Service Manual 360 Fully Automatic Hematology Analyzer Service
Vapi, Gujarat
Web: www.merillife.com
Phone: +91 260 305 2100
Chapter 1 Foreword
1
Chapter 2 Function and operation
2
2.1 Simple operation
2
2.2 Founction Introduction
3
2.2-1 Founction Module Diagram
4
2.2-2 Connection Port of Complete Machine
7
Chapter 3 Maintenance
14
3.1
14
Routine Maintenance
3.1.1 Clean the Appearance
14
3.2 Daily Maintenace
15
3.3 Weekly Maintenance
15
3.4 Monthly Maintenace
15
3.5 Yearly Maintenance
16
3.6 Maintenance Before Transport or For the Instrument that will not be used for along time
16
Chapter4 Troubleshooting
17
Chapter5 Liquid System diagram
25
Chapter6 Mechanical Assembly
26
Chapter 1 Foreword This manual includes the information about functionality, daily operation and maintenance. To achieve the optimal performance, be sure to operate and maintain the instrument according to this manual. Functionality Operation Maintenance. In the chapter 2 There are diagrams related to express function and operation. In the chapter 3 Maintenance is mentioned. In the chapter 4 This includes Errors encountered during operation of the machine and trouble shooting.
1
Chapter 2 Function and operation 2.1 Simple Operation First: Front panel
Display
Sample Needle
Power Light Printer Cover
Aspiration Switch
2-1 Diagram Front Panel
2
Chapter 2 Function and operation Second:Operation Process 1) Test blank once,after that it will show the data for the background check .Repeat blank test till the values of all the five parameters namely WBC, RBC, HgB,PLT and MCV are within the acceptable range. 2) Once Blank is in acceptable range patient samples can be tested . Put the patient sample under the sample probe and press the start key . Ensure that bubble is not entering during the aspiration procedure . Move the sample tube along with the sample probe so that sample enters correctly into the measuring chamber. 3) Before the test is completed no other operation can be executed . After completion the needle come back to the home position and is ready for the next sample aspiration. 4) Press the key “print”. The results are printed on the thermal printer paper 5) After finishing measure sample click the right key,the menu will pop-up last bar of the menu is “shutdown” key,press this key,it will show “please shutdown power “ a few minutes later, operator can turn off the power.
3
2.2 ) Function Introduction 2.2-1. Function Module Diagram 1. Interface board
P104 bus
Motor control and
CPLD--->printer driver One-way
Printer driver
CPLD total WBC-RBC
drive unit
control unit
CPLD