount and align diaphragm module –see 3.1.11 and 4.2.12 5) Perform a System Checkout in accordance with the instrument checklist.

Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 16

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.1.15 Magnification changer (Sensor) Spare part:

1693-867

Optical sensor changer X207

Removal: 1) Shut down and switch off the instrument. 2) Remove both head covers – see 3.1.1 3) Disconnect connector X207 on the Digital IOL PCB. 4) Remove 2 securing screws (2.5 mm hex key) and take out the sensor. Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount new sensor and loosely secure it with screws. 2) Re-connector X207 on the Digital IOL PCB. 3) Align the sensor position – see 4.2.15, section “Sensor”. 4) Perform a System Checkout in accordance with the instrument checklist.

Sensor Fig. 3 – 21: Changer sensor

3.1.16 Interferometer Motor Spare part:

1755-837

Motor interferometer, complete

Removal: 1) Shut down and switch off the instrument. 2) Remove both head covers – see 3.1.1 3) Disconnect motor cable from X208 on the Digital IOL PCB. 4) Move the prism slide to the bottom 5) Remove the two screws (2.5 mm hex key) and remove the motor.

Reassembly:

Fig 3 – 22: Interferometer motor

Reassemble in reverse order observing the following instructions. 1) Install motor ensuring that smooth drive motion is observed. 2) Adjust the interferometer motor speed - see 4.2.9 3) Perform a System Checkout in accordance with the instrument checklist.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 17

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.1.17 Interferometer edge detector (Sensor) Spare part:

1693-879 1694-348

Optical sensor interferometer X203 Optical sensor interferometer X204

Removal: 1) Shut down and switch off the instrument. 2) Remove both head covers – see 3.1.1. 3) Disconnect the cable of the damaged sensor, either X203 (blue/upper) or X204 (red/lower), see Fig. 3 - 23 4) Move the prism slide in a good position, where the screws are reachable 5) Remove the two fixation screws (2.5 mm hey key) and remove the sensor

Fig 3 – 23: Interferometer sensor

Reassembly: Reassemble in reverse order observing the following instructions. 1) Install the new sensor. Ensure that the flag of the interferometer sled can easily pass through the sensor. 2) Reconnect the sensor cable at the appropriate place 3) Align the ALM measurement range- see 4.2.10 4) Perform a System Checkout in accordance with the instrument checklist.

Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 18

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

3.2

Parts Removal / Replacement

Computer Base

Fig. 3 – 24: IOLMaster Computer base

3.2.1 Opening the Computer Base Removal: 1) Shut down, switch off the instrument and unplug the main power cord. 2) Disconnect control cable from the base. 3) Remove four screws (2 mm hex key) on the connector panel. 4) Remove eight M5 hex socket head screws (3 mm hex key). Note: It is recommended to position the instrument over the edge of its table in order to access the screws under the base. After removing the screws from one side, then position the instrument over the other side of the table to access the remaining screws. 5) Place the computer base with the head rest upright again. 6) Lift the base cover with the head rest and measuring head off the base plate and put it aside. Fig. 3 – 25: Opening the base

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 19

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

Reassembly: Reassemble in reverse order observing the following instructions. 1) When reattaching the computer base, take care not to bend the left/right switch (sensor). 2) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 3) Perform a System Checkout in accordance with the instrument checklist.

3.2.2 Complete computer base Spare part:

1773-042

Computer IOLMaster with cover

Removal: 1) If possible, back up the existing lens and user database before exchanging the computer base - see user manual for details. 2) Shut down and switch off the instrument. 3) Open the computer base – see 3.2.1. Reassembly: Reassemble in reverse order observing the following instructions. 1) When reattaching the new computer base, take care not to bend the left/right switch (sensor). 2) Tighten the four screws (2 mm hex key) on the Fig. 3 – 26: Computer connector panel before the screws on the bottom of the base. 3) Restore regional settings – see 4.3.1 4) Restore program settings – see 4.3.1 5) Install the backed up lens- and user database – see 4.3.1 6) Re-install printer drivers – see 4.3.1 7) Re-install software-options– see 4.3.1 8) Restore network settings – see 4.3.1 9) Create a local ghost image – see 4.3.1 10) Perform a System Checkout in accordance with the instrument checklist.

3.2.3 Hard drive (HDD) Spare part:

0509-551

Fig 3 – 27: computer base (bottom view)

Hard disk 2.5” SATA

Removal: 1) If possible, back up the existing lens and user database before exchanging the HDD - see user manual for details. 2) Shut down and switch off the instrument. 3) Open the computer base – see 3.2.1 4) Turn the computer base around 180 degrees Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 20

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

5) Remove the four mounting screws (2.5 mm hex key), see Fig. 3 - 27 (item A). Exercise caution so that the screws do not drop out (Place tape over the holes). 6) Turn again the computer base and carefully unplug the HDD in direction of the control cable connector Reassembly: Reassemble in reverse order observing the following instructions. 1) Connect the new HDD and secure it with the four screws (2.5 mm hex key) 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) Image the HDD from the related Service-USB flash drive – see 4.3.1 5) Restore regional settings – see 4.3.1 6) Restore program settings – see 4.3.1 7) Install the backed up lens and user database – see 4.3.1 8) Re-install printer drivers – see 4.3.1 9) Re-install software-options– see 4.3.1 10) Restore network settings – see 4.3.1 11) Create a local ghost image – see 4.3.1 12) Perform a System Checkout in accordance with the instrument checklist.

3.2.4 Power supply Spare part:

0508-752

Power supply 90 W 12 V (DC)/ 7.5 A

Removal: 1) 2) 3) 4) 5) 6)

Shut down and switch off the instrument. Disconnect the main power cord. Open the computer base – see 3.2.1 Cut all of the zip ties from the power supply. Disconnect all cables. Remove the four screws (2 mm hex key) from the bottom side (see Fig. 3 – 27, item B). Exercise caution as the power supply may drop down.

Reassembly: Reassemble in reverse order observing the following instructions. 1) Do not forget to reconnect any and all ground connections! The latch of the blade connector should be set in the direction to the screw head. 2) Check and if necessary adjust the output voltage – see 4.3.6 3) When reattaching the computer base, take care not to bend the left/right switch (sensor). 4) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 5) Perform a System Checkout inclusive the technical safety check in accordance with the instrument checklist and Section 2. SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 21

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.2.5 Fan Spare part:

1693-872

Fan complete

Removal: 1) 2) 3) 4)

Shut down and switch off the instrument. Open the computer base – see 3.2.1 Disconnect the cable connection of the fan and remove the cable clamp. Remove two screws (2 mm hex key) from the bottom side (see Fig. 3 – 27, item C) and replace the fan.

Reassembly: Reassemble in reverse order observing the following instructions. 1) Tighten the screws loosely. 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) If the computer base is closed, slide the fan (after loosening the screws from the bottom side) as far as possible in the direction of the cover (avoid any vibration or noise), then tighten all screws. 5) Perform a System Checkout in accordance with the instrument checklist.

3.2.6 Power ON/OFF PCB Spare part:

1671-563 1694-129 1694-354

Power ON/OFF PCB Cable for power ON/OFF (X853 to X602) Power ON/OFF - mains cable

Removal: 1) 2) 3) 4) 5)

Shut down and switch off the instrument. Disconnect the main power cord. Open the computer base – see 3.2.1 Disconnect the X851, X885 and X853 connections Remove the two M3 screws with 2.5 mm hex key and one M3 screw with an 2.0 mm hex key and remove the Power ON/OFF PCB

Reassembly: Reassemble in reverse order observing the following instructions. 1) Re-connect all connection. 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) Perform a System Checkout inclusive the technical safety check in accordance with the instrument checklist and Section 2. Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 22

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.2.7 Motherboard Spare part:

1528-161

IOLMaster MED-COM PCB

Removal: 1) Shut down and switch off the instrument. 2) Open the computer base – see 3.2.1

1

Note: Observe ESD protection measures when working on electronic components. 3) Disconnect all cables shown in the illustration (Fig. 3 – 28). 4) Remove the mounting bolts for the VGA and COM port (4.5 mm hex socket wrench) 5) Remove the other screws (2.5 mm hex key) and remove the board 6) Remove the HDD – see 3.2.3 Reassembly: Reassemble in reverse order observing the Fig 3 – 28: MED-COM Motherboard following instructions. 1) Assemble the used HDD and the new motherboard 2) Insert new motherboard and secure it in position. 3) Re-connect all cables. 4) When reattaching the computer base, take care not to bend the left/right switch (sensor). 5) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 6) Perform a System Checkout in accordance with the instrument checklist.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 23

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.2.8 Battery Spare part:

0509-839

Lithium coin cell 3V/230mAh 3,2x20mmØ

Removal: 1) Shut down and switch off the instrument. 2) Open the computer base – see 3.2.1 3) Remove the battery with a tool, e.g. a screw driver (see Fig. 3 – 29) Note: Place the defective battery in an approved depleted battery receptacle for proper disposal.

Reassembly:

Fig 3 – 29: BIOS CMOS backup Battery

1) Mount the new battery 2) Verify the proper battery voltage of 3.0 V DC or higher using a calibrated digital volt meter. 3) When reattaching the computer base, take care not to bend the left/right switch (sensor). 4) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 5) Perform a System Checkout in accordance with the instrument checklist.

3.2.9 Inductive Sensor (left/right eye switch) Spare part:

1693-874

Inductive sensor switch

Removal: 1) Shut down and switch off the instrument. 2) Open the computer base – see 3.2.1 3) Disconnect the sensor (X603) and remove two screws (3 mm hex key) Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount the new sensor at the bar 2) When reattaching the computer base, take care not to bend the left/right switch (sensor). It is recommended to open the white plate at the base cover to check the position of the sensor. 3) Tighten the four screws (2 mm hex key) on the connector panel before the screws on the bottom of the base. 4) Perform a System Checkout in accordance with the instrument checklist.

Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 24

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.2.10 Master switch and power entry module Spare part:

1693-874 0505-635

Master switch, 2 pins, green power entry module medical version 3A

Removal: 1) Shut down and switch off the instrument. 2) Disconnect the main power cord 3) Open the computer base – see 3.2.1 Master switch: 1) Disconnect the four cables 2) Compress the snap latches and remove the switch Power entry module: 1) 2) 3) 4)

Remove the fan – see 3.2.5 Remove the master switch Disconnect the three cables Carefully compress the lower and upper snap latch with an appropriate screw driver and remove the power entry module

Reassembly: Reassemble in reverse order observing the following instructions. 1) Mount the new part Note: Take care of the grounding measures and connect all grounding cables at the right place!

Fig 3 – 30: master switch and power entry module

2) Reinstall the fan– see 3.2.5 3) Perform a System Checkout inclusive the technical safety check in accordance with the instrument checklist and Section 2.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 25

Service Manual Part No. 1836-541

Parts Removal / Replacement

3.3

IOLMaster 5.5

Instrument Base / Cross Stage

Fig. 3 – 31: Instrument base

3.3.1 Opening the instrument base 1) Shut down and switch off the instrument. 2) Move the instrument to the highest position (by using the joystick). Note: Always use the right screw driver size and be very careful. Do not scratch the paint. 3) 4) 5) 6)

Remove the two screws on the side of the housing and remove the housing. Remove both head covers – see 3.1.1 Remove the two screws on the side of the cover and remove the cover. Remove the front plate (four screws)

Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 26

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.3.2 Complete instrument base Spare part:

266101-4501-927

Cross Stage for IOLMaster

Removal: 1) Shut down and switch off the instrument. 2) Remove the complete measurement head – see 3.1.3 3) Remove the four screws of both gear racks (2.5 mm hex key), see Fig. 3 - 32 4) Remove the instrument base Reassembly: Reassemble in reverse order observing the following instructions. 1) Install the gear rack onto the computer base, insert the instrument base and install the second toothed rack. Fig. 3 – 32: gear rack Note: Make sure that the gears are identically positioned on the left and right. The shaft must have an endplay clearance of approx. 1/10 mm in each position. 2) Check upper and lower stop of the horizontal movement range and readjust if necessary.

3.3.3 Joystick with release button Spare part:

266101-4560-514

IOLMaster joystick with release button

Removal: 1) 2) 3) 4)

Shut down and switch off the instrument. Open the instrument base – see 3.3.1 Move instrument head to the lowest position (by using the joystick) Mark the position of the joysticks stop-pin and the stop at the spur gear. Note: Before exchanging gears, mark their precise position and insert the new part accordingly. The subsequent adjustment of the upper and lower stops may take some time.

5) Disconnect joystick cable on X209 of Digital IOL PCB and remove cable fixation if present. 6) Remove the four fixation screws (2.5 mm hex key) and remove joystick assembly. Reassembly: Reassemble in reverse order observing the following instructions. 1) Re-position the joystick as the markings show. 2) Check upper and lower stop of the horizontal movement range and readjust if necessary.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 27

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.3.4 Spur gear Spare part:

266101-4533-010

Spur gear

Removal: 1) 2) 3) 4)

Shut down and switch off the instrument. Open the instrument base – see 3.3.1 Move instrument head to the lowest position (by using the joystick). Mark the position of the joysticks stop-pin and the stop at the spur gear. Note: Before exchanging gears, mark their precise position and insert the new part accordingly. The subsequent adjustment of the upper and lower stops may take some time. 5) Mark the position of the defective gear, and do not move the joystick afterwards. 6) Remove the spur gear (loosen the screw with 3 mm hex key). Reassembly:

Reassemble in reverse order observing the following instructions. 1) Apply the marking to the new spur gear. 2) Insert the new spur gear in accordance with the position marking and tighten the fixation screw. 3) Test the upper and lower stops of the movement range, and readjust if necessary.

3.3.5 Spindle nut / spindle complete Spare part:

266101-4502-514 266101-4510-010

Spindle, complete Spindle nut

Removal: 1) Shut down and switch off the instrument. 2) Open the instrument base – see 3.3.1 3) Move instrument head to the lowest position (by using the joystick). Note: Before exchanging gears, mark their precise position and insert the new part accordingly. The subsequent adjustment of the upper and lower stops may take some time. 4) Remove the two screws (2.5 mm hex key) and then remove the spindle nut. 5) Remove the screws (2.5 mm hex key) and then remove the spindle.

Reassembly: Reassemble in reverse order observing the following instructions. 1) When positioning the new spindle nut, take care that the same horizontal position in relation to the spindle will be chosen. 2) Test the upper and lower stops of the movement range, and readjust if necessary. Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 28

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.3.6 Cross table locking knob Spare part:

1775-095

Service Set cross table locking knob IOL

Removal: 1) Remove the cap with an appropriate tiny screw driver 2) Loosen the screw and remove the turning knob. Reassembly: Reassemble in reverse order observing the following instructions. 1) Be careful to replace the parts in their proper order. 2) Press the knob down as far as it will go and fasten the fixation screw. 3) Test the proper function. Fig. 3 – 33: Service Set locking knob

3.4

Head Rest

Fig. 3 – 34: Head rest SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 29

Service Manual Part No. 1836-541

Parts Removal / Replacement

IOLMaster 5.5

3.4.1 Complete head rest Spare part:

1025-670

Head rest

Removal: 1) Shut down and switch off the instrument. 2) Remove head rest from the fixtures, loosen the two locking screws with a 4 mm hex key (Fig. 3 - 34 – item 1) and pull the head rest out.

