84 1 1MB
![HI VISION Avius Specifications (Rev5.5) [PDF]](https://vdoc.tips/img/200x200/hi-vision-avius-specifications-rev55.jpg)
Ultrasound Diagnostic Scanner
 
 Specifications
 
 Step5.5 (Jun, 2017 Revised) Rev5.5
 
 Hitachi, Ltd. CONFIDENTIALITY: This material contains proprietary information of Hitachi and it shall not be disclosed to the third party or reproduced without the prior written permission of Hitachi. Copyright ©Hitachi, Ltd. 2017. All rights reserved.
 
 SN-U391E
 
 Trademarks or the registered trademarks The following is the trademarks or the registered trademarks in this manual or in this equipment. 
 
 HI VISION Avius, Avius, Real-time Tissue Elastography, HdTHI, HI REZ, Ultra BE/Ultrasound Broadband Engine, HINT-NAVI, Tool Bar, Fine Flow are registered trademarks of Hitachi in Japan.
 
 
 
 HI VISION is registered trademarks of Hitachi in the U.S.A. and Europe.
 
 
 
 Windows, Windows XP, Windows Embedded Standard 2009, Windows Vista, Windows 7, Windows 8, Windows 8.1, Windows NT, Windows 2000, Windows 98, Microsoft Excel, Office, PowerPoint, Windows Media, DirectX are trademarks or the registered trademarks of Microsoft Corporation in the U.S.A. and other countries.
 
 
 
 Linux® is a trademark registered to Linus Torvalds.
 
 
 
 CinePlayer Decoder Pack is used in Digital Video Archiving Software (Option). CinePlayer Decoder Pack is trademark of Sonic Solutions.
 
 
 
 In this manual or in this equipment, other brand names and product names are trademarks or registered trademarks of their respective companies.
 
 Abbreviations 
 
 In this document, Microsoft Windows Operating System is called Windows.
 
 
 
 In this document, the Windows® XP operating system is called Windows XP.
 
 
 
 In this document, Microsoft® Windows® Vista Operating System is called Windows Vista.
 
 
 
 In this document, Microsoft® Windows® 7 Operating System is called Windows 7.
 
 
 
 In this document, Microsoft® Windows® 8 Operating System is called Windows 8.
 
 
 
 In this document, Microsoft® Windows® 8.1 Operating System is called Windows 8.1.
 
 
 
 In this document, the Microsoft(R) Windows(R) NT Workstation Operating System or Microsoft(R) Windows(R) NT Sever Network is called Windows NT.
 
 
 
 In this document, Microsoft(R) Windows(R)2000 Professional Operating System or Microsoft(R) Windows(R)2000 Server Operating System or Microsoft(R)
 
 Windows(R)2000 Advanced Server
 
 Operating is called Windows NT.
 
 
 
 In this document, Microsoft® Windows® 98 Operating System is called Windows 98.
 
  No part of this publication may be transmitted into without the prior written permission of HITACHI.
 
  This publication is subject to change without notice.  Although HITACHI has attempted to ensure accuracy throughout this publication, HITACHI welcomes customer input on corrections and suggestions for improvements to this publication.
 
 SN-U391E
 
 Contents 1. General Description ...................................................................................................... 1 1.1.
 
 General ........................................................................................................................................ 1
 
 1.2.
 
 Features ...................................................................................................................................... 1
 
 2. System Composition ..................................................................................................... 6 2.1.
 
 Main System Composition ........................................................................................................ 6
 
 2.1.1.
 
 Ultrasound Scanner Main Unit .......................................................................................... 6
 
 2.1.2.
 
 Accessories ........................................................................................................................... 6
 
 2.2.
 
 Probes ......................................................................................................................................... 6
 
 2.2.1.
 
 Convex array probes ............................................................................................................ 6
 
 2.2.2.
 
 Linear array probes ............................................................................................................. 7
 
 2.2.3.
 
 Phased array probes ............................................................................................................ 7
 
 2.2.4.
 
 Other probes ........................................................................................................................ 8
 
 2.3.
 
 Options ....................................................................................................................................... 8
 
 2.4.
 
 Peripheral Devices ................................................................................................................... 10
 
 3. Specifications .............................................................................................................. 11 3.1.
 
 General ...................................................................................................................................... 11
 
 3.1.1.
 
 System ................................................................................................................................ 11
 
 3.1.2.
 
 Basic Function ................................................................................................................... 11
 
 3.2.
 
 Options ..................................................................................................................................... 22
 
 3.2.1.
 
 17-inch Digital LCD Monitor (EZU-MT29-S1) ................................................................. 22
 
 3.2.2.
 
 17-inch Digital LCD Monitor (Flexible arm) (EZU-MT30-S1) ........................................ 22
 
 3.2.3.
 
 Biomedical Waveform Display Unit (EZU-EK28) ............................................................ 22
 
 3.2.4.
 
 CW Doppler Unit (EZU-ST8) ............................................................................................ 22
 
 3.2.5.
 
 Real-time Tissue Elastography Unit (EZU-TE6) ............................................................. 23
 
 3.2.6.
 
 Real-time Tissue Elastography software (Strain Histogram) (EZU-TESH1) ................. 24
 
 3.2.7.
 
 Digital Video Archiving Software (EZU-DV4) .................................................................. 25
 
 3.2.8.
 
 Logon Management Software (EZU-LG2) ........................................................................ 26
 
 3.2.9.
 
 DICOM Software(Transfer and Media Storage) (EZU-FC11) ......................................... 27
 
 3.2.10. DICOM Software (WorkList) (EZU-FC11W) / DICOM Software (Printer) (EZU-FC11P)31 3.2.11.
 
 DICOM Software (Query / Retrieve) (EZU-FC11Q) ......................................................... 31
 
 3.2.12. Omni-directional M mode display function (EZU-OD3) .................................................. 32 3.2.13. Advanced Measurement Software (EZU-AM7) ................................................................ 32 3.2.14. IMT Measurement Software (EZU-IM2) .......................................................................... 33 3.2.15. Software Specified for Contrast Agents (EZU-CH8) ........................................................ 33 3.2.16. WideView Software (EZU-WS8) ........................................................................................ 35 3.2.17. Stress Echo Software (EZU-SE6)...................................................................................... 36 3.2.18. Three-dimensional display unit (Real-Time) (EZU-4D7)................................................. 36 3.2.19. Three-dimensional Display Software (STIC) (EZU-4DST2) ............................................ 39 
 
 SN-U391E
 
 3.2.20. Three-dimensional Display Software (Freehand) (EZU-3D8) ......................................... 40 3.2.21. Remote Control Unit (EZU-RH4) ..................................................................................... 42 3.2.22. Remote Control Interface Unit (EZU-RiF5) ..................................................................... 42 3.2.23. Picture in Picture Software (EZU-PP5) ............................................................................ 42 3.2.24. Network Interface Unit (EZU-NW1) ................................................................................ 43 3.2.25. Analog Capture Card Unit (EZU-ACP2) .......................................................................... 43 3.2.26. Digital Capture Card Unit (EZU-DCP3) .......................................................................... 43 3.2.27. Video Interface Unit (EZU-ViF2) ...................................................................................... 43 3.2.28. Expanded USB Unit (EZU-UP2) ...................................................................................... 43 3.2.29. External LCD Monitor Unit (EZU-EL2)........................................................................... 44 3.2.30. Jelly Warmer Unit (EZU-JW1) ......................................................................................... 44 3.2.31. Alpha numeric keyboard Unit (EZU-AK1E) .................................................................... 44 3.2.32. High Neck Unit (EZU-HN1) ............................................................................................. 44 3.2.33. Probe Hanger (EZU-TH7) ................................................................................................. 44 3.2.34. Processor Tray (EZU-TY3) ................................................................................................ 44 3.2.35. Junction Box (PUN-JBP1) ................................................................................................ 44 3.3.
 
 Power requirements ................................................................................................................. 44
 
 3.4.
 
 Environment conditions .......................................................................................................... 45
 
 3.5.
 
 Exterior view ............................................................................................................................. 45 
 
 SN-U391E
 
 1. General Description 1.1. General Model HI VISION Avius is a multi-functional ultrasound diagnostic scanner in which Doppler, ColorFlow Mapping, etc. are provided and all circuits related to image quality are fully digitalized. As not only linear and convex scan type probes, but also phased array scan type probe can be operated with HI VISION Avius, it is possible to use it for any field of application. Furthermore, High-definition dTHI (HdTHI) mode (Hitachi unique technology) are provided in standard function. This mode provide “Clear image” to reduce artifact such as vibration echo, and to improve spatial resolution and contrast resolution. HI Com function, HI REZ+ function, Trapezoidal display function, Oblique B-mode image function, real-time Doppler trace, cine memory, patient data management function, and Real-time Archiving function can be used on the standard model. Wide view software, Stress echo software, Software specified Contrast Agents, cardiac measurement software, Real-time Tissue Elastography, omni-directional M (ODM) mode, Remote control unit, Picture in Picture software can be used optional function. It is convenience for user to adopt Ergonomic design and 17-inch digital LCD Monitor and more improved High-definition screen-displayed format.
 
 1.2. Features  True digital architecture Due to beamformer which is decisively concerned to the image quality and all image processing are digitalized, high spatial resolution, wide dynamic range and high contrast resolution are realized. Ultrasound transmit circuit is improved by high performance transmit amplifier, and ultrasound receiving circuit is improved by high performance pre-amplifier. Frame rate is improved much more faster comparing with conventional system. High quality image is digital image processing application. This architecture that is achieved high quality image is good for new generation.  Quick response Mode changing, image condition settings, probe changing and application selection can be performed quickly so that operator can pay attention to the patients.  Usable for wide field of application Linear, convex and phased array scan type probe are usable. It is possible to cover from abdominal and obstetric to cardiovascular fields.
 
 -1-
 
 SN-U391E
 
  17-inch digital LCD Monitor It is adopted 17-inch digital LCD Monitor. LCD Monitor provides high quality image. The LCD simple arm that optimizes the position of the LCD monitor and the operational panel provides ease of use. The LCD flexible arm that can adjust height and four-way position provides freely position of monitor.  High-definition screen-displayed format The resolution of the monitor is increased and the layout of the screen is displayed to the high-definition. The inspection efficiency is improved by additional information and the operation assistance on the screen. - Design The screen design which looks big ultrasound image. - Tab menus It is easy to use the tab menus to switch applications, perform measurements, and activate annotations. Because it is positioned out of diagnosis area. - Brightness Setting It is adjustable character and background brightness level on the screen depends on setting condition. - Image size change function The image size can be changed under certain conditions.  Ergonomic design - Design Design is adopted to focus on user’s “easiness to use”, like front layout of probe connector, based on ergonomics. - Easily operable keyboard As key arrangement is taken in consideration of frequency of use and operational has improved. An optional alphanumeric keyboard that is adopted backlight keyboard improves key response and visibility. - Excellent mobility with four swivel casters As all the four casters are of swivel type, it is easy to move in narrow space. Due to reduced friction force to floor it is easy to move from ward to ward. - Probe rest It connects 3 probes simultaneously and has 1 dummy connector.  Use interface Workflow is improved because of new GUI. It is a few procedures for many functions to use screen buttons.  Online Manual It is applicable to browse instruction manual on your HI VISION Avius.
 
 -2-
 
 SN-U391E
 
  Various B-mode scan - HI Com function Compound function provides high spatial resolution, wide dynamic range and high contrast resolution. - Trapezoidal display function Trapezoidal display is applicable for linear probe. It is improved diagnosis area for linear probe. - Oblique B-mode image function This function enables you to use oblique function not only CFI/CFA mode but B mode.  High quality image technique. - High-definition dynamic tissue harmonic imaging (HdTHI) Mode High-definition dynamic tissue harmonic imaging (Hitachi unique technology) provide “Clear image” to reduce artifact such as vibration echo, and to improve spatial resolution and contrast resolution. - Adaptive Imaging (HI REZ+) Adaptive imaging is the technique to generate optimum image with changing filter characteristic depending on input signal. This technique which installed this ultrasound diagnostic scanner reduces speckle noise on B-mode images and provides uniformly and high contrast images. - Fine Flow function It is the function to display high frame rate and fine high-definition color Doppler image. It is possible to display small vessel less blooming than original system. - Automatic image optimization (HI Support) It is the function to adjusts B mode parameters (Gain, TGC and Depth-specific PDC) and PW mode parameters (Gain, PRF and Baseline) so that it may become the optimal image automatically, and leads to shortening of inspection time.  Real time Doppler measurement function Automatic tracing function and measuring function in Doppler-mode have been allowed to apply only to frozen images, but with this function it is possible to perform it to real time images which contributes to shorten the time for examination. It is displayed position of Doppler waveform and Vp, Vd values of Doppler waveform in real time. After Freeze, you can select the trace and measuring results of any past waveform by playing cine memory images frame-by-frame and turning the PW Gain encoder.
 
