Automatic Biochemical Analyzer Maintenance Manual FACA-261 FACA-301 FACA-401 LABOMED, INC. www.labomed.com

1

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Contents

Chapter 1, The basic principles of instrument ............................................................................. 1 Chapter 2 Working environment ................................................................................................ 1 Chapter 3 Safety matters .......................................................................................................... 1 Chapter 4. Maintenance personnel requirements and tools the maintenance personnel need ....... 2 4.1 Maintenance personnel requirements....................................................................... 2 4.2 The knowledge maintenance personnel should know ...............................................2 4.3 Maintenance tools .................................................................................................2 Chapter 5 Instrument maintenance .......................................................................................... 3 5.1Day maintenance condition ....................................................................................3 5.1.1 Boot maintenance ..............................................................................................3 5.1.2 Shutdown maintenance ......................................................................................3 5.2Weekly maintenance .............................................................................................3 5.3Months maintenance condition ...............................................................................4 5.4Years maintenance condition .................................................................................4 Chapter 6 Instrument basic structure and major components .................................................... 5 6.1Main control board ................................................................................................5 6.1.1 Basic functions .................................................................................................... 5 6.1.2 Basic structure ...................................................................................................5 6.2Motor driver module ..............................................................................................6 6.2.1 Function .............................................................................................................. 6 6.2.2 Basic structure ...................................................................................................6 6.3Power supply system ............................................................................................ 11 6. 4 Analysis system .................................................................................................. 12 6.4.1Optical system ................................................................................................. 12 6.4.2Heating system ................................................................................................ 12 6.4.3Detection system ............................................................................................. 16 6.5Add sample system ............................................................................................. 20 6.5.1Function ............................................................................................................ 20 6.5.2 Reagent sample trays installation method .......................................................... 21 6.6Liquid road system ............................................................................................... 23

2

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 6.6.1 liquid road system functions ............................................................................... 23 6.6.2 Liquid road system structure ............................................................................. 23

3

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 7 Unworn components correction replacement ........................................................... 24 7.1 The replacement to lamp............................................................................................ 25 7.2 The replacement for the piston of injector .................................................................... 26 7.3 The replacement to probes ......................................................................................... 27 7.4 The replacement to cuvettes ........................................................................................ 29 7.5 The replacement to fuse .............................................................................................. 29 7.6 Adjustments for GAIN & OFFSET................................................................................. 30 7.6.1 The adjustment to OFFSET ............................................................................... 30 7.6.2 The adjustment to GAIN ................................................................................... 31 Chapter 8. Common malfunctions and processing.................................................................... 32 Appendix 1: FACA-261-301-401 instrument wiring diagram ...................................................... 39 1.SX400 motor drive system wiring ................................................................................... 39 2.SX300 motor drive system wiring diagram ...................................................................... 40 3.SX261 motor drive system wiring diagram ...................................................................... 40 Appendix 2. The maintenance manuals of electrolyte part............................. 1. The basic principle of electrolyte analysis module............................... 2. Electrolyte analysis module for maintenance...................................... 3. Main components. .......................................................................... 4.Electrolyte parameter Settings .......................................................... 5. The instrument replacement of parts ................................................. 6. Common fault and processing ..........................................................

4

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 1

The

Basic principles of instrument Biochemistry analyzer’s test principle mainly uses Lambert - Beer's law, and mix reagents and patient's serum samples according to certain mixing ratio, the mixture in certain temperatures than color in the dish, and continuous measurement are incubated on certain wavelengths light absorbed, finally, according to the absorbency value (change) calculated automatically determine the concentration of matter. Chapter 2 Working environment Normal working environment should satisfy following conditions: 1. The environment temperature: 10℃ ~ 30℃; 2. Relative humidity: 30% ~ 70%; 3. The atmospheric pressure: 860hPa ~ 1060hPa; 4. Power supply requirements: equipped with Stabilizer, 220V ~ + 10%, 50Hz + 1 Hz; 5. No strong electromagnetic interference and mechanical vibration, no corrosive gas, no glare direct illuminate; 6. The instrument should connect to the ground. Chapter 3 Safety matters This instrument in the process of installation and maintenance should pay attention to safety matters: 1. We use 220V power supply for the machine, so you should cut off the power before operating to avoid the risk of electric shock. 2. When doing liquid road maintenance, you can't let the fluid to the circuit boards and power, and prevent burning circuit. 3. When operating circuit, you must cut off the power, and we forbid operating with power. 4. In the maintenance of equipment, beware of biological pollution. 5. Instruments must connect to the ground in an excellent way. 6. Heating circle is forbidden filling with water.

5

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 4 Maintenance requirements 4.1 Necessary of maintenance This instrument for precision instrument, equipment maintenance personnel must be trained by company’s after-sales engineer or after my company training professional personnel. 4.2 The knowledge maintenance personnel should know 1. The basic knowledge of analog and digital circuit 2. Get to know the performance and test method of the basic components 3. Compute knowledge 4. Some safety maintenance knowledge Tools of maintenance

Tools list: cross screwdriver medium 1, cross a screwdriver to trumpet 1, a word screwdriver medium 1, a word screwdriver trumpet 1, pliers, 150 x 19 wrench a, hex wrench set, 150W electric iron

a, soldering 1 ,digital multi-meter 1 only.