Reassembly: Reassemble in reverse order observing the following instructions. 1) When putting the head rest back into the fixtures, take care that the tubes are completely in before tightening the fixation screws.

3.4.2 Chin rest unit Spare part:

1294-613

Chin rest unit, assembled

Removal: 1) Shut down and switch off the instrument. 2) Remove head rest - see 3.4.1 3) Slide the chin rest unit off from the head rest tube.

Reassembly: Reassemble in reverse order observing the following instructions. 1) When putting the head rest back into the fixtures, take care that the tubes are completely in before tightening the fixation screws.

3.4.3 Chin rest Spare part:

319752-8016-000

Chin rest

Removal: 1) Shut down and switch off the instrument. 2) Remove the fixation screw (3 mm hex key), see Fig. 3 - 34 item 3 3) Remove the chin rest.

Reassembly: Reassemble in reverse order. Service Manual Part No. 1836-541

Confidential and Proprietary

3 - 30

SM-30-4054-en Section 3, Revision A, Feb. 2010

IOLMaster 5.5

Parts Removal / Replacement

3.4.4 Forehead strap Spare part:

319752-0024-000

Forehead strap

Removal: 1) Shut down and switch off the instrument. (2.5 mm hex key) 2) Remove the two fixation screws (2.5 mm hex key), see Fig. 3 - 34 item 2 3) Remove the forehead strap

Reassembly: Reassemble in reverse order observing the following instructions. 1) Fasten the two screws with great care, do not break the plastic of the strap. Note: Do not use any adhesives for locking the two screws. Chemical reactions may weaken the forehead strap and lead to breakage under load.

SM-30-4054-en Section 3, Revision A, Feb. 2010

Confidential and Proprietary

3- 31

Service Manual Part No. 1836-541

Section 4 – Adjustment and Calibration 4.1 Service Mode..................................................................................................................3 4.1.1 Entering the Service Mode............................................................................................ 3 4.1.2 Calibration Tabs (ALM, ACD, KER, WTW)...................................................................... 4 4.1.3 Grabber Tab................................................................................................................. 4 4.1.4 Prog Tab ...................................................................................................................... 5 4.1.5 System Tab .................................................................................................................. 5 4.1.6 Delete Data .................................................................................................................. 6 4.1.7 Speed Test ................................................................................................................... 6 4.1.8 Read / Write EEPROM................................................................................................... 7 4.1.9 PIO-Test ....................................................................................................................... 7 4.2 Adjustments / Alignments of the Measuring Head .........................................................9 4.2.1 Optics cleaning ............................................................................................................ 9 4.2.2 Power supply voltage Digital IOL board ...................................................................... 14 4.2.3 Laser power “Adjustment power” .............................................................................. 15 4.2.4 Laser power "100% power" ....................................................................................... 16 4.2.5 Laser Monitoring........................................................................................................ 17 4.2.6 CCD camera sensitivity ............................................................................................... 18 4.2.7 Surrounding field illumination..................................................................................... 18 4.2.8 Keratometer LED illumination ..................................................................................... 19 4.2.9 Interferometer-motor ................................................................................................. 20 4.2.10 ALM Measurement range alignment .......................................................................... 22 4.2.11 Keratometer alignment .............................................................................................. 24 4.2.12 Aperture alignment .................................................................................................... 24 4.2.13 Slit group alignment................................................................................................... 25 4.2.14 APD alignment ........................................................................................................... 26 4.2.15 Magnification changer alignment ............................................................................... 26 Sensor........................................................................................................................ 27 Flag............................................................................................................................ 29 Changer Steps............................................................................................................ 30 4.3 Adjustments / Alignments of the computer..................................................................31 4.3.1 Re-imaging................................................................................................................. 31 Re-imaging options .................................................................................................... 31 Re-imaging procedure ................................................................................................ 32 Installation of Software Version 5.5............................................................................ 33 Settings after re-imaging ............................................................................................ 34 4.3.2 HDD Structure (Databases and Logfiles)...................................................................... 36 4.3.3 Database Exchange .................................................................................................... 36 4.3.4 Deleting of corrupted Databases ................................................................................ 36 4.3.5 Deleting the administrator-password .......................................................................... 37 4.3.6 Update the Software .................................................................................................. 37 4.3.7 Power supply Voltages ............................................................................................... 37 4.4 Calibration....................................................................................................................38 4.4.1 ALM calibration.......................................................................................................... 38 4.4.2 ACD calibration .......................................................................................................... 39 4.4.3 Keratometer calibration.............................................................................................. 40 4.4.4 WTW calibration ........................................................................................................ 41 4.4.5 Spot size calibration ................................................................................................... 41 SM-30-4054-en Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4-1

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Notes:

Service Manual Part No. 1836-541

Confidential and Proprietary

4-2

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

4.1 Service Mode The IOLMaster has an embedded service mode which can be entered with special passwords. The passwords are case sensitive and are not permitted to distribute to any non-authorized person. The usage of the service mode is restricted to certified service engineers, trained from an authorized Carl Zeiss Meditec trainer.

4.1.1 Entering the Service Mode 1) The service mode can be activated via Æ “Options” Fig. 4 - 1: Entering Options Æ “Service” 2) A password prompt appears. Enter the related password and confirm. 3) Depending on whether the “Service” password or “Calibration” password is used, the Service Menu will open differently. While all calibration pages are able to be viewed with the service password, the calibration password also allows changes within these pages and so the sections are highlighted.

Fig. 4 - 2: Entering Service mode

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 3

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Fig. 4 - 4: Calibration mode

Fig. 4 - 3: Service mode

Note: Usage of the calibration password is only recommended for the purpose of calibration. Exiting the calibration mode by clicking “OK” will cause the EEPROM to be re-written. Because of the delays which the rewriting causes, the rewriting of the EEPROM should be minimized to extend its lifetime.

4.1.2 Calibration Tabs (ALM, ACD, KER, WTW) Æ See “4.4 Calibrations” for detailed information.

4.1.3 Grabber Tab The frame grabber settings are of vital importance in relation to all image analyzing measurements of the IOLMaster (ACD, KER and WTW). These settings are to remain at the factory default values and must not be changed at any time.

Fig. 4 - 5: Grabber Tab

Service Manual Part No. 1836-541

Confidential and Proprietary

4-4

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

4.1.4 Prog Tab The Prog Tab provides the “Save” section, the “Output” section and the “Database” section. The defaults are shown in Fig. 4 - 6. “Save” section The “Grab Images” checkbox can be activated to highlight the checkboxes next to it. These checkboxes can be individual activated to save the images during the measurement (KER, ACD or WTW). The related images will be saved in the IOLMaster folder and may be used later for detailed troubleshooting

Fig. 4 - 6: Prog Tab

Note: After the images have been made, the feature must be deactivated manually again. Otherwise, the speed of the measurements will slow down and the HDD will be overloaded with images. “Output” section Activating the “Detailed Output” checkbox will show the detailed measurement of the KER. Instead of an average measurement, the five individual results will be displayed. “Manual R/L Switch” allows switching via touchpad between the left and right eye side, without moving the instruments head. The magnetic sensor which normally detects the right and left side is temporary disabled. To switch between both sides, just click in the related column (OD or OS). “Save and show Raw Signal” is only used at the factory. It shows the raw signal in ALM mode. “Database” section With both of these checkboxes checked, older patient files will be deleted from the HDD after the time specified in “Program Settings” menu has elapsed. Unchecking these boxes should not be done, as this would allow the HDD to fill up.

4.1.5 System Tab The system tab page provides links to the Windows® Explorer, the Control Panel, the PIO-Test, the Speed Test and the EEPROM. It is also possible to erase some data via the “Delete Data” link

Fig. 4 - 7: System tab SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 5

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

4.1.6 Delete Data The Delete Data submenu enables several options. – see Fig. 4 - 8 If the related checkboxes are checked, the corresponding data will be erased by clicking “OK”. The following window shows the selection again and needs to be confirmed to finally delete the data.

Fig. 4 - 8: Delete data

4.1.7 Speed Test The speed test allows direct access and control to the interferometer motor. Æ See 4.2.9 Interferometer motor for details. The menu buttons also allows for opening any of the other speed files, which are stored at C:\ (root directory) as “vmess_hsp3.dat”. Speed test files can also be printed.

Service Manual Part No. 1836-541

Confidential and Proprietary

4-6

Fig. 4 - 9: Speed-Test Navigation

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

4.1.8 Read / Write EEPROM The EEPROM is the main location of all calibration data. There is also a backup file for this data, which can be found in the IOLMaster folder. The name of the file is “iol.ini”. “Write INI File” This option will manually create a backup file (ini-file). The destination can be chosen within a popup-window. Please note that the destination must be entered by hand (in the “File name” field) and can not be selected by browsing folders. Note: The changer steps will also be written into the *.ini-file. Fig. 4 - 10: EEPROM

“Read INI File” This option reads an existing ini-file. The destination can be chosen within a popup-window. Please note that the destination must be entered by hand and can not be selected by browsing folders. “Write to EEPROM” This option re-writes the EEPROM with the data currently displayed in the sections of the EEPROMscreen (see Fig. 4 - 10). The instrument must be restarted after the EEPROM has been rewritten. It is highly recommended to compare the data of the ini-file and the EEPROM afterwards to verify the success of the writing procedure. Open the related ini-file via “Notepad” and compare the sections with those of the EEPROM screen. If they are not the same, repeat the procedure.

4.1.9 PIO-Test The PIO Test contains three tabs which can be used for field-service. The “Inbetriebnahmetest” tab is only for the manufacturing. For efficient use of the PIO-Test, the service-mode must be closed. Therefore click outside of the PIOTest window to get back to the main service mode. Exit the service mode and get back to the PIOTest window by using the + key-combination on the keyboard.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 7

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

PIO-Ports-tab The PIO-Ports tab contains several sections for manual control of the instruments functions. Channel 0: Channel 0 provides information about laser and APD. By checking the available radio buttons, the related information will be displayed as increments (numbers) within the numeric section within this channel. Channel 1: Clicking the check box switches the overview illumination LED’s on. Channel 2: The check boxes and radio buttons within this channel control the laser diode. Also the fixation LED can be controlled within this channel.

Fig. 4 - 11: PIO-ports

Channel 3: The LED’s of the Keratometer and those of the slitgroups (ACD = VKT) can be controlled with the radio buttons of this channel. Channel 4: The interferometer motor movement (up and down) can be controlled within this channel. Channel 5: Several radio buttons allow a manual frame grabber exposure control.

Motors tab Section Step motors Both of the step motors of the aperture and changer can be controlled manually within this section. Furthermore, the changer steps can be set here.

Fig. 4 - 12: Motors-tab

Test Hardware Pressing this button will start a complete hardware test of the system. Any errors which may occur will be displayed in the window next to it. All error messages are also logged in a file named as “czgif.log” and is located in D:\IOLMaster folder. Permanent Test This button will run a permanent test of the systems hardware. The test can be stopped by clicking the button again. At the end of the test a log window will popup containing information about start and end time and errors.

Service Manual Part No. 1836-541

Confidential and Proprietary

4-8

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

EEPROM tab The EEPROM tab contains a section for the “Serial Number” of the instrument, which accepts 7 digits. In this example, the serial number which is saved will be printed on the speed test printouts. (See 4.1.7).

4.2

Adjustments / Alignments of the Measuring Head Fig. 4 - 13: EEPROM

4.2.1 Optics cleaning Tools and Material required o

Alpha Swab with Long Handle TX 761

part-no.: 000000-0433-973

Fig. 4 - 14: Micro fiber swab

OR o

Cotton Swab with wooden handle

part-no.: 000000-0500-895

Fig. 4 - 15: Cotton swab

o

Optics cleaning mixture “L” (50ml Bottle)

part-no.: 000000-1526-414

Fig. 4 - 16: Optic cleaning mixture SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 9

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Optics Cleaning Procedure Note: The instrument must be switched off for cleaning – do not switch the instrument on before the cleaning mixture has been dried off. Use Carl Zeiss cleaning mixture “L” for cleaning only. This cleaning agent is approved. Be carefully while cleaning components near painted surfaces. Use small quantities at one time – do not “flood” the instrument. Keep the cleaning mixture clean (do not dip the dirty wiper into the mixture bottle). The cleaning should be made in the following sequence: -

Cleaning of the test eye Cleaning of the optical surfaces of the Keratometer assembly (outside and inside) Cleaning of the optical surfaces (inside the head).

1) Cleaning the test eye

Fig. 4 - 17: Cleaning the Test eye

2) Cleaning the optical surfaces of the Keratometer assembly

Fig. 4 - 18: Cleaning Keratometer Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 10

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

3) Cleaning the optical surfaces (inside the head). Note: It is recommended to perform inside the head cleaning with compressed-air, as the inside components normally are only contaminated with dust (no grease contaminants). If using canned compressed air, do not allow liquid propellant to strike optical surfaces, as spotting may result on the surfaces. If liquid cleaning is necessary (e.g. fingerprints), proceed as showed in the following sections.

ƒ Beam splitter

Fig. 4 - 19: Beam splitter

a) Top side

b) Bottom side

ƒ Interferometer triple prism

Fig. 4 - 20: Triple prism

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 11

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

ƒ Prism 26

Fig. 4 - 21: Prism 26

a) Backside

b) Bottom side

a) Frontside

b) Backside

ƒ Prism 34

Fig. 4 - 22: Prism34

ƒ Lens (behind prism 34)

Fig. 4 - 23: lens

Service Manual Part No. 1836-541

a) Frontside

b) Backside

Confidential and Proprietary

4 - 12

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

Magnification changer lenses

Fig. 4 - 24: Changer lenses

a) upper lens

b) lower lens

ƒ Camera mirror (small)

Fig. 4 - 25: Camera mirror

Caution: After successfully cleaning the optics, the laser output power must be checked and adjusted. See 4.2.3 and 4.2.4 for detailed procedure. Laser power adjustment also requires adjustment of the laser monitoring.

+++ Perform a function test in accordance with the checklist. +++

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 13

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

4.2.2 Power supply voltage Digital IOL board Procedure: 1) 2) 3) 4)

Remove the six hex socket screws (2.5 mm hex key) from the left cover only. Remove the cover. Switch on the instrument. Measure the supply voltages as Fig. 4 - 26 shows. Correct values will ensure a proper functionality of the measurement head. Note: If the voltage at pin 1 (see Fig. 4 - 26) is too low, an adjustment can be made at the power supply (computer base). See 4.3.6 for details.

´

Service connector X220: X220: X220: X220:

1 +12V 2 + 9V 3 + 5V 4 GND

Fig. 4 - 26: Power supply voltage / digital IOL board

+++ Perform a function test in accordance with the checklist. +++

Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 14

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

4.2.3 Laser power “Adjustment power” Note: After adjustment of the laser power, the laser monitoring must also be re-adjusted. Special Tools: ƒ Power meter (such as Coherent Fieldmaster or Ophir Nova) ƒ Adequate probe for the wavelength of 780 nm (such as LM2 or PD300) ƒ Optional recommended: Measuring adapter 1094-546 for the LM2 measuring head. Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. 3) Enter service mode and start the PIO Test – see 4.1. 4) Set power meter to 780 nm as well as in an adequate range for the expected power of 40 μW. 5) Remove any present filters from the probes (e.g. PD300) and offset the probe for ambient light. 6) Switch off the fixation LED. It is further recommended to switch off also the overview LEDs and KER LEDs and slit groups if the laser measuring head is positioned not directly on the front lens. 7) Select "Adjustment power" in Channel 2 and check the "Laser On/Off" checkbox for “On”. 8) Position the probe close to the front lens and move slightly for the best measurement result. 9) Use “Motor forward” / “Motor backward” in Channel 4 to determine the position with the highest measured power.

