GAS CHROMATOGRAPHY

CLARUS 690 GC

Hardware and Customer Service Guide

Release History

Part Number

Release

Publication Date

09931373

C

April 2020

Any comments about the documentation for this product should be addressed to: User Assistance PerkinElmer 710 Bridgeport Avenue Shelton, Connecticut 06484-4794 U.S.A. EU Importer: PerkinElmer Ltd Chalfont Road Seer Green Beaconsfield Bucks HP9 2FX United Kingdom or http://www.perkinelmer.com/contactus/ Notices The information contained in this document is subject to change without notice. Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no warranty of any kind with regard to this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material. Copyright Information This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc. Copyright © 2020 PerkinElmer, Inc. Trademarks Registered names, trademarks, etc., used in this document, even when not specifically marked as such, are protected by law. PerkinElmer is a registered trademark of PerkinElmer, Inc.

Table of Contents . 3

Table of Contents Table of Contents .............................................................................................................. 3 1 Introduction .................................................................................................................... 9 About This Manual ............................................................................................................ 11 Conventions Used in this Manual ........................................................................................ 12 Notes, Cautions, and Warnings ...................................................................................... 12 Customer Service ......................................................................................................... 14 Electromagnetic Compatibility (EMC) ................................................................................... 15 United States ............................................................................................................... 15 European Union............................................................................................................ 15 European Union Industrial Environment ..................................................................... 15 South Korea ................................................................................................................. 15 Susceptibility to RF Interference .................................................................................... 16 Warning Signs on the Instrument ....................................................................................... 17 Label locations on instrument ............................................................................................. 20 Front of Instrument ...................................................................................................... 20 2 Safety Practices .............................................................................................................. 23 Chapter Overview.............................................................................................................. 25 Generic Warnings ......................................................................................................... 25 Heated Zones ............................................................................................................... 25 Precautions ....................................................................................................................... 27 Environmental Conditions................................................................................................... 29 Operating Conditions .................................................................................................... 29 Pollution Degree ...................................................................................................... 29 Clarus 690 GC Touch Screen ..................................................................................... 29 Normal Operating Conditions .................................................................................... 29 Safe Operating Conditions ........................................................................................ 29 Storage Conditions ................................................................................................... 29 Decontamination ............................................................................................................... 31 Decontamination .......................................................................................................... 31 Cleaning the Instrument................................................................................................ 31 General Laboratory Safety ............................................................................................. 31 Electrical Safety ................................................................................................................ 32 Moving the Clarus 690 GC .................................................................................................. 34 ECD Radioactive Hazards ................................................................................................... 35 United States Government Regulations for ECDs ............................................................. 35 Labels ......................................................................................................................... 35 Leak Testing ................................................................................................................ 36 ECD Cell Failure or Damage ........................................................................................... 36 Reporting Radiation Incidents, Theft or Loss ................................................................... 36 Other ECD Requirements............................................................................................... 36 United Kingdom Regulations .......................................................................................... 36 Safe Handling of Gases ...................................................................................................... 38 Ventilation ................................................................................................................... 38 Using Hydrogen ............................................................................................................ 38 Identification of Gas Cylinders ....................................................................................... 40 Storing Gas Cylinders .................................................................................................... 40 Handling of Gas Cylinders.............................................................................................. 40 Hazardous Chemicals ......................................................................................................... 42 Definitions in Warning for Hazardous Chemicals ................................................................... 42

4 . Clarus 690 GC Hardware and Customer Service Guide WEEE Instructions for PerkinElmer Products ........................................................................ 43 3 System Description ......................................................................................................... 45 Overview of the Clarus 690 GC ........................................................................................... 47 About the Touch Screen..................................................................................................... 49 4 Before You Install a Column ........................................................................................... 51 Column Installation Information ......................................................................................... 53 Injector and Detector Fittings ........................................................................................ 53 Column Hangers ........................................................................................................... 53 Protecting Your Column ..................................................................................................... 54 5 Installing a Packed Column ............................................................................................ 57 About this Chapter ............................................................................................................ 59 Summary ..................................................................................................................... 59 Materials and Tools Required ......................................................................................... 59 Packed Column Injector Overview....................................................................................... 60 About the Wide-Bore Adapter ........................................................................................ 60 Step 1: Turn off the Heaters .............................................................................................. 62 Step 2: Set the Carrier Gas Flow ......................................................................................... 62 Setting the Carrier Gas Flow Using a Soap Bubble or Electronic Flowmeter.................... 63 Step 3: Connect One End of the Column to the Packed Injector ............................................ 66 Step 4: Leak Test .............................................................................................................. 66 Step 5: Condition the Column ............................................................................................. 67 Step 6: Attach the Other End of the Column to the Detector ................................................. 70 Step 7: Leak Test the Column/Detector Connection .............................................................. 71 6 Installing a Capillary Column .......................................................................................... 73 About this Chapter ............................................................................................................ 75 Summary ..................................................................................................................... 75 Step 1: Turn the Heaters Off .............................................................................................. 76 Step 2: Connect the Column to the Injector ......................................................................... 77 Step 2A: Connect the Column to the Split/Splitless (CAP) Injector ................................. 77 About the Injector Liners .......................................................................................... 78 Splitless Injections ................................................................................................... 78 Split Injections......................................................................................................... 79 About the Pneumatics Control ................................................................................... 79 Connecting a Column to the CAP Injector................................................................... 79 Step 2B Connect the Column to the Programmed Split/Splitless (PSS) Injector: ............. 89 About the Injector Liners .......................................................................................... 89 About the Pneumatics Control ................................................................................... 91 Connect the Column to the PSS Injector .................................................................... 91 Step 2C Connect the Column to the Programmed On-Column Injector (POC): ............... 106 About the Pneumatics Control ................................................................................... 107 Step 3: Set the Carrier Gas ............................................................................................. 114 Setting the Carrier Gas Pressure for the Split/Splitless Injector (CAP) and Programmed Split/Splitless Injector (PSS)...................................................................................... 114 Setting the Carrier Gas Flow for the Programmed On-Column Injector (POC) ................ 116 Suggested Capillary Column Pressures............................................................................ 118 Setting the Carrier Gas Flow for the Programmed On-Column Injector (POC) Using the Optional Flow Readout............................................................................... 118 Set the Carrier Gas Flow for the Programmed On-Column Injector (POC) Using the Soap Bubble Flowmeter ............................................................................. 120

Table of Contents . 5 Step 4 Leak Test All New Connections: ........................................................................... 122 PPC Pneumatics (POC Injector Only) .............................................................................. 122 Step 5: Condition the Column and the Mechanical Joint Between the Pre-column and Column: ..................................................................................................................... 124 Step 6: Connect the Column to the Detector .................................................................... 127 Step 7: Leak Test All New Connections ............................................................................ 131 Step 8: Set up the Split Mode for a CAP or PSS Injector .................................................... 131 PSS and POC Operating Hints: ............................................................................................ 132 Oven Programming Mode .............................................................................................. 132 Inlet Programming Mode ............................................................................................... 132 Solvent Purge Mode ...................................................................................................... 133 Calculating a Capillary Column Split Ratio ............................................................................ 135 7 Prevent ............................................................................................................................ 137 About this Chapter ............................................................................................................ 139 Installing PreVent on an Injector ........................................................................................ 140 Summary ..................................................................................................................... 140 Step 1. Install the Restrictor in the PreVent Adapter ....................................................... 140 Step 2. Connect the PreVent Adapter to the Injector....................................................... 141 Step 3. Connect the Column to the PreVent Adapter ....................................................... 142 Step 4. Leak-Check the System..................................................................................... 144 Step 5. Connect the Column to the Detector .................................................................. 146 Step 6. Set the Initial Pressures .................................................................................... 149 Step 7. Condition the System........................................................................................ 152 Installing PreVent on a Detector ......................................................................................... 153 Summary ..................................................................................................................... 153 Step 1. Install the Restrictor in the PreVent Adapter ....................................................... 153 Step 2. Connect the PreVent Adapter to the Detector ..................................................... 154 Step 3. Connect the Column to the PreVent Adapter ....................................................... 155 Step 4. Leak-Check the System..................................................................................... 156 Step 5. Connect the Column to the Injector ................................................................... 158 Step 6. Set the Initial Pressures .................................................................................... 160 Step 7. Condition the System........................................................................................ 164 Installing PreVent on a Clarus SQ 8 MS Detector .................................................................. 165 Summary ..................................................................................................................... 165 Step 1. Prepare the Clarus SQ 8 MS .............................................................................. 165 Step 2. Install the PreVent Injector Adapter in the Clarus SQ 8 MS Transfer Line............... 165 Step 3. Connect the Column to the PreVent Injector Adapter ........................................... 167 Step 4. Connect the Transfer Line to the Clarus SQ 8...................................................... 168 Step 5. Leak-Check the System..................................................................................... 171 Step 6. Connect the Column to the Injector ................................................................... 174 Step 7. Set the Initial Pressures .................................................................................... 175 Step 8. Condition the PSS System ................................................................................. 178 Step 9. Replacing the Restrictor .................................................................................... 178 PreVent Operating Techniques ........................................................................................... 180 Summary of the PreVent Techniques .............................................................................. 180 Column Isolation Technique .......................................................................................... 180 Solvent Purge Technique ............................................................................................... 182 Large Volume Injections (LVI) ....................................................................................... 184 Sample Residue Purge Technique .................................................................................. 184 Time Saver Technique .................................................................................................. 186 8 PPC Fundamentals ........................................................................................................... 189

6 . Clarus 690 GC Hardware and Customer Service Guide Introduction...................................................................................................................... 191 Carrier Gas Control ............................................................................................................ 192 Packed Column and Programmed-Temperature On-Column Inlets..................................... 192 Carrier Gas Mass-Flow Controller Operating Range ..................................................... 192 Mass-Flow Controller Setup....................................................................................... 192 Mass-Flow Controller Calibration................................................................................ 193 Carrier Gas Mass-Flow Controller Configuration .......................................................... 193 Headspace Analysis .................................................................................................. 196 Capillary and Programmed Split/Splitless Inlets ............................................................... 196 Split Pneumatic Setup .............................................................................................. 196 Split Pneumatic Control ............................................................................................ 197 Split Pneumatic Calibration ....................................................................................... 198 Split Pneumatic Carrier Gas Configuration .................................................................. 200 Column Dimensions ................................................................................................. 202 Vacuum Compensation ............................................................................................. 202 Split Control Mode Configuration ............................................................................... 202 Split Flow Offset ...................................................................................................... 203 Split Pneumatic Operation ........................................................................................ 204 Theory of Capillary Column Control ................................................................................ 204 Column Temperature Effects ..................................................................................... 205 Flow Programmed Operation..................................................................................... 206 Velocity Programmed Operation ................................................................................ 207 Pressure / Oventrack Mode ....................................................................................... 208 Vacuum Compensation ............................................................................................. 208 Detector Gas Flow Control.................................................................................................. 210 Detector Gas Flow Controller Operating Range ................................................................ 210 Detector Gas Flow Controller Setup ................................................................................ 210 Detector Gas Flow Controller Calibration ......................................................................... 211 Detector Gas Mass-Flow Controller Operation .................................................................. 211 Flame Gases ................................................................................................................ 211 Flame Ignition ......................................................................................................... 211 Makeup and Reference Gases ........................................................................................ 212 Auxiliary Pressure and Flow Control .................................................................................... 213 Auxiliary Carrier Gas Control .......................................................................................... 213 Auxiliary Carrier Gas Setup ....................................................................................... 213 Auxiliary Carrier Gas Mass-Flow Controller Calibration ................................................. 213 Auxiliary Carrier Gas Zone Operation ......................................................................... 213 Auxiliary Detector Gas Control ....................................................................................... 214 Auxiliary Detector Gas Setup ..................................................................................... 214 Auxiliary Detector Gas Flow Controller Calibration ....................................................... 214 Auxiliary Detector Gas Zone Operation ....................................................................... 214 PPC Tips and Techniques ................................................................................................... 215 Effect of Flow on Pressurization Rate .............................................................................. 215 Correcting the Column Dimensions ................................................................................. 215 Correcting the Column Length................................................................................... 215 Splitless Injection ......................................................................................................... 216 Zero Initial Split Flow ............................................................................................... 216 Zero Split Flow by Pre-Run Event .............................................................................. 216 Pressure-Pulse Injection ................................................................................................ 217 Reducing Carrier-Gas Consumption ................................................................................ 217 9 Maintenance..................................................................................................................... 219 Overview .......................................................................................................................... 221

Table of Contents . 7 Autosampler Maintenance .................................................................................................. 222 Changing a Syringe....................................................................................................... 222 Removing a Syringe ................................................................................................. 223 Installing a Syringe .................................................................................................. 224 Replacing the Vial-Locator Mechanism ............................................................................ 225 Cleaning the Autosampler Tray ...................................................................................... 226 Syringe Maintenance ......................................................................................................... 227 Cleaning the 5-µL and 50-µL Syringe Plungers ................................................................ 227 Servicing Idle Syringes .................................................................................................. 227 Injector Maintenance ......................................................................................................... 228 Changing Septa ............................................................................................................ 228 Replacing a CAP Injector Gold Seal ................................................................................ 229 Changing and Repacking Packed Column Injector Liners .................................................. 232 Changing the Hourglass Needle Guide on the Programmed On-Column (POC) Injector ....... 233 Changing and Repacking Capillary Split/Splitless (CAP) and Programmed Split/Splitless (PSS) Injector Liners .............................................................................................................. 234 Selecting an Appropriate PSS Injector Liner .................................................................... 234 Packing the PSS Injector Liner with Quartz Wool ............................................................. 235 Packing a CAP Injector Liner for Split Operation.......................................................... 235 Packing a CAP Injector Liner for Splitless Operation .................................................... 235 Reinstalling the Liner in the PSS Injector ........................................................................ 236 Removing a Broken Liner from the PSS Injector Body ...................................................... 236 Changing the Charcoal Trap on the Split/Splitless (CAP) Injectors ..................................... 238 Removing a Charcoal Trap ........................................................................................ 238 Changing the Charcoal Trap or Replacing Charcoal on PSS Injectors ................................. 240 Removing a Charcoal Trap ........................................................................................ 240 Installing a New Charcoal Trap.................................................................................. 241 FID Maintenance ............................................................................................................... 243 Replacing a FID Jet....................................................................................................... 243 Cleaning a FID Jet ........................................................................................................ 245 Replacing the O-Ring in the FID Collector ....................................................................... 245 Cleaning the FID Collector and Cap ................................................................................ 246 ECD Maintenance .............................................................................................................. 247 Baking the ECD ............................................................................................................ 247 Changing the Charcoal Traps ......................................................................................... 247 Cleaning the ECD Anode ............................................................................................... 248 Wipe Testing an ECD Cell .............................................................................................. 249 Disposal and Refurbish/Refoil of an ECD Cell ................................................................... 251 PID Maintenance ............................................................................................................... 252 Changing a PID Lamp ................................................................................................... 252 Cleaning PID Lamp Windows ......................................................................................... 253 Changing PID Lamp Window Seals and Positioning Disks ................................................. 255 Leak-Test ................................................................................................................ 255 NPD Maintenance .............................................................................................................. 256 Changing the NPD Bead ................................................................................................ 256 NPD Bead Activation Procedure...................................................................................... 257 Packed Column Setup: ............................................................................................. 258 Capillary Column Setup: ........................................................................................... 258 Reactivating a Used NPD Bead .................................................................................. 258 Reactivate a Used NPD Bead..................................................................................... 261 Operating Notes ...................................................................................................... 261 Replacing an NPD Jet .................................................................................................... 262 FPD Maintenance .............................................................................................................. 265

8 . Clarus 690 GC Hardware and Customer Service Guide Cleaning/Replacing an Optical Filter Assembly ................................................................. 265 Cleaning/Replacing the Detector Liner ............................................................................ 266 Cleaning/Replacing the Detector Window........................................................................ 267 Removing the Window ............................................................................................. 268 Cleaning and Replacing the Window .......................................................................... 269 Replacing the Photomultiplier Tube ................................................................................ 269 Replacing the FPD Jet ................................................................................................... 272 PPC Maintenance .............................................................................................................. 275 Replacing a Restrictor ................................................................................................... 275 Zeroing the Pressure ..................................................................................................... 275 PPC Restrictor Information ............................................................................................ 275 Available Restrictors ................................................................................................. 275 Restrictors for Carrier Gas Control ............................................................................. 276 Restrictors for Detector Gas Control........................................................................... 276 Practical Hints ................................................................................................................... 277 Reversing TCD Polarity .................................................................................................. 277 Optimizing FID Performance .......................................................................................... 277 Filtering Detector Output ............................................................................................... 277 Attenuation vs. Detector Output..................................................................................... 278 Prolonging the Life of a PID UV Lamp............................................................................. 279 Optimizing ECD Performance ......................................................................................... 279 Optimizing FPD Performance ......................................................................................... 279 About Ionization Gas Flows ....................................................................................... 279 Optimizing FPD Gas Flows ........................................................................................ 279 Quenching .............................................................................................................. 280 PSS and POC Operating Hints ........................................................................................ 280 Oven Programming Mode ......................................................................................... 280 Inlet Programming Mode .......................................................................................... 280 PPC Restrictor Information ............................................................................................ 281 Available Restrictors ................................................................................................. 281 Restrictors for Carrier Gas Control ............................................................................. 282 Restrictors for Detector Gas Control........................................................................... 282 10 Troubleshooting ............................................................................................................. 283 About this Chapter ............................................................................................................ 285 Error Messages ................................................................................................................. 285 GC Troubleshooting ........................................................................................................... 289 Spare Components. ...................................................................................................... 289 Logical Troubleshooting Steps. ...................................................................................... 289 Dual Identical Channels Only ......................................................................................... 290 Appendix 1: U.S. Nuclear Regulations ................................................................................ 291 Appendix I - SUPPLEMENT 2 Agreement States.................................................................... 292 U.S. Nuclear Regulatory Commission Regional Offices .......................................................... 296 Nuclear Regulatory Commission Regulations ........................................................................ 297 Appendix II: Ionization Potentials ..................................................................................... 318 Index ................................................................................................................................... 323

1 I ntroduction

10 . Clarus 690 GC Customer Hardware and Service Guide

About This Manual . 11

About This M anual This hardware guide is divided into following chapters:

Chapter 1 I ntroduction This chapter contains a brief introduction on the instrument, the conventions and warnings used in the manual.

Chapter 2 Safety Practices Important safety information for the Clarus 690 GC is provided in this chapter.

Chapter 3 System Description This chapter contains information on the components of the instrument, how it works and instrument specifications.

Chapter 4 Before You I nstall a Colum n This chapter contains general information regarding column installation and the basic procedures you should know in order to install a column.

Chapter 5 I nstalling a Packed Colum n This chapter contains procedures for connecting a packed column to the packed column injector and setting the carrier gas flow using PPC.

Chapter 6 I nstalling a Capillary Colum n This chapter contains procedures for connecting a capillary column to a Capillary Injector (CAP); a Programmed Split/Splitless Capillary Injector (PSS); and a Programmed On-Column Capillary Injector (POC). It also describes how to set the gas flows using PPC.

Chapter 7 PreVent PreVent is an enhanced capillary inlet system for the Clarus 690 GC that uses columns with an inside diameter (i.d.) between 0.25 mm and 0.53 mm.

Chapter 8 PPC Fundam entals This chapter describes the fundamentals of Programmable Pneumatic Control (PPC) and how to use PPC to control your injectors.

Chapter 9 M aintenance Maintenance and cleaning procedures for the various components of your instrument are provided.

Chapter 10 Troubleshooting Performance checks, troubleshooting information, software and system error messages along with practical hints for running for the instrument are provided.

Appendix I : U.S. Nuclear R egulations Regulations from the Nuclear Regulatory Commission.

Appendix I I : I onization Potential Ionization potentials for representative compounds.

12 . Clarus 690 GC Customer Hardware and Service Guide

Conventions Used in this M anual Normal text is used to provide information and instructions. Bold text refers to button or tab page that is displayed on the touch screen. All eight digit numbers are PerkinElmer part numbers unless stated otherwise.

N otes, Cautions, and W arnings Three terms, in the following standard formats, are also used to highlight special circumstances and warnings.

N OTE: A note indicates additional, significant information that is provided with some procedures.

Conventions Used in this Manual . 13

CAUTION D

We use the term CAUTI ON to inform you about situations that could result in serious damage to the instrument or other equipment. Details about these circumstances are in a box like this one. Caution (Achtung) Bedeutet, daß die genannte Anleitung genau befolgt w erden m uß, um einen Geräteschaden zu verm eiden.

DK

E

F

I

NL

P

Caution (Bemærk)

Dette betyder, at den nævnte vejledning sk al overholdes nøje for at undgå en beskadigelse af apparatet. Caution (Advertencia)

Utilizam os el térm ino CAUTION (ADVERTENCI A) para advertir sobre situaciones que pueden provocar averías graves en este equipo o en otros. En recuadros éste se proporciona inform ación sobre este tipo de circunstancias. Caution (Attention)

N ous utilisons le term e CAUTION (ATTENTI ON) pour signaler les situations susceptibles de provoquer de graves détériorations de l'instrument ou d'autre m atériel. Les détails sur ces circonstances figurent dans un encadré sem blable à celui-ci. Caution (Attenzione)

Con il term ine CAUTION (ATTENZI ON E) vengono segnalate situazioni che potrebbero arrecare gravi danni allo strumento o ad altra apparecchiatura. Troverete inform azioni su tali circostanze in un riquadro com e questo. Caution (Opgelet)

Betek ent dat de genoem de handleiding nauw k eurig m oet w orden opgevolgd, om beschadiging van het instrument te voork om en. Caution (Atenção)

Significa que a instrução referida tem de ser respeitada para evitar a danificação do aparelho.