6

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 5 Instrument maintenance 5.1 Day maintenance condition In use, maintenance is instrument must be carried out every day; the process is divided into a boot maintenance and shutdown maintenance. 5.1.1 Boot maintenance Turn on the host the power switch, and computer

ensure there is sufficient deionized water in

water bottle, and empty the waste bottle. Ensure that all pipe no twisting instrument, and warm up 30 minutes before running the instruments maintenance icon

click “wash all reaction cup”

Turn on the click daily

go to” cup blank” test, and fill water,

Check blank value and save it (three times, each time all saved)

directly pumped

You also can run the default boot process “one key maintenance” also when doing boot maintenance, you need to check importantly: 1) Water which fill into reaction cuvette should reach to half of total volume, otherwise you need to adjust injection time 2) If there are more than 30% cuvette marked in blue, recommend replacing all the cups. 3) Water level in cuvette should be higher than the light beam center, which will ensure get correct AD value. 5.1.2 Shutdown maintenance: Wash all reaction cuvettes

wash needle 4times

fill water

shutdown

or execute the default shutdown maintenance process 5.2 Weekly maintenance 1. Every week do needle maintenance one time, you can use alcohol cotton cleaning tip of probe. 2. Cleaning all reagent bottles and pure bucket.

7

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 3. Cleaning the dust and dirt on the surface of the instrument.

5.3 Monthly maintenance condition 1. Check the use of reaction cup reaction cup, use cleaner undertakes immersing, remove reaction cup, wash with clean lens paper surface of the glass of reaction, and manual is wiped clean wiped process prevent scratch inspection face and, when necessary, replace reaction cup. 2. 3 months after the metal rod of dilution is added lubricant, (such as Vaseline, lubricant) lubricant usage not too much, otherwise, it will affect mechanical components operation. 3. After 3 months to dilute the piston overtime lubricant, is like Vaseline, can increase the abrasion resistance, increase air tightness and reduce resistance. 5.4 Years maintenance condition Years of maintain must professional engineers or by the company after company training for personnel operation. 1. Clean attached in the instrument of the components inside the dust, such as fans, sensors, etc. 2. Check of all parts of the belt running situation of necessary adjustment of strap firmness. 3. Inspection instruments, liquid and use with aging line for replacement, after two years of all pipelines must be replaced 4. Replace syringe piston. 5. Inspection, control board AD voltage values, put the passage below 45,000 adjusted to get the 55,000, in 20 -- 100's value. 6. I represent ABAD backup database files, and after the backup to the letter "S" beginning in the clean the database tables. 7. Light bulb two years later, the need for replacement

8

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 6 Instrument basic structure and major components 6.1 Main control board

6.1.1 Basic functions Main control board is the core components, is instrument for the brain, main instrument can communicate with a PC software various instructions issued by the execution. Control through the various moving parts, acquisition amplification, and signal detection and issued AD transform PC for software processing analysis. 6.1.2 Basic structure The motherboard major chip and interface for DSP processors, CPLD programmable logic chips, memory chips, and AD transform chip, AD amplifier chip, RS232 interface, reaction plate control parallel interface RS485 parallel bus, as shown in figure 1 shows: Check line

OFFSET Adjust potentiometer 

Reset light‐coupler indicator

Counting light‐coupler indicator 

Power  socket 

DSP chip 

RS232

CPLD

Figure 1

RS485

9

Reaction plate control interface

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 6.2 motor driver boards

6.2.1 Function Receive the instructions from main control board, collected signals from the sensor board, drive control motor and electromagnetic valve movements, etc. 6.2.2 Basic structure It has Motor driver module, motor cage, bus board, general motor boards, high-power motor boards, reaction plate motor boards; 5V power makes parts board, etc. As figure2shows Liquid detection board                       

  Syringe  motor boards 

                                 

10

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] V3.0  reaction  plate  motor  boards 

5V power board 

General motors boards 

  

  

11

 

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Figure 2

12

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 6.2.2.1 Motor cage is to install each function motor board stents mechanical components, cooling fan with above. 6.2.2.2 Bus plate surface are motors board slots and motor line and light coupling line connector, connects to a motor, sensors, electromagnetic valves, etc. Fig.3 shows

Valve and pump interface

RS485 

Motor line interface

Light‐coupler interface

Figure 3

13

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

6.2.2.3 General motors board drive add kind arm, reagent sample plate etc motor running, and the solenoid switch. This motor board may, according to needs, set up its address (1-15 as shown in figure 4 shows between), and software Settings would need to address plus 16. Address of setting method: Motor board microcontroller chips four platoon insert needle below for 8421 yards, the 1,2,4,8 from extroversion order. Use short sub plug will its connection, as shown in figure 4 can be connected motor board address for 6, software Settings is 6 + 16 = 22, 1