Fig. 4 - 27: Laser power adjustment

Caution: If the laser power is determined as too low, ensure that the optical path is clean before adjusting the laser power – see 4.2.1. Note: If the laser power is unstable, the most likely cause is that the laser diode is defective. Caution: The highest reading of the both positions must be adjusted according the default (40 μW). It is absolutely normal that the lower reading differ from the higher. This difference can be up to 5 μW. It is not required that the lower reading must lie in the tolerance range. 10) Use R7 potentiometer on the Digital IOL PCB (see Fig. 4 - 27) to set the adjustment laser power to (40 ± 2) μW.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 15

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Note: If once the laser power has been adjusted, the laser monitoring must be adjusted as well. See 4.2.5. +++ Perform a function test in accordance with the checklist. +++

4.2.4 Laser power "100% power" Note: After adjustment of the laser power, the laser monitoring must also be re-adjusted. Special Tools: ƒ Power meter (such as Coherent Fieldmaster or Ophir Nova) ƒ Adequate probe for the wavelength of 780 nm (such as LM2 or PD300) ƒ Optional recommended: Measuring adapter 1094-546 for the LM2 measuring head. Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. 3) Enter service mode and start the PIO Test – see 4.1. 4) Set the power meter to 780 nm as well as in an adequate range for the expected 100% power of 400 μW. 5) Remove any present filters from the probes (e.g. PD300) and offset the probe for straight light. 6) Select "100% power" in Channel 2 and check the "Laser On/Off" box for “On”. 7) Switch off the fixation LED. It is further recommended to switch off also the overview LEDs and KER LEDs and slit groups if the laser measuring head is positioned not directly on the front lens. Caution: Do not operate the laser diode at "100% Power" for more than 5 minutes.

8) Position the probe close to the front lens and move slightly for the best measurement result. 9) Use “Motor forward” / “Motor backward” in Channel 4 to determine the highest measurement position. Caution: If the laser power is determined as too low, ensure that the optical path is clean before adjusting the laser power – see 4.2.1 Note: If the laser power is unstable, the most likely cause is that the laser diode is defective.

Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 16

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

Caution: The highest reading of the both positions must be adjusted according the default (400 μW). It is absolutely normal that the lower reading differ from the higher. This difference can be up to 50 μW. It is not required that the lower reading must lie in the tolerance range. 10) Use R9 potentiometer on the Digital IOL PCB (see Fig. 4 - 27) to set the adjustment laser power to (400 ± 5) μW. Note: If once the laser power has been adjusted, the laser monitoring must be adjusted as well. See 4.2.5 +++ Perform a function test in accordance with the checklist. +++

4.2.5 Laser Monitoring Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. 3) Enter service mode and start the PIO Test – see 4.1. 4) Switch off the fixation LED in Channel 2. 5) Click on "Laser monitoring" (Channel 0) and "100% Power" (Channel 2); if necessary, click "Laser on/off" to switch the box inactive. 6) Note down displayed value as "Laser-off reading", depends on ambient conditions (typical reading approx. 128) – see Fig. 4 - 28 part a) 7) Switch on the laser ("Laser on/off" checkbox active) and note down measurement result (Channel 0) as "Value 100% Power" (typical reading approx. 48). – see Fig. 4 - 28 part b) Caution: Do not operate the laser diode at "100% Power" for more than 5 minutes!

a)

b)

c)

8) The difference between the measured values ("Value dark signal" minus "Value 100% Power") must be 80 ± 3 increments. 9) In the case of deviations, adjust the R26 potentiometer (at the APD –PCB, see Fig. 4 - 29) accordingly.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 17

Fig. 4 - 28: Laser monitoring

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Caution: The APD board has more than one potentiometer. Ensure that you adjust only R26. A wrong adjustment of any of the other pots can not be corrected in the field. Do not change the settings of the two other potentiometers on the APD PCB. 10) After completion of the potentiometer adjustment, switch to the "Adjustment Power" function and note down the measurement result as "Adjustment power reading (typical reading approx. 120). – see Fig. 4 - 28 part c) 11) The difference between the measured values ("Value dark signal" minus "Value adjustment power") must be 8 ± 2 increments.

Fig. 4 - 29: Monitoring adjustment

+++ Perform a function test in accordance with the checklist. +++

4.2.6 CCD camera sensitivity Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Switch on the instrument and measure the sensitivity voltage between any of the groundpins and the pin (P1) below R75 on the Digital IOL PCB. - see Fig. 4 - 30 3) Adjust the sensitivity voltage at DC (2.8 ± 0.3) V

+++ Perform a function test in accordance with the checklist. +++ Fig. 4 - 30: reading points for camera sensitivity

4.2.7 Surrounding field illumination Preconditions: ƒ Frame grabber settings according the defaults – see 4.1.3. ƒ CCD camera sensitivity adjusted according the specifications (and according the camera generation). Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 18

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only. 2) Switch on the instrument and select overview mode. Note: It is of vital importance, that the “Check, Device” patient has been selected and the Test Eye is in front of the instrument. 3) Focus the Test Eye (surface!) in overview mode, so that the grainy structure appears in focus. 4) Increase the illumination with R3 on the Digital IOL PCB (see Fig. 4 - 31) until the image seems to be too bright (white areas are visible at the Test Eye’s surface). 5) Now reduce the illumination until the surface of the Test Eye is well illuminated and the image is without glare (no white areas). Note: If it is impossible to increase/decrease the surrounding field illumination in a way to get a well illuminated Test Eye surface, the camera sensitivity voltage can be increased/decreased slightly within the tolerance range.

Fig. 4 - 31: Potentiometer for adjustment of Field and KER illumination

+++ Perform a function test in accordance with the checklist. +++

4.2.8 Keratometer LED illumination Preconditions: • • •

Frame grabber settings according the defaults – see 4.1.3. CCD camera sensitivity adjusted according the specs (and according the camera generation). Aperture alignment OK – see 4.2.12.

Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. Fig. 4 - 32: Calibration setting 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 3) Position the Test Eye in front of the instrument and focus the spots, so that also the halo ring will be visible. Note: It is of vital importance, that no external ambient light strikes the instrument or Test Eye while adjusting the illumination.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 19

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

4) Use potentiometer R40 on the Digital IOL PCB (see Fig. 4 - 31) to adjust the brightness of the Keratometer spots in such a way that they appear without glare - with a bright center and a clearly visible, but weaker halo ring. 5) Enter Service Mode, KER-Tab (see Fig. 4 - 32) and uncheck the box for “adjustment aid calibrated” – see 4.1 6) After closing the “Program Setting” dialog with “Save” and change to the Dual mode. You will be prompted to calibrate the Keratometer aid. 7) Position the measure-head so that the spots appear blurry. Ensure that the horizontal and vertical position is OK – the six spots must be visible and appear equal in the cross-hair. 8) Focus the spots by moving the joystick in the direction where the image becomes sharper. Move further in the same direction until the spots appear blurry again before pressing the joystick-button. 9) Now you will be prompted to repeat the process once again for comparison. Please do so. 10) If the calibration process was successful, a message to this effect will be displayed. 11) If a message appears, stating that variance between the calibration values of the two calibrations was too wide, the process (starting at item 8) must be repeated. Fig. 4 - 33: KER-aid calibration 12) After the calibration process, a prompt shows the spot size of the Keratometer-LED’s in focus. The value must lie in a range of 50 – 55. If the value is in range then confirm and calibration procedure would be successful 13) Use R 40 potentiometer to adjust the brightness of the Keratometer spots and repeat the procedure, starting with item 6 until the spot size lies within 50-55. +++ Perform a function test in accordance with the checklist. +++

4.2.9 Interferometer-motor Special Tools: ƒ Thermometer (min. range 10°C - 35°C (50°F – 95°F) / accuracy ± 0,5°C / resolution 0,2°C). ƒ Recommended thermometer: TESTO “testo 608-H1“ – See 5.3 (Toollist).

Procedure: 1) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 2) Enter Service-Mode and run Speed-Test (System – Tab). Run motor about 10 times for „warm up“ and stabilisation of sledge speed – use Æ

Fig. 4 - 34: Speed adjustment Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 20

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

3) Measure the surrounding temperature and determine the corresponding sledge speed which must be adjusted – see diagram below (Fig. 4 - 35). 4) Enter Service-Mode and run Speed-Test (System – Tab) 5) If not already present, change to “Speed-View“, by pressing the “Speed” button. use Æ Adjust first the determined sledge speed in forward direction (up) using potentiometer R64 on the Digital IOL PCB (see Fig. 4 - 34). Adjusting tolerance is ± 0,2 mm/s. 6) Ensure that the displayed speed is reached after 80 ms (first dotted vertical line – black). 74,0

sledge speed [mm/s]

73,0 72,0

71,6 71,0 70,0 69,0

35°C (95°F)

34°C (93,2°F)

33°C (91,4°F)

32°C (89,6°F)

31°C (87,8°F)

30°C (86°F)

29°C (84,2°F)

28°C (82,4°F)

27°C (80,6°F)

25°C (77°F)

26°C (78,8°F)

24°C (75,2°F)

23°C (73,4°F)

22°C (71,6°F)

20°C (68°F)

21°C (69,8°F)

19°C (66,2°F)

18°C (64,4°F)

17°C (62,6°F)

16°C (60,8°F)

15°C (59°F)

14°C (57,2°F)

13°C (55,4°F)

12°C (53,6°F)

11°C (51,8°F)

10°C (50°F)

68,0

temperature [ °C (°F) ]

Fig. 4 - 35: Temperature graph with example

Note: If the determined slide speed can’t be reached after 80ms, the present motor might be hard-steering. In that case the motor must be exchanged! If the problem still persists, the IOLMaster must be sent for in-house repair! 7) Change to “Oscillation-View”, by pressing the Osillation button (see Fig. 4 - 37). Check the maximum and minimum value of the speed: Max value < 81 mm/s Fig. 4 - 36: Speed graph Min value > 61 mm/s Following that, check the speed-oscillation of the motor. If any peak is in the red zones, the slide motor must be re-adjusted / repositioned. While doing the re-positioning, also the motor gear can be manually turned (90°…180°) before placing the motor back in the rail. This may eliminate the peaks in the red area. SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 21

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Note: If the motor cannot be adjusted so that all peaks lie within the white zone, the motor must be replaced! If the problem still persists, the entire measurement head must be replaced, or the instrument must be sent to a CZM repair center for in– house repair. 8) If not already present, change back to „Speed-View“, by pressing the “Speed” button. Afterwards use Æ Adjust the determined sledge speed in Fig. 4 - 37: Speed oszillation backward direction (down) using potentiometer R77 on the Digital IOL PCB (see Fig. 4 - 34). Adjusting tolerance is again ± 0,2 mm/s. 9) Perform the same check as in item 7 described. +++ Perform a function test in accordance with the checklist. +++

4.2.10 ALM Measurement range alignment Preconditions: ƒ Interferometer motor speed adjusted – see 4.2.9 Caution: The ALM measurement range alignment must only be performed with IOLMaster Software Version 7 or higher. Software Version 7 has to be installed before the range can be aligned! Please use the Service USB flash drive (Option 2) and perform all necessary steps before. – see 4.3.1 Procedure: 1) Check the IOLMaster software version and install the Version 7 or higher, if necessary. 2) Remove the six hex socket screws (2.5 mm hex key) from the left cover only and remove the cover. 3) Position the Test Eye in front of the instrument (ALM sphere), choose “Check, Device” patient and select Dual mode press the key until the ALM mode is reached. Fig. 4 - 38: Measurement range (lower) Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 22

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

4) Mark an individual measurement (not the composite measurement) from the active column of measurements and move the white spot manually over the left end of the red graph using the right mouse button. 5) Mark another individual measurement (not the composite measurement) from the active column of measurements and move the white spot manually over the right end of the red graph using the right mouse button. Note: Do not use the left mouse button because the left button will automatically determine any peak in the area of positioning the white spot. In that case you may not be at the end of the measurement range!

Fig. 4 - 39: Measurement range (upper)

Note: Sometimes, if the white spot extends past the end of the range, instead a logical value (e.g. 13.92 mm), a dash “-“ which symbolizes an “error” or a high numeric value appears. In this case the spot is out of the available range. Slightly move the spot (using the right mouse button) in direction to the inside of the measurement range until an mm-value appears. 6) If needed, adjust the measurement range to ≤ 13.5 mm and ≥ 38.5 mm (lower than 13.5 mm and greater than 38.5 mm). It is not needed to widen the range much more than that. In the case of the lower end (13.5 mm) can’t be reached, you have to adjust the lower sensor (edge detector) of the interferometer - you have to move it upwards. Note: After each adjustment of the sensor, at least two new measurements must be taken before verifying the range-values again. ƒ Caution: If the alignment has been completed successfully, please install the Software Version 5.5 again, if it was the previous installed Software Version. – see 4.3.1

Fig. 4 - 40: Sensor alignments

+++ Perform a function test in accordance with the checklist. +++ SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 23

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

4.2.11 Keratometer alignment Preconditions: ƒ Frame grabber settings according the defaults – see 4.1.3 ƒ CCD camera sensitivity adjusted according the specs (and according the camera generation). Procedure: 1) Remove both head covers – see 3.1.1 2) Cover the optical path, to avoid any light interference. 3) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 4) Position the Test Eye is in front of the instrument and focus the spots, so that also the halo rings will be visible. 5) Loosen the three fixation-screws of the Keratometer unit – see 3.1.7 6) Position the Keratometer is such a way, that the

Fig. 4 - 41: Keratometer alignment

spots appear centred in their surrounding halo rings. The two middle spots must also be parallel to and bisected by the centre line of the crosshair. 7) Recalibrate, if necessary, the KER and ACD – see 4.4.3 and 4.4.2 +++ Perform a function test in accordance with the checklist. +++

4.2.12 Aperture alignment Preconditions: ƒ Frame grabber settings according the defaults – see 4.1.3 ƒ CCD camera sensitivity adjusted according the specs (and according the camera generation).

Fig. 4 - 42: Bad Aperture alignment

Procedure: 1) Remove both head covers – see 3.1.1. 2) Cover the optical path, to avoid any light interference. 3) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 4) Position the Test Eye is in front of the instrument and focus the spots, so that the halo rings will also be visible. 5) Loosen the two screws of the aperture-assembly.

Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 24

Fig. 4 - 43: Good Aperture alignment

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

6) Position the aperture in a way that no flare (see Fig. 4 - 42 and Fig. 4 - 43) is shown around the Keratometer spots. The halo rings must be clearly visible around the spots. +++ Perform a function test in accordance with the checklist. +++

4.2.13 Slit group alignment Preconditions: ƒ Frame grabber settings according the defaults – see 4.1.3. ƒ CCD camera sensitivity adjusted according the specs. ƒ No CCD camera focus shift present – see 3.1.9 section “Focus shift determination / compensation”. Procedure: 1) Remove both head covers – see 3.1.1. 2) Cover the optical path, to avoid any light interference. 3) Position the Test Eye (ACD Sphere) in front of the instrument. 4) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select KER mode. 5) Run PIO-Test (service-mode). Go to “Chanel 3” of the PIOPorts tab and switch on the slit group which needs to be adjusted. Ensure that the slit image will be projected to the black surface of the Test Eye (left from the ACD sphere). 6) Focus the black surface of the Test eye by moving the instruments head forward/backward.

Fig. 4 - 44: Slit group alignment

Note: Please align only one slit group at same time, otherwise the second group as a reference get lost. 7) Now align the related slit group in a way, that the projected slit becomes as thin as possible. The position of the slit must be in the middle of the crosshair. The inner circle can be seen as the tolerance range for the adjustment. 8) Double check both slit groups. The projected slit images should virtually “stay” in the same position while switching between both slit groups. If necessary also adjust the other slit group, too. SM-30-4054-de Section 4, Revision A, Feb. 2010

Fig. 4 - 45: Slit group adjustment

Confidential and Proprietary

4- 25

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Note: If one or both slit groups have been adjusted, a re-calibration of the ACD might be necessary. – see 4.4.2 +++ Perform a function test in accordance with the checklist. +++

4.2.14 APD alignment Æ The APD alignment must be performed in our in-house repair center.

4.2.15 Magnification changer alignment

Note: There are three possible alignments of the magnification changer – the sensor, the flag and the changer steps (software setting). The alignments, if necessary, should be done in order. Overview: Magnification optics Direction of movement towards sensor Signal of sensor

Stepp cycle of 2-phase stepper motor

011001100 001100110 Signal change of sensor

Magnification optics

Adjusting and measurement position of magnification changer

Magnification optics

Overview position

Measurement position Overview optics Adjustable actuating flag

Fixed actuating flag

Overview optics Fixed actuating flag

Direction of movement towards sensor

Movement direction 290-300 steps

Adjustable actuating flag

Sensor

Sensor

Fig. 4 - 46: Changer – schematic overview

Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 26

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

Sensor Note: The Flag alignment primarily changes the measurement-position. The overview-position will also be affected by this alignment and must be re-aligned afterwards. The sensor must be adjusted according to the following criteria: ƒ the position according the optical path (rough positioning) and ƒ the position in relation to the changer motor phase (fine positioning) Procedure: 1) Remove both head covers – see 3.1.1. 2) Cover the optical path, to avoid any light interference. 3) Position the Test Eye (ALM Sphere) in front of the instrument. 4) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. Position the measurement head according the sphere. Run PIO-Test (service-mode). Go to “Chanel 2” of the PIO-Ports tab and switch on the laser (100%). Caution: Do not operate the laser diode at "100% Power" for more than 5 minutes.

Fig. 4 - 47: Laser Reflections

5) Change back to ALM mode (click outside the PIO-Test) and position the measurement head in a way, that both, the bright laser spot (primary reflection) and the secondary reflection (see Fig. 4 - 47) are over each other and fix the position of the measurement head. Note: If this position is not roughly aligned with the crosshair, this can be taken as prove that the sensor of the magnification changer must be aligned. 6) Switch between measurement-position and overview-position back and forward. At the moment check only the measurement position. If the position where both laser reflexes lie over each other don’t match with the crosshair position, the sensor must be re-aligned. 7) Loosen the sensor (two screws) and change the position slightly. Depending if the crosshair stays left or right from the laser spot, the sensor must be aligned to the left or right hand side. Note: Whatever will be done at the sensor, do not move the position of the measurement head. It is of vital importance the measurement head keeps positioned as the laser reflections stay over each other.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 27

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

8) Switch again between measurement-position and overview-position back and forward to see the effect of the alignment. Repeat the procedure as long as the crosshair will stay over the laser spot in measuring mode (not overview-mode!). 9) Go to KER-Mode and bring back to the PIO-Test ( + ) and switch the laser back to adjustment power. Change to the Motor-tab and slide the window to bottom of the screen, so that you are able to see the KER spots as well as the control of the changer motor.

Fig. 4 - 48: Changer Steps

Note: You are now in KER mode and should see the spots and the crosshair perfectly aligned. If not, manually move the changer once again (click the checkbox next to “Changer”) to overview- and back to measurement position. If necessary also re-position the head. 10) Click “Step +” several times (at least 8 times). The motor will move the optics (image) now to the left hand side (out of the crosshair). 11) Click “Step –“ until the “End” position switches from 1 to 0 (indicates that the sensor has detected the flag). Continue pressing “Step –“until the “phase” shows 00 and count the number of motor-steps in between of the switch of the end-position and the 00 (The switch itself from 1 to 0 does not count). 12) The accepted range for this “switch-gap” is 2 or 3 steps. Any other number requires a fine adjustment of the sensor position. Loosen the sensor again and slightly move it. Be carefully, the necessary alignment requires a minor positioning only. Note: The fine adjustment may take some time and some attempts. Be accurate with this alignment, because it will ensure the stable positioning of the measurement lens. Remember: All measurements taken by the instrument are related to this positioning! 13) Use the manual control of the changer (PIO-Test / Motor tab) to switch between the measurement-position and overview-position and back to see the effect of the sensor-alignment. Always check the alignment according the laser spots before checking the fine-alignment according the steps (switch-gap). 14) If the sensor is successfully aligned, finally tighten the screws. The sensor doesn’t need to be aligned for the overview position of the magnification changer. +++ Perform a function test in accordance with the checklist. +++ Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 28

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

Flag Note: The Flag alignment will only change the overview-position. The measurement position will not affected by this alignment. Procedure: 1) Remove both head covers – see 3.1.1 2) Cover the optical path, to avoid any light interference. 3) Position the Test Eye (ALM Sphere) in front of the instrument. 4) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. Position the measurement head according the sphere. Run PIO-Test (service-mode). – see 4.1 Change to the Motor-tab and slide the window to bottom of the screen, so that you are able to see the KER spots as well as the control of the changer motor. See Fig. 4 - 48. Note: You are now in KER mode and should see the spots and the crosshair perfectly aligned. If not, manually move the changer once again (click the checkbox next to changer) to overview and back to measurement position. If necessary also re-position the instruments head. 5) Manually move the changer now (click the checkbox next to “Changer”) to the overview-position. Note: Depending upon the step-information (setting in the “Steps” window), the position of the crosshair according the image might be correct or not. Due to this “offset” will be adjusted later, it’s not necessary to be correct for the flag-alignment. Anyhow, do not move the position of the instruments head. It is of vital importance the instruments head keeps positioned according the previously aligned measurement position. 6) Use “Step +“/ “Step –“to manual position the crosshair according the image. 7) If in this position the “End” position indicates 1, the flag must be roughly adjusted first. Loosen the screws and move the flag to the left. If the “end position” indicates 0, skip to section 9. 8) Move the changer manually to the measurement-position and back to overview to see the effect of the alignment. Use again “Step +“/ “Step –“to manual position the crosshair according the image. If still the “end position” shows 1, the flag must be adjusted again. 9) Now click “Step –” several times (at least 8 times). The motor will move the optics (image) now to the right hand side (out of the crosshair). The “end position” should switch from 0 to 1 somewhere in between. 10) Now, click “Step +“as long as the “end position” switches from 1 to 0 (indicates that the sensor has detected the [adjustable] flag). Continue pressing “Step +“until the crosshair matches with the image and count the number of motor-steps in between of the switch of the end-position and the position when the crosshair matches (The switch itself from 1 to 0 does not count). 11) The accepted range for this “switch-gap” is 2 or 3 steps. Any other number requires a fine adjustment of the flag position. Loosen the flag and slightly move it. Be carefully, the necessary alignment requires a minor positioning only.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 29

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Note: The fine adjustment may take some time and some attempts. Be accurate with this alignment, because it will ensure the stable positioning of the overview lens. +++ Perform a function test in accordance with the checklist. +++

Changer Steps Note: The changer steps will ensure a good overview-position only in combination with a well aligned sensor and flag. It is vital to check (or if necessary align) the sensor and flag before setting the changer steps. Procedure: 1) Position the Test Eye (ALM Sphere) is in front of the instrument. 2) Switch on the instrument, choose the “Check, Device” patient or activate the Test Eye function and select ALM mode. Position the measurement head according the sphere. Run PIO-Test (service-mode). – see 4.1 Change to the Motor-tab and slide the window to bottom of the screen, so that you are able to see the KER spots as well as the control of the changer motor. See Fig. 4 - 48 Note: You are now in KER mode and should see the spots and the crosshair perfectly aligned. If not, manually move the changer once again (click the checkbox next to changer) to overview and back to measurement position. If necessary also re-position the instruments head. 3) Manually move the changer now (click the checkbox next to “Changer”) to the overview-position. Listen to the sound of the changer motor. Note: If the motor stops in between very shortly but continues straight away in direction (or exactly) to the overview-position, the number of steps is too less. Most of the instruments have their setting around the 295. It is recommended to set this as a default, before continuing with the adjustment. 4) If, in overview, with this default the crosshair stands too far right of the image, the number of steps must be increased. If it stands too far left of the image, the number of steps must be decreased. Note: Setting (changing) of the changer steps is only possible in overview-mode. The Set button won’t work in measurement mode.

+++ Perform a function test in accordance with the checklist. +++

Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 30

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

4.3

Adjustment and Calibration

Adjustments / Alignments of the computer

4.3.1 Re-imaging Re-imaging options The service USB flash drive menu provides the following options: 1) “Recover Windows XP on system partition from USB-Stick” Note: This option will delete all data on the C:\ partition. Customized data, such as printer drivers, regional settings, computer name, network settings and installed software options will be lost after the re-imaging. 2) “Recover Windows XP including all partitions from USB-Stick” Note: This option will delete all data on the C:\ and D:\ partition. Customized data, such as printer drivers, regional settings, computer name, network settings and installed software options will be lost after the re-imaging. Furthermore all lens, user and patient data will be lost as well as the IOLMaster program settings, such as refractive index, clinic name, language, export settings and so on. Please be prepared for redoing all customized data (C:\ partition) after the re-imaging process. Lens and user data should be available from the customer’s backup If not, these data can be backed up before the re-imaging (as long as the HDD and the related database are not corrupt). Patient data (visits) can NOT be restored after re-imaging. 3) “Recover system backup from local drive” Note: With this option the C:\ partition will be restored from an image normally existing at the D:\ partition (D:\IOLMSYS.GHO). All data on the C:\ partition will be replaced by this from the image file. Customized data, such as printer drivers, regional settings, computer name, network settings and installed software options will be restored as at the time as the image was created. Please be prepared for redoing some customized data if they were not correctly restored (in case, they have not been included once in the image-backup). 4) “Create system backup onto local drive” Note: By choosing this option, the C:\ partition will be backed up into an image file at the D:\ partition (D:\IOLMSYS.GHO). This backup should always been made after changes into the C:\ partition (system) such as printer drivers, regional settings, computer name, network settings, installed software options and so on.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 31

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Re-imaging procedure Note: Before performing the HDD re-imaging, use the following checklist to ensure the completion after the re-imaging. Printer drivers (Option 1, 2, maybe also 3) Software option installation media (Option 1, 2, maybe also 3) Backup of lens and user data (Option 2) Program setting information (Option 2)

Customer property – only approved printers are allowed to be installed. Customer property (Option N, Option T) Customers backup USB flash drive According customers requests

1) Switch off the instrument and switch on again while pressing F2 alternately (do NOT keep pressing) until the BIOS program is started. 2) Enter the BIOS password and change the boot-sequence for “USB HDD” as first priority. Connect the right service USB flash drive (see Tool list 5.3), save settings and quit BIOS (F10). 3) The instrument automatically reboots now from the service USB flash drive. Please make your selection according the needs. While imaging an empty HDD, “Option 2” must be chosen. 4) If option 2 has been chosen, a prompt will appear, asking for the right partition sizes. Check carefully that the full space of the HDD will be used depending upon the size. 5) HDD will be re-imaged now. When finished, please disconnect the USB flash drive and reboot twice. If the re-imaging has been made from the well-kept local system-backup (Option 3) the following steps might not be necessary. Please continue with item 9).

Fig. 4 - 49: Imaging menu

Fig. 4 - 50: Imaging process

Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 32

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

Installation of Software Version 5.5 Note: The installation of Software Version 5.5 is only mandatory, if the reimaging procedure Option 1 or Option 2 has been performed. 1) After either Option 1 or Option 2 always restart the instrument 2) Connect the IOLMaster Software Version 5.5 USB flash drive (see Tool list 5.3) with the IOLMaster 3) Enter the Settings menu, click on the wrench symbol ¼ 4) Click on the Program tab 5) Click on the Update button (see Fig. 4 - 51) 6) Choose the appropriate language 7) Follow the instructions on the screen 8) If a password prompt appears, enter the IOLMaster XP service password and continue with OK (see Fig. 4 - 53) 9) After a successful installation the system will be restarted.

Fig. 4 - 51: Update button with SW Vers. 7

Fig. 4 - 52: Update button with SW Vers. 5.5

Fig. 4 - 53: Password SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 33

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Settings after re-imaging 1) Set “Date & Time” as well as “Time Zone” according to the customers needs. Æ “Options” Æ “Setup”  “Date/Time”

Fig. 4 - 54: Date and Time

2) Set regional settings, such as language and keyboard according to the customer and type of the keyboard used with the instrument. Æ “Options“  “Setup” Æ “Regional Settings” Æ “Languages” tab Æ “Details”

Fig. 4 - 55: Input language

3) Re-install all options (N / T) which are the customer’s property for the related instrument. The option USB flash drive should be provided by the customer. Æ “Options“  “Setup” Æ “SW Option”  “Install” Follow the instructions on the screen. The installed options are displayed after installation (at the system information page).

Fig. 4 - 56: Options

4) Append the serial number to the existing computer name “iolm”. Æ “Options“ Æ “Service“ (password required) Æ “System“ tab Æ “Control Panel“ Æ “System“ Æ “Computer Name“ tab Æ “Change“ Æ Format “IOLMSerial-Number“ (e.g. IOLM987654) Æ “OK“ Fig. 4 - 57: Computer name Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 34

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

5) Re-install the printer. Use only printer according the approved specifications! Unapproved printers are not supported by Carl Zeiss Meditec and may lead into serious problems using the instrument. Æ “Options“  “Setup” Æ “Printer” Note: See user manual for detailed installation of network-printers.

