Radiography X-ray System

DM-5125 Operation Manual

509-ho, Unitechvil, 1141-2, Baegsuk-dong, Ilsandong-ku, Goyang-si, Gyeonggi-do, Republic of Korea Tel : 82-31-907-7222, Fax : 82-31-904-2231, E-mail : [email protected]

Operation Manual: DM-5125

TABLE OF CONTENTS

PART 1. Generator 1.

DM-100[12.5kW]

----------------------------------------------------------------

2.

DM-200, 300[15kW]

----------------------------------------------------------------

3.

20kW/32kW/40kW/50kW/80kW

----------------------------------------------------------------

4.

DIG-650RF[50kW 2-Tube]

----------------------------------------------------------------

PART 2. X-ray Tube 1.

E7132X

----------------------------------------------------------------

2.

E7239X

----------------------------------------------------------------

3.

E7240X

----------------------------------------------------------------

4.

E7252X

----------------------------------------------------------------

PART 3. Bucky Table 1.

DM-FT-1 [Fixed Table-Top Table]

---------------------------------------------------------------

2.

DM-BT-1 [4-way Table]

---------------------------------------------------------------

3.

DM-RFT-5[RF Tilting Table]

---------------------------------------------------------------

PART 4. Tube Stand 1.

DM-FCT-1[Floor to Ceiling Mounted]

----------------------------------------------------------------

PART 5. Bucky Stand 1.

VB-1[Bucky Stand]

----------------------------------------------------------------

PART 6. High Tension Cable 1.

High Tension Cable

----------------------------------------------------------------

PART 7. Collimator 1.

LB-50[Collimator]

----------------------------------------------------------------

2/146

Operation Manual: DM-5125

PART 1. Generator

1. DM-100[12.5kW Generator]

3/146

Operation Manual: DM-5125

NOTICE 1 This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines thoroughly before your using the generator. The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which many results from the use of this manual. Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. Always keep the manual at hand for your reference.

NOTICE 2 1. This equipment should be used only by the legally qualified persons and practitioners. 2. For the safe installation a. Please avoid wet and humid locations. b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. c. Protect the product from slope, vibration, and shock. d. Do not install the product where chemicals are stored or gases are generated. e. Check the frequency, voltage, and current of power supply. f. Check the earth ground power about the electric discharge and polarity. 3. Before the operation of the machine a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates. b. Check the safety of the ground power connections. c. Check the correctness and safety of the cable and plug connections. d. Be careful that mixed use of the machine may give wrong diagnostic results and damages. e. Check the exterior circuits, which occur direct physical contact to the patient. f. Check the earth ground power about the electric discharge and polarity. 4. During the operation of the machine a. Do not exceed the designated time for diagnosis and cures. b. Monitor the conditions of device and influence on the patient. c. When you find any troubles about generators or patients, stop the operation of the machine and take appropriate measures for the safety of the patient. d. Be careful that the patient should not touch the device.

4/146

Operation Manual: DM-5125

5. After the operation of the machine a. Turn off the power of the machine after returning the operation switches and scales to the original position in accordance with the prescribed regulation. b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections. c. For the safe storage and installation d. Please avoid wet and humid locations. e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. f. Protect the product from slope, vibration, and shock. g. Do not install the product where chemicals are stored or gases are generated. 6. If the device is broken, mark proper description and ask repairs to the engineer. 7. Do not modify the machine. 8. Inspections and maintenance a. All device and parts should be checked on the regular basis. b. If you operate the device after long period of non-use, check the safety and conditions of the machine. 9. Operate the machine in correct ways according to the manuals. 10. Be careful about following cautions a. Wear the lead apron in order to prevent the x-ray irradiation. b. Regular inspection should be executed about leakage dose and irradiation meter. c. Be careful about the x-ray exposures. d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1 1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate measures. 2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off the power during operation. 3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions.

5/146

Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2 1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic drugs, oxygen and hydrogen near the working locations. 2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must be checked after the occurrence of the earthquake. 3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use this machine” and ask repairs to the engineers. 4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE 1. Power on and off Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the resupply of the power. 2. Regarding the operation environment a. X-ray room Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is designed to be protective against humidity and dust. However, if you use the machine at the place with high humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts. b. Do not leave the machine exposed to the strong direct sunlight. 3. Exposure of the x-rays a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions. b. Take enough cares in order not to make excessive radiation on the same body locations. 4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS Do not use the machine with following equipment. *The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The officials in hospital ought to give enough advises and education for the operator of the machine, doctors, patient and other the third parties.

6/146

Operation Manual: DM-5125

SYMBOLS AND TERMS The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that

DANGER

cause heavy damages to people and even the death. It warns of the conditions and situations, if not taken over or avoided, that

WARNING

could cause heavy damages to people or cause unrepeatable damages to the equipment. It warns of the conditions and situations, if not taken over or avoided, that

CAUTION

could cause damages to people, or to the equipment.

WAIVER CLAUSE Dongmun do not take responsibility for the following causes for damage and repairs. 1. Any installation, movement, modification and repairs by other engineering services, which was not designated by Dongmun. 2. Troubles and causes to our machine incurred by other equipment. 3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun. 4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual. 5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the powers, installation and working places. 6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders. 7. Accident and damage happened by the use beyond the original purpose of the machine. 8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the conditions and results of the medical prescriptions.

7/146

Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY 1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from normal use and operation of product. 2. Service fee is charged for the following cases. - Malfunctioning after the period of warranty - Disorder caused by fire, earthquake, or lightening - Inappropriate movement or negligence of product - Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones. 3. Note that defects not related to the main functions of the product are not eligible for this warranty. 4. Requirements for Service - Stop using the product when a problem is found, and check the manual for a solution. - Turn off the product before service request, and check the model number, serial number, the date of purchase before calling the service center. 5. Defects or depreciation in appearance is not eligible for refund or change. 6. Dongmun is not responsible for indirect damage related to the use of products. 7. Dongmun is not responsible for injury or damage after the end of warranty period. 8. All detailed adaptability warranties and other warranties are subjected to this statement.

8/146

Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION 1-1. COMPOSITION The electric composition block diagram of DIG-Series is shown in Fig 1-1.

AC Input

EMI Filter

AD/DC Converter

HF DC/AC Inverter

HV Tank X-Ray Tube

IGBT Driver

Starter Operation Panel

HT-Controller

Transformer

Power

Collimator

Filament Interface

Fig 1-1 COMPOSITION OF THE HFG AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit.

This DC voltage is

inverted in high-frequency voltage of about 20kHz by IGBT inverter. The high-frequency voltage is converted into the 40kV to 125kV DC high voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current. The operation panel can operate selection in each mode of operation and control, selection and setting of conditions. These operation conditions are transmitted by the communication system from OP-CONT to HTCONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.

9/146

Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel

High Voltage Generator Main CPU

A/D Converter

D/A Converter

kV F/B

PWM

kV & mA Feedback Amplifier

Filament mA F/B

PWM

RMS DC Converter

Drive Circuit

Filament Inverter

IGBT

HV Transformer Filament Transformer

A

kV mA

C FILAMENT

X-RAY Tube

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

10/146

Operation Manual: DM-5125

1-2. SPECIFICATION 1

Rating Out-put

12.5 kW

2

Tube Voltage kV

40 ~ 125kV

3

Tube Current mA

20S, 40S, 75S,100S mA

4

‘kV Accuracy

±10 %

5

‘mA Accuracy

±15 %

6

mAs Range

0.4 ~ 320 mAs

7

Digital Display

Digital display (LED & 7 - Segment)

8

Setting Method

Sheet Push Switch

9

Self-Diagnosis

Code display in LED digital

10

Selection

Manual / APR

11

APR

15 Region

(1kV step incr./Decr.)

Display (OP panel)

12

‘kV

3 Digital LED

‘mAs

3 Digital LED

Selection / Setting

LED (12)

Alarm

LED (1)

Weight 13

HFG

14 kg

Op panel

1.5 kg

1-3. ELECTRICAL REQUIREMENT Model

DIG SERIES

Line Nominal Voltage

Single Phase 200V / 220V / 240V

Automatic Line Compensation Range

±10 %

Frequency

50 Hz / 60 Hz

Line Impedance

Less Than 0.12 Ω

Grounding conditions

Based on the local regulation or 3 independent ground circuit

1-4. PROPER ENVIRONMENT 1. Avoid following places for the normal operation and safe storage. a. Where the equipment is exposed to water vapor. b. Where the equipment is exposed to direct sunlight. c. Where the equipment is exposed to dust. d. Where the equipment is exposed to high humidity. e. Where there is a ventilation problem. f. Where the equipment is exposed to salty atmosphere.

11/146

Operation Manual: DM-5125

g. Where the equipment is exposed to chemicals or gas. 2. Keep away from the place with strong vibration and maintain proper environment and conditions. OPERATION ENVIRONMENT Temperature range

+ 10°C ~ + 40°C,

Relative Humidity Range

30% ~ 75% RH (No dews)

Others

No Direct Sunlight No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere. STORAGE AND TRANSPORTATION Temperature range

-25°C ~ +60°C

Relative Humidity Range

10% ~ 95% RH(No dews)

Air Pressure

700hPa ~ 1060hPa

Others

No corrosive gases

12/146

Operation Manual: DM-5125

1-5. WIRE CONNECTION 1) FRONT PART

① (CN204) PIN No.

Connect to OP

② (TB502)

③ (TB3)

④ (TB1)

Connect to External

Main Power

Connect to Main

Device

Selection

Power

1

Rx+

AC 12V

AC 240V

AC INPUT

2

Rx-

AC 12V RTN

AC 220V

F.G

3

Tx+

AC 100V

AC 200V

AC INPUT

4

Tx-

AC 100V RTN

5

EXP

BUCKY

6

GND

BUCKY RTN

7

GND

EXT

8

DC 24V

EXT RTN

9

DC 24V

10

DC 24V RTN

11

SWITCH

12

SWITCH RTN

13/146

Operation Manual: DM-5125

2) REAR PART

⑤ (CN107)

⑥ (CN108)

⑦ (TB102)

Connect to

Connect to

Connect to TUBE

TANK

TANK

(ROTOR)

1

EP F/B

COM

ROTOR1

OUTPUT

2

IP F/B

L1

ROTOR2

OUTPUT

3

GND

L2

ROTOR3

4

GND

S1

5

INTERLOCK

S2

6

NC

PIN No.

14/146

⑧ (TB2) Connect to TANK

Operation Manual: DM-5125

1-6. ADDITIONAL FUNCTIONS DIP S/W FUNCTION 1) REMOTE HT-CONTROLLER DIP S/W No. DIP S/W 1

Function

Remark

Not defined Focus Selection

DIP S/W 2

On

Large Focus

Off

Small Focus

Apply to mass chest x-ray DIP S/W 3

On

Use mass chest x-ray

Off

Not use mass chest x-ray

Bucky Type Selection DIP S/W 4

On

Motor Type Bucky

Off

Solenoid Type Bucky

2) OPERATION PANEL DIP S/W No.

Function

DIP S/W 1

Current Data Calibration Mode

DIP S/W 2

Preheat Data Calibration Mode

DIP S/W 3

Serial EEPROM Initialize Mode

DIP S/W 4

RAM Monitor Mode

15/146

Remark

Operation Manual: DM-5125

SECTION 2. DISCRIPTIONS / FUNCTIONS 2-1. OUTLINE AND DIMENSIONS

FIG 2-1. DIMENSION OF HFG

FIG 2-2. DISPLAY OF OP PANEL

16/146

Operation Manual: DM-5125

2-2. OPERATION PANEL DISPLAY

FIG. 2-3 OP PANEL DISCRIPTION

1

Size Display / Selection Switch (Large)

12

Exposure wait display (Yellow)

2

Size Display / Selection Switch (Middle)

13

Error Display (Red)

3

Size Display / Selection Switch (Small)

14

Tube kV Display

4

APR-1 Display / Selection Switch

15

mAs Display

5

APR-2 Display / Selection Switch

16

Tube kV Down Switch

6

APR-3 Display / Selection Switch

17

Tube kV Up Switch

7

APR-4 Display / Selection Switch

18

mAs Down Switch

8

APR-5 Display / Selection Switch

19

mAs Up Switch

9

APR Storage Switch

20

Bucky Display/Selection Switch

10

X-ray exposure ready display (Green)

21

X-ray exposure ready Switch

11

X-ray exposure Display (Yellow)

22

X-ray exposure Switch

17/146

Operation Manual: DM-5125

1) SIZE DISPLAY/SELECTION SWITCH (LARGE)

Make selection for body size “large” under APR radiography. Choose the APR No. 1 to 5 after the selection of Body size. 2) SIZE DISPLAY/SELECTION SIWTCH (MIDDLE)

Make selection for body size “Middle” under APR radiography. Choose the APR No. 1 to 5 after the selection of Body size. 3) SIZE DISPLAY/SELECTION SIWTCH (SMALL)

Make selection for body size “Small” under APR radiography. Choose the APR No. 1 to 5 after the selection of Body size. 4) APR-1 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography. The saved kV and mAs will be displayed on each segment. 5) APR-2 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography. The saved kV and mAs will be displayed on each segment. 6) APR-3 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography. The saved kV and mAs will be displayed on each segment.

18/146

Operation Manual: DM-5125

7) APR-4 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography. The saved kV and mAs will be displayed on each segment. 8) APR-5 DISPLAY / SELECTION SWITCH

Choose after the selection of body size under the APR radiography. The saved kV and mAs will be displayed on each segment. 9) APR STORAGE SWITCH

Save the changed value of kV and mAs for APR. If this switch is pushed longer than 0.5 sec after the selection of kV and mAs, the data will be saved with the buzzer sound. 10) X-RAY EXPOSURE READY LED (GREEN LED) This LED is lighted when x-ray exposure is ready. 11) X-RAY EXPOSURE LED (YELLOW LED) This LED is lighted during the x-ray exposure. 12) EXPOSURE WAIT LED (YELLOW LED) No function. 13) ERROR LED This LED is lighted during the occurrence of Error. 14) TUBE kV DISPLAY Indicate the kV of radiography. 15) mAs DISPLAY Indicate the mAs of radiography.

19/146

Operation Manual: DM-5125

16) TUBE kV DOWN SWITCH Operate the setting of tube kV. Move 1 kV by pressing the switch one time and will be changed consecutively by keep pressing the button. Setting range is 40~125kV. 17) TUBE kV UP SWITCH Operate the setting of tube kV. Move 1 kV by pressing the switch one time and will be changed consecutively by keep pressing the button. Setting range is 40~125kV. 18) mAs DOWN SWITCH Operate the setting of mAs for radiography. Setting range is 0.4~320 mAs. 19) mAs UP SWITCH Operate the setting of mAs for radiography. 20) BUCKY DISPLAY / SELECTION SWITCH

Select and display the sort of Bucky. The LED will be turned on during Bucky 1 and turned off at non-bucky. 21) X-RAY EXPOSURE READY SWITCH

When the ready switch is pushed, it rotate the rotor, control the heating of filament and make the exposure at ready condition. 22) X-RAY EXPOSURE SWITCH (X-RAY SWITCH)

Make the exposure of x-rays by pressing this siwtch after the ready LED is displayed. Touch off from the button after the radiography is completed.

20/146

Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE 3-1. SELECTION OF THE RADIOGRAPHY METHOD There are radiography modes shown below in this equipment. 1. Manual radiography mode Radiography conditions are performed by manual setup. 2. APR radiography mode You can select the radiography region; direction and patient size on the control panel and make x-ray exposure with automatically set radiography conditions. Each radiography conditions can be changed and saved.

