Heat Sea l Te s ter

HST-H3

INSTRUCTION MANUAL for Heat Seal Tester HST-H3

Labthink

1

Heat Sea l Te s ter

HST-H3

Notice The information contained in this manual is derived from proprietary data from Labthink. This manual has been prepared expressly for the purpose of assisting operation and maintenance personnel in their use and general maintenance of HST-H3 Heat Seal Tester. Publication of this information does not imply any rights to reproduce or use it for purposes other than installing, operating, or maintaining the instrument described herein. These provisions are intended to state all of the rights and responsibilities between Labthink and the customer. They take place of and supersede all warranties, expressed or implied, and whether of merchantability, fitness or otherwise. The remedies contained in this instruction manual are exclusive.

2

Heat Sea l Te s ter

HST-H3 Contents Preface…………………………………………………………………….5 1. Introduction…………………………………………………………...6  

Specifications…………………………………………………………………..6 Unpacking………………………………………………………………………7

2. Instrument …………………………………………………………….8       

Instrument Description………………………………………………………..8 Operation Panel…………………………………………………………………..9 Installation………………………………………………………………………10 Instrument System…………………………………………………………….11 Temperature Control…………………………………………………………..13 Pressure Control……………………………………………………………….15 Heat Sealing Jaws……………………………………………………………15

3. Software………………………………………………………………16     

System Requirements………………………………………………………..16 Installation…………………………………………………………………..16 Run the Software……………………………………………………………22 Delete the Software…………………………………………………………22 How to Use the Software……………………………………………………22

4. Operation………………………………………………………………24   

Overview………………………………………………………………………..24 Sample Preparation……………………………………………………………24 Operation Procedure…………………………………………………………..24

5. Maintenance…………………………………………………………..26   

General Maintenance………………………………………………………….26 Diagnostics and Troubleshooting……………………………………………..27 Technical Support………………………………………………………………27

3

Heat Sea l Te s ter

HST-H3 Appendix…………………………………………………………………..28 Ⅰ Unit Conversion Table………………………………………………28 Ⅱ Sealing Temperature and Dwell Time for General Materials…29 Ⅲ Customer Service Regulation……………………………………..30

4

Heat Sea l Te s ter

HST-H3

Preface About This Guide

Welcome to the Labthink PARAM® HST-H3 Heat Seal Tester. This manual contains a description of the operation and maintenance of your Heat Seal Tester.

Safety Precautions

Observe the following safety precautions when you operate or perform service on the HST-H3 Heat Seal Tester: Do not touch the area around heat sealing jaws during testing process and DO NOT replace the sealing jaws until they are cooled down. Power supply MUST be AC 220V 50/60Hz, and well grounded. DO NOT operate the instrument with other high-power electrical appliances. DO NOT place anything other than the test material in the sealing area.

This instrument SHOULD NOT be operated in intensely corrosive, combustible, explosive, or similar environment, or under intensive vibration or strong electromagnetic interference. DO NOT start the test until the desired temperature is reached and steady, or without the gas supply. Lock and DO NOT adjust the pressure knob when the desired pressure is set. The compressed gas should be clean and without any visible liquid drop or solid substances which may cause damage to the instrument.

DO NOT touch any region that labeled “Danger”

. 5

Heat Sea l Te s ter

HST-H3

1. Introduction In accordance with ASTM F 2029, YBB 00122003, and QB/T 2358 (ZBY 28004) standards, HST-H3 Heat Seal Tester is designed to optimize the sealing temperature, pressure, and dwell time of plastic films and composite materials for packaging in order to determine the heatsealability of such substances. The sealing pressure is controlled by adjusting the pressure gauge of the gas cylinder. The micro-computer regulates the up-and-down movement of the sealing jaw in order to seal the test materials under certain temperature, pressure and dwell time, from which the appropriate sealing conditions are generated to assist industrial production.