6) Restore lens and user data (Option 2 only) Æ “Options“ Æ “Lens database” Æ Administrator logon required Æ “Restore”

Fig. 4 - 58: Restore lens data

7) Program settings (Option 2 only). Configure the application settings according the customers needs. Æ “Options“ Æ “Setup” Æ “Program Settings” Customize the settings of “Program”, “Export”, “Network” and “User management” tabs as necessary. 8) Create / update local backup-image. Reboot the instrument with the service USB flash drive. The menu shown in Fig. 4 - 49 will appear again. Select option 4 for creating a new image from the C:\ partition. This image will be saved at the D:\ partition (D:\IOLMSYS.GHO)

Fig. 4 - 59: Program settings

9) IMPORTANT: After the image has been successfully created, please enter the BIOS again to change back the boot sequence. Restart the instrument while pressing F2 alternately (do NOT keep pressing) until the BIOS program is started. Enter the password and change boot-sequence for “Hard Drive” as first priority. Save settings and quit BIOS (F10). 10) Remove the service USB flash drive. +++ Perform a function test in accordance with the checklist. +++ SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 35

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

4.3.2 HDD Structure (Databases and Logfiles) IOLMaster : Patient Database: Patient visits User and lens Database Keratometer logfile ALM logfile Error message logfile

D:\DATA\CZM_BaseDB.MDB D:\DATA\ *.iol D:\DATA\ IolmXP.MDB D:\IOLMaster\Keratometer.log D:\IOLMaster\Alm.log D:\IOLMaster\czgif.log

4.3.3 Database Exchange Procedure: 1) Enter service mode / system-tab and run Windows® Explorer. 2) Browse to the D:\ partition. 3) Copy the whole DATA folder to a media (USB flash drive, USB HDD). 4) Now, enter service mode / system-tab and run Windows® Explorer of that instrument where the copied data folder should be replaced. 5) Rename the existing DATA folder (“D:\” partition) of the related instrument to DATAOLD. Note: Without renaming the existing DATA folder, the paste of the backed up DATA folder won’t work. This is because opened files can’t be replaced while they are open. 6) Now copy the backed up DATA folder to the “D:\” partition of the instrument. 7) Reboot the instrument and check the patient data, the user data and lens data. 8) If the exchange was successfully, the DATAOLD folder can be deleted. +++ Perform a function test in accordance with the checklist. +++

4.3.4 Deleting of corrupted Databases 1) Enter service mode / system-tab and run Windows® Explorer – see 4.1 2) Browse to the D:\Tools folder and run “taskmgr”. 3) End the IOLMaster task. 4) Browse back to the D:\Data folder and delete the related database (CZM_BaseDB.MDB and/or IolmXP.MDB). Note: All data will be finally lost if deleting a database. The automatically recreated database will be empty. 5) Also delete the related database (CZM_BaseDB.MDB and/or IolmXP.MDB) in the D:\Data\Save folder. 6) After rebooting the instrument, the deleted database will be automatically re-created (empty!) +++ Perform a function test in accordance with the checklist. +++ Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 36

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

4.3.5 Deleting the administrator-password Procedure: 1) Enter Service mode (system tab). Click “Erase Data” and check the “administrator password” checkbox (see 4.1.6). Confirm with OK. +++ Perform a function test in accordance with the checklist. +++

4.3.6 Update the Software Procedure: 1) Switch on the instrument. 2) Connect the USB flash drive containing the adequate update software 3) Click on “Options”  “Setup”  “Update” and follow the instruction on the screen – See Fig. 4 - 60. +++ Perform a function test in accordance with the checklist. +++

4.3.7 Power supply Voltages

Fig. 4 - 60: Software Update

Procedure: 4) Open the computer base – See 3.2.1 5) Measure the voltage output at the recommended points for the power supply – See Fig. 4 - 61. 6) The output voltage range must be DC (+12 ± 0.1) V. 7) Adjust the voltage with a long and appropriate screw driver at the related potentiometer Vadj – See Fig. 4 - 61. Caution: Be careful with the screw driver and do not damage parts closeby the pot. There is no high voltage within this region, but there is still the risk of electric shock. SM-30-4054-de Section 4, Revision A, Feb. 2010

GND

Vadj +12V

Fig. 4 - 61: Power supply IOLMaster

Confidential and Proprietary

4- 37

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

Note: It is recommended to check the supply voltage at the Digital IOL board, too. This will ensure a proper functionality of the measurement head. See 4.2.2 for details. +++ Perform a function test in accordance with the checklist. +++

4.4

Calibration Note: No other changes as described are permitted to be performed. Changes of other values may lead into wrong measurements.

4.4.1 ALM calibration Procedure: Note: The re-calibration must be done only if the readings of the ALM are out of the tolerance range (± 0.05 mm) of the test-eyes imprinted value. 1) 2) 3) 4)

Install Test Eye (ALM sphere). Take some ALM measurements. Enter Service mode (calibration password), ALM tab – see 4.1 Run the calibration spread-sheet at an external PC (Microsoft® Excel required) and perform the recalibration as Fig. 4 - 62 shows.

Fig. 4 - 62: ALM calibration Service Manual Part No. 1836-541

Confidential and Proprietary

4 - 38

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

5) After determining the new Origin Offset, enter it as the Origin Offset value on the ALM tab. The new ALM calibration Origin Offset value will be saved in the EEPROM and iol.ini-file by clicking OK to leaving the service-mode. 6) Verify the calibration by measuring the test-eye again (ALM measurement). Note: A new patient (visit) must be created to take new readings. If the same patient (as with the old calibration values) would be used, the average reading or composite signal would not give the correct output according the new calibration values. Note: If the deviation between the Test Eye value and the measurement result was quite big, the calibration process (Excel calculation) must be repeated several times. +++ Perform a function test in accordance with the checklist. +++

4.4.2 ACD calibration Preconditions: ƒ ƒ ƒ ƒ ƒ ƒ

Clean Test eye sphere without scratches. Optical path of ACD clean. Frame grabber settings OK (see 4.1.3). Ensure that no ambient light strikes the test eye or enters the exit lens of the IOLMaster head. Well adjusted slit modules (see 4.2.13). CCD camera sensitivity voltage well adjusted (see 4.2.6).

Procedure: 1) Install Test Eye (ACD sphere). 2) Take ACD measurements (with “Check, Device” patient). 3) Enter Service mode (calibration password), ACD tab - see 4.1 4) Run the calibration spread-sheet at an external PC (Microsoft® Excel required). Note: The individual conversion factors for right and left side are always viewed from the patient and not from the user. Do not mix the values up.

Fig. 4 - 63: ACD calibration

5) Enter the values as requested: ƒ Test Eye Value is the value of the test eye which is printed on the test eye label as VKT ƒ Measurement is the value that the instrument reports for the measured ACD. ƒ Conversion Old is taken from the Service Menu ACD tab (before calibration). ƒ Conversion New is the value to correct the ACD calibration.

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 39

Service Manual Part No. 1836-541

Adjustment and Calibration

IOLMaster 5.5

6) Enter the calculated conversion factor into the conversion section of the ACD tab. 7) Verify the calibration by taking new measurements. If necessary repeat the calibration process (Excel calculation). +++ Perform a function test in accordance with the checklist. +++

4.4.3 Keratometer calibration Preconditions: ƒ ƒ ƒ ƒ ƒ ƒ ƒ ƒ

Clean Test eye sphere without scratches. Optical path of KER clean. Frame grabber settings OK (see 4.1.3). Ensure that no ambient light strikes the test eye or enters the exit lens of the IOLMaster head. Well adjusted Keratometer assembly (see 4.2.11). Well adjusted Aperture assembly (see 4.2.11). CCD camera sensitivity voltage well adjusted (see 4.2.6). Keratometer LEDs illumination well adjusted (see 4.2.8).

Procedure: 1) Enter Service mode (calibration password), KER tab – see 4.1 2) Enter the default values in the conversion section and the section for the factors (0°, 60°, and 120°). They are 0.035 / 1 / 1 / 1 (see Fig. 4 - 64) 3) Save the values by leaving the service mode with “OK” 4) Install Test Eye (KER sphere). 5) Take KER measurements. 6) Enter the service mode again, run Microsoft® Explorer (System-Tab) – see 4.1

Fig. 4 - 64: Keratometer calibration

7) Browse to the IOLMaster folder. 8) Open “Keratometer.log” and scroll down to the last block of the log file. 9) Run the calibration spread-sheet at an external PC (Microsoft® Excel required). 10) Enter 5 values each for Kappa 1 / 2 / 3 (from the log file) into the calibration spread-sheet fields highlighted in gray. 11) Enter now the two average values for radius R1 and R2 (from the log file) into the calibration spread-sheet fields highlighted also in gray.

Service Manual Part No. 1836-541

Fig. 4 - 65: Keratometer calibration spread-sheet

Confidential and Proprietary

4 - 40

SM-30-4054-en Section 4, Revision A, Feb. 2010

IOLMaster 5.5

Adjustment and Calibration

12) Enter the imprinted radius value from the Test Eye as the “Radius Test Eye” in the calibration spread-sheet. 13) Open service mode, KER tab and replace the default values with those calculated from the calibration spread-sheet (highlighted in green). 14) Verify the calibration by taking new measurements. If necessary repeat the calibration process (Excel calculation). +++ Perform a function test in accordance with the checklist. +++

4.4.4 WTW calibration Preconditions: ƒ White-to-white measurement activated (“Settings”  “Measurement” tab  “Activate WTW” checkbox) ƒ Clean Test eye sphere without scratches, WTW calibration scale attached. ƒ Optical path of WTW clean ƒ Frame grabber settings OK (see 4.1.3) ƒ Ensure that no ambient light strikes the test eye or enters the exit lens of the IOLMaster head. ƒ CCD camera sensitivity voltage well adjusted (see 4.2.6). ƒ Surrounding field illumination well adjusted (see 4.2.7).

Procedure: 1) Enter Service mode (calibration password), WTW tab. 2) Uncheck the “calibrated” checkbox. 3) Install Test Eye (WTW scale must face to the instrument). 4) Save setting by leaving the service mode with “OK”. 5) Chose WTW – the calibration prompt appears. 6) Position the head in a way to see the full calibration scale in the live image. Focus the scale and press joystick button.

Fig. 4 - 66: WTW calibration

7) Repeat the focusing as prompted once more and press joystick button again. Note: If the calibration has been accepted, the “calibration” checkbox is automatically checked from the system. This procedure must not be made manually. +++ Perform a function test in accordance with the checklist. +++

4.4.5 Spot size calibration Æ See 4.2.8 Keratometer LED illumination

SM-30-4054-de Section 4, Revision A, Feb. 2010

Confidential and Proprietary

4- 41

Service Manual Part No. 1836-541

Section 5 – Troubleshooting 5.1 Troubleshooting Image ..................................................................................................3 5.1.1 General image troubleshooting ....................................................................................... 3 5.1.2 Detailed troubleshooting I (KER – image) ......................................................................... 4 5.1.3 Detailed troubleshooting II (ACD – image) ....................................................................... 5 5.1.4 Detailed troubleshooting III (WTW – image)..................................................................... 6 5.2 Error messages...............................................................................................................7 5.2.1 Error messages during powering up ................................................................................ 7 5.2.2 Error messages during operation ................................................................................... 11 5.2.3 Typical error messages and causes therefore ................................................................. 13 5.3 Tool list IOLMaster .......................................................................................................14

SM-30-4054-en Section 5, Revision A, Feb. 2010

Confidential and Proprietary

5-1

Service Manual Part No. 1836-541

Troubleshooting

IOLMaster 5.5

Notes:

Service Manual Part No. 1836-541

Confidential and Proprietary

5-2

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

5.1

Troubleshooting

Troubleshooting Image

5.1.1 General image troubleshooting Preconditions: - Instrument switched on - Boot procedure complete successfully - No error messages - Test Eye installed and “Test Eye“ Mode active or “Check,Device“ patient choosen

OK

Is Patient Manager visible?

No

Control cable maybe improperly connected. A general troubleshoot for the computerbase would be neccessary.

No

Check the connections of the related cables. If necessary, a slightly warm up of the camera may help. (See 3.1.8)

No

Frame grabber settings must be adjusted. (See 4.1.3)

Yes

Is the live camera image in Overview mode present?

Yes

Are the frame-grabber settings according the defaults?

Yes

Is the CCD camera sensitivity voltage according the defaults (2.9 V DC)?

No

Adjust the sensitivity voltage. (See 4.2.6)

No

Adjust the brightness of the surrounding field illumination LEDs. (See 4.2.7)

Yes

Is the surrounding field illumination well adjusted? (“Test Eye“ Mode active or “Check,Device“ patient choosen)

Yes

Are all optical components clean and free of dust?

No

Clean the optics. (See 4.2.1)

Yes

Replace the camera. (See 3.1.8)

No

Instrument head exchange or in-house repair necessary (prism 26 must be replaced)

Yes

Are there signs of a “worn out“ CCD camera?

No

Coating of prism 26 OK?

Continue with detailed troubleshoot of KER image

SM-30-4054-en Section 5, Revision A, Feb. 2010

Continue with detailed troubleshoot of ACD image

Confidential and Proprietary

5- 3

Continue with detailed troubleshoot of WZW image

Service Manual Part No. 1836-541

Troubleshooting

IOLMaster 5.5

5.1.2 Detailed troubleshooting I (KER – image) Detailed troubleshoot of KER image

Are ambient lights too bright?

Yes

Avoid direct or indirect light interference. Dim room lights.

No

Replace Keratometer assembly. (See 3.1.6)

No

Adjustment of Keratometer LEDs intensity. (See 4.2.8)

No

Adjustment of the aperture. (See 4.2.12)

No

Instrument head exchange or in-house repair necessary (prism 26 must be replaced)

No

Are all KER LEDs working?

Yes

Are KER LEDs intensity (spot size) adjusted?

Yes

Does the aperture move in and out of position correctly?

Yes

Coating of prism 26 OK?

Yes

Further troubleshoot by the next Carl Zeiss Meditec in-house repair center. Service Manual Part No. 1836-541

Confidential and Proprietary

5-4

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

Troubleshooting

5.1.3 Detailed troubleshooting II (ACD – image)

SM-30-4054-en Section 5, Revision A, Feb. 2010

Confidential and Proprietary

5- 5

Service Manual Part No. 1836-541

Troubleshooting

IOLMaster 5.5

5.1.4 Detailed troubleshooting III (WTW – image)

Service Manual Part No. 1836-541

Confidential and Proprietary

5-6

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

5.2

Troubleshooting

Error messages

5.2.1 Error messages during powering up Error message (plain text) Changer: coupler always on.

Possible Causes

Error Correction

The signal of the changer light barrier always shows "0" (on). - Defective light barrier. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective motherboard MED-COM. - Defective control cable connection. Changer: coupler The signal of the changer light not found. barrier always shows "1" (off). - Defective light barrier or connection interrupted. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective control cable connection. - Defective motherboard MED-COM. Changer: - Faulty alignment of changer. step error in forward - Defective stepper motor. direction - Stiff motion of changer. Changer: - Faulty alignment of changer. step error in - Defective stepper motor. backward direction - Stiff motion of changer. Aperture: coupler The signal of the aperture light always on. barrier always shows "0" (on). - Defective light barrier. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective motherboard MED-COM. - Defective control cable connection. Aperture: step error

- Defective stepper motor. - Stiff motion of aperture.

