24 1 809KB
LLC “PKF Geotekh”
Down-hole self-contained combined tool GEO-6
Technical description Operating instruction Edition from 06/27/2017
Neftekamsk
2017
Content. 1. Introduction .............................................................................................................................................. 3 2. Technical characteristics of the tool ........................................................................................................ 3 3. Design and description of the tool. .......................................................................................................... 4 4. Testing of working ability, start/stop record of tool. ............................................................................... 5 5. Order of data reading. .............................................................................................................................. 6 6. Location of the tool’s channels. ............................................................................................................... 6 7. Rules of operation and transporting the tool. ......................................................................................... 6 8. Power supply. ............................................................................................................................................ 7 9. Package Contents ................................................................................................................................... 10 10. Warranty. .............................................................................................................................................. 10 11. Possible faults and troubleshooting. .................................................................................................... 10 12. Questions and Answers. ....................................................................................................................... 11 Appendix 1 Assembly of GEO-6................................................................................................................. 12 Appendix 2. Flowmeter. ............................................................................................................................. 13 Appendix 3. Procedure for filling of wire to the wire tip of the tool. ....................................................... 14 Appendix 4. The scheme of interface cables. ............................................................................................ 15
2
1. Introduction Down-hole self-contained combined tool GEO-6 makes registration of 6 parameters: pressure P, temperature T, gamma-ray GR, fluid flow-meter Q, collar locator CCL, thermoindication STI. Using the tool allows you to: To produce a research of injection wells in operation at a pressure at the wellhead without stopping of the injection and the formation water spout. Research the wellbore, including the upper fresh water aquifers. Receive the log well data from PC directly at the well. Reduce the duration of the well research significantly due to the exclusion of works to pressure relief and upload of recovery injection mode. Information is received from the sensors to the memory block of down-hole tool where stored for measuring time. Power is supplied by a lithium battery or accumulator. After working the tool information is read into the PC. Then you can explore the results and control data on the screen as a standard log data in the time format. After combining of tool GEO-6 data with depth data it can be formed a single file of depth format (LAS). If necessary, all the data is issued to the printer in the standard form.
2. Technical characteristics of the tool Number of recording channels ...................................................................................... from 1 to 6 Pressure channel (P): Range of measuring, kgf/cm2 ........................................................................... 0-400,0-600,0-800 Number of ADC bits ................................................................................................................... 16 Absolute error, kgf/cm2 ............................................................................................... 0.6, 0.9, 1.2 Relative error, % ..................................................................................................................... 0.15 Resolution, kgf/cm2 ........................................................................................ 0.006, 0.009, 0.012 Resolution, % ...................................................................................................................... 0.0015 Temperature channel (T): Range of measuring, °С ................................................................................. 0-100, 0-120, 0-150 Number of ADC bits ................................................................................................................... 16 Absolute error, °С ..................................................................................................................... 0.5 Resolution, °С ....................................................................................................................... 0.003 Time constant, с ....................................................................................................................... 1.2 Gamma ray channel (GR): Measuring range, mR/h ........................................................................................................ 0-100 Sensitivity, imp/min per 1 mR/h, not less ................................................................................. 150 Relative error +/-, % ................................................................................................................... 15 Conversion rate imp/min to mR/h (for 38mm) ...................................................................... 0.006 Case Collar Locator channel (CCL): Signal/noise ratio, not less ........................................................................................................ 5:1 Thermoindication channel (STI): Supply voltage, V ...................................................................................................................... 5-6 The time constant, sec ................................................................................................................. 3 Current consumption, mA .................................................................................................. 80-100 An overheat sensor in the still water, °С not less ........................................................................ 1 Upper channel inline flowmeter (Q optional): Measuring range, m3/h ......................................................................................................... 3-200 Relative error, % .......................................................................................................................... 5 Bottom channel flowmeter (Q optional): Measuring range, m3/h ......................................................................................................... 3-200 Relative error, % .......................................................................................................................... 5 Water capacity hold-up meter (optional): Measuring range, % .................................................................................................... от 0 до 100 Absolute error, % ................................................................................................................... +/-10 3
Tool memory: The amount of tool memory ...................................... 4224Kb or 2162688 points (if one channel) The time of information storage, years ...................................................................................... 10 Number of record cycles to the internal memory ................................................................. 10000 Time characteristics: Cycle for 4 firmware version (measurement period), sec............................................... 0.125-127 Cycle for 3 firmware version (measurement period), sec............................................ 0.300-0.992 Delay time before beginning of record, min ....................................................................... to 9999 Interface parameters: PC port for communication *. ............................................................................................. RS-232 Speed rate, bit/sec. ............................................................................................................ 115200 * the delivery includes an cable-adapter for the USB port of a PC Tool power: Power supply: (two lithium batteries 3V with the capacity of 1-1.5 A/h or accumulator 3.6-4.2V with the capacity of 2.6-2.9 A/h) The minimum work voltage, V........................................................ 5.0 batteries(3.1-accumulator) Current consumption in delayed mode, mА .............................................................................. 0.1 Current consumption in recording mode, Without GR and STI, mА .................................................................................................. 5 With GR without STI, mA ............................................................................................... .25 All channel, mA ...................................................................................................... 105-125 Time of working tool (with 1.5A/h batteries) Without GR and STI, hours........................................................................................... 140 With GR without STI, hours ............................................................................................ 56 All channel, hours ........................................................................................................... 11 The software allows you to work in the Windows operating system and provides the following main functions: -switching on(with delay)/off the tool in the recording mode using a computer (by Geo6.exe); -show the actual values of all channels to check the tool computer (by Geo6.exe); -specifying different measurement programs (