3-2. SETTING MANUAL RADIOGRAPHY 1. Select the manual radiography mode Check if the APR mode is chosen. If they are selected in APR mode, please press APR setting switch, and release the setup conditions. 2. Setting tube voltage "kV" Set up the tube voltage with “kV” setting switch. Setting range: 40kV – 125 kV (1kv step) 3. Setting tube current “mA” Set up the tube current with “mA” setting switch. Setting range: 0.4~320 mAs 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.4, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320 mAs

3-3. SETTING APR RADIOGRAPHY 1. Selection of APR mode APR mode is selected by size and number of APR. 2. Selection of Radiography Regions Choose the [L], [M], [S] radiography with setting switch. 3. Selection of APR numbers Choose the number out of No. 1~5 after the selection of the size

21/146

Operation Manual: DM-5125

4. Change and storage Change the settings with tube kV down and up switch after the selection of APR according to the size and APR no. 3-4. RADIOGRAPHY 1. PREPARATION After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is ended. 2. X-RAY RADIOGRAPHY BY RADIOGRAPHY SWITCH ON THE PANEL 1) Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is completed. 2) If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be started. 3) X-rays exposure LED will be lighted on and the buzzer sounds during the exposure. 3. X-ray Radiography Using Hand-Switch 1) Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when it is prepared to make exposure. 2) When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started. 3) During exposure, the exposure LED will be lighted on and the buzzer sounds. 4) If the x-ray exposure LED is lighted or the buzzer is off, touch off from the switch and close the radiography procedures.

CAUTION

If you touch out from the hand switch during exposure, the x-ray shooting is stopped and you cannot get clear pictures. Please touch off your hand after the x-ray exposure signs and the buzzer sound are turned off except the case you want to stop the x-ray exposure urgently

22/146

Operation Manual: DM-5125

PART 1. Generator

2. DM-200[15kW Generator]

23/146

Operation Manual: DM-5125

NOTICE 1 This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines thoroughly before your using the generator. The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which many results from the use of this manual. Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. Always keep the manual at hand for your reference.

NOTICE 2 1. This equipment should be used only by the legally qualified persons and practitioners. 2. For the safe installation a. Please avoid wet and humid locations. b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. c. Protect the product from slope, vibration, and shock. d. Do not install the product where chemicals are stored or gases are generated. e. Check the frequency, voltage, and current of power supply. f. Check the earth ground power about the electric discharge and polarity. 3. Before the operation of the machine a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates. b. Check the safety of the ground power connections. c. Check the correctness and safety of the cable and plug connections. d. Be careful that mixed use of the machine may give wrong diagnostic results and damages. e. Check the exterior circuits, which occur direct physical contact to the patient. f. Check the earth ground power about the electric discharge and polarity. 4. During the operation of the machine a. Do not exceed the designated time for diagnosis and cures. b. Monitor the conditions of device and influence on the patient. c. When you find any troubles about generators or patients, stop the operation of the machine and take appropriate measures for the safety of the patient. d. Be careful that the patient should not touch the device.

24/146

Operation Manual: DM-5125

5. After the operation of the machine a. Turn off the power of the machine after returning the operation switches and scales to the original position in accordance with the prescribed regulation. b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections. c. For the safe storage and installation d. Please avoid wet and humid locations. e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. f. Protect the product from slope, vibration, and shock. g. Do not install the product where chemicals are stored or gases are generated. 6. If the device is broken, mark proper description and ask repairs to the engineer. 7. Do not modify the machine. 8. Inspections and maintenance a. All device and parts should be checked on the regular basis. b. If you operate the device after long period of non-use, check the safety and conditions of the machine. 9. Operate the machine in correct ways according to the manuals. 10. Be careful about following cautions a. Wear the lead apron in order to prevent the x-ray irradiation. b. Regular inspection should be executed about leakage dose and irradiation meter. c. Be careful about the x-ray exposures. d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1 1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate measures. 2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off the power during operation. 3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions.

25/146

Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2 1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic drugs, oxygen and hydrogen near the working locations. 2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must be checked after the occurrence of the earthquake. 3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use this machine” and ask repairs to the engineers. 4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE 1. Power on and off Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the resupply of the power. 2. Regarding the operation environment a. X-ray room Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is designed to be protective against humidity and dust. However, if you use the machine at the place with high humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts. b. Do not leave the machine exposed to the strong direct sunlight. 3. Exposure of the x-rays a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions. b. Take enough cares in order not to make excessive radiation on the same body locations. 4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS Do not use the machine with following equipment. *The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The officials in hospital ought to give enough advises and education for the operator of the machine, doctors, patient and other the third parties.

26/146

Operation Manual: DM-5125

SYMBOLS AND TERMS The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that

DANGER

cause heavy damages to people and even the death. It warns of the conditions and situations, if not taken over or avoided, that

WARNING

could cause heavy damages to people or cause unrepeatable damages to the equipment. It warns of the conditions and situations, if not taken over or avoided, that

CAUTION

could cause damages to people, or to the equipment.

WAIVER CLAUSE Dongmun do not take responsibility for the following causes for damage and repairs. 1. Any installation, movement, modification and repairs by other engineering services, which was not designated by Dongmun. 2. Troubles and causes to our machine incurred by other equipment. 3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun. 4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual. 5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the powers, installation and working places. 6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders. 7. Accident and damage happened by the use beyond the original purpose of the machine. 8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the conditions and results of the medical prescriptions.

27/146

Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY 1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from normal use and operation of product. 2. Service fee is charged for the following cases. - Malfunctioning after the period of warranty - Disorder caused by fire, earthquake, or lightening - Inappropriate movement or negligence of product - Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones. 3. Note that defects not related to the main functions of the product are not eligible for this warranty. 4. Requirements for Service - Stop using the product when a problem is found, and check the manual for a solution. - Turn off the product before service request, and check the model number, serial number, the date of purchase before calling the service center. 5. Defects or depreciation in appearance is not eligible for refund or change. 6. Dongmun is not responsible for indirect damage related to the use of products. 7. Dongmun is not responsible for injury or damage after the end of warranty period. 8. All detailed adaptability warranties and other warranties are subjected to this statement.

28/146

Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION 1-1. COMPOSITION The electric composition block diagram of DIG-Series is shown in Fig 1-1.

AC Input

EMI Filter

AD/DC Converter

HF DC/AC Inverter

HV Tank X-Ray Tube

IGBT Driver

Starter Operation Panel

HT-Controller

Transformer

Power

Collimator

Filament Interface

Fig 1-1 COMPOSITION OF THE HFG AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit.

This DC voltage is

inverted in high-frequency voltage of about 20kHz by IGBT inverter. The high-frequency voltage is converted into the 40kV to 125kV DC high voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current. The operation panel can operate selection in each mode of operation and control, selection and setting of conditions. These operation conditions are transmitted by the communication system from OP-CONT to HTCONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.

29/146

Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel

High Voltage Generator Main CPU

A/D Converter

D/A Converter

kV F/B

PWM

kV & mA Feedback Amplifier

Filament mA F/B

PWM

RMS DC Converter

Drive Circuit

Filament Inverter

IGBT

HV Transformer Filament Transformer

A

kV mA

C FILAMENT

X-RAY Tube

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

30/146

Operation Manual: DM-5125

1-2. SPECIFICATION Output Power

15 kW

kV Range

40 ~ 125kV

mA Range

Small Focus

50, 100, 150

Large Focus

200

(1kV step incr./Decr.)

Time Range

0.004 ~ 6.0sec in 33 Steps

mAs Range

0.4 ~ 320mAs in 30 Steps

kV Accuracy

±10 %

mA Accuracy

±15 %

Digital Display

Digital Display (LED & 7 - Segment)

Self-Diagnosis

Code Display in LED

APR

30 Memories

Display

Dimension

Weight

kV

3 Digital LED

mA

3 Digital LED

Sec/mAs

4 Digital LED

Generator

280W X 370H X 452D

Operator Panel

161W X 53H X 367D

Generator

14Kg

Operator Panel

2Kg

1-3. ELECTRICAL REQUIREMENT Model

DIG SERIES

Line Nominal Voltage

Single Phase 200V / 220V / 240V

Automatic Line Compensation Range

±10 %

Frequency

50 Hz / 60 Hz

Line Impedance

Less Than 0.12 Ω

Grounding conditions

Based on the local regulation or 3 independent ground circuit

1-4. PROPER ENVIRONMENT 1. Avoid following places for the normal operation and safe storage. a. Where the equipment is exposed to water vapor. b. Where the equipment is exposed to direct sunlight. c. Where the equipment is exposed to dust. d. Where the equipment is exposed to high humidity. e. Where there is a ventilation problem. f. Where the equipment is exposed to salty atmosphere.

31/146

Operation Manual: DM-5125

g. Where the equipment is exposed to chemicals or gas. 2. Keep away from the place with strong vibration and maintain proper environment and conditions. OPERATION ENVIRONMENT Temperature range

+ 10°C ~ + 40°C,

Relative Humidity Range

30% ~ 75% RH (No dews)

Others

No Direct Sunlight No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere. STORAGE AND TRANSPORTATION Temperature range

-25°C ~ +60°C

Relative Humidity Range

10% ~ 95% RH(No dews)

Air Pressure

700hPa ~ 1060hPa

Others

No corrosive gases

32/146

Operation Manual: DM-5125

1-5. WIRE CONNECTION 1) FRONT PART

12

CN204

1 TB502 TB1

TB3 1

1

3

3

① (CN204) PIN No.

Connect to OP

② (TB502)

③ (TB3)

④ (TB1)

Connect to External

Main Power

Connect to Main

Device

Selection

Power

1

Rx+

AC 12V

AC 240V

AC INPUT

2

Rx-

AC 0V

AC 220V

F.G

3

Tx+

SWITCH

AC 200V

AC INPUT

4

Tx-

SWITCH RTN

5

EXP

AC 100V(Bucky 2)

6

GND

AC 0V(Bucky 2)

7

GND

AC 100V(Bucky 1)

8

DC 24V

AC 0V(Bucky 1)

9

EXP 100V

10

EXP 0V

11

DC 24V

12

DC 0V

33/146

Operation Manual: DM-5125

2) REAR PART

3

TB102

1 CN108 CN107

TB2 2 1

⑤ (CN107)

⑥ (CN108)

⑦ (TB102)

Connect to

Connect to

Connect to TUBE

TANK

TANK

(ROTOR)

1

EP F/B

COM

ROTOR1

OUTPUT

2

IP F/B

L1

ROTOR2

OUTPUT

3

GND

L2

ROTOR3

4

GND

S1

5

INTERLOCK

S2

6

NC

PIN No.

34/146

⑧ (TB2) Connect to TANK

Operation Manual: DM-5125

1-6. ADDITIONAL FUNCTIONS DIP S/W FUNCTION 1) REMOTE HT-CONTROLLER DIP S/W No.

Function

DIP S/W 1

Not Defined

DIP S/W 2

Not Defined

Remark

Apply to mass chest x-ray DIP S/W 3

ON

Use mass chest x-ray

OFF

Not use mass chest x-ray

Bucky Type Selection DIP S/W 4

ON

Motor Type Bucky

OFF

Solenoid Type Bucky

2) OPERATION PANEL DIP S/W No. DIP S/W 1 DIP S/W 2 DIP S/W 3 DIP S/W 4 DIP S/W 5 DIP S/W 6 DIP S/W 7 DIP S/W 8

Function ON

Calibration Mode

OFF

Normal Mode

ON

-

OFF

-

ON

-

OFF

-

ON

RAM Monitor Mode

OFF

Normal Mode

ON

-

OFF

-

ON

-

OFF

-

ON

-

OFF

-

ON

X-ray END Buzzer

OFF

X-ray ON Buzzer

35/146

Remark

Operation Manual: DM-5125

SECTION 2. DISCRIPTIONS / FUNCTIONS 2-1. OUTLINE AND DIMENSIONS

FIG 2-1. DIMENSION OF HFG

FIG 2-2. DISPLAY OF OP PANEL

36/146

Operation Manual: DM-5125

2-2. OPERATION PANEL DISPLAY

1

12

2 15

3

4 16

5

6

7

8

9

10

14

13

11

17

FIG. 2-3 OP PANEL DISCRIPTION

1

Patient Size Selection Switch

10

APR Region Selection Switch & Lamp

2

“Over” Lamp

11

APR Direct Selection Switch

3

“Error” Lamp

12

kV Setting Switch & Display

4

Small Focus Lamp

13

Backup Release Switch

5

Large Focus Lamp

14

Bucky Selection Switch & Display

6

X-ray Ready Lamp

15

mA Selection Switch & Display

7

X-ray Exposure Lamp

16

mAs/Sec Setting Switch & Display

8

mA/Sec Selection Switch & Lamp

17

X-ray Ready Switch

9

APR Mode Selection Switch

18

X-ray Exposure Switch

37/146

18

Operation Manual: DM-5125

1) Patient Size Selection Switch It operates the selections of patient size. The selected size will be displayed on the segment. Save the changed APR conditions by pressing the size switches. You can set the exposure conditions (kV, mA and mAs/Sec) by selecting the exposure regions, direction and patient size. Each condition can be changed and stored by the exposure size switch. 2) “Over” Lamp The over lamp is lighted on when the exposure conditions exceeds the normal standards. You cannot make exposure while the over lamp is lighted on. 3) “Error” Lamp The lamp is lighted in case of errors. 4) Small Focus Lamp The lamp is lighted on when the small focus is selected. 5) Large Focus Lamp The lamp is lighted on when the large focus is selected. 6) X-ray Ready Lamp The X-ray ready lamp is lighted on when it is at the ready condition. 7) X-ray Exposure Lamp The X-ray exposure lamp is lighted on during the X-ray exposure. 8) mAs/Sec Selection Switch & Lamp The mAs or Sec radiography mode is selected. The selected radiography mode is displayed with lamp. 9) APR Mode Selection Switch It operates the on/off of the APR mode. 10) APR Region Selection Switch & Lamp It operates the selection of APR regions and the selected location will be displayed on the lamp. 11) APR Direction Selection Switch It operates the selection of ARP directions the selected direction will be displayed in the lamp. 12) kV Setting Switch & Display Up-button increases the kV value of X-ray tube whereas down-button decreases kV. If you keep pressing buttons, the value of kV will increase or decrease continuously.

38/146

Operation Manual: DM-5125

13) Backup Release Switch Release the back-up conditions. 14) Bucky Selection Switch & Display Select and display Bucky 1 (Table) or Bucky 2 (Stand). 15) mA Setting Switch & Display Up-button increases the mA value of X-ray tube whereas down-button decreases mA. If you keep pressing buttons, the value of mA will increase or decrease continuously. 16) mAs/Sec Setting Switch & Display Up-button increases the mAs/Sec value of X-ray tube whereas down-button decreases mAs/Sec. If you keep pressing buttons, the value of mAs/Sec will increase or decrease continuously. 17) X-ray Ready Switch It circulates the rotor and controls the heat of the filament that makes ready actions for the X-ray exposure. 18) X-ray Exposure Switch Press the X-ray exposure switch after the [6] X-ray ready lamp is lighted on.

39/146

Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE 3-1. SELECTION OF THE RADIOGRAPHY METHOD There are radiography modes shown below in this equipment. 1. Manual radiography mode Radiography conditions are performed by manual setup. 2. APR radiography mode You can select the radiography region; direction and patient size on the control panel and make x-ray exposure with automatically set radiography conditions. Each radiography conditions can be changed and saved.