SM-30-4054-en Section 5, Revision A, Feb. 2010

- Change light barrier, see 3.1.15 - Change Digital IOL PCB or stepper motor - Change Digital IOL PCB, see 3.1.10 - Change motherboard MED-COM, see 3.2.7 - Repair or change control cable, see 3.1.4

- Change light barrier, repair connection, see 3.1.15 - Change Digital IOL PCB or stepper motor - Change Digital IOL PCB, see 3.1.10 - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7. -

Align changer, see 4.2.15 Change stepper motor, see 3.1.14 Repair changer, see 3.1.15 Align changer, see 4.2.15 Change stepper motor, see 3.1.14 Repair changer, see 3.1.15

- Change light barrier, see 3.1.15 - Change Digital IOL PCB or stepper motor. - Change Digital IOL PCB, see 3.1.10 - Change motherboard MED-COM, see 3.2.7. - Repair or change control cable, see 3.1.4 - Change stepper motor, see 3.1.12 - Repair aperture, see 3.1.12

Confidential and Proprietary

5- 7

Service Manual Part No. 1836-541

Troubleshooting

Error message (plain text) Aperture: coupler not found.

IOLMaster 5.5

Possible Causes

Error Correction

The signal of the aperture light barrier always is "1" (off). - Defective light barrier or connection interrupted. - Stepper motor does not run defective motor or Digital IOL PCB. - Defective Digital IOL PCB. - Defective control cable connection. - Defective motherboard MED-COM. Slider: coupler does The signal of the lower light barrier not switch. on the prism slider always is "1" (off). No AL measurement - Defective light barrier or possible, prism slider connection interrupted. does not move - Defective Digital IOL PCB. Interferometer does not run up. - Defective motor or connection.

- Change light barrier or repair connection, see 3.1.13 - Change Digital IOL PCB or stepper motor. - Change Digital IOL PCB, see 3.1.10 - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7.

- Change light barrier or repair connection, see 3.1.17 - Change Digital IOL PCB, see 3.1.10 - Change motor, repair connection, see 3.1.16 - Change Digital IOL PCB, see 3.1.10

- Defective Digital IOL PCB. The signal of the lower light barrier always is "1". - Defective light barrier. - Change lower light barrier (right) , see 3.1.17 - Defective Digital IOL PCB. - Change Digital IOL PCB, see 3.1.10 Slider: upper coupler The signal of the upper light barrier not found. always is "1". - Defective light barrier. - Change upper light barrier (left) , see 3.1.17 - Defective Digital IOL PCB. - Change Digital IOL PCB, see 3.1.10 LIE4: Motor of prism slider does not run. no zero pulse - Defective motor or connection. - Change interferometer motor, see 3.1.16 - Change Digital IOL PCB, see 3.1.10 backward - Defective Digital IOL PCB. Slider: bottom coupler not found.

Motor of prism slider runs. - Defective Digital IOL PCB. - Defective encoder or connection. - Defective control cable connection. - Defective motherboard MEDCOM.

Service Manual Part No. 1836-541

- Change Digital IOL PCB, see 3.1.10 - Change measuring head or repair connection. - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7.

Confidential and Proprietary

5-8

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

Troubleshooting

Error message Possible Causes (plain text) LIE4: - Incorrect motor speed. measurement length - Defective interferometer. backward too short - Stiff motion of prism slider/motor - Defective Digital IOL PCB. - Upper or lower light barrier out of alignment. LIE4: no zero pulse forward

Laser: adjustment power too high.

Laser: 100% power too low.

- Adjust speed, see 4.2.9 - Change measuring head, see 3.1.3 - Change measuring head or change interferometer motor - Change Digital IOL PCB, see 3.1.10 - Move light barrier concerned as far upward or downward as possible, increasing the path for the slider as far as possible.

Motor of prism slider does not run. - Defective motor or connection. - Change motor, see 3.1.16 - Defective Digital IOL PCB. - Change Digital IOL PCB, see 3.1.10

Motor of prism slider runs. - Defective Digital IOL PCB. - Defective encoder or connection. - Defective control cable connection. - Defective motherboard MED-COM. LIE4: - Defective interferometer. measurement length - Stiff motion of prism forward too short slider/motor - Incorrect motor speed. - Defective Digital IOL PCB. - Upper or lower light barrier out of alignment. Laser: adjustment power too low.

Error Correction

- Faulty adjustment of laser power - Defective reflecting diode. - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode. - Faulty adjustment of laser power - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode. - Faulty adjustment of laser power - Defective reflecting diode. - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode.

SM-30-4054-en Section 5, Revision A, Feb. 2010

- Change Digital IOL PCB, see 3.1.10 - Change measuring head or repair connection. - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7. - Change measuring head, see 3.1.3 - Change measuring head or change interferometer motor. - Adjust speed, see 4.2.9 - Change Digital IOL PCB, see 3.1.10 - Move light barrier concerned as far upward or downward as possible, increasing the path for the slider as far as possible. - Adjust laser power, see 4.2.3 - Change measuring head, see 3.1.3 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3 - Adjust laser power, see 4.2.3 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3 - Adjust laser power, see 4.2.4 - Change measuring head, see 3.1.3 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3

Confidential and Proprietary

5- 9

Service Manual Part No. 1836-541

Troubleshooting

Error message (plain text) Laser: 100% power too high.

No laser switch off. + 12 V too low

+ 12 V too high

+ 5 V too low. + 5 V too high - 5 V too low. - 5 V too high. - 12 V too low. - 12 V too high. +12V: too low +5V: too low -5V: too high -12V: too high

IOLMaster 5.5

Possible Causes

Error Correction

- Faulty adjustment of laser power - Defective Digital IOL PCB. - Incorrect adjustment of laser monitoring - Defective laser diode. - Defective Digital IOL PCB. - Defective APD PCB. - Defective power supply unit. - Defective APD PCB. - Defective motherboard MED-COM. - Defective control cable connection. - Defective Digital IOL PCB. - Defective power supply unit. - Defective APD PCB. - Defective measuring head/PC connection - Defective APD PCB. - Defective motherboard MED-COM. - Defective measuring head/PC connection - Defective power supply unit. - Defective APD PCB.

- Defective motherboard MED-COM. - Defective measuring head/PC connection Power limit reached. Laser power too high Please call service. - Incorrect adjustment of laser power - Defective Digital IOL PCB. - Defective APD PCB.

Changer: step error in forward direction +12 V: too low + 5 V: too low - 5 V too high. - 12 V too high.

Service Manual Part No. 1836-541

- Defective laser diode. - Incorrect adjustment of laser monitoring - Defective contol cable - Defective control cable connector - Defective motherboard MED-COM

- Adjust laser power, see 4.2.4 - Change Digital IOL PCB, see 3.1.10 - Adjust laser monitoring, see 4.2.5 - Change measuring head, see 3.1.3 - Change Digital IOL PCB, see 3.1.10 - Change head or send instrument to next in-house repair to change the APD PCB. - Change power supply unit, see 3.2.4 - Change head or send instrument to next in-house repair to change the APD PCB. - Change motherboard MED-COM, see 3.2.7. - Repair or change control cable, see 3.1.4 - Change Digital IOL PCB, see 3.1.10 - Change power supply unit, see 3.2.4 - Change head or send instrument to next in-house repair to change the APD PCB. - Repair connection. - Change head or send instrument to next in-house repair to change the APD PCB. - Change motherboard MED-COM, see 3.2.7. - Repair connection. - Change power supply unit, see 3.2.4 - Change head or send instrument to next in-house repair to change the APD PCB. - Change motherboard MED-COM, see 3.2.7. - Check all connections - Perform adjustment. - Change Digital IOL PCB, see 3.1.10 - Change head or send instrument to next in-house repair to change the APD PCB. - Change measuring head - Adjusting laser monitoring - Repair or change control cable, see 3.1.4 - Change motherboard MED-COM, see 3.2.7 - Change motherboard MED-COM, see 3.2.7

Confidential and Proprietary

5 - 10

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

Troubleshooting

5.2.2 Error messages during operation Error message (plain text) ERR: no image possible

ERR: 0x%x laser adjustment power too high.

Possible Causes

Error Correction

Image recording could not be started - Defective motherboard MEDCOM (frame grabber). - Problems in IOL Master Software. Laser power too high - Incorrectly set laser power of 40 μW - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.

-

Change motherboard MED-COM, see 3.2.7. Perform IOL Master Recovery/Re-imaging.

-

Set laser power

-

Set laser monitoring

-

-

Defective laser diode. Laser power too low. - Incorrectly set laser power of 40 μW. - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.

-

Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3

-

Set laser power

-

Set laser monitoring

-

Defective laser diode. Laser power too high during measurement. - Incorrectly set laser power of 400 μW. - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.

-

Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3

-

Set laser power

-

Set laser monitoring

-

Defective laser diode. ERR: Laser power too low during 0x%x laser measurement. measurement power - Incorrectly set laser power of 400 μW. too low. - Incorrect adjustment of laser monitoring - Defective Digital IOL PCB. - Defective APD PCB.

-

Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3

-

Set laser power

-

Set laser monitoring

-

Change Digital IOL PCB, see 3.1.10 Change head or send instrument to next inhouse repair to exchange the APD PCB. Change measuring head, see 3.1.3

-

ERR: 0x%x laser adjustment power too low.

-

ERR: 0x%x laser adjustment power too high, measurement interrupted.

-

-

Defective laser diode.

SM-30-4054-en Section 5, Revision A, Feb. 2010

-

-

-

-

Confidential and Proprietary

5- 11

Service Manual Part No. 1836-541

Troubleshooting

IOLMaster 5.5

Error message Possible Causes (plain text) ERR: Laser power much too high Power limit reached. (end of AD conversion range Please call service. reached) - Defective Digital IOL PCB. - Defective APD PCB.

Error Correction

- Change Digital IOL PCB, see 3.1.10 - Change head or send instrument to next inhouse repair to exchange the APD PCB. - Change measuring head, see 3.1.3

- Defective laser diode. Zero pulse of LIE4 path0x00000xxx , no measuring encoder is missing. zero pulse - Defective path-measuring - Change measuring head, see 3.1.3 encoder. - Change Digital IOL PCB, see 3.1.10 (xxx stands for the - Defective Digital IOL PCB. - Repair or change control cable, see 3.1.4 relevant error code) - Connection in control cable interrupted. - Defective motherboard MED- - Change motherboard MED-COM, see 3.2.7 COM. 0x00000180 // images could not be grabbed 0x00000181 // No zero pulse during AL measurement detected 0x00000182 // Data-file could not be read 0x00000183 // Incorrect distance between zero-pulses

Service Manual Part No. 1836-541

Confidential and Proprietary

5 - 12

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

Troubleshooting

5.2.3 Typical error messages and causes therefore Error message (plain text) changer: step error in forward direction +12V: too low +5V: too low -5V: too high -12V: too high

Possible Causes - Incorrect Control-cable (connection!) - Defect motherboard MED-COM

additional "krrrrrrr" sound from the changer - changer-sensor defect / not well plugged without "krrrrrr" sound changer: optocoupler always on from the changer changer: step error in forward - changer-motor defect / direction not well plugged - sensor aperture defect / not well plugged aperture: optocoupler not found - motor aperture defect / not well plugged - lower sensor Interferometer defect / not well plugged sledge: optocoupler don't switch - motor Interferometer defect / not well plugged System error! No measurement AL measurement possible – possible. but no result laser: adjustment power too low - APD defect / not well laser: 100% power too low plugged +12V: too low +5V: too low -5V: too high -12V: too high AL measurement shows LIE4: no zero pulse backward laser: adjustment power too low "ERR: 0x181, no zero pulse. : 0" laser: 100% power too low - motherboard MED-COM +12V: too low defect +5V: too low -5V: too low -12V: too low Difference during determination of the increments (Interferometer-scale). ERR: 0x183, measurement system "99" is the offset as a nu: 99 meric value. - damaged scale or scale reader AL-graph looks like a sine wave or - defective laser diode otherwise abnormal Laser-Power (400 μW) unstable or - defective laser diode unable to adjust changer: optocoupler not found changer: step error in forward direction

SM-30-4054-en Section 5, Revision A, Feb. 2010

Confidential and Proprietary

5- 13

Error Correction - Check the control cable connection - Change motherboard MED-COM

- Check connection or change sensor, see 3.1.15 - Check connection or change the changer motor, see 3.1.14 - Check connection or change sensor, see 3.1.13 - Check connection or change aperture module, see 3.1.12 - Change sensor or repair connection, see 3.1.17 - Change motor, repair connection, see 3.1.16 - Check connection - Change head or send instrument to next in-house repair to exchange the APD PCB.

- Change the motherboard MEDCOM

- Change head or send instrument to next in-house repair. - Change head or send instrument to next in-house repair. - Change head or send instrument to next in-house repair. Service Manual Part No. 1836-541

Troubleshooting

5.3

IOLMaster 5.5

Tool list IOLMaster Tool Designation

Order Number

Fieldmaster power meter with power adapter

Service Manual Part No. 1836-541

Use of Tool Indicating device of laser power

115V 230V

0115-811 0115-810

Detector head LM 2

0113-503 Sensor of metering laser power

Measuring Adapter for LM2

1094-546 To use with the LM2 detector head for exact positioning.

Ophthalmometer Headrest Clamp

302086-0000-000 Holder for detector head on head rest

Ophir NOVA power meter

320866-0209-022 Indicating device of laser power

Confidential and Proprietary

5 - 14

SM-30-4054-en Section 5, Revision A, Feb. 2010

IOLMaster 5.5

Troubleshooting

Tool Designation

Order Number

Ophir power head PD 300 – 3W

SM-30-4054-en Section 5, Revision A, Feb. 2010

Use of Tool

On request Sensor of metering of laser power

Hygrometer TESTO 608-H1

0492-815 Speed adjustment

Service USB flash drive and IOLMaster Software Version 5.5 USB flash drive

1778-631 Re-imaging the HDD of the IOLMaster 1790-501

Alignment device for IOLMaster

1676-641 CCD camera alignment

Laser short circuit plug

1078-821 For protecting the laser diode (ESD) during service activities.