Specifications Sealing Temperature

Room Temperature~300°C

Accuracy

±0.2°C

Dwell Time

0.1~999.9s

Sealing Pressure

0.05~0.7 MPa

Heating Mode

Single heat surface/Double heat surface

Sealing Area

330 mm×10 mm

Gas Supply Pressure

0.5MPa~0.7MPa (not in supply scope)

Power supply

AC 220 V 50/60 Hz

Instrument dimensions

536 (L)mm×335(W) mm×413(H)mm

Net Weight

43 kg

6

Heat Sea l Te s ter

HST-H3 Unpacking Remove the instrument from the shipping carton and inspect for any signs of damage. Contact the delivery agent immediately if any damage observed. Leave the instrument in the shipping carton for the delivery agent to examine the damage. If there is no visible damage, check the contents of your shipment against the packing list. Notify Labthink immediately if there are any discrepancies. The instrument is shipped in a special designed box to prevent damage. If re-shipment is necessary, the instrument should be repacked in the original box. Please keep the box properly.

7

Heat Sea l Te s ter

HST-H3

2. Instrument Instrument Description Appearance

Figure 1 Instrument Appearance 

Upper Sealing Jaw



Lower Sealing Jaw

Controlled by the gas cylinder to move up and down and to provide pressure and heat for sealing. The surface of sealing jaws is covered with special coatings to isolate the test specimen to prevent sticking of specimen to the jaw. The coating should not be dismantled for safety concerns. Equipped with silicone rubber for the placement of test 8

Heat Sea l Te s ter

HST-H3

 

Pressure Knob Temperature Control Meter



LCD Screen

specimens and to evenly distribute sealing heat and pressure. The silicone rubber should not be dismantled for safety concerns. To adjust sealing pressure. To control and display sealing temperature of upper sealing jaw. To display test parameters.

Operation Panel

Figure 2 Operation Panel

Function keys “RESET” “BACK” “SETUP” “TEST” “Arrow key”

Force the system to enter standby mode Return to superior menu Confirm to enter a selected submenu, or save the settings Start a test Press “

” and “

” to move among options; press “

” and “

” to modify

the values of test parameters ※Pedal Switch: HST-H3 Heat Seal Tester is featured with a pedal switch. In “Test” interface, the function of this pedal switch is equivalent to the “Test” Key 9

Heat Sea l Te s ter

HST-H3 Installation Environmental Requirements:     

The instrument should be kept in an environment of 23±2 ℃ and 50% RH. The power supply should be AC 220 V 50/60 Hz and well grounded. Do not operate this instrument with other high-power equipments. The working environment should be clean and free of corrosive gas and electro-magnetic interference. Place the instrument on a working table which is stable and kept away from heat source.

Installation Procedures:     

Take the instruments out of the packing box and check all the items according to packing list. Before installing the instrument, please read this instruction manual carefully to understand all requirements for installation and operations. Place the instrument on a stable worktable carefully and leave at least 0.5m space for operation. Connect all the power cables of the instrument. Remove the φ6 PU tubing (100mm) connected with the gas inlet, then connect the instrument with the gas supply with the suppliedφ6 PU tubing.

Note: Please keep the shipping carton in case the instrument is returned for repair.

10

Heat Sea l Te s ter

HST-H3 Instrument System The instrument is controlled by micro-computer integrated in the circuit board, and displays on a LCD screen. 1. Start up the instrument and the welcome screen shows.

HST-H3 Heat Seal Tester Labthink 2. Press Setup key to enter main screen.

Main Test Cal. 25 / 03 / 2010

Preset Contact us 14: 50

3. In the “Main” screen, move the cursor to “Test” and click to enter into “Test” screen.

Test Time: 1.8 s Pressure: 281 KPa 25 / 03 / 2010 14: 50

This screen displays the preset dwell time and sealing pressure which could be set in “Preset” menu before test. The current date and time are shown in the bottom line. In “Test” interface, the heat sealing jaw starts working. Note: The operator should keep off from the instrument during the testing process. 11

Heat Sea l Te s ter

HST-H3 4. Return to main screen and move the cursor to “Preset” and click “Setup” key to enter into “Preset” screen.