3-2. SETTING MANUAL RADIOGRAPHY 1. Select the manual radiography mode Check if the APR mode is chosen. If they are selected in APR mode, please press APR setting switch, and release the setup conditions. 2. Setting tube voltage "kV" Set up the tube voltage with “kV” setting switch. Setting range: 40kV – 125 kV (1kv step) 3. Setting tube current “mA” Set up the tube current with “mA” setting switch. Setting range: 50, 100, 150mA (Small Focus),

200, 300mA (Large Focus)

4. Setting the tube current-time “mAs” and exposure-time “sec” Select the radiography mode with mAs/Sec selection switch. The “Sec” radiography makes X-ray exposure by setting time. The “mAs” radiography makes X-ray exposure by setting current-time. A setting of mAs values in mAs mode. mAs value is chosen with the mAs/Sec setting switch. Setting range: 0.4 ~ 320mAs in 30 Steps Step:

0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10.0, 12.5, 16.0, 20.0, 25.0, 32.0, 40.0, 50.0, 63.0, 80.0, 100, 125, 160, 200, 250, 320mAs

40/146

Operation Manual: DM-5125

A setting of Sec values in Sec mode. Sec value is chosen with the mAs/Sec setting switch. Setting range: 0.004 ~ 6.0sec in 33 Steps Step:

0.004, 0.005, 0.006, 0.008, 0.010, 0.013, 0.016, 0.020, 0.025, 0.032, 0.040, 0.050, 0.063, 0.080, 0.100, 0.125, 0.160, 0.200, 0.250, 0.320, 0.400, 0.500, 0.630, 0.800, 1.00, 1.25, 1.60, 2.00, 2.50, 3.20, 4.00, 5.00, 6.00sec

NOTICE About “OVER” display of radiographic conditions When set-up radiographic condition “kV”, “mA” and “mAs” or “sec” exceeds the permissible value of equipment, OVER sign lights up. X-ray exposure is prohibited when the over sign is turned on. Please reset the appropriate radiographic conditions.

3-3. SETTING APR RADIOGRAPHY 1. Selection of APR mode Choose APR mode with the [9] APR selection switch and the lmap displayed APR mode will be lighted on the upper left of the OP Panel. 2. Selection of Radiography Regions Choose the radiographic regions with [10] APR region selection switch and you can move the lamp display with up/down button. Available option is 5. 3. Selection of Radiographic Directions Choose the radiographic directions with [11] APR direction selection switch out of 2 options. 4. Selection of Radiographic Size Choose the radiographic size with [1] Patient Size selection switch out of 3 options, [L]. [M], [S]. 5. Change of the Radiographic Conditions You can change the selected radiographic conditions with the same method of the 3-2 manual radiographic condition setting. The over lamp can be lighted from time to time when the radiographic conditions change. Please be careful in setting of the radiographic conditions since the X-ray exposure is prohibited under the case. 6. Memory of the Changed Radiographic Conditions You can save the changed conditions above. How to store is pressing the selected switch for 2 second until the buzzer beeps.

CAUTION

Memory of the changed radiographic condition Be careful that other radiographic conditions can be selection if you choose the selected radiographic size switch. You cannot save the memory while the over lamp is lighted on.

41/146

Operation Manual: DM-5125

3-4. RADIOGRAPHY 1. PREPARATION After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is ended. 2. X-RAY RADIOGRAPHY BY RADIOGRAPHY SWITCH ON THE PANEL 1) Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is completed. 2) If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be started. 3) X-rays exposure LED will be lighted on and the buzzer sounds during the exposure. 3. X-ray Radiography Using Hand-Switch 1) Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when it is prepared to make exposure. 2) When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started. 3) During exposure, the exposure lamp will be lighted on and the buzzer sounds. 4) If the X-ray exposure lamp is lighted or the buzzer is off, it means the radiography work is completed. Then, you can stop pressing the hand switch. 5) If you get your hands off and stopped X-ray shooting during exposure, the check code will be displayed.

CAUTION

If you touch out from the hand switch during exposure, the x-ray shooting is stopped and you cannot get clear pictures. Please touch off your hand after the x-ray exposure signs and the buzzer sound are turned off except the case you want to stop the x-ray exposure urgently

CAUTION

You cannot make radiography while the over lamp is lighted on. Change the conditions for the radiography in those cases.

42/146

Operation Manual: DM-5125

PART 1. Generator

3. DIG-325/525/650/1150 [20/32/40/50/80kW Generator]

43/146

Operation Manual: DM-5125

NOTICE 1 This manual is designed to ensure correct use and operation of DIG-Series HFG. Please read all the lines thoroughly before your using the generator. The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which many results from the use of this manual. Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. Always keep the manual at hand for your reference.

NOTICE 2 1. This equipment should be used only by the legally qualified persons and practitioners. 2. For the safe installation a. Please avoid wet and humid locations. b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. c. Protect the product from slope, vibration, and shock. d. Do not install the product where chemicals are stored or gases are generated. e. Check the frequency, voltage, and current of power supply. f. Check the earth ground power about the electric discharge and polarity. 3. Before the operation of the machine a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates. b. Check the safety of the ground power connections. c. Check the correctness and safety of the cable and plug connections. d. Be careful that mixed use of the machine may give wrong diagnostic results and damages. e. Check the exterior circuits, which occur direct physical contact to the patient. f. Check the earth ground power about the electric discharge and polarity. 4. During the operation of the machine a. Do not exceed the designated time for diagnosis and cures. b. Monitor the conditions of device and influence on the patient. c. When you find any troubles about generators or patients, stop the operation of the machine and take appropriate measures for the safety of the patient. d. Be careful that the patient should not touch the device.

44/146

Operation Manual: DM-5125

5. After the operation of the machine a. Turn off the power of the machine after returning the operation switches and scales to the original position in accordance with the prescribed regulation. b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections. c. For the safe storage and installation d. Please avoid wet and humid locations. e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. f. Protect the product from slope, vibration, and shock. g. Do not install the product where chemicals are stored or gases are generated. 6. If the device is broken, mark proper description and ask repairs to the engineer. 7. Do not modify the machine. 8. Inspections and maintenance a. All device and parts should be checked on the regular basis. b. If you operate the device after long period of non-use, check the safety and conditions of the machine. 9. Operate the machine in correct ways according to the manuals. 10. Be careful about following cautions a. Wear the lead apron in order to prevent the x-ray irradiation. b. Regular inspection should be executed about leakage dose and irradiation meter. c. Be careful about the x-ray exposures. d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1 1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate measures. 2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off the power during operation. 3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions.

45/146

Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2 1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic drugs, oxygen and hydrogen near the working locations. 2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must be checked after the occurrence of the earthquake. 3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use this machine” and ask repairs to the engineers. 4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE 1. Power on and off Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the resupply of the power. 2. Regarding the operation environment a. X-ray room Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is designed to be protective against humidity and dust. However, if you use the machine at the place with high humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts. b. Do not leave the machine exposed to the strong direct sunlight. 3. Exposure of the x-rays a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions. b. Take enough cares in order not to make excessive radiation on the same body locations. 4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS Do not use the machine with following equipment. *The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The officials in hospital ought to give enough advises and education for the operator of the machine, doctors, patient and other the third parties.

46/146

Operation Manual: DM-5125

SYMBOLS AND TERMS The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that

DANGER

cause heavy damages to people and even the death. It warns of the conditions and situations, if not taken over or avoided, that

WARNING

could cause heavy damages to people or cause unrepeatable damages to the equipment. It warns of the conditions and situations, if not taken over or avoided, that

CAUTION

could cause damages to people, or to the equipment.

WAIVER CLAUSE Dongmun do not take responsibility for the following causes for damage and repairs. 1. Any installation, movement, modification and repairs by other engineering services, which was not designated by Dongmun. 2. Troubles and causes to our machine incurred by other equipment. 3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun. 4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual. 5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the powers, installation and working places. 6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders. 7. Accident and damage happened by the use beyond the original purpose of the machine. 8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the conditions and results of the medical prescriptions.

47/146

Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY 1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from normal use and operation of product. 2. Service fee is charged for the following cases. - Malfunctioning after the period of warranty - Disorder caused by fire, earthquake, or lightening - Inappropriate movement or negligence of product - Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones. 3. Note that defects not related to the main functions of the product are not eligible for this warranty. 4. Requirements for Service - Stop using the product when a problem is found, and check the manual for a solution. - Turn off the product before service request, and check the model number, serial number, the date of purchase before calling the service center. 5. Defects or depreciation in appearance is not eligible for refund or change. 6. Dongmun is not responsible for indirect damage related to the use of products. 7. Dongmun is not responsible for injury or damage after the end of warranty period. 8. All detailed adaptability warranties and other warranties are subjected to this statement.

48/146

Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION 1-1. COMPOSITION The electric composition block diagram of DIG-Series is shown in Fig 1-1.

EM swich Power Supply

HF DC/AC Inverter

AD/DC Convertor

Delay EM swich

HV TANK X-ray TUBE

IGBT Driver Stater OP PANEL

HT CONT Collimator PWR

FILAMENT IF

AEC

ION Chamber

Fig 1-1 COMPOSITION OF THE HFG AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is inverted in high-frequency voltage of about 20kHz by IGBT inverter. The high-frequency voltage is converted into the 40 (+20kV/-20kV) to 125kV (+62.5kV/-62.5kV) DC high voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current. . The operation panel can operate selection in each mode of operation and control, selection and setting of conditions. These operation conditions are transmitted by the communication system from OP-CONT to HTCONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CONT.

49/146

Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel High Voltage Generator

Main CPU

A/D CONV

D/A CONV

kV FB

PWM

PWM Filament mA FB

kV, mA FB Amp

Drive circuit

rms/DC CONV

AEC

IGBT

HV Trans FIL. Trans.

A X-ray TUBE

kV mA

C FIL.

ION Chamber

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

50/146

FIL. INV.

Operation Manual: DM-5125

1-2. SPECIFICATION

1

Rating Output

2

Tube Voltage kV

DIG-325 DIG-325 DIG-525 DIG-650 DIG-1150

20kW 32kW 40KW 50kW 80KW

Radiography

40-150kV (1kV STEP)

Fluoroscopy

40-125kV (1kV STEP) 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160mA

3

Tube Current mA

Radiography

(Small Focus) 200, 250, 320, 400, 500, 630, 800mA (Large Focus)

Fluoroscopy

0.1~4.0mA (0.1mA Step)

4

kV Accuracy

±10%

5

mA Accuracy

±15%

6

Timer Range

0.001 sec – 10 sec (38 step), 0.003 sec ~

7

mAs Range

DIG-325 DIG-525 DIG-650 DIG-1150

8

mAs Accuracy

9

Display Method

LED digital indication

10

Setting Method

Sheet Push Switch

11

Self-diagnostic Function

Error Code Indication

12

Technique Selection

Manual, APR, AEC

13

APR

216 exam techniques

14

AEC (Option)

Ionization Chamber 5 (Option)

0.1mAs - 630mAs 0.1mAs - 630mAs 0.1mAs - 800mAs 0.1mAs - 999mAs

20mAs or more: Less than 15% Less than 20mAs: Less than 20%

Display (OP Panel)

15

16

kV

3 Digit LED

mA

3 Digit LED

mAs

4 Digit LED (change into sec)

Time

4 Digit LED (change into mAs)

Selection/setting

LED (36)

Alarm

LED (2)

Weight

62 kg

Operation Panel

2 kg

51/146

6 sec (PXR-161 Only)

Operation Manual: DM-5125

1-3. ELECTRICAL REQUIREMENT Model

DIG SERIES

Line Nominal Voltage

Single Phase 200V / 220V / 240V

Automatic Line Compensation Range

±10 %

Frequency

50 Hz / 60 Hz

Line Impedance

Less Than 0.12 Ω

Grounding conditions

Based on the local regulation or 3 independent ground circuit

1-4. PROPER ENVIRONMENT 1. Avoid following places for the normal operation and safe storage. a. Where the equipment is exposed to water vapor. b. Where the equipment is exposed to direct sunlight. c. Where the equipment is exposed to dust. d. Where the equipment is exposed to high humidity. e. Where there is a ventilation problem. f. Where the equipment is exposed to salty atmosphere. g. Where the equipment is exposed to chemicals or gas. 2. Keep away from the place with strong vibration and maintain proper environment and conditions. OPERATION ENVIRONMENT Temperature range

+ 10°C ~ + 40°C,

Relative Humidity Range

30% ~ 75% RH (No dews)

Others

No Direct Sunlight No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere. STORAGE AND TRANSPORTATION Temperature range

-25°C ~ +60°C

Relative Humidity Range

10% ~ 95% RH(No dews)

Air Pressure

700hPa ~ 1060hPa

Others

No corrosive gases

52/146

Operation Manual: DM-5125

SECTION 2. DISPLAY / FUNCTIONS 2-1. STRUCTURES AND SIZE

FIG 2-1. STRUCTURE AND FIGURES (16kW ~ 50kW)

FIG 2-2. STRUCTURE AND FIGURES (50kW ~ 80kW 3Phase)

53/146

400. 0 m m

2 0 0 .0m m

Operation Manual: DM-5125

36. 0 m m

42. 0 m m

FIG 2-3. CONFIGURATION OF OPERATION PANEL

54/146

Operation Manual: DM-5125

2-2. OPERATION PANEL 22 10 16

12

3

9

4

23

13

14

15

18

19

5

11

1

2

24

20

17

21

25

7

FIG. 2-4 OPERATION PANEL

1

Power on switch

14

APR exposure region selection/display

2

Power on switch

15

APR exposure direction selection

3

kV Display Segment

16

Patient size selection and memory

4

kV Setting Switch

17

AEC Reset/ mode display

5

mAs/sec Setting Switch

18

Backup release switch

6

mAs/sec Setting Switch

19

Density setting/display switch

7

X-ray Exposure Ready Switch

20

Bucky selection/display

8

X-ray Exposure Switch

21

AEC mode and field selection switch

9

“OVER” Display (Red)

22

X-ray exposure hand switch

10

“ERROR” Display (Red)

23

Small focus display

11

X-ray Exposure ready sign

24

Large focus display

12

X-ray Exposure Display (Yellow)

25

AEC film/screen combination selection

13

APR Menu Selection Switch

55/146

8

6

Operation Manual: DM-5125

1) Power ON Switch Turn on the power of generator. 2) Power OFF Switch Turn off the power of generator. 3) kV Setting/Display Segment Up-Button increases the kV value of tube voltage whereas Down-button decreases kV. If you keep buttons pressed, the value of kV will be increased or decreased continuously. 4) mA Setting/Display Segment Up-Button increases the mA value of tube voltage whereas Down-button decreases mA. If you keep buttons pressed, the value of mA will be increased or decreased continuously. 5) mAs/Sec Setting/Display Segment Up-Button increases the mAs/Sec value of tube voltage whereas Down-button decreases mAs/Sec. If you keep buttons pressed, the value of mAs/Sec will be increased or decreased continuously. 6) mAs/Sec Selection Switch The mAs or sec radiography mode is selected. The selected radiography mode is displayed with LED. 7) X-ray Exposure Ready Switch It circulates the rotor and controls the heat of the filament that makes ready actions for the x-ray exposure. 8) X-ray Exposure switch Press the exposure switch after the ready sign is lighted on. 9) “Over” LED (Red) The over LED is lighted on when the exposure conditions exceeds the normal standards. You cannot make exposure while the over LED is lighted on. 10) “Error” Indication (Red) The LED is lighted in case of errors. 11) X-ray Exposure Ready LED (Green) The exposure ready sign is lighted on when it is at the ready condition. 12) X-ray Exposure LED (Yellow) The X-ray exposure sign is lighted on during the X-ray exposure.

