Customer - EPX-Seires Service Manual Rev.E (USA) - 03 - 2017 [PDF]

  • 0 0 0
  • Gefällt Ihnen dieses papier und der download? Sie können Ihre eigene PDF-Datei in wenigen Minuten kostenlos online veröffentlichen! Anmelden
Datei wird geladen, bitte warten...
Zitiervorschau

EPX-SERISE SERVICE MANUAL Rev. E

EPX-Series Diagnostic X-ray Systems Service Manual

1

EPX-SERISE SERVICE MANUAL Rev. E (This page intentionally left blank)

2

EPX-SERISE SERVICE MANUAL Rev. E

SERVICE CHAPTER

1. CALIBRATION------------------------------------------------------------------------------------------- 7 2. DIP SWITCH SETTING--------------------------------------------------------------------------------31 3 .HARD-WIRED hand switch control---------------------------------------------------------------32 4. COLLIMATOR-------------------------------------------------------------------------------------------33 5. TROUBLE SHOOTING-------------------------------------------------------------------------------35 6. SERVICE CHECK LIST-------------------------------------------------------------------------------36

REVISON HISTORY 3

EPX-SERISE SERVICE MANUAL Rev. E Revision Number

Date

Description

A

SEP 11, 2013

First Edition

B

SEP 30, 2014

European Authorized Representative Change

C

JAN 12, 2015

Manual Updates

D

SEP 01, 2015

Manual Updates

E

JUN 21, 2016

Manual Updates

F

4

EPX-SERISE SERVICE MANUAL Rev. E (This page intentionally left blank)

5

EPX-SERISE SERVICE MANUAL Rev. E

Service chapter This manual is designed to ensure correct use and operation of EPX model diagnostic x-ray units. Please read all the lines thoroughly before you use the equipment. Incorrect use and operation exceeding described conditions in this manual may cause damage of the equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. This equipment should be used only by the legally qualified persons and practitioners. Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be used. Only trained maintenance staff may remove the covers of the x-ray system. Do not modify the equipment at your discretion and in case any modification is required unavoidably, ask the help of Ecotron or its authorized dealer for the service. Information for repair will be provided by Ecotron. (Circuit diagram, Component part list, descriptions, Calibration instructions, or other information) No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties which many results from the use of this manual. Always keep the manual at hand for your reference.

NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following rules and methods described in this manual.

CAUTION The ionizing radiation is dangerous for the operator if the following safety measures are not strictly observed.

6

EPX-SERISE SERVICE MANUAL Rev. E

1. Calibration 1.1 EPX-F1600 1.1.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.1.2 Calibration Procedure 1.1.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(30mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 90KV and 5mAs(16mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 4.5V, adjust KV by turning VR3 on the OP board.

7

EPX-SERISE SERVICE MANUAL Rev. E

1.1.2.2

Tube current confirmation and calibration.

a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(30mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C09 are calibration code of tube current at AC110V model and C10~C19 are calibration code of tube current at AC220V model. C00 or C10 is calibration code of tube current (30mA at 40kV~50kV) at AC110V or AC220V and 30.0 is calibration data of tube current (30mA at 40kV~50kV). a. Adjust mA by using the mAs Up-Down switch (see operation sheet).

8

EPX-SERISE SERVICE MANUAL Rev. E

b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,

Where, C01 or C11 is calibration code of tube current (25mA at 40kV~50kV) and 25.0 is calibration data of tube current (25mA at 40kV~50kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

l. If it is not 5V, set the calibration data again.

Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,

9

EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P07 are calibration code of filament preheat current at AC110V, P08~P15 are calibration code of filament preheat current at AC220V, P00 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA), and 1.50 or 1.40 is Filament preheat calibration data (Preheat calibration data of 30mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,

Where, P01 or P09 is Filament preheat calibration code (Preheat calibration code of 25mA) and 1.45 or 1.35 is Filament preheat calibration data (Preheat calibration data of 25mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

10

EPX-SERISE SERVICE MANUAL Rev. E

1.2 EPX-F2400 1.2.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.2.2 Calibration Procedure 1.2.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(40mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100KV and 5mAs(20mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5V, adjust KV by turning VR3 on the OP board.

11

EPX-SERISE SERVICE MANUAL Rev. E

1.2.2.2

Tube current confirmation and calibration.

a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 (GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(40mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration code of tube current at AC220V model. C00 or C08 is calibration code of tube current (40mA at 40kV~60kV) at AC110V or AC220V and 40.0 is calibration data of tube current (40mA at 40kV~60kV).

12

EPX-SERISE SERVICE MANUAL Rev. E

a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,

Where, C01 or C09 is calibration code of tube current (35mA at 40kV~60kV) and 35.0 is calibration data of tube current (35mA at 40kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

l. If it is not 5.25V, set the calibration data again.

Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,

13

EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P04 are calibration code of filament preheat current at AC110V, P05~P09 are calibration code of filament preheat current at AC220V, P00 or P05 is Filament preheat calibration code (Preheat calibration code of 40mA), and 0.70 or 0.72 is Filament preheat calibration data (Preheat calibration data of 40mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,

Where, P01 or P06 is Filament preheat calibration code (Preheat calibration code of 35mA) and 0.67 or 0.69 is Filament preheat calibration data (Preheat calibration data of 35mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

14

EPX-SERISE SERVICE MANUAL Rev. E

1.3 EPX-F2800 1.3.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.3.2

Calibration Procedure

1.3.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14(GND) on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(35mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 120KV and 5mAs(16mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 6V, adjust KV by turning VR3 on the OP board.

15

EPX-SERISE SERVICE MANUAL Rev. E

1.3.2.2 Tube current confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(35mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C15 are calibration code of tube current at AC110V model and C16~C31 are calibration code of tube current at AC220V model. C00 or C16 is calibration code of tube current (35mA at 40kV~60kV) at AC110V or AC220V and 35.0 is calibration data of tube current (35mA at 40kV~60kV).

16

EPX-SERISE SERVICE MANUAL Rev. E

a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,

Where, C01 or C17 is calibration code of tube current (30mA at 61kV~70kV) and 30.0 is calibration data of tube current (30mA at 61kV~70kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

l. If it is not 5.15V, set the calibration data again.

Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,

17

EPX-SERISE SERVICE MANUAL Rev. E

where, P00~P06 are calibration code of filament preheat current at AC110V, P07~P13 are calibration code of filament preheat current at AC220V, P00 or P07 is Filament preheat calibration code (Preheat calibration code of 35mA), and 0.70 or 0.70 is Filament preheat calibration data (Preheat calibration data of 35mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,

Where, P01 or P08 is Filament preheat calibration code (Preheat calibration code of 30mA) and 0.67 or 0.67 is Filament preheat calibration data (Preheat calibration data of 30mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

18

EPX-SERISE SERVICE MANUAL Rev. E

1.4 EPX-F3200 1.4.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.4.2

Calibration Procedure

1.4.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(60mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 100KV and 5mAs(30mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5V, adjust KV by turning VR3 on the OP board.

19

EPX-SERISE SERVICE MANUAL Rev. E

1.4.2.2

Tube current confirmation and calibration.

a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(60mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 6.0 V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C07 are calibration code of tube current at AC110V model and C08~C15 are calibration code of tube current at AC220V model. C00 or C08 is calibration code of tube current (60mA at 40kV~50kV) at AC110V and 60.0 is calibration data of tube current (60mA at 40kV~50kV). a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet).

20

EPX-SERISE SERVICE MANUAL Rev. E

c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 6.0V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,

Where, C01 or C09 is calibration code of tube current (50mA at 51kV~60kV) and 50.0 is calibration data of tube current (50mA at 51kV~60kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

l. If it is not 5.0V, set the calibration data again.

Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,

21

EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P05 are calibration code of filament preheat current at AC110V, P06~P11 are calibration code of filament preheat current at AC220V, P00 or P06 is Filament preheat calibration code (Preheat calibration code of 60mA), and 1.65 or 1.65 is Filament preheat calibration data (Preheat calibration data of 60mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,

where, P01 or P07 is Filament preheat calibration code (Preheat calibration code of 50mA) and 1.60 or 1.60 is Filament preheat calibration data (Preheat calibration data of 50mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

22

EPX-SERISE SERVICE MANUAL Rev. E

1.5 EPX-F4000 1.5.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.5.2

Calibration Procedure

1.5.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 50KV and 5mAs(80mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2.5V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 110KV and 5mAs(25mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.

23

EPX-SERISE SERVICE MANUAL Rev. E

1.5.2.2

Tube current confirmation and calibration.

a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 50kV and 5mAs(80mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.44V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C14 are calibration code of tube current at AC110V model and C15~C29 are calibration code of tube current at AC220V model. C00 or C15 is calibration code of tube current (80mA at 50kV) at 80.0 is calibration data of tube current (80mA at 50kV). a. Adjust mA by using the mAs Up-Down switch (see operation sheet). b. After it is adjusted, push the STORE button during 3sec(see operation sheet). 24

EPX-SERISE SERVICE MANUAL Rev. E

c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 4V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,

Where, C01 or C16 is calibration code of tube current (70mA at 50kV) and 70.0 is calibration data of tube current (70mA at 50kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

l. If it is not 3.0V, set the calibration data again.

Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,

25

EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P09 are calibration code of filament preheat current at AC110V, P10~P19 are calibration code of filament preheat current at AC220V, P00 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA), and 2.50 or 2.50 is Filament preheat calibration data (Preheat calibration data of 80mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,

Where, P01 or P11 is Filament preheat calibration code (Preheat calibration code of 70mA) and 2.20 or 2.20 is Filament preheat calibration data (Preheat calibration data of 70mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

26

EPX-SERISE SERVICE MANUAL Rev. E

1.6 EPX-F5000 1.6.1 Line Power Switch Line switch should be at off position. Connect power line cord to the power outlet. Turn on the line switch. The FND on the control panel should illuminate. 1.6.2

Calibration Procedure

1.6.2.1 Tube voltage confirmation and calibration. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40KV and 5mAs(100mA), respectively. h. Then observe the KV waveform of the oscilloscope channel 2. i. If it is not 2V, adjust KV by turning VR3 on the OP board.

j. Set the kV and the mAs at 110KV and 5mAs(40mA), respectively. k. Then observe the KV waveform of the oscilloscope channel 2. l. If it is not 5.5V, adjust KV by turning VR3 on the OP board.