14 . Clarus 690 GC Customer Hardware and Service Guide

WARNING D

DK

We use the term W ARNI NG to inform you about situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one. Warning (Warnung)

Bedeutet, daß es bei Nichtbeachten der genannten Anw eisung zu einer Verletzung des Benutzers k om m en k ann. Warning (Advarsel)

Betyder, at brugeren k an blive kvæstet, hvis anvisningen ikke overholdes. Warning (Peligro)

E

F

I

NL

P

Utilizam os el térm ino WARNING (PELI GRO) para inform arle sobre situaciones que pueden provocar daños personales a usted o a otras personas. En los recuadros com o éste se proporciona inform ación sobre este tipo de circunstancias. Warning (Danger)

N ous utilisons la form ule WARNING (DAN GER) pour avertir des situations pouvant occasionner des dommages corporels à l'utilisateur ou à d'autres personnes. Les détails sur ces circonstances sont données dans un encadré sem blable à celui-ci. Warning (Pericolo)

Con il term ine WARNING (P ERI COLO) vengono segnalate situazioni che potrebbero provocare incidenti alle persone. Troverete inform azioni su tali circostanze in un riquadro com e questo. Warning (Waarschuwing)

Betek ent dat, w anneer de genoem de aanw ijzing niet in acht w ordt genom en, dit k an leiden tot verwondingen van de gebruik er. Warning (Aviso)

Significa que a não observância da instrução referida poderá causar um ferimento ao usuário.

Custom er Service PerkinElmer Instruments 710 Bridgeport Avenue Shelton, CT 06484-4794 U.S.A Tel: 1 (800) 762-4000 or (203) 762-4000 Internet: http://www.perkinelmer.com

Electromagnetic Compatibility (EMC) . 15

Electrom agnetic Com patibility (EM C) This product has been designed and manufactured, having regard to the state of the art, to ensure that: (a) the electromagnetic disturbance generated does not exceed the level above which radio and telecommunications equipment or other equipment cannot operate as intended; (b) it has a level of immunity to the electromagnetic disturbance to be expected in its intended use which allows it to operate without unacceptable degradation of its intended use.

United States This product is classified as a digital device used exclusively as industrial, commercial, or medical test equipment. It is exempt from the technical standards specified in Part 15 of the FCC rules and regulations, based on Section15.103(c). Use of this product in residential areas may cause interference and such use should be avoided unless special measures are taken by the user to restrict emissions to a level that allows the reception of broadcast emissions.

European Union All information concerning EMC standards will be in the Declaration of Conformity and these standards will change as the European Union adds new requirements.

European Union I ndustrial Environm ent The 230 V/50 Hz. Clarus GC manufactured for use in the European Union is intended for the industrial environment. The instrument is to be connected to a mains power network supplied from a high or medium-voltage transformer dedicated for the supply of an installation feeding a manufacturing or similar plant. Industrial environments are characterized by the existence of one or more of the following conditions: •

industrial, scientific and medical (ISM) apparatus are present



heavy inductive or capacitive loads are frequently switched



currents and associated magnetic fields are high

These are the major contributors to the industrial electromagnetic environment and as such distinguish the industrial from other environments. The instrument is not intended for connection to a public mains network supplying residential, commercial and light-industrial locations.

South K orea This device complies with MSIP (Ministry of Science, ICT, and Future Planning) EMC Registration requirements. This instrument is registered as a Class A instrument for business use only. Product seller and user should notice that this equipment is not for household use. Class A

A급 기기 (업무용 방송통신기자재) 이 기기는 업무용(A급) 전자파적합기기로서 판 매자 또는 사용자는 이 점을 주의하시기 바라 며, 가정외의 지역에서 사용하는 것을 목적으 로 합니다.

16 . Clarus 690 GC Customer Hardware and Service Guide

Susceptibility to RF I nterference With the exception of the Flame Ionization Detector (FID), a RF field strength of 10 V/m between 80 MHz. and 1000 MHz. with 80% modulation at 1 kHz. may cause a deflection on the chromatographic detector baseline that exceeds its normal pattern. This implies that if a transmitting device, such as a walkie-talkie carried by a security guard, is used near the detector, a spike or peak on the chromatographic baseline may occur. If you are concerned that such an event may occur, PerkinElmer recommends that walkie-talkie restriction notices be posted in the vicinity. Cell phones, beepers, and other similar devices operate in a much higher frequency range and do not cause interference.

Warning Signs on the Instrument . 17

W arning Signs on the I nstrum ent Alternating current. Courant alternative.

Protective conductor terminal. Terminal de conducteur de protection.

Off position of the main power source. Position hors tension de la source d'alimentation principale.

On position of the main power source. Sur la position de la source d'alimentation principale.

W arning, Hot surface. Avertissem ent, Surface chaude.

W arning, Risk of electric shock. Avertissem ent, Risque d'électrocution.

Caution, Risk of Danger. Documentation must be consulted to determine the nature of the potential hazard and any actions which have to be taken. Attention, Risque de danger. La documentation doit être consultée afin de déterminer la nature du danger potentiel et les mesures à prendre.

18 . Clarus 690 GC Customer Hardware and Service Guide

This unit contains protective circuitry. Contact PerkinElmer qualified service personnel before performing any AC line tests. Cet appareil contient des circuits de protection. Contactez le personnel de service qualifié PerkinElmer avant d'effectuer des essais en ligne AC. Caution If the instrument top is opened, the alignment of the Autosampler tower must be verified. Refer to your Clarus Manual. Attention Si le haut de l'instrument est ouvert, l'alignement de la tour du collecteur d'échantillons automatiques doit être vérifié. Consultez votre manuel Clarus. (The CAUTION above is only visible if detector compartment door is open.) (L'ATTENTION ci-dessus n'est visible que si la porte du compartiment détecteur est ouverte.)

W arning Disconnect AC power cord from the outlet before removing any covers or parts. Do not operate the instrument with any cover or parts removed. Avertissem ent Débranchez le cordon d'alimentation secteur de la prise avant de retirer les couvercles ou les pièces. Ne pas utiliser l'instrument avec un couvercle ou des pièces retirées. (The WARNING above is only visible if detector compartment door is open.) (L'AVERTISSEMENT ci-dessus n'est visible que si la porte du compartiment détecteur est ouverte).

W arning Grounding circuit continuity is vital for safe operation of equipment. Never operate equipment with grounding conductor disconnected. Avertissem ent La continuité du circuit de mise à la terre est essentielle pour un fonctionnement sûr de l'équipement. N'utilisez jamais d'équipement avec le conducteur de mise à la terre déconnecté. W arning Shut off hydrogen and cap unused flame detector fittings to prevent accumulation of hydrogen in oven and possible explosion. Avertissem ent Couper l'hydrogène et fermer les raccords détecteurs de flamme inutilisés pour éviter l'accumulation d'hydrogène dans le four et une éventuelle explosion. (The WARNING above is only visible if detector compartment door is open.) (L'AVERTISSEMENT ci-dessus n'est visible que si la porte du compartiment détecteur est ouverte).

Warning Signs on the Instrument . 19

For ECD Instruments only the following certificate appears:

CERTIFICATE 63 Ni Electron Capture Detector

_____ N6100134 ____N61000063 Save for proof of tests. The attached source complies with the requirements of ANSI N542 (ISO 2919) Classification C42211. This source has been leak tested and is certified to have less than 0.005 μCi, 180 Bq of removable contamination. Serial Number Assembly Number Isotope 63 Ni Activity: 15 mCi Pressurization test at 30 psi N2: Passed: Initial Test Performed by: Wipe Test Date

20 . Clarus 690 GC Customer Hardware and Service Guide

Label locations on instrum ent Front of I nstrum ent CAUTION DETECTOR CONTAINS RADIOACTIVE MATERIAL (SEE TAG ON DETECTOR)

Warning

Caution

Disc onne ct supply cord and turn off inlet gases before opening top. Recheck integrity of rea r gas connections aft er clos ing.

If th e instr um ent top is o pened , the align men t o f A utosam pler tower m ust be verified . Refer to you r Claru s 600 Manual.

Warning Hot Surfac e

Warning Shut off hy drogen and cap unused flame detector fittings to prevent accumula tion of hydrogen in oven and possible explosion.

ECD Certificate Label

Wa rning Do not depre ss the ove n safe ty switches.

Wa rning Do not touch the hot me ta l surfa ce of the oven.

Warning Do not touch the oven fan while it is moving.

Label locations on instrument . 21

Back of I nstrum ent PerkinElmer

precisely. Shelton, CT 06484 USA

Warning Hot Surfac e

Clarus 600 Gas Chromatograph Part No. Nx xx xxx x

Serial No. Xxx xx xx

230 VAC ~ 50/60Hz MAX POWER 2400 VA EN 55011-Class A, Group 1

Caution

Caution

Consult the proper section in the manual for the allowable gases and maximum permitted pressures

Caution

Chimney Vent

Power Cord

Do n ot in stall closer th an 6mm to a wall o f co mb ustible m aterial.

Caution D o not r estrict air in take o r exhaust.

Warning Disc onnect AC power cord from outlet before removing any c over or parts . Do not opera te the instrument with any cover or parts removed.

Caution This unit contains protective circuitry. Contact PerkinElmer qualified service personnel before performing any AC line tests.

Warning Grounding circuit continuity is vital for safe operation of equipment. Never operate equipment with grounding conductor disconnected

22 . Clarus 690 GC Customer Hardware and Service Guide

Warning Hot Surfac e

Label locations on instrument . 23

2 Safety Practices

24 . Clarus 690 GC Customer Hardware and Service Guide

Chapter Overview . 25

Chapter Overview This chapter describes the general safety practices and precautions that must be observed when operating the Clarus 690 GC. This advice is intended to supplement, not supersede, the normal safety codes in the user's country. It is also a supplement to the PerkinElmer standard Safety and Health Policy. The information provided does not cover every safety procedure that should be practiced. Ultimately, maintenance of a safe laboratory environment is the responsibility of the analyst and the analyst's organization. Please consult all manuals supplied with the Clarus 690 GC and accessories before you start working with the instrument. Carefully read the safety information in this chapter and in the other manuals supplied. When setting up the instrument or performing analyses or maintenance procedures, strictly follow the instructions provided. The Clarus 690 GC should be used in accordance with the instructions provided in this manual. If used otherwise, the protection provided by the instrument may be impaired.

Generic W arnings Before installing or operating the Clarus 690 GC, read the following information concerning hazards and potential hazards. You should ensure that anyone involved with installation and/or operation of the Clarus 690 GC is knowledgeable in both general safety practices for the laboratory and safety practices for the Clarus 690 GC. Get advice from your safety engineer, industrial hygienist, environmental engineer, or safety manager before you install or use this instrument.

Heated Zones Heated zones should be treated with caution, for example, injector caps and detectors. Avoid physical contact with the injector caps. The detector cover may get hot, especially if flame ionization detectors are operated at high temperatures. As a general rule, allow heated zones to cool before attempting to work in the oven, injector, or detector areas.

THERM AL RUNAW AY PROTECTI ON: The Clarus 690 GC software shuts down the instrument if any heated zone exceeds 470 °C. Should this occur, the following error message is displayed: INSTRUMENT SHUTDOWN

xxx THERM RUNAWAY where xxx is the heated zone

CAUTION

Call your PerkinElmer Representative. Instrument shutdown also occurs if there is a PRT (Platinum Resistance Thermometer) or MPU (Micro Processor Unit) failure. In these cases the following error message is displayed: INSTRUMENT SHUTDOWN xxx PRT ERROR where xxx is the failed zone.

Call your PerkinElmer Representative.

26 . Clarus 690 GC Customer Hardware and Service Guide

P ROTECTI ON THERM I QUE RUN AW AY: Le logiciel Clarus 690 GC ferme l'instrument si une zone chauffée dépasse 470 ° C. Dans ce cas, le message d'erreur suivant s'affiche: ARRÊT DE L'I NSTRUM ENT Xxx THERM RUNAWAY où xxx est la zone chauffée ATTENTION

Appelez votre représentant PerkinElmer. L'arrêt de l'instrument se produit également s'il existe une défaillance PRT (Platinum Resistance Thermometer) ou MPU (Micro Processor Unit). Dans ces cas, le message d'erreur suivant s'affiche: ARRÊT DE L'INSTRUMENT Xxx PRT ERROR où xxx est la zone défectueuse. Appelez votre représentant PerkinElmer.

Precautions . 27

P recautions

WARNING AVERTISSEMENT

CAUTION ATTENTION

WARNING AVERTISSEMENT

WARNING AVERTISSEMENT CAUTION ATTENTION

Be sure that all instrument operators read and understand the precautions listed below. It is advisable to post a copy of the precautions near or on the instrument shelf. Assurez-vous que tous les opérateurs d'instruments lisent et comprennent les précautions indiquées ci-dessous. Il est conseillé de poster une copie des précautions à proximité ou à l'étagère de l'instrument. The protection provided by this equipment may be impaired if the equipment is used in a manner not specified by PerkinElmer. La protection fournie par cet équipement peut être altérée si l'équipement est utilisé de manière non spécifiée par PerkinElmer. To reduce the chance of electrical shock, do not remove covers that require tool access. No user serviceable parts are inside. Refer servicing to PerkinElmer or other qualified service personnel. Pour réduire les risques de choc électrique, ne retirez pas les couvercles qui nécessitent un accès à l'outil. Aucune pièce pouvant être réparée par l'utilisateur n'est à l'intérieur. Reportez-vous à l'entretien de PerkinElmer ou d'un autre personnel qualifié.

To avoid electrical shock, disconnect power cord before servicing. Pour éviter les chocs électriques, débranchez le cordon d'alimentation avant de procéder à l'entretien.

Do not restrict air intake or exhaust. Ne pas limiter l'admission d'air ou les gaz d'échappement.

The following precautions must be observed when using the Clarus 690 GC: •

PerkinElmer provides a number of safety interlocks on its instruments that monitor such things as door locks, gas pressures, and the temperatures of various components. Do not attempt to defeat these interlocks; you may compromise your own or someone else’s safety.



Be sure that the power line voltage of the Clarus 690 GC corresponds to the voltage used in your laboratory.



Never remove the side panels of the Clarus 690 GC without shutting down the instrument and disconnecting the instrument power cord from line power.



Do not immerse the purge gas exit line in a liquid as the liquid may be drawn back into the sample holder.

28 . Clarus 690 GC Customer Hardware and Service Guide •

Only high-quality purge gases should be used with the Clarus 690 GC. Minimum purity of 99.995% is recommended. A high-quality filter-dryer accessory is recommended for the removal of any moisture from the purge gases.

Environmental Conditions . 29

Environm ental Conditions Operating Conditions CAUTION ATTENTION

WARNING AVERTISSEMENT

The Clarus 690 GC is designed for indoor use only. Le Clarus 690 GC est conçu pour une utilisation à l'intérieur seulement.

The Clarus 690 GC is not designed for operation in an explosive environment. Le Clarus 690 GC n'est pas conçu pour fonctionner dans un environnement explosif.

Pollution Degree The Clarus 690 GC will operate safely in environments that contain nonconductive foreign matter up to Pollution Degree 2 in EN/IEC 61010-1.

WARNING AVERTISSEMENT

Dispose of waste in accordance with the regulations applicable to your locality, state and/or country. Éliminer les déchets conformément à la réglementation applicable à votre localité, votre état et / ou votre pays.

Clarus 690 GC Touch Screen For optimum performance, the Clarus 690 GC’s touch screen may require periodic re-calibration. The interval between re-calibration may be affected by exposure to combined heat and humidity conditions (ambient conditions between 30 °C / 50% RH and 35 °C / 80% RH).

Norm al Operating Conditions •

Indoors



Temperature: +10 °C to +35 °C (+50 °F to +95 °F)



Relative humidity: 20% to 80%, without condensation



Altitude: in the range -400 m to 2,000 m (sea level to 6,562 feet)

Safe Operating Conditions Do not operate in a Cold Room or a refrigerated area. •

Indoors



Temperature: +5°C to +40°C (+41°F to +104°F)



Relative humidity: 20% to 80%, without condensation



Altitude: in the range -400 m to 2,000 m (sea level to 6,562 feet)

Storage Conditions The Clarus 690 GC may be stored under the following conditions: •

Ambient temperature is -20 °C to +60 °C (-4 to 140 °F)



Ambient relative humidity is 20 to 80%, non-condensing

30 . Clarus 690 GC Customer Hardware and Service Guide •

Altitude is in the range -400 m to 12 000 m (sea level to 39 370 feet).

Decontamination . 31

Decontam ination Decontam ination Before using any cleaning or decontamination methods except those specified by PerkinElmer, users should check with PerkinElmer that the proposed method will not damage the equipment. Customers wishing to return instrumentation and/or associated materials to PerkinElmer for repair, maintenance, warranty or trade-in purposes are advised that all returned goods must be certified as clean and free from contamination. The customer's responsible body is required to follow the “Equipment Decontamination Procedure” and complete the “Certificate of Decontamination”. These documents are available on the PerkinElmer public website: http://www.perkinelmer.com/Content/technicalinfo/dts_instrumentdeconprocedure.pdf If you do not have access to the internet contact Customer Care:

USA (8:30 a.m. – 7 p.m. EST) Canada Mexico EU (Brussels) EU (Monza) Brazil India China Hong Kong Japan Macau Philippines South Korea Taiwan Singapore Malaysia Thailand Australia

1-800-762-4000 (inside the USA)/(+1) 203-925-4602 (outside the USA) 800-561-4646 52-55-5337-0700 0800-40-858 0800-90-66-42 55-11-3868-6200 1800-22-1660/1800-266-1660 86-800-820-5046 /+86-400-820-5046 +852-26201881 +81-45-339-5866 +852-26201881 + 63-8220511 +82-2-868-5350 886-0800-211134/886-2-87912589 ext 620 +65-6868-1662 (Sales)/+ 65 68681668 (Services) +60-3-79491118 1-800-898-282 1-800-066-776 (Service)/1 800-033-391 (Sales)

If you are located outside of these regions, please call your local PerkinElmer sales office for more information.

Cleaning the I nstrum ent Exterior surfaces may be cleaned with a soft cloth, dampened with a mild detergent and water solution. Do not use abrasive cleaners or solvents.

General Laboratory Safety Your laboratory should have all equipment ordinarily required for the safety of individuals working with chemicals (fire extinguishers, first-aid equipment, safety shower and eye-wash fountain, spill cleanup equipment, etc.).

32 . Clarus 690 GC Customer Hardware and Service Guide

Electrical Safety The Clarus 690 GC contains high voltage. To prevent the risk of shock, unplug the line cord from the AC outlet and wait at least one minute before opening or removing any instrument panels. The instrument has been designed to protect the operator from potential electrical hazards. This section describes some recommended electrical safety practices.

CAUTION ATTENTION

WARNING AVERTISSEMENT

WARNING AVERTISSEMENT

WARNING AVERTISSEMENT

This unit contains protective circuitry. Contact PerkinElmer Service before performing any AC line tests. Cet appareil contient des circuits de protection. Contactez PerkinElmer Service avant d'effectuer des tests de ligne AC. Connect the GC to an AC line power outlet that has a protective ground connection. To ensure satisfactory and safe operation of the GC, it is essential that the protective ground conductor (the green/yellow lead) of the line power cord is connected to a true electrical ground. Any interruption of the protective ground conductor, inside or outside the GC, or disconnection of the protective ground terminal may impair the protection provided by the GC. Connectez le GC à une prise de courant AC qui a une connexion de terre de protection. Pour assurer un fonctionnement satisfaisant et sûr du GC, il est essentiel que le conducteur de terre de protection (le fil vert / jaune) du cordon d'alimentation de la ligne soit connecté à un sol électrique réel. Toute interruption du conducteur de terre de protection, à l'intérieur ou à l'extérieur du GC, ou la déconnexion de la borne de terre de protection peut nuire à la protection fournie par le GC.

Do not operate the GC with any covers or parts removed. N'utilisez pas le GC avec des couvertures ou des pièces retirées.

To avoid electrical shock, disconnect the power cord from the AC outlet before servicing. Servicing on the GC is to be performed only by a PerkinElmer service representative or similarly trained and authorized person. Pour éviter les chocs électriques, débranchez le cordon d'alimentation de la prise secteur avant de procéder à l'entretien. L'entretien du GC doit être effectué uniquement par un représentant du service PerkinElmer ou une personne formée et autorisée de manière similaire.

Electrical Safety . 33

WARNING AVERTISSEMENT

CAUTION ATTENTION

Do not attempt to make adjustments, replacements or repairs to this GC except as described in the user documentation. N'essayez pas de faire des ajustements, des remplacements ou des réparations à ce CG, sauf comme décrit dans la documentation de l'utilisateur.

To ensure adequate cooling of the instrument electronics, do not obstruct the gap at the base of the GC, and leave at least a 6-inch clearance between instruments. Pour assurer un refroidissement adéquat de l'électronique de l'instrument, ne pas obstruer l'espace à la base du GC et laisser au moins un dégagement de 6 pouces entre les instruments.

Ensure that the power cord is correctly wired and that the ground leads of all electrical units (for example, recorders, integrators) are connected together via the circuit ground to earth. Use only three-prong outlets with common earth ground connections. For customers in the United Kingdom, the power plug provided with the Clarus 400/480 GC contains a 13 Amp fuse mounted within the plug body. This fuse complies with standard BS1362 and has a breaking capacity of 6000 Amps at 250 VAC. Any replacement fuse must be of the same type and amperage rating. Servicing of incoming AC line components in your laboratory should be performed only by a licensed electrician.

Lethal voltages are present at certain areas w ithin the instrum ent. I nstallation and internal m aintenance of the instrum ent should only be perform ed by a P erk inElm er service engineer or sim ilarly authorized and trained person. W hen the instrum ent is connected to line pow er, opening the instrum ent covers is likely to ex pose live parts. Even w hen the pow er sw itch is off, high voltages can still be present. Capacitors inside the instrum ent m ay still be charged even if the instrum ent has been disconnected from all voltage sources.

WARNING AVERTISSEMENT

Des tensions létales sont présentes dans certaines zones de l'instrum ent. L'installation et la m aintenance interne de l'instrum ent ne doivent être effectuées que par un ingénieur de service P erkinElm er ou une personne autorisée et form ée de m anière sim ilaire. Lorsque l'instrum ent est connecté à la ligne, l'ouverture des capots de l'instrum ent risque d'ex poser les pièces en direct. M êm e lorsque l'interrupteur d'alim entation est éteint, des tensions élevées peuvent encore être présentes. Les condensateurs à l'intérieur de l'instrum ent peuvent toujours être chargés m êm e si l'instrum ent a été déconnecté de toutes les sources de tension.

The instrument must be correctly connected to a suitable electrical supply. The supply must have a correctly installed protective conductor (earth ground) and must be installed or checked by a qualified electrician before connecting the instrument.