56/146

Operation Manual: DM-5125

13) APR Mode Setting Switch It operates the on/off of the APR mode. 14) APR Exposure Region Selection/Display switch It operates the selection of exposure regions and the selected location will be displayed on the LED. 15) APR Exposure Direction Setting Switch It operates the selection of exposure directions the selected direction will be displayed on the LED. 16) Patient Size Selection and Memory Switch It operates the selections of patient size. The selected size will be displayed on the segment. Save the changed APR conditions by pressing the size switches. 17) AEC Reset/Mode Display Switch If the x-ray exposure is closed early by the AEC backup timer, the AEC mode LED will blink off and the buzzer sounds. The next exposure is forbidden until the reset switch is pushed. 18) Backup Release Switch Release the back-up conditions. 19) Density Setting/Display Switch (AEC/APR mode) It selects the density level of the exposure and increases or decreases the mAs/sec under the APR mode. 20) Bucky Selection/Display Select and display BUCKY1 (table) and BUCKY 2(stand). 21) AEC Mode and Field Selection Switch It selects the field on the AEC mode. Each LED represents the physical location of irradiation detector and the combination of designated field is available. 22) X-ray Exposure Hand Switch It circulates the rotor and controls the heat of the filament that makes ready actions for the x-ray exposure. Press the exposure switch after the ready sign is lighted on. Please touch off from the button after the LED is lighted on. 23) Small Focus Display (Green LED) It is displayed when the small focus is selected. 24) Large Focus Display (Green LED) It is displayed when the large focus is selected. 25) Film/Screen Combination Selection It operates the adjustment of kVp and divided by low speed, medium speed and high speed (200, 400, 800)

57/146

Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE 3-1. POWER ON & OFF 1. Power-on Switch a. Press the Power on switch. b. If the power is turned on, initial condition of equipment will be checked. c. During the checking actions, "-" mark will be displayed on kV, mA, and mAs/sec display segment. d. After the check, it displays radiographic conditions. e. Radiographic conditions display the final status of the previous operation. 2. Power-off Switch a. Press the Power off switch. b. Please turn off the power supply except the time of use.

Do not turn on and off the power of the equipment continuously except

CAUTION

emergent cases. Have the interval of 30 seconds or more between ON-OFF operations of a power supply.

3-2. SELECTION OF THE RADIOGRAPHY METHOD There are radiography modes shown below in this equipment. 1. Manual radiography mode Radiography conditions are performed by manual setup. 2. APR radiography mode You can select the radiography region; direction and patient size on the control panel and make x-ray exposure with automatically set radiography conditions. Each radiography conditions can be changed and saved. 3. AEC radiography mode It helps to produce the optimal images by AEC (Automatic Exposure Control Option) Menu selection.

58/146

Operation Manual: DM-5125

3-3. SETTING MANUAL RADIOGRAPHY 1. Select the manual radiography mode Check if the APR mode or AEC mode are chosen. If they are selected in APR or AEC mode, please press APR or AEC setting switch, and release the setup conditions. 2. Setting tube voltage "kV" Set up the tube voltage with “kV” setting switch. Setting range: 40kV – 125 or 150kV (1kv step) 3. Setting tube current “mA” Set up the tube current with “mA” setting switch. Setting range: 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160

m: Large Focus

200, 250, 320, 400, 500, 630, 800, 999 Model

mA Range

DIG-325

10~400mA

DIG-525

10~500mA

DIG-650

10~630(800)mA

DIG-1150

10~999mA

: Small Focus

Remark

4. Setting the radiography mode (mAs/sec radiography) Select the radiography mode with mAs/sec selection switch (display). "sec" radiography makes x-ray exposure by set-up time "mAs" radiography makes x-ray exposure by the setup time of current. - A setup of mAs value in “mAs" mode - mAs value is chosen with the mAs/sec setting switch. Setting range: 0.1mAs ~ 800mAs Step : 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160, 200, 250, 320, 400, 500, 630, 800, 999mAs Model

mAs Range

DIG-325

0.1~630mAs

DIG-525

0.1~630mAs

DIG-650

0.1~800mAs

DIG-1150

0.1~999mAs

Remark

59/146

Operation Manual: DM-5125

- A setup of sec value in “sec” mode - Sec Value is chosen with the mAs/sec [5] setting switch. Setting range: 0.001sec ~ 10sec Step : 0.001, 0.002, 0.003, 0.004, 0.005, 0.006, 0.008, 0.010, 0.013, 0.016, 0.020, 0.025, 0.032, 0.040, 0.050, 0.063, 0.080, 0.100, 0.125, 0.160, 0.200, 0.250, 0.320, 0.400, 0.500, 0.630, 0.800, 1.00, 1.25, 1.60, 2.00, 2.50, 3.20, 4 .00, 5.00, 6.30, 8.00, 10.00sec About "OVER" display of radiography conditions When set-up radiography condition "kV", "mA", and "mAs or sec" exceeds the permissible value of equipment, OVER sign lights up. X-ray exposure is prohibited when the over sign is turned on. Please reset the appropriate radiography conditions.

NOTICE

Model

sec Range

DIG-325

0.001~10 sec

DIG-525

0.001~10 sec

DIG-650

0.001~10 sec

DIG-1150

0.001~10 sec

Remark

3-4. SETTING APR 1. Selection of APR mode Choose APR mode with the (13) APR selection switch and the LED displaying APR mode will be lighted on the upper left of the OP panel. 2. Selection of Radiography Regions Choose the radiography regions with (14) radiography region selection switch and you can move the LED display with up/down button. The available option is 9. 3. Selection of Radiography Directions Choose the radiography directions with (15) radiography direction switch out of 4 options. 4. Selection of the Radiograph Size Choose the radiography size with (16) patient size selection switch out of 6 options, [LL] [L] [M] [S] [SS] [BABY]. The radiography condition will be selected after the selection of the radiography symbol, radiography region, and radiography method. As well the order for the selection will not be changed.

60/146

Operation Manual: DM-5125

5. Selection of the Density Adjust the mAs/sec settings with the density switch. Increase the conditions of mAs/sec if the Density is +1 ~ +3. Increase the conditions of mAs/sec if the Density is -1 ~ -3. 6. Change of the Radiography Conditions You can change the selected radiography conditions with the same method of the 3-3 manual radiography condition setting. The over LED can be lighted from time to time when the radiography conditions are changed. Please be careful about the setting of the radiography conditions since the x-ray exposure is prohibited under the case. 7. Memory of the Changed Radiography Conditions You can save the conditions changed above. The method for store is pressing the selected switch for 2 seconds until the buzzer sounds.

Memory of the changed Radiography condition Be careful that other radiography conditions can be selected if you choose

CAUTION

the selected radiography size switch. You cannot save the memory while the over LED is lighted and when the density setting is not 0.

3-5. AEC RADIOGRAPHY 1. Selection of AEC radiography mode In order to use AEC mode, the DIP S/W 5 of operation panel should be turned on. When DIP S/W 5 is turned off: AEC mode is not available. 2. Switch [17] is the AEC mode selection switch. Ion-chamber is chosen with switch [18] after you select the AEC mode with switch [17]. 3. The selection of density in the AEC mode can adjust the level of x-ray exposure levels. 4. “Ready” action is not operated if the bucky and field are not chosen after the AEC mode selection.

3-6. RADIOGRAPHY 1. Photography Preparation After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is ended.

61/146

Operation Manual: DM-5125

2. X-ray Radiography Using Operation Panel a. Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is completed. b. If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be started. c. X-rays exposure LED will be lighted on and the buzzer sounds during the exposure. 3. X-ray Radiography Using Hand-Switch a. Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when it is prepared to make exposure. b. When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started. c. During exposure, the exposure LED will be lighted on and the buzzer sounds. d. If the X-ray exposure LED is turned off or buzzer does not sound, it means the radiography work is completed. Then, you can stop pressing the hand switch. e. If you touched out and stopped x-ray shooting during exposure, the check code will be displayed. If you touch out from the hand switch during exposure, the x-ray shooting is

CAUTION

stopped and you cannot get clear pictures. Please touch off your hand after the x-ray exposure signs and the buzzer sound are turned off except the case you want to stop the x-ray exposure urgently.

62/146

Operation Manual: DM-5125

SECTION 4. TROUBLES 4-1. ERROR CODE AND CHECK CODE In case any problems happen during the operation of the device, they will be indicated on the kV/mA segment. 1. Error Code The error code is displayed when problems, such as abnormalities, occur to the equipment. While the error code is displayed, radiography operation is forbidden for safety. A power supply should be switched off and on again and an error code will be cleared. If the problem of equipment is not solved at the time of the power-supply reswitched on, the error code will be displayed again. Since it is the failure of equipment when an unusual state is not recovered, please inform the service section of our company or our agent. 2. Check Code Unlike an error code, a check code is the warning sign when wrong operations are applied by the user. X-ray radiography operation is forbidden for the safety during the check code. The way to release a check code is pressing the WR switch for 2 seconds. CH01 will be released if thermal switch is off during the re-supply of the power.

63/146

Operation Manual: DM-5125

ERROR CODE LIST Error Code

DESCRIPTION

Err 01

CHARGE MONITOR ERROR (HT-CONT)

Err 02

OVER TIME ERROR (OP PANEL) AT STAND BY

Err 03

READY-OUT SIGNAL ERROR AT SAND BY

Err 04

X-RAY OUT SIGNAL ERROR AT STAND BY

Err 05

READY SWITCH INPUT ERROR AT STAND BY

Err 06

X-RAY SWITCH INPUT ERROR AT STAND BY

Err 07

HAND SWITCH INPUT ERROR AT STAND BY

Err 08

HAND SWITCH X-RAY INPUT ERROR AT STAND BY

Err 09

ROTOR DETECTER SIGNAL AT STANDBY

Err 10

FILLAMENT CURRENT DETECTER ERROR AT STANDBY

Err 11

‘kV FEEDBACK ERROR AT STANDBY

Err 12

‘mA FEEDBACK ERROR AT STANDBY

Err 13

ROTOR DETECTER ERROR AT EXPOSURE

Err 14

FILLAMENT CUREENT DETECTER ERROR AT EXPOSURE

Err 15

‘kV FEEDBACK ERROR DETECTER ERROR AT EXPOSURE

Err 16

‘mA FEEDBACK DETECTIVE ERROR EXPOSURE

Err 17

NOT DEFINED

Err 18

X-RAY TIME OVER IN SEC MODE

Err 19

X-RAY TIME OVER IN mAs MODE

Err 20

X-RAY TIME OVER IN AEC MODE

Err 21

HV INTERLOCK ERROR

Err 22

IGBT FEEDBACK ERROR

Err 23

‘kV FEEDBACK OVER DETECTER ERROR AT EXPOSURE

Err 24

‘mA FEEDBACK OVER DETECTER ERROR AT EXPOSURE

Err 25

NO ZERO-CROSS OR NO INPUT

Err 26

ZERO-CROSS OVER OR WRONG INPUT FREQUENCY

Err 27

BUCKY 1 OK SIGNAL ERROR (HT-CONT)

Err 28

BUCKY 2 OK SIGNAL ERROR (HT-CONT)

Err 29

AEC A/D DATA-IN ERROR (HT-CONT)

Err 30

AEC A/D DATA-IN ERROR (HT-CONT) AT NON-EXPOSURE

Err 31

I/F BOARD CONDITION ERROR

Err 32

AEC BOARD CONDITION ERROR

Err 33

SMALL-FOCUS SELECTION ERROR (HT-CONT)

Err 34

LARGE-FOCUS SELECTION ERROR (HT-CONT)

64/146

ACTION

RESUPPLY OF THE POWER

Operation Manual: DM-5125

CHECK CODE LIST Check Code

DESCRIPTION

Measure

Ch 01

TUBE THERMOSTAT OPERATION

Reset

Ch 02

When another switch is pushed during exposure

Release by “WR” switch

Ch 03

When ready switch on OP PANEL is detached during exposure

Ch 04

When X-RAY switch on OP PANEL is detached during exposure

Ch 05

When a hand switch (ready) is detached during exposure

Ch 06

When a hand switch (X-RAY) is detached during exposure

Ch 07

Shortage of power-supply capacity for exposure

Power-supply re-switch

Ch 08

Simultaneous exposure prohibition

Reset

Ch 09

Warning to X-rays room door opening

Ch 10

AEC Level Error

# CH01 is will be released if the TUBE terminal switch is off, in case resupply of the power.

65/146

Operation Manual: DM-5125

SECTION 5. MAINTANENCE 5-1. MAINTANANCE AND PERIODIC OPERATOR Maintenance has check menus carried out daily by the user and periodic inspection by technical service engineers who completed special training programs on the repair of the generator. User checks items based on " Daily Checking Items by the operator " and should keep data records. The following checks and maintenance procedures, together with the suggested intervals, are the manufacturer’s recommendation for most effective Periodical Maintenance schedule for this generator. Exclusively service personnel specifically trained on medical X-ray generator must perform service tasks here described. The first periodic maintenance service should be performed six (6) months after installation, and the subsequent service every twelve (12) months. When a trouble and a poor part are discovered, please stop use of equipments, turn off a power supply, and contact a near service station.

5-2. DAILY CHECKING ITEMS BY THE OPERATOR Check following items according to the regular inspection plans. Please refer to the daily inspection method regarding the way of checking.

In order to avoid the danger of an electroshock, please do not remove the cover of equipment by any means.

CAUTION

Only the expertise person who received training performs the check and repair accompanied by removal of a cover.

INTERVAL

Check item

Daily

Operation Panel

○ ○ ○

Radiographic Condition



Radiation field lamp



Warning light in use



Air Conditioner Temperature and Humidity

66/146

Weekly

Monthly

Operation Manual: DM-5125

5-3. WAYS OF DAILY CHECKING (CHEKING MENUS) Please check clear environment inside the operation room and x-ray room. The system is designed to operate with normal conditions under following environment. Please keep following temperature and humidity. The recommended operating environment is follow. Temperature

: +10 - +35 ℃

Relative humidity

: 45 – 85 %, 45~65 % (X-ray room)

In the environment with a rapid change of temperature and humidity, it is easy

CAUTION

to generate dew condensation. Installation of an air conditioner is recommended

The ingredient that makes silver etc. sulfurate may be contained in the exhaust

CAUTION

gas of an automatic development machine. Please install the independent air conditioner.

Operation Panel Please check whether each switch of an operation panel lights up. When you clean the surface of an operation panel, it should be wiped with the soft and fine cloth. When using a cleaner, it is made to contain in cloth and is used.

Don't blow a liquid upon the operation panel directly.

CAUTION

It goes into the inside of equipment from a crevice, and becomes the cause of failure.

(The check of radiographic condition Before the commencement of work, trial radiography is performed and film density, unevenness, and etc. are checked. Radiation field lamp It confirms whether a radiation field lamp lights up before the commencement of work. Warning light in use It confirms whether warning light in use switch on before the commencement of work.

67/146

Operation Manual: DM-5125

PART 1. Generator

4. DIG-650RF [50kW 2-TUBE Generator]

68/146

Operation Manual: DM-5125

NOTICE 1 This manual is designed to ensure correct use and operation of DIG-Series HFG (2-TUBE). Please read all the lines thoroughly before your using the generator. The responsibility about the use of medical apparatus and management of maintenance is in a user’s side. No responsibility is taken by Dongmun for any infringement of patients or other rights of the third parties which many results from the use of this manual. Incorrect use and operation exceeding described conditions in this manual may occur damage of the device and shorten its life. Particular attention must be paid to all the warnings, cautions and notes incorporated herein. Always keep the manual at hand for your reference.