27

EPX-SERISE SERVICE MANUAL Rev. E

1.6.2.2

Tube current confirmation and calibration.

a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. Connect the channel one (1) probe of the oscilloscope to TP8 (IP F/B). d. Connect the channel two (2) probe of the oscilloscope to TP11 (EP F/B). e. Connect the ground to TP13 or TP14 or on the OP board. f. Turn on the line switch. g. Set the kV and the mAs at 40kV and 5mAs(100mA), respectively. h. Then observe the mA waveform of the oscilloscope channel 1. i. If it is not 5.1V, adjust mA using the calibration data on the OP BOARD. j. DIP SWITCH #4 on the OP BOARD PCB is set to ON position.

Then the display should indicate as follows,

Where, C00~C20 are calibration code of tube current at AC110V model and C21~C41 are calibration code of tube current at AC220V model. C00 or C21 is calibration code of tube current (100mA at 50kV) at A0.0(=100) is calibration data of tube current (100mA at 50kV). a. Adjust mA by using the mAs Up-Down switch (see operation sheet).

28

EPX-SERISE SERVICE MANUAL Rev. E

b. After it is adjusted, push the STORE button during 3sec(see operation sheet). c. DIP SWITCH #4 on the PCB OP BOARD should be set to OFF position and the expose X-ray. e. If it is not 5.1V, set the calibration data again. f. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. g. Then push the kV UP switch (see operation sheet). h. The display should indicate as follows,

Where, C01 or C22 is calibration code of tube current (80mA at 50kV) and 80.0 is calibration data of tube current (80mA at 50kV). i. Adjust mA by using the mAs Up-Down switch (see operation sheet). j. After it is adjusted, push the STORE switch during 3sec(see operation sheet). k. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

l. If it is not 4.08V, set the calibration data again.

Filament preheat calibration This is preheating filament for making tube current. The overshooting or the undershooting may occur from the tube current if the Preheat calibration data is set by mistake. Changing the calibration data is possible in accordance with the tube condition. a. Turn off the line switch. b. Detach the control panel from the unit to access the PCB OP BOARD. c. DIP SWITCH #4 on the PCB OP BOARD is set to ON position. d. Turn on the line switch. f. Push APR S/W ① and then the display should indicate as follows,

29

EPX-SERISE SERVICE MANUAL Rev. E

Where, P00~P08 are calibration code of filament preheat current at AC110V, P09~P15 are calibration code of filament preheat current at AC220V, P00 or P09 is Filament preheat calibration code (Preheat calibration code of 100mA), and 2.40 or 2.40 is Filament preheat calibration data (Preheat calibration data of 100mA). g. Adjust data by using the mAs Up-Down switch (see operation sheet). h. After it is adjusted, push the STORE button during 3sec(see operation sheet). i. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then expose X-ray.

j. If overshooting or undershooting occurs in the waveform, set the calibration data again. k. Then push the KV UP switch (see operation sheet). l. The display should indicate as follows,

Where, P01 or P10 is Filament preheat calibration code (Preheat calibration code of 80mA) and 2.25 or 2.25 is Filament preheat calibration data (Preheat calibration data of 80mA). m. Adjust data by using the mAs Up-Down switch on the Control Panel. n. After it is adjusted, push the Store switch during 3sec(see operation sheet). o. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray. p. If overshooting or undershooting occurs in the waveform, set the calibration data again.

30

EPX-SERISE SERVICE MANUAL Rev. E

2. DIP Switch setting

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position.

31

EPX-SERISE SERVICE MANUAL Rev. E

3. Hard-wired hand switch control 3.1 Introduction The Exposure Hand Switch consists of a two step hand switch. The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage. The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.

SWITCH 1 SWITCH 2

3.2 Exposure process There are two ways to use the hand switch to get the exposure.

a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or Ready” status on, then press Switch #2 to get the “Exposure”.

b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1 second. This action will result in “Prep or Ready” status and “Exposure” at the same time.