 -3-
 
 SN-U391E
 
  Filing function. - Examination data management Examination data (images measurement results and other data) can manage and operate together, so it is easy to refer data before an examination and after an examination. - Multiple functions to save data No image degradation because of digital image recording. This filing function can record still images and multi images not only HDD include system) but also DVD-RAM, DVD-R, USB memory and USB HDD. - Easy-to-use filing function In the Filing window, you can view examination data. It is consideration for easy-to-use in the routine examination by filtering Patient ID or Patient name or keyword search. Thumbnail view provides easy access to destination image. Review images can display 1/4/9/16 screen depend on choosing images. - Patient data management You can select the patient in the list view who had examination once.  Omni-Directional M mode (ODM) function Usually the normal M-mode image is controlled by the beam direction. With this function, it is possible to display the M-mode image on an optional line on the B-mode image, so it enables to perform more accurate left ventricle volume measurement.  Real time Tissue Elastography Real time Tissue Elastography imaging is a qualitative imaging technique that displays the relative tissue stiffness by detecting strain.  Digital video archiving function It is the function to record a B-mode and a color mode image as a digital movie file on real time. The operational of the monitoring of contrast agents and recording of digital motion picture is improved by this function. It is possible to use Clip capture mode that is suitable for clip analyses and Long time recording mode.  Contrast harmonic imaging function With this function enabling to produce images further effectively from signal coming out of the contrast agent it is possible to improve to diagnose the existence of tumor with ease. Two or more ROIs can be set up to the video recorded by contrast echo inspection, the Time-Average (Time-Intensity) value in ROI can be indicated by graph, and quantitative analysis is possible.
 
 -4-
 
 SN-U391E
 
  DICOM function This software can transfer ultrasound images and measurement reports to a storage server. It has a Storage Commitment function for ultrasound images. This software can output and read ultrasound images in DICOM file format. This software has a function to obtain WorkLists from the MWM server. It has a function to send examination start and completion notifications for ultrasound images to the MPPS server. And it can be sent Ultrasound images to grayscale or color DICOM printers.  4D function 3D ultrasound images can be viewed in almost real-time. As the scanning part of the probe swings automatically in a fan-like fashion, it sends signals to the three-dimensional display unit, which rapidly builds a 3D image based on the probe’s signal.  STIC function Multiple 3D images representing a single heartbeat can be displayed by automatically swinging the scanning part of the probe in a fan-like fashion and estimating the cardiac cycle from the obtained images.  3D function The software is able to collect sequential 2D image data by using a probe to scan at a steady rate and then display the collected data as 3D images.  Advanced Measurement function - %WT measurement Measurement useful to heartbeat diagnosis of thickness change of heart muscle etc. is enabled by tracing automatically the movement toward a heart tissue in arbitrary positions. - Color tissue tracking measurement The changes of myocardium thickness are automatically tracked and their strain amount is displayed as colored movies on the screen. The graph of the changes of myocardium, caliper on the graphs, movie outputs, tracking data outputs can be possible. Bull’s eye can be created by loading the tracking data which three cross-sectional surface of APICAL, MID and BASAL. - LA tracking measurement The motion of left atrium wall is automatically tracked so that the volume of left atrium and other parameters against time are measured. - IMT measurement IMT (Intima-Media Thickness) measurement is enabled by detecting the position of intima and adventitia of carotid artery.  Remote control function Freeze, REF, mode and etc. keys of the operation panel can be switched by remote control. -5-
 
 SN-U391E
 
 2. System Composition 2.1. Main System Composition 2.1.1. Ultrasound Scanner Main Unit  Ultrasound transmitting / receiving unit  Digital scan converter unit  Pulsed Doppler unit  Color flow mapping unit  HI Com function  Adaptive Imaging function (HI REZ+)  TDI (Tissue Doppler imaging) function  Filing/Patient data management function  Tissue harmonic imaging function  Cine-memory unit  Automatic image optimization (HI Support)  High density zoom function (HI Zoom)  Power supply unit  DVD (DVD-RAM, DVD-R) drive
 
 2.1.2. Accessories  System CD
 
 2
 
  Hexagonal wrench bar
 
 1 set
 
  Spare fuses
 
 1 set
 
  Cords
 
 1 set
 
  Instruction manual (Include CD)
 
 1 set
 
  DVD-RAM (4.7GB)
 
 1
 
 2.2. Probes 2.2.1. Convex array probes  4-2MHz
 
 10R
 
 EUP-C511
 
  5-2 MHz
 
 40R
 
 EUP-C514
 
  5-1 MHz
 
 50R
 
 EUP-C715
 
  5-2 MHz
 
 60R
 
 EUP-C516
 
  6-3 MHz
 
 40R
 
 EUP-C524
 
  5-2 MHz
 
 40R (Real-time 3D)
 
 EUP-CV714
 
  7-2MHz
 
 40R (Real-time 3D)
 
 EUP-CV724
 
  6-3 MHz
 
 40R (Real-time 3D)
 
 EUP-CV524
 
  8-4 MHz
 
 20R
 
 EUP-C532 -6-
 
 SN-U391E
 
  5-2 MHz
 
 20R (Biopsy)
 
 EUP-B512
 
  6-1 MHz
 
 20R (Biopsy)
 
 EUP-B712
 
  5-2 MHz
 
 40R (Biopsy)
 
 EUP-B514
 
  5-1MHz
 
 50R(Biopsy)
 
 EUP-B715
 
  8-4 MHz
 
 10R (Transvaginal wide view)
 
 EUP-V53W
 
  10-2 MHz
 
 10R (Transvaginal wide view)
 
 EUP-V73W
 
  8-4MHz
 
 10R (Transvaginal, Real-time 3D)
 
 EUP-VV531
 
  8-4 MHz
 
 10R (Transvaginal, Real-time 3D)
 
 EUP-VV731
 
  8-4 MHz 10R Convex / 8-4 MHz 10R Convex (Intracavital)
 
 EUP-CC531
 
  8-4 MHz 10R Convex / 8-4 MHz 10R Convex (Intracavital)
 
 EUP-CC531S
 
  8-4 MHz 10R Convex / 10-5 MHz Linear (Intracavital)
 
 EUP-U533
 
  8-4 MHz
 
 10R (Transrectal)
 
 EUP-U531
 
  8-4 MHz
 
 10R (Finger top)
 
 EUP-F531
 
  7.5 MHz
 
 40R (Finger top)
 
 EUP-F334
 
  10-5 MHz
 
 10R (Laparoscopic)
 
 EUP-OL531
 
  7.5 MHz
 
 40R (Laparoscopic)
 
 EUP-OL334
 
  10-4MHz
 
 20R (Intraoperative)
 
 EUP-O732T
 
  10-5 MHz
 
 6R (Transrectal, 19cm, 360°)
 
 EUP-R54AW-19
 
  10-5 MHz
 
 6R (Transrectal, 33cm, 360°)
 
 EUP-R54AW-33
 
 2.2.2. Linear array probes  7-3 MHz
 
 EUP-L52
 
  9-4 MHz
 
 EUP-L73S
 
  10-5 MHz
 
 EUP-L53
 
  10-5 MHz
 
 (Wide view)
 
 EUP-L53L
 
  13-6 MHz
 
 EUP-L54MA
 
  13-5MHz
 
 EUP-L74M
 
  18-5MHz
 
 EUP-L75
 
  14-6 MHz
 
 EUP-L65
 
  13-5MHz
 
 (Real-time 3D)
 
 EUP-LV74
 
  10-5 MHz
 
 (Intraoperative)
 
 EUP-O53T
 
  13-7MHz
 
 (Intraoperative)
 
 EUP-O54J
 
 2.2.3. Phased array probes  4-2 MHz
 
 (Steerable CW)
 
 EUP-S50A
 
  5-1 MHz
 
 (Steerable CW)
 
 EUP-S70
 
  7-3 MHz
 
 (Steerable CW)
 
 EUP-S52
 
  9-2MHz
 
 (Steerable CW)
 
 EUP-S72
 
  8-3 MHz
 
 (Steerable CW, Electric multiplane transesophageal)
 
 EUP-ES52E
 
  6-3.5 MHz
 
 (Steerable CW, Multiplane transesophageal)
 
 EUP-ES52M
 
 -7-
 
 SN-U391E
 
 2.2.4. Other probes  7.5 MHz
 
 10R (Upper G.I. Fiberscope, made by HOYA)
 
 FG-34UX
 
  7.5 MHz
 
 10R (Upper G.I. Fiberscope, made by HOYA)
 
 FG-36UX
 
  7.5 MHz
 
 10R (Upper G.I. Fiberscope, made by HOYA)
 
 FG-38UX
 
  7.5 MHz
 
 10R (Upper G.I. Videoscope, made by HOYA)
 
 EG-3630U
 
  10-5MHz
 
 6R (Upper G.I. Videoscope, made by HOYA)
 
 EG-3270UK
 
  10-5 MHz
 
 6R (Upper G.I. Videoscope,made by HOYA)
 
 EG-3630UR
 
  10-5 MHz
 
 10R (Upper G.I. Videoscope, made by HOYA)
 
 EG-3830UT
 
  10-5 MHz
 
 6R (Upper G.I. Videoscope, made by HOYA)
 
 EG-3670URK
 
  10-5 MHz
 
 10R (Upper G.I. Videoscope, made by HOYA)
 
 EG-3870UTK
 
  10-5 MHz
 
 6R (Videobronchoscope, made by HOYA)
 
 EB-1970UK
 
  10-5 MHz
 
 6R (Videobronchoscope, made by HOYA)
 
 EB19-J10U
 
 2.3. Options  17-inch Digital LCD Monitor Unit
 
 EZU-MT29-S1
 
  17-inch Digital LCD Monitor Unit (Flexible arm)
 
 EZU-MT30-S1
 
  Biomedical Waveform Display Unit
 
 EZU-EK28
 
  CW Doppler Unit
 
 EZU-ST8
 
  Software Specified for Contrast Agents
 
 EZU-CH8
 
  DICOM Software (Transfer and Media Storage)
 
 EZU-FC11
 
  DICOM Software (Printer)
 
 EZU-FC11P
 
  DICOM Software (Worklist)
 
 EZU-FC11W
 
  DICOM Query/Retrieve Software
 
 EZU-FC11Q
 
  Three-dimensional display unit (Real-Time)
 
 EZU-4D7
 
  Three-dimensional Display Software (STIC)
 
 EZU-4DST2
 
  Three-dimensional Display Software (Freehand)
 
 EZU-3D8
 
  Real-time Tissue Elastography Software
 
 EZU-TE6
 
  Real Time Tissue Elastography Software (Strain Histogram)
 
 EZU-TESH1
 
  ODM Software
 
 EZU-OD3
 
  Advanced Measurement Software
 
 EZU-AM7
 
  IMT Measurement Software
 
 EZU-IM2
 
  WideView Software
 
 EZU-WS8
 
  Stress Echo Software
 
 EZU-SE6
 
  Digital Video Archiving Software
 
 EZU-DV4
 
  Remote Control Unit
 
 EZU-RH4
 
  Remote Control I/F Unit
 
 EZU-RiF5
 
  Picture in Picture Software
 
 EZU-PP5 -8-
 
 SN-U391E
 
  Logon Management Software
 
 EZU-LG2
 
  Alpha numeric keyboard Unit
 
 EZU-AK1E/F
 
  Network Interface Unit
 
 EZU-NW1
 
  Digital Capture Card Unit
 
 EZU-DCP3
 
  Analog Capture Card Unit
 
 EZU-ACP2
 
  Video Interface Unit
 
 EZU-ViF2
 
  Foot Switch (for Freeze)
 
 EZU-FS1
 
  Foot Switch (for Freeze / User Define)
 
 EZU-FS2
 
  Expanded USB Unit
 
 EZU-UP2
 
  External LCD Monitor Unit
 
 EZU-EL2
 
  Jelly Warmer Unit
 
 EZU-JW1
 
  High Neck Unit
 
 EZU-HN1
 
  Probe Hanger
 
 EZU-TH7
 
  Processor Tray
 
 EZU-TY3
 
  5MHz Pencil CW Probe Adapter
 
 EZU-AC1
 
 - 5MHz Pencil CW Probe Sonde Crayon 5MHz P050CN20W00+AD (with BNC cabling) (by AXON’CABLE S.A.S)  10MHz Pencil CW Probe Adapter
 
 EZU-AC2
 
 - 10MHz Pencil CW Probe Sonde Crayon 10MHz 10D4 (with BNC cabling) (by AXON’CABLE S.A.S)  5MHz/10MHz pencil CW probe Interface
 
 EZU-AC3
 
  Junction Box
 
 PUN-JBP1
 
 -9-
 
 SN-U391E
 
 2.4. Peripheral Devices Peripheral device Black & white printer
 
 Manufacturer SONY
 
 Manufacturer’s model, type UP-D897 UP-897MD
 
 (Notice II)
 
 UP-X898MD MITSUBISHI
 
 P93DW P93E/W
 
 (Notice II)
 
 P95DE/DW Color printer
 
 SONY
 
 UP-D23MD UP-D25MD
 
 MITSUBISHI
 
 CP30DW
 
 DVD Video Recorder
 
 SONY
 
 DVO-1000MD
 
 PC printer
 
 HP
 
 HP Photosmart D5460 *
 
 (Notice I)
 
 HP OFFICEJET 6000 * Wireless LAN adapter
 
 CISCO
 
 Aironet 1130AG *
 
 * These devices meet IEC60950 compliance. (Notice I): EZU-ViF2 is required to record. EZU-ViF2 and EZU-ACP2 is required to playback. (Notice II): EZU-ViF2 is required to print. (NoticeⅢ): It is necessary to detach a rear handle to carry out simultaneous loading of a color printer and a DVD video recorder.
 