NOTICE 2 1. This equipment should be used only by the legally qualified persons and practitioners. 2. For the safe installation a. Please avoid wet and humid locations. b. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. c. Protect the product from slope, vibration, and shock. d. Do not install the product where chemicals are stored or gases are generated. e. Check the frequency, voltage, and current of power supply. f. Check the earth ground power about the electric discharge and polarity. 3. Before the operation of the machine a. Examine contact of switch, polarity, dial setting, and display, and check if the product properly operates. b. Check the safety of the ground power connections. c. Check the correctness and safety of the cable and plug connections. d. Be careful that mixed use of the machine may give wrong diagnostic results and damages. e. Check the exterior circuits, which occur direct physical contact to the patient. f. Check the earth ground power about the electric discharge and polarity. 4. During the operation of the machine a. Do not exceed the designated time for diagnosis and cures. b. Monitor the conditions of device and influence on the patient. c. When you find any troubles about generators or patients, stop the operation of the machine and take appropriate measures for the safety of the patient. d. Be careful that the patient should not touch the device.

69/146

Operation Manual: DM-5125

5. After the operation of the machine a. Turn off the power of the machine after returning the operation switches and scales to the original position in accordance with the prescribed regulation. b. Do not pull out the plugs and lines with compelling forces and it can hurt the connections. c. For the safe storage and installation d. Please avoid wet and humid locations. e. Install the product where there is no adverse effect from pressure, temperature, humidity, ventilation, dust, sulfur and other air pollutants. f. Protect the product from slope, vibration, and shock. g. Do not install the product where chemicals are stored or gases are generated. 6. If the device is broken, mark proper description and ask repairs to the engineer. 7. Do not modify the machine. 8. Inspections and maintenance a. All device and parts should be checked on the regular basis. b. If you operate the device after long period of non-use, check the safety and conditions of the machine. 9. Operate the machine in correct ways according to the manuals. 10. Be careful about following cautions a. Wear the lead apron in order to prevent the x-ray irradiation. b. Regular inspection should be executed about leakage dose and irradiation meter. c. Be careful about the x-ray exposures. d. Do not exceed the maximum weight for the x-ray tables.

NOTICE FOR THE OPERATION 1 1. If you find any troubles about the conditions of patients or machine, stop the operation and take appropriate measures. 2. For the need of emergency power cut-off, use the power switch on the operation panel. Do not randomly cut off the power during operation. 3. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions.

70/146

Operation Manual: DM-5125

NOTICE FOR THE OPERATION 2 1. As this device is not protective against exposure, do not use combustible or blasting gases such as narcotic drugs, oxygen and hydrogen near the working locations. 2. Do not use the machine during the warning signs of the earthquake. As well, the security of the machine must be checked after the occurrence of the earthquake. 3. In case any troubles happen to the machine, turn off the power and mark appropriate sign such as “Do not use this machine” and ask repairs to the engineers. 4. Regular inspection is required to the device and ask any questions to the sales division of the Dongmun.

OTHER NOTICE AND ADVICE 1. Power on and off Do not turn off the machine for 30 seconds after turning on the power. Please wait at least 30 seconds for the resupply of the power. 2. Regarding the operation environment a. X-ray room Practitioners should take enough cares for the release of the x-rays out of the x-ray room. This device is designed to be protective against humidity and dust. However, if you use the machine at the place with high humidity and dust, it can shorten the life of electronic insult and grow rust on the metal parts. b. Do not leave the machine exposed to the strong direct sunlight. 3. Exposure of the x-rays a. Be careful about using the exposure and hand switches on the operation panel. You can make exposures anytime under normal conditions. b. Take enough cares in order not to make excessive radiation on the same body locations. 4. Each cable has enough length and flexibility but do not bend the cable forcibly.

DANGEROUS Do not use the machine with following equipment. *The equipment, which release electronic waves such as mobile phone, radios, wireless toys, and etc. The officials in hospital ought to give enough advises and education for the operator of the machine, doctors, patient and other the third parties.

71/146

Operation Manual: DM-5125

SYMBOLS AND TERMS The following symbols will be used in this manual: their meanings and applications are described further below.

It warns of the conditions and situations, if not taken over or avoided, that

DANGER

cause heavy damages to people and even the death. It warns of the conditions and situations, if not taken over or avoided, that

WARNING

could cause heavy damages to people or cause unrepeatable damages to the equipment. It warns of the conditions and situations, if not taken over or avoided, that

CAUTION

could cause damages to people, or to the equipment.

WAIVER CLAUSE Dongmun do not take responsibility for the following causes for damage and repairs. 1. Any installation, movement, modification and repairs by other engineering services, which was not designated by Dongmun. 2. Troubles and causes to our machine incurred by other equipment. 3. Any troubles and defects by the use of other parts, which is not designated and recommended by Dongmun. 4. Damages or accident caused by the disobedience to the cautions and regulations in user’s manual. 5. Troubles and damages created by inappropriate environment prescribed in this manual regarding use of the powers, installation and working places. 6. Any defects and causes occurred by natural calamity such as fire, earthquake, windfall and thunders. 7. Accident and damage happened by the use beyond the original purpose of the machine. 8. Diagnosis for the patient should be under the doctor’s responsibility. Dongmun do not take responsible for the conditions and results of the medical prescriptions.

72/146

Operation Manual: DM-5125

SCOPE AND PERIOD OF WARRENTY 1. All Dongmun products are guaranteed warranty of 1(one) year for any disorder or discrepancy occurred from normal use and operation of product. 2. Service fee is charged for the following cases. - Malfunctioning after the period of warranty - Disorder caused by fire, earthquake, or lightening - Inappropriate movement or negligence of product - Disorder from repair or reconstruction by other than Dongmun service agent or other designated ones. 3. Note that defects not related to the main functions of the product are not eligible for this warranty. 4. Requirements for Service - Stop using the product when a problem is found, and check the manual for a solution. - Turn off the product before service request, and check the model number, serial number, the date of purchase before calling the service center. 5. Defects or depreciation in appearance is not eligible for refund or change. 6. Dongmun is not responsible for indirect damage related to the use of products. 7. Dongmun is not responsible for injury or damage after the end of warranty period. 8. All detailed adaptability warranties and other warranties are subjected to this statement.

73/146

Operation Manual: DM-5125

SECTION 1. COMPOSITION / SPCECIFICATION 1-1. COMPOSITION The electric composition block diagram of DIG-Series is shown in Fig 1-1.

EM swich Power Supply

HF DC/AC Inverter

AD/DC Convertor

Delay EM swich

HV TANK X-ray TUBE

IGBT Driver Stater OP PANEL

HT CONT Collimator PWR

FILAMENT IF

AEC

ION Chamber

Fig 1-1 COMPOSITION OF THE HFG AC input voltage is converted into DC voltage by a single-phase all wave rectifier circuit. This DC voltage is inverted in high-frequency voltage of about 20kHz by IGBT inverter. The high-frequency voltage is converted into the 40 (+20kV/-20kV) to 125kV (+62.5kV/-62.5kV) DC high voltage by a high-voltage transformer and the multi-voltage rectifier circuit. This DC high voltage is supplied to an X-ray tube through a high-voltage cable, and PWM INV controls this X-ray tube voltage. Modulating the Filament voltage controls the tube current. . The operation panel can operate selection in each mode of operation and control, selection and setting of conditions. These operation conditions are transmitted by the communication system from OP-CONT to HTCONT. HT-CONT controls the output of a high-voltage generator according to the instructions from OP-CPU.

74/146

Operation Manual: DM-5125

The flow of the control signal of X-ray exposure is shown in the following Fig 1-2.

Operation Panel High Voltage Generator

Main CPU

A/D CONV

D/A CONV

kV FB

PWM

PWM Filament mA FB

kV, mA FB Amp

Drive circuit

rms/DC CONV

AEC

IGBT

HV Trans FIL. Trans.

A X-ray TUBE

kV mA

C FIL.

ION Chamber

Fig 1-2 CONTROL SIGNAL OF X-RAY EXPOSURE

75/146

FIL. INV.

Operation Manual: DM-5125

1-2. SPECIFICATION 1

Rating Output

DIG-650RF

50kW

2

Tube Voltage kV

DIG-650RF

40-150kV (1kV STEP) 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160mA

3

Tube Current mA

Radiography

(Small Focus) 200, 250, 320, 400, 500, 630, 800mA (Large Focus)

4

kV Accuracy

±10%

5

mA Accuracy

±15%

6

Timer Range

0.001 sec – 10 sec (38 Step)

7

mAs Range

DIG-650

8

mAs Accuracy

9

Display Method

LED digital indication

10

Setting Method

Sheet Push Switch

11

Self-diagnostic Function

Error Code Indication

12

Technique Selection

Manual, APR, AEC, ABC

13

APR

240 Exam Techniques

14

AEC (Option)

Ionization Chamber 5 (Option)

Display (OP Panel)

Radiography OP

Fluoroscopy OP

kV

3 Digit LED

3 Digit LED

mA

3 Digit LED

2 Digit LED

mAs

4 Digit LED (change into sec)

-

Time

4 Digit LED (change into mAs)

2 Digit LED

Selection/setting

LED (36)

LED (6)

Alarm

LED (2)

LED (1)

Weight

62 kg

Operation Panel

2 kg

15

16

0.1mAs - 800mAs (37 Step)

20mAs or more: Less than 15% Less than 20mAs: Less than 20%

1-3. ELECTRICAL REQUIREMENT Model

DIG SERIES

Line Nominal Voltage

Single Phase 200V / 220V / 240V

Automatic Line Compensation Range

±10 %

Frequency

50 Hz / 60 Hz

Line Impedance

Less Than 0.12 Ω

Grounding conditions

Based on the local regulation or 3 independent ground circuit

76/146

Operation Manual: DM-5125

1-4. PROPER ENVIRONMENT 1. Avoid following places for the normal operation and safe storage. a. Where the equipment is exposed to water vapor. b. Where the equipment is exposed to direct sunlight. c. Where the equipment is exposed to dust. d. Where the equipment is exposed to high humidity. e. Where there is a ventilation problem. f. Where the equipment is exposed to salty atmosphere. g. Where the equipment is exposed to chemicals or gas. 2. Keep away from the place with strong vibration and maintain proper environment and conditions. OPERATION ENVIRONMENT Temperature range

+ 10°C ~ + 40°C,

Relative Humidity Range

30% ~ 75% RH (No dews)

Others

No Direct Sunlight No combustible and corrosive gases

3. For safe storage and transportation, you must keep following range of temperature, humidity and atmosphere. STORAGE AND TRANSPORTATION Temperature range

-25°C ~ +60°C

Relative Humidity Range

10% ~ 95% RH(No dews)

Air Pressure

700hPa ~ 1060hPa

Others

No corrosive gases

77/146

Operation Manual: DM-5125

SECTION 2. DISPLAY / FUNCTIONS 2-1. STRUCTURES AND SIZE

FIG 2-1. STRUCTURE AND FIGURES

ERROR

Sec mAs

Sec mAs

Size

kV

mA

Density

Non

1

AEC

Field Select

2

3

WR

Bucky

Film/ Screen

FIG 2-2. CONFIGURATION OF OPERATION PANEL

78/146

Operation Manual: DM-5125

ERROR

kV

mA

min

Reset

Density

RAuto

ABC

R/F Table

FIG 2-3. CONFIGURATION OF FLUOROSCOPIC OPERATION PANEL

79/146

Operation Manual: DM-5125

2-2. RADIOGRAPHIC OPERATION PANEL DESCRIPTION

18

19

20

21

22

23

24

25

17

ERROR

Sec mAs

Sec mAs

Size

kV

mA

Density

1

6 2

3

4 5

7

9

8

Non

1

AEC

Field Select

10

11

2

3

WR

Bucky

Film/ Screen

12

13

FIG. 2-4 RADIOGRAPHIC OPERATION PANEL

1

Power on switch

14

X-ray exposure ready switch

2

Power off switch

15

X-ray exposure switch

3

mAs/sec display/setting switch

16

APR mode store switch

4

kV display/setting switch

17

APR menu selection switch

5

mA display/setting switch

18

“Over” LED (Red)

6

mAs/sec selection switch

19

Small focus LED (Green)

7

Density display/setting switch

20

Large focus LED (Green)

8

APR patient size setting switch

21

High speed rotor LED

9

Bucky selection switch

22

X-ray exposure ready LED (Green)

10

AEC display/setting

23

X-ray exposure LED (Yellow)

11

Field display/setting switch

24

LCD 1

12

Film/Screen selection switch

25

LCD 2

13

Page up/down & write switch

80/146

14

15

16

Operation Manual: DM-5125

1) Power ON Switch Turn on the power of generator. 2) Power OFF Switch Turn off the power of generator. 3) mAs/Sec Display/Setting Switch It operates the setting of mAs or Sec. Up-Button increases and Down-button decreases the values. When you keep pushing the buttons, the values will be changed consecutively. The selected [mAs] [Sec] will be displayed on the [mAs/Sec] segment. 4) kV Display /Setting Switch It operates the setting of kV. Up-Button increases and Down-button decreases the values. When you keep pushing the buttons, the values will be changed consecutively. The selected kV will be displayed on the kV segment. 5) kV Display /Setting Switch It operates the setting of mA. Up-Button increases and Down-button decreases the values. When you keep pushing the buttons, the values will be changed consecutively. The selected mA will be displayed on the mA segment. 6) mAs/Sec Selection Switch Select mAs/Sec exposure mode. - mAs/Sec selection [mAs] mode. - [Sec] mode for the exposure time. Indicate the selected exposure mode on the menu. 7) Density Display/Setting Switch Set the select density level of exposure. 8) APR Size Selection and Memory Switch Select the patient size for radiography. 9) Bucky Selection Switch Select BUCKY1, BUCKY2, BUCKY3 or Non-BUCKY. 10) AEC Display/Setting Switch Set or release the AEC mode. 11) Field Display/Setting Switch Select the field in AEC mode. Each LED indicate the physical location of AEC, and combination of designated fields are possible.

81/146

Operation Manual: DM-5125

12) Film/Screen Display/Selection Switch Adjust the kVp, which is the combination of programmed film and screen and display as low, middle, high speed (200, 400, 800) 13) Page Up/Down Switch & Write Switch Operate the page up/down of APR menu displayed on LCD 2(25). 14) X-ray Exposure Ready Switch It circulates the rotor and controls the heat of the filament that makes ready actions for the x-ray exposure. 15) X-ray Exposure switch After the ready LED is turned on, the exposure is executed by pressing X-ray exposure switch. 16) Write Switch Store the changed conditions on the selected APR menu. 17) APR Menu Selection Switch Select the APR Menu using the switch located below LCD 2. 18) “Over” LED (Red) The over LED is lighted on when the exposure conditions exceeds the normal standards. You cannot make exposure while the over LED is lighted on. 19) Small Focus Display (Green LED) Indicate that the small focus is selected. 20) Large Focus Display (Green LED) Indicate that the large focus is selected. 21) High-Speed Rotor LED(Green) Indicate that high-speed rotor is selected. 22) X-ray Exposure Ready LED (Green) Lighted when the exposure for radiography is ready. 23) X-ray Exposure LED (Yellow) Lighted during the exposure of X-ray. 24) LCD 1 - Display the error no and description under the error conditions - Display over when the conditions are over the standards. - Display the selected menu in APR Mode. 25) LCD 2 - Display the APR Menu. - Select the displayed menu by using switch 17.

82/146

Operation Manual: DM-5125

2-3. FLUOROSCOPIC OPERATION PANEL DESCRIPTION

1 3 8 2 4

9

10

ERROR

kV

mA

min

Reset

Density

RAuto

ABC

R/F Table

5

6

7

FIG. 2-5 FLUOROSCOPIC OPERATION PANEL 1

kV display/setting switch

6

ABC Mode display / selection

2

mA display/setting switch

7

R/F table mode display / selection

3

MAs/Sec display/reset switch

8

Error display (Red)

4

Density display/setting switch

9

X-ray exposure ready LED (Green)

5

R-Auto Mode display/selection

10

X-ray exposure LED (Yellow)

1) kV display/setting Switch It operates the setting of kV. Up-button increases and Down-button decreases the values. When you keep pushing the buttons, the values will be changed consecutively. The selected kV value will be displayed on the segment.