Preset Time: 1.8 s Pressure: 281 KPa 25 / 03 / 2010 14: 50 Press “

” and “

Print: Y

” to move to choose the parameter to set and then press “

” and “

” to

modify the values of test parameters or change the status of the selected options. Press “Setup” to save the settings. For pressure setting, the procedures are as follows:  Make sure there is no other objects or substances between upper sealing jaw and lower sealing jaw 

Press “



Press “Test” key so that the upper sealing jaw will come down to touch the lower sealing jaw Pull the pressure knob and adjust it to increase or decrease the value of the “Pressure”. When the adjustment is finished, lock the pressure knob Press “Setup” key to save the settings Press “Back” key so that the upper sealing jaw will return to original position

  

” and “

” to move to “Pressure”

5. Return to the “Main” screen and select “Cal.” to view the calibration settings.

Calibration Pressure: 281 KPa Zero End 000

Calibration can be achieved from this screen. Due to the complexity of the calibration 12

Heat Sea l Te s ter

HST-H3 process, recalibration is not recommended for the user. For this purpose, a password is applied, which shows in the last line as “000”. The default password is “555”. The calibration procedure is followed:  



Connect the pressure sensor of the instrument with a pressure gauge featured with accuracy of 0.25% of reading value and resolution of 1kPa. Check zero point. Move the cursor to “Zero”. When the gas cylinder holds the upper sealing jaw in the upper position, the output pressure should be 0. Set it to 0 if not and press “SETUP” key to confirm. Check end point. Move the cursor to “End” and press “TEST” key. The upper sealing jaw moves down to meet the lower sealing jaw by the control of the gas cylinder. The displayed pressure should agree with the value on the pressure calibrator. Set to the correct value if not and press “SETUP” key to confirm.

Note: The instrument has been calibrated before shipment. If recalibration is required, please contact us for technical support. DO NOT change the calibration parameters or calibrate without instruction. 6. Return to the “Main” screen, and choose “Contact Us” to view contacting information.

Contact Us Tel: +8653185864214 Fax: +8653185812140 Web: www.labthink.cn

Temperature Control The temperature is controlled via P.I.D technology to achieve a precision of ±0.5℃. The parameters have been optimized by the manufacturer and customers have no need to adjust for any purposes. 

Operation panel. The operation panel of the temperature control meter is shown in Figure 4. 13

Heat Sea l Te s ter

HST-H3



The “PV” window displays the present value of temperature; “SV” shows preset value of temperature. Under the “SV” window, there are six signal lights: MAN has no function; PRG indicates the status of the meter; COM turns on when instrument is connected with the computer; OP1, AU1, and AU2 correspond to respective input/output operation. The four function keys in the lower line are “Setup”, “Decimal Shift”, “Value Decrease” (incorporated with Run/Pause function), and “Value Increase” (incorporated with Stop function), respectively. Temperature setup. You can adjust the temperature with the “Value Increase” and “Value Decrease” keys. For your convenience, you can adjust the position of the decimal by the “Decimal Shift” key to set the value of corresponding digit. For example, if the desired temperature is 250℃ and current temperature is only 25℃, you can move the decimal to the hundred digit and change it from 0 to 2. Then set the rest digits.

Figure 4 Temperature Control Meter 

Parameter setup. You can perform temperature calibration and self-calibration by adjusting the corresponding parameters. Hold the “Setup” key for about 2 seconds to enter the parameter setup status. Employ “Value Increase” and “Value Decrease” to adjust parameters and press “Setup” one more time to move to the next parameter. Hold “Decimal Shift” key for 2 seconds to return the previous parameter. To exit the parameter setup 14

Heat Sea l Te s ter

HST-H3





status, hold the “Decimal Shift” key and press “Setup”. The system will automatically exit this status if no operation for 25 seconds. Self-calibration. Self-calibration is usually not active unless the temperature accuracy has been compromised after long time use. To activate it, hold “Decimal Shift” for 2 seconds, “AT” displays and blinks and the system enters self-calibration status. When “AT” stops blinking, the process finishes and the CTRL parameter will be set to 3 or 4 (default value 1). Temperature calibration. Place the instrument in an environment with ±2℃ temperature fluctuation for more than 2 hours. The current temperature is T1, where the temperature displayed on the meter is T2. Set the Sc parameter to be the difference between T1 and T2 (ΔT) and make sure the sign of ΔT is correct.

Pressure Control The pressure control knob is located on the upper right side of the front panel of the instrument. To set the pressure, pull out the knob and adjust it to the desired pressure. Then push in to lock it when setup is finished.