34 . Clarus 690 GC Customer Hardware and Service Guide Any interruption of the protective conductor (earth ground) inside or outside the instrum ent or disconnection of the protective conductor term inal is lik ely to m ak e the instrum ent dangerous. I ntentional interruption is prohibited. WARNING AVERTISSEMENT

Toute interruption du conducteur de protection (terreuse) à l'intérieur ou à l'ex térieur de l'instrum ent ou la déconnex ion de la borne du conducteur de protection risque de rendre l'instrum ent dangereux . L'interruption intentionnelle est interdite.

When working with the instrument: •

Do not position the equipment so that it is difficult to operate the AC power switch.



Disconnect the instrument from all voltage sources before opening it for any adjustment, replacement, maintenance, or repair. If afterwards, the opened instrument must be operated for further adjustment, maintenance, or repair, this must only be done by a PerkinElmer Service engineer.



Whenever it is possible that the instrument is no longer electrically safe for use, make the instrument inoperative and secure it against any unauthorized or unintentional operation. The electrical safety of the instrument is likely to be impaired if, for example, the instrument shows visible damage, has been subjected to prolonged storage under unfavorable conditions, or has been subjected to severe stress during transportation.

M oving the Clarus 690 GC Risk of injury through lifting. The instrum ent w eighs 49 k g (108 lb). I m proper lifting can cause painful and som etim es perm anent back injury.

WARNING AVERTISSEMENT

Use proper lifting procedures and/or mechanical lifting aids to lift or move the instrument. If the instrument must be moved, we recommend that at least two people carefully lift the instrument in order to move it. Risque de blessures par levage. L'instrum ent pèse 49 k g (108 lb). Un levage inadéquat peut causer des blessures douloureuses et parfois perm anentes au dos. Utilisez des procédures de levage appropriées et / ou des dispositifs de levage mécanique pour soulever ou déplacer l'instrument. Si l'instrument doit être déplacé, nous recommandons qu'au moins deux personnes soulevent soigneusement l'instrument pour le déplacer.

ECD Radioactive Hazards . 35

ECD R adioactive Hazards To assure that rem ovable radioactive contam ination on the ex ternal parts of the ECD rem ains at a safe level, the United States Nuclear Regulatory Com m ission requires that: The ECD be w ipe tested at least once every six m onths. A record of the results m ust be m aintained for NRC inspection. WARNING AVERTISSEMENT

P our s'assurer que la contam ination radioactive am ovible sur les parties ex ternes du DPE dem eure à un niveau sûr, la Com m ission de réglem entation nucléaire des États-Unis ex ige que: Le DPE doit être testé au m oins une fois tous les six m ois. Un registre des résultats doit être conservé pour l'inspection du CN RC.

United States Governm ent R egulations for ECDs N OTE: To repair an Electron Capture Detector cell requires a specific license issued by the U.S. Nuclear Regulatory Commission (NRC) and/or in some states by the equivalent state agency. For further information on obtaining a license, contact the Customer Service Department at PerkinElmer, Shelton, Connecticut, or the NRC Material Branch, Office of Nuclear Materials, Safety and Safeguards, Washington, DC 20555. All USNRC regulations can be obtained through the internet at www.nrc.gov/reading-rm/

N OTE: These instructions are for ECD cell purchasers who are not specifically licensed to handle radioactive materials. The Clarus 690 GC Electron Capture Detector model P/N N6100063) contains a maximum of 15 mCi of Nickel 63 (Ni 63), a radioactive material. Your possession and use of this detector is governed by 10 C.F.R. Section 31.5 which is reproduced in Appendix I. Under the provisions of that regulation you are deemed a "General Licensee." Your possession and use of the detector cell may also be regulated by the state where you are located. The requirements of state regulatory agencies are substantially similar to those contained in NRC regulation 10 C.F.R. Section 31.5, but they may differ in some respects. It is suggested that you procure a copy of the regulations of your particular state. (Supplement 2 in Appendix I contains a list of the "Agreement States" which have been granted authority by the U.S. Nuclear Regulatory Commission to regulate the possession and use of radioactive material.) It is required that you be familiar with regulation 10 C.F.R. Section 31.5 (Appendix 1 in the Hardware Guide 09931371). Following are summaries of its requirements.

Labels Do not remove any of the labels attached to the ECD cell or any of the labels attached to your Clarus 690 Gas Chromatograph that refer to the ECD cell. Follow all instructions and abide by all precautions provided by the labels and in user instruction manuals referred to by the labels.

36 . Clarus 690 GC Customer Hardware and Service Guide

Leak Testing You are obligated under U.S. federal and state regulations to make certain that the ECD cell is wipe-tested for leakage of radioactive materials at intervals of no longer than six months, and that the analysis of these wipe tests is conducted by a person specifically licensed to do so, either by the U.S. Nuclear Regulatory Commission or by an Agreement State. The analyses can be performed by the firm listed below: National Leak Test Center P.O. Box 486 North Tonawanda, New York 14120

ECD Cell Failure or Dam age If a leak test detects more than 0.005 µCi (microcurie) of removable radioactive material on the surface of an ECD cell, or if the cell itself is damaged in such a way as to indicate that it may no longer adequately shield the radioactive material inside, you must immediately suspend operation of your chromatograph until the cell has been repaired or disposed of by a person specifically licensed to do so . Any such incident must be reported by you to the Regional Office, Inspection and Enforcement, U.S. Nuclear Regulatory Commission.

Reporting Radiation I ncidents, Theft or Loss Please read Regulation 10 C.F.R. Section 20.2201 and 20.2202. These describe your duties should the radioactive material (Ni 63) in the ECD cell be lost, stolen, or released, or should any person be exposed to radiation.

Other ECD R equirem ents Regulation 10 C.F.R. Section 31.5 (see Appendix I) does not permit you to abandon the ECD cell or export it. It may not be transferred except to a person specifically licensed to receive it. Within thirty days of such a transfer, you must report to the Director of Nuclear Material Safety and Safeguards, U.S. Nuclear Regulatory Commission, Washington, D.C. 20555, the name and address of the transferee. However, no report is needed to transfer a defective ECD cell to PerkinElmer in order to obtain a replacement. You may transfer the ECD cell to another general licensee, like yourself, only when it remains at the same location to which it was shipped by PerkinElmer. Give the transferee a copy of these instructions and the regulations in Appendix I, and report to the commission as required in Regulation C.F.R. Section 31.5.

N EVER DI SM ANTLE THE ECD CELL!! You can rem ove the ECD cell from the GC for repair . WARNING AVERTISSEMENT

N E JAM AI S DÉM ON TER LA CELLE ECD!! Vous pouvez supprim er la cellule ECD du GC pour réparation.

United K ingdom Regulations In the U.K., registration is required under the Radioactive Substances Act of 1960, for anyone keeping or using radioactive materials. Application should be made to any one of the following governing bodies: ENGLAND

Department of the Environment Queen Anne's Chambers

WALES

Welsh Office Cathay's Park

ECD Radioactive Hazards . 37 Tothill Street London, SW1H 9J4 SCOTLAND

Scottish Development Department 21 Hill Street Edinburgh, EH2 3J4

Cardiff, CF1 3NG NORTHERN IRELAND

Ministry of Development Parliament Building Storemont Belfast, Northern Ireland

38 . Clarus 690 GC Customer Hardware and Service Guide

Safe Handling of Gases When using hydrogen, either as the combustion gas for a flame ionization detector or as a carrier gas, special care must be taken to avoid buildup of explosive hydrogen/air mixtures. Ensure that all hydrogen line couplings are leak-free and do not allow hydrogen to vent within the oven. Warning: Compressed Gases

WARNING AVERTISSEMENT

High pressure gas cylinders can be dangerous if mishandled or misused. Always handle gas cylinders with caution and observe your local regulations for the safe handling of gas cylinders. Avertissem ent: Gaz com prim és Les bouteilles à gaz à haute pression peuvent être dangereuses si elles sont mal utilisées ou mal utilisées. Toujours manipuler les bouteilles de gaz avec précaution et respecter la réglementation locale pour la manipulation sécurisée des bouteilles de gaz.

Ventilation Adequate ventilation must be provided. When analyzing hazardous compounds, such as pesticides, it may be necessary to arrange for venting of detector effluent into a fume hood.

Using Hydrogen Flame Ionization Detectors (FID) uses hydrogen as fuel. If the hydrogen is turned on without a column attached to the injector and detector fittings inside the oven, hydrogen could diffuse into the oven creating the possibility of an explosion. To avoid possible injury, DO NOT TURN ON THE HYDROGEN UN LESS A COLUM N I S ATTACHED AN D ALL JOI NTS HAVE BEEN LEAK TESTED. Before disconnecting a colum n, m ak e certain that the hydrogen has been turned OFF.

WARNING AVERTISSEMENT

If two FIDs are installed and only one has a column attached to it, make certain that you cap off the unused detector inlet fitting with a 1/8-inch stainless steel plug (P/N N9300061). Les détecteurs d'ionisation de flamme (FID) utilisent de l'hydrogène comme carburant. Si l'hydrogène est allumé sans une colonne attachée aux raccords de l'injecteur et du détecteur à l'intérieur du four, de l'hydrogène pourrait se diffuser dans le four en créant la possibilité d'une explosion. P our éviter d'éventuelles blessures, N E P AS ALLUM ER SUR L'HYDROGÈN E À M OI NS QUE LA COLON NE EST ATTACHÉE ET TOUTES LES JOI NTS ON T ÉTÉ FÉM ES TESTÉS. Avant de déconnecter une colonne, assurez-vous que l'hydrogène a été éteint. Si deux FID sont installés et qu'un seul a une colonne qui y est attachée, assurez-vous que vous avez coupé le raccord d'entrée du détecteur inutilisé avec une prise en acier inoxydable de 1/8 pouces (P/N N9300061).

Safe Handling of Gases . 39

Contact the gas supplier for a m aterial safety data sheet (M SDS) containing detailed inform ation on the potential hazards associated w ith the gas. Carefully use, store, and handle com pressed gases in cylinders. Gas cylinders can be hazardous if they are m ishandled. WARNING AVERTISSEMENT

Contactez le fournisseur de gaz pour une fiche de données de sécurité (fiche signalétique) contenant des inform ations détaillées sur les dangers potentiels associés au gaz. Utiliser, stock er et m anipuler avec précaution des gaz com prim és dans des cylindres. Les bouteilles de gaz peuvent être dangereuses si elles sont m al usurées.

Warning: - EXPLOSION HAZARD

WARNING AVERTISSEMENT

The use of hydrogen as a carrier gas is dangerous. Hydrogen is potentially explosive and must be used with extreme care. Any use of hydrogen gas must be reviewed by appropriate health and safety staff and all installations of hydrogen systems must be performed to applicable codes and standards. It is important that the system be carefully checked on a routine basis for leaks before operation. It is highly recommended that the gas chromatograph be equipped with a hydrogen sensor if you use hydrogen as a carrier gas. When the system is not in use turn off all hydrogen tanks and vent the lines. PerkinElmer assumes no liability for the improper use of hydrogen as a carrier gas. Avertissem ent: RI SQUE D'EXPLOSI ON L'utilisation d'hydrogène comme gaz vecteur est dangereuse. L'hydrogène est potentiellement explosif et doit être utilisé avec une extrême prudence. Toute utilisation d'hydrogène gazeux doit être examinée par le personnel de santé et de sécurité approprié et toutes les installations de systèmes d'hydrogène doivent être effectuées conformément aux codes et normes applicables. Il est important que le système soit soigneusement vérifié régulièrement pour détecter les fuites avant utilisation. Il est fortement recommandé que le chromatographe en phase gazeuse soit équipé d'un capteur d'hydrogène si vous utilisez l'hydrogène comme gaz vecteur. Lorsque le système n'est pas utilisé, fermez tous les réservoirs d'hydrogène et purgez les conduites. PerkinElmer décline toute responsabilité en cas d'utilisation incorrecte de l'hydrogène comme gaz vecteur.

N OTE: The permanent installation of gas supplies is the responsibility of the user and should conform to local safety and building codes. Consult the following references for more detailed information and additional guidelines about gas cyclinders. •

Compressed Gas Association (USA), "Safe Handling of Compressed Gases in Containers," pamphlet no. P-1, 1984.

40 . Clarus 690 GC Customer Hardware and Service Guide •

Compressed Gas Association (USA), "The Inert Gases – Argon, Nitrogen and Helium," pamphlet no. P-9, 1992.

I dentification of Gas Cylinders •

Legibly mark cylinders to identify their contents. Use the chemical name or commercially accepted name for the gas.

Storing Gas Cylinders Review the following precautions with the customer to ensure the safe use and storage of gas cylinders. •

Cylinders should be stored in accordance with the regulations and standards applicable to the customer’s locality, state, and country.



When cylinders are stored indoors in storage rooms, the storage room should be well ventilated and dry. Ensure that the ventilation is adequate to prevent the formation of dangerous accumulations of gas. This is particularly important in small or confined areas.



Do not store cylinders near elevators, gangways, or in locations where heavy moving objects may strike or fall against them.



Use and store cylinders away from exits and exit routes.



Locate cylinders away from heat sources, including heat lamps. Compressed gas cylinders should not be subjected to temperatures above 52 °C (126 °F).



It is recommended that gas cylinders be stored and placed outside the laboratory and connected to the instrument through copper lines.

Handling of Gas Cylinders •

Do not allow ignition sources in the storage area and keep cylinders away from readily ignitable substances such as gasoline or waste, or combustibles in bulk, including oil.



Store cylinders standing upright, fastened securely to an immovable bulkhead or permanent wall.



When storing cylinders outdoors, they should be stored above ground on a suitable floor and protected against temperature extremes (including the direct rays of the sun).



Arrange gas hoses where they will not be damaged or stepped on and where things will not be dropped on them.



Take care not to kink or stress the gas lines. For safety, cylinders should be firmly clamped in position.



If it becomes necessary to move the cylinders, do so with a suitable hand truck after insuring that the container cap is secured and the cylinder is properly fastened to the hand truck.



Use only regulators, tubing and hose connectors approved by an appropriate regulatory agency.



Do not refill cylinders.



Check the condition of pipes, hoses and connectors regularly. Perform gas leak tests at all joints and seals of the gas system regularly, using an approved gas leak detection system.

Safe Handling of Gases . 41 •

When the equipment is turned off, close all gas cylinder valves tightly at the cylinder. Bleed the remainder of the line before turning the exhaust vent off.

42 . Clarus 690 GC Customer Hardware and Service Guide

Hazardous Chem icals Before using samples, you should be thoroughly familiar with all hazards and safe handling practices. Observe the manufacturer’s recommendations for use, storage and disposal. These recommendations are normally provided in the Material Safety Data Sheets (MSDS) supplied with the solvents. Be aware that the chemicals that you use in conjunction with the GC may be hazardous. DO NOT store, handle, or work with any chemicals or hazardous materials unless you have received appropriate safety training and have read and understood all related Material Safety Data Sheets (MSDS). MSDSs provide information on physical characteristics, precautions, first aid, spill clean up, and disposal procedures. Familiarize yourself with the information and precautions contained in these documents before attempting to store, use or dispose of the reagents. Comply with all federal, state, and local laws related to chemical storage, handling, and disposal. You must work under a suitable hood when handling and mixing certain chemicals. The room in which you work must have proper ventilation and a waste collection system. Always wear appropriate safety attire (full-length laboratory coat, protective glasses, gloves, etc.), as indicated on Material Safety Data Sheets.

WARNING AVERTISSEMENT

Some chemicals used with this GC may be hazardous or may become hazardous after completion of an analysis. The responsible body (for example, the Lab Manager) must take the necessary precautions to ensure that the GC operators and the surrounding workplace are not exposed to hazardous levels of toxic substances (chemical or biological) as defined in the applicable Material Safety Data Sheets (MSDS) or OSHA, ACGIH, or COSHH documents. Venting for fumes and disposal of waste must be in accordance with all national, state and local health and safety regulations and laws. Certains produits chimiques utilisés avec ce GC peuvent être dangereux ou peuvent devenir dangereux après l'achèvement d'une analyse. L'organisme responsable (par exemple, le gestionnaire de laboratoire) doit prendre les précautions nécessaires pour s'assurer que les opérateurs du GC et le lieu de travail environnant ne sont pas exposés à des concentrations dangereuses de substances toxiques (chimiques ou biologiques) telles que définies dans les fiches de données de sécurité applicables MSDS) ou les documents OSHA, ACGIH ou COSHH. L'évacuation des fumées et l'élimination des déchets doivent être conformes à toutes les réglementations et lois nationales, nationales et locales en matière de santé et de sécurité.

Definitions in W arning for Hazardous Chem icals Responsible body. “Individual or group responsible for the use and maintenance of equipment, and for ensuring that operators are adequately trained.” [per IEC 1010-1]. Operator. “Person operating equipment for its intended purpose.” [per IEC 1010-1]. OSHA:

Occupational Safety and Health Administration (United States)

ACGIH:

American Conference of Governmental Industrial Hygienists

COSHH:

Control of Substances Hazardous to Health (United Kingdom)

WEEE Instructions for PerkinElmer Products . 43

W EEE I nstructions for P erkinElm er P roducts

or A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive and is not to be disposed of as unsorted municipal waste. Any products marked with this symbol must be collected separately, according to the regulatory guidelines in your area. The objectives of this program are to preserve, protect and improve the quality of the environment, protect human health, and utilize natural resources prudently Requirements for waste collection, reuse, recycling, and recovery programs vary by regulatory authority at your location. Contact your local responsible body (for example, your laboratory manager) or authorized representative for information regarding applicable disposal regulations. Contact PerkinElmer at the web site listed below for information specific to PerkinElmer products. Web address: http://www.perkinelmer.com/WEEE For Customer Care telephone numbers select “Contact us” on the web page. Products from other manufacturers may also form a part of your PerkinElmer system. These other producers are directly responsible for the collection and processing of their own waste products under the terms of the WEEE Directive. Please contact these producers directly before discarding any of their products. Consult the PerkinElmer web site (above) for producer names and web addresses.

44 . Clarus 690 GC Customer Hardware and Service Guide

WEEE Instructions for PerkinElmer Products . 45

3 System Description

46 . Clarus 690 GC Customer Hardware and Service Guide

Overview of the Clarus 690 GC . 47

Overview of the Clarus 690 GC The Clarus 690 Gas Chromatograph is a dual-channel, temperature-programmable stand-alone gas chromatograph (GC). It is available in many configurations, such as with or without, an autosampler, programmable pneumatic control (PPC), and a variety of injector/detector combinations to provide you with total GC flexibility. The Clarus 690 GC is microprocessor controlled. You use the touch screen to operate this instrument.

Figure 1. The Clarus 690 GC. Your Clarus 690 GC may have none, one, or two of the following detectors installed: 

Flame Ionization (FID)



Photoionization (PID)



Electron Capture (ECD)



Nitrogen Phosphorus (NPD)



Thermal Conductivity (TCD)



Flame Photometric (FPD)

The FID, ECD, TCD, NPD, or FPD may be installed in either the front or the rear detector position. But only one PID can be installed and it must be installed only in the rear position (Channel 2). Each installed detector has one analog output which may be attached to either an integrator or recorder. Signals may be routed under instrument control. Either none, one, or two packed column injectors; none, one, or two capillary column injectors; or one of each injector type may be installed. Capillary column injectors consist of the conventional split/splitless injector (CAP), a temperature-programmed split/splitless injector (PSS), and a temperature programmed on-column injector (POC).

48 . Clarus 690 GC Customer Hardware and Service Guide Up to two gas sampling valves may be installed. In programmable pneumatic control (PPC), the carrier gas and detector gases are monitored and controlled by the microprocessor, thereby producing a fully automated system that is capable of managing all pneumatic functions within the gas chromatograph. The Clarus 690 GC can store up to five GC methods. Methods can be generated, copied, deleted, edited, set up, and printed. The autosampler can run up to 15 injections per vial from as many as 108 vials and one priority vial using one or two autosampler programs. In the latter case, a different GC method can be used by each program if desired. The autosampler operates in one of two program modes, single or multi. Real-time digital readouts are provided to simplify setting carrier gas pressures and flows, if flow readout or PPC is installed. An additional digital readout can be displayed to show the detector output (in mV).

About the Touch Screen . 49

About the Touch Screen You control the Clarus 690 GC by the touch screen. The touch screen is your link to the software. It is designed to provide the easiest access to the most frequently used features. The touch screen allows quick access to settings and functions, such as the Oven program and Ignite button. The System Status screen provides access to all heated zones. The icons represent configured components. These icons will guide you through the application. The System Status screen provides you with one-click access to the oven settings. For detailed information on the touch screen see the Clarus 690 User’s Guide (P/N 09931374).

Figure 2. The Clarus System Status screen.

50 . Clarus 690 GC Customer Hardware and Service Guide

About the Touch Screen . 51

4 Before You I nstall a Colum n

52 . Clarus 690 GC Customer Hardware and Service Guide

Column Installation Information . 53

Colum n I nstallation I nform ation

WARNING AVERTISSEMENT

At the moment you turn on the Clarus 690 GC, the oven, injector(s), and detector(s) begin to heat up rapidly. To avoid injury while installing a column, all heaters should be turned off and their respective zones allowed to cool before touching the injector septum caps or any of the fittings inside the oven. Au moment où vous allumez le Clarus 690 GC, le four, l'injecteur (s) et le (s) détecteur (s) commencent à réchauffer rapidement. Pour éviter toute blessure lors de l'installation d'une colonne, tous les radiateurs doivent être désactivés et leurs zones respectives sont autorisées à refroidir avant de toucher les capuchons de cloison de l'injecteur ou l'un des raccords à l'intérieur du four.

This chapter contains general column installation information, and the following procedures: •

Installing your column



Protecting your column

I njector and Detector Fittings Columns are installed inside the oven. The injector fittings are on the left side and the detector fittings are on the right side of the oven ceiling. Figure 3 shows a capillary injector fitting in the front position and a packed injector fitting in the rear position.

WARNING AVERTISSEMENT

Before installing a column, make certain the oven is OFF (by opening the oven door), the oven fan has stopped, and the oven is cool. Avant d'installer une colonne, assurez-vous que le four est éteint (en ouvrant la porte du four), le ventilateur du four s'est arrêté et le four est frais.

Figure 3. Injector and detector fittings.

Colum n Hangers Capillary columns are supported on column hangers. The left and right sides of the oven walls each have two rectangular slots into which column hangers are inserted. The two rear slots are used to install a column hanger in the rear position. The two front slots are used to install a column in the front position.