83/146

Operation Manual: DM-5125

2) mA display/setting Segment It operates the setting of mA. Up-button increases and Down-button decreases the values. When you keep pushing the buttons, the values will be changed consecutively. The selected mA value will be displayed on the segment. 3) mAs/Sec display/reset switch It displays the accumulated time under fluoroscopic mode. You can reset the accumulated time with reset switch. (4) Density display/setting Switch It operates the selections of density under the ABC mode and controls the brightness of monitor. 5) R-auto mode display/selection switch Selects or release R-auto mode. (6) ABC mode display/selection switch It selects or release ABC mode. 7) R/F table mode display/selection switch It selects or release R/F table mode. 8) Error LED (Red) The error LED is lighted on, when errors occur. 9) X-ray Exposure Ready LED (Green) The exposure ready sign is lighted on when it is at the ready condition. 10) X-ray exposure LED (Yellow) The X-ray exposure sign is lighted on during the X-ray exposure.

84/146

Operation Manual: DM-5125

SECTION 3. OPERATION / EXPOSURE 3-1. POWER ON & OFF 1. Power-on Switch a. Press the Power on switch. b. If the power is turned on, initial condition of equipment will be checked. c. During the checking actions, "-" mark will be displayed on kV, mA, and mAs/sec display segment. d. After the check, it displays radiographic conditions. e. Radiographic conditions display the final status of the previous operation. 2. Power-off Switch a. Press the Power off switch. b. Please turn off the power supply except the time of use.

Do not turn on and off the power of the equipment continuously except

CAUTION

emergent cases. Have the interval of 30 seconds or more between ON-OFF operations of a power supply.

3-2. SELECTION OF THE RADIOGRAPHY METHOD There are radiography modes shown below in this equipment. 1. Manual radiography mode Radiography conditions are performed by manual setup. 2. APR radiography mode You can select the radiography region; direction and patient size on the control panel and make x-ray exposure with automatically set radiography conditions. Each radiography conditions can be changed and saved. 3. AEC radiography mode It helps to produce the optimal images by AEC (Automatic Exposure Control Option) Menu selection. 4. Fluoroscopy mode 5. ABC mode This is the control mode to gain the best image under the fluoroscopy mode.

85/146

Operation Manual: DM-5125

3-3. SETTING MANUAL RADIOGRAPHY 1. Select the manual radiography mode Clear up the conditions by pushing bucky switch [13], and [16] switch when the APR mode and AEC mode are selected. 2. A set up of tube voltage "kV" Set up the tube voltage with “kV” setting switch. Setting range: 40kV – 150kV (1kv step) Model

kV Range

DIG-525

40~125kV

DIG-650

40~150kV

Remark

3. A set up of tube current “mA” Set up the tube current with “mA” [5] setting switch. Setting range: 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160

mA: Small Focus

200, 250, 320, 400, 500, 630, 800,

mA: Large Focus

4. A set up of mAs/Sec mAs/sec value is chosen with the mAs/sec setting switch. Setting range: 0.1mAs ~ 800mAs Step : 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.8, 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125, 160, 200, 250, 320, 400, 500, 630, 800mAs Model DIG-650

mAs Range

Remark

10~800mAs

Sec Value is chosen with the mAs/sec [5] setting switch. Setting range: 0.001sec ~ 10sec Step : 0.001, 0.002, 0.003, 0.004, 0.005, 0.006, 0.008, 0.010, 0.013, 0.016, 0.020, 0.025, 0.032, 0.040, 0.050, 0.063, 0.080, 0.100, 0.125, 0.160, 0.200, 0.250, 0.320, 0.400, 0.500, 0.630, 0.800, 1.00, 1.25, 1.60, 2.00, 2.50, 3.20, 4 .00, 5.00, 6.30, 8.00, 10.00sec

NOTICE

About "OVER" display of radiography conditions When set-up radiography condition "kV", "mA", and "mAs or sec" exceeds the permissible value of equipment, OVER sign lights up. X-ray exposure is prohibited when the over sign is turned on. Please reset the appropriate radiography conditions.

86/146

Operation Manual: DM-5125

3-4. SETTING APR MODE 1. APR mode selection APR mode is set up with the selection of Bucky, symbol on LCD 2 and patient size. 2. Radiography symbol selection Select the radiography symbol by switch [19]. Options are 7 parts and selected symbol is displayed on the LCD. 3. Radiography Region Selection Select the radiography region by switch [19] after the symbol is chosen. Selectable region is different according to each symbol. The selected region is displayed on the LCD 1. 4. Radiography method Selection Select the radiography region by switch [19] after the region is chosen. Selectable region is different according to each region. The selected method is displayed on the LCD 1. 5. Page up/down Selectable by page up/down key when the displayed menu on the LCD 2 is more than 7. 6. Density Selection Increase or decrease the mAs/sec selection by density selection switch. The selected mAs/sec will be increased when the density is +1~+5. The selected mAs/sec will be increased when the density is -1~-5. 7. Change of radiography condition The radiography conditions can be changed with the same method for “manual radiography” The over LED can be lighted when the conditions are changed. The x-ray exposure is prohibited in this case and special cautions are required. 8. Memory of the Changed Radiography Conditions The changed radiography conditions can be saved by pushing the “WR” switch for 2 sec until the buzzer sounds.

Memory of the changed Radiography condition Be careful that other radiography conditions can be selected if you choose

CAUTION

the selected radiography size switch. You cannot save the memory while the over LED is lighted and when the density setting is not 0.

87/146

Operation Manual: DM-5125

9. Method to save the APR data (example) When you save the head (symbol), skull (region), LAT (method). a. Choose the Bucky. (Select the bucky out of four) b. If you push SW27 (Page Up), SW28 (Page Down), SW29 (WR) at the same time APR store menu will be displayed on the LCD2 as following. c. SW19 LCD2

1. Modify

SW21 2. Write

SW20

SW23

SW25

3. Delete

SW22

SW24

SW26

If you select the “Write”(SW21), the numbers from 1 to 8 will be displayed on the LCD2. SW19 LCD2

SW21

SW23

SW25

1

2

3

4

5

6

7

8

SW20

SW22

SW24

SW26

d. If you select SW19, SW21, and SW23 in order, the numbers on the LCD1 will be displayed as following. The saving places for SW19, SW21, SW23 will be allotted by order. e. The following symbols will be displayed on the LCD2.

LCD1

1

2

SW19 LCD2

3

SW21

SW23

HEAD

CHEST

ABDOMEN

HIP

UPPER EXT

LOWER EXT

SW20

SW22

SW24

SW25 SPINE SW26

Select the “HEAD”(SW19) of the LCD2. The head will be displayed on the number of the LCD1 and the region will be indicated on the LCD2.

88/146

Operation Manual: DM-5125

LCD1

1 HEAD

2

SW19 LCD2

3

SW21

SW23

SKULL

SINUS

MANDIBLE

NASAL CAV

TM JOINTT

DENTAL JAW

SW20

SW22

SW25 MAXILLA /→

SW24

SW26

[ /→] mark on the right bottom of LCD2 means that it has the next page. Move the previous and next page with page up/down switch. f. If you select “SKULL”(SW19) on the LCD 2, skull will be indicated on the No. 2 of LCD1 and method will be displayed on the LCD2.

LCD1

1 HEAD

2 SKULL

SW19 LCD2

3

SW21

SW23

AP

PA

LAT

OBL

TOWNE

WATERS

SW20

SW22

SW25 AXI

SW24

/→ SW26

g. If you select “LAT”(SW23) on the LCD2, the LAT will be displayed on the No. 3 of LCD1 and “wirte ok?” will be screened when all the selection work of the menu is completed.

LCD1

1 HEAD

SW19 LCD2

2 SKULL

SW21

3 LAT

SW23

SW25

SW24

SW26

Write OK? SW20

SW22

h. If you push “WR”(SW29) after you select the kV, mA, sec, Body Size, they will be stored in the locations above.

89/146

Operation Manual: DM-5125

i. If you push the SW27 (Page Up), SW28(Page Down) at the same out of APR menus, they will be returned to the conditions before selection. 10. The change of the selected APR data When you modify the APR settings where the HEAD (Symbol), SKULL (Region), LAT (Method) are stored on the location 1,2,3. a. Push the SW27(Page Up), SW28(Page Down), SW29(WR) at the same time. Then, the LCD 2 will display as following. SW19 LCD2

1 Modify

SW21 2 Write

SW20

SW22

SW23

SW25

3 Delete SW24

SW26

b. Press SW19 and select “modify”. If you select the “modify” function, stored APR menus will be displayed according to the selected bucky. SW19 LCD2

SW21

SW23

SW25

SW22

SW24

SW26

HEAD SW20

If only the word “head” is displayed, it means there is only “head” on the APR menu. c. Press SW19 and select the “HEAD”(Symbol). If you select the “head” it will be displayed on the LCD1 and the skull (region) will be displayed on the LCD2.

LCD1

HEAD

SW19

SW21

SW23

SW25

SW24

SW26

SKULL

LCD2 SW20

SW22

d. Push the SW21 and select the skull (region). If you selelct the skull it will be displayed on the LCD1 and LCD2 as following.

90/146

Operation Manual: DM-5125

LCD1

HEAD

SW19

SKULL

SW21

SW23

SW25

SW24

SW26

LAT

LCD2 SW20

SW22

e. Press SW23 and select “LAT”(Method). After the selection of the LAT, change the value of kV, mA, sec and store the changed menus by pressing the WR button.

LCD1

HEAD

SW19

SKULL

SW21

LAT

SW23

SW25

SW24

SW26

Modify OK?

LCD2 SW20

SW22

11. The deletion of the stored APR setting (example) You use this key when you intend to delete the stored APR data. a. If you want to delete the stored APR data above. b. If you push the SW 27,28, 29 at the same time, they will be displayed as following and select the “delete”.

SW19 LCD2

1 Modify SW20

SW21 2 Write SW22

SW23

SW25

3 Delete SW24

SW26

c. If you select the “delete”, the stored APR menus will be displayed on the LCD2 according to the selected Bucky.

91/146

Operation Manual: DM-5125

SW19 LCD2

SW21

SW23

SW25

SW22

SW24

SW26

HEAD SW20

d. If you push the SW19 and select the Head (symbol), the “head” will be demonstrated on the LCD1 and skull (region) will be on the LCD2.

LCD1

HEAD

SW19

SW21

SW23

SW25

SW24

SW26

SKULL

LCD2 SW20

SW22

e. If you select the “skull” by pressing SW21, the LCD1 and LCD2 will be displayed as following.

LCD1

HEAD

SW19

SKULL

SW21

SW23

SW25

SW24

SW26

LAT

LCD2 SW20

SW22

f. If you push the SW23 and choose the LAT, the question “Delete OK?” will be screened on the LCD2 to ask weather you want to delete.

LCD1

HEAD

SW19

SKULL

SW21

LAT

SW23

SW25

SW24

SW26

Delete OK?

LCD2 SW20

SW22

92/146

Operation Manual: DM-5125

g. If you want to delete the selected APR menu, press the WR (SW29). If you select the wrong menu on the APR selections, you can move to the previous steps by pressing the SW27 and SW28 at the same time.

3-5. RADIOGRAPHY 1. PHOTOGRAPHY PREPARATION After setting up radiography conditions, assemble the cassette equipped with the film on the x-ray table. If the X-ray irradiation field lamp switch attached to the variable X-ray beam- limiting device is pushed, the lamp will be lighted up. X-ray irradiation field is adjusted with a radiography position, and radiography preparation is ended. 2. X-ray radiography using operation panel a. Press the X-ray exposure preparation switch. Ready LED will be lighted on when all preparation work is completed. b. If you press the x-ray preparation switch and exposure switch at the same time, the x-ray shooting will be started. c. X-rays exposure LED will be lighted on and the buzzer sounds during the exposure. 3. X-ray radiography using hand-switch a. Press the hand switch one time and make it at ready condition. The x-ray ready LED will be lighted on when it is prepared to make exposure. b. When the equipment is ready to make exposure, press the hand switch again and the x-ray shooting is started. c. During exposure, the exposure LED will be lighted on and the buzzer sounds. d. If the X-ray exposure LED is turned off or buzzer does not sound, it means the radiography work is completed. Then, you can stop pressing the hand switch. e. If you touched out and stopped x-ray shooting during exposure, the check code will be displayed.

If you touch out from the hand switch during exposure, the x-ray shooting is

CAUTION

stopped and you cannot get clear pictures. Please touch off your hand after the x-ray exposure signs and the buzzer sound are turned off except the case you want to stop the x-ray exposure urgently.

You cannot make radiography while the over LED is lighted on. Change the

CAUTION

conditions for the radiography in those cases.

93/146

Operation Manual: DM-5125

SECTION 4. TROUBLES 4-1. ERROR CODE AND CHECK CODE In case any problems happen during the operation of the device, they will be indicated on the kV/mA segment. 1. Error Code The error code is displayed when problems, such as abnormalities, occur to the equipment. While the error code is displayed, radiography operation is forbidden for safety. A power supply should be switched off and on again and an error code will be cleared. If the problem of equipment is not solved at the time of the power-supply reswitched on, the error code will be displayed again. Since it is the failure of equipment when an unusual state is not recovered, please inform the service section of our company or our agent. 2. Check Code Unlike an error code, a check code is the warning sign when wrong operations are applied by the user. X-ray radiography operation is forbidden for the safety during the check code. The way to release a check code is pressing the WR switch for 2 seconds. CH01 will be released if thermal switch is off during the re-supply of the power.

94/146

Operation Manual: DM-5125

ERROR CODE LIST Error Code

DESCRIPTION

Err 01

CHARGE MONITOR ERROR (HT-CONT)

Err 02

OVER TIME ERROR (OP PANEL) AT STAND BY

Err 03

READY-OUT SIGNAL ERROR AT SAND BY

Err 04

X-RAY OUT SIGNAL ERROR AT STAND BY

Err 05

READY SWITCH INPUT ERROR AT STAND BY

Err 06

X-RAY SWITCH INPUT ERROR AT STAND BY

Err 07

HAND SWITCH INPUT ERROR AT STAND BY

Err 08

HAND SWITCH X-RAY INPUT ERROR AT STAND BY

Err 09

ROTOR DETECTER SIGNAL AT STANDBY

Err 10

FILLAMENT CURRENT DETECTER ERROR AT STANDBY

Err 11

‘kV FEEDBACK ERROR AT STANDBY

Err 12

‘mA FEEDBACK ERROR AT STANDBY

Err 13

ROTOR DETECTER ERROR AT EXPOSURE

Err 14

FILLAMENT CUREENT DETECTER ERROR AT EXPOSURE

Err 15

‘kV FEEDBACK ERROR DETECTER ERROR AT EXPOSURE

Err 16

‘mA FEEDBACK DETECTIVE ERROR EXPOSURE

Err 17

NOT DEFINED

Err 18

X-RAY TIME OVER IN SEC MODE

Err 19

X-RAY TIME OVER IN mAs MODE

Err 20

X-RAY TIME OVER IN AEC MODE

Err 21

HV INTERLOCK ERROR

Err 22

IGBT FEEDBACK ERROR

Err 23

‘kV FEEDBACK OVER DETECTER ERROR AT EXPOSURE

Err 24

‘mA FEEDBACK OVER DETECTER ERROR AT EXPOSURE

Err 25

NO ZERO-CROSS OR NO INPUT

Err 26

ZERO-CROSS OVER OR WRONG INPUT FREQUENCY

Err 27

BUCKY 1 OK SIGNAL ERROR (HT-CONT)

Err 28

BUCKY 2 OK SIGNAL ERROR (HT-CONT)

Err 29

AEC A/D DATA-IN ERROR (HT-CONT)

Err 30

AEC A/D DATA-IN ERROR (HT-CONT) AT NON-EXPOSURE

Err 31

I/F BOARD CONDITION ERROR

Err 32

AEC BOARD CONDITION ERROR

Err 33

SMALL-FOCUS SELECTION ERROR (HT-CONT)

Err 34

LARGE-FOCUS SELECTION ERROR (HT-CONT)

95/146

ACTION

RESUPPLY OF THE POWER

Operation Manual: DM-5125

CHECK CODE LIST Check Code

DESCRIPTION

Measure

Ch 01

TUBE THERMOSTAT OPERATION

Reset

Ch 02

When another switch is pushed during exposure

Release by “WR” switch

Ch 03

When ready switch on OP PANEL is detached during exposure

Ch 04

When X-RAY switch on OP PANEL is detached during exposure

Ch 05

When a hand switch (ready) is detached during exposure

Ch 06

When a hand switch (X-RAY) is detached during exposure

Ch 07

Shortage of power-supply capacity for exposure

Ch 08

Simultaneous exposure prohibition

Ch 09

Warning to X-rays room door opening

Ch 10

AEC Level Error

Power-supply re-switch

Reset

# CH01 is will be released if the TUBE terminal switch is off, in case resupply of the power.

96/146

Operation Manual: DM-5125

SECTION 5. MAINTANENCE 5-1. MAINTANANCE AND PERIODIC OPERATOR Maintenance has check menus carried out daily by the user and periodic inspection by technical service engineers who completed special training programs on the repair of the generator. User checks items based on " Daily Checking Items by the operator " and should keep data records. The following checks and maintenance procedures, together with the suggested intervals, are the manufacturer’s recommendation for most effective Periodical Maintenance schedule for this generator. Exclusively service personnel specifically trained on medical X-ray generator must perform service tasks here described. The first periodic maintenance service should be performed six (6) months after installation, and the subsequent service every twelve (12) months. When a trouble and a poor part are discovered, please stop use of equipments, turn off a power supply, and contact a near service station.