3/32" hex key

0118-751 For unscrewing the power supply

Torx tool kit T5-T50

1209-366 For service on the LC display

Confidential and Proprietary

5- 15

Service Manual Part No. 1836-541

Section 6 – Diagrams 6.1 Diagrams........................................................................................................................3 6.1.1 Overview of functional modules ...................................................................................... 3 6.1.2 Measuring voltages and signals ....................................................................................... 4 Voltages on the power supply unit .................................................................................. 4 Voltages and signals on the Power ON/OFF PCB .............................................................. 4 Voltages and signals on APD 3 PCB ................................................................................. 5 Voltages and signals on Digital IOL PCB........................................................................... 6

SM-30-4054-en Section 6, Revision A, Feb. 2010

Confidential and Proprietary

6-1

Service Manual Part No. 1836-541

Diagrams

IOLMaster 5.5

Notes:

Service Manual Part No. 1836-541

Confidential and Proprietary

6-2

SM-30-4054-en Section 6, Revision A, Feb. 2010

IOLMaster 5.5

6.1

Diagrams

Diagrams

6.1.1 Overview of functional modules Measuring head

Interferometer

Control Digital IOL

X202 X202 X160

EEPROM

Instrument base

X208 X208

speed down

speed up

Fire button

Light detector end

X203 X203

X209

X1

Light detector beginning M Motor IOL

X204 X204

R77

R64

R40

6 IR LED cycled(KER)

Aperturte diaphragm X205 X205

X401 X214

6 IR LED overview illum.

Measuring system IOLMaster

Light detector M Motor aperture

X212 X212

R3

5 LED ACD left X402

X402

X217 X217

5 LED ACD right X402

X402

X218 X218

Changer X207 X207

Light detector M Motor changer

X211 X211

R7 40μW R75

CCD camera

X150

X150

X219

Gain CCD

X219

X216 X170

R9 400μW

X222 X502

+9V +5V

LED converter unit for backlight

LC-Display 640x480

+12V

+12V

CN1 CN1

X231 X201

X213

X581 X580

X231 X201

X213

IR receiver X501

X304 Power

X304

Isolating transformer 115/230 VAC 115VA

Monitor Laser Diode

Reflex Fixier Diode

+12 V

+12V X301 +5V1/2/3 -5V1/2/3 -12V R26

+5V -5V -12V

MUX

X302 X302

Laser Monitoring

APD Diode

APD-Module IOLMaster

Power

Power INPUT 100...240 VAC Assym. table

S1

Power

1a 1 2 2a

Keyboard

X6 X5X2 X3

DISPLAY optional

Touch pad

USB flash drive optional

X855 X851 Power on/off X853

PS2 PS2 X611 Mouse Keyboard

PSU 100-240V AC

X601

VGA

USB (4x)

X601

X609

X608

+12V +5V A

Power IN +12V

X600

LAN optional

Network isolator (optional)

Ethernet X607

Computer MED-COM D

X602

X604

X603

Fan

Inductive position switch

X7

Printer optional

SATA X606

COM X610

HDD

Computer base IOLMaster SM-30-4054-en Section 6, Revision A, Feb. 2010

Confidential and Proprietary

6- 3

Service Manual Part No. 1836-541

Diagrams

IOLMaster 5.5

6.1.2 Measuring voltages and signals Voltages on the power supply unit +12V +12V GND GND GND

DC output voltage of the PSU

Power Supply 100 -240 V (AC)

Input voltage of the PSU 100 –240V AC Line input voltage of the Power ON/OFF PCB

Line Input 100 -240 V (AC)

Voltages and signals on the Power ON/OFF PCB S1

Power Supply 100-240V AC S1

Line input 100–240V AC

F1/F2 T 3.15A X853:1 ◄ +5V of MED-COM X853:2 ► ON/OFF X853:3 ◄ Delete X853:4 ◄ GND of MED-COM

risk of electric shock

X602 Motherboard MED-COM

Service Manual Part No. 1836-541

Confidential and Proprietary

6-4

SM-30-4054-en Section 6, Revision A, Feb. 2010

IOLMaster 5.5

Diagrams

Voltages and signals on APD 3 PCB

R26 Laser Monitoring

◄ X304:1 selection MUX ◄ X304:2 selection MUX ◄ X304:3 selection MUX ◄ X304:4 0V ◄ X304:5 +12V ◄ X304:6 0V X213 Digital IOL

► X302:1 ► X302:2 ► X302:3 ► X302:4

0V Analog signal MUX -5V2 +5V2

X601 MED-COM ◄ X301:1 anode ◄ X301:2 cathode X501 IR Laser Monitoring Diode

SM-30-4054-en Section 6, Revision A, Feb. 2010

Confidential and Proprietary

6- 5

Service Manual Part No. 1836-541

Diagrams

IOLMaster 5.5

Voltages and signals on Digital IOL PCB

40μW laser power R7 Service connector

400μW laser power R9

X220: 1 +12V X220: 2 + 9V X220: 3 + 5V X220: 4 GND

Brightness Keratometer R40 Surrounding field illumination R3 CCD Camera sensitivity R75 Interf. motor speed (down) R77 Interf. motor speed (up) R64

CCD Camera Module ► X219:4 +9V ► X219:5 GND X201:12 +12V ◄MED-COM X601 X201: 9 GND ◄MED-COM X601 X231: 9 +5V

◄MED-COM X601

APD Module X304 ► X213:5 +12V ► X213:6 GND

X209:1 NC X209:2 GND X209:3 NO FIRE

Service Manual Part No. 1836-541

Confidential and Proprietary

6-6

SM-30-4054-en Section 6, Revision A, Feb. 2010

Section 7 - Parts IOLMaster V5.5 IOLMaster V5.5

4

Zubehör IOLMaster Accessories IOLMaster

SM-30-4054 Section 7, Revision A, Feb. 2010

50

Confidential and Proprietary

7-1

Service Manual Order No. 1836-541

SM-30-4054 IOLMaster V5.5 000000-1836-541 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1

IOLMaster V5.5

IOLMaster V5.5

2

Zubehör IOLMaster

Accessories IOLMaster

~ 000000-1790-501

SW-Paket IOLMaster V5.5

SW-Paket IOLMaster V5.5

~ 000000-1778-631

SW-Paket IOLMaster Service Vs 7.1.2

SW-Packet IOLMaster Service 7.1.2

Service Manual Order No. 1836-541

Confidential and Proprietary

7-2

4 50

SM-30-4054 Section 7, Revision A, Feb. 2010

SM-30-4054 IOLMaster V5.5 000000-1836-541

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-3

Service Manual Order No. 1836-541

IOLMaster V5.5 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1 000000-1836-516

Set Ersatzhauben IOLMaster V5.5

Set of spare covers for IOLMaster V5.5

2

ISO 4762-M 3 x 10-A2-70

Screw M3x10 (screws for scanner)

3

ISO 4762-M 3X 20-A2-70

ISO 4762-M 3X 20-A2-70

6

Untersatz IOLMaster

Overview IOLMaster base

12

7

Gerätekopf IOLMaster m. Verpackung

IOLMaster head incl. packaging

32

Service Manual Order No. 1836-541

Confidential and Proprietary

7-4

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-5

Service Manual Order No. 1836-541

IOLMaster V5.5 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

50 000000-1692-640

Kabel MS (X160) - DIGIT (X202)

Cable MS (X160) - DIGIT (X202)

51 000000-1078-821

Kurzschlussstecker

Short Circuit Connector IOL Master

52 000000-1534-730

FBG DIGITAL IOL (DIGIT)

Digital IOL (DIGIT)

53 000000-1693-879

Koppler Interferometer X203

Interferometer Coupler X203

54 000000-1694-348

Koppler Interferometer X204

Optical sensor interferometer X204

55 000000-1761-131

Kabel IOLMaster (X601) - Gerätekopf

Cable IOLMaster (X601) - head

44

57 000000-1692-644

Kabel DIGIT (X213) - APD (X304)

Cable DIGIT (X213) - APD (X304)

46

58 000000-1693-867

Koppler Wechsler X207

Optical sensor changer X207

59 000000-1692-642

Kabel CCD (X150) - DIGIT (X219)

Cable CCD (X150) - DIGIT (X219)

Service Manual Order No. 1836-541

Confidential and Proprietary

7-6

42

40

48

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-7

Service Manual Order No. 1836-541

IOLMaster V5.5 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

60 000000-1693-816

Motor Wechsler

Changer motor

61 000000-1693-881

Koppler Blende X205

Optical sensor diaphragm X205

62 000000-0510-510

TFT-Color-Display 6,5-inch

TFT-Color-Display 6,5-inch

63 000000-0510-511

LED-Converter Unit für Backlight

LED-Converter Unit for Backlight

Service Manual Order No. 1836-541

Confidential and Proprietary

7-8

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-9

Service Manual Order No. 1836-541

IOLMaster V5.5 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

6

Untersatz IOLMaster

Overview IOLMaster base

12

7

Gerätekopf IOLMaster m. Verpackung

IOLMaster head incl. packaging

32

50 000000-1692-640

Kabel MS (X160) - DIGIT (X202)

Cable MS (X160) - DIGIT (X202)

42

52 000000-1534-730

FBG DIGITAL IOL (DIGIT)

Digital IOL (DIGIT)

40

53 000000-1693-879

Koppler Interferometer X203

Interferometer Coupler X203

54 000000-1694-348

Koppler Interferometer X204

Optical sensor interferometer X204

55 000000-1761-131

Kabel IOLMaster (X601) - Gerätekopf

Cable IOLMaster (X601) - head

44

57 000000-1692-644

Kabel DIGIT (X213) - APD (X304)

Cable DIGIT (X213) - APD (X304)

46

58 000000-1693-867

Koppler Wechsler X207

Optical sensor changer X207

59 000000-1692-642

Kabel CCD (X150) - DIGIT (X219)

Cable CCD (X150) - DIGIT (X219)

60 000000-1693-816

Motor Wechsler

Changer motor

61 000000-1693-881

Koppler Blende X205

Optical sensor diaphragm X205

62 000000-0510-510

TFT-Color-Display 6,5-inch

TFT-Color-Display 6,5-inch

63 000000-0510-511

LED-Converter Unit für Backlight

LED-Converter Unit for Backlight

Service Manual Order No. 1836-541

Confidential and Proprietary

7-10

48

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-11

Service Manual Order No. 1836-541

Overview IOLMaster base Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1 000000-1025-670

Kopfstütze IOL

Head Rest

14

2 266101-4501-927

Kreuztisch IOL

Cross Stage for IOL Master

16

3

DIN 6912-M5X 16-A2-70

DIN 6912-M5X 16-A2-70

4

Computer IOLMaster mit Haube

Computer IOLMaster with cover

5

AT Computer IOLMaster mit Haube

EC IOLMaster computer with cover

6

ISO 7380-M 5x 8 A2-70

ISO 7380-M 5x 8 A2-70

7 266101-4510-310

Gleitplatte

Sliding plate for Cross stage

8 000000-1210-002

Zahnradpaar mit Rollen

Pair of gear wheels with rollers

9

ISO 7380-M 4x12 A2-70

Screw M4x12 (screws for cover)

Zahnstange RAL U713

Gear rack RAL U713

10 266101-4500-562

Service Manual Order No. 1836-541

Confidential and Proprietary

7-12

20

SM-30-4054 Section 7, Revision A, Feb. 2010

Overview IOLMaster base

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-13

Service Manual Order No. 1836-541

Head Rest 000000-1025-670 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1 319752-0024-000

Stirnband

Forehead strap

2

ISO 7380-M 4x 6-A2-70

ISO 7380-M 4x 6-A2-70

3 319752-8016-000

Kinnschale

Chin rest

4 266181-7000-551

Papierhalter

Paper holder

5

ISO 4762-M 5 x 20-A2-70

ISO 4762-M 5 x 20-A2-70

6

ISO 4762-M 4 x 8-A2-70

Screw M4x8

7 000000-1294-613

Kinntraverse montiert IOL

Chin rest unit, assembled

Service Manual Order No. 1836-541

Confidential and Proprietary

7-14

SM-30-4054 Section 7, Revision A, Feb. 2010

Head Rest 000000-1025-670

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-15

Service Manual Order No. 1836-541

Cross Stage for IOL Master 266101-4501-927 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1

ISO 1580-M3 x 5-A2-70

ISO 1580-M3 x 5-A2-70

2

DIN 84-M 2,5x 4-A2-70

DIN 84-M 2,5x 4-A2-70

3 000000-1775-095

Set Kleinteile KT

Service Set cross table locking knob IOL

4

DIN 920-M3 x8 -A1-70

DIN 920-M3 x8 -A1-70

5

ISO 4762-M 3 x 10-A2-70

Screw M3x10 (screws for scanner)

6

DIN 125-A 3,7-A2

Disk DIN 125-A 3,7-A2

7 266101-4510-010

Spindelmutter

Spindle nut

8 266101-4502-514

SPINDEL. KOMPL.

Spindle, complete

9

ISO 4762-M 4 x 20-A2-70

ISO 4762-M 4 x 20-A2-70

10

4,2X12X 0,8 CZN790 X12CRNI17.7K

4,2X12X 0,8 CZN790 X12CRNI17.7K

11 266101-4533-010

Stirnrad

Spur gear

12 266101-4560-514

Steuergriff

IOL joystick with release button

13

AT Steuergriff IOLMaster

EC IOLMaster Joystick w. release button

14

ISO 4762-M 3 x 8-A2-70

ISO 4762-M 3 x 8-A2-70

15

DIN 125-A 3,2-A2

DIN 125-A 3.2-A2

16

DIN 125-A 5,3-A2

DIN 125-A 5,3-A2

17

DIN 6912-M 5x 12-A2-70

DIN 6912-M 5x 12-A2-70

Service Manual Order No. 1836-541

Confidential and Proprietary

7-16

18

SM-30-4054 Section 7, Revision A, Feb. 2010

Cross Stage for IOL Master 266101-4501-927

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-17

Service Manual Order No. 1836-541

IOL joystick with release button 266101-4560-514 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1

AT Steuergriff IOLMaster

EC IOLMaster Joystick w. release button

2 266101-4515-110

Kappe

Cap

Service Manual Order No. 1836-541

Confidential and Proprietary

7-18

SM-30-4054 Section 7, Revision A, Feb. 2010

IOL joystick with release button 266101-4560-514

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-19

Service Manual Order No. 1836-541

Computer IOLMaster with cover Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1

AT Computer IOLMaster mit Haube

EC IOLMaster computer with cover

2 000000-1693-874

induktiver Näherungsschalter

Induct. Sensor Switch

3

ISO 4762-M 4X 10-A2-70

ISO 4762-M 4X 10-A2-70

4

ISO 4762-M 3 x 5-A2-70

ISO 4762-M 3 x 5-A2-70

5

DIN 125-A 3,2-EP-HP S2000E/FR3NATUR

DIN 125-A 3,2-EP-HP S2000E/FR3NATUR

6 000000-0509-551

Festplatte 2,5" SATA 80GB

Hard disk 2,5" SATA 80GB

7 000000-1528-161

FBG MED-COM IOLMaster (MED-COM)

IOLMaster MED-COM PCB

8

AT FBG MED-COM IOLMaster (MED-COM)

EC PCB MED-COM IOLMaster (MED-COM)

9

ISO 4762-M 3 x 8-A2-70

ISO 4762-M 3 x 8-A2-70

FBG POWER ON/OFF 2 (POWER)

ASSEMBLY POWER ON/OFF 2

24

10A 000000-1694-354

Kabel Power - On/Off Netz

Power On/Off - mains cable

26

11 000000-0508-752

Netzteil 90W 12Vdc/7,5A

Power supply 90W 12VDC/7,5A

12

ISO 7380-M 3x 5-A2-70

ISO 7380-M 3x 5-A2-70

13

DIN 125-A 3,2-A2

DIN 125-A 3.2-A2

14 000000-1693-872

Lüfter kpl.

Fan complete

15 000000-1531-678

Schalter RSM 2pol. grün Sonderdruck

Master switch, 2 pins, green

16 000000-0505-635

Netzmodul/MED 3A Fi

Power entry modul medical version 3A

10 000000-1671-563

Service Manual Order No. 1836-541

steck6308

Confidential and Proprietary

7-20

SM-30-4054 Section 7, Revision A, Feb. 2010

Computer IOLMaster with cover

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-21

Service Manual Order No. 1836-541

Computer IOLMaster with cover Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

2 000000-1693-874

induktiver Näherungsschalter

Induct. Sensor Switch

6 000000-0509-551

Festplatte 2,5" SATA 80GB

Hard disk 2,5" SATA 80GB

7 000000-1528-161

FBG MED-COM IOLMaster (MED-COM)

IOLMaster MED-COM PCB

8

AT FBG MED-COM IOLMaster (MED-COM)

EC PCB MED-COM IOLMaster (MED-COM)

FBG POWER ON/OFF 2 (POWER)

ASSEMBLY POWER ON/OFF 2

24

10A 000000-1694-354

Kabel Power - On/Off Netz

Power On/Off - mains cable

26

11 000000-0508-752

Netzteil 90W 12Vdc/7,5A

Power supply 90W 12VDC/7,5A

14 000000-1693-872

Lüfter kpl.