54 . Clarus 690 GC Customer Hardware and Service Guide To install a column hanger, simply insert one end into the left slot and the other end into the right slot. If you are installing two capillary columns, install the rear hanger and the rear column before installing the front hanger and the front column.

Figure 4. A column hanger installed in the oven rear position.

P rotecting Your Colum n The Clarus 690 GC provides a means for protecting your column(s) from overheating. It does this by not allowing the oven to heat up beyond the Oven Maximum Temperature Limit (OMTL), a value that you set in the active method. You should set the OMTL equal to or less than the maximum permissible operating temperature recommended in the specifications for your column. If you are installing two columns, use the lower of the two permissible maximum operating temperatures. To protect the column, the OMTL value (that you enter) works in conjunction with the Oven Temperature specified in the Active Method. Should you (or someone else) attempt to set an Oven Temperature in the Active Method to a value greater than the OMTL, the system displays an error message and will not allow you to continue until an appropriate new oven temperature is set. To set the oven temperature: 1. Go to the System Status screen and touch the oven icon:

2. Touch the field you need to update. It will appear black and then using the plus and minus buttons input the correct oven temperatures and time.

Protecting Your Column . 55 For information on turning the heaters and detectors on and off see the Clarus 690 User’s Guide

(P/N 09931374) chapter on the Active Method.

56 . Clarus 690 GC Customer Hardware and Service Guide

Protecting Your Column . 57

5 I nstalling a Packed Colum n

58 . Clarus 690 GC Customer Hardware and Service Guide

About this Chapter . 59

About this Chapter This chapter describes procedures for installing a packed column and setting the carrier gas flow using programmed pressure controlled (PPC). After installing a packed column as described in this chapter, you should set up the detector(s) to be used with this injector.

Sum m ary The following steps summarize how to install a new packed column and get it ready for use: Step 1: Turn off the heaters. Step 2: Set the carrier gas flow. Step 3: Attach one end of the column to the packed injector. Step 4: Leak test. Step 5: Condition the column Step 6: Attach the other end of the column to the detector. Step 7: Leak test the column/detector connection.

N OTE: A new packed column should not be used until it is conditioned overnight with carrier gas flowing through the column. Do not connect the column to the detector during the conditioning process.

M aterials and Tools R equired •

Two 7/16-inch open end wrenches when a normal 1/8-inch column is being installed. Two 9/16-inch open end wrenches will be required in addition if a 1/4-inch column is being installed.



Two 1/8-inch to 1/4-inch adapters (P/N 00080100) if a 1/4-inch column is being installed.



A flow meter or the optional flow readout.



A packed column of your choice.



A leak test solution or an electronic leak tester.

60 . Clarus 690 GC Customer Hardware and Service Guide

P acked Colum n I njector Overview The packed column injector consists of a septum cap, needle guide, quartz injector liner, and the injector body. This injector is used with 1/8-inch or 1/4-inch glass or metal packed columns. In addition, by installing the 530 Micron Wide-Bore Adapter Kit (P/N N6120001) you can convert the injector to accept wide-bore capillary columns.

Figure 5. Packed Column Injector.

About the W ide-Bore Adapter If you are operating in the off-column mode at above optimum flow rates (>10 mL/min), you may not need to install the wide-bore quartz injector liner. Depending on your sample or solvent, the solvent profile (tail) may be acceptable for your application with the standard liner (the illustration at the left in Figure 6) and the addition of the adapter fitting (P/N N6100083). However, if the solvent profile is not acceptable, install the wide-bore quartz injector liner. The off-column or on-column flash vaporization mode of operation is determined by the position of the hourglass portion of the wide-bore quartz injector liner in the packed column injector. When installed correctly, this liner produces improved solvent profiles, especially at optimum flow rates. For complete installation instructions, refer to the Installation Instructions: 530 Micron Wide-Bore Adapter Kit for the AutoSystem GC and Clarus GC (P/N 09938661). Insert the wide-bore quartz injector liner (P/N N6121003) into the packed column injector with the hourglass portion in the correct position for your desired mode of operation. Figure 6 shows a cross section of a packed column injector containing a standard liner and a cross section of a packed column injector containing a wide-bore quartz injector liner installed in the off-column position and the on-column position.

Packed Column Injector Overview . 61 For off-colum n flash vaporization (hourglass end first) : To avoid contaminating the quartz wool, wear vinyl, powder-free disposable gloves (the same type used to perform maintenance on Clarus SQ 8). Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch rod (P/N N610T100), push the quartz wool into the liner. Loosely pack some quartz wool in the top portion of the liner to wipe the syringe needle upon injection. Insert the wide-bore quartz injector liner into the packed column injector with the hourglass end first. Or

For on-colum n flash vaporization (hourglass end last):

Insert the wide-bore quartz injector liner into the packed column injector with the hourglass end of the liner last. Do not pack the wide-bore quartz injector liner with silanized quartz wool. You must use a 0.47-mm O.D. syringe in this mode. If you are using the autosampler, install a 0.47-mm O.D. syringe and use the “SLOW" injection mode. If the Clarus 690 GC software is Rev. 1.6 or lower, use the “ON-COL" injection mode.

Figure 6. Cross sections of three packed injector configurations with a wide-bore column.

62 . Clarus 690 GC Customer Hardware and Service Guide

Step 1: Turn off the Heaters

WARNING AVERTISSEMENT

The moment the Clarus 690 GC is turned on, the oven, injector(s), and detector(s) begin to heat up rapidly. To avoid burns and injury while installing a column, all heaters should be turned off and their respective zones allowed to cool before touching the injector septum caps or any of the fittings inside the oven. Au moment où le Clarus 690 GC est allumé, le four, l'injecteur (s) et le (s) détecteur (s) commencent à réchauffer rapidement. Pour éviter les brûlures et les blessures lors de l'installation d'une colonne, tous les radiateurs doivent être éteints et leurs zones respectives sont autorisées à refroidir avant de toucher les capuchons de cloison de l'injecteur ou l'un des raccords à l'intérieur du four.

N OTE: See the Clarus 690 GC User’s Guide for detailed procedures for turning heaters off and on. N OTE: It is recommended that you remove the injector liner shipped with the packed injector and pack it with a small amount of silanized glass wool before performing analyses. Please refer to the Maintenance chapter later in this manual.

Step 2: Set the Carrier Gas Flow This step describes how to set the carrier gas using PPC pneumatics modules. 1. Turn on the carrier gas at the tank and adjust the line pressure between 60 to 90 psig.

2. From the System Status screen select either A or B injector (the example here shows the A injector is the packed column injector). 3. Touch the Pkd button

and the following screen appears:

Step 2: Set the Carrier Gas Flow . 63

4. In the above window touch the Program button under the Carrier Gas field. 5. Touch a field to select it and using the up and down arrow buttons or the keypad button to enter the desired values

Setting the Carrier Gas Flow Using a Soap Bubble or Electronic Flow m eter The procedure below assumes that you know how to measure carrier gas flow using a soap bubble or electronic flowmeter and the built-in stopwatch. If you need instructions, please read “Using Tools,” in the Clarus 690 GC User’s Guide (P/N 09931374) before proceeding. 1. Locate the packed injector fitting inside the oven. 2. Attach a soap bubble flowmeter to the packed injector fitting.

64 . Clarus 690 GC Customer Hardware and Service Guide

Figure 7. A packed injector fitting. 3. Turn on the carrier gas at the tank and adjust the line pressure to 90 psig. 4. From the System Status Screen select the Tools button and then select Utilities from the drop down menu.

. 6. In the Utilities screen select the Stopwatch icon.

Step 2: Set the Carrier Gas Flow . 65

7. Using the up and down arrow buttons or the keypad

button enter the

appropriate values and press Start.

8. Start the carrier gas flowing by turning the flow controller knob counterclockwise and measure the flow.

N OTE: For best accuracy, use a soap bubble flowmeter volume that gives a reading of at least 30 seconds. 9. Adjust the flow to the desired set point by repeatedly measuring the flow and turning the flow controller knob counterclockwise to increase the flow, clockwise to decrease the flow, until the desired flow is obtained. 10. Disconnect the soap bubble flow meter before proceeding to the next step.

66 . Clarus 690 GC Customer Hardware and Service Guide

Step 3: Connect One End of the Colum n to the P acked I njector N OTE: If you are installing a 1/4-inch column, attach a 1/8-inch to 1/4-inch adapter to the packed injector fitting before continuing. Finger-tighten the adapter, then while holding the packed injector fitting steady with a 7/16-inch wrench, tighten the adapter with a 9/16-inch wrench. 1. Insert one end of the column into the packed injector fitting until it bottoms, then finger tighten the column nut onto the packed injector fitting (see below). 2. While holding the packed injector fitting with one 7/16-inch wrench, tighten the column nut an additional 1/8 to 1/4 turn with the other wrench.

N OTE: Do not overtighten column nuts. Overtightening causes permanent damage to the fittings.

Figure 8. Packed column connected to a packed injector fitting.

Step 4: Leak Test Test the connection to the packed injector fitting for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To avoid contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

Step 5: Condition the Column . 67

Step 5: Condition the Colum n This section contains a suggested temperature program for conditioning a column. The program starts off by holding the oven temperature at a medium value for 10 minutes, gradually increasing the oven temperature at a fixed rate (5 °C/min) to the column operating temperature, then holding that temperature overnight with the carrier gas flowing.

CAUTION ATTENTION

The temperatures shown in the following examples should only be used as guidelines. Please refer to the column manufacturer's operating instructions for specific temperature recommendations. Les températures indiquées dans les exemples suivants ne doivent être utilisées que comme directives. Reportez-vous au mode d'emploi du fabricant de la colonne pour des recommandations de température spécifiques.

To condition the column: 1. Close the oven door, go to the System Status screen, and touch the Oven icon.

2. Select the Init. Temp field. Then using the up and down arrow or keypad buttons enter the initial (Init) oven temperature to a set point of 50.

68 . Clarus 690 GC Customer Hardware and Service Guide

3. Enter an Init. TIME of 10. 4. To add another program step, enter a RATE of 5 (°C/min). 5. Using the up and down arrow or keypad

buttons enter for Temp 2, a set

point 25 °C to 50 °C above your planned analytical operating temperature. For example, enter a set point of 150.

6. Set Time 2 to 999. This displays HOLD in the field.

CAUTION

To avoid damaging the column, do not enter a temperature higher than the maximum operating temperature specified by the column manufacturer. Pour éviter d'endommager la colonne, n'entrez pas une température supérieure à la température de fonctionnement maximale spécifiée par le fabricant de la colonne.

7. Select the B injector by clicking the B-Pkd tab.

Step 5: Condition the Column . 69

8. In the above screen touch the temp field, then using the up and down arrow or keypad buttons set an Injector Temp about 50 °C higher than the Oven Temp 2 setting. See the chapter “Active Method” in the Clarus 690 GC User’s Guide for information on setting Temp 2. 9. Turn Detector Temperature off and run the system overnight. 10. The next morning go to the System Status screen and select the Oven icon. 11. Reset the oven temperature set point to that specified for TEMP 1 at the beginning of the temperature program. 12. Open the oven door and allow the oven to cool until the oven fan goes off. This occurs when the oven cools down to 40 °C.

N OTE: Condition a new column before using it in an analysis. Once it is conditioned, you will not need to recondition it.

70 . Clarus 690 GC Customer Hardware and Service Guide

Step 6: Attach the Other End of the Colum n to the Detector 1. Insert the free end of the column into the detector fitting, then finger tighten the column nut onto the detector fitting.

Figure 9. Packed column attached to the rear detector fitting. 2. While holding the detector fitting with one of the 7/16-inch wrenches, tighten the column nut an additional 1/8 to 1/4 turn with the other wrench. CAUTION ATTENTION

Make certain that no part of the column touches the bottom or sides of the oven once it is installed. Assurez-vous qu'aucune partie de la colonne ne touche le fond ou les côtés du four une fois qu'il est installé.

N OTE: If you are installing a 1/4-inch column, attach a 1/8-inch to 1/4-inch adapter to the detector fitting before continuing. Finger-tighten the adapter, then while holding the detector fitting steady with a 7/16-inch wrench, tighten the adapter with a 9/16-inch wrench.

Step 7: Leak Test the Column/Detector Connection . 71

Step 7: Leak Test the Colum n/ Detector Connection The following procedures describe leak testing the column to detector connections using PPC Pneumatics. 1. From the System Status screen select either A or B injector (the example here shows the B packed injector ready for setting the carrier gas). In this example, touch the Channel B packed injector icon.

2. In the following screen touch the Program button under the Carrier Gas field.

3. Using the up and down arrow or keypad 100 psig for minimum and maximum.

buttons enter pressure 0.1 to

72 . Clarus 690 GC Customer Hardware and Service Guide

The Clarus 690 GC monitors the column head pressure with a carrier-gas mass-flow controller in use. After selecting the carrier-gas Flow mode in configuration, you enter the minimum and maximum values. If the inlet pressure drops below or rises above the set limits for more than one consecutive minute, an error message will appear: Carrier (with specific carrier number) unable to maintain pressure.

N OTE: Refer to the Clarus 690 GC User’s Guide (P/N 09931374) for information on minimal and maximum values. You should correct the leak; the most common area of a leak is the injector septum. Clear the message by touching the OK button before continuing to use the GC. This will stop an autosampler program from continuing. If you do not make any entries, leak checking will not be done. If you are temperature programming, then the maximum value you enter must be above the column head pressure at the highest oven temperature. You can just enter a value for the minimum and leave the maximum blank. See the Clarus 690 Users Guide (P/N 09931374) for more information.

Step 7: Leak Test the Column/Detector Connection . 73

6 I nstalling a Capillary Colum n

74 . Clarus 690 GC Customer Hardware and Service Guide

About this Chapter . 75

About this Chapter This chapter describes how to install a capillary column in the following three injectors: 

Capillary Split/Splitless (CAP)



Programmed Split/Splitless (PSS)



Programmed On-Column (POC)

The information in this chapter is presented as one sequential procedure (Steps 1 through 7) for all three injectors (CAP, PSS, and POC) with the following procedural steps common to all three injectors: 

Setting carrier gas flow using PPC modules

N OTE: For a thorough understanding of PPC, refer to Chapter 8, “PPC Fundamentals.” 

Leak testing



Conditioning the column



Attaching the column to the detector and leak checking

N OTE: If you are analyzing reactive compounds, appropriately deactivate injector liners and wool for your sample type.

CAUTION ATTENTION

All three capillary injectors; CAP, PSS, and POC use a 1/16-inch fitting for the column connection. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the column nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut. Les trois injecteurs capillaires; CAP, PSS et POC utilisent un raccord de 1/16 pouces pour la connexion de la colonne. Ce raccord est fragile. Pour préserver l'intégrité du raccord, connectez soigneusement l'écrou de la colonne pour empêcher le filetage croisé du raccord et / ou le serrage excessif de l'écrou sur le raccord. Vous pouvez également préserver l'intégrité du raccord en permettant à l'injecteur de refroidir avant de brancher un écrou.

Sum m ary The following steps summarize how to install a capillary column and get it ready for use: Step 1: Turn the heaters off. Step 2: Connect the column to the CAP, PSS, or POC injector. 1. Split/Splitless (CAP) injector 2. Programmed Split/Splitless (PSS) injector 3. Programmed On-Column (POC) injector Step 3: Set the carrier gas. (Set the pressure for the CAP and the PSS, or the flow for the POC, using the optional flow readout or a flowmeter.) Step 4: Leak test all new connections.

76 . Clarus 690 GC Customer Hardware and Service Guide Step 5: Condition the column (to the manufacturers specifications) and the mechanical joint between the column and pre-column. Step 6: Connect the column to the detector. Step 7: Leak test all new connections.

Step 1: Turn the Heaters Off

CAUTION ATTENTION

The moment the Clarus 690 GC is turned on, the oven, injector(s), and detector(s) begin to heat up rapidly. To avoid injury while installing a column, all heaters should be turned off and their respective zones allowed too cool before touching the injector septum caps or any of the fittings inside the oven. Au moment où le Clarus 690 GC est allumé, le four, l'injecteur (s) et le (s) détecteur(s) commencent à réchauffer rapidement. Pour éviter les blessures lors de l'installation d'une colonne, tous les radiateurs doivent être éteints et leurs zones respectives sont trop froides avant de toucher les capuchons de la cloison de l'injecteur ou l'un des raccords à l'intérieur du four.

CAUTION ATTENTION

Before you install a column, follow the detailed procedures for turning heaters off and on, in Chapter 4 of this manual “Before You Install a Column.” If you have not read this chapter, please do so before proceeding. Avant d'installer une colonne, suivez les procédures détaillées pour éteindre et éteindre les réchauffeurs, au chapitre 4 de ce manuel "Avant d'installer une colonne". Si vous n'avez pas lu ce chapitre, faites-le avant de continuer.

Step 2: Connect the Column to the Injector . 77

Step 2: Connect the Colum n to the I njector This step contains a separate procedure that describes how to connect a column to each of the following three Clarus 690 GC injectors: Step 2A:

Connect the Column to the Split/Splitless (CAP) Injector.

Step 2B:

Connect the Column to the Programmed Split/Splitless (PSS) Injector.

Step 2C:

Connect the Column to the Programmed On-Column (POC) Injector.

Proceed to the section that contains the procedure for your injector.

Step 2A: Connect the Colum n to the Split/ Splitless (CAP ) I njector The Split/Splitless injector (CAP) consists of a septum purge assembly and the injector body. Carrier gas enters the injector body at the point just above the O-ring and flows through the quartz liner past the column tip.

Figure 10. Cutaway view of the Split/Splitless injector (CAP).

78 . Clarus 690 GC Customer Hardware and Service Guide

About the I njector Liners The CAP injector uses the following quartz liners: •

Splitless Narrow Bore Liner (P/N 09200696)



Split Wide Bore Liner (P/N 09200624)

The narrow-bore liner is generally used for splitless injections and the wide-bore liner is generally used for split injections. Due to the small internal volume (0.3 mL) of the narrow-bore liner, you can prevent overfilling the liner with vapor (caused by solvent expansion upon injection) by limiting the amount of sample injected to 0.5 µL. This prevents the solvent expansion upon injection from overfilling the liner with vapor. The wide-bore liner is used for splitless injection volumes over 0.5 µl since its internal volume is 1.25 mL. The sample size should be limited to a maximum of 2 µL for hydrocarbon solvents and less than that for high-expansion solvents, such as water or CH2Cl. Refer to Table 6-1. If the wide-bore liner is used for splitless injection, the splitless sampling time (vent-on time) should be more than one minute. Also, lower initial oven temperatures may be required to give good resolution in the first few minutes after the solvent peak. The wide-bore liner should be used with columns having an i.d. of 0.32 mm or greater. To wipe the syringe needle, we recommend packing a small amount of quartz wool in the top portion of all liner types or injection modes (for example, split or splitless). Each liner should be packed with the quartz wool as described later in this chapter.

Splitless I njections In the splitless injection mode, the narrow-bore quartz liner is typically used without quartz wool. The narrow-bore decreases the sample residence time in the liner, making it useful for trace analysis with smaller sample volumes (0.5 µL or less). By closing the split vent, most of the sample mixture enters the column. Then, opening the split vent clears the inlet of residual solvent. For splitless injection volumes over 0.5 µL, the wide-bore liner with an internal volume of 1.25 mL should be used. However, the amount of sample should be limited to a maximum of 2 µL for hydrocarbon solvents and less than that for high-expansion solvents such as water or CH2Cl2. Refer to Table 6-1 for examples of gas volumes formed upon sample injection for selected solvents. If the wide-bore liner is used for splitless injection, the splitless sampling time (vent-on time) should be at least one minute or more. Also, lower initial oven temperatures may be required to produce good solute resolution in the first few minutes after the solvent peak. The wide-bore liner should be used with columns having an i.d. of 0.32 mm or greater. Table 6-1. Gas Volumes Formed Upon Sample Injection (Injector 250°C, Inlet Pressure 10 psig) Solvent Methylene Chloride Methanol Water

Volume Injected (µL)

Gas Volume Generated (µL)

1

333

2

571

1

475

2

768

1

823

Step 2: Connect the Column to the Injector . 79 2

1166

Split I njections In the split injection mode, the wide-bore quartz liner is packed with quartz wool to ensure thorough mixing of the sample and carrier gas before they encounter the column tip. The split vent is open at the time of injection so that a fraction of the sample mixture enters the column while the remainder is routed out through the split vent.

About the Pneum atics Control The injector pneumatics are PPC (programmable pneumatic controlled) modules that regulate the inlet flow and pressure of the gases using software controlled by an Clarus 690 GC method.

CAUTION ATTENTION

The CAP injector is shipped with the wide-bore liner installed without quartz wool packing. Before using the injector, remove the liner and pack it with quartz wool. If you are using the injector in the splitless mode, you may want to install the narrow-bore liner. L'injecteur CAP est livré avec la doublure à large éventail installée sans emballage en laine de quartz. Avant d'utiliser l'injecteur, retirer la doublure et l'emballer avec de la laine de quartz. Si vous utilisez l'injecteur en mode sans partage, vous voudrez peut-être installer le revêtement étroit.

Connecting a Colum n to the CAP I njector The following five steps describe how to connect a column to the CAP injector: Step 1. Remove the CAP injector liner. Step 2. Select an appropriate CAP injector liner. Step 3. Pack the CAP injector liner with quartz wool. Step 4. Reinstall the liner in the CAP injector. Step 5. Connect a column to the CAP injector.

80 . Clarus 690 GC Customer Hardware and Service Guide Step 1. Remove the CAP Injector Liner. To remove a CAP injector liner:

Figure 11. Location of the CAP injector. 1. Ensure that the injector heater has been turned off. Allow the injector to cool until it is slightly warm to the touch. 2. Remove the injector cover and bezel.

Figure 12. Removing the injector cover and bezel. 4. Locate the septa purge assembly lock nut highlighted in blue.

Step 2: Connect the Column to the Injector . 81

Figure 13. Septum purge assembly locknut. 5. Unscrew the septa purge lock nut and carefully tilt the assembly out of the way as shown below to access the liner and liner O-ring.

Figure 14. Tilting the septum purge assembly out of the way. The carrier gas inlet line and septa purge line may be bent slightly. 6. Pull the liner up out of the injector body. The O-ring may stay on the liner or may stay in the top of the injector.