32

EPX-SERISE SERVICE MANUAL Rev. E

4. Collimator a. Halogen Lamp * 1 b. Laser pointer * 2 c. Horizontal and vertical lighting range control i. Block diagram

ii. Layout a. External Layout

33

EPX-SERISE SERVICE MANUAL Rev. E

b. Internal Layout

c. Knob Layout

34

EPX-SERISE SERVICE MANUAL Rev. E

5. Trouble shooting If you have any error message when you are using portable x-ray, please check below service check list

5.1 EPX-F1600/F2400/F2800/F3200/F4000/F5000 Error Code Table ERROR Code

Description

Indication

2

THERMAL ERR(INTERLOCK)

Excess from 60℃

7

READY SW ERR

READY switch malfunction

8

EXPOSURE SW ERR

EXPOSURE switch malfunction

10

standby FIL FB ERROR

Excess from 200mA (10% of max. current, 2A)

11

standby EP FB ERROR

Excess from 12KV (10% of max. voltage, 120KV)

12

standby IP FB ERROR

Excess from 3.5mA (10% of MAX current, 35mA)

14

run FIL FB ERR

Out of range from reference±10%

15

run EP FB or AC POWER LOW ERR

Out of range from reference-10% or AC power line poor

16

run IP FB HIGH ERR

Out of range from reference+10%

17

run EP FB HIGH ERR

Out of range from reference+10%

18

run IP FB LOW ERR

Out of range from reference-10%

28

Pulse by Pulse ERR

FET board fault(Hardware)

29

INC_OC ERR

Feedback error

30

KV_OVP ERR

kV over voltage(Hardware)

31

FIL_OC ERR

FIL over current(Hardware)

32

IP_OCP ERR

mA over current(Hardware)

35

EPX-SERISE SERVICE MANUAL Rev. E

6. Service Check List 6.1 EPX-F1600 Please fill in the blank. kV/mAs condition(When error is occurred)

Product S/N

Error code

※ Open upper cover of portable x-ray below picture.

6.1.1 Voltage check 1) Turn on AC switch

VDC Range Pass/Fail

TP26

TP24

TP25

TP6

4.8V~5.0V

11.5V~12.5V

-11.5V~-12.5V

+3V~3.6V

Pass

Fail

Pass

Fail

Measured Voltage If it is fail case

36

Pass

Fail

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.1.2. kV, mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch.

4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage

40msec/div

EP_FB 2V Pass

IP_FB 6V Fail

If kV, mA waveforms are fail, please attach your kV, mA waveform

37

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 2.5V Pass

IP_FB 6V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 3V Pass

IP_FB 5V Fail

are fail, please attach your kV, mA waveform

38

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 3.5V Pass

IP_FB 4V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

100msec/div

EP_FB 4V Pass

IP_FB 3.6V Fail

are fail, please attach your kV, mA waveform

39

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

100msec/div

EP_FB 4.5V Pass

IP_FB 3.1V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

6.1.3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.

40

EPX-SERISE SERVICE MANUAL Rev. E

6.1.4 kV, mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture. 3) Turn on AC switch.

4) kV / mA PWM frequency CH 1: kV

2V/div

CH Frequency Pass/Fail Measured frequency

CH 2 : mA

2V/div

4μsec/div

CH 1 mA PWM 200kHz Pass

41

Fail

CH 2 kV PWM 140kHz Pass Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.1.5 mA calibration data check.

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C09(AC110V), C10~C19(AC220V) Please check Input data.

42

EPX-SERISE SERVICE MANUAL Rev. E Input data and Tube current should be the same. Tube Current Parameter

Tube Voltage

mAs

Tube Current

INPUT DATA

AC110V

AC220V

KV

MIN

MAX

mA

mA

C00

C10

40~50 (1)

0.4

20

30.0

30.0

C01

C11

40~50 (2)

25

32

25.0

25.0

C02

C12

51~60 (1)

0.4

20

25.0

25.0

C03

C13

51~60 (2)

25

32

20.0

20.0

C04

C14

61~70 (1)

0.4

20

20.0

20.0

C05

C15

61~70 (2)

25

32

15.0

15.0

C06

C16

71~80 (1)

0.4

20

18

18

C07

C17

71~80 (2)

25

32

14

14

C08

C18

81~90 (1)

0.4

20

16

16

C09

C19

81~90 (2)

25

32

12

12

4) If mA calibration is done, Please dip switch 4 set to OFF position.

6.1.6 Filament Preheat Calibration Code and Data check 1) Turn off

AC switch

2) Dip switch on op board

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off)

43

EPX-SERISE SERVICE MANUAL Rev. E For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P07(AC110V), P08~P15(AC220V) Please check Input data. Input data and Tube current should be the same.

Fill in the blank below table because P_DATA is different for each product.

Model

1.6kW

Current(mA)

P_Mode(110V)

P_Mode(220V)

30

P00

P08

25

P01

P09

20

P02

P10

18

P03

P11

16

P04

P12

15

P05

P13

14

P06

P14

12

P07

P15

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

44

P_DATA

EPX-SERISE SERVICE MANUAL Rev. E

6.2 EPX-F2400 Please fill in the blank. kV/mAs condition(When error is occurred)

Product S/N

Error code

※ Open upper cover of portable x-ray below picture.

6.2.1 Voltage check 1) Turn on AC switch

VDC Range Pass/Fail

TP26

TP24

TP25

TP6

4.8V~5.0V

11.5V~12.5V

-11.5V~-12.5V

+3V~3.6V

Pass

Fail

Pass

Fail

Measured Voltage If it is fail case

45

Pass

Fail

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.2.2. kV, mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch.