1

2

2

4

4

8

8

14

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

ON=1 OFF=1

---This kind Switch, we start to use since 11.2014 6.2.2.4 Syringe motor plate is used to control the dilution of operation, and for the solenoid switch. This motor board has four address fixed address, for the 16th and 17th respectively, 18, 41. As shown in figure 4 shows:

Figure 4

15

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

6.2.2.5 Reaction plate motor plate is controlling reaction plate operation, motor board directly through parallel serial no SCM by mainboard control. Its address software set to 0, as shown in figure 5 shows:

Figure 5

16

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] V2.0 reaction plate motor boards, used to for the old-fashioned motherboards V3.0 reaction plate motor boards, used to FACA-401\301\261 etc. need 24V power supply for these boards. V4.0 reaction plate motor boards, used to FACA-401, use, directly put into slot

V4.0

V3.0  Figure 5.1 6.2.2.6

5V power plate is used to supply power for the sensor and motor boards. If instrument

reattachment, motor unlocked, check the 5V power board and 33V power, that 5V power is supplied by 33V transformation. as shown in figure 5.2 shows:

Figure 5.2

17

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 6.3 power supply system Power system mainly have the power switch, filter, the isolated transformer, switch power, wire etc parts composition, the main function is to provide instruments work voltage, and control instruments switch machine. Circuit structure figure 6 shows:

Graphics ~220V output

~220V input

+5V

8A fuls fuse

+5V

switch

I O

220V transform

150

Cool/on off

150

Waste led Water led Waster overflow Absence water

Power switch

+24V3A GND +12V2A GND

220v

220v

220v

+15V1 0A Gnd

220V +12V5A Gnd -12V3A Gnd

+33v5 A Gnd

+13.8V Gnd

Temperature controll meter

10K 10K

1 2

cool2

Power lamp

cool

TM2

Red TM4

TM2

1 2

fan

TM2

fan

Pump 24v

24v Power control board

Black Cool lamp

Green

TM6

1 2 3 4 5 6

1 2

fan

1 TM2 2

Fan

4

TM2

1 2

fan

cool 1 TM2 2

TM2

2 3

1 2

cool TM2

Lam p

1

cool1

1 2

cool TM2

T e m p

Red purple orange Black Black Red

Red purple orange Black Black Red

DS200 without this part

18

1 2 3 4 5 6

220v

TM6 up

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Figure 6

19

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Various specifications power usage: Power

+12V 5A -12V 3A

12V 24V

33V 6A

13.8V 6A

the mainboard, bus board,

pump and

motors and

bulbs

temperature control board

fan

heating

specifications Use

When the instrument can't work, should check whether the corresponding have electricity power, or damaged. 6. 4 Analysis system Analysis system means reaction plate system and it is the core of an instrument, it is also the component that is used for chemical reaction and analysis the changing ABS of chemical reaction. All of the components are around analysis system to work. 6.4.1optical system The optical system, the spectrometer system after halogen lamp, optical fiber and filter is composed of, which filters used for 13.8 instruments V50W, light. 12V20W grating instrument use Structure as shown in figure 7 shows:

Receiver 

Light

Cuvette

A/D 

Filter Figure 7

6.4.2 Heating system Heating circle, temperature control board, temperature sensors, power conversion board (using for thermometer) and so on, circuit diagram as shown in figure 8 shown:

20

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Figure 8

If heating circle without temperatures, then check: 1) Whether there is resistance ,about 18 ohms around heat circle ,such as position ④show 2)In switch power board if there is 33V power input, such as position ③show 3) Examine whether there is 33V power for heating circle to work, such as position ④ show 4) Check the input of switch power board whether the voltage is 12V, such as position ②show If heating alarm, then check switching power board, and maybe chip is bad.

21

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Temperature control board adjusting instructions J5 

J4

KEYMODE

TENS

UNITS 

SITES 

J3

J13

J8 

J6

J14  Figure 9

22

S1

S2

S3 

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Name Function key

Function Display of PT100 channel temperature : reaction pan

P

(mode key)

temperature currently Enactment of PT100 channel control temperature:reaction

S

pan enactment temperature(41℃ as a rule) Enactment of PT100channel alarm temperature:upper

A

limit of reaction pan alarm temperature(50℃ as a rule) Enactment of AD590 channel control temperature :

F

enactment of heating water temperature(45℃ as a rule) Display of AD590 channel temperature: heating water

D

temperature currently Tens

Tens of displaying data

Units

Units of displaying data

Sites

The first bit behind the radix point of displaying data

2. Three key S1, S2, S3, functional see table 2.1 Serial number

Key-press

Function

1

S1

Mode key:The switch of function key P, S, A, F, D

2

S2

Decrease key:brief press,decrease 0.1;lasting press, decrease 1.0 per second

3

S3

Increase key:brief press,increase 0.1;lasting press, increase 1.0 per second

3.The interface in common use Serial number

Interface

J5

AD590 sensor(detect of heating water temperature)

J4

PT100 sensor(detect of heating loop temperature)