Fan complete

15 000000-1531-678

Schalter RSM 2pol. grün Sonderdruck

Master switch, 2 pins, green

16 000000-0505-635

Netzmodul/MED 3A Fi

Power entry modul medical version 3A

17 000000-1694-129

Kabel POWER(X853)-MED-COM(X602)

Cable POWER(X853)-MED-COM(X602)

28

18 000000-1694-203

Kabel MED-COM (X600) - NT (X7)

Cable MED-COM (X600) - NT (X7)

30

19 000000-0509-839

Knopfzelle Lithium 3V/230mAh 3,2x20mmØ

Lithium coin cell 3V/230mAh 3,2x20mmØ

10 000000-1671-563

Service Manual Order No. 1836-541

steck6308

Confidential and Proprietary

7-22

SM-30-4054 Section 7, Revision A, Feb. 2010

Computer IOLMaster with cover

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-23

Service Manual Order No. 1836-541

ASSEMBLY POWER ON/OFF 2 000000-1671-563

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-24

SM-30-4054 Section 7, Revision A, Feb. 2010

ASSEMBLY POWER ON/OFF 2 000000-1671-563

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-25

Service Manual Order No. 1836-541

Power On/Off - mains cable 000000-1694-354

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-26

SM-30-4054 Section 7, Revision A, Feb. 2010

Power On/Off - mains cable 000000-1694-354

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-27

Service Manual Order No. 1836-541

Cable POWER(X853)-MED-COM(X602) 000000-1694-129

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-28

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Cable POWER(X853)-MED-COM(X602) 000000-1694-129

Confidential and Proprietary

7-29

Service Manual Order No. 1836-541

Cable MED-COM (X600) - NT (X7) 000000-1694-203

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-30

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Cable MED-COM (X600) - NT (X7) 000000-1694-203

Confidential and Proprietary

7-31

Service Manual Order No. 1836-541

IOLMaster head incl. packaging Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1

ISO 4762-M 5 X16-A2-70

ISO 4762-M 5 X16-A2-70

2

DIN 125-A 5,3-A2

DIN 125-A 5,3-A2

3 000000-1770-344

LCD, ohne Kabel

LCD, without cable

4 000000-1757-148

Spaltgruppe VKT

ACD slit module

5

DIN 125-A 2,2-A2

DIN 125-A 2,2-A2

6

ISO 4762-M 2 X 8-A2-70

ISO 4762-M 2 X 8-A2-70

7 000000-1757-077

Keratometer,kpl.

Keratometer, compl.

8

ISO 4762-M 3 x 6-A2-70

ISO 4762-M 3 x 6-A2-70

9 000000-1756-302

Blendengruppe Keratometer

Keratometer diaphragm module

10

DIN 433- 3,2-A1

Washer DIN 433-3,2-A1

11

ISO 4762-M 3 x 8-A2-70

ISO 4762-M 3 x 8-A2-70

12 000000-1693-816

Motor Wechsler

Changer motor

13

ISO 4762-M 2,5x 4-A2-70

ISO 4762-M 2,5x 4-A2-70

14

DIN 125-A 2,7-A2

DIN 125-A 2,7-A2 disc

15

APD-LP montiert

APD PCB, mounted

16

DIN 125-A 3,2-A2

DIN 125-A 3.2-A2

17

ISO 4762-M 3 x 10-A2-70

Screw M3x10 (screws for scanner)

18 000000-1770-340

Set, CCD Kamera 9V

CCD camera set, 9V

19

ISO 4762-M 2,5x 6-A2-70

ISO 4762-M 2,5x 6-A2-70

20

DIN 9021- 2,7-140 HV-A2

DIN 9021- 2,7-140 HV-A2

21

Prismengruppe 26-er N-SF2

Prism module 26, N-SF2

22

DIN 137-A 3 -A2

DIN 137-A 3 -A2

Service Manual Order No. 1836-541

Confidential and Proprietary

7-32

36

38

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster head incl. packaging

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-33

Service Manual Order No. 1836-541

IOLMaster head incl. packaging Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

ISO 4762-M 3 x 6-A2-70

ISO 4762-M 3 x 6-A2-70

10

DIN 433- 3,2-A1

Washer DIN 433-3,2-A1

11

ISO 4762-M 3 x 8-A2-70

ISO 4762-M 3 x 8-A2-70

23 000000-1693-867

Koppler Wechsler X207

Optical sensor changer X207

24

ISO 4762-M 3 x 5-A2-70

ISO 4762-M 3 x 5-A2-70

25

Distanzhülse Pa/sw 3,2x5,0x3,0

distance bush Pa/bk 3,2x5,0x3,0

26

DIN 125-A 3,2-EP-HP S2000E/FR3NATUR

DIN 125-A 3,2-EP-HP S2000E/FR3NATUR

27 000000-1078-821

Kurzschlussstecker

Short Circuit Connector IOL Master

28 000000-1534-730

FBG DIGITAL IOL (DIGIT)

Digital IOL (DIGIT)

29 000000-1755-837

Motor Interferometer, kpl.

Motor Interferometer, complete

30 000000-1693-879

Koppler Interferometer X203

Interferometer Coupler X203

31 000000-1694-348

Koppler Interferometer X204

Optical sensor interferometer X204

32

Interferometer mit Verpackung

Interferometer incl. packaging

33

ALM-Wechsler

ALM converter

8

Service Manual Order No. 1836-541

Confidential and Proprietary

7-34

40

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster head incl. packaging

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-35

Service Manual Order No. 1836-541

ACD slit module 000000-1757-148

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-36

SM-30-4054 Section 7, Revision A, Feb. 2010

ACD slit module 000000-1757-148

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-37

Service Manual Order No. 1836-541

Keratometer diaphragm module 000000-1756-302 Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1 000000-1693-881

Koppler Blende X205

Optical sensor diaphragm X205

2

ISO 4762-M 3 x 5-A2-70

ISO 4762-M 3 x 5-A2-70

3

DIN 433- 3,2-A1

Washer DIN 433-3,2-A1

Service Manual Order No. 1836-541

Confidential and Proprietary

7-38

SM-30-4054 Section 7, Revision A, Feb. 2010

Keratometer diaphragm module 000000-1756-302

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-39

Service Manual Order No. 1836-541

Digital IOL (DIGIT) 000000-1534-730

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-40

SM-30-4054 Section 7, Revision A, Feb. 2010

Digital IOL (DIGIT) 000000-1534-730

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-41

Service Manual Order No. 1836-541

Cable MS (X160) - DIGIT (X202) 000000-1692-640

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-42

SM-30-4054 Section 7, Revision A, Feb. 2010

Cable MS (X160) - DIGIT (X202) 000000-1692-640

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-43

Service Manual Order No. 1836-541

Cable IOLMaster (X601) - head 000000-1761-131

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-44

SM-30-4054 Section 7, Revision A, Feb. 2010

Cable IOLMaster (X601) - head 000000-1761-131

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-45

Service Manual Order No. 1836-541

Cable DIGIT (X213) - APD (X304) 000000-1692-644

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-46

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Cable DIGIT (X213) - APD (X304) 000000-1692-644

Confidential and Proprietary

7-47

Service Manual Order No. 1836-541

Cable CCD (X150) - DIGIT (X219) 000000-1692-642

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-48

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Cable CCD (X150) - DIGIT (X219) 000000-1692-642

Confidential and Proprietary

7-49

Service Manual Order No. 1836-541

Accessories IOLMaster Pos. Item

Bestell-Nr. Order no.

IOLMaster V5.5 Seite Page

Benennung

Designation

1 000000-0433-973

Polyester-Tupfer TX 761 -Basan-

Alpha Swab with Long Handle TX 761

2 000000-0500-895

neo-Lab Wattestäbchen, 150mm, 1Pck=100St Cotton swab 150mm (100 pcs)

3 000000-1046-387

Halterung (Prüfkugel)

Holder Test Eye Compl for IOL Master

52

4 000000-1313-258

Tastatur ACK 540, deutsch

Keyboard ACK 540, German

54

5 000000-1313-260

Tastatur ACK-540, Englisch

Keyboard ACK-540, English

56

6 000000-1526-414

Optikputzmischung L , Flasche 50ml

Optics cleaning mixture L, 50ml bottle

7 000000-1676-641

Prüfvorrichtung für Kamera

Alignment device for IOL master

Service Manual Order No. 1836-541

Confidential and Proprietary

7-50

SM-30-4054 Section 7, Revision A, Feb. 2010

Accessories IOLMaster

IOLMaster V5.5

KEINE ZEICHNUNG VORHANDEN NO DRAWING AVAILABLE

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-51

Service Manual Order No. 1836-541

Holder Test Eye Compl for IOL Master 000000-1046-387

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-52

SM-30-4054 Section 7, Revision A, Feb. 2010

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Holder Test Eye Compl for IOL Master 000000-1046-387

Confidential and Proprietary

7-53

Service Manual Order No. 1836-541

Keyboard ACK 540, German 000000-1313-258

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-54

SM-30-4054 Section 7, Revision A, Feb. 2010

Keyboard ACK 540, German 000000-1313-258

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-55

Service Manual Order No. 1836-541

Keyboard ACK-540, English 000000-1313-260

IOLMaster V5.5

KEINE STÜCKLISTE VORHANDEN NO PARTS LIST AVAILABLE

Service Manual Order No. 1836-541

Confidential and Proprietary

7-56

SM-30-4054 Section 7, Revision A, Feb. 2010

Keyboard ACK-540, English 000000-1313-260

IOLMaster V5.5

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-57

Service Manual Order No. 1836-541

pé~êÉ=m~êíë=iáëí Order No.

Designation

000000-0433-973

Alpha Swab with Long Handle TX 761

000000-0500-895

Cotton swab 150mm (100 pcs)

000000-0505-635

Power entry modul medical version 3A

000000-0505-635

Power entry modul medical version 3A

000000-0508-752

Power supply 90W 12VDC/7,5A

000000-0508-752

Power supply 90W 12VDC/7,5A

000000-0509-551

Hard disk 2,5" SATA 80GB

000000-0509-551

Hard disk 2,5" SATA 80GB

000000-0509-839

Lithium coin cell 3V/230mAh 3,2x20mmØ

000000-0510-510

TFT-Color-Display 6,5-inch

000000-0510-510

TFT-Color-Display 6,5-inch

000000-0510-511

LED-Converter Unit for Backlight

000000-0510-511

LED-Converter Unit for Backlight

000000-1025-670

Head Rest

000000-1046-387

Holder Test Eye Compl for IOL Master

000000-1078-821

Short Circuit Connector IOL Master

000000-1078-821

Short Circuit Connector IOL Master

000000-1210-002

Pair of gear wheels with rollers

000000-1294-613

Chin rest unit, assembled

000000-1313-258

Keyboard ACK 540, German

000000-1313-260

Keyboard ACK-540, English

000000-1526-414

Optics cleaning mixture L, 50ml bottle

000000-1528-161

IOLMaster MED-COM PCB

000000-1528-161

IOLMaster MED-COM PCB

000000-1531-678

Master switch, 2 pins, green

000000-1531-678

Master switch, 2 pins, green

000000-1534-730

Digital IOL (DIGIT)

000000-1534-730

Digital IOL (DIGIT)

000000-1534-730

Digital IOL (DIGIT)

000000-1671-563

ASSEMBLY POWER ON/OFF 2

000000-1671-563

ASSEMBLY POWER ON/OFF 2

000000-1676-641

Alignment device for IOL master

000000-1692-640

Cable MS (X160) - DIGIT (X202)

000000-1692-640

Cable MS (X160) - DIGIT (X202)

000000-1692-642

Cable CCD (X150) - DIGIT (X219)

000000-1692-642

Cable CCD (X150) - DIGIT (X219)

000000-1692-644

Cable DIGIT (X213) - APD (X304)

000000-1692-644

Cable DIGIT (X213) - APD (X304)

000000-1693-816

Changer motor

000000-1693-816

Changer motor

000000-1693-816

Changer motor

000000-1693-867

Optical sensor changer X207

000000-1693-867

Optical sensor changer X207

Service Manual Order No. 1836-541

Confidential and Proprietary

7-58

SM-30-4054 Section 7, Revision A, Feb. 2010

pé~êÉ=m~êíë=iáëí Order No.

Designation

000000-1693-867

Optical sensor changer X207

000000-1693-872

Fan complete

000000-1693-872

Fan complete

000000-1693-874

Induct. Sensor Switch

000000-1693-874

Induct. Sensor Switch

000000-1693-879

Interferometer Coupler X203

000000-1693-879

Interferometer Coupler X203

000000-1693-879

Interferometer Coupler X203

000000-1693-881

Optical sensor diaphragm X205

000000-1693-881

Optical sensor diaphragm X205

000000-1693-881

Optical sensor diaphragm X205

000000-1694-129

Cable POWER(X853)-MED-COM(X602)

000000-1694-203

Cable MED-COM (X600) - NT (X7)

000000-1694-348

Optical sensor interferometer X204

000000-1694-348

Optical sensor interferometer X204

000000-1694-348

Optical sensor interferometer X204

000000-1694-354

Power On/Off - mains cable

000000-1694-354

Power On/Off - mains cable

000000-1755-837

Motor Interferometer, complete

000000-1756-302

Keratometer diaphragm module

000000-1757-077

Keratometer, compl.

000000-1757-148

ACD slit module

000000-1761-131

Cable IOLMaster (X601) - head

000000-1761-131

Cable IOLMaster (X601) - head

000000-1770-340

CCD camera set, 9V

000000-1770-344

LCD, without cable

000000-1775-095

Service Set cross table locking knob IOL

000000-1778-631

SW-Packet IOLMaster Service 7.1.2

000000-1790-501

SW-Paket IOLMaster V5.5

000000-1836-516

Set of spare covers for IOLMaster V5.5

266101-4500-562

Gear rack RAL U713

266101-4501-927

Cross Stage for IOL Master

266101-4502-514

Spindle, complete

266101-4510-010

Spindle nut

266101-4510-310

Sliding plate for Cross stage

266101-4515-110

Cap

266101-4533-010

Spur gear

266101-4560-514

IOL joystick with release button

266181-7000-551

Paper holder

319752-0024-000

Forehead strap

319752-8016-000

Chin rest

SM-30-4054 Section 7, Revision A, Feb. 2010

Confidential and Proprietary

7-59

Service Manual Order No. 1836-541

Section 8 – Appendices

IOLMaster Instrument Checklist and Maintenance Report Service Bulletins

SM-30-4054-en Section 8, Revision A, Feb. 2010

Confidential and Proprietary

8-1

Service Manual Part No. 1836-541