 - 10 -
 
 SN-U391E
 
 3. Specifications 3.1. General 3.1.1. System  OS
 
 Windows Embedded Standard 2009
 
  DVD
 
 DVD-RAM DVD-R
 
  USB memory USB HDD
 
 Connectable
 
  Network
 
 10/100 BaseT (EZU-NW1 is required to network connection)
 
  Starting time
 
 About 100 seconds (About 70 seconds in Hibernation mode)
 
 3.1.2. Basic Function 3.1.2.1
 
 Type of probes
 
  Convex array probes
 
 1.0~10 [MHz]
 
  Linear array probes
 
 3.0~18 [MHz]
 
  Phased array probes
 
 1.0~9.0 [MHz]
 
 3.1.2.2
 
 Image examination modes
 
  B, M mode
 
 B, B/B, B/M, M
 
  Doppler mode
 
 B/D,D
 
  CFM mode
 
 CFM-B, CFM-B/CFM-B, CFM-B/CFM-M, CFM-M, CFM-B/D
 
  Tissue Doppler mode
 
 TDI-B, TDI-M, TDI-PW
 
  Biplane mode
 
 B/B (Real-time biplane display)
 
  Harmonic
 
 dTHI-R,G,P
 
 : Filter method
 
 dTHI-W-R,P : WPI (Wideband Pulse Inversion) method HdTHI-R,P  Compound
 
 : HdTHI (High-definition dynamic tissue harmonic imaging)
 
 HI Com Applicable in Doppler and CFM mode
 
  Trapezoid
 
 Max display angle ±25°(Depend on probes)
 
  Oblique B-mode image
 
 Right and lest each 3 steps (Depend on probes)
 
 3.1.2.3
 
 Gain adjustment
 
  Depth Gain Control (DGC)
 
 8 steps, 30dB variable, selectable Fixed depth/Variable
 
  B mode gain
 
 60dB (adjustable in real-time and frozen)
 
  M mode gain
 
 60dB (adjustable in real-time and frozen)
 
  Doppler mode gain
 
 60dB (adjustable in real-time and frozen)
 
  CFM mode gain
 
 60dB
 
 3.1.2.4
 
 Focus
 
  Transmission
 
 Max 10 steps (applicable to the setting of Focus width at multiple focus) - 11 -
 
 SN-U391E
 
  Reception
 
 Continuous dynamic focusing with variable aperture
 
  Other
 
 2-dimentional apodization
 
 3.1.2.5
 
 Frequency selection
 
  Fundamental
 
 Max. 5 steps
 
  Harmonic
 
 Max 6 steps (Depend on probes)
 
  HI Com: fundamental
 
 Max. 5 steps (Depend on probes)
 
  HI Com: harmonic
 
 Max. 6 steps (Depend on probes)
 
 3.1.2.6
 
 Acoustic power control
 
  Switching steps
 
 5 ~ 100% (5% steps)
 
 3 steps (CW mode)  Applicable independent setup
 
 3.1.2.7
 
 B mode image display
 
  Single image  Dual image
 
 Different probe and different depth can be displayed. Up-and-down can be selected for linear probe.
 
 3.1.2.8
 
 Depth range
 
  20mm~360 mm, (5mm steps) (Depend on probes)
 
 3.1.2.9
 
 Display format change in frozen
 
 Pan zoom, HI zoom, orientation (flip horizontal, flip vertical), 90 deg rotation, Image size change
 
 3.1.2.10
 
 Zoom function
 
  Pan zoom
 
 Adjustable position and size, display magnification: max. 8 times
 
  HI zoom
 
 Adjustable position and size depend on the probe shape
 
 3.1.2.11
 
 Image Size 80~100% (1% steps)
 
 3.1.2.12
 
 Display angle (width) 6 steps
 
 3.1.2.13
 
 Scan line density 4 steps
 
 3.1.2.14
 
 Biopsy guideline 6 selections (color: 2 types, size: 3 types) - 12 -
 
 SN-U391E
 
 (Notice) Biopsy guideline can not be displayed, when an application that image split setting of B-mode is up-and-down is selected.
 
 3.1.2.15
 
 Marking assist display function
 
  Applicable Probe
 
 EUP-L75
 
  Condition of use
 
 A B-mode image or a CFM-mode image appears in single display
 
  Display method
 
 The Screen button pushing
 
  Displayed line
 
 A solid line and a dash line
 
 3.1.2.16
 
 Orientation
 
  Flip vertical and horizontal
 
 B mode, Doppler mode
 
  Flip vertical
 
 B mode: 90 deg steps rotation
 
 3.1.2.17
 
 Image processing (B mode and M mode)
 
  Dynamic range
 
 Max. 90dB 45~ 90 dB, 5 dB steps (can be changed in real-time and frozen)
 
  Enhancement
 
 - Enhance
 
 4 steps (can be changed in real-time and frozen without M mode)
 
 - Weak-signal enhance
 
 Echo enhancement function, 8 steps
 
  Texture
 
 2 steps (without M mode)
 
  Persistence
 
 8 steps, including OFF
 
  AGC (auto gain control)
 
 4 steps
 
  Gray map
 
 8 steps (can be changed in real-time and frozen)
 
  Adaptive imaging
 
 HI REZ+ function, ON (8 steps)/OFF (can be changed in real-time and frozen)
 
  B(M)color
 
 29 maps (include B/W)
 
  PDC
 
 29 steps
 
  Receiving Filter
 
 3 steps
 
  Post processing
 
 γ-correction, rejection and saturation (can be changed in real-time and frozen)
 
 3.1.2.18
 
 M mode image display
 
  Display method
 
 - Real-time
 
 Moving bar
 
 - Cine playback
 
 Scroll
 
  Sweep speed
 
 1,1.5,2,3,4,6,8,10 s/screen (can be changed in real-time and frozen)
 
  Vertical shift
 
 Applicable
 
  Display area
 
 Side-by-side or up-and-down (can be changed in real-time and frozen), split ratio: 1:1 (side-by-side), 1:1, 1:2, 2:1 (up-and-down)
 
 - 13 -
 
 SN-U391E
 
 3.1.2.19
 
 Doppler function
 
  Display mode
 
 B/D (B-Freeze, B/D real), D
 
  Display area
 
 Side-by-side or up-and-down (can be changed in real-time and frozen), split ratio: 1:1 (side-by-side), 1:1, 1:2, 2:1 (up-and-down)
 
  Analysis
 
 FFT method, 3 steps, max. 128 points
 
  Pulse repetition frequency (PRF)
 
 1, 1.3, 1.5, 1.8, 2, 2.2, 2.5, 3, 3.5, 4, 4.5, 5, 5.5, 6, 6.5, 7, 7.5, 8, 9, 10, 11, 12, 14, 16, 18, 20 [kHz], total 26 steps  Reference frequency (REF)
 
 3 frequencies: (Depend on probes)
 
  Wall filter
 
 12 steps (maximum value : 1/16 of PRF)
 
  Baseline shift
 
 Applicable (in real-time and frozen)
 
  Detectable depth
 
 0mm ~360 mm
 
  Sample length
 
 0.5 mm ~ 5.0 mm: 0.5 mm step, 5.0 mm ~ 15.0 mm: 1.0 mm step
 
  Measurement
 
 - Real-time
 
 Velocity
 
 - Frozen
 
 Velocity, Acceleration, Index for stenosis, Pressure gradient, Flow volume, Time average velocity, Ratio-time, Ratio-velocity, Average velocity, PHT (Pressure Half Time), HR (Heart Rate), TVI (Time Velocity Integral), Auto trace is available
 
 - Limitations  Audio input / output
 
 HR: 25 ~ 300, measurement is not applicable in cine playback. 2ch speaker mounted (for forward / reverse flow), VCR audio input / output (for forward / reverse flow)
 
  Maximum detectable velocity
 
 ±3.65m/s (2.0 [MHz], 20 [kHz])  B/D simultaneous
 
 Up to 20 [kHz]
 
  Display method
 
 - Real-time
 
 Moving bar
 
 - Cine playback
 
 Scroll
 
  Sweep speed
 
 1, 1.5, 2, 3, 4, 6, 8, 10 s /screen (can be changed in real-time and frozen)
 
  Angle correction
 
 Applicable (in real-time and frozen)
 
  FFT-γ
 
 8 steps (can be changed in real-time and frozen)
 
  Oblique function
 
 Linear probe 0°,±10°,±15°,±20° (Depend on probes)
 
  D-Color
 
 29maps (can be changed in real-time and frozen)
 
  Post-processing
 
 γ-correction, rejection and saturation (can be changed in real-time and frozen)
 
  Unit
 
 3types (cm/s, m/s, kHz)
 
  Smoothing
 
 PW simultaneous mode only
 
  Doppler real-time measurement
 
 - Upper trace
 
 Trace upper the baseline - 14 -
 
 SN-U391E
 
 - Down trace
 
 Trace bottom the baseline
 
 - Both trace
 
 comparing for peak of both upper and bottom baseline, and trace either one higher
 
 3.1.2.20
 
 Color flow mapping function
 
  Display mode
 
 CFM-B, CFM-M, CFM-B/D (B-freeze, ALT), CFM-B/CFM-M Applicable dual simultaneous display of CFM and B/W
 
  CFM-B
 
 - Color box angle
 
 Continuously selectable in B-mode image display
 
 - Color box depth
 
 Continuously selectable in B-mode image display
 
 - Color ROI
 
 Color ROI guideline display on off selectable
 
 - Oblique scan
 
 Linear probes: 0, ±10, ±15, ±20 deg (Depend on prbes) Operations that are linked with or independent of Doppler cursor angles can be performed.
 
  CFM-M
 
 - Display method 
 
 Real-time
 
 Moving bar
 
 
 
 Cine playback
 
 Scroll
 
 - Sweep speed
 
 1, 1.5, 2, 3, 4, 6, 8, 10 s /screen (can be changed in real-time and frozen)
 
  Color display
 
 - Velocity - Velocity variance The direction of blood and velocity is displayed in brightness of red and blue, and add in brightness of green in the place of velocity variance. - Velocity
 
 The direction and velocity is displayed in contrast of red and blue variation.
 
 - CFA
 
 Angle dependence is small and color is displayed uniformly in blood vessels.
 
 - Directional CFA
 
 The direction is displayed red and blue and angle dependence is small and color is displayed uniformly in blood vessels.
 
 - Fine Flow
 
 It displays blood flow power with higher definition than CFA mode. It is possible to display microscopic vessels. (Include directional mode)
 
 - Velocity variance
 
 The velocity variance is displayed in brightness of yellow.
 