J3

MCU_JATG

J13

Interface of relay control

J8

Power supply of Temp-control

J14

Power supply of heating loop

J6

Input of 33V power supply

23

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 6.4.3 detection system Detection system mainly constitutes by detection board, detection lines and main control board. detection board as figure 10 shows:

Photoelectric diode

Figure 10 Resistance list ,Corresponding resistance as shown on the below list 340:4.7M 405:3M 450:1.5M

546:510K, 578:510K 620:300:K 670:200K

505:510K Note: When assembling detection board, the light may be not in the center of the cups, so you can use a few eccentric optical fibers to adjust. 6.4.4 analysis system maintenance of note points and replacement of parts 6.4.4.1 reaction plate code salver the adjustment 1. Reaction plate code salver top in after the position control silk, tighten yards disk. Reaction plate yards dish, shaft dish, is bearing hoop to pressure in place, the axis dish not fluctuation string move. As shown in figure 11 shows:

axis

screw 

 Gear disk 

Figure 11

24

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 2. Heating circle after adjusting the do, must tighten the screws nuts; Rotating reaction sets heating circle friction-free reaction plate phenomenon.

screw  Washing part

Reaction plate tray Reaction plate  PT100 Temperature sensing    fiber 

Figure 12 6.4.4.2 Fiber adjustments 1. Adjust fiber bronze pillars end in 2.5 mm in diameter to flare to the left. 2. Fiber each stocks have the Numbers, 1 340nm inspection channels for use, 2 405nm detection channels for use, other strands of optical fiber can be used at the remaining wavelength detection channels use 3. Use M3 * 3 top silk fiber, when fixed head top wire cannot prep above fixed heat laps will affect reaction plate surface or the rotation. 4. Finally use layering of the fiber axis pressure, check whether the fiber axis wear plate (figure 13 shows)   

Fiber pillars 

       

 

   Heating  loop 

screw      

Fiber cover        

 

 

25

   Fix on fiber 

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Figure 13

26

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 5. After the install fiber, beams happens to be located on reaction cup central; Debugging of 5mm can be used when a wide hard white cardboard observation light path, if flare is too high, try doses used; Low may take cup bottom, causing detection is not accurate. Figure 14 shows:

Figure 14 6.4.4.3 Sensor replacement Reaction dish light-coupler has two, divided into long light-coupler and short light-coupler. Long light the decoupling control reaction dish reset, if long light-coupler to malfunction, reaction plate will not reset; Short for counting light-coupler to light the decoupling control AD sampling, if short light-coupler to malfunction, cleaning reaction sets, washing needle collide reaction plate, instrument testing abs would appear 0.0000 1. The short and long with code salver light-coupler to the position of the two relations, code salver light-coupler to slit center in turns friction-free light-coupler phenomenon. 2. Short and long time coupling light-coupler to install the stent with code salver tangent parallel. 3. Dish reattachment the length of the control board light-coupler two lights must at the same time, if two indicator light is not the same as the bright, there will be detected for the jagged response curve, or clean needles collide reaction plate. As shown in figure 15, 16 shows:

27

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Figure 15

Countin g sensor

Reset ssensor

图 16

4 core bipolar machine

Reaction 3M belt Figure 16

28

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

6.4.4.3 Filter replacement Filters in damp environment filter moldy mass and after a long time the job also can appear the phenomenon of aging, causing machine detecting absorbency is not accurate. 1. Will have damaged filters removed. 2. Will filter installation in filter set, mark arrow outward, test set, and then in pressure filter, prevent its loose. Note cannot use tools or hand direct contact with filter glass surface, cannot use water or other organic solution to wipe the surface of the filter. 3. Will filter suit in heating ring, install filters with heating circle of inner surface level, otherwise affect reaction plate rotation or damage reaction cup. Use hex silk lock, roof top of silk cannot prep above heating circle will affect reaction plate surface or the rotation. As shown in figure 17 shows:  

Photoelectric diode covert 

Heating  circle 

   

Filter     

 

     

Filter cover 

Screw 

  Figure 17 SX260 from the reaction plate 24 began installing 340 filters, 27 installation 405 filters in turn on. SX301 from the reaction plate 4 began installing 340 filters, 7 installation 405 filter in turn on. SX401 from the reaction plate 90 # began installing 340 filters, 87 installation 405 filters, clockwise analogy 6.5 Add sample system 6.5.1 Function Add sample system, main function in the process of response analysis instrument in the reaction automatically adding need reagent and samples.