5-2. DAILY CHECKING ITEMS BY THE OPERATOR Check following items according to the regular inspection plans. Please refer to the daily inspection method regarding the way of checking.

In order to avoid the danger of an electroshock, please do not remove the cover of equipment by any means.

CAUTION

Only the expertise person who received training performs the check and repair accompanied by removal of a cover.

INTERVAL

Check item

Daily

Operation Panel

○ ○ ○

Radiographic Condition



Radiation field lamp



Warning light in use



Air Conditioner Temperature and Humidity

97/146

Weekly

Monthly

Operation Manual: DM-5125

5-3. WAYS OF DAILY CHECKING (CHEKING MENUS) Please check clear environment inside the operation room and x-ray room. The system is designed to operate with normal conditions under following environment. Please keep following temperature and humidity. The recommended operating environment is follow. Temperature

: +10 - +35 ℃

Relative humidity

: 45 – 85 %, 45~65 % (X-ray room)

In the environment with a rapid change of temperature and humidity, it is easy

CAUTION

to generate dew condensation. Installation of an air conditioner is recommended

The ingredient that makes silver etc. sulfurate may be contained in the exhaust

CAUTION

gas of an automatic development machine. Please install the independent air conditioner.

Operation Panel Please check whether each switch of an operation panel lights up. When you clean the surface of an operation panel, it should be wiped with the soft and fine cloth. When using a cleaner, it is made to contain in cloth and is used.

Don't blow a liquid upon the operation panel directly.

CAUTION

It goes into the inside of equipment from a crevice, and becomes the cause of failure.

The check of radiographic condition Before the commencement of work, trial radiography is performed and film density, unevenness, and etc. are checked. Radiation field lamp It confirms whether a radiation field lamp lights up before the commencement of work. Warning light in use It confirms whether warning light in use switch on before the commencement of work.

98/146

Operation Manual: DM-5125

PART 2. X-ray Tube

1. E7132X

Refer to the Attached E7132X Data Sheet.

99/146

Operation Manual: DM-5125

PART 2. X-ray Tube

2. E7239X

Refer to the Attached E7239X Data Sheet.

100/146

Operation Manual: DM-5125

PART 2. X-ray Tube

3. E7240X

Refer to the Attached E7240X Data Sheet.

101/146

Operation Manual: DM-5125

PART 2. X-ray Tube

4. E7252X

Refer to the Attached E7252X Data Sheet.

102/146

Operation Manual: DM-5125

PART 3. Bucky Table

1. DM-FT-1[Fixed Table-Top Table]

103/146

Operation Manual: DM-5125

RADIATION WARNING INFORMATION X-ray and Gamma-ray and dangerous to both patient and operator unless Established safe exposure procedures are strictly observed. The useful beam can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an unskilled operator. Adequate precautions must always taken to avoid exposure to useful beam, as well as to leakage radiation from within the source housing or to scattered radiation resulting from the passage through matter. The authorized to operator, participate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the currently established safe exposure factors and procedures described in the National Concil on Radiation Protection and Measurement ( NCRP ) “ Medical X-ray and Gamma-ray protection for energies up to 10 Mev- Equipment design and Use “NCRP Report-33 as revised or replaced in the future. Those responsible for the planning of X-ray and Gamma-ray equipment installation must be thoroughly familiar and comply completely with the structural shielding requirements outlined in NCRP-34 as revised or replaced in the further. Failure to observe these warnings may cause serious or fatal bodily injuries to the operator, patient or those in attendance.

OUTLINE DATA LEGAL OBLIGATION You have legal obligation that failure to follow the manufacturer’s instructions or manufacturer’s will affect radiation of any component by user or installer which well affect radiation safety, causes the user or installer to assume full responsibility for that product. The manufacturer is required to prescribe maintenance information and a schedule of performance to ensure equipment complies with specified para-meters After installation the installer must supply manufacturer’s data including this manual to the equipment purchaser or user. The equipment purchaser or user must follow maintenance instructions.

104/146

Operation Manual: DM-5125

SECTION 1. INTRODUCTION This manual provides information relative to the fixed table-top table. The table system, hereinafter to as the DM-FT-1. The apparatus describe in this a heavy-duty, fixed table-top table is designed for general radiographic examinations.

SECTION 2. DESCRIPTION This table system is designed for general diagnostic radiography. The table base fixed the table top. The bucky is electromagnetically locked.

2-1. EQUIPMENT SUPPLIED 1. Table Leg 2. Table Top 3. Bucky

SECTION 3. SPECIFICATION Table top

:

2000 mm ( 78.74” ) X 740 mm ( 29.13” )

Table top height :

620 mm ( 24.4” )

Table top – to – film distance :

65 mm ( 3.34” )

Table top maximum load :

136Kg

Table top inherent filtration :

1.0 mm Aluminum equivalent at 100 KVP

Bucky travel :

230mm ( 20.66” )

Weight :

40 Kg

Power :

Voltage :

24 VAC

Frequency :

50 or 60 Hz

105/146

Operation Manual: DM-5125

SECTION 4. INSTALLATION 4-1. PRELIMINARY CONSIDERATION The following factors must be consider, in order to planning an efficient Installation site layout. - Location of tube stand. - Location of auxiliary equipment (bucky stand, cassette holder, etc.). - Incoming power for X-ray generator and auxiliary equipments. The tube stand, floor-ceiling tube stand should be installed before installation of DM-FT-1 table system, and tube stand must be aligned to longitudinal travel. The floor should be leveled the area where the system is being installed.

This table shoule not be used in the presence of explosive or flammable gases

WARNING

as might be found in an operating room.

4-2. Electrical Wiring Interconnecting cable is supplied with this system, it is highly recommended to use adequate size of floor trough from control console to high tension transformer and rear of the DM-BT-1 table. The trough should have removable cover flush with the finished surface of the floor. If the power is being supplied from wall disconnect box, #12 AWG stranded wire should be used for table power and #10 AWG for equipment ground.

4-3. System Calibration 1. Set the table into position under the tube stand. 2. A plum line can be used from the X-ray, align the table center line along the longitudinal travel of the movement. Note : To align the table bucky verses tube stand, the table-top should be removed from the DM-FT-1 table. 3. The table-top should be level (Longitudinally and transversely) at this time, if necessary the table should be shimmed. Note : Shims are supplied by manufacturer.

106/146

Operation Manual: DM-5125

4. Secure the table to the floor. Note : Use adequate hardwares for that particular installation. See location of the table anchor bracket in the major component layout. 5. Check all connections between each components before apply the power to the table. 6. Check bucky lock release function by pulling the cassette tray toward to the front direction. 7. Check the table-top center at the center of the table-top is aligned to the table bucky. 8. The table bucky function should be checked, make an exposure and verify that bucky grid starts oscillating during the X-ray generator exposure cycle.

107/146

Operation Manual: DM-5125

SECTION 5. MAINTENANCE At least once a month external surfaces and exposed tracks and bearing surfaces should be wiped a cloth slightly moistened with light oil to provide sufficient lubrication for smooth operation and protect against rust. The following checks for proper operation should be made: - Bucky travel free and smooth, and grid operating properly. - All locks should be operating properly. - Spacing between locks and bearings should be checked. - All electrical connections and coil cords in table should be proper condition.

DM-FT-1 DIAGRAM

Bucky (Generator)

DC24V

Solenoid

108/146

Operation Manual: DM-5125

PART 3. Bucky Table

2. DM-BT-1[4-way Table]

109/146

Operation Manual: DM-5125

RADIATION WARNING INFORMATION X-ray and Gamma-ray and dangerous to both patient and operator unless Established safe exposure procedures are strictly observed. The useful beam can produce serious or fatal bodily injuries to any persons in the surrounding area if used by an unskilled operator. Adequate precautions must always taken to avoid exposure to useful beam, as well as to leakage radiation from within the source housing or to scattered radiation resulting from the passage through matter. The authorized to operator, participate in or supervise the operation of the equipment must be thoroughly familiar and comply completely with the currently established safe exposure factors and procedures described in the National Concil on Radiation Protection and Measurement ( NCRP ) “ Medical X-ray and Gamma-ray protection for energies up to 10 Mev- Equipment design and Use “NCRP Report-33 as revised or replaced in the future. Those responsible for the planning of X-ray and Gamma-ray equipment installation must be thoroughly familiar and comply completely with the structural shielding requirements outlined in NCRP-34 as revised or replaced in the further. Failure to observe these warnings may cause serious or fatal bodily injuries to the operator, patient or those in attendance.

OUTLINE DATA LEGAL OBLIGATION You have legal obligation that failure to follow the manufacturer’s instructions or manufacturer’s will affect radiation of any component by user or installer which well affect radiation safety, causes the user or installer to assume full responsibility for that product. The manufacturer is required to prescribe maintenance information and a schedule of performance to ensure equipment complies with specified para-meters After installation the installer must supply manufacturer’s data including this manual to the equipment purchaser or user. The equipment purchaser or user must follow maintenance instructions.

110/146

Operation Manual: DM-5125

SECTION 1. INTRODUCTION This manual provides information relative to the fixed table-top table. The table system, hereinafter to as the DM-BT-1. The apparatus describe in this a heavy-duty, fixed table-top table is designed for general radiographic examinations.

SECTION 2. DESCRIPTION This table system is designed for general horizontal diagnostic radiograpy. The table base support the table top. The transverse and longitudinal table-top locks are controlled by a foot SW at the front the table base. The table motious and bucky are electromagnetically locked. The bucky station AEC (Automatic Exposure Control) are option. A single station AEC is available as an potion on the bucky by using chamber.

2-1. EQUIPMENT SUPPLIED 1. Table Leg 2. Table Top 3. Bucky (Speed is adjustable to the max. 5 Step) 4. Compression device (Option)

SECTION 3. SPECIFICATION Table base :

1262 mm ( 49.68” ) X 650 mm ( 25.59” )

Table top :

2000 mm ( 78.74” ) X 740 mm ( 29.13” )

Table top height :

700 mm ( 27.55” )

Table top longitudinal :

350 mm ( 13.77” )

Table top transverse travel :

150 mm ( 5.9” )

Table top – to – film distance :

65 mm ( 2.55” )

Table top maximum load :

136Kg

Table top inherent filtration :

1.0 mm Aluminum equivalent at 100 KVP

Bucky travel :

230 mm ( 9.05” )

Weight :

70 Kg

Power :

Voltage :

220 VAC

Frequency :

50 or 60 Hz

Current :

3A

111/146

Operation Manual: DM-5125

See Outline Data

112/146

Operation Manual: DM-5125

SECTION 4. INSTALLATION 4-1. PRELIMINARY CONSIDERATION The following factors must be consider, in order to planning an efficient Installation site layout. - Location of tube stand. - Location of auxiliary equipment (bucky stand, cassette holder, etc.). - Incoming power for X-ray generator and auxiliary equipments. The tube stand, floor-ceiling tube stand should be installed before installation of DM-FT-1 table system, and tube stand must be aligned to longitudinal travel. The floor should be leveled the area where the system is being installed.

This table shoule not be used in the presence of explosive or flammable gases

WARNING

as might be found in an operating room.

4-2. Electrical Wiring Interconnecting cable is supplied with this system, it is highly recommended to use adequate size of floor trough from control console to high tension transformer and rear of the DM-BT-1 table. The trough should have removable cover flush with the finished surface of the floor. If the power is being supplied from wall disconnect box, #12 AWG stranded wire should be used for table power and #10 AWG for equipment ground.

4-3. System Calibration 1. Set the table into position under the tube stand. 2. A plum line can be used from the X-ray, align the table center line along the longitudinal travel of the movement. Note : To align the table bucky verses tube stand, the table-top should be removed from the DM-FT-1 table. 3. The table-top should be level (Longitudinally and transversely) at this time, if necessary the table should be shimmed. Note : Shims are supplied by manufacturer.

113/146

Operation Manual: DM-5125

4. Secure the table to the floor. Note : Use adequate hardwares for that particular installation. See location of the table anchor bracket in the major component layout. 5. Check all connections between each components before apply the power to the table. 6. Turn SW 1 (ON & OFF) on. 7. All locks should functioning. 8. Verify that power on & off indication green lamp should be on. 9. Depress the foot-SW, the longitudinal and transverse locks should be released. 10. Check bucky lock release function by pulling the cassette tray toward to the front direction. 11. Check the table-top transverse center, depress the foot-SW and move the table-top is aligned to the table bucky, the BZ1 (table center indication Buzzer) should be sounded 12. The table bucky function should be checked, make an exposure and verify that bucky grid starts oscillating during the X-ray generator exposure cycle.

114/146

Operation Manual: DM-5125

SECTION 5. MAINTENANCE At least once a month external surfaces and exposed tracks and bearing surfaces should be wiped a cloth slightly moistened with light oil to provide sufficient lubrication for smooth operation and protect against rust. The following checks for proper operation should be made: - Bucky travel free and smooth, and grid operating properly. - All locks should be operating properly. - Spacing between locks and bearings should be checked. - All electrical connections and coil cords in table should be proper condition.