82 . Clarus 690 GC Customer Hardware and Service Guide

Figure 15. Removing a capillary injector liner.

CAUTION ATTENTION

If the O-ring adheres to the injector body, use a small screwdriver to loosen the O-ring so that you can remove the liner and O-ring. Be careful not to scratch the barrel w here the O-ring seals. Discard this O-ring and install a new O-ring. Si le joint torique adhère au corps de l'injecteur, utilisez un petit tournevis pour desserrer le joint torique afin que vous puissiez enlever le revêtement et le joint torique. Veillez à ne pas rayer le baril où le joint torique obture. Jeter ce joint torique et installer un nouveau joint torique.

N OTE: If the liner breaks inside the CAP injector, it can be removed by first removing the column. Then using a 9/16-inch wrench, remove the 1/4-inch injector fitting inside the oven. The liner should fall out. If the liner is stuck, you can push it out from the top or bottom. Step 2. Select an Appropriate CAP Injector Liner. Select the appropriate injector liner for your application. Step 3. Pack the CAP Injector Liner with Quartz Wool. To wipe the syringe needle, we recommend packing a small amount of quartz wool in the top portion of the liner regardless of the liner type or injector mode used (for example, split or splitless). This packing assures that reproducible volumes are injected because it wipes the syringe needle every time the needle is inserted. Remove the liner and replace the quartz wool packing on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest, tailing, and loss of sensitivity. Remove the wool with a small hook on the end of a thin wire, or blow it out using compressed air.

N OTE: To avoid contaminating the quartz wool when packing the injection liner, wear vinyl, powderfree, disposable.

Step 2: Connect the Column to the Injector . 83 Packing CAP Injector Liner for Split Operation Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch rod (P/N N610T100), push the quartz wool into the liner. Pack a one-inch piece (2.5 cm) of quartz wool tightly 1 from about one inch (2.5 cm) from the bottom upwards. Loosely pack quartz wool in the top portion of the liner to wipe the syringe needle upon injection. Packing a CAP Injector Liner for Splitless Operation Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch rod (P/N N610T100), push the quartz wool into the liner. Pack a one-inch piece (2.5 cm) of quartz wool loosely below the top ground portion of the liner (see Figure 17). The sample is then injected into the wool, thereby preventing the delivery of sample beyond the column. The wool also wipes the syringe needle upon injection. Remove the wool by making a small hook on the end of a thin wire and using that to pull it out, or blow it out using compressed air.

Figure 16. Examples of CAP injector liners packed with quartz wool.

N OTE: As you can see in Figure 17, each CAP injector liner has an O-ring installed. If the O-ring has adhered to the liner, it may not be easy to remove the liner. (If this is the case, use a small screwdriver to dislodge the O-ring before removing the liner and O-ring). Injector O-Rings

Recommended Maximum Temperature

CAP Injector 09200725 Kalrez (pkg. of 1)

450 °C

The recovery of high molecular weight components (e.g., C40) may be improved if the liner is loosely packed.

1

84 . Clarus 690 GC Customer Hardware and Service Guide 09200714 Viton (pkg. of 1)

250 °C (not recommended for use with ECD)

Step 4. Reinstall the Liner in the CAP Injector. To reinstall the liner: 1. Install a new O-ring on the liner. 2. Insert the liner in the injector body. 3. Place the septum purge lock nut back over the injector and secure it in place. 4. Replace the injector cover and bezel. Step 5. Connect a Column to the CAP Injector. CAUTION ATTENTION

Injector must be cooled to less than 100 ºC before removing the column nut from the gold seal nut. L'injecteur doit être refroidi à moins de 100 ºC avant de retirer l'écrou de colonne de l'écrou d'étanchéité en or.

To connect a column: 1. Insert a 1/16-inch column nut (P/N N6552084) and one of the following three 1/16inch graphite ferrules over one end of the column as shown below. For example, 0.4 mm i.d. for 0.25-mm i.d. columns, P/N 09200685; 0.5 mm i.d. for 0.32-mm i.d. columns, P/N 09200686; or 0.8 mm i.d. for 0.53-mm i.d. columns, P/N 09200687.

Figure 17. Narrow-bore capillary column, nut, and ferrule on the injector end of a column.

N OTE: Verify that the tapered end of the ferrule is facing towards the nut as shown above. 2. Cut off about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N9301376, pkg. of 10 scribes). 3. Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to the examples shown in the following figure:

Step 2: Connect the Column to the Injector . 85

Bad Cuts

Good Cut

Figure 6-11. Example of a good cut and bad cuts. 4. To automatically calculate the position of the column nut on the column so that the back of the nut is the correct distance from the end of the column, go to the system Status Screen and touch the Tools button.

5. Touch Utilities from the drop-down menu. The above Utilities screen appears. 6. In the Utilities screen touch the Column Length Calc icon

.

The Column Insertion Length screen appears. 7. Using the drop-down menu touch the type of Detector and injector you are using. Select whether you want the calculation in mm or inches. The proper distance will be calculated automatically.

86 . Clarus 690 GC Customer Hardware and Service Guide

Figure 18. Positioning the column in the column nut (P/N N6552084). 8. Using a septum position it on the column just beyond the back edge of the column nut (see Figure 19). You can also use “white-out” or a felt-tipped pen to make a mark on the column just beyond the back edge of the column nut.

CAUTION ATTENTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne contactent pas la marque sur la colonne.

9. Locate the capillary injector fitting inside the oven. (highlighted in blue)

Step 2: Connect the Column to the Injector . 87

Figure 19. Capillary injector fitting inside the oven. 10. Carefully insert the column and column nut into the capillary injector fitting. Then handtighten the column nut onto the capillary injector fitting.

Figure 20. Inserting the column and nut into the capillary injector fitting inside the oven. 11. Carefully push/slide the column into the gold seal nut until the septum (or mark) is aligned with the back of the column nut. 12. Using one 1/4-inch wrench, tighten the column nut only until the column cannot be pulled out of the fitting. CAUTION ATTENTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

88 . Clarus 690 GC Customer Hardware and Service Guide

Figure 21. Capillary column attached to capillary injector fitting.

TIP

To ensure a leak-free connection, cycle the oven temperature to the maximum temperature in the method, then cool. Once cool recheck the tightness of the column nut then retighten if necessary.

Step 2: Connect the Column to the Injector . 89

Step 2B Connect the Colum n to the P rogram m ed Split/ Splitless (P SS) I njector: The Programmed Split/Splitless injector (PSS) consists of a septum purge assembly and the injector body. Carrier gas enters the injector at a point just above the O-ring and flows through the quartz liner past the column tip.

Figure 22. Cutaway view of the Programmed Split/Splitless injector (PSS).

About the I njector Liners The PSS injector uses the following three quartz liners: 

2.0-mm i.d. (wide-bore) liner (P/N N6121004)



1-mm i.d. (narrow-bore) liner (P/N N6121006)



Hourglass (on-column) liner (P/N N6101539)

In general, for split or splitless injections, use the 2-mm or 1-mm i.d. liner and operate the PSS in the inlet-programmed mode. For on-column operation, use the hourglass liner and the oven program mode. The 2-mm i.d. liner should be packed with quartz wool as described in this chapter and used for either split or splitless operation. The 1-mm i.d. liner may give better resolution of the early-eluting peaks in the split or splitless mode and it is better for labile compounds; however, it should be used for those samples with early-eluting peaks for which

90 . Clarus 690 GC Customer Hardware and Service Guide additional solute trapping focusing cannot be obtained by lowering the initial oven temperature or by using a column with a thicker stationary phase film. To wipe the syringe needle, in all liner types (wide-bore or narrow-bore) or injection modes (split or splitless), we recommend packing a small amount of quartz wool in the top portion of the liner. The sample is injected into the PSS injector at a "cool" temperature. The injector temperature is then programmed to increase. This is useful for samples that are thermally labile and/or have a wide molecular weight range. The PSS injector can also be used in a programmed on-column mode by replacing the quartz liner with the hourglass liner and closing the split vent flow.

CAUTION ATTENTION

CAUTION ATTENTION

The PSS can be used in the "hot" split or splitless mode. This, however, is not recommended for use with the 1 mm i.d. liner; it could cause solvent flashback in the injector. This mode should be used with caution depending upon the solvent and temperatures you choose. See Table 6-1. Le PSS peut être utilisé dans le mode «chaud» ou fractionné. Ceci, cependant, n'est pas recommandé pour une utilisation avec i.d. 1 mm. doublure; Cela pourrait provoquer un retour de solvant dans l'injecteur. Ce mode doit être utilisé avec prudence en fonction du solvant et des températures que vous choisissez. Voir le tableau 6-1.

When using the PSS in the on-column mode with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N6101253 or N6101380). Refer the Active Method Chapter in the Clarus 690 Users Guide, “Controlling the Autosampler,” for more detail. You must use only the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle; the "Slow" injection speed may produce double peaks due to the momentary stoppage of column flow during the longer injection. You can achieve better precision in the on-column mode when sample volumes of 1.0 µL or greater are injected. Lorsque vous utilisez le PSS en mode colonne avec l'autoéchantillonneur, vous devez utiliser une seringue spéciale qui a une aiguille o.d. De 0,47 mm (P/N N6101253 ou N6101380). Reportez-vous au chapitre Méthode active dans le Guide de l'utilisateur Clarus 690, «Contrôle de l'échantillonneur automatique», pour plus de détails. Vous devez utiliser uniquement la vitesse d'injection "Norm" avec cette seringue en mode colonne. La vitesse d'injection "rapide" plie cette fine aiguille; La vitesse d'injection "lent" peut produire des pics doubles en raison de l'arrêt momentané du flux de colonne lors de l'injection prolongée. Vous pouvez obtenir une meilleure précision dans le mode en colonne lorsque des échantillons de 1,0 μL ou plus sont injectés.

Step 2: Connect the Column to the Injector . 91 Split Injections In the split mode, the wide-bore quartz liner is packed with quartz wool to insure thorough mixing of the sample and carrier gas before they reach the column tip. The split vent is open at the time of injection so that a fraction of the sample mixture enters the column while the remainder is routed out through the split vent. Splitless Injections In the splitless mode, the narrow-bore quartz liner is typically used without tightly packed quartz wool for mixing. Instead, a small amount of quartz wool in the center is recommended. This is useful for trace analysis with smaller sample volumes (less than 1 µL). By closing the split vent, most of the sample mixture enters the column, then the split vent is opened to clear the injector inlet of residual solvent. Solvent Purge Injections In applications where the sample is in a trace concentration and has a high molecular weight, this sample can be injected into the PSS with the vent open and the starting temperature near the boiling point of the solvent. In this way, only the solvent is purged out through the vent, then the vent is closed and the injector temperature is programmed up to elute the peaks of interest onto the column. This technique allows larger quantities of sample to enter the column without large solvent effects.

About the Pneum atics Control The PSS pneumatics consist of PPC (programmable pneumatics controlled) modules that regulate the inlet flow and pressure of the gases using software controlled by an Clarus 690 GC method.

CAUTION ATTENTION

The PSS injector is shipped with the wide-bore liner installed but without the quartz wool packing. Before using the injector, remove the liner and pack it with quartz wool. L'injecteur PSS est livré avec la doublure à large éventail installée mais sans l'emballage en laine de quartz. Avant d'utiliser l'injecteur, retirer la doublure et l'emballer avec de la laine de quartz.

Connect a column to the programmed split/splitless injector in either one of the following two modes: •

Split/Splitless Mode



On-Column Mode using the hourglass liner

Connect the Colum n to the PSS I njector The following five steps summarize how to connect a column to the PSS injector in the split/splitless mode: Step 1. Remove the PSS injector liner. Step 2. Select an appropriate PSS injector liner. Step 3. Pack the PSS injector liner with quartz wool. Step 4. Reinstall the liner in the PSS injector.

92 . Clarus 690 GC Customer Hardware and Service Guide Step 5. Connect a column to the PSS injector. Step 1. Remove the PSS Injector Liner. To remove the PSS injector liner: 1. Ensure that the injector heater has been turned off. Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to too low a temperature (less than 80 °C) will make it difficult to remove the injector liner. 2. Remove the septum cap.

Septum Cap (P/N N6100153)

Figure 23. Removing the septum cap. 3. Remove the injector cover.

Figure 24. Removing the injector cover. 4. Loosen the threaded collar by using the spanner (P/N N6101359) provided, then remove the threaded collar.

Step 2: Connect the Column to the Injector . 93

Spanner (P/N N6101359)

Figure 25. Loosening the threaded collar. 5. Replace the septum cap on the injector. 6. Pull the septum cap upwards to remove the septum purge assembly.

Figure 26. Removing the septum purge assembly. The carrier gas inlet line is coiled to allow you to pull the septum purge assembly over to the side and gain access to the liner. 7. Ensure that the PSS injector liner is cool, then insert the end of the PSS liner-removal tool (P/N 02506247) over the end of the injector liner that is shipped with PSS injector. Remove the liner by lifting it up and out of the injector. The PSS liner must be cool (no hotter than 100 °C) or the liner-removal tool will melt! The end of the PSS liner-removal tool may flare out with use. If this happens, cut off the flared end with a razor blade or scissors.

94 . Clarus 690 GC Customer Hardware and Service Guide The PSS injector does not have a removable fitting at the bottom of the assembly. Be very careful when removing this liner to prevent breaking it.

CAUTION ATTENTION

As the injector cools, the O-ring adheres to the metal base of the injector body. Use a small screwdriver to loosen the O-ring, then remove the liner and O-ring. Be careful not to scratch the barrel where the O-ring seals. Replace the O-ring with a new O-ring when you replace the liner. L'injecteur PSS n'a pas de montage amovible au bas de l'assemblage. Faites très attention lorsque vous retirez cette doublure pour éviter de la casser. Lorsque l'injecteur refroidit, le joint torique adhère à la base métallique du corps de l'injecteur. Utilisez un petit tournevis pour desserrer le joint torique, puis retirez le doublure et le joint torique. Veillez à ne pas rayer le baril où le joint torique obture. Remplacez le joint torique par un nouveau joint torique lorsque vous remplacez le revêtement.

PSS Liner-Removal Tool (P/N 02506247)

Figure 27. Removing a PSS injector liner. PSS Injector O-Rings

Recommended Maximum Temperature

N6101751 Graphite (pkg. of 5)

450 °C

09921004 Kalrez (pkg. of 1)

450 °C

N6101747 Viton (pkg. of 10)

250 °C (not recommended for use with ECD) Step 2. Select an Appropriate PSS Injector Liner.

The PSS injector uses the following three liners: 

2-mm i.d. PSS injector wide-bore liner (P/N N6121004)



1-mm i.d. PSS injector narrow-bore liner (P/N N6121006)



PSS injector on-column (hourglass) liner (P/N N6101539)

Step 2: Connect the Column to the Injector . 95 The PSS injector is operated in the inlet-programmed mode for split or splitless injection with either the 2-mm or 1-mm i.d. liner. If used in the “hot” split or splitless mode, the 2-mm i.d. liner should be used. It is recommended that you do not inject more than approximately 0.5 µL in the “Hot” mode. Refer to Table 6-1. For PSS on-column operation, use the hourglass liner and the oven-program mode. The 2-mm i.d. PSS liner that is used for either split or splitless operation should be packed with quartz wool as described in this chapter. The 1-mm i.d. PSS liner may give better early-eluting peak resolution in the split or splitless mode. This liner should be used for samples with early-eluting peaks that cannot be resolved by additional solute trapping/focusing (by lowering the initial oven temperature or by using a column with a thicker stationary-phase film). Step 3. Pack the PSS Injector Liner with Quartz Wool.

N ever pack the hourglass liner with wool. N e jam ais emballer la doublure de sablier avec de la laine.

CAUTION ATTENTION

To wipe the syringe needle, we recommend packing a small amount of quartz wool in the top portion of the liner regardless of the liner type (wide-bore or narrow-bore but never the hourglass) or injector mode (split or splitless). Quartz wool assures that reproducible volumes are injected by wiping the syringe needle every time it is inserted. Remove the liner and replace the quartz wool on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest, tailing, and loss of sensitivity. Remove the wool with a small hook on the end of a thin wire, or blow it out using compressed air. SPLITLESS PACKING

SPLIT PACKING

ON-COLUMN O-Ring

O-Ring

O-Rings

Quartz Wool Loosely Packed

Quartz Wool Loosely Packed Quartz Wool Tightly Packed Dimple

Dimple

2-mm i.d. Wide-Bore Liner (P/N N6121004)

1-mm i.d. Narrow-Bore Liner (P/N N6121006)

2-mm i.d. Wide-Bore Liner (P/N N6121004)

Figure 28. PSS liners packed with quartz wool.

Hourglass Liner (P/N N6101539)

96 . Clarus 690 GC Customer Hardware and Service Guide Packing a PSS Injector Liner for the Splitless Mode

CAUTION ATTENTION

To prevent contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable gloves (for example, the same type of gloves used to perform maintenance on Clarus SQ 8). Pour éviter de contaminer la laine de quartz lors de l'emballage de la doublure d'injection, porter des gants jetables en vinyle, sans poudre (par exemple, le même type de gants utilisé pour la maintenance sur Clarus SQ 8).

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch o.d. rod (P/N N610T100), push the quartz wool into the liner. Pack a 2.5-cm (one-inch) piece of quartz wool loosely below the top ground portion of the liner (see Figure 29). The sample is then injected into the wool, thereby preventing the delivery of sample beyond the column. The wool also wipes the syringe needle upon injection.

N OTE: The narrow-bore liner is more difficult to pack because of its small i.d. However, there is a dimple in the middle of the liner to hold the wool in place. Do not pack the wool too tightly! When the narrow-bore liner is installed, use a 5-cm needle-length syringe when making manual injections. Packing PSS Injector Liner for the Split Mode CAUTION ATTENTION

To prevent contaminating the quartz wool when packing the injection liner, wear vinyl, powder-free, disposable gloves. Pour éviter de contaminer la laine de quartz lors de l'emballage de la doublure d'injection, porter des gants jetables en vinyle, sans poudre.

Take a small piece of quartz wool and twist it into an elongated shape so that you can insert it into the liner. Then using a 1/16-inch o.d. rod (P/N N610T100), push the quartz wool into the liner. Pack the wool tightly 2 from the dimple upwards (about one inch [2.5 cm]). Loosely pack quartz wool in the top portion of the liner to wipe the syringe needle upon injection.

N OTE: As you can see in Figure 29, each liner has an O-ring installed on the part furthest away from the dimple on the PSS injector liner. If the O-ring has adhered to the liner, you may not be able to easily remove the liner. Use a small screwdriver to dislodge the O-ring before removing the liner and O-ring. Be careful not to scratch the barrel where the O-ring seals. Step 4: Reinstall the Liner in the PSS Injector. To reinstall the liner: 1. Install a new O-ring on the top portion of the liner. 2. Insert the liner in the injector body. 3. Place the septum purge assembly over the liner. 4. Press the septum purge assembly down to correctly position the liner in the injector. Make sure that you tightly secure the septum purge assembly to the injector base with the spanner. 2

The recovery of high molecular weight components (e.g., C40) may be improved if the liner is loosely packed.

Step 2: Connect the Column to the Injector . 97 Step 5A. Connect a Column to the PSS Injector. To connect a column: 1. Insert a 1/16-inch column nut (P/N 09903392) and 1/16-inch graphite ferrule (0.8 mm i.d., P/N 09920141 or 0.5 mm i.d., P/N 09903700) over one end of the column as shown below:

CAUTION ATTENTION

Figure 29. column.

This injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut. Cet injecteur se termine par un raccord de 1/16 de pouce. Ce raccord est fragile. Pour préserver l'intégrité du raccord, connectez soigneusement l'écrou pour empêcher le filetage croisé du raccord et / ou le serrage excessif de l'écrou sur le raccord. Vous pouvez également préserver l'intégrité du raccord en permettant à l'injecteur de refroidir avant de brancher un écrou.

Column, nut, and ferrule on the injector end of a narrow-bore capillary

N OTE: Make certain that the tapered end of the ferrule (in Figure 30) is facing towards the nut. 2. Cut off about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N9301376, pkg. of 10 scribes). Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to Figure 31.

Figure 30.

Example of a good cut and bad cuts.

98 . Clarus 690 GC Customer Hardware and Service Guide 3. Position the column nut on the column so that the back of the nut is 3.8 cm to 4.4 cm (1 1/2 to 1 3/4 inches) from the end of the column. 4. Using typewriter "white-out" or a felt tipped pen, make a mark on the column just beyond the back edge of the column nut (see Figure 30).

To avoid contaminating the system, make certain that the nut and ferrule do not come into contact with the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne sont pas en contact avec la marque sur la colonne.

CAUTION ATTENTION

PSS Injector Fitting 1/16" Column Nut Capillary Column

Figure 31. Step 5B.

Capillary column attached to PSS injector fitting. Connect the Column to the PSS Injector in the On-Column Mode Using the Hourglass Liner

The on-column mode is not recommended for use with columns having an internal diameter of 0.1 mm. This is due to the flow differences between the pre-column (0.53 mm i.d.) and the column. The following three steps summarize how to connect a column to the PSS injector in the oncolumn mode: Step 1. Remove the PSS split/splitless injector liner. Step 2. Install the hourglass liner in the PSS injector. Step 3. Connect a column to the PSS injector. Step 1. Remove the PSS Split/Splitless Injector Liner. To remove the PSS injector liner: 1. Ensure that the injector heater has been turned off. Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to too low a temperature (less than 80 °C) will make it difficult to remove the injector liner. 2. Remove the septum cap.

Step 2: Connect the Column to the Injector . 99

Septum Cap (P/N N6100153)

Figure 32. Removing the septum cap. 3. Remove the injector cover.

Figure 33. Removing the injector cover. 4. Loosen the threaded collar using the spanner (P/N N6101359) provided, then remove the threaded collar.

Spanner (P/N N6101359)

Figure 34. Loosening the threaded collar. 5. Replace the septum cap on the injector. 6. Pull the septum cap upwards to remove the septum purge assembly.