29

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] A reagent dish, sample trays, plus sample arm, plus sample needle, needle, stirring device, diluted makes parts such as pipeline composition. FACA-261 each needle to respond to dish corresponding cup bit: reagent for reaction plate 1 cup, stirring to 1 cup needle also FACA-301 each needle to respond to dish corresponding cup bit: reagent 1 for reaction plate 1 cup for reaction plate, reagent 2 for no. 76 cup, stirring, cup 74 FACA-401, each needle to respond to dish corresponding cup bit: reagent 1 for reaction plate 1 cup, sample for reaction plate 86 # 2, reagent for reaction cup dish no. 76 cup, stirring 1 aimed at 40, 41 cup, stir 2 aimed at 30 and 31 cup 6.5.2 Reagent sample trays installation method. 6.5.2.1 Motor replacement method: Motor is a stable parts generally not damaged. 1. The four screws loose motor stents, make strap in flabby condition, remove damaged from motor stent. 2. Will belt outfit synchronization pulleys, fixed on the four corners of the screw motor. 3. Tighten belt, the four screws fastening motor stents. 4. Adjust synchronization pulleys, make the belt without friction code salver and synchronization pulleys, and tighten synchronization pulleys top silk. As shown in figure 18 shows

Motor fixed 

motor

Stents fixed screw   

Figure 18

30

Synchronization pulleys 

belt 

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] How to judge fault or motor fault: can use good board this way to control a control problem of all the way, if there is not normal, it can diagnose the problem; motor whereas the judgment is panel also 6.5.2.2 Belt adjustment and replace: Strap in long after the job, relax and the phenomenon of ageing. Need adjustment firmness or replacement of the belt. 1. Adjust tightness, loosen your belt of the four screws, motor stents to tighten belt, and the four screws fixed motor stents can. 2. Replace belt need to loosen the motor of the four screws, dismantling scaffolds the code salver light-coupler, remove aging or damaged belt 3. Put new belt, taut, fixed motor stents, four screws light-coupler to recover. 4. Check without friction yards dish, leather belt snaking phenomenon. As shown in figure 19 shows:

Belt 

  

Synchronization pulleys gear disk

light-coupler Figure 19

6.5.2.4 The replacement of Peltier 1. Will damage refrigeration piece removed. 2. Will face, refrigeration reagent plate surface and radiator of the tablet surface to smear evenly heat conduction silicon grease. 3. Will refrigeration piece refrigeration clingy reagent plate metal plate surface clingy heating surface, refrigeration pad, finally installation radiator. 4. Electrify affixed to the test its cooling effect, 10 minutes after 10 ℃ temperature refrigeration board obvious above.

31

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] As shown in figure 20 shows: Heating and refrigeration blocks face divisional distinguish method: The black line in the left, refrigeration blocks for refrigeration noodles, or upper hand touch with short electricity, heat, distinguish face

Metal plate  Refrigeration piece 

Radiator 

Cooling pad

 

Figure 20 6.6 liquid road systems

6.6.1 Liquid road system functions Liquid road system function is added in the instrument during sample automatic cleaning plus sample needle, and after completion of the instrument response analysis be automatic cleaning colorimetric container. 6.6.2 Liquid road system structure Liquid road system mainly is composed by the water pump, liquid pump, the 3-way magnet valve, clinic solenoid valve, cleaning pool, washing parts. In the process of pipeline connection, the move is natural, and piping doesn't appear the following wrong: 1. Pipeline 90 degrees to bend. 2. Have crushed piping. 3. Cleaning pool of liquid pipe bending into U fonts. 4. Pass long, curved pipe in the cabinet

32

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Figure 21

33

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 7 Unworn Components Correction Replacement In order to ensure the normal functioning and reliability, when the instrument requires maintenance and the timely and effective for some parts, it may change the necessary calibration. Note: ● The following maintenance and replacement of some parts can be completed by the users after engineers train you on the instrument.

34

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 7.1 The replacement to lamp Being damaged or running for 2 years. The steps as follows: 1, Turn off the analyzer and wait for at least 15 minutes before next step 2, Open the back cover of the analyzer. 3, Unscrew and remove the lamp cover and then remove the lamp housing. 4, Unscrew four screws fixing the lamp and remove the damaged lamp. 5, Connect the chinaware housing of the lamp again and fix the lamp with the four screws. And ensure the spring piece is in the right direction. 6, Fix the cover with four screws. 7, Close the back cover. As shown in figure 22 shows:

Lamp 

Lampshade 

 

Fixed shrapnel  Ceramic socket 

Figure 22                  

35

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

 

36

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]  

  Lamp holder Figure 22.1 Replacement of 12V lamp Normally, we need to change the lamp after 1000 working hours 1, Turn off power and wait 15 minutes until lamp is cool down. 2, Open the rear panel 3, Release the screw of the lamp 4, Replace a new lamp If you notice the looseness or deformation in replacement of lamp, you need to change lamp holder, In cause of light changes. 7.2 The replacement for the piston of injector The steps as follows: 1, Enter into “Device run/ device maintenance”, and then click “Device reset’ 2, Open the window of the injector. 3, Unscrew the tail screw of injector’s piston, as A in figure 23. 4, Unscrew the screw for fixing the glass of syringe and then remove the glass and the piston, as A in figure 24. 5, Take out the old piston from the glass and clean the new piston with alcohol. Smear the grease at the middle of the piston and insert the piston into the glass carefully and ensure that the piston reach the top of the glass. 6, Adjust and fix carefully with screw. Ensure the piston plumbs down to the center of the glass. As figure 24 shows. 7, Draw the piston down about 3mm along the glass gently 8, Screw the screws on the two ends of piston tightly. 9, Close the window of syringe well with screws. Connecting two valve 

Connection reagent or sample 

 

Connection reagent or sample 

     

  Connecting two valve 

1~2mm

37

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]    

Tight nuts   

  Fixed nuts 

  Figure 24 

Do not scratch the piston during installation. Or else, it is easy to make the piston leak so that testing results are not accurate.