 - Color map
 
 CFI
 
 : 22maps
 
 CFA
 
 : 13maps
 
 Fin Flow : 9maps - Color invert
 
 Applicable (can be changed in real-time and frozen)
 
  Pulse repetition frequency (PRF)
 
 120, 250, 350, 400, 500, 650, 800, 1k, 1.3k, 1.5k, 2k, 2.5k, 3k, 3.5k, 4k, 4.5k, 5k, 6k, 8k, 10k, 12k, 16k, 20k [Hz], total 23 steps  Persistence
 
 8 steps
 
  Motion artifact suppression
 
 OFF + 3 steps
 
  Scan line density
 
 8 steps - 15 -
 
 SN-U391E
 
  Color preference
 
 Selectable 2 modes
 
 - Auto
 
 Brighter signal is displayed between B and color
 
 - Manual (16 steps)
 
 B image is displayed where B image intensity is bigger than preference bar level on the gray scale.
 
  Peak hold
 
 [Not ECG display] Overlay color data in real-time, Refresh time: 1s, 2s, 4s, 8s, ∞ [ECG display] Overlay color data in real-time, Refresh time: 1beat, 2beat, 4beat, 8beat, ∞
 
  Velocity tagging
 
 Applicable
 
  Color filter
 
 3 steps
 
  Interleave
 
 Max.32 lines (depending on PRF)
 
  Packet size
 
 3 steps
 
  CFA dynamic range
 
 8 steps (available to Fine flow)
 
  Reference frequency (REF)
 
 Max.3 frequencies (Depend on probes)
 
 3.1.2.21
 
 Compound function (HI Com function)
 
  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 B, dTHI, B/M, CFM-B, CFM-B/M, CFM-B/D, B/D, Elasto, Fine Flow
 
  Number of compound
 
 2 steps (Max. 5 images)
 
  Oblique angle
 
 Max. 5 directions (include 0 deg)
 
  Scan line density
 
 4 steps (Applicable independent setup)
 
  Steering HI Com
 
 Right and left (each 3steps) (can be changed to 3 step on linear array probes only.)
 
 3.1.2.22
 
 Trapezoid display function
 
  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 B, CFM,PW ,Fine Flow
 
  Display angle
 
 Max. ±25°(Depend on probes)
 
  HI Zoom
 
 Applicable
 
  Prohibit mode
 
 HI Com, B oblique, PW/CFM oblique
 
 3.1.2.23
 
 Oblique B mode image function
 
  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 B, dTHI, B/M, CFM-B, CFM-B/M, CFM-B/D, B/D mode
 
  Oblique angle
 
 Right and left (each 3 steps) Angle is depend on probes
 
  HI Zoom
 
 NA
 
 3.1.2.24
 
 Automatic Image Optimization (HI Support)
 
  Applicable probes
 
 All probes(*1)
 
  Display mode
 
 All mode (Except as M, CFI/CFA/Fine Flow, TDI-B and TDI-M mode) - 16 -
 
 SN-U391E
 
  Automatic optimization parameter
 
 - B mode
 
 Gain, TGC, Depth-specific PDC
 
 - PW mode
 
 Gain, PRF, Baseline
 
  Display
 
 - B mode(Gain,TGC)
 
 ON : Reverse display of BG
 
 - B mode(Depth-specific PDC)
 
 ON : Display “#” on PDC Screen button
 
 - PW mode
 
 ON : Reverse display of DG and PRF
 
  Reset
 
 Available B mode only
 
  Brightness offset
 
 21 steps (available B mode only)
 
 *1) HI Support for PDC is available at linear probes only.
 
 3.1.2.25
 
 TDI (Tissue Doppler Imaging) function
 
  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 TDI-B, TDI-M, TDI-PW
 
  TDI-B, TDI-M mode
 
 - Color map
 
 8 maps (for TDI)
 
 - Pulse repetition frequency (PRF) 120, 250, 350, 400, 500, 650, 800, 1k, 1.3k, 1.5k, 2k, 2.5k, 3k, 3.5k, 4k, 4.5k, 5k, 6k, 8k, 10k, 12k, 16k, 20k [Hz], total 23 steps - Persistence
 
 8 steps
 
 - Smoothing
 
 3 steps
 
 - Blending
 
 8 steps
 
 - Velocity Tagging
 
 Applicable
 
  TDI-PW mode
 
 - Analysis
 
 FFT method, 3 steps, Max. 128 points
 
 - Pulse repetition frequency (PRF) 400, 500, 650, 800, 1k, 1.3k, 1.5k, 1.8k, 2k, 2.2k, 2.5k, 3k, 3.5k, 4k, 4.5k, 5k, 5.5k, 6k, 6.5k, 7k, 7.5k, 8k, 9k, 10k, 11k, 12k, 14k, 16k, 18k, 20k [Hz], total 30 steps - Baseline shift
 
 Applicable (in real-time and frozen)
 
 - Detectable depth
 
 0mm ~360 mm
 
 - Sample length
 
 0.5 mm ~ 5.0 mm
 
 : 0.5 mm step,
 
 5.0 mm ~ 15.0 mm : 1.0 mm step - FFT-γ
 
 3.1.2.26
 
 Real-time Biplane mode
 
  Applicable probe
 
 3.1.2.27
 
 8 steps (can be changed in real-time and frozen)
 
 Convex type probe EUP-CC531S only
 
 Cine memory function
 
  Recording mode
 
 B, M/D (include CFM) - 17 -
 
 SN-U391E
 
  B-image Frame storage capacity
 
 Max. 19,000 frames  M/D recording time
 
 Max. 300s
 
  ODM recording time
 
 Max. 150s
 
  Cine memory split
 
 B/B split, B/MD split
 
  Playing images
 
 Continuous playback (B) Rewinding frame by frame (B, M/D) Automatically detected heartbeats (B) (Mounted a Biomedical Waveform Display Unit)
 
  Play-back speed
 
 B
 
 : ±1/1, ±1/2, ±1/4, ±1/8, 0
 
 M/D : Continuous playback is not applicable  Phase bars
 
 Applicable (B)
 
  Changing the time axis (Freeze on)
 
 Applicable (M/D)
 
 3.1.2.28
 
 Preset
 
  Application
 
 Max 22 sets of operating parameters can be registered for each probes Max 10 set of user registered
 
  PSS function
 
 Registered for each application Max 8 set of preset
 
  Region Data Setting
 
 Max 37 set of region data Max 30 set of user registered
 
 3.1.2.29
 
 Character display
 
  Comments
 
 102 characters×30 lines maximum.4 kinds of character size, 4 character style (Normal / Bold / Italic / Bold italic) , 3 entry mode (Insert / Letter Overwrite / Single Word Overwrite), Auto Complete Entry, 2 initial cursor position ( Home Position / Previous Position)
 
  ID name
 
 Regular 15 characters
 
  Patient name
 
 Regular 26 characters
 
  Hospital name
 
 Regular 29 characters
 
  Automatically displayed items
 
 Data and time (year, month, day, hour, minute, and second), Physician, age, sex, Probe name, frequency of the probes, gain settings, acoustic power level, post-processing parameters, dynamic range (DYN), delay times from R-wave, heart rate, measurement values, Doppler function display, focus point, view of depths display.  Annotation
 
 Preregistered 402 annotations (abbreviated medical terms)
 
 - 18 -
 
 SN-U391E
 
 3.1.2.30
 
 Body marks
 
  Display position
 
 Adjustable
 
  Display size
 
 Large or Middle
 
  Types
 
 - Abdomen
 
 8 types
 
 - Breast
 
 10 types
 
 - Head/Neck
 
 15 types
 
 - Urology
 
 10 types
 
 - Cardiovascular
 
 14 types
 
 - Organ
 
 11 types
 
 - Upper limb
 
 18 types
 
 - Lower limb
 
 10 types
 
 - Obstetrics
 
 21 types
 
 - Gynecology
 
 4 types
 
 3.1.2.31
 
 Measurement function
 
  General measurements
 
 Distance, area, circumference, volume, velocity, time,Doppler trace, Accel., Decel., velocity, average velocity, histogram, RI, PI, FlowVolume, Doppler profile, velocity ratio, time ratio, distance ratio, area ratio, heart rate, and CFM velocity,Angle
 
  Abdomen, urology,
 
 Hip joint angle, urology, SVolume, and Vascularity
 
 and others  Cardiovascular
 
 measurements
 
 Left ventricle measurements, LV, MV, Ao, TV, CO, EF, regurgitant
 
 volume,
 
 dP/dt, Qp/Qs, blood flow measurements, MVG, and ventricular mass
 
  Vascular measurements
 
 Carotid artery, upper extremities, lower extremities, and %Steno
 
  Obstetrics and gynecology
 
 Fetal measurements, blood flow measurements, fetal weight estimates,
 
 measurements
 
 gestational age estimates, fetal growth evaluations, ovarian follicle measurements, and gynecology measurements
 
  Real-time Doppler
 
 measurements
 
 Blood flow waveforms are traced in real time and the measurement results are displayed. The average heart rate can be calculated based on several measurements.
 
  Measurements of retrieved
 
 images  Obstetric tables
 
 Measurements can be performed only for images recorded using the image filing function. (Video playback measurements are not supported.) Data specific to the United States, Europe, and Japan are incorporated. You can create a new table or edit an existing table. Tables can be loaded to or from external media.
 
  Report function
 
 You can display a list of measurement results, edit results, enter comments, print reports, or display obstetrics graphs. You can record the measurement results to a hard disk or other supported media.
 
  User-defined functions
 
 You can register unique measurement packages by combining measurement - 19 -
 
 SN-U391E
 
 items, measurement orders, and formulas.
 
 3.1.2.32
 
 Media Management
 
  Folder or file list display
 
 Applicable
 
  Free space display
 
 Applicable
 
  Delete and copy of deleted image data
 
 Applicable ⁃ Copy destination
 
 3.1.2.33
 
 USB memory, USB HDD, DVD-RAM, Network folder
 
 Storage to hard disk
 
  Capacity
 
 Maximum 60GB
 
  Number of images
 
 B/W
 
 : approx. maximum 72,000
 
 Color
 
 : approx. maximum 24,000
 
 3.1.2.34
 
 Filing Function / Patient information window
 
  Image transfer
 
 Digital
 
  Transferring image
 
 Motion image / Still image
 
  Motion image transfer method
 
 - Frame to Frame (F-F)
 
 All image or 10, 20, 30, 40, 50, 60, 70, 80, 90, 100 images from last image
 
 - R-Wave to R-Wave (R-R) All image or 1, 2, 3, 4, 5, 6, 7, 8, 9, 10 beats from last image (Mounted a Biomedical Waveform Display Unit) - All (Notice) : When the number of images in cine memory exceeds the record upper limit, all image transfer is not available.  Image format
 
 BMP (Bitmap), TIFF, AVI, AVI (Compression), JPEG, DICOM, DICOM JPEG
 
  TIFF compression format
 
 None, Pack Bits
 
  AVI compress
 
 Microsoft Video 1
 
  Number of retrieve images
 
 1, 4, 9, 16
 
  Retrieve mode
 
 Frame by frame review Continuous review, Changing the time axis
 
  Retrieve speed
 
 -1/1, -1/2, -1/4, -1/8, 0, 1/8, 1/4, 1/2, 1/1
 
  Divided frame display
 
 Available with motion image review
 
  Review range specification
 
 Applicable
 
  Playback measurement
 
 Applicable excluding D-Profile, CFM-Velo, histogram, auto trace and PISA, MVG-M, vascularity in MV Regurg and TV Regurg
 
  Filing with
 
 Each image, Each patient
 
  Masking of ID and patient name
 
 Applicable (in transferring to external media) - 20 -
 
 SN-U391E
 
  Patient information window
 
 - Patient data list display Applicable - Patient ID assigned automatically (Emergency situation) Applicable - Pre-input patient information Applicable - Single Image recording Applicable - Auto complete setting
 
 Applicable
 
 - The number of the patients who can register 10,000 (It is assumed that for 1 year and half.)  Filing window
 
 - Keyword search
 
 Today's Study, Search All, Patient ID, Patient name, Study Date, Study Term, Body Part Exam., Examined by, Ref. Physician, Additional Patient History, Application, Comment, Transfer Status
 
 - Thumbnail images sorting Applicable - Image Data
 
 copy, delete applicable
 
 - Correction of the patient information for the saved examination data like an image. Not applicable (Notice) : Other patients’ examination data cannot be referred to during examination. (Notice) : To record data onto media that was created by using equipment that uses different software version, select the target location list in the Filing window, and then update the database before recording the data.  OS of the network computer which can perform file sharing.
 