100 . Clarus 690 GC Customer Hardware and Service Guide

Figure 35. Removing the septum purge assembly. The carrier gas inlet line is coiled to allow you to pull the septum purge assembly over to the side and gain access to the liner. 7. Ensure that the liner is cool, then insert the end of the PSS liner-removal tool (P/N 02506247) over the end of the wide-bore or narrow-bore PSS injector liner. Remove the liner by lifting it up. Gently probe the O-ring if it sticks to the injector body. The PSS liner must be cool (no hotter than 100 °C) or the tubing will melt! The end of the PSS liner-removal tool may flare out with use. If this happens, cut off the flared end with a razor blade or scissors.

PSS Liner-Removal Tool (P/N 02506247)

Figure 36. Removing an injector liner. Step 2. Install the Hourglass Liner in the PSS Injector. To install the hourglass liner: 1. Replace the existing liner with the hourglass liner. Install the liner with the hourglass end closest to the septum and use a new O-ring.

Step 2: Connect the Column to the Injector . 101

Figure 37. Hourglass liner (P/N N6101539). 2. Replace the septum purge assembly removed previously. DO NOT secure it with the collar at this time. We recommend the following procedure to connect a column and pre-column to the universal connector. You may find that changing the order of the steps is more convenient for you. However, the critical concerns are: •

Making straight, clean, even cuts on the column/pre-column.



Wetting the ends of the column/pre-column.



Wetting the fused-silica universal connector/hourglass needle guide.



GENTLY twisting the column/pre-column into the hourglass fused-silica universal connector.



Conditioning the mechanical joint (this is critical).

In the on-column mode, you will install a column (0.53 mm i.d.) or a deactivated fused silica precolumn (0.53 mm i.d.) in the injector body. This will enable the sample to be deposited directly into the column or the pre-column. Step 3. Connect a Column to the PSS Injector To connect a column to the PSS in the on-column mode using the hourglass liner: 1. Insert a 1/16-inch column nut (P/N 09903392) and 1/16-inch graphite ferrule (P/N 09920141) over the other end of the 0.53-mm i.d. column or pre-column as shown below:

102 . Clarus 690 GC Customer Hardware and Service Guide

CAUTION ATTENTION

If you are using a 0.53-mm i.d. column, insert it directly into the injector. If you are using a column with an i.d. less than 0.53 mm, insert a one-meter piece of deactivated 0.53-mm i.d. fused silica (pre-column) into the injector. This precolumn will also serve as a retention gap. In addition, the long length of this tubing will enable you to change columns and still have enough of the pre-column to make numerous new column connections. Connect a column to the end of this pre-column (in the oven) with a universal connector (P/N N9302149). Si vous utilisez un i.d. 0,53 mm. Colonne, insérez-la directement dans l'injecteur. Si vous utilisez une colonne avec un i.d. Moins de 0,53 mm, insérez un morceau d'un mètre de 0.5.-i.d. Silice fondue (pré-colonne) dans l'injecteur. Cette pré-colonne servira également d'écart de conservation. De plus, la longue longueur de ce tube vous permettra de modifier les colonnes et d'avoir suffisamment de pré-colonnes pour créer de nombreuses nouvelles connexions de colonnes. Connectez une colonne à la fin de cette pré-colonne (au four) avec un connecteur universel (P/N N9302149).

Figure 38. Column, Nut, and Ferrule on the injector end of a narrow-bore capillary column.

CAUTION ATTENTION

Ensure that the tapered end of the ferrule is facing towards the nut as shown above. We have found that graphite/Vespel ferrules used to connect the column to the injector fitting loosen after the injector temperature cycles several times. Use graphite ferrules to eliminate this problem. Assurez-vous que l'extrémité conique de la virole est tournée vers l'écrou comme indiqué ci-dessus. Nous avons constaté que les ferrilles de graphite / Vespel utilisées pour relier la colonne au raccord de l'injecteur se desserrent après plusieurs cycles de température de l'injecteur. Utilisez des ferrilles en graphite pour éliminer ce problème.

2. Cut off about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N9301376, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark so that the break is clean and square. If you are using a 0.53-mm i.d. pre-

Step 2: Connect the Column to the Injector . 103 column, make a clean square cut on both ends. Examine the cut with a magnifying glass and compare it to the following figure:

CAUTION ATTENTION

A clean, square cut is especially critical to produce a leak-free seal in the fused-silica universal connector. Une coupe propre et carrée est particulièrement critique pour produire un joint d'étanchéité sans fuite dans le connecteur universel à base de silice fondue.

Figure 39. Example of a good cut and bad cuts. 3. Locate the PSS injector fitting inside the oven.

PSS Injector Fitting 1/16" Column Nut Capillary Column

Figure 40. PSS injector fitting in the column oven. 4. Insert the 0.53-mm i.d. column or pre-column into the PSS injector fitting. Tighten the column nut so the column/pre-column can still be moved but with some resistance. 5. Push the column/pre-column up into the injector until it can not go any further. You may notice the septum purge assembly move upward. 6. Mark the column with typewriter "white out" just below the column nut. 7. Push the septum purge assembly down into position, place the threaded collar on the assembly, and tighten the threaded collar with the spanner provided.

104 . Clarus 690 GC Customer Hardware and Service Guide

Figure 41. Replacing the septum purge assembly. 8. Using a 1/4-inch wrench, tighten the fitting until the column cannot be pulled out of the fitting.

CAUTION ATTENTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

9. Replace the septum cap but do not tighten it. Manually insert the 0.47-mm o.d. syringe needle into the 0.53-mm i.d. column/pre-column. With the syringe still inserted in the injector, tighten the septum cap completely. This aligns the column and septum cap to ensure a smooth movement of the syringe through the septum cap and into the 0.53-mm i.d. column/pre-column. 10. If your column i.d. is smaller than 0.53 mm, connect your column to the 0.53-mm i.d. pre-column as follows: a)

Make a clean, even cut on the end of your column and examine it with a magnifying glass.

b)

Drop some solvent (for example, methanol) into both sides of the fused-silica universal connector (P/N N9302149, package of 5).

c)

Wet a tissue with solvent and wipe the end of the column.

d)

Insert the column end into the universal connector and gently twist it. The column should seal in the universal connector.

11. Wet a tissue with solvent (for example, methanol) and wipe the end of the pre-column. 12. Carefully push the universal connector, with the attached column, up into the end of the pre-column. Gently twist the universal connector to seal the connection between precolumn and universal connector. This connection is called the mechanical joint.

Step 2: Connect the Column to the Injector . 105

POC Injector Fitting 1/16-inch Column Nut Pre-Column (0.53mm o.d.) Universal Connector (P/N N9302149) Column

Figure 42. Connecting a column and universal connector to the pre-column to create the mechanical joint.

Typically the split vent is closed in this mode of operation. Condition the mechanical joint between the pre-column and column prior to making any analytical runs. Condition the joint by slowly temperature programming the oven up to 200 °C and holding it at 200 °C for one hour. Do not exceed the recommended temperature of your column. This can be incorporated when you condition the column, see Step 5 in this chapter. CAUTION

If you replace the column, do not reuse the universal connector. Cut the column and pre-column from the universal connector. Replace the universal connector with a new one. The pre-column ought to be long enough so that you can reuse it. If you replace the pre-column, it will also require some conditioning. If the entire installation is new, the column conditioning procedure will condition the pre-column, the column, and the mechanical joint. If you are just replacing the pre-column, you can condition it and the mechanical joint by the above procedure. If a high background persists with a fresh pre-column, condition the pre-column by only increasing the injector temperature for one hour.

106 . Clarus 690 GC Customer Hardware and Service Guide En règle générale, le ventilateur divisé est fermé dans ce mode de fonctionnement. Conditionnez l'articulation mécanique entre la pré-colonne et la colonne avant de réaliser des essais analytiques. Conditionnez l'articulation en programmant lentement le four jusqu'à 200 ° C et en le tenant à 200 °C pendant une heure. Ne pas dépasser la température recommandée de votre colonne. Cela peut être incorporé lorsque vous conditionnez la colonne, voir l'étape 5 de ce chapitre. ATTENTION

Si vous remplacez la colonne, ne réutilisez pas le connecteur universel. Couper la colonne et la pré-colonne à partir du connecteur universel. Remplacez le connecteur universel par un nouveau. La pré-colonne doit être suffisamment longue pour pouvoir la réutiliser. Si vous remplacez la pré-colonne, cela nécessitera également un certain conditionnement. Si l'installation complète est nouvelle, la procédure de conditionnement de la colonne conditionnera la précolonne, la colonne et l'articulation mécanique. Si vous remplacez simplement la pré-colonne, vous pouvez la conditionner et l'articulation mécanique selon la procédure ci-dessus. Si un fond élevé persiste avec une pré-colonne fraîche, conditionnez la pré-colonne en augmentant seulement la température de l'injecteur pendant une heure.

CAUTION ATTENTION

When using the PSS injector in the on-column mode or with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N6101253 or N6101380). Refer to Chapter 9 for more details. You must use only the "Norm" injection speed with this syringe when you are in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break up or distorted peaks. You can achieve better precision in the on-column mode when sample volumes of 1.0 µL or greater are injected. Lorsque vous utilisez l'injecteur PSS en mode colonne ou avec l'automateur, vous devez utiliser une seringue spéciale qui a une aiguille o.d. De 0,47 mm (P/N N6101253 ou N6101380). Reportezvous au Chapitre 9 pour plus de détails. Vous devez utiliser uniquement la vitesse d'injection "Norm" avec cette seringue lorsque vous êtes en mode colonne. La vitesse d'injection "rapide" plie cette fine aiguille et la vitesse d'injection "lente" peut produire des pics de pointe ou des pics déformés. Vous pouvez obtenir une meilleure précision dans le mode en colonne lorsque des volumes d'échantillons de 1,0 μL ou plus sont injectés.

Step 2C Connect the Colum n to the P rogram m ed On-Colum n I njector (P OC): The Programmed On-Column injector (POC) consists of an hourglass adapter, a deactivated 0.53mm i.d. fused-silica pre-column, and a fused-silica universal connector. The sample is injected into the POC at a “cool" temperature; the injector temperature is then programmed to increase.

Step 2: Connect the Column to the Injector . 107 This is helpful for samples that are thermally labile and/or of a wide molecular weight range. This injector is used only for trace analysis or diluted solutions.

CAUTION ATTENTION

Only use a syringe that has a 0.47-mm o.d. needle (P/N N6101380) with this injector. Utilisez uniquement une seringue ayant un 0.47 mm o.d. Aiguille (P/N N6101380) avec cet injecteur.

Figure 43. Cutaway view of a Programmed On-Column injector (POC).

About the Pneum atics Control The POC pneumatics consist of PPC (programmable pneumatics controlled) modules that regulate the inlet flow and pressure of the gases using software controlled by an Clarus 690 GC method.

108 . Clarus 690 GC Customer Hardware and Service Guide

CAUTION ATTENTION

This injector terminates in a 1/16-inch fitting. When connecting a nut to this fitting, take special care not to crossthread the fitting or overtighten the nut. Cet injecteur se termine par un raccord de 1/16 de pouce. Lorsque vous raccordez un écrou à ce raccord, prenez soin de ne pas croiser le raccord ou trop serrer l'écrou.

On-column injection is not recommended for columns with an internal diameter of 0.1 mm. This is because of flow differences between the pre-column (0.53 mm) and the column. We recommend that you use the following procedure to connect a column and pre-column with the universal connector. You may find that changing the order of the steps is more convenient for you. However, the critical concerns are: 

Making straight, clean, even cuts on the column/pre-column.



Wetting the ends of the column/pre-column with solvent.



Wetting the universal connector/hourglass needle guide with solvent.



GENTLY twisting the column/pre-column into the hourglass universal connector.



Conditioning the mechanical joint (this is critical).

To connect a column to the programmed on-column injector (POC): 1. Remove the septum cap.

Figure 44. Removing a septum cap. 2. Remove the septum shield (P/N N6101702) with the large end of the liner-removal tool (P/N N6100102).

Figure 45. Liner-removal tool (P/N N6100102). 3. Remove the hourglass needle guide (P/N N6101703) with a pair of small pliers or tweezers.

Step 2: Connect the Column to the Injector . 109

Figure 46. Hourglass needle guide.

CAUTION ATTENTION

If you are using a 0.53-mm i.d. column, insert it directly into the injector. If you are using a column with an i.d. of less than 0.53 mm, insert a one-meter piece of deactivated 0.53-mm i.d. fused silica into the injector to act as a pre-column. This pre-column will also serve as a retention gap. In addition, the long length of this tubing will enable you to change columns and still have enough of the pre-column to make numerous new column connections. Connect a column to the end of this pre-column (in the oven) with a universal connector (P/N N9302149). Si vous utilisez un i.d. 0,53 mm. Colonne, insérez-la directement dans l'injecteur. Si vous utilisez une colonne avec un i.d. De moins de 0,53 mm, insérer un morceau d'un mètre de 0.5.-i.d. La silice fondue dans l'injecteur pour agir comme pré-colonne. Cette pré-colonne servira également d'écart de conservation. De plus, la longue longueur de ce tube vous permettra de modifier les colonnes et d'avoir suffisamment de pré-colonnes pour créer de nombreuses nouvelles connexions de colonnes. Connectez une colonne à la fin de cette pré-colonne (au four) avec un connecteur universel (P/N N9302149).

4. Insert the column nut (P/N 09903392) and ferrule (P/N 09920141) on the end of the column/pre-column.

Figure 47. Column, nut, and ferrule on the injector end of a narrow-bore capillary column.

110 . Clarus 690 GC Customer Hardware and Service Guide

CAUTION ATTENTION

We have found that graphite/Vespel ferrules used to connect the column to the injector fitting loosen after the injector temperature cycles several times. Use graphite ferrules to eliminate this problem. Make certain that the tapered end of the ferrule is facing towards the nut as shown above. Nous avons constaté que les ferrilles de graphite / Vespel utilisées pour relier la colonne au raccord de l'injecteur se desserrent après plusieurs cycles de température de l'injecteur. Utilisez des ferrilles en graphite pour éliminer ce problème. Assurez-vous que l'extrémité conique de la virole est tournée vers l'écrou comme indiqué ci-dessus.

5. Make a clean, square cut on the end of the 0.53-mm column or on both ends of the 0.53-mm pre-column. Wipe the end with a tissue soaked in methanol to remove fragments of polyimide or silica.

CAUTION ATTENTION

A clean, square cut is especially critical to obtain a leak-free seal in the hourglass needle guide and fused-silica universal connector. Une coupe propre et carrée est particulièrement critique pour obtenir un joint étanche aux fuites dans le guide de l'aiguille du sablier et le connecteur universel à base de silice fondue.

Figure 48. Example of a clean cut and bad cuts. 6. Locate the injector fitting in the oven. Then insert the column/pre-column until it protrudes out the septum end of the injector. 7. Thread the nut onto the injector fitting. Then tighten it so that the column/pre-column can still be moved, but you feel some resistance. 8. Push the column/pre-column into the hourglass needle guide and twist it slightly. 9. Place the metal septum shield over the hourglass needle guide and push it into the injector until it is completely seated. Replace the septum cap. 10. Tighten the column nut inside the oven until the column cannot be pulled out of the fitting.

Step 2: Connect the Column to the Injector . 111

CAUTION ATTENTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

You may want to mark the column, just behind the nut, with typewriter "white-out." The mark will act as a guide when you test to see if the fitting is tight enough.

CAUTION ATTENTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne contactent pas la marque sur la colonne.

11. Loosen the septum cap and manually insert the 0.47-mm o.d. syringe needle into the column/pre-column. With the syringe in the injector, tighten the septum cap completely. This ensures smooth movement of the syringe through the septum cap and into the column/pre-column. 12. If your column i.d. is smaller than 0.53 mm, connect your column to the 0.53-mm i.d. pre-column as follows: a) Make a clean, even cut on the end of your column and examine it with a magnifying glass. b) Drop some solvent (for example, methanol) into both sides of the fused-silica universal connector (P/N N9302149, package of 5). c) Wet a tissue with solvent and wipe the end of the column. d) Insert the column end into the universal connector and gently twist it. The column should seal in the universal connector. 13. Wet a tissue with solvent (for example, methanol) and wipe the end of the pre-column. 14. Carefully push the universal connector, with the attached column, up into the end of the pre-column. Gently twist the universal connector to seal the connection between the pre-column and universal connector. This connection is called the mechanical joint.

POC Injector Fitting 1/16-inch Column Nut Pre-Column (0.53mm o.d.) Universal Connector (P/N N9302149) Column

112 . Clarus 690 GC Customer Hardware and Service Guide Figure 49. Connecting the column and universal connector to the pre-column to create the mechanical joint.

Condition the mechanical joint between the pre-column and column prior to making any analytical runs. Condition the joint by slowly temperature programming the oven up to 200 °C and holding it at 200 °C for one hour. Do not exceed the recommended temperature of your column. This can be incorporated when you condition the column, see Step 5 in this Chapter.

CAUTION

If you replace the column, do not reuse the universal connector. Cut the column and pre-column from the universal connector. Replace the universal connector with a new one. The pre-column ought to be long enough so that you can reuse it. If you replace the pre-column, it will also require some conditioning. If the entire installation is new, the column conditioning procedure will condition the pre-column, the column, and the mechanical joint. If you are just replacing the pre-column, you can condition it and the mechanical joint by the above procedure. If a high background persists with a fresh pre-column, condition the pre-column by only increasing the injector temperature for one hour

Conditionnez l'articulation mécanique entre la pré-colonne et la colonne avant de réaliser des essais analytiques. Conditionnez l'articulation en programmant lentement le four jusqu'à 200 ° C et en le tenant à 200 °C pendant une heure. Ne pas dépasser la température recommandée de votre colonne. Cela peut être incorporé lorsque vous conditionnez la colonne, voir l'étape 5 de ce chapitre.

ATTENTION

Si vous remplacez la colonne, ne réutilisez pas le connecteur universel. Couper la colonne et la pré-colonne à partir du connecteur universel. Remplacez le connecteur universel par un nouveau. La pré-colonne doit être suffisamment longue pour pouvoir la réutiliser. Si vous remplacez la pré-colonne, cela nécessitera également un certain conditionnement. Si l'installation complète est nouvelle, la procédure de conditionnement de la colonne conditionnera la précolonne, la colonne et l'articulation mécanique. Si vous remplacez simplement la pré-colonne, vous pouvez la conditionner et l'articulation mécanique selon la procédure ci-dessus. Si un fond élevé persiste avec une pré-colonne fraîche, conditionnez la précolonne en augmentant seulement la température de l'injecteur pendant une heure.

Step 2: Connect the Column to the Injector . 113

CAUTION ATTENTION

When using the POC with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N6101253 or N6101380). Refer to Chapter 9 for more detail. You must only use the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break up or distorted peaks. You can achieve better precision in the on-column mode when injecting sample volumes of 1.0 mL or greater. Lorsque vous utilisez le POC avec l'auto-échantillonneur, vous devez utiliser une seringue spéciale qui a une aiguille o.d. De 0,47 mm (P/N N6101253 ou N6101380). Reportez-vous au chapitre 9 pour plus de détails. Vous devez uniquement utiliser la vitesse d'injection "Norm" avec cette seringue en mode colonne. La vitesse d'injection "rapide" plie cette fine aiguille et la vitesse d'injection "lente" peut produire des pics de pointe ou des pics déformés. Vous pouvez obtenir une meilleure précision en mode colonne lors de l'injection de volumes d'échantillons de 1,0 mL ou plus.

114 . Clarus 690 GC Customer Hardware and Service Guide

Step 3: Set the Carrier Gas This step describes how to set the carrier gas for PPC modules. The following procedure describes how to: 

Set the Carrier Gas Pressure for the Split/Splitless Injector (CAP) and Programmed Split/Splitless Injector (PSS)



Set the Carrier Gas Flow for the Programmed On-Column Injector (POC)

Setting the Carrier Gas Pressure for the Split/ Splitless I njector (CAP) and Program m ed Split/ Splitless I njector (PSS) N OTE: The line pressure should be at least 15 psi greater than the highest carrier gas pressure setting, up to a maximum of 100 psig. To set the carrier gas pressure: 1. Turn on the carrier gas at the tank. Then adjust the line pressure to 60 - 90 psig. 2. Since a capillary injector is in position A, and Capillary Control is selected the Carrier Control drop down menu in the Configuration screen allows you to select Pressure, Flow, or Velocity. The following screen show Pressure selected.

3. From the System Status screen select either A or B injector (the example here shows the A PSS injector ready for setting the carrier gas). In this example, touch the Channel

A PSS injector icon

.

Step 3: Set the Carrier Gas . 115

4. In the following window touch the Program button under the Carrier Gas field. 5. Touch the program button in the Carrier Gas section and the following screen will be displayed.

6. Using the up and down arrow or keypad

buttons enter an Initial pressure

(0.1 to 75 psig). The following screen shows 2.0 psi.

116 . Clarus 690 GC Customer Hardware and Service Guide 7. Using the up and down arrow or keypad

buttons enter the appropriate

values. Once entered select Close to exit.

N OTE: The pressure readout display is factory configured to display the actual pressure as psig. 8. Enter a pressure set point. (See Suggested Capillary Column Pressures on Table 6-5.).

Setting the Carrier Gas Flow for the Program m ed On-Colum n I njector (POC) To set the carrier gas flow: 1. Turn on the carrier gas at the tank. Adjust the line pressure to 60 - 90 psig. 2. From the System Status screen select either A or B injector (the example here shows the B injector ready for setting the carrier gas).

3. In the following window touch the Program button under the Carrier Gas field.

Step 3: Set the Carrier Gas . 117

4. Using the up and down arrow or keypad

buttons enter a pressure (0.1 to

100 psig).

N OTE: If you wish to generate a flow program method, refer to the Clarus 690 Users Guide (P/N 09931374).

118 . Clarus 690 GC Customer Hardware and Service Guide

Suggested Capillary Colum n P ressures The following tables are applicable to PPC pneumatics. Table 6-2. Calculated Pressure Drops (psig) for 10m Column 3 _

u

Column I.D. (µm) 4

320

250

100

10 20 30 40 60 80

1.0 2.1 3.1 4.1 6.2 8.3

2.4 4.9 7.3 9.8 14.6 19.5

10.0 21.2 31.8 42.3 63.5 84.7

Table 6-3. Calculated Pressure Drops (psig) for 25m Columns4 _

Column I.D. (µm)

u5

320

250

100

10 20 30 40 60 80

2.6 5.2 7.8 10.3 15.5 20.7

6.1 12.2 18.3 24.4 36.6 48.8

26.5 52.9 79.4 -

Table 6-4. Calculated Pressure Drops (psig) for 50m Columns4 Column I.D. (µm) _ u5 320 250 100 10 20 30 40 60 80

5.2 10.3 15.5 20.7 31.0 41.3

12.2 24.4 36.6 48.8 73.2 -

52.9 -

Setting the Carrier Gas Flow for the Program m ed On-Colum n I njector (POC) Using the Optional Flow R eadout To set the carrier gas flow with the optional flow readout: 1. Turn on the carrier gas at the tank. Adjust the line pressure to 90 psig. 2. From the System Status screen select either A or B injector (the example here shows the B injector ready for setting the carrier gas).