4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage

40msec/div

EP_FB 2V Pass

IP_FB 6V Fail

If kV, mA waveforms are fail, please attach your kV, mA waveform

46

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 2.5V Pass

IP_FB 6V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 3V Pass

IP_FB 6V Fail

are fail, please attach your kV, mA waveform

47

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 3.5V Pass

IP_FB 4.5V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 4V Pass

IP_FB 4.5V Fail

are fail, please attach your kV, mA waveform

48

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 4.5V Pass

IP_FB 3.75V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 5V Pass

IP_FB 3V Fail

are fail, please attach your kV,mA waveform

49

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.2.3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.

6.2.4 kV, mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture. 3) Turn on AC switch.

50

EPX-SERISE SERVICE MANUAL Rev. E 4) kV / mA PWM frequency CH 1: kV

2V/div

CH 2 : mA

2V/div

CH Frequency Pass/Fail Measured frequency

4μsec/div

CH 1 mA PWM 200kHz Pass

Fail

6.2.5 mA calibration data check.

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

51

CH 2 kV PWM 140kHz Pass Fail

EPX-SERISE SERVICE MANUAL Rev. E 3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C07(AC110V), C08~C15(AC220V) Please check Input data.

Input data and Tube current should be the same.

Tube Current Parameter

mAs

Tube Voltage

Tube Current

INPUT DATA

AC110V

AC220V

KV

MIN

MAX

mA

mA

C00

C08

40~60(1)

0.4

64

40.0

40.0

C01

C09

40~60(2)

80

100

35.0

30.0

C02

C10

61~70(1)

0.4

100

30.0

25.0

C03

C11

71~80(1)

0.4

32

30.0

30.0

C04

C12

71~80(2)

40

80

25.0

25.0

C05

C13

81~90(1)

0.4

40

25.0

25.0

C06

C14

81~90(2)

50

80

20.0

20.0

C07

C15

91~100(1)

0.4

80

20.0

20.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

52

EPX-SERISE SERVICE MANUAL Rev. E

6.2.6 Filament Preheat Calibration Code and Data check 1) Turn off

AC switch

2) Dip switch on op board

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P04(AC110V), P05~P09(AC220V) Please check Input data. Input data and Tube current should be the same.

53

EPX-SERISE SERVICE MANUAL Rev. E Fill in the blank below table because P_DATA is different for each product.

Model

2.4kW

Current(mA)

P_Mode(110V)

P_Mode(220V)

40

P00

P05

35

P01

P06

30

P02

P07

25

P03

P08

20

P04

P09

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

54

P_DATA(A)

EPX-SERISE SERVICE MANUAL Rev. E

6.3 EPX-F2800 Please fill in the blank. kV/mAs condition(When error is occurred)

Product S/N

Error code

※ Open upper cover of portable x-ray below picture.

6.3.1 Voltage check 1) Turn on AC switch

VDC Range Pass/Fail

TP26

TP24

TP25

TP6

4.8V~5.0V

11.5V~12.5V

-11.5V~-12.5V

+3V~3.6V

Pass

Fail

Pass

Fail

Measured Voltage If it is fail case

55

Pass

Fail

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.3.2. kV, mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch.

4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage

40msec/div

EP_FB 2V Pass

IP_FB 6V Fail

If kV, mA waveforms are fail, please attach your kV, mA waveform

56

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 2.5V Pass

IP_FB 6V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 3V Pass

IP_FB 6V Fail

are fail, please attach your kV, mA waveform

57

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 3.5V Pass

IP_FB 4.5V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 4V Pass

IP_FB 5.25V Fail

are fail, please attach your kV, mA waveform

58

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 4.5V Pass

IP_FB 4.5V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 5V Pass

IP_FB 3.75V Fail

are fail, please attach your kV,mA waveform

59

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 11) 110kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 110kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 110kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 5.5V Pass

IP_FB 2.7V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

12) 120kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 120kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 120kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

100msec/div

EP_FB 6V Pass

IP_FB 2.4V Fail

are fail, please attach your kV,mA waveform

60

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.3.3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.

6.3.4 kV, mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture. 3) Turn on AC switch.

61

EPX-SERISE SERVICE MANUAL Rev. E 4) kV / mA PWM frequency CH 1: kV

2V/div

CH 2 : mA

2V/div

CH Frequency Pass/Fail Measured frequency

4μsec/div

CH 1 mA PWM 200kHz Pass

Fail

6.3.5 mA calibration data check.

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

62

CH 2 kV PWM 140kHz Pass Fail

EPX-SERISE SERVICE MANUAL Rev. E 3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C015(AC110V), C16~C31(AC220V) Please check Input data.

Input data and Tube current should be the same.