 Windows XP Professional, Windows Vista Ultimate/Home Premium Windows 7 Ultimate/Professional/Home Premium Windows 8/Windows 8 Pro Windows 8.1/Windows 8.1 Pro (Notice) : The simultaneous reference of network folder cannot be performed from two or more system. (Notice) : File sharing is available when using a password-protected account.
 
 3.1.2.35
 
 Others
 
  Brightness Setting
 
 Character 6 level Background 6 level
 
  Screensaver
 
 Applicable
 
  Online Manual
 
 Applicable
 
  Screen Keyboard
 
 Enter alphanumeric characters and symbols (If the optional alphanumeric keyboard is not mounted)
 
 - 21 -
 
 SN-U391E
 
 3.2. Options 3.2.1. 17-inch Digital LCD Monitor (EZU-MT29-S1)  Standard component
 
 LCD monitor, LCD simple arm, cables
 
  Screen
 
 17 inch, Non-interlace
 
  Input voltage
 
 DC12V
 
  Power consumption
 
 Max.34VA
 
  Video signal
 
 DVI-D
 
 3.2.2. 17-inch Digital LCD Monitor (Flexible arm) (EZU-MT30-S1)  Standard component
 
 LCD monitor, LCD flexible arm, cables
 
  Screen
 
 17 inch, Non-interlace
 
  Input voltage
 
 DC12V
 
  Power consumption
 
 Max.34VA
 
  Video signal
 
 DVI-D
 
 3.2.3. Biomedical Waveform Display Unit (EZU-EK28)  Components
 
 BIO PCB, ECG panel
 
  Display signal
 
 ECG, Ext-ECG
 
  Display method
 
 - Real-time
 
 Moving bar
 
 - Cine playback
 
 Scroll
 
  Display control
 
 Vertical movement, gain control, auto gain: 3 steps
 
  Sweep speed
 
 1, 1.5, 2, 3, 4, 6, 8, 10 s/screen (can be changed in real-time and frozen)
 
  R-R Detection
 
 RRp/RRpp≒1 interval is detected automatically in freeze mode. (RRp: preceding R-R interval, RRpp: prepreceding R-R interval)
 
 (Notice) Biomedical waveform cannot be displayed, when an application that image split setting of B-mode is up-and-down is selected.
 
 3.2.4. CW Doppler Unit (EZU-ST8)  Components
 
 CW PCB, total 1pcbs
 
  Applicable probes
 
 EUP-S50A, S70, S52, S72, ES52E, ES52M, Single Element CW Doppler probe
 
  Display mode
 
 CW, B/CW (B, B-freeze), CFM-B/CW (CFM-B, B-freeze)
 
  Reference frequency (REF)
 
 - EUP-S50A
 
 2.0 [MHz]
 
 - EUP-S70
 
 2.0 [MHz]
 
 - EUP-S52
 
 3.5 [MHz]
 
 - EUP-S72
 
 3.0 [MHz]
 
 - EUP-ES52E
 
 3.5 [MHz] - 22 -
 
 SN-U391E
 
 - EUP-ES52M
 
 3.5 [MHz]
 
  FFT pulse repetition frequency (PRF)
 
 4, 4.5, 5, 6, 8, 10, 12, 16, 20, 25, 35, 50 [kHz], total 12 steps  Analysis
 
 FFT method, 3Steps, Max. 256point
 
  Maximum detectable velocity
 
 8.91m/s (2.0 [MHz], 50 [kHz])  Sweep speed
 
 1, 1.5, 2, 3, 4, 6, 8, 10 s/screen (can be changed in real-time and frozen)
 
  Wall filter
 
 12 steps (maximum value : 1/8 of PRF)
 
  Transmission-reception focus
 
 8 steps (automatic follow-up the sampling gate)  Baseline shift
 
 Applicable (in real-time and frozen)
 
  Angle correction
 
 Applicable (in real-time and frozen)
 
  Post-processing
 
 γ-correction, rejection and saturation (can be changed in real-time and frozen)
 
  D-Color
 
 29maps (can be changed in real-time and frozen)
 
  Echo Enhancement
 
 4 steps, including OFF
 
 3.2.5. Real-time Tissue Elastography Unit (EZU-TE6)  Components
 
 Stabilizers for compression assistance
 
  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 - Elasto mode
 
 Overlapping display of Elastography on B-mode image
 
 - Dual Elasto mode
 
 Dual image display of Elasto mode, possible to select different color map on each side, also possible to display B-mode image on the right and Elasto image on the left side
 
 - Real-time biplane Elasto mode
 
 Overlapping display of Elastography on B-mode image of either cross-section, in the function to display B-mode image with the one of different cross-section simultaneously
 
 Specifications  Reference frequency (REF)
 
 2 frequencies (L65, L54MA, L75 and O54J: 1 frequency)
 
  Persistence
 
 8 steps
 
  Smoothing
 
 4 steps
 
  Dynamic range
 
 8 steps
 
  Frame rejection
 
 8 steps
 
  Noise rejection
 
 8 steps
 
  Scan line density
 
 4 steps
 
  Brightness Control
 
 11steps
 
  Color map
 
 8 maps(red color and blue color can be inverted)
 
  Tagging
 
 Max. 65 steps in both upper and lower limit values - 23 -
 
 SN-U391E
 
  Rejection
 
 65 steps
 
  Blending
 
 0% ~ 100%, 2% steps
 
  Blend Type
 
 2 types
 
  Frame rate
 
 5steps
 
  Dual mode
 
 Applicable
 
  Biplane mode
 
 Applicable (EUP-CC531S)
 
  Biopsy guide line
 
 Applicable
 
  Press Guide function
 
 - Strain Graph
 
 Displays the variation over time of tissue strain caused by compression. Available graph to exclude low strain area.
 
 - Press Indicator
 
 Display the compression status
 
  Frame select function
 
 Possible to select images under stable compression automatically
 
  Strain ratio measurement
 
 - Number of ROI
 
 Max. 10, possible to position the ROIs on B mode image
 
 - ROI shape
 
 Circle, oval, rectangle and closed spline curve
 
 - Averaging function
 
 Max. 5 frames
 
 - Auto ROI Setting function Clicking center of the tumor automatically sets the measurement ROI to the tumor and the fatty layer (the mammary gland region and lesion are targeted). - Measurements of retrieved images
 
 3.2.6. Real-time Tissue (EZU-TESH1)
 
 Applicable
 
 Elastography
 
 software
 
  Components
 
 Software
 
  Applicable Probes
 
 Same probe as Real-time Tissue Elastography
 
 (Strain
 
 Histogram)
 
 Specifications 
 
 Strain Histogram Measurement - Strain Graph
 
 1 point
 
 - ROI Shape
 
 rectangle
 
 - Measurements of retrieved images 
 
 Data export
 
 Applicable
 
 Applicable (CSV format)
 
 (Notice) : EZU-TESH1 is required for EZU-TE6. (Notice) : If B-mode images are displayed at the top and bottom part of the screen in dual display, Real Time Tissue Elastography software (Strain Histogram) cannot be used.
 
 - 24 -
 
 SN-U391E
 
 3.2.7. Digital Video Archiving Software (EZU-DV4)  Components
 
 3.2.7.1 
 
 Software
 
 Clip capture mode
 
 Common specification - Preset
 
 8 steps
 
 - Compression level
 
 2 steps (Standard/HI Compression)
 
 - Image format
 
 AVI (uncompressed, Motion JPEG Compression), DICOM, DICOM JPEG
 
 - Playback speed ・ Play
 
 1x
 
 ・ Rewind
 
 -1x
 
 ・ Slow-forward
 
 1/8x, 1/4x, 1/2x
 
 ・ Slow-rewind
 
 -1/8x, -1/4x, -1/2x
 
  Specifications – in real-time
 
 - Transfer mode
 
 Free capture, Beat Duration
 
 - Capture rate
 
 Free capture
 
 : 6, 8, 10, 15, 20, 30 fps 6steps
 
 Beat Duration : 30 fps - Capture time
 
 Free capture
 
 : 10 sec (30 fps), 90 sec (20 fps), 120 sec (15 fps), 180 sec (10 fps), 225 sec (8 fps), 300 sec (6 fps)
 
 Beat Duration : 1, 2, 3, 4, 5, 6, 7, 8, 9 and 10 heartbeat - Measurements on clip capture Applicable However, there are following limitations for a trimmed clip. ・ Necessary to set all image parameter before measurement ・ Impossible to perform the following functions %WT measurement, Color Tissue Tracking measurement, LA Tracking measurement, Time Intensity Curve function  Specifications – in frozen
 
 - Transfer mode
 
 Frame to Frame (F-F) / R-Wave to R-Wave (R-R) / All
 
 - Capture time
 
 Free capture
 
 :
 
 AVI(Motion JPEG Compressiona) or DICOM JPEG format 10 sec (30 fps), 90 sec (20 fps), 120 sec (15 fps),180 sec (10 fps), 225 sec (8 fps), 300 sec (6 fps) AVI(uncompressed) or DICOM format 10 sec (30 fps), 18 sec (20 fps), 24 sec (15 fps), 36 sec (10 fps), 45 sec (8 fps), 60 sec (6 fps) R-Wave to R-Wave : ALL, 1, 2, 3, 4, 5, 6, 7, 8, 9 and 10 heartbeat/manual (Notice) : EZU-DCP3 is required for EZU-DV4. (Notice) : EZU-EK28 is required for R-wave to R-wave mode (Notice) : EZU-FC11 is required for selecting DICOM JPEG as the image format. - 25 -
 
 SN-U391E
 
 (Notice) : When the number of images in cine memory exceeds the record upper limit, all image transfer is not available. (Notice) : Quality of AVI(uncompressed) or DICOM format and AVI(Motion JPEG Compression) or DICOM JPEG format are same in real-time.
 
 3.2.7.2
 
 Long time recording mode
 
  Recording format
 
 MPEG2 (can be played on image filing function)
 
  Destination
 
 DVD-RAM / Hard Disk / USB HDD
 
  Continuous recordable time
 
 Maximum 1 examination
 
  Recordable time to DVD
 
 - High-quality
 
 Approx. 1 hour (8Mbps)
 
 - Standard
 
 Approx 2 hours (4Mbps)
 
 - High-compression
 
 Approx 4 hours (2Mbps)
 
  Recording area
 
 Fundamental area
 
  Record and play of Doppler sound
 
 Applicable  Playback speed
 
 - Play
 
 1×
 
 - Fast-forward
 
 2×, 8×, 32×
 
 - Fast-rewind
 
 -1×, -2×, -8×, -32×
 
 - Slow-forward
 
 1/8×, 1/4×, 1/2×
 
 - Slow-rewind
 
 -1/8×, -1/4×, -1/2×
 
  Playback to the start by chapter
 
 Applicable  Playback function
 
 Only MPEG2 clip file that is captured this software
 
 (Notice) : EZU-DCP3 is required for EZU-DV4.
 
 3.2.8. Logon Management Software (EZU-LG2)  Components
 
 software
 
  Registerable account number
 
 - Service
 
 1 (undeletable)
 
 - Administrator
 
 max. 30
 
 - General user
 
 max. 30
 
  User licensing function
 
 - User ID
 
 Alphanumeric half size of character, symbols (except space), 1~16 characters
 
 - Password
 
 Alphanumeric half size of character, 8~16 characters, (Not applicable to input fewer than 8 characters and same letter string of user ID)
 
 - Password changing
 
 Applicable
 
  Account management (permitted to administrator only) - 26 -
 
 SN-U391E
 
 - Display of information
 
 Displaying a list of account
 
 - Registration function
 
 New registration of account
 
 - Delete function
 
 Deregistration of account
 
  Logoff function
 
 - Authentication release
 
 Applicable
 
 - Auto logoff
 
 Applicable
 
 3.2.9. DICOM Software(Transfer and Media Storage) (EZU-FC11)  Components
 
 Software
 
  Functions
 
 - Transfer
 
 Transfer to DICOM server
 
 - Media transfer
 
 Save to the DVD-RAM, DVD-R, USB memory or USB HDD as DICOM format
 
 - Connection check
 
 C-ECHO function (carries out a connection check with server)
 
 - Storage commitment function The function to query whether to be preserved normally after transferring for ultrasound images - DICOM SR
 
 The function to transfer the measurement results to report server as DICOM SR (Structured report) A measurement item name is modifiable. It is possible to copy a measurement which does not support DICOM SR for the measurement using DICOM SR. It is possible to add a measurement which does not support DICOM SR for the measurement region using DICOM SR.
 