3 4

In psig, using helium as a carrier gas at 100 °C. Average linear velocity (cm/sec).

Step 3: Set the Carrier Gas . 119

3. In the following window touch the Program screen under the Carrier Gas field.

4. Using the up and down arrow or keypad 100 psig).

5. Type the desired flow set-point value.

buttons enter a pressure (0.1 to

120 . Clarus 690 GC Customer Hardware and Service Guide

Set the Carrier Gas Flow for the Program m ed On-Colum n I njector (POC) Using the Soap Bubble Flow m eter To perform this procedure, you must know how to measure carrier gas flow using a soap bubble flowmeter and the Clarus 690 GC built-in stopwatch. To set the carrier gas flow using the soap bubble flowmeter: 1. Locate the POC injector fitting inside the column oven. The POC injector fitting is shown below:

Figure 50. POC injector fitting in the column oven. 2. Attach a soap bubble flowmeter to the POC injector fitting. 3. Turn on the carrier gas at the tank. Adjust the line pressure to 90 psig. 4. From the System Status Screen select the Tools button and then select Utilities.

5. In the Utilities window select the Stopwatch icon.

Step 3: Set the Carrier Gas . 121

6. Using the up and down arrow or keypad

buttons enter the appropriate

values and press Start.

7. Start the carrier gas flowing by turning the flow controller knob counterclockwise and measure the flow.

N OTE: For best accuracy, use a soap bubble flowmeter volume that gives a reading of at least 30 seconds. 8. Adjust the flow to the desired set point by repeatedly measuring the flow and turning the flow controller knob counterclockwise to increase the flow, clockwise to decrease the flow, until the desired flow is obtained. 9. Disconnect the soap bubble flowmeter before proceeding to the next step to Step D “Leak Test all New Connections.” Table 6-5. Suggested Capillary Column Flow Rates Using Helium As Carrier Column i.d. (µm) Flow mL/min

250

320

530

0.6 - 0.8

1.0 - 1.5

2.5 - 4.0

122 . Clarus 690 GC Customer Hardware and Service Guide

Step 4 Leak Test All New Connections: P P C P neum atics (P OC I njector Only) 1. In the following example select the POC from the System Status screen.

2. In the next window select the Program icon.

3. In the program screen enter a minimum value between 0.1 and 100 psig. Enter a maximum value between 0.1 and 100 psig.

Step 4 Leak Test All New Connections: . 123

The Clarus 690 GC monitors the inlet column head pressure with a carrier-gas mass-flow controller in use. You enter the minimum and maximum values after selecting the carriergas Flow mode in configuration. If the inlet column head pressure drops below or rises above the set limits for more than one minute, a warning message appears on the display. The following error message will appear: Carrier (with specific carrier number) unable to maintain pressure. You should correct the leak; the most common area would be the injector septum. Then clear the message by touching OK before continuing to use the GC. This will stop an autosampler program from continuing. If you do not make any entries, leak checking will not be done. If you are temperature programming, then the value you enter for the maximum must be above the column head pressure at the highest oven temperature. You can just enter a value for the minimum and leave the maximum off (zero).

124 . Clarus 690 GC Customer Hardware and Service Guide

Step 5: Condition the Colum n and the M echanical Joint Betw een the P re-colum n and Colum n: This section contains a suggested temperature program for conditioning a column. The program starts off by holding the oven temperature at a medium value for 10 minutes, gradually increasing the oven temperature at a fixed rate (5 °C/min) to the column operating temperature, then holding that temperature overnight with the carrier gas flowing.

CAUTION ATTENTION

The temperatures shown in the examples which follow should be used as guidelines. Please refer to the column manufacturer's operating instructions for specific temperature recommendations. Les températures indiquées dans les exemples qui suivent doivent être utilisées comme lignes directrices. Reportezvous au mode d'emploi du fabricant de la colonne pour des recommandations de température spécifiques.

CAUTION ATTENTION

To keep the injector clean, open the split vent to direct more gas through the injector. Pour garder l'injecteur propre, ouvrez l'évent de rupture pour diriger plus de gaz à travers l'injecteur.

To condition the column: 1. Close the oven door and from the System Status screen press the Oven icon.

2. In the oven screen enter an oven temperature set point of 50 °C and enter a (Hold) time of 10.

Step 5: Condition the Column and the Mechanical Joint Between the Pre-column and Column: . 125

3. To add another program step, enter a RATE of 5 (ºC/min). 4. For TEMP 2, enter a set point 25 to 50 ºC above your planned analytical operating temperature. For example, enter a set point of 200.

CAUTION ATTENTION

To avoid damaging the column, do not enter a temperature higher than the maximum recommended temperature specified by the column manufacturer. Pour éviter d'endommager la colonne, n'entrez pas une température supérieure à la température maximale recommandée spécifiée par le fabricant de la colonne.

5. Configure the injector for the oven mode. From the System Status screen touch the Tools button and select the Configuration menu

. 6. In the Configuration menu select the injector icon. In the following example injector B has been selected to be configured.

126 . Clarus 690 GC Customer Hardware and Service Guide

7. Select the Program button. Touch the OK button to close this screen then touch the Close button to close the Configuration screen. 8. Touch the A detector button (for example FID) and turn the Detector Temperature off. Allow the system to run overnight.

9. In the morning, reset the oven the oven temperature set point to the initial TEMP at the beginning of the temperature program. See step 4 earlier in this procedure. 10. Open the oven door. Allow the oven to cool until the oven fan turns off. This occurs when the oven cools down to 40 °C.

Step 6: Connect the Column to the Detector . 127

Step 6: Connect the Colum n to the Detector 1. Place the column over the hanger so that no part of the column touches the bottom or sides of the oven. 2. For a FID, insert a 1/16-inch column nut (P/N 09903392) and 1/16-inch graphite ferrule (P/N 09920141) over the other end of the 0.53-mm i.d. column or pre-column as shown below:

Figure 51. Column, Nut, and Ferrule on the injector end of a narrow-bore capillary column. For a PID, insert a 1/8-inch column nut (P/N 09903453) and 1/8-inch graphite ferrule (for 0.32/0.25-mm i.d. columns (P/N 09903981) or for 0.53-mm i.d. columns (P/N 09903394)) over the free end of the column as shown below:

Figure 52. Nut and ferrule on the detector end of a narrow-bore capillary column. 3. Cut about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N9301376, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 53. Example of a good cut and bad cuts. 4. Mark the column the following distances from the end using typewriter "white-out" or a felt-tipped pen:

128 . Clarus 690 GC Customer Hardware and Service Guide

Column Diameter

Detector to which you are attaching the column

Distance from back of nut

≤ 0.53 mm i.d.

FID

2.75 inches (70 mm)

≤ 0.53 mm i.d.

FID with Adapter

3.74 inches (95 mm)

≤ 0.53 mm i.d.

ECD 5

2.5 inches (64 mm)

≤ 0.53 mm i.d.

ECD with Adapter

3.46 inches (88 mm)

≤ 0.53 mm i.d.

TCD

4 inches (103 mm)

≤ 0.53 mm i.d.

NPD

3 inches (77 mm)

0.53 mm i.d.

FPD

2.5 inches (64 mm)

≤ 0.32 mm i.d.

FPD

4 inches (103 mm)

0.53 mm i.d.

PID

4 inches (103 mm)

≤ 0.32 mm i.d.

PID

5.25 inches (135 mm)

5. From the System Status screen select the Tools menu. From the tools menu select Utlitities.

6. In the Utilities screen select the Column Length Calc icon. 7. Using the drop down menus you will be able to select the proper detector and injector for the calculator. The using the plus and minus buttons input the length from the back of the nut and the Column Length Calculator will automatically calculate the column length in millimeters or inches.

5

A glass-lined receiver (P/N N6000968) is available to reduce high background readings.

Step 6: Connect the Column to the Detector . 129

CAUTION ATTENTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne contactent pas la marque sur la colonne.

6. Locate the detector fitting protruding from the right side of the oven roof. 7. When using capillary columns with the FID or ECD, install the adapter fitting (P/N N6120020) on the FID or ECD fitting.

N OTE: The adapter fitting provides the proper column connection to the FID for optimum performance. 8. Insert the column into the detector adapter fitting, keeping the mark just behind the column nut. 9. While holding the column in position, hand-tighten the column nut. 10. Hold the detector fitting steady with one of the 7/16-inch wrenches as you gradually tighten the column nut with the other wrench. Tighten the nut only until you cannot pull the column out of the nut. DO NOT OVERTIGHTEN THE NUT!

Figure 54. Capillary column connected to the detector fitting.

130 . Clarus 690 GC Customer Hardware and Service Guide N OTE: Before connecting a wide-bore column (0.53-mm i.d.) to a Photoionization Detector (PID), remove the 1/8-inch diameter receiver from the PID base.

PID/ELCD Series Tubing PID Receiver

Figure 55. PID receiver viewed inside the column oven

CAUTION ATTENTION

Make certain that no part of the column touches the walls or bottom of the oven. Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Assurez-vous qu'aucune partie de la colonne ne touche les parois ou le fond du four. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

Step 7: Leak Test All New Connections . 131

Step 7: Leak Test All New Connections Test the detector connection for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To avoid contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections.

Step 8: Set up the Split M ode for a CAP or PSS I njector The Split Mode is used to analyze concentrated samples. In this mode only part of the sample enters the column; the remainder is split and vented through a charcoal filter to atmosphere. This step describes how to set the split mode for PPC.

I M PORTANT: For information on Setting the Split Mode Using PPC, Setting up the Splitless Mode for a CAP or PSS Injector, Setting the Splitless Mode Using PPC Modules, refer to the Clarus 690 GC Users Guide (P/N 09931374) Using the Active Method chapter. N OTE: In the Split Mode, the split vent is always open. N OTE: The injector is shipped with an unpacked wide-bore injector liner installed. Remove the liner and pack it with quartz wool before running your analysis. Refer to Step 2, Connect the Column to the Injector, in this chapter.

132 . Clarus 690 GC Customer Hardware and Service Guide

P SS and P OC Operating Hints: The Programmed Split/Splitless (PSS) and Programmed On-Column (POC) injectors can be operated in either the oven-programming mode or the inlet-programming mode. For specific instructions, refer to Chapter 9 of this manual.

CAUTION ATTENTION

If you have the subambient option, the POC and PSS injectors are linked to the oven subambient option; therefore, you cannot operate the injectors below the oven subambient temperature. Si vous avez l'option subambient, les injecteurs POC et PSS sont liés à l'option subambient du four; Par conséquent, vous ne pouvez pas utiliser les injecteurs sous la température du sous-ambiant du four.

Oven P rogram m ing M ode This is the default mode for both the POC and the PSS injectors. This mode is the easiest to operate since only the oven temperature program needs to be entered into the method. In this mode, the injector will follow the oven temperature profile plus five degrees Celsius. In this mode, the sample is introduced into the injector when the temperature of the inlet is at the boiling point of the solvent. Then start the injector and oven temperature program. If the initial temperature of the oven is above the boiling point of the solvent you are using, then it would be better to modify the oven program to start at a lower temperature or to configure the injector for the Inlet mode and set a temperature for the injector separate from the oven temperature.

I nlet P rogram m ing M ode This mode permits the use of independent injector temperatures and rates that you define in the method. The injector will be programmed for injector temperature 1, injector time 1, injector rate 1, injector temperature 2, injector time 2, etc. You can program up to three temperatures and two ramps for each PSS or POC configured in the inlet mode. It is important to set the initial injector temperature to approximately the boiling point of the solvent you are using.

CAUTION ATTENTION

The PSS can be used in the "hot" split or splitless mode. This, however, is not recommended for use with the 1 mm i.d. liner; it could cause solvent flashback in the injector. This mode should be used with caution depending upon the solvent and temperatures you choose. Always use the 2-mm i.d. injector liner. See Table 6-1at the beginning of this chapter. Le PSS peut être utilisé dans le mode «chaud» ou fractionné. Ceci, cependant, n'est pas recommandé pour une utilisation avec i.d. 1 mm. doublure; Cela pourrait provoquer un retour de solvant dans l'injecteur. Ce mode doit être utilisé avec prudence en fonction du solvant et des températures que vous choisissez. Utilisez toujours le i.d. 2 mm. Doublure injecteur. Voir le tableau 6-1 au début de ce chapitre.

PSS and POC Operating Hints: . 133

CAUTION ATTENTION

When using the PSS in the on-column mode, or the POC with the autosampler, you must use a special syringe that has a needle o.d. of 0.47 mm (P/N N6101253 or N6101380). You must only use the "Norm" injection speed with this syringe in the on-column mode. The "Fast" injection speed will bend this thin needle and the "Slow" injection speed may produce peak break up or distorted peaks. You can achieve better precision in the on-column mode when sample volumes of 1.0 uL or greater are injected. Lorsque vous utilisez le PSS en mode colonne ou le POC avec l'autoéchantillonneur, vous devez utiliser une seringue spéciale qui a une aiguille o.d. De 0,47 mm (P / N N6101253 ou N6101380). Vous devez uniquement utiliser la vitesse d'injection "Norm" avec cette seringue en mode colonne. La vitesse d'injection "rapide" plie cette fine aiguille et la vitesse d'injection "lente" peut produire des pics de pointe ou des pics déformés. Vous pouvez obtenir une meilleure précision en mode colonne lorsque des volumes d'échantillons de 1,0 uL ou plus sont injectés.

If a column is used extensively at high temperatures (350 °C or greater), the polyimide may become very brittle. This brittleness will cause the column to fracture when you try to seal it in the universal adapter. If you wish to continue using a brittle column, use a low dead-volume union instead of the universal adapter.

Solvent P urge M ode The PSS injector can be operated in the solvent purge mode without any modification. This technique is an enhanced mode of the splitless injection and may use either the 0.5-µL or 5.0-µL syringe. To obtain better detection limits, you may also use a 50-µL syringe to make larger volume injections. In the solvent purge mode, the split vent is open at the time of injection and the PSS injector temperature is held below the boiling point of the solvent. Once most of the solvent has vented, the split vent closes and the PSS injector is heated so that it can transfer the analytes onto the column. You must select the starting temperature of the PSS injector based on the solvent you are using. In addition, to eliminate most of the solvent without losing any of the compounds of interest, you must experimentally determine the minimum split flow and vent time. The initial value of the split vent can be ON or OFF, depending upon your solvent volatility. When you make your injection, the split vent OPENS (turns ON) and stays open for the time you have selected. After that time, the split vent CLOSES (turns OFF via a timed event) and the PSS injector heats to the selected temperature. To help maintain good peak shape and quantitative recovery, we recommend holding the Clarus 690 GC oven temperature at its starting value until the PSS injector temperature reaches its final temperature and all of the sample has been transferred onto the column. To minimize tailing of the residual solvent and ensure that the system is completely cleaned for the next injection, start the oven temperature program and reopen the split vent. The charcoal trap's capacity is too small to handle the large volumes of solvent and it will saturate quickly. When making large volume injections, we recommend disconnecting the charcoal trap from the PSS injector. For details, refer to “Replacing the Charcoal Trap on the Split/Splitless, CAP, and PSS Injectors” in Chapter 9, Maintenance. To retain the vented solvent, we strongly recommend a larger capacity charcoal trap after (downstream from) the vent on the Clarus GC.

134 . Clarus 690 GC Customer Hardware and Service Guide N OTE: When using the PSS injector in the split mode, the charcoal trap is required to maintain the split ratio throughout the injection.

Calculating a Capillary Column Split Ratio . 135

Calculating a Capillary Colum n Split R atio The following procedure and example show how to calculate the capillary column split ratio using PPC pneumatics. 1. Enter the column length, I.D., and vacuum compensation values. 2. Measure the unretained peak time and correct the column length, see Chapter 8. 3. Read the column and split flows from the screen on the Clarus 690 GC. 4. Calculate the Split Ratio: Flow rate from the split vent in mL/min + Flow rate of the column Split Rate = _______________________________________________________________________________

Flow rate of the column

N OTE: Remember that the split ratio determines how much sample is placed into the capillary column. A larger split ratio means that less sample is placed in the column, therefore less sample is analyzed by the detector. In order to reproduce the same chromatographic conditions in the future save the oven temperature program, column head pressure value, and split vent flow rate value. In the PPC version, you can directly control the split ratio.

136 . Clarus 690 GC Customer Hardware and Service Guide

Calculating a Capillary Column Split Ratio . 137

7 Prevent

138 . Clarus 690 GC Customer Hardware and Service Guide

About this Chapter . 139

About this Chapter PreVent is an enhanced capillary inlet system for the Clarus 690 GC that uses columns with an inside diameter (i.d.) between 0.25 mm and 0.53 mm. You can use the pre-column venting technique to: 

Manage difficult samples



Protect the column and gas chromatograph



Lower detection limits



Increase analytical throughput



PreVent is used in the following five Modes of Operation:



Column Isolation



Solvent Purge (PSS injector only)



Large Volume Injection (PSS injector only)



Sample Residue Purge (PSS injector only)



Time Saver

This chapter describes how to install the PreVent adapter and restrictor on an injector or detector. Based on your mode of operation, the PreVent adapter is installed on either the injector or detector (including Clarus SQ 8 MS). Turn to the procedure in this chapter that describes how to install PreVent: •

Installing PreVent on an Injector



Installing PreVent on a Detector



Installing PreVent on a Clarus SQ 8 MS Detector

N OTE: The instructions in this chapter assume that the PreVent option has been installed in the front injector position (Channel A). The mid-point pressure for Channel A is controlled by valve 3, and the mid-point pressure for Channel B is controlled by valve 4. N OTE: The PreVent option contains a bleed flow of nominally 15 mL/min of carrier gas that constantly flows and vents to atmosphere when you set the pressure on the PreVent mid-point module. If your carrier gas is hydrogen, you should connect tubing to the barbed fitting on the bleed controller (on the underside of the Clarus 690 GC top cover), and direct it to a venting hood.

140 . Clarus 690 GC Customer Hardware and Service Guide

I nstalling P reVent on an I njector The following procedure describes how to install the PreVent adapter on a programmable split/splitless injector (PSS) and how to make it ready for an analysis.

N ote:

This capability can be achieved with a capillary column injector (CAP) by using Swafer. Please reference P/N 09936965A Swafer Cookbook and P/N 09936966A Swafer Utility Software Users Guide.

Sum m ary The following steps summarize how to install to an injector and get it ready for use: Step 1: Install the restrictor in the PreVent adapter. Step 2: Connect the PreVent adapter to the injector. Step 3: Connect the column to the PreVent adapter. Step 4: Leak-check the system. Step 5: Connect the column to the detector. Step 6: Set the initial pressures. Step 6: Condition the system.

Step 1. I nstall the Restrictor in the P reVent Adapter The following restrictors are available for the injector: 

0.06 mm i.d. restrictor (P/N N6103080)



0.06 mm i.d. to 0.250 mm i.d. restrictor (P/N N6100342)

Select the appropriate restrictor for your analysis and cut it to the proper length. Refer to the following table: Table 7-1 Recommended Restrictors Injector

Injection Mode

Restrictor

(P/N)

Part Number

Length (mm)

PSS

Split

0.06 mm i.d. to 0.250 mm i.d 6.

N6100342

104

PSS

Splitless

0.06 mm i.d. or

N6103080

104

0.06 mm i.d. to 0.250 mm i.d.

N6100342

Thread the selected restrictor first through a 0.5 mm graphite ferrule (P/N 09903700) and then through the PreVent adapter (P/N N6100328) until 5 mm protrudes through the column connection on the bottom. When using the 0.06 mm i.d. to 250 mm i.d. restrictor, the narrow end should protrude through the bottom.

6

This restrictor is pre-cut to the correct length for PSS.

Installing PreVent on an Injector . 141

CAUTION ATTENTION

To prevent contamination problems and the need for extensive conditioning procedures, ensure that all PreVent components are thoroughly clean. Baking them in an oven prior to assembly is recommended or wiping with a tissue dampened with alcohol. Always wear clean gloves when assembling the PreVent components. Pour éviter les problèmes de contamination et la nécessité de procédures de conditionnement étendues, assurez-vous que tous les composants PreVent sont bien nettoyés. La cuisson dans un four avant l'assemblage est recommandée ou l'essuyage avec un tissu humidifié avec de l'alcool. Portez toujours des gants propres lors de l'assemblage des composants PreVent.

Restrictor Tubing (Wide end 0.25 mm i.d Injector End of Adaptor Mid-Point Pressure Supply

1/16-inch x 0.5 mm Graphite Ferrule (P/N 09903700)

PreVent Adapter (P/N N6100328)

Column Connection 06 mm i.d. Restrictor protrudes 5 mm

Figure 56. Installing the restrictor in the PreVent adapter.

Step 2. Connect the P reVent Adapter to the I njector •

Verify that the injector liner is correct for your application. If necessary, pack the injector liner with quartz wool. For details, refer to packing a PSS injector liner in Chapter 6.

5. Carefully connect the PreVent adapter by screwing the injector end of the adapter into the capillary injector fitting until it is finger tight. 6. Using a 1/4-inch wrench, tighten the PreVent adapter until the graphite ferrule just starts to tighten on the restrictor. 7. Locate the stainless steel tubing (that supplies the mid-point pressure) inside the oven. 8. Insert a 1/16-inch nut (P/N 09903392) and a graphite/Vespel ferrule (P/N 09920107) over the end of this tubing. 9. Measure and mark the tubing 1/2-inch from the back of the nut. 10. Insert the tubing into side arm on the PreVent adapter, keeping the mark just behind the nut. Then tighten the nut and ferrule fingertight.