Tube Current Parameter

Tube Voltage

mAs

Tube Current

INPUT DATA

AC110V

AC220V

KV

MIN

MAX

mA

mA

C00

C16

40~60

0.4

100

35.0

35.0

C01

C17

61~70(1)

0.4

64

30.0

30.0

C02

C18

61~70(2)

80

100

25.0

25.0

C03

C19

71~80(1)

0.4

10

35.0

35.0

C04

C20

71~80(2)

13

32

30.0

30.0

C05

C21

71~80(3)

40

80

25.0

25.0

C06

C22

81~90(1)

0.4

10

30.0

30.0

C07

C23

81~90(2)

13

40

25.0

25.0

C08

C24

81~90(3)

50

80

20.0

20.0

C09

C25

91~100(1)

0.4

13

25.0

25.0

C10

C26

91~100(2)

16

50

20.0

20.0

C11

C27

91~100(3)

64

80

16.0

16.0

C12

C28

101~110(1)

0.4

32

18.0

18.0

C13

C29

101~110(2)

40

64

16.0

16.0

C14

C30

111~120(1)

0.4

20

16.0

16.0

C15

C31

111~120(2)

25

50

14.0

14.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

63

EPX-SERISE SERVICE MANUAL Rev. E

6.3.6 Filament Preheat Calibration Code and Data check 1) Turn off

AC switch

2) Dip switch on op board

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P06(AC110V), P07~P13(AC220V) Please check Input data. Input data and Tube current should be the same.

64

EPX-SERISE SERVICE MANUAL Rev. E Fill in the blank below table because P_DATA is different for each product.

Model

2.8kW

Current(mA)

P_Mode(110V)

P_Mode(220V)

35

P00

P07

30

P01

P08

25

P02

P09

20

P03

P10

18

P04

P11

16

P05

P12

14

P06

P13

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

65

P_DATA(A)

EPX-SERISE SERVICE MANUAL Rev. E

6.4 EPX-F3200 Please fill in the blank. kV/mAs condition(When error is occurred)

Product S/N

Error code

※ Open upper cover of portable x-ray below picture.

6.4.1 Voltage check 1) Turn on AC switch

VDC Range

TP26

TP24

TP25

TP6

4.8V~5.0V Pass Fail

11.5V~12.5V Pass Fail

-11.5V~-12.5V Pass Fail

+3V~3.6V Pass Fail

Pass/Fail Measured Voltage If it is fail case

66

EPX-SERISE SERVICE MANUAL Rev. E

6.4.2. kV, mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch.

4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage

20msec/div

EP_FB 2V Pass

IP_FB 6V Fail

If kV, mA waveforms are fail, please attach your kV, mA waveform

67

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

20msec/div

EP_FB 2.5V Pass

IP_FB 6V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

20msec/div

EP_FB 3V Pass

IP_FB 5V Fail

are fail, please attach your kV, mA waveform

68

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 3.5V Pass

IP_FB 4V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 4V Pass

IP_FB 4V Fail

are fail, please attach your kV, mA waveform

69

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 4.5V Pass

IP_FB 3V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 5V Pass

IP_FB 3V Fail

are fail, please attach your kV,mA waveform

70

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.4.3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.

6.4.4 kV, mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture. 3) Turn on AC switch.

71

EPX-SERISE SERVICE MANUAL Rev. E 4) kV / mA PWM frequency CH 1: kV

2V/div

CH 2 : mA

2V/div

CH Frequency Pass/Fail Measured frequency

4μsec/div

CH 1 mA PWM 200kHz Pass

Fail

6.4.5 mA calibration data check.

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

72

CH 2 kV PWM 140kHz Pass Fail

EPX-SERISE SERVICE MANUAL Rev. E 3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C07(AC110V), C08~C15(AC220V) Please check Input data.

Input data and Tube current should be the same.

Tube Current Parameter

mAs

Tube Voltage

Tube Current

INPUT DATA

AC110V

AC220V

KV

MIN

MAX

mA

mA

C00

C08

40~50

0.4

100

60.0

60.0

C01

C09

51~60

0.4

80

50.0

50.0

C02

C10

61~70

0.4

100

40.0

40.0

C03

C11

71~80(1)

0.4

32

40.0

40.0

C04

C12

71~80(2)

40

80

35.0

35.0

C05

C13

81~90

0.4

80

30.0

30.0

C06

C14

91~100(1)

0.4

50

30.0

30.0

C07

C15

91~100(2)

64

80

25.0

25.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

73

EPX-SERISE SERVICE MANUAL Rev. E

6.4.6 Filament Preheat Calibration Code and Data check 1) Turn off

AC switch

2) Dip switch on op board

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P05(AC110V), P06~P11(AC220V) Please check Input data. Input data and Tube current should be the same.

74

EPX-SERISE SERVICE MANUAL Rev. E Fill in the blank below table because P_DATA is different for each product.

Model

Current(mA)

P_Mode(110V)

P_Mode(220V)

60

P00

P06

50

P01

P07

40

P02

P08

35

P03

P09

30

P04

P10

25

P05

P11

3.2kW

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

75

P_DATA(A)

EPX-SERISE SERVICE MANUAL Rev. E

6.5 EPX-F4000 Please fill in the blank. kV/mAs condition(When error is occurred)

Product S/N

Error code

※ Open upper cover of portable x-ray below picture.