Heat Sealing Jaws HST-H3 employs two heat sealing jaws of which the upper one is the main sealing jaw providing accurate heat sealing of specimen and the lower one is an auxiliary sealing jaw whose main function is to lessen the temperature difference between the upper sealing jaw and lower sealing jaw to protect the accuracy of the upper sealing jaw. The shape of the sealing jaw surface can be customized based on the need of customers. Standard Material of Sealing Jaws: Upper Sealing Jaw: smooth metal surface Lower Sealing Jaw: silicon rubber surface

15

Heat Sea l Te s ter

HST-H3

3. Software System Requirement Hardware requirement

CPU: Pentium 2.6 GHz or higher RAM: 128 MB Hard disk: 40 G CD-ROM: required Display: 1024x768, 75 Hz refresh rate

Operating system

Windows 98 or above

Installation 1.

Start Windows, and exit from other programs.

2.

Insert the software CD of Heat Seal Tester into CD-ROM.

3.

See the content of CD from “My computer” icon on desktop. Enter document “HST-H3”.

4.

Double click “Setup” icon

to start the installation of the Heat Seal Tester program.

Firstly, select language to install the software.

16

Heat Sea l Te s ter

HST-H3

You can stop the installation at any step by clicking “Cancel” .

Click “Next” to continue. 17

Heat Sea l Te s ter

HST-H3

Select “I accept the terms in the license agreement”. Then click “Next” to continue.

Input “User Name” and “Organization”, and select account authorization for this application, then click “Next” to continue. 18

Heat Sea l Te s ter

HST-H3

Select “Complete” setup type and click “Next” to continue installation. Or select “Custom” setup type and choose the program features you want installed and where the program will be installed, then click “Next”. If you select “Custom” setup type and click “Next”, the following interface will appear.

19

Heat Sea l Te s ter

HST-H3 Select the program features you want installed. If you want to change the path of the program installation, please click “Change”.

Click “OK” and back to previous interface, then click “Next” to continue.

Click “Install” to continue. 20

Heat Sea l Te s ter

HST-H3

Click “Finish” to finish installation. The system will automatically create a shortcut on desktop. 21

Heat Sea l Te s ter

HST-H3 Run the Software The software automatically creates a desktop shortcut “Launch HST-H3” when the installation is completed. Double click this icon can start the program, or click “Start” menu of Windows, point to “Program”, later point to “Labthink”, “HST-H3”, then point to “Launch HST-H3” and click it to start.

Delete the Software Click “Start” menu of Windows, point to “Program”, later point to “Labthink”, “HST-H3”, choose to delete this software “Uninstall HST-H3”, and then select “Yes”. Later continue the delete process under the program pilot till the software has been deleted completely.

How to Use the Software 1. 2.

Double click the desktop shortcut “HST-H3” to start the software. The main interface will appear after 2 seconds automatically.

1

3

2

22

Heat Sea l Te s ter

HST-H3 The main interface includes 3 sections: 1. Parameters of upper and lower sealing jaws according to current test. 2. Temperature chart of upper and lower sealing jaws. 3. Parameters list to all tests.

3.

Set the user parameters by clicking the “Preset” button

in toolbar. The user parameters

include sample name, test item, temperature and operator etc. Press “SETUP” to save the settings. 4.

Review the test report. Click the “Report”

in Toolbar to review the test reports which include all details of the

sample tests for users’ reference.

23

Heat Sea l Te s ter

HST-H3

4. Operation Overview 1. Open the gas supply 2. Power on the instruments 3. Set up the test parameters 4. Allow the heat sealing jaws to reach required temperature 5. Mount the samples and press “Test” key or use pedal switch to start the test 6. Test is finished 7. Shut down the instrument 8. Shut off the gas supply

Sample Preparation Please follow the corresponding standards to prepare test samples before testing.

Operation Procedure 1.

Open the gas supply and mount the prepared sample on the lower sealing jaw.

2.

Turn on the instrument and press any key to enter into the welcome screen shows.

HST-H3 Heat Seal Tester Labthink 3.

Before test, you should preset the test parameters in preset screen. Press any key except Reset to enter into Main screen. Move the cursor to “Preset”, and press “SETUP” key to 24

Heat Sea l Te s ter

HST-H3 enter the “Preset” menu.