  Transfer image
 
 Single images, motion images (It is applicable the collectively transferring by patient)
 
  Brightness adjustment
 
 The brightness of images can be adjusted when transferring them to the storage server. The way of setting the corrected information to a DICOM attribute, and the way of changing the brightness of the image pixel value to selection is possible.
 
  Anonymize patient info.
 
 The patient information printed in image can be masked when transferring them to the storage server.
 
  Motion images transfer
 
 Frame to frame transfer, R-R transfer [ECG is used], all images transfer in cine memory
 
  Retrieve
 
 DICOM format images saved in the folder on network, DVD-RAM, DVD-R,USB memory and USB HDD It is applicable to play after DICOM server transfer by EZU-FC11Q.
 
  Image conversion
 
 BMP to DICOM, DICOM JPEG AVI to DICOM, DICOM JPEG
 
  LAN
 
 10/100BaseT - 27 -
 
 SN-U391E
 
  DICOM log acquisition
 
 Applicable (DICOM log in communication)
 
  SR transfer operation
 
 DICOM SR is transferred by pushing "SR Transfer" button in the Filing Window, "Save" button in the Report Winodw, Patient Key or End Exam key.
 
  Output measurement item to DICOM SR
 
 - OB-GYN measurement (OB_SR) - OB measurement GS, CRL, BPD, OFD, HC, LVW, HW, TCD, IOD, OD, APTD, TTD, AC, FTA, HL, ULNA, RAD, FL, TIB, FIB, EFBW, CI, HC/AC, FL/AC, FL/BPD, APTD*TAD AFI measurement: Q1, Q2, Q3, Q4, AFI - Gynecology measurement Uterus measurement: Uterus L, Uterus H, Uterus W, Endomet. T., Cervix L L-Ovary measurement: L, AP, W, V R-Ovary measurement: L,AP,W,V - Follicular measurement L-Ovary measurement: L, AP, W, V R-Ovary measurement: L, AP, W, V L-Follicular measurement: L, AP, W, V R-Follicular measurement: L, AP, W, V Uterus measurement: Uterus L, Uterus AP, Uterus W, V, Endomet. T.,Cervix L - OB-GYN measurement (OB_SR_II) - Vascular measurment Fetal_Vasc measurment: Umbil.A. measurment, MCA measurment, OB-DV measurment, OB-IVC measurment, OB-Daorta measurment, OB-Aorta measurment, OB-PA measurment, CIA measurment, L-RA measurment, R-RA measurment, Umbil.V. measurment Pelvic_Vasc measurment: L-Ovary.A. measurment, R-Ovary.A. measurment, L-Uterine A. measurment, R-Uterine A. measurment - OB measurement GS, CRL, BPD, OFD, HC, LVW, HW, TCD, IOD, OD, APTD, TTD, AC, FTA, HL, ULNA, RAD, FL, TIB, FIB, EFBW, CI, HC/AC, FL/AC, FL/BPD, APTD*TAD, Foot, NB, NT AFI measurement: Q1, Q2, Q3, Q4, AFI - Gynecology measurement Uterus measurement: Uterus L, Uterus H, Uterus W, Endomet. T., Cervix L - 28 -
 
 SN-U391E
 
 L-Ovary measurement: L, AP, W, V R-Ovary measurement: L,AP,W,V - Follicular measurement L-Ovary measurement: L, AP, W, V R-Ovary measurement: L, AP, W, V L-Follicular measurement :L,AP,W,V R-Follicular measurement :L,AP,W,V Uterus measurement: Uterus L, Uterus AP, Uterus W, V, Endomet. T., Cervix L - Vascular measurement
 
 Extracranial Arteries measurement: Bulb, CCA, ECA, ICA, PROX, ICA MID, ICA DIST, ICA(RATIO), Subclav.A, Vertebral A Lower Extremity Arteries measurement: CIA, ATA, CFA, DPA, EIA, IIA, Pero.A, Pop.A, PTA, PFA, SFA Lower Extremity Veins measurement: ATV, CFV, CIV, EIV, GV, GSV, LSV, Pero.V, Pop.V, PTV, PFV, SFJ, SV, SFV, SPJ, Hunter, Cocket, Boyd
 
 - Cardiac measurement (Cardiac_SR) Left Ventricle measurement: HR, LVEdD, LVEsD, FS, LVEdV, LVEsV, EF, Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI, IVRT, e', E/e', A, S, SV, CO, CI, IVSd, LVPWs, LVPWd, LVMass-d, IVSd/LVPWd Right Ventricle measurement: HR, Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI, SV, CO, CI Aortic Valve measurement: Avcusp Open, ET, Ao/ET, Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI Mitral Valve measurement: A, E, E/A, EFslope, EPSS, Adur, Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI, SV, CO, CI, dp/dt Pulmonic Valve measurement: Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI Tricuspid Valve measurement: Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI Pulmonary Veins measurement: S, D, A, Adur, HR Congenital measurement: HR, Regurg Vol., Flow Rate, PISA, PISA-R, Vp, TVI, Qp/Os - Cardiac measurement (Cardiac_SR_II) Echo LV measurement : B-LV, M-LV, M.Simpson, Dia.Func, Dop Index, CO, BI-Plane, Area Length, LVMass-AL, LVMass-TE, HR Echo LA measurement : B-LA, M-LA, TEE-LA Echo RV measurement : CO, B-RV, M-RV, HR Echo RA measurement : B-RA Echo AV measurement : AV Regurg, Velocity, PHT, D-Trace, - 29 -
 
 SN-U391E
 
 B-Area, Area-Vp, Area-Vm, Area-VTI, HR Echo MV measurement : Dia.Func, dP/dt, MV Regurg, Velocity, PHT, D-Trace, B-Area, Area-Vp, Area-Vm, Area-VTI, MV-RegVol, HR Echo PV measurement : PV Regurg, Velocity, PHT, D-Trace, HR Echo TV measurement : TV Regurg, Velocity, PHT, D-Trace, HR Echo Ao measurement : B Ao, Ao, HR Echo PArtery measurement : B-PArtery, D-PArtery, HR Echo IVC measurement : B-IVC, HR Echo PVein measurement : D-PVein, HR Echo CS measurement : Qp/Qs, HR Echo Cong measurement : Cong Regurg, HR (Notice) : EZU-NW1 is required for EZU-FC11.
 
 - 30 -
 
 SN-U391E
 
 3.2.10. DICOM Software (WorkList) (EZU-FC11W) / DICOM Software (Printer) (EZU-FC11P)  Components
 
 Software
 
  Functions
 
 - Printer function
 
 Printout to a DICOM printer on network
 
 - WorkList
 
 Send/receive a patient data to/from a DICOM WorkList server
 
  Specifications for printer
 
 - Print format
 
 B/W: MONOCHROME2, color: RGB
 
 - Images in a film
 
 Max. 16 images
 
 - Start of printing
 
 Each image : Press REC key Each patient : Press End Exam key or click Patient key, Click Image Print button on the Filing Dialog
 
 - Film format setup
 
 Film size, image display format, film orientation, film destination
 
 - Other setup
 
 Number of copies, magnification type, print priority
 
 - Simultaneous transfer
 
 Simultaneous transfer to DICOM server and printer
 
  Specifications for WorkList
 
 Acquisition of patient data Retrieval of patient data Display of detailed patient data The patient information which is indicating by the list using the mobile function can be held, and use is possible at off-line Setup is freely possible in the tag which quer(ies) to a work list server It is possible automatic calculation of the age based on the birth date obtained from Worklist Possible to query automatically when the Patient Information window is displayed.
 
 (Notice) : EZU-NW1 and EZU-FC11 are required for EZU-FC11P and EZU-FC11W.
 
 3.2.11. DICOM Software (Query / Retrieve) (EZU-FC11Q)  Components
 
 Software
 
  Specifications
 
 - Retrieve modality
 
 US, CT, MR, MG, EC
 
 - Query key
 
 ID, Name, Date, Accession #, Modality
 
 - Multi-display
 
 1, 2, 4, 6 display
 
 - Save format
 
 BMP, JPEG
 
 - Display format
 
 Original size, window size
 
 - Image size
 
 Zoom in, zoom out, panning
 
 - Window level
 
 Window level, ROI window level, B/W inversion
 
 - Rotation, inversion
 
 90 deg rotation to right and left, 180 deg rotation, flip horizontal, flip vertical
 
 - Review
 
 By frame, loop, synchronization - 31 -
 
 SN-U391E
 
 - Measurement
 
 Caliper, distance ratio, angle, area, average velocity, acceleration, heart rate (Notice: If there is no tag for calibrations, average velocity, acceleration and heart rate cannot be measured.)
 
 - Distance calibration
 
 Applicable in caliper and area
 
 - Annotation
 
 Applicable (text and rectangle on display)
 
 - Hard disk storage
 
 Images acquired from DICOM server can be preserved up to 3GB.
 
 (Notice) : EZU-NW1 and EZU-FC11 are required for EZU-FC11Q.
 
 3.2.12. Omni-directional M mode display function (EZU-OD3)  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 - Real-time ODM mode
 
 B (Real) / ODM, B (Real) / ODM / ODM, ODM,ODM / ODM
 
 - Cine ODM mode
 
 B (Cine) / ODM, B (Cine) / ODM / ODM
 
  Settings
 
 - Number of ODM beam line 1 direction / 2 directions - ODM beam line center control Common center mode / Independent center mode - Angle rotation
 
 0 ~ 359 deg, 1 deg steps
 
 - Center position movement 1 pixel steps - Sweep speed
 
 1, 1.5, 2, 3, 4, 6, 8, 10 s/screen (can be changed in real-time and frozen)
 
 - ODM dynamic range
 
 45~ 90 dB, 5 dB steps (can be changed in real-time and frozen)
 
 - ODM enhancement
 
 4 steps
 
 - Gray map
 
 8 steps (can be changed in real-time and frozen)
 
 - B image scan density
 
 Fixes
 
 - B image display angle
 
 Selectable (available for Real-time ODM mode)
 
  Display method
 
 - Real-time
 
 Moving bar
 
 - Cine
 
 Moving bar
 
  Measurement
 
 M mode measurement
 
  Others
 
 Not applicable with Doppler and CFM mode
 
 3.2.13. Advanced Measurement Software (EZU-AM7)  Components
 
 Software
 
  Specifications - %WT (wall thickness) measurement
 
 - Image format
 
 AVI, Motion JPEG (B/W B mode image)
 
 - Analysis point
 
 Max. 20 points
 
 - WT graph display
 
 Applicable - 32 -
 
 SN-U391E
 
 - Data export
 
 Applicable (CSV format)
 
  Specifications - IMT (Intima-Media Thickness) measurement
 
 - Image format
 
 Still image (Frozen, playback image)
 
 - Measurement results 
 
 Trace method
 
 Max, mean, value of 3 points, average value of 3 points
 
 
 
 Simple 3 points method Value of 3 points, average value of 3 points
 
 - Manual correction of trace line Applicable (in trace method) - Detection position correction Applicable (in simple points method)  Specifications - Color tissue tracking measurement
 
 - Image format
 
 AVI, Motion JPEG (B/W B mode image)
 
 - Specified measurement segments APICAL, MID, BASAL - Graphs for transition
 
 Applicable
 
 - Data export
 
 Applicable (CSV format)
 
 - Creating Bull’s eye
 
 Applicable
 
  Specifications - LA tracking measurement
 
 - Image format
 
 AVI, Motion JPEG (B/W B mode image)
 
 - Measurement results
 
 Volume, dV/dt, area, circulation, long axis
 
 - Volume measurement
 
 Modified Simpson method
 
 - Biplane measurement
 
 Applicable
 
 - Graphs for transition
 
 Applicable
 
 - Data export
 
 Applicable (CSV format)
 
 3.2.14. IMT Measurement Software (EZU-IM2) This is the same function as IMT measurement in EZU-AM7. Please refer to EZU-AM7 for details.
 