6.5.1 Voltage check 1) Turn on AC switch

VDC Range Pass/Fail

TP26

TP24

TP25

TP6

4.8V~5.0V

11.5V~12.5V

-11.5V~-12.5V

+3V~3.6V

Pass

Fail

Pass

Fail

Measured Voltage If it is fail case

76

Pass

Fail

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.5.2. kV, mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3)Turn on AC switch.

4) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

20msec/div

EP_FB 2.5V Pass

IP_FB 5.44V Fail

are fail, please attach your kV, mA waveform

77

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

20msec/div

EP_FB 3V Pass

IP_FB 4.08V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

20msec/div

EP_FB 3.5V Pass

IP_FB 3.4V Fail

are fail, please attach your kV, mA waveform

78

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

20msec/div

EP_FB 4V Pass

IP_FB 3.4V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 4.5V Pass

IP_FB 2.72V Fail

are fail, please attach your kV, mA waveform

79

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 5V Pass

IP_FB 2.38V Fail

Pass

Fail

are fail, please attach your kV,mA waveform

11) 110kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 110kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 110kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 5.5V Pass

IP_FB 2.04V Fail

are fail, please attach your kV,mA waveform

80

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.5.3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.

6.5.4 kV, mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture. 3) Turn on AC switch.

81

EPX-SERISE SERVICE MANUAL Rev. E 4) kV / mA PWM frequency CH 1: kV

2V/div

CH 2 : mA

2V/div

CH Frequency Pass/Fail Measured frequency

4μsec/div

CH 1 mA PWM 200kHz Pass

Fail

6.5.5 mA calibration data check.

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

82

CH 2 kV PWM 140kHz Pass Fail

EPX-SERISE SERVICE MANUAL Rev. E 3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C014(AC110V), C15~C29(AC220V) Please check Input data.

Input data and Tube current should be the same.

Tube Current Parameter

Tube Voltage

mAs

Tube Current

INPUT DATA

AC 110V

AC 220V

KV

MIN

MAX

mA

mA

C00

C15

50 (1)

0.4

16

80.0

80.0

C01

C16

50 (2)

20

32

70.0

70.0

C02

C17

50 (3)

40

100

60.0

60.0

C03

C18

51~60 (1)

0.4

32

60.0

60.0

C04

C19

51~60 (2)

40

100

50.0

50.0

C05

C20

61~70 (1)

0.4

32

50.0

50.0

C06

C21

61~70 (2)

40

100

40.0

40.0

C07

C22

71~80 (1)

0.4

10

50.0

45.0

C08

C23

71~80 (2)

13

80

40.0

40.0

C09

C24

81~90 (1)

0.4

32

40.0

35.0

C10

C25

81~90 (2)

40

80

30.0

30.0

C11

C26

91~100 (1)

0.4

32

35.0

30.0

C12

C27

91~100 (2)

40

80

25.0

25.0

C13

C28

101~110 (1)

0.4

32

30.0

25.0

C14

C29

101~110 (2)

40

80

20.0

20.0

4) If mA calibration is done, Please dip switch 4 set to OFF position.

83

EPX-SERISE SERVICE MANUAL Rev. E

6.5.6 Filament Preheat Calibration Code and Data check 1) Turn off

AC switch

2) Dip switch on op board

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P09(AC110V), P10~P19(AC220V) Please check Input data. Input data and Tube current should be the same.

84

EPX-SERISE SERVICE MANUAL Rev. E Fill in the blank below table because P_DATA is different for each product.

Model

4.0KW

Current(mA)

P_Mode(110V)

P_Mode(220V)

80

P00

P10

70

P01

P11

60

P02

P12

50

P03

P13

45

P04

P14

40

P05

P15

35

P06

P16

30

P07

P17

25

P08

P18

20

P09

P19

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

85

P_DATA

EPX-SERISE SERVICE MANUAL Rev. E

6.6 EPX-F5000 Please fill in the blank. kV/mAs condition(When error is occurred)

Product S/N

Error code

※ Open upper cover of portable x-ray below picture.

6.6.1 Voltage check 1) Turn on AC switch

VDC Range Pass/Fail

TP26

TP24

TP25

TP6

4.8V~5.0V

11.5V~12.5V

-11.5V~-12.5V

+3V~3.6V

Pass

Fail

Pass

Fail

Measured Voltage If it is fail case

86

Pass

Fail

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.6.2. kV, mA waveform check 1) Turn off AC switch 2) Please connect probe to EP_FB (kV feedback_TP11), IP_FB (mA feedback_TP8) test point on operation board below picture. 3) Turn on AC switch.