Ensure the piston is situated in the center of the glass. Or else, it may damage the glass or reduce lifespan of the piston.



While the piston has been used for three months, it must be smeared grease.



New Syringe don’t need change piston, it is a ceramics piston

7.3 The replacement to probes The method of replacing samples probe is the same as the regent probe’s replacement. Concrete steps as follows: 1, First of all, open the cover of the probe and then cut off swathe for fixing the probe with pliers. 2, Unscrew the pinching screws with fixing wire and the compacting screw of the probe. 3, Remove the adding sample probe and the soft tube covered the adding sample probe. 4, Install a new adding sample probe and a soft tube. The structure of the soft tube is as the below figure 25 shows. 5, Fix the probe and the wire well with the compacting screw and swathe 6, Close the cover.

Need flux   

图 26

Needle block the ray

38

  Keying‐in position

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Figure 25

Figure 25-1

Five core level of wiring

Connect   to   main   wire 

Level detection board 

39

Connect to sample needle 

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Level  oscillating board 

40

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Figure 26 In the figure above the second part is concussion board, this V2.2: concussion board is improved by the version 2.1 concussion board. Improvement methods: directly change the C3 capacitance to R0 resistance in the V2.1 board. Note: V2.2 board and V2.1 doesn't allow exchange use, please production workshop, after engineers, warehouse must be no mistake. Connections top-down, as shown in figure shown: 26 (1) Blue wire for level signal (2) The yellow line for oscillation board (3) Black wire for ground (4) Red is + 5V power supply line (5) White is impact-proof signal line; 7.4 The replacement to cuvettes Cuvettes are dirty during using; it needs to test “cuvettes water blank”. If the blank absorbency of the cuvettes exceeds 0.025A, clean again or replace them. 

Do not touch the two detecting surface of cuvettes.



The top surface of the cuvettes placed must be level. Or else, it is easy to leave water in the cuvettes during cleaning so that the testing results are not accurate.



Use the same batch cuvettes as possible.

7.5 The replacement to fuse Steps are as follows: 1. Take the fuse out from the accessories kit. To mention that the fuse of mainframe is 8A and water heating system is 4A. 2. Turn off the power of analyzer and pull out the plug. 3. Pull the power wire from the power socket of mainframe and elicit the fuse housing. 4. Take out the damaged fuse housing and install a new fuse into fuse housing and then plug the fuse housing into power socket. 5. Switch on the power plug.

41

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 7.6 Adjustments for GAIN & OFFSET 7.6.1 The adjustment to OFFSET Generally speaking, the reaction curve is not desired and the absorbency of reagent is unusual (more than 2.2) or repeatability is not good, it is necessary to check the OFFSET value. 1, Enter into “Blank test” interface of the software after analyzer is reset. As the Figure 27 shows: 2, Choose “Monitoring” and take out the cuvettes where each wavelength lies, and then put them into the black cups which are used for sheltering light. As the Figure 28 shows: 3, Observe the real-time voltage and adjust the resistor of every channel of the right arrange to the numerical value needed.(20-50)

Figure 27

42

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Figure 28 7.6.2 The adjustment to GAIN 1. Enter into “Blank test” interface of the software after analyzer is reset. 2. Click on the water injection, waiting for water injection completed. 3. Observe real time voltage and adjust the resistor of every channel to reach about value 54000.

43

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Chapter 8 Common malfunctions and processing Malfunction

Possible reasons

phenomenon The

analyzer

doesn’t

work

when

the

power

is

on(the indicator

light

is off)

The

power

cable

is

disconnected. 2.

Main program

1. Examine connection whether is in good condition.

is

not

2. Turn off the analyzer and restart the

started up.

analyzer after 5 minutes

3. Fuse is broken.

3. Examine the fuse.

4. The a l t e r n a t i n g

4. Examine the socket whether has

c u r r e n t socket has no

power.

power.

The

liquid

leaks

from

analyzer

1.

Disposal steps

1. The tube is damaged.

Turn off the power firstly; wipe the

2. Connections break off.

liquid leaked away, then examine

3. Dilutor syringe leak

whether the connections break off, the

4.3-way valve is jammed.

tube is undamaged and the pump &

5. The pump leaks water

the air bubble release case exist

6. The b u l b -release

leakage, and check 3-way valve is

c a s e leaks water.

jammed

44

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 1.

The

analyzer

hasn’t 1. Reset the COM1 of the analyzer

chosen the right COM port, 2. Examine the RS232 cable MAINCOM=O (COM1).

2. RS232 is not connected item “ analyzer’s running/device”

The

well or it’s inside wires is 4. Replace the COM of computer

connection between

the

computer and the analyzer is failed

3. Reset the testing channel under the

(the

indicator power is on)

of

5. Examine the signal line of motor’

connected well

3. The setting of testing control channel of the analyzer is 6. Turn off the analyzer and restart the analyzer after 5 minutes.

wrong.