Preset Time: 1.8 s Print: Y Pressure: 281 KPa 25 / 03 / 2010 14: 50 For Time Setting Employ “

” and “

” keys (Arrow keys) to set the desired dwell time.

For Pressure Setting Without anything in between the sealing jaws, move the cursor to “Pressure” and press “TEST” key. The upper sealing jaw moves down and meets the lower sealing jaw. Pull out the pressure knob and turn counterclockwise/clockwise to increase/decrease the pressure. Push in when finished adjusting and press “SETUP” key to confirm settings. The upper sealing jaws moves back to the upper position afterward. 4.

For temperature setup (please refer to the introduction to temperature control in last chapter), the red digitals in the upper row of the temperature meter stand for the current temperature, where the lower displays the preset temperature. The signal light shows the heating status: on for heating and off for standby. It takes about 30 minutes to reach the preset temperature.

5.

Start sealing. Once the temperature is steady for more than 5 min, place the test specimen on the lower sealing jaw wherein the sealant layer is parallel with the sealing jaws. Press “TEST” key or use the pedal switch to start the test and the upper layer moves down to seal the specimen. When the preset dwell time is reached, the upper sealing jaw lifts and sealing is finished. Remove the sealed specimen for further seal strength test.

Test Time: 1.8 s Pressure: 281 KPa 25 / 03 / 2010 14: 50

6.

Shut off the instrument and the gas supply. The system will save and memorize the current test parameters. 25

Heat Sea l Te s ter

HST-H3

5. Maintenance General Maintenance 1. 2. 3. 4.

Power off the instrument when tests are finished. Keep the instrument clean. Cover the instrument when it is not in use. It is normal the temperature of outer surface of the instrument rises during the testing process.

Notice: 

Keep off the sealing jaws during the testing process and DO NOT touch any region that labeled “

      

”.

Operate the instrument in a ventilated area. Do not replace the sealing jaws when they are still not cool down. Start testing only when the preset temperature is reached and stable. The temperature of lower sealing jaw should not exceed 200℃ for a long time. Once the pressure is set, please lock the pressure knob and do not adjust it again. Stop heating after tests are finished. Drain and dry the air compressor regularly

Diagnostics and Troubleshooting Problem

Possible Reasons

Solution Press “RESET” key.

No response during a test. Temperature displays 0 or above 600℃.

Pt resistor failure caused by sealing jaw movement, or calibration data messed up.

Replace Pt resistor and recalibrate.

No heating.

Faulty Pt resistor, fuse, SSR, or heating tube.

Replace corresponding component.

Insufficient Pressure.

Gas supply failure.

Check gas supply assembly.

System dead.

Short circuit or damaged fuse.

Check power supply or replace

26

Heat Sea l Te s ter

HST-H3 fuse. Desired seal strength not reached.

Sealing conditions not optimized.

Adjust sealing temperature, pressure, and dwell time

Contaminated sealing jaw.

Carefully remove contaminants or replace sealing jaw; DO NOT scratch the contaminant with hard objects.

Unevenly distributed pressure.

Adjusting the position of the upper sealing jaw via the screw nuts.

Technical Support Web support Please visit our website for more details of general problems and technical supports. http://www.labthinkinternational.com.cn

Contact us You can also reach us by telephone or e-mail: Tel: +86-531-85063398 E-mail: [email protected]

NOTICE: Before you call us for help, please make sure: Read the Instruction Manual thoroughly and check all possible problems; Place instrument and Instruction Manual at hand; Prepare the following information: Model, purchase date and manufacturing code.