 3.2.15. Software Specified for Contrast Agents (EZU-CH8)  Components
 
 Software
 
  Applicable probes
 
 Refer to Appendix
 
  Display mode
 
 WPI (Wideband pulse inversion) mode CWPI (Color wideband pulse inversion) mode
 
  Intermittent mode
 
 Monitor mode (possible to shot manually) Multi-shot mode (2~16 shot, total 15 steps)
 
  Trigger timing
 
 - System clock trigger timing (ECG display off) 
 
 Abdominal
 
 32 steps 0.1s ~ 1.0s: 0.1s step - 33 -
 
 SN-U391E
 
 1.5s ~ 10.0s: 0.5s step 11.0s ~ 14.0s: 1s step 
 
 Cardiac
 
 31 steps 0.2s ~ 1.0s: 0.1s step 1.5s ~ 10.0s: 0.5s step 11.0s ~ 14.0s: 1s step
 
 - ECG trigger timing (ECG display on) 32 steps 1 ~ 32 beats, 1 beat step  Acoustic pressure
 
 %Control and MI Control are selectable MI Control : 0~0.5 0.5~
 
 0.01 step 0.05 step
 
 (Max MI value: Depend on probes and focus setting and so on) %Control : 0~100% 34 steps Automatic acoustic pressure control on fundamental image in monitor mode and alternate mode is possible.  Reference frequency (REF)
 
 2 frequencies
 
  Stopwatch function
 
 -30s ~ 59m59s, for monitoring the time after injection of contrast agent For monitoring the time of image recorded in cine memory.
 
  Application function
 
 Possible to independently set up each application parameter for contrast imaging
 
  Multi-step trigger (MST) mode
 
 Possible to make a desirable protocol for contrast imaging by combining freely an intermittent transmission interval and execute it. - Number of registrable protocol Max. 8 protocols - Number of Steps for each protocol Max. 8 steps - Items for registration
 
 Start time, Interval time, Image numbers
 
 - Data Reading, Saving
 
 Applicable
 
  Flash function
 
 Breaking contrast agent for observing replenishment, Number of flash: 1, 2, 4, 8, 16, 32 frames, total 6 steps
 
  Intermittent flash mode
 
 - System clock trigger timing (ECG display off) 5, 10, 15, 20, 25, 30, 35, 40s, total 8 steps - ECG trigger timing (ECG display on) 32 steps 1 ~ 32 beats, 1 beat step, total 32 steps  UCI (uniform contrast imaging) function
 
 This function uses the technology of a different aspheric focusing from the conventional focal technique, it can generate uniform acoustic field in the body - 34 -
 
 SN-U391E
 
  Alternate mode
 
 In dual screen displays in alternate mode, contrast image (left screen) and fundamental image (right screen) are transmitted and displayed alternately, ultrasound contrast agents and anatomy can be observed on real time (applicable in middle and low acoustic pressure transmission mode)
 
  MTI (microbubble trace imaging) mode
 
 To improve the visibility of running of the contrast agents, the maximum intensity of each frame of the contrast image is overlaid (applicable in middle and low acoustic pressure transmission mode)  Color wideband pulse inversion mode
 
 - Selectable Color overlay mode / Color stand-alone mode - Color map
 
 8 maps
 
 - Smoothing
 
 3 steps
 
 - Dynamic range
 
 45~80 dB, 5 dB steps (can be changed in real-time and frozen)
 
 - Gain
 
 60dB (adjustable in real-time and frozen)
 
 - Scan line density
 
 4 steps
 
 - Receiving filter
 
 3 steps
 
 - Frame rate
 
 3 steps
 
 - Display mode
 
 Selectable triplet ON/OFF (in high acoustic pressure transmission mode)
 
  Time intensity curve function
 
 - Motion image format
 
 AVI, Motion JPEG
 
 - Applicable mode
 
 B, B/B
 
 - Number of ROIs
 
 Max. 12
 
 - Shapes of ROI
 
 Circle, ellipse, square, rectangle, banana, freehand
 
 - Operation of ROI
 
 Copy, rotate, delete, interpolation
 
 - Display magnification
 
 ×1, ×2, ×4
 
 - Review of motion image Review (forward), frame review (forward / backward), review speed (5 steps for each direction), move frame (start frame / end frame) - Analysis result
 
 Display of ROI, graphs and data
 
 - Graph display
 
 Averaged intensity, fitting curve (2 types)
 
 - Data export
 
 Applicable (CSV format)
 
 (Notice) :Up and Down display of B mode can not be selected, when an applicable application of EZU-CH8 is selected.
 
 3.2.16. WideView Software (EZU-WS8)  Components
 
 Software
 
  Applicable probes
 
 Refer to Appendix
 
  Operating condition
 
 B / CFM-B single display
 
  Display mode
 
 WideView display, cine playback display
 
  Display functions
 
 Zoom, Panning, Rotation - 35 -
 
 SN-U391E
 
  Measurement function
 
 B mode measurement (except for histogram and CFM VELO)
 
 (Notice) : EZU-DCP3 is required for EZU-WS8. (Notice) : Capture operation by foot switch can be done by connecting EZU-FS2.
 
 3.2.17. Stress Echo Software (EZU-SE6)  Components
 
 Software
 
  Specifications
 
 - Applicable mode
 
 B mode (single image display)
 
 - Database
 
 Patient Information, Stress examination information (Images, WMS report)
 
 - Protocol
 
 Conformity with ASE (American Society of Echocardiography), user programmable
 
 - Capture functions
 
 Max. 12 stages, Max. 10 loop/time, selectable capture area, R-R capture
 
 - Playback functions
 
 Multiple-display simultaneous playback (automatic synchronizer), standard playback, frame by frame playback, reverse display, display of selected images
 
 - Full disclosure mode
 
 User can continuously acquire loops for a certain time (mostly 2 minutes) Later, in select mode, the user can select the appropriate views
 
 - Gain and ROI save function Gain setting and the ROI position of the previous stage are saved and applied for the next stage - Image input signal
 
 Digital YUV
 
 - Measurement function
 
 LV volume measurement
 
 - Report function
 
 WMS report, LV volume measurement report, user comment
 
 - Print
 
 B/W, color
 
 - Protocol setting
 
 By protocol editor
 
 - Image data export
 
 Applicable (BMP, AVI)
 
 - Setting for destination of the image to save By archive tool (Notice) : EZU-DCP3 and EZU-EK28 are required for EZU-SE6. (Notice):
 
 Not available to EZU-SE6 when selecting U/D the Image Split.
 
 (Notice) : Capture operation by foot switch can be done by connecting EZU-FS2.
 
 3.2.18. Three-dimensional display unit (Real-Time) (EZU-4D7)  Components
 
 Software (Real time 3D software), MTDRV PCB, MTBP PCB
 
  Applicable probe
 
 Refer to Appendix
 
  Display mode
 
 - ROI setting mode
 
 Displays the ROI that specifies the 3D image creation area on the 2D image
 
 - 3D mode
 
 Real-time 3D image is displayed on 1 screen. - 36 -
 
 SN-U391E
 
 - 2D-3D mode
 
 Screen is vertically divided into two sides. The ROI setting mode is displayed on the left side and the 3D mode on the right side
 
 - MPR 4 screen mode
 
 Screen is vertically and horizontally divided into four. 3D image is displayed on the lower right, while the random orthogonal 3 cross-section surface images are displayed on the other sections
 
 - MPR 1 screen mode
 
 One of the orthogonal 3 cross-section surface image in MPR 4 screen mode is displayed on 1 screen
 
 - Multi Slice multi screen mode Screen is divides into multi (2, 4, 9, 16) sections. A reference image is
 
 displayed on the upper left section, and the slice images are displayed on the other sections. - Multi Slice 1 screen mode
 
 The selected slice image from Multi Slice multi screen mode is displayed on 1 screen.  HI Definition 3D
 
 By one swing scanning at a speed that is slower than real time 3D, and high-definition 3D image or MPR image is displayed.
 
 Applicable to CFM mode (CFI, CFA and Fine Flow except for color maps for displays of velocity variance)  Rendering mode
 
 B/W - Opacity - MIP (Maximum Intensity Projection) - mIP (minimum Intensity Projection) - Transparency - Inversion - Orthogonal - Opac. Advanced - 4Dshading Color - Opacity - MIP (Maximum Intensity Projection) - Transparency - Orthogonal  Quality settings
 
 - Image quality
 
 Real-time 3D
 
 6 steps
 
 HI Definition 3D 3 steps B/W - Brightness 
 
 3D image
 
 17 steps (-8 ~ 8, except rendering mode: Orthogonal)
 
 
 
 MPR image
 
 17 steps (-8 ~ 8) - 37 -
 
 SN-U391E
 
 - Color 
 
 3D color
 
 37 types (except rendering mode: Orthogonal, Opac.Advanced, and 4Dshading)
 
 
 
 MPR color
 
 37 types
 
 
 
 Depth color
 
 14 types (rendering mode: Opac.Advanced)
 
 - Shading
 
 6 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
 
 - Filter
 
 20 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced,, or 4Dshading)
 
 - Threshold
 
 256 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced, or 4Dshading)
 
 - Smoothing
 
 8 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
 
 - P-Filter
 
 8 steps (rendering mode: MIP or mIP)
 
 - BW 3D Smooth.
 
 3 types (rendering mode: 4Dshading)
 
 - 3D Smooth Coef
 
 11 types (rendering mode: Opac.Advanced, or
 
 - Depth Contrast
 
 40 types (rendering mode: Opac.Advanced)
 
 - Depth Color Shift
 
 127 types (rendering mode: Opac.Advanced)
 
 - Region Map
 
 9 types (rendering mode: 4Dshading)
 
 - Absorption
 
 10 types (rendering mode: 4Dshading)
 
 - Hue
 
 4 types (rendering mode: 4Dshading)
 
 - Shadow Smooth
 
 10 types (rendering mode: 4Dshading)
 
 - Light Intensity
 
 5 types (rendering mode: 4Dshading)
 
 - Lighting Reso.
 
 4 types (rendering mode: 4Dshading)
 
 - Transp. Level
 
 8 types (rendering mode: 4Dshading)
 
 4Dshading)
 
 Color - Combination
 
 B/W and CFM image combination / only B/W image / only CFM image (only CFM image is impossible when rendering mode is Orthogonal)
 
 - Blend
 
 100 steps (except rendering mode: Orthogonal)
 
 - Brightness 
 
 3D image
 
 17 steps (-8 ~ 8, except rendering mode: Orthogonal)
 
 
 
 MPR image
 
 17 steps (-8 ~ 8) (except CFI mode)
 
 - Shading
 
 6 steps (rendering mode: opacity except CFI mode)
 
 - Filter
 
 5 steps (rendering mode: only opacity)
 
 - Threshold
 
 256 steps (rendering mode: only opacity )
 
 - Smoothing
 
 8 steps (rendering mode: only opacity)
 
 - P-Filter
 
 8 steps (rendering mode: only MIP)
 
  Operations
 
 - Rotation
 
 By trackball and encoder
 
 - View Line
 
 The cut-line of arbitrary curvature can be specified on 2D image
 
 - Sliding, Plane cut
 
 Max. 14 planes by trackball (only in frozen, except rendering mode: Orthogonal) - 38 -
 
 SN-U391E
 
 Image operations: both B/W and CFM images / only B/W image - Cut function
 
 The arbitrarily specified part of 3D image can be dug (only in frozen, except rendering mode: Orthogonal) Image operations: both B/W and CFM images / only B/W image
 
 - Zoom
 
 1.0~4.0 (0.25 steps)
 
 - Z Adjust
 
 10 steps
 
 - Image filing
 
 3D volume data, Real-time 3D mode single images (such as BMP) and multi images (such as AVI) can be saved Saved images can be played back from the image filing screen, as with regular 2D images Image pre-cutted by ViewLine and full image can be switched
 
 - Cine memory
 
 3D images can be saved in the cine memory and played back
 
 - PSS function
 
 Using the 3D PSS (3D Preset Scanning Selector) function
 
 - Intelligent Surface Tracking Set ViewLine automatically - Volume Slice Imaging
 
 Display a thick cross-sectional image From 0mm (OFF) to 20mm. 24 steps.
 
  Measurement
 
 Distance, area, circumference, volume etc. can be measured on 2D image, MPR image or Multi Slice image.
 
  User-defined screen button
 
 The items of 3D screen button menu and image adjustment screen button menu are registerable
 
 3.2.19. Three-dimensional Display Software (STIC) (EZU-4DST2)  Components
 
 Software (STIC software)
 
  Applicable probe
 
 Refer to Appendix
 
  Display mode
 
 Same as EZU-4D7 (except 2D-3D mode)
 
  Rendering mode
 
 Same as EZU-4D7
 
  Quality settings
 
 - Scan time
 
 4 steps (7.5, 10, 12.5, 15s)
 
 - Color MAP
 
 Variance maps are available.
 