4. The computer’s COM is 7. Copy the backup documents or use the function “the restore of

wrong.

the

5. The signal line of motor’ hardware parameter” control breaks off. (6 line)

8. Close one window of them , or

6.

restart computer

Main program is not

started up and the mother

45

9. Choose the appropriate channel

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] board is something wrong。 7. of“

The

according to the mother board

parameter

Hardware”

of

the

analyzer has missed 8. The software runs two windows at the same time. 9.12- channel is set wrongly in the setting of parameter of the hardware

1. Sample probe jammed

1. Get it through by a thin needle and

2. Dilutor syringe doesn’t

run “probe maintenance”

work

2. Examine the dilutor syringe and

3. Dilutor syringe leaks

motor and check the setting of address

4. Sample probes touches

code of the dilutor

the bottom of cuvettes

3. Replace the piston of dilutor syringe

5. There’s something wrong

4. Adjust the height of probe

46

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

Sample is not absorbed in.

with liquid level sensor so 5. Get down the sensitivity of liquid that sample probe can’t get level into the cup

sensor

and

examine

the

connection, make sure the probe

6. The tube is damaged and touches the surface of sample every the connections break off.

time

7. Relevant magnetism valve 6. Check the tube is broken and or is damage or out of work

connections break off. 7. Examine the electromagnetic valve by the action testing program whose voltage is 12V.

1. The line of liquid level 1. Check the line of liquid level sensor Samples probe sensor is broken or is not 2. Adjust the sensitivity of liquid level touches

sensor(8-14)

the connected well.

bottom of cup 2. The sensitivity is too high 3. Get down the position to make the or

can’t

into the cup

distance

get or low.

between

the

reagent

3. The parameter that the

&sample probe and the bottom of

high of probe is wrong set.

reaction cuvettes as 3mm and the

47

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 4.

The

motor

board

is distance between sample probe and reaction cuvettes and the bottom of

damaged 5. Sample

cups

is

serum cuvettes are both 3mm.

anamorphic, can’t put into 4. Examine electromagnetic valve by the motion testing program

the bottom of the holder.

5. Choose the eligible sample cup 1.

Water suspends

in

the tip of probe after cleaning.

Tube

is

damaged

or

1. Examine the tube.

broken ,which leaks air

2. Get it through by a thin needle or

2. Reagent probe is jammed.

run “probe maintenance”

3. Relevant electromagnetic

3. Examine electromagnetic valve by

valve

the motion testing program

4.

Dilutor

syringe

exist

4. Replace the piston or glass tube of

leakage.

dilutor syringe.

5. Sample probe pricks

5. Adjust the position of sample probe

the bottom of the cuvettes.

and the sensitivity of liquid level

6. There are air bubbles in

sensor.

tube.

6. Take “probe cleaning”5 times.

48

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 1. If the short probe drops 1. Take the tube away, clean and press water

after

cleaning,

it it

in

the

opposite direction

with

indicates the one-way valve syringe to make it closed completely is not closed completely.

2.

Adjust the down depth of cleaning

2. If the short probe drops block to make it touch the bottom water after

cleaning,

the properly.

down depth of the probe is 3. Adjust the long needles in level line the needle of too washer water

touch

the and the cleaning block is 1mm lower

drops bottom of the cuvettes.

than other long needle; test it by

3.

deep

The

needles

to

longer drop

cleaning maximum motor step, the cleaning

water:

the block can touch the bottom properly,

seven needles of cleaning but the other needle don’t touch the device a r e n o t i n l e v e l line;

they

bottom.

absorb water 4. Open electromagnetic valve and

incompletely and leak.

clean membrane.

4.

5. Fix the cleaning block with pastern.

When

reposition

the or

analyzer’s the

needle

49

6. Adjust the vertical position of

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] cleaning is taken, the short cleaning block and ensure cleaning needle drops water. It should block lie centre of cuvettes after be

that

3-way

adjusting the position of holder.

electromagnetic valve is not closed well. 5. Cleaning block is loose. 6. Cleaning block is not in the middle 1. The reaction cuvettes are 1. Replace the reaction cuvettes. dirty

2. Adjust the facula of fiber to make it

2. The facula of optical path lie in the middle and the distance doesn’t locate in the middle between the facula and the bottom of reaction cuvettes is 1.5—2mm。

of cuvettes.

3. The voltage of reaction 3. Adjust the GAIN of signal and offset cuvettes is not in the usual 4. Check the tube range.

5. Replace the piston or glass tube and

4. The tube is damaged.

add grease

50

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] \

The

testing 5. The dilutor syringe exists

samples probe.

results is not for leakage. accurate

6. Adjust the position and height of

6. The position and height of

7.

sample probe are unsuitable.

sample。

7.

8. Check the electromagnetic valve by

Reagents

and

control

Replace the reagents and control

sample are out of date.

the motion testing program。

8. Electromagnetic valve is

9. Adjust the sensitivity of liquid level

damaged

detecting board

9. The liquid level sensor is

10. check and reset the parameters

out of work.