27

Heat Sea l Te s ter

HST-H3

Appendix ⅠUnit Conversion Table 

1 bar = 105 Pa = 1000 mbar



1kgf/cm2 = 0.981×105 Pa



1 mm Hg = 133.3 Pa



1 mm H2O = 9.81 Pa



1 mm Hg = 13.6 mm H2O



1 atm = 760 mmHg = 1013 mbar = 101.3 kPa



1 s = 1000 ms = 106 us



1 at=1 kgf/cm2



1 atm=760 mm Hg



1 psi = 1 pound per square inch



1 lb. = 4.45 N



1 inch= 2.54 cm



1 psi = 6.89 kPa

28

Heat Sea l Te s ter

HST-H3

ⅡSealing Temperature and Dwell Time for General Materials

Material

Thickness

Suggested

Suggested

Temperature

Dwell Time

CPP

20-70μ

120℃---160℃

HDPE

20-90μ

125℃---150℃

MDPE

20-120μ

115℃---145℃

LDPE

20-160μ

105℃---150℃

NY/CPE

55-80μ

140℃---160℃

NY/CPP

55-80μ

145℃---170℃

OPP/CPE

40-80μ

130℃---155℃

OPP/CPP

40-80μ

135℃---160℃

OPP/MCPP

45-60μ

140℃---155℃

OPP/PE

40-80μ

130℃---150℃

OPP/PET/CPE

55-110μ

145℃---160℃

OPP/PET/CPP

55-90μ

105℃---160℃

OPP/PP

40-60μ

135℃---160℃

PET/AL/CPE

55-90μ

105℃---160℃

0.7s-1.0s

29

Heat Sea l Te s ter

HST-H3

III Customer Service Regulation  General Provisions ※ Purpose: This regulation is instituted to satisfy the need of customers with professional and timely service. ※ Content: This regulation includes items of general provisions, service instructions and response time. ※ Service Aim: Provide customers with professional, timely and continuous service. Try our best to satisfy the requirements of customers. ※ Service Principle: Integrity and sense of responsibility, sophisticated skill and customer oriented. ※ Performance Pledge: 1. Notice of lack of configuration should be provided in written form within thirty days from AWB date. Otherwise, it is considered as a conform information. Responses will be made within one workday after receiving written notification. 2. Products are warranted with free service and replace components of the same brand, type and specifications for a period of one year from the AWB data against defects of quality. 3. Replaced parts are warranted with free service of replacement or maintenance for a period of 90 days from the date of purchase against defects of quality. 4. Technical whole-life service will be provided outside warranty period. While cost of maintenance is free, cost of material as well as freight out and home is charged. 5. Repaired instruments are warranted with free service of replacement or maintenance for a period of 180 days from the return date of the repaired instrument against defects of quality. 6. Responses will be made within one hour after receiving service application. Disposing suggestions and solution program will be provided within one working day. ※ Contact information (8:00-17:00 Monday to Friday Peking time):  Tel: +86-531-85068566  Service Tel: +86-531-85063398  Fax: +86-531-85812140  E-mail: [email protected]  Http://www.labthinkinternational.com.cn  Zip Code: 250031  Address: No.144 Wuyingshan Road Jinan China 30

Heat Sea l Te s ter

HST-H3 ※ Exclusion: For products not sold by LABTHINK directly, the sales regulation abides by the service agreement and quality guarantee regulations between distributors and customers. Division of authority and responsibility as well as the quality guarantee regulations between corresponding distributor and LABTHINK follow the content of contract (oral agreement) between these two sides.

 Service instructions ※ Instructions for Initial Use 1. Open-package Inspection: Check the required configuration of product in contract with that of packing list. Make sure that all the parts mentioned in the packing list are included. (When taking out the products from its package, pay attention to the caution sign such as “Do not hold here”.) Be sure all seals on the instrument are intact. Otherwise please contact LABTHINK as soon as possible. 2. Product Commissioning: Please review the instruction manual and other materials supplied with the instruments carefully. Test the instruments under specified environment, parameters and operating mode. If there is any problem about our products, please don't hesitate to clarify it with LABTHINK before commissioning. (Software of instrument computer has been installed before delivery. Users can use the computer directly. Discs supplied with instrument should be kept carefully for later use) ※ Confirmation of Warranty Period Warranty period is supported from AWB date (Buyer and seller as well as the product mode are given clear indication). Without such supporting document, the warranty start date will be Ex-factory Date of the purchased products recorded by LABTHINK. In any of the following cases, there will be no maintenance free of charge. 1. Products are not purchased from LABTHINK or LABTHINK’s agents and distributors are not covered in the service regulation. 2. Defects of instrument or other components caused by the use of accessories or peripheral equipments not produced or authorized by LABTHINK. 3. Defects or damages of instruments or other components resulted from abnormal reasons (including unsatisfied power supply, working environment, not the first delivery, long term storage, outside force and foreign body entering instrument). 4. Defects or damages of instruments or other components caused by natural influences (such as earthquake or fire) or accidents (such as stealing and lost) 31