 Same as EZU-4D7 excluding the above-mentioned  Operations
 
 - Display 3D
 
 Selectable display or hide 3D images
 
 Same as EZU-4D7 excluding the above-mentioned  Measurement
 
 Same as EZU-4D7
 
  User-defined screen button
 
 The items of image adjustment screen button menu and mode-specific screen button menu are registerable
 
 (Notice) : EZU-4D7 is required for EZU-4DST2.
 
 - 39 -
 
 SN-U391E
 
 3.2.20. Three-dimensional Display Software (Freehand) (EZU-3D8)  Components
 
 Software (Freehand 3D software)
 
  Applicable probe
 
 Refer to Appendix
 
  Display mode
 
 - ROI setting mode
 
 Displays the ROI that specifies the 3D image creation area on the 2D image
 
 - 3D mode
 
 3D image is displayed on 1 screen.
 
 - 2D-3D mode
 
 Screen is vertically divided into two sides. The ROI setting mode is displayed on the left side and the 3D mode on the right side
 
 - MPR 4 screen mode
 
 Screen is vertically and horizontally divided into four. 3D image is displayed on the lower right, while the random orthogonal 3 cross-section surface images are displayed on the other sections
 
 - MPR 1 screen mode
 
 One of the orthogonal 3 cross-section surface image in MPR 4 screen mode is displayed on 1 screen
 
  Rendering mode
 
 B/W - Opacity - MIP (Maximum Intensity Projection) - mIP (minimum Intensity Projection) - Transparency - Inversion - Orthogonal - Cube - Opac. Advanced - 4Dshading Color - Opacity - MIP (Maximum Intensity Projection) - Transparency - Orthogonal - Cube  Quality settings
 
 B/W - Brightness 
 
 3D image
 
 17 steps (-8 ~ 8, except rendering mode: Orthogonal and Cube)
 
 
 
 MPR image
 
 17 steps (-8 ~ 8)
 
 - Color 
 
 3D color
 
 37 types (except rendering mode: Orthogonal, Cube and, Opac.Advanced, 4Dshading) - 40 -
 
 SN-U391E
 
 
 
 MPR color
 
 37 types
 
 - Shading
 
 6 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
 
 - Filter
 
 20 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced,, or 4Dshading)
 
 - Threshold
 
 256 steps (rendering mode: opacity, Opac.Invert, Opac.Advanced, or 4Dshading)
 
 - Smoothing
 
 8 steps (rendering mode: opacity, Opac.Invert, or Opac.Advanced)
 
 - P-Filter
 
 8 steps (rendering mode: MIP or mIP)
 
 - BW 3D Smooth.
 
 3 types (rendering mode: 4Dshading)
 
 - 3D Smooth Coef
 
 11 types (rendering mode: Opac.Advanced, or
 
 - Depth Contrast
 
 40 types (rendering mode: Opac.Advanced)
 
 - Depth Color Shift
 
 127 types (rendering mode: Opac.Advanced)
 
 - Region Map
 
 9 types (rendering mode: 4Dshading)
 
 - Absorption
 
 10 types (rendering mode: 4Dshading)
 
 - Hue
 
 4 types (rendering mode: 4Dshading)
 
 - Shadow Smooth
 
 10 types (rendering mode: 4Dshading)
 
 - Light Intensity
 
 5 types (rendering mode: 4Dshading)
 
 - Lighting Reso.
 
 4 types (rendering mode: 4Dshading)
 
 - Transp. Level
 
 8 types (rendering mode: 4Dshading)
 
 4Dshading)
 
 Color - Combination
 
 B/W and CFM image combination / only B/W image / only CFM image (only CFM image is impossible when rendering mode is
 
 Orthogonal or
 
 Cube) - Blend
 
 100 steps (except rendering mode: Orthogonal and Cube)
 
 - Brightness 
 
 3D image
 
 17 steps (-8 ~ 8, except rendering mode: Orthogonal and Cube)
 
 
 
 MPR image
 
 17 steps (-8 ~ 8) (except CFI mode)
 
 - Shading
 
 6 steps (rendering mode: opacity except CFI mode)
 
 - Filter
 
 5 steps (rendering mode: only opacity)
 
 - Threshold
 
 256 steps (rendering mode: only opacity )
 
 - Smoothing
 
 8 steps (rendering mode: only opacity)
 
 - P-Filter
 
 8 steps (rendering mode: only MIP)
 
  Operations
 
 - Rotation
 
 By trackball and encoder
 
 - View Line
 
 The cut-line of arbitrary curvature can be specified on 2D image
 
 - Sliding, Plane cut
 
 Max. 14 planes by trackball (only in frozen, except rendering mode: Orthogonal) Image operations: both B/W and CFM images / only B/W image (both B/W and CFM image is only possible when rendering mode is Cube)
 
 - Cut function
 
 The arbitrarily specified part of 3D image can be dug (only in frozen, except - 41 -
 
 SN-U391E
 
 rendering mode: Orthogonal and Cube) Image operations: both B/W and CFM images / only B/W image - Zoom
 
 1.0~4.0 (0.25 steps)
 
 - Image filing
 
 3D volume data, Real-time 3D mode single images (such as BMP) and multi images (such as AVI) can be saved Saved images can be played back from the image filing screen, as with regular 2D images Image pre-cutted by ViewLine and full image can be switched
 
 - PSS function
 
 Using the 3D PSS (3D Preset Scanning Selector) function
 
  Measurement
 
 Not applicable
 
  User-defined screen button
 
 The items of 3D screen button menu and image adjustment screen button menu are registerable
 
 3.2.21. Remote Control Unit (EZU-RH4)  Components
 
 Remote control, battery charger
 
  Size
 
 185 (H) × 76 (W) × 22.5 (T) mm
 
  Weight
 
 220 g
 
  Continuous use time
 
 Approx.5 hours (on a full charge)
 
  Charging time
 
 Approx. 3 hours
 
  Battery
 
 Lithium-ion battery
 
  Transmission method
 
 Infrared communication
 
  Coverage range
 
 Within 3 m
 
  Waterproofing
 
 Applicable
 
  Plasma sterilization
 
 Applicable
 
  Remote control function
 
 B/CFM/PW mode changes, B/CFM/PW gain adjustment, REF changes, focus position changes, PRF changes, depth changes, freeze ON/OFF, caliper, Single Image recording, movie recording, print out, orientation, trackball priority, display changes,Picture in Picture settings, Remote user key1, Remote user key2 (User defined)
 
 (Notice) : EZU-RiF5 is required for EZU-RH4. (Notice) : EZU-PP5 is required for “Picture in Picture display change.”
 
 3.2.22. Remote Control Interface Unit (EZU-RiF5) This is the optical receiver installed in the ultrasound diagnostic scanner (main unit) for making to work remote control unit EZU-RH4.  Components
 
 KBIF PCB
 
 3.2.23. Picture in Picture Software (EZU-PP5)  Components
 
 Software - 42 -
 
 SN-U391E
 
  Input signal
 
 Y/C
 
  Applicable probes
 
 All probes
 
  Display mode
 
 - Full screen display - Dual screen display
 
 Probe : FG series, EG series, EB series, EUP-OL531 Mode : B, CFM, Elasto
 
 - Small screen display  Display position
 
 - Dual screen display
 
 L/R preset can be possible
 
 - Small screen display
 
 Four corners preset can be possible, display position can be changed in real-time
 
  Display ON/OFF
 
 Applicable
 
  Cine playback
 
 Applicable only US images
 
  Movie recording
 
 Applicable
 
 (Notice) : EZU-ACP2 and EZU-ViF2 are required for EZU-PP5.
 
 3.2.24. Network Interface Unit (EZU-NW1)  Components
 
 LAN cable
 
  Specifications
 
 Ethernet 10/100 Base-T (IEEE802.3)
 
 3.2.25. Analog Capture Card Unit (EZU-ACP2)  Components
 
 Analog capture card, cables, Driver software
 
 3.2.26. Digital Capture Card Unit (EZU-DCP3)  Components
 
 Digital capture card, cables, Driver software
 
 3.2.27. Video Interface Unit (EZU-ViF2)  Components
 
 TVIF unit, cables
 
  Specifications
 
 RGC output Y/C input and output (Only connector for input) Brightness signal Remote control output for BE printer Audio input and output
 
 3.2.28. Expanded USB Unit (EZU-UP2)  Components
 
 Expanded USB Unit
 
  Specifications
 
 - Number of USB ports
 
 2 ports
 
 - 43 -
 
 SN-U391E
 
 3.2.29. External LCD Monitor Unit (EZU-EL2)  Components
 
 DVI Connector Unit
 
  Specifications
 
 - Video signal
 
 DVI-D
 
 - Resolution
 
 SXGA(1280x1024)
 
 3.2.30. Jelly Warmer Unit (EZU-JW1)  Components
 
 Jelly Warmer, Power cable
 
 3.2.31. Alpha numeric keyboard Unit (EZU-AK1E)  Components
 
 Alphanumeric keyboard unit, Lock screw, Hexagon socket head bolt M6X30
 
 3.2.32. High Neck Unit (EZU-HN1)  Components
 
 High Neck Unit
 
 3.2.33. Probe Hanger (EZU-TH7)  Components
 
 Probe Hanger
 
 3.2.34. Processor Tray (EZU-TY3)  Components
 
 Processor tray
 
 (Notice):
 
 EZU-HN1 is required for EZU-TY3.
 
 (Notice):
 
 When you install EZU-TY3 (processor tray) onto the equipment, peripherals other than the video processor made by HOYA cannot be installed.
 
 3.2.35. Junction Box (PUN-JBP1)  Components
 
 Junction box
 
  Outside dimension
 
 150 (W) × 118.5 (D) × 110.5 (H) mm
 
  Weight
 
 Approx. 900 g
 
 3.3. Power requirements  Input voltage
 
 100/110 120, 220~240VAC ± 10%
 
  Frequency
 
 50 / 60 Hz
 
  Auxiliary AC outlet
 
 350VA (peripheral devices)
 
  Consumption
 
 Max. 1.5kVA (include 350VA for AC outlet)
 
  Outlet
 
 The 3-wire power outlet which has a protective earth terminal (with less than 10Ω grounding resistance) of the grounding facility.
 
 - 44 -
 
 SN-U391E
 
 3.4. Environment conditions  Temperature
 
 10 ~ 35°C
 
  Relative humidity
 
 30 ~ 85% RH (No dew condensation)
 
  Atmospheric pressure
 
 700 ~ 1060hP
 
 3.5. Exterior view  Outside dimension - 450 (W) x 700 (D)
 
 (with EZU-MT29-S1)
 
 x 1350 (H) mm - 450 (W) x 700 (D)
 
 (with EZU-MT29-S1 and EZU-HN1)
 
 x 1430 (H) mm - 450 (W) x 700 (D)
 
 (with EZU-MT30-S1)
 
 x 1400 (H) mm - 450 (W) x 700 (D)
 
 (with EZU-MT30-S1 and EZU-HN1)
 
 x 1480 (H) mm  Weight - Approx. 125kg
 
 (standard configuration with EZU-MT29-S1)
 
 - Approx. 127kg
 
 (standard configuration with EZU-MT29-S1 and EZU-HN1)
 
 - Approx. 130kg
 
 (standard configuration with EZU-MT30-S1)
 
 - Approx. 132kg
 
 (standard configuration with EZU-MT30-S1 and EZU-HN1)
 
  Monitor
 
 17 inch digital LCD monitor 1280 x 1024 Non interlace
 
  LCD Monitor Arm - simple arm
 
 Angle:
 
 - flexible arm
 
 Height: adjustable approx. 18cm Angle:
 
 adjustable 2 directions adjustable 4 directions
 
  Caster
 
 Swivel・lock /brake
 
  Probe connector
 
 4 probes (3 probes available, 1 probe rest)
 
  Jelly wormer
 
 Applicable (Option)
 
  Handle for movement
 
 Rear side .
 
 - 45 -
 
 SN-U391E
 
 1350 820 450
 
 700
 
 900
 
 1430
 
 Body dimension (with EZU-MT29-S1)
 
 450
 
 700 Unit: mm Body dimension (with EZU-MT29-S1 and EZU-HN1)
 
 - 46 -
 
 SN-U391E
 
 1400 820 450
 
 700
 
 900
 
 1480
 
 Body dimension (with EZU-MT30-S1)
 
 450
 
 700
 
 Unit: mm
 
 Body dimension (with EZU-MT30-S1 and EZU-HN1) E - 47 -
 
 SN-U391E