115/146

Operation Manual: DM-5125

DM-BT-1 SCHEMATIC

116/146

Operation Manual: DM-5125

DM-BT-1 BUCKY SCHEMATIC

117/146

Operation Manual: DM-5125

PART 3. Bucky Table

3. DM-RFT-5[RF Tilting Table]

118/146

Operation Manual: DM-5125

SECTION 1. APPLICATIONS Used in combination with an X-Ray high tension apparatus, an X-Ray tube unit, An X-ray tube support unit , this table is applied for the following X-Ray examinations: 1. Fluoroscopy 2. Spot Filming 3. Bucky Radiography

119/146

Operation Manual: DM-5125

SECTION 2. CONSTITUTIONS Dimensions : See the following drawing

1. DM RFT-1 consists of Radiography table, bucky device unit, spot film device, Image intensifier, Camera, Monitor with trolley 2. The relative accessories : compression band, hand grips and foot rest. 3. Standard Accessories - Built-in Bucky Device

1Set

- Foot Rest

1Set

- Hand Grips

1Set

- Foot Switch

1Set

120/146

Operation Manual: DM-5125

SECTION 3. SPECIFICAIONS Parts Table tilting

DM-RFT-5 type

motor driven

range

standard : +90˚ ∼ -15˚

speed

90˚/20 sec., 50/60Hz

type

motor driven

option : +90˚ ∼ -20˚

horizontal : ±50cm Tabletop sliding

movement range

lateral : ±11cm(option) speed

Tabletop

Bucky device

Spot film device

Image assembly

focal distance

41cmH

filtration

less than 1.0mm AL. Eq. at 100KV

tabletop to film

6.5cm

type

solenoid reciprocating

grid

standard : 40LPC 10:1, option : 60LPC, 70LPC

cassette range

13x18cm ∼ 35x43cm(5x7" ∼ 14x17")

tarvel range

100cm

AEC mode

Ion chamber(option)

parking

slide parking from left to right

locking

electromagnetic lock

longitudinal movement

60cm

transversal movement

±12cm

vertical movement

14cm ∼ 50cm

grid

standard : 40LPC 10:1, option : 60LPC, 70LPC

AEC mode

Ion chamber(option) manual operation

under couch

motor driven

Image intensifier

12", 9", 6" selectable

TV Camera

CCD

Monitor

23", 17", 15", 14" selectable

Trolley

four wheels, two shelves, 51cmLx44cmWx138cmH 220VAC, 3KVA, 50/60Hz, single phase

net weight

Dimensions Installation space

5cm/sec., 50Hz Lminated 60mmT

Power requirement Weight

3.6cm/sec., 60Hz

material

Compression cone Field collimation

upward(tilting position) : 50cm

650kg except image assembly approx. 200cmL x 72cmW x 85cmH

minimum

310cmL x 158cmW x 260cmH

121/146

Operation Manual: DM-5125

SECTION 4. BASIC TECHNICAL CONCEPT Model DM-RFT-5 is designed for performing Radiography & Fluoroscopy as well as Castro Intestinal examinations. This unit not only makes your patients feel comfortable and safe during the operation, but also makes your diagnosis extremely efficient and reliable.

4-1. Safe and High Operability of Table 1. Table-top sliding switch and table tilting switch are provided on the tower, near fluoroscopy screen or I/I unit. 2. 81cm high table-top is suitable for treating decubit us patients.

122/146

Operation Manual: DM-5125

SECTION 5. INSTALLATION

123/146

Operation Manual: DM-5125

SECTION 6. OPERATION SWITCHES The table tilting switches are provided on the tower, near the fluoroscopy screen, the tilting angle can be read by the indicator provided on the front cover. The table tilting switch is provided on the spot film device.

6-1. Table Switches

124/146

Operation Manual: DM-5125

6-2. Switches of Fluoroscopic Operation Panel (Spot Film Device)

SPOT

Collimator long

Collimator long

X-RAY

: Tube open

: Tube close

FLOURO.

Collimator lat

ON

: Tube close

TILT DOWN Collimator lat TILT UP

TILT SLIDE

SLIDE

UP

DOWN

: Tube open

DOWN -15°

VERTICAL LOCK

ALL LOCK REALSE

125/146

Operation Manual: DM-5125

PART 4. Tube Stand

1. DM-FCT-1 [Floor to Ceiling Mounted]

126/146

Operation Manual: DM-5125

SECTION 1. INTRODUCTION Congratulations on your purchasing our floor – to – ceiling tube support. This manual will guide you how to use our tube support effectively Please read the contents thoroughly.

SECTION 2. FEATURES Dongmun Tube support Units series are designed for heavy duty usage with all tables and erect bucky stands, assuring a wide and precise tube positions with stability, rigidity of tube arm and, support capacity. Its fully counter balance providing all positions and angulations.

SECTION 3. SPECIFICATIONS Model Rails Stand

DM-FCT-1 : Floor To Ceiling Floor

2,440mmL

Ceiling

2,440mmL

vertical

650-1,900mm

Transversal

880-1,230mm

Longitudinal

Min. 1,400mm

Rotating of Vertical Column

±180º manual detents at each 90º positions

Rotating of Horizontal X-ray Beam

±180º, manual lock

Lock Release System

Electromagnetic Lock

Lock Release System

Vertical, Transverse, Longitudinal, Rotating

Components

Tube Stand, Tube Arm & Ring Base Assembly Floor/Ceiling Rails Angle Indicator

Power requirements

DC 24V

Dimensions

2100x1,020x2,150Hmm

127/146

Operation Manual: DM-5125

SECTION 4. CONSTRUCTIONS

128/146

Operation Manual: DM-5125

129/146

Operation Manual: DM-5125

SECTION 5. OPERATION This tube support unit is operated by operation to move longitudinally, Transversally, and vertically, and locked electromagnetically.

1. Special care must be taken not to let the X-Ray tube unit come in contact with the table or other objects during the traveling. 2. Avoid rough handling and giving a jerk. 3. The functions of Operation switches: Angle Indicator

Rotating Lock S/W Longitudinal Lock S/W

Vertical Lock S/W

All Lock S/W

Horizontal Lock S/W

130/146

Operation Manual: DM-5125

SECTION 6. MAINTENANCE AND INSPECTION 1. The rails should be kept clean from dust 2. Check the parts regularly. 3. External inspection a. Any defect in the unit b. Lack of lubrication c. Any deformation or corrosion d. Connecting parts (Pins and screws): Wear, deformation, breakage, loosening or missing of those parts. e. Bearings: Wear, rusting or lack of lubrication f. Other check Points Any failure or defects in the section related with the X-Ray Tube support.

131/146

Operation Manual: DM-5125

PART 5. Bucky Stand

1. VB-1 [Bucky Stand]

132/146

Operation Manual: DM-5125

SECTION 1. DESCRIPTION Dongmun Vertical Bucky Stand , DM-VB-1 is designed for space efficiency and adopt free standing mounting and counter weight balance system, It is engineered for use of radiography of Chest Skull, Thoracic, spinal. The height is widely adjusted from 50 cm to 165 cm above the floor level.

SECTION 2. SPECIFICATIONS Vertical Column

Counter weight balance system, Floor mounted

Vertical Travel

Approximate 120cm, Center to floor distance 50cm – 165 cm above floor level.

Front Plate to Film distance

Approximate 3 cm

Inherent Filteration of Front plate

Less than 0.5 mm Al. eq. at 100kVp

Vertical Locking System

Electromagnetic lock. DC 24V

Cassette Loading

Left or Right loading Reciprocating

Standard : Solenoid Option : Motor

Bucky Device Unit

Film Formats

Up to 14 X 17 “ ( 35 X 43 cm )

Grid

Standard

Bucky 103LP1 10;1 F.D = 40-72”

Option

Bucky 150LPI 10;1 F.D = 40-72”

Dimensions

590W

x 570D x 1800mmH

Power requirements

110/ 220 VAC , 50/60Hz

Options

Bucky device for 15angle, upward tilting

133/146

Operation Manual: DM-5125

SECTION 3. OPEARTION 3-1. Components

Stand

Jaw Rest Vertical Movement Bucky Device

Lock Switch

Base

3-2. How to Operator 1. Power ON: 1. Tube Stand power will be supplied automatically by Power ON of console. 2. Turn ON the Console Power. a. To operator the system, push the power ON switch of Console. b. After power on, system will diagnosing automatically during 15 seconds. 2. How to Use a. Movement of Electromagnetic lock - This product uses electromagnetic lock method. - Electromagnetic lock will be released when push the lock switch right side of Cassette tray. Electromagnetic lock will be locked when released the lock switch.

134/146

Operation Manual: DM-5125

- Bucky device will be release if pull the hand grip of cassette tray and push the cassette tray and then bucky device will be locked. b. Movement of Bucky device - When turn ON the system, bucky device will be locked by electromagnetic and if push the right side of lock switch, electromagnetic lock will be released and if release the lock switch, electromagnetic lock will be locked again. - Do not move bucky device in locked condition. c. How to use Bucky device - Bucky device controls (reduce) X-ray scattering and grid was installed. - When turn ON the system, bucky device will be locked by electromagnetic, when pull cassette tray hand grip, electromagnetic lock will be released. - Pull out cassette tray hand grip and make a position with moving left or right. d. How to install cassette on cassette tray - Insert cassette to the cassette tray as below image. - Take cassette out from cassette tray as reverse procedure. 1

2

3

135/146

Operation Manual: DM-5125

Circuit Diagram

136/146

Operation Manual: DM-5125

PART 6. High Tension Cable

1. High Tension Cable

Refer to the Attached Specifications Data Sheet.

137/146

Operation Manual: DM-5125

PART 7. Collimator

1. LB-50 [Collimator]

138/146

Operation Manual: DM-5125

SECTION 1. PRODUCT DESCRIPTION 1-1. General Description This X-ray collimator is designed for such applications as shown in the specifications of this manual and must be sued with the X-ray tube to restrict and define the area covered by the X-ray beam and also to reduce stray radiation. The beam that is collimated by the two control knobs shall over the smallest area of the patient necessary to produce the required radiograph. The collimator that allows a step less adjustment of a field size utilizes a light localizer to project a visible beam with coverage equivalent to the coverage of the X-ray beam.

1-2. Features and Performance 1. Provided with a push button switch timer to automatically switch off the lamp of light localizer in 30 seconds. 2. The lamp of light localizer easily replaceable with minimum error of the same lamp position and the lamp filament position can be easily adjusted. 3. Maximum shutter opening of collimator is 48cm x 48cm at SID 100cm 4. Minimum X-ray field size at SID 100cm is 0cm x 0cm (blades overlapped) 5. The average illuminance of light field at SID 100cm is 160Lux or more. 6. The contrast ratio on the light field edge is 4 : 1 or more, which allows an easy identification of light field and its size. 7. Leakage radiation is within 35mR/h at a position 100cm from the focal spot.

139/146

Operation Manual: DM-5125

1-3. Specification Applications

: Manual-driven adjustable collimator for radiography

Maximum kVp

: 150kVp

Maximum field size

: 48cm x 48cm at SID 100cm

Minimum field size

: 0cm x 0cm at SID 100cm

Project lamp

: Halogen lamp, 24VAC, 150W

Lamp timer

: Push button type 30sec, electronic timer

Light projection

: More than 160Lux at SID 100cm

Inherent filtration

: 1.2mm Al equivalent

Tube focal distance

: 60mm as the standard tube focal distance from tube focus to the bottom surface of range

Field coverage scale

: 13, 15, 18, 20, 23, 25, 28, 30, 36 and 40cm

SID indicate scale

: 90, 100, 120 and 180cm

Leakage radiation

: Not more than 35mR/h

Rotation of radiation field

: +180

Cone rail tracks

: Mounted (148W x 164L cm)

Dimension

: 183W x 188D x 178H mm

Weight

: 5Kg

1-4. Dimension and Function

1

X-ray field adjust knobs

7

Installation plate of flange

2

Push button switch of lamp

8

Stoper knob

3

X-ray field coverage scale

9

Front cover

4

Back cover

10

Cone rails

5

Body case

11

Center cross marked acrylic plate

6

Flange ring

140/146

Operation Manual: DM-5125

SECTION 2. INSTGALLATION AND OPERATION 2-1. Responsibility of Installer It is the responsibility of the installer to correctly install this collimator as instructed in this manual. The installer must certify that the X-ray equipment on which the housing assembly the collimator is installed is in compliance with KFDA regulations.

2-2. Electrical Supply Requirement It is imperative that a sufficient power is supplied to the collimator lamp to obtain the specified voltage under onload condition being measured at the lamp socket.

2-3. Installation Method The following instruction should be read and thoroughly understood attemption to installed collimator. It is assumed that all persons installing, adjusting and maintaining this collimator have the knowledge and experience necessary to achieve the objective outlined. 1. Unpacking Carefully unpacking the collimator and check each piece for visible damage incurred during shipment. Any damage should be referred to the agency that delivered the collimator. 2. Collimator mounting procedures Collimator mounting consists of removing the flange ring from the collimator, fastening the flange ring to the X-ray tube port bottom and then fastening the collimator to the flange ring assembly. a. Remove the flange ring (A) from the top of the collimator by remove the two installation plates of ring (C). b. Use the proper screw length to insure that the mounting screw engage at least 5 threads into the X-ray tube port bottom, or mounting, surface tapped holes if adaptor plate is used. c. Attach the flange ring assembly to the X-ray tube (or tube support) with mounting spacer (no supply) inserted between the ring and the X-ray tube support. d. Thread the side safety screw (B2) in about 6mm, leaving the flange screw (B1) flush with the inside of the ring. e. Lift the collimator up to the installed flange ring. f. While the supporting the collimator, thread the side safety screw (B1) in, then tighten both safety screw (B1 and B2). g. Check the installation thoroughly for correct and safe mounting.

141/146

Operation Manual: DM-5125

2-4. Operation 1. Push lamp timer and confirm that light field is illuminated. 2. Adjust collimator position so that the centering cross mark of collimator is aligned with the center of an image receptor. 3. Collimator light field to the image receptor size operating the 2 manual control knobs.

2-5. Cares to be taken on Operation 1. Do not fail to use within the specified maximum tube voltage. 2. Do not use installing on those X-ray tube which focal spot distance are different from the ones specified. 3. If the lamp replacement become necessary, do it after the inside temperature decreased sufficicently. 4. Never fail to use the same type of lamp as the specified. 5. Do not touch on the projection lamp, glass mirror and front acryil plate except it become necessary to do so.

142/146

Operation Manual: DM-5125

6. In case that it is illuminated successively 5 times or so, keep it off for about 2 minutes till next operation for cooling purpose.

2-6. Identification of Illuminance. 1. Set up Mount the collimator on the tube housing assembly. Locate the photometer on the measurement field, and adjust so that the distance between the tube focal spot and input surface (image receptor) of photometer is 100cm. 2. Measurement Adjust the blade to light field of 35cm x 35cm. Switch off the room light, and switch on the collimator lamp at the adjusted supply voltage. Measure the illuminance in the center of the 4 quadrants of the light field. The average illuminance of the center of the 4 quadrants shall not be less than 160Lux.

143/146

Operation Manual: DM-5125

SECTION 3. INSTRUCTION FOR MAIN ADJUSTMENT 3-1. Lamp replacement and adjustment method of lamp filament position. 1. For replacement of lamp, first remove the back cover (4), but never move the front acrylic plate (11). 2. Mark sure first that the lamp surface temperature sufficiently decreased, and pull lamp out of the lamp socket. 3. Wrap over the replacement lamp surface with a soft cloth or the like so as not to allow a finger to touch on the surface, and insert the lamp into the lamp socket so that the lamp pins will reach to the utmost depth of lamp socket. 4. After the replacement work, illuminate the lamp by pushing lamp timer, and measure if the centering cross mark is in the center of light field. 5. If the centering cross mark is deviated from the center, adjust the lamp filament position as follows. (See the drawing below.) a. Measure A size B size Increase A size is larger, loosen the adjustment screw and move the adjustment plate (Z) op-down adjustment then fasten the screw. b. Measure C size D size Increase C size is larger, loosen the adjustment screw and move the adjustment plate (X, Y) to right side then fasten the screws. In case D is larger, move the adjustment plate to left side.

144/146

Operation Manual: DM-5125

3-2. Glass Mirror Replacement For replacement of glass mirror, first remove the back cover and body cover, and open the collimator shutter to the maximum positions in both longitudinal and transverse.

145/146