Heat Sea l Te s ter

HST-H3 5. Defects or damages of instruments or other components caused by dismantling, repair, and refit without LABTHINK permission. 6. Defects or damages of instruments or other components caused by plugging and connecting power cord , typing cord, communication cord and other connecting wire without cutting the power supply. 7. According to the international convention, correlation insurance company is directly responsible for charges of maintenance or complete damage resulting from serious damage during international transportation. Customers must inform LABTHINK within one working day after receiving the product. (For uninsured cargo, it is recommended that customers take out insurance) ※ Expiration of Warranty Period Warranty period is expired in one year from the AWB date (indicating the transaction sides and instrument type). Without above supporting document, the warranty start date will be the Ex-factory Date of the purchased instrument recorded by LABTHINK. ※ Charge of Service 1. Training a. LABTHINK provides customers with demonstration video of the sold products free of charge (attached with instrument or transmitted through Internet) b. On the premise that customers need and are willing to bear correlative expenses, LABTHINK engineer may arrive at the nominated places for product installation, debugging, training and commission. While maintenance service is free of charge, Users must undertake relating travel expense of the engineer (including round-trip air ticket, ticket, accommodation and other travel expense). With the signed (or reached) service contract (oral agreement) between two sides and prepaid relating expenditures, LABTHINK will send its engineer for contract service. 2. Maintenance a. For problem caused by product quality within warranty period, LABTHINK will provide product maintenance, technical support, freight out and home as well as cost of replaced components free of charge. b. Problem not caused by product quality within warranty period or maintenance outside warranty period is only charged for replaced components. However, freights home and out is at the customer’ side (only the replaced component is charged if transportation of instrument is not needed). c. Door-to-door service outside warranty period requires signed (or reached) service contract (oral agreement) between two sides. Users must undertake relating travel expense of the 32

Heat Sea l Te s ter

HST-H3 engineer (including round-trip air ticket, ticket, accommodation and other travel expense). With the signed (or reached) service contract (oral agreement) between two sides and prepaid relating expenditures, LABTHINK will send its engineer for contract service.

 Service Content 1. Please contact LABTHINK as soon as possible if there is problem during the process of operation. To help service man providing accurate and timely solution for your problem or trouble, be sure to make the following preparation before contact: (1)Check product type on product certification (or in instruction manual and manufacturer certificate); (2)Having reviewed instruction manual carefully and having it at hand at the time of contact (3)Make sure the protective tube and all the connecting wire are in good condition when there is malfunction. (4)Collect information about symptoms of problem (including displaying content and operation condition), problem time and other detailed information (customers who conceal actual information should be responsible for the lost) 2. LABTHINK will make telephone call or door-to-door return visit about using condition of product periodically.

 Response Time 1. LABTHINK’s personnel will response to help users solving the problem within one hour after receiving telephone call or written notice. It will give customers technical support through remote technologies such as telephone call, E-mail, and demonstration video. If component replacement is needed, the required component will be sent out within one working day (if there is charged fees, the required component will be sent out after receiving fax of transferred account). 2. If instruments need to be sent back for maintenance, LABTHINK will notify the user within one working day after receiving the product about the inspection information. Instruments will be repaired within two working days and send out within one working day after repairing (if there is charged fees, the product will be sent out after receiving fax of transferred account). 3. If technical support cannot be realized through remote technology while it is inconvenient to send instruments back, LABTHINK can offer field service. Charges and regulations are as above stated. Unless there are special circumstances, the response time is: 33

Heat Sea l Te s ter

HST-H3   

Notify customers the personal information of LABTHINK engineer within one workday after receiving customer notice and reaching agreement. On the premise that customers provide corresponding invitation and other necessary help, LABTHINK will get the engineer’s visa as soon as possible. LABTHINK engineer reach the conferred place within one week of the visa date.

 Others    

This regulation is the only open service regulations for customers. Herein other ones with special purpose or beyond the above items are refused. The right to explain this regulation is reserved by LABTHINK INSTRUMENTS CO., LTD. The right to revise this regulation is reserved by LABTHINK INSTRUMENTS CO., LTD. Distributors (agents) should refer to cooperation agreement additionally.

34

Heat Sea l Te s ter

HST-H3