11. Check whether connection to

10 The setting of testing

ground is well AD voltage is stable and

parameters is wrong

the testing board is loose or leak light.

11 The testing voltage and

Examine whether the filter is damp

absorbency are unstable.

and the sides of fiber is fixed well.

12. Sample & reagent probes

Besides, the range of facula isФ2—

are jammed.

Ф2.5.

13.

The

temperature

51

of

12, Get it through by a needle and take

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] reaction plate cann ot be “probe maintenance” controlled well. 14. RS232 cable is loose.

13. Check or adjust control of the

15. There’s water left behind temperature of reaction plate in

the

because

reaction the

cuvettes 14. Fix the RS232 cable reaction 15. Adjust the position of reaction

cuvettes are unstable or the cuvettes and the height of cleaning tip cleaning tip is not in the to keep them level. appropriate height.

16. Adjust the encoding disk to make it

16. Encoding disk rubs the locate in the middle and clean the sensor

sensor with alcohol or blow, or replace

17, The testing cable isn’t it. 17. Connect the testing cable again.

connected well.

18 Not short pin2,3 of no 18. Short pin2,3 of no used channel used channel

19. Fix the counting sensor(short

19. The sensor used for

sensor)

counting is loose.

20. Extend the time of water-in and

20. The water used for make the capacity as a half of the cleaning cuvettes are not cuvettes. enough

52

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

The voltage is zero or rather lower the water

when cuvettes blank

value is tested

1. The lamp is damaged

1. Replace the lamp

2. The wire connected with

2. Examine t h e c o n n e c t i o n o f

lamp is loose.

t h e lamp.

3. There’ is something wrong 3. Examine or change the power with

voltage

of

power supply 4. Examine the signal connection

supply.

4. There is something wrong 5. Replace the mother board with the cable connected 6. Replace the filter with signal device 5.

There

is

7. Examine whether the

RS232

something connection is loose or change it.

wrong with mother board 6.

The filter is damp

7.

RS232

connection

53

is

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] useless. 1. Sample & reagent probe is

1. Get it through by a needle and take

jammed.

“probe maintenance”

reagents 2. Leakage exists in Dilutor

The

and the water syringe. can’t

be

3. Adjust the sensitivity of liquid level

3, Liquid level detecting is

distributed

detecting board 4. Examine electromagnetic valve by

and useless.

absorbed

Change the piston or glass tube

4. Relevant electromagnetic

the motion testing program

valve is damaged.

5.

Examine the tube

5. The tube breaks off. The setting of

The

attribute

of

file

device

“Hardware” is read only.

Cancel the “read-only” of “Hardware”

parameter can’t be saved. All

red hints

appear

when

The item “blankVotage” of

Set the item “blankVotage” by the

“Hardware” hasn’t been set.

sequence below: Open hardware

54

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected]

testing

the

“blankODFilter=0.025

water blank of

blankVotageMax=62000

cuvette

blankVotageMin=30000” 1.

The red hints indicate its 1, If voltage is lower than 30000, it

The blue and voltage is out of range of means the reaction cuvettes should be the red hints 30000—62000. appear

changed. If voltage is higher than

when 2.The blue hints indicate OD 62000.it means the detecting voltage

testing

the value exceeds the range of needs adjusting.

water blank of the blank absorbency, the 2. clean or change cuvettes cuvette.

discrepancy among cuvettes is large

The

reagent 1.

blank

is

range wrong.

The

setting

of 1.

out wavelength is wrong. or 2.

There

are

When characters

again.

disorderly 2. Open the “cup blank” files and in

the delete the content. Repeat test the

result marked

“cupblank.ini”

file,

with ‘*’

affects

results

the

Choose the right wavelength over

which

55

of

“cuvette blank” and saved it. 3. check biochemical item parameter

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] wavelengths. 3. The reagent blank setting is wrong 1.

computer’s 1.

The

Replace

a

high-configuration

configuration is some poor computer. matching with the software condition or its running is 2. Reinstall. the system unstable.

3. Antivirus computer

The computer 2. Windows is unstable or 4. Don’t switch the interface too fast. or the software not reliable.

5. Copy or use the backup documents

has 3. The system has been

system

attacked by virus.

crashed.

4.

The

switch

about

interfaces is too fast. 5. Configuration files has been damaged or missed. 1, database have redundant Software

can

not run, error

characters 2, not to install southeast Asian language 1,, Hardware set wrong

Cannot

enter

Hardware parameter Settings, or not work normally

56

1, database compression, repair 2, installation southeast Asian languages 1, check Hardware Settings

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] Appendix 1: FACA-261-301-401 instrument wiring diagram 1. FACA-401 motor drive system wiring

57

2728 S. La Cienega Blvd. Los Angeles, CA 90034 Tel.: (310) 202-0811/14 Fax: (310) 202-7286 Website: www.labomed.com Email: [email protected] 2. FACA-301 motor drive system wiring diagram

3. FACA-261 motor drive system wiring diagram

jump

8421

L : 0-15 =16

H : 16-31=

32

4+16=20

58