4) 40kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 40kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 40kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

20msec/div

EP_FB 2V Pass

IP_FB 5.1V Fail

are fail, please attach your kV, mA waveform

87

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 5) 50kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 50kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 50kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

20msec/div

EP_FB 2.5V Pass

IP_FB 5.1V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

6) 60kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 60kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 60kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

20msec/div

EP_FB 3V Pass

IP_FB 4.08V Fail

are fail, please attach your kV, mA waveform

88

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 7) 70kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 70kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 70kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

20msec/div

EP_FB 3.5V Pass

IP_FB 3.57V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

8) 80kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 80kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 80kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

20msec/div

EP_FB 4V Pass

IP_FB 3.06V Fail

are fail, please attach your kV, mA waveform

89

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 9) 90kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 90kV/5mAs CH 1: kV

2V/div

CH 2 : mA

2V/div

Setting value Ref voltage 90kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

40msec/div

EP_FB 4.5V Pass

IP_FB 2.04V Fail

Pass

Fail

are fail, please attach your kV, mA waveform

10) 100kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 100kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 100kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 5V Pass

IP_FB 1.79V Fail

are fail, please attach your kV,mA waveform

90

Pass

Fail

EPX-SERISE SERVICE MANUAL Rev. E 11) 110kV/5mAs waveform Please set kV, mAs below table and take exposure Exposure condition : 110kV/5mAs CH 1: kV

2V/div

CH 2 : mA

Setting value Ref voltage 110kV Pass/Fail Measured kV feedback voltage If kV,mA waveforms

2V/div

40msec/div

EP_FB 5.5V Pass

IP_FB 1.53V Fail

Pass

Fail

are fail, please attach your kV,mA waveform

6.6.3. Waveform of poor AC power condition If AC power is under 100V or 200VAC, you have below abnormal waveform. Error 015,016,017,018 is occurred by poor AC power.

91

EPX-SERISE SERVICE MANUAL Rev. E

6.6.4 kV, mA PWM frequency check 1) Turn off AC swtich 2) Please connect probe to 200kHz(TP15), 140kHz(TP16) test point on operation board below picture. 3) Turn on AC switch.

4) kV / mA PWM frequency CH 1: kV

2V/div

CH Frequency Pass/Fail Measured frequency

CH 2 : mA

2V/div

4μsec/div

CH 1 mA PWM 200kHz Pass

92

Fail

CH 2 kV PWM 140kHz Pass Fail

EPX-SERISE SERVICE MANUAL Rev. E

6.6.5 mA calibration data check.

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch.

Calibration code

Input data

Up/Down switches increase and decrease C00~C20(AC110V), C21~C41(AC220V) Please check Input data.

93

EPX-SERISE SERVICE MANUAL Rev. E Input data and Tube current should be the same.

Tube Current Parameter AC110V AC220V

Tube Voltage

mAs

Tube Current

INPUT DATA

KV

MIN

MAX

mA

mA

C00

C21

50(1)

0.4

10

100

100

C01

C22

50(2)

13

32

80

80

C02

C23

50(3)

40

100

70

70

C03

C24

51~60(1)

0.4

10

80

80

C04

C25

51~60(2)

13

32

70

70

C05

C26

51~60(3)

40

100

60

60

C06

C27

61~70(1)

0.4

10

70

70

C07

C28

61~70(2)

13

32

60

60

C08

C29

61~70(3)

40

100

50

50

C09

C30

71~80(1)

0.4

10

60

60

C10

C31

71~80(2)

13

32

50

50

C11

C32

71~80(3)

40

100

40

40

C12

C33

81~90(1)

0.4

10

50

50

C13

C34

81~90(2)

13

32

40

40

C14

C35

81~90(3)

40

100

30

30

C15

C36

91~100(1)

0.4

10

45

45

C16

C37

91~100(2)

13

32

40

40

C17

C38

91~100(3)

40

100

30

30

C18

C39

101~110(1)

0.4

10

40

40

C19

C40

101~110 (2)

13

32

30

30

C20

C41

101~110 (3)

40

100

20

20

4) If mA calibration is done, Please dip switch 4 set to OFF position.

94

EPX-SERISE SERVICE MANUAL Rev. E

6.1.6 Filament Preheat Calibration Code and Data check 1) Turn off

AC switch

2) Dip switch on op board

DIP Switches are used for following setting. SW #1: ERROR Enable (ON)/Disable (OFF) SW #2: Reserved SW #3: Initialize memories (reserved) SW #4: Calibration mode (ON)/Normal mode (Off) For normal operation, only the DIP SW#1 must be set to ON position. DIP SWITCH #4 on the OP BOARD PCB is set to ON position. Then the display should indicate as follows,

3) Turn on AC switch. 4) In mA calibration mode, Press the number 1 switch as shown in figure. Mode is converted.

P_Calibration code

Input data

Up/Down switches increase and decrease P00~P04(AC110V), P05~P09(AC220V) Please check Input data. Input data and Tube current should be the same.

95

EPX-SERISE SERVICE MANUAL Rev. E Fill in the blank below table because P_DATA is different for each product.

Model

5kW

Current(mA)

P_Mode(110V)

P_Mode(220V)

100

P00

P09

80

P01

P10

70

P02

P11

60

P03

P12

50

P04

P13

45

P05

P14

40

P06

P15

30

P07

P16

25

P08

P17

5) If Preheat calibration is done, Please dip switch 4 set to OFF position.

96

P_DATA(A)