142 . Clarus 690 GC Customer Hardware and Service Guide 11. Make a leak-free seal by tightening the PreVent adapter, on the capillary injector fitting, another 1/4 to 1/2 turn. Mid-Point Pressure Tubing

Detector Fitting

PreVent Adapter (P/N N6100329)

1/16-inch Nut (P/N 09903392)

Graphite Vespel Ferrule (P/N 09920107)

Figure 57. Capillary injector fitting inside the oven.

Step 3. Connect the Colum n to the P reVent Adapter 1. Place the column on the hanger so that no part of the column touches the bottom or sides of the oven. 2. Insert a 1/16-inch extended nut (P/N 09903392, pkg. of 5), a graphite/Vespel ferrule (1/16-inch x 0.5 mm P/N 09920105, pkg. of 10 for 0.25/0.32 i.d. column, or a 1/16inch x 0.8 mm P/N 09920107, pkg of 10 for a 0.53 mm i.d. column), and a spacer (P/N N6103079), over one end of the column.

CAUTION ATTENTION

Do not use a graphite ferrule since it will extrude up into the body of the PreVent adapter. N'utilisez pas une virole en graphite puisqu'elle sera extrudée dans le corps de l'adaptateur PreVent.

1/16-inch Extended Nut (P/N 09903392, pkg. of 5)

Spacer (P/N N6103079)

1/16-inch x 0.5 mm Graphite/Vespel Ferrule (P/N 09920105, pkg. of 10) 1/16-inch x 0.8 mm Graphite/Vespel Ferrule (P/N 09920107, pkg. of 10)

Figure 58. Extended nut, graphite/Vespel ferrule, and spacer. 3. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N9301386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Installing PreVent on an Injector . 143

Figure 59. Example of a good column cut and bad cuts. 4. Insert the column end (as shown in, Figure 61 Step 1) into the hourglass guide (P/N N6103082). 5. Align the end of the restrictor tubing that protrudes from the PreVent adapter with the open end of the hourglass guide. Then, carefully push the hourglass guide, with the attached column, up into the PreVent adapter as shown in Figure 61, Step 2. Notice that the restrictor tubing fits inside the column.

NOTE: Slightly twist the column as you push the hourglass guide into the PreVent adapter. 6. Loosely connect the column nut to the PreVent adapter. 7. Carefully push the column up and into the PreVent adapter until it bottoms in the adapter. Then pull the column back 3 mm as shown in the following figure, Step 3. 8. Use two 1/4-inch wrenches to tighten the nut. DO NOT OVERTIGHTEN THE NUT!

Figure 60. Connecting a column to the PreVent injector adapter.

144 . Clarus 690 GC Customer Hardware and Service Guide

Step 4. Leak -Check the System 1. Locate the detector end of the column. 2. Seal the column end by firmly inserting the detector end of the column into the side of a septum to about the mid-point as shown in the following figure.

Figure 61. Column inserted into a septum. 3. Turn on the Clarus 690 GC. When the Clarus 690 GC is turned on, it is set to the initial default settings. 4. From the System Status screen touch the injector to get to the following screen. Touch the numeric field until it turns black and use the up and down arrow or keypad buttons to get the proper split flow of 1.5 and a ratio of 25.

5. From the Events tab turn off the mid-point carrier gas supply by closing Valve 3.

Installing PreVent on an Injector . 145

6. Allow the system to equilibrate for several minutes, and then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

7. Return to the inject screen to view the carrier gas pressure read out which should be at 0 psig.

146 . Clarus 690 GC Customer Hardware and Service Guide 8. Set the injector pressure to 75 psig. After approximately 5 minutes, if no leaks exist, the Aux 1 reading should be stable at approximately 75 psi. If it is low, then leak-check the fitting connections. If you suspect a leak, test the connections for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To prevent contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections. If you detect a leak, check and tighten the fittings on the mid-point pressure tee and/or the connections on the rear of the Clarus 690 GC.

CAUTION ATTENTION

Do not overtighten the column nut. If this fitting has a persistent leak, replace the graphite/Vespel ferrule. Ne pas trop serrer l'écrou de la colonne. Si ce raccord présente une fuite persistante, remplacez la ferrule graphite/Vespel.

9. Turn off the two gas pressures and remove the septum from the end of the column.

Step 5. Connect the Colum n to the Detector CAUTION ATTENTION

Make certain that no part of the column touches the walls or bottom of the oven. Assurez-vous qu'aucune partie de la colonne ne touche les parois ou le fond du four.

N OTE: To connect a column to the SQ 8 MS, refer to the instructions in the Clarus SQ 8 MS Hardware Guide (P/N 09931017). 1. Insert a 1/8-inch column nut and graphite ferrule over the free end of the column as shown in the following figure. 1/8-inch Column Nut White-Out

Graphite Ferrule (1/8-inch x 0.5 mm i.d. P/N 0990398 (1/8-inch x 1.0 mm i.d. P/N 09903394

Figure 62. Nut and ferrule on the detector end of a narrow-bore capillary column. 2. Cut about 1 cm (3/8 inch) from the column end using a wafer scribe (P/N N9301376, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark so that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Installing PreVent on an Injector . 147

Figure 63. Example of a good cut and bad cuts. 3. Mark the column using typewriter "white-out" or a felt-tipped pen. The distances are calculated using Column Length Calc procedure that follows. 4. From the System Status screen select the Tools menu. From the tools menu select Utilities.

5. In the Utilities screen select the Column Length Calc icon.

6. Using the drop-down menus you will be able to select the proper detector and injector for the calculator. The using the plus and minus buttons input the length from the back

148 . Clarus 690 GC Customer Hardware and Service Guide of the nut and the Column Length Calculator will automatically calculate the column length in millimeters or inches.

CAUTION ATTENTION

To prevent contaminating the system, ensure that the nut and ferrule do not contact the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne contactent pas la marque sur la colonne.

7. Locate the detector fitting that protrudes through the right side of the oven roof. 8. Insert the column into the detector fitting, keeping the mark just behind the column nut. 9. While holding the column in position, hand-tighten the column nut. 10. Holding the detector fitting steady with one of the 7/16-inch wrenches, gradually tighten the column nut with the other wrench only until you cannot pull the column out of the nut. DO NOT OVERTIGHTEN THE NUT!

Figure 64. Capillary column connected to the detector fitting.

Installing PreVent on an Injector . 149

CAUTION ATTENTION

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

Step 6. Set the I nitial P ressures To set the mid-point pressure: 1. From the System Status screen select either A or B injector (the example here shows the B injector ready for setting the carrier gas).

2. In the following window touch the Program Graph screen under the Carrier Gas field.

3. Using the up and down arrow buttons or keypad enter the initial value of 30 psig for the Carrier gas. Press Close to go back to the previous screen (as seen in step 2).

150 . Clarus 690 GC Customer Hardware and Service Guide

4. In the Events tab touch the Split event and then press Edit. Set Split 1 to 100 mL/min.

5. From the Oven tab set the oven temperature to its highest programmed temperature.

Installing PreVent on an Injector . 151 6. From the Events tab turn off the mid-point carrier gas supply by closing Valve 3.

Allow the system to equilibrate for several minutes, and then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

N OTE: If the Aux 1 pressure is very low, then the graphite ferrule that holds the restrictor inside the PreVent adapter may be leaking. Remedy this by tightening the adapter on the injector and observe if there is an improvement. 7. Adjust the pressure until the Aux pressure is maintained at about 1.5 psig below the required column inlet pressure. Wait a few minutes for the system to stabilize to ensure that no further changes are required. 8. Open Valve 3 and set the Aux 1 pressure to the required column inlet pressure; for example, 1.5 psig higher than the Aux 1 reading established in step 9 above. This ensures a positive gas flow through the Aux 1 supply to the mid-point tee. CAUTION ATTENTION

Valve 3 MUST remain open (ON) during PreVent operation. La vanne 3 DOIT rester ouverte (ON) pendant l'opération PreVent.

9. Condition the system to ready it for an analysis.

152 . Clarus 690 GC Customer Hardware and Service Guide

Step 7. Condition the System 1. Write down the pressures established for Press 1 and Aux 1 in step 6. 2. Set Press 1 to 2 psig, Split 1 to 100 mL/min, and Aux 1 to 80 psig. 3. Set the PSS injector temperature to the maximum programmed temperature. 4. Set the oven to the maximum programmed temperature. 5. Allow the system to bake at this temperature for several hours  overnight if possible. 6. Shorter conditioning times may be possible, but that depends on the condition of the PreVent components and condition of the column. 7. Cool the oven and cool the PSS injector). 8. Reset the initial settings of Press 1 and Aux 1. The system is now ready to run an analysis.

Installing PreVent on a Detector . 153

I nstalling P reVent on a Detector This section describes how to install PreVent on a detector (except for Clarus SQ8 MS). The next section describes how to install PreVent on Clarus SQ MS.

Sum m ary The following steps summarize how to install a PreVent adapter to a detector and make it ready for an analysis: Step 1.

Install the restrictor in the PreVent adapter.

Step 2.

Connect the PreVent adapter to the detector.

Step 3.

Connect the column to the PreVent adapter to the detector.

Step 4.

Leak-check the system.

Step 5.

Connect the column to the injector.

Step 6.

Set the initial pressures.

Step 7.

Condition the system.

Step 1. I nstall the Restrictor in the P reVent Adapter 1. Cut the 0.075 mm i.d. restrictor (P/N N6103081) to the proper length for your detector. Refer to the following table: Table 7-1. Recommended Restrictor Lengths for Detectors Detector to which you are attaching the adapter

Restrictor Length

PID

195 mm (7.66 inches)

TCD

163 mm (6.40 inches)

FID

137 mm (5.38 inches)

ECD

7

124 mm (4.87 inches)

NPD

137 mm (5.38 inches)

FPD

163 mm (6.40 inches)

2. Insert a 1/8-inch graphite ferrule over one end of the restrictor tubing. 3. Insert the end of the restrictor tubing into the top of the PreVent detector adapter (P/N N6100329). 4. Feed the restrictor tubing through the PreVent adapter until 5 mm protrudes through the column connection on the bottom.

7

A glass-lined receiver (P/N N600-0968) is available to reduce high background readings.

154 . Clarus 690 GC Customer Hardware and Service Guide

Figure 65. Installing the restrictor in the PreVent detector adapter.

Step 2. Connect the P reVent Adapter to the Detector 1. Carefully connect the PreVent adapter by screwing the detector end of the adapter into the detector fitting until it is finger tight. 2. Using a 7/16-inch wrench, tighten the PreVent adapter until the graphite ferrule just starts to tighten on the restrictor. 3. Locate the stainless steel tubing (that supplies the mid-point pressure) inside the oven. 4. Insert a 1/16-inch nut (P/N 09903392) and a graphite/Vespel ferrule (P/N 09920107) over the end of this tubing. 5. Measure and mark the tubing 1/2-inch from the back of the nut. 6. Insert the tubing into side arm on the PreVent adapter, keeping the mark just behind the nut. Then tighten the nut and ferrule fingertight. 7. Make a leak-free seal by tightening the PreVent adapter on the detector fitting another 1/4 to 1/2 turn. Mid-Point Pressure Tubing

Detector Fitting

PreVent Adapter (P/N N6100329)

1/16-inch Nut (P/N 09903392)

Graphite Vespel Ferrule (P/N 09920107)

Figure 66. Location of the detector fitting.

Installing PreVent on a Detector . 155

Step 3. Connect the Colum n to the P reVent Adapter 1. Place the column on the hanger so that no part of the column touches the bottom or sides of the oven. 2. Insert a 1/16-inch extended nut (P/N 09903392, pkg. of 5), a graphite/Vespel ferrule (1/16-inch x 0.5 mm P/N 09920105, pkg. of 10 for 0.25/0.32 i.d. column or a 1/16-inch x 0.8 mm P/N 09920107, pkg of 10 for a 0.53 mm i.d. column), and a spacer (P/N N6103079) over one end of the column.

CAUTION ATTENTION

Do not use a graphite ferrule since it will extrude up into the body of the PreVent adapter. N'utilisez pas une virole en graphite puisqu'elle sera extrudée dans le corps de l'adaptateur PreVent.

1/16-inch Extended Nut (P/N 09903392, pkg. of 5)

Spacer (P/N N6103079)

1/16-inch x 0.5 mm Graphite/Vespel Ferrule (P/N 09920105, pkg. of 10) 1/16-inch x 0.8 mm Graphite/Vespel Ferrule (P/N 09920107, pkg. of 10)

Figure 67. Extended nut, graphite/Vespel ferrule, and spacer. 3. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N9301386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 68. Example of a good tubing cut and bad cuts. 4. Insert the column end into the hourglass guide (P/N N6103082). 5. Align the end of the restrictor tubing that protrudes from the PreVent adapter with the open end of the hourglass guide. Then carefully push the hourglass guide, with the attached column, up into the PreVent adapter as shown in the following figure. Notice that the restrictor tubing fits inside the column.

N OTE: Slightly twist the column as you push the hourglass guide into the PreVent adapter.

156 . Clarus 690 GC Customer Hardware and Service Guide 6. Loosely tighten the column nut. 7. Carefully push the column up into the PreVent adapter until it bottoms in the detector. Then pull the column back 3 mm. 8. Use two 1/4-inch wrenches to tighten the nut. DO NOT OVERTIGHTEN THE NUT! Detector Fitting

Graphite Ferrule Restricto Tubing

PreVent Adapter

Restrictor Tubing Hourglass Guide (P/N N6103082)

Space

Spacer (P/N N6103079)

Column Tubing

Column Column

1

2

3

Figure 69. Connecting a column to the PreVent detector adapter.

Step 4. Leak -Check the System 1. Turn on the Clarus 690 GC. When the Clarus 690 GC is turned on, it is set to the initial default settings. 2. From the System Status screen touch the injector to get to the following screen. Turn the detector temperature off.

Installing PreVent on a Detector . 157

3. From the Events tab turn off the mid-point carrier gas supply by opening Valve 3.

4. Allow the system to equilibrate for several minutes, then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

158 . Clarus 690 GC Customer Hardware and Service Guide 5. Return to the inject screen to view the carrier gas pressure read-out which should be at 0 psig.

6. After approximately 5 minutes, if no leaks exist, the Aux 1 reading should be stable at approximately 75 psi. If it is low, then leak-check the fitting connections. 7. If you suspect a leak, test the connections for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To prevent contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections. 8. If you detect a leak, check and tighten the fittings on the mid-point pressure tee and/or the connections on the rear of the Clarus 690 GC.

CAUTION ATTENTION

Do not overtighten the column nut. If this fitting has a persistent leak, replace the graphite/Vespel ferrule. Ne pas trop serrer l'écrou de la colonne. Si ce raccord présente une fuite persistante, remplacez la ferrule graphite/Vespel.

Step 5. Connect the Colum n to the I njector CAUTION ATTENTION

Make certain that no part of the column touches the walls or bottom of the oven. Assurez-vous qu'aucune partie de la colonne ne touche les parois ou le fond du four.

1. Insert a 1/16-inch extended nut (P/N 09903392, pkg. of 5) and graphite/Vespel ferrule (1/16-inch x 0.5 mm P/N 09920105, pkg. of 10 for 0.25/0.32 i.d. column or a 1/16-inch x 0.8 mm P/N 09920107, pkg of 10 for a 0.53 mm i.d. column) over one end of the column.

Installing PreVent on a Detector . 159 1/16-inch Extended Nut (P/N 09903392, pkg. of 5)

1/16-inch x 0.5 mm Graphite/Vespel Ferrule (P/N 09920105, pkg. of 10) 1/16-inch x 0.8 mm Graphite/Vespel Ferrule (P/N 09920107, pkg. of 10)

Figure 70. Extended nut and graphite/Vespel ferrule. 2. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N9301386, pkg of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 71. Example of a good column cut and bad cuts. 3. Position the column nut on the column so that the back of the nut is 1 1/2 inches to 1 3/4 inches from the end of the column for the PSS injector. 4. Using typewriter "white out" or a felt-tipped pen, make a mark on the column just beyond the back edge of the column nut.

CAUTION ATTENTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne contactent pas la marque sur la colonne.

5. Locate the capillary injector fitting inside the oven. 6. Refer to the following figures to connect the column to the PSS injector.

160 . Clarus 690 GC Customer Hardware and Service Guide

CAUTION ATTENTION

CAUTION ATTENTION

The injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut. L'injecteur se termine par un raccord de 1/16 de pouce. Ce raccord est fragile. Pour préserver l'intégrité du raccord, connectez soigneusement l'écrou pour empêcher le filetage croisé du raccord et / ou le serrage excessif de l'écrou sur le raccord. Vous pouvez également préserver l'intégrité du raccord en permettant à l'injecteur de refroidir avant de brancher un écrou.

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

Figure 72. Capillary column attached to PSS injector fitting.

Step 6. Set the I nitial P ressures To set the mid-point pressure: 1. From the System Status screen select either A or B injector (the example here shows the B PSS injector ready for setting the carrier gas).

Installing PreVent on a Detector . 161

2. In the following window touch the Program Graph screen under the Carrier Gas field.

3. Using the up and down arrow buttons or keypad to enter the initial value of 30 psig for the Carrier gas. Press Close to go back to the previous screen (as seen in step 2).

162 . Clarus 690 GC Customer Hardware and Service Guide

4. In the Events tab touch the Split event and then press Edit. Set Split 1 to 100 mL/min.

5. From the Oven tab set the oven temperature to its highest programmed temperature.

Installing PreVent on a Detector . 163

6. From the Events tab turn off the mid-point carrier gas supply by closing Valve 3.

7. Allow the system to equilibrate for several minutes, then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

164 . Clarus 690 GC Customer Hardware and Service Guide

N OTE: If the Aux 1 pressure is very low, then the graphite ferrule that holds the restrictor inside the PreVent adapter may be leaking. Remedy this by tightening the adapter on the injector and observe if there is an improvement. 8. Adjust the pressure until the Aux pressure is maintained at about 1.5 psig below the required column inlet pressure. Wait a few minutes for the system to stabilize to ensure that no further changes are required. 9. Open Valve 3 and set the Aux 1 pressure to the required column inlet pressure; for example, 1.5 psig higher than the Aux 1 reading established in step 9 above. This ensures a positive gas flow through the Aux 1 supply to the mid-point tee. CAUTION ATTENTION

Valve 3 MUST remain open (ON) during PreVent operation. La vanne 3 DOIT rester ouverte (ON) pendant l'opération PreVent.

10. Condition the system to ready it for an analysis.

Step 7. Condition the System 1. Write down the pressures established in step F for Press 1 and Aux 1. 2. Set Press 1 to 2 psig, Split 1 to 100 mL/min, and Aux 1 to 80 psig. 3. If you are using a PSS, set its temperature to the maximum programmed temperature. Set the oven to the maximum programmed temperature. 4. Allow the system to bake for several hours  overnight if possible. Shorter conditioning times may be possible, but that depends on the condition of the PreVent components and condition of the column. Cool the oven (and if you are using a PSS injector, cool the injector.) 5. Reset the initial settings of Press 1 and Aux 1. The system is now ready to use for an analysis.

Installing PreVent on a Clarus SQ 8 MS Detector . 165

I nstalling P reVent on a Clarus SQ 8 M S Detector This procedure describes how to install PreVent on a Clarus SQ 8 MS (GCMS) detector.

Sum m ary The following steps summarize how to install a PreVent injector adapter to a GCMS detector and make it ready for an analysis: Step 1.

Prepare the Clarus SQ 8 MS Detector

Step 2.

Install the restrictor in the Clarus SQ 8 MS Transfer Line.

Step 3.

Install the PreVent injector adapter.

Step 4.

Connect the column to the PreVent injector adapter.

Step 5.

Connect the Transfer Line to the Clarus SQ 8 MS

Step 6.

Leak-check the system.

Step 7.

Connect the column to the injector.

Step 8.

Set the initial pressures.

Step 9.

Condition the system

Step 1. P repare the Clarus SQ 8 M S 1. Install the Clarus SQ 8 MS and Clarus 690 GC according to the instructions given in their installation manuals. 2. If a column is installed, cool the Gas Chromatograph oven, the ionizer and the transfer line to ambient temperature. Vent the Clarus SQ 8 MS and remove the column from the transfer line.

Inner Transfer Tube

Pull Back

Column

1/4-inch Nut

Figure 73. Removing a column from the Clarus SQ 8 MS transfer line.

Step 2. I nstall the P reVent I njector Adapter in the Clarus SQ 8 M S Transfer Line 1. Cut a 40cm length of the 0.075mm i.d. fused silica tubing (P/N 6103081). Make sure that the two ends are cut clean and square. Wipe the outside of the tubing with a tissue dampened with methanol.

166 . Clarus 690 GC Customer Hardware and Service Guide 2. Feed the fused silica tubing through the PreVent injector adapter (P/N N6100328) and a graphitized Vespel ferrule (P/N 09920105) as shown in the following figure. Fuse silica tubing (P/N 6103081)

Ferrule (P/N 09920105)

PreVent Injector Adapter (P/N N6100328)

Figure 74. Feeding the fused silica tubing through the injector adapter. 3. Introduce the fused silica tubing into the GCMS transfer line until it reaches a stop. Loosely screw the PreVent adapter onto the 1/16” threaded union at the end of the GCMS transfer line as shown in the following figure.

Figure 75. Normal connection of column to Clarus SQ 8 transfer line. 4. Withdraw the fused silica tubing by about 5mm and using two 1/4 inch wrenches tighten up the PreVent adapter on the union at the end of the GCMS transfer line. 5. Cut the fused silica tubing so that about 7mm is exposed beyond the PreVent adapter as shown in the following figure. The cut should be clean and square.

CAUTION ATTENTION

Take care when working near the exposed end of the fused silica tubing. It can be easily broken off. Faites attention lorsque vous travaillez près de l'extrémité exposée du tube de silice fondu. Il peut être facilement interrompu.

Installing PreVent on a Clarus SQ 8 MS Detector . 167 Figure 76. Example of a good column cut and bad cuts. 6. Locate the stainless steel tubing (that supplies the mid-point pressure) inside the oven. 7. Insert a 1/16-inch nut (P/N 09903392) and a graphite/Vespel ferrule (P/N 0992-0107) over the end of this tubing. 8. Measure and mark the tubing 1/2-inch from the back of the nut. 9. Insert the tubing into side arm on the PreVent adapter, keeping the mark just behind the nut. Then tighten the nut and ferrule fingertight. 10. Make a leak-free seal by using a two 1/4-inch wrenches to tighten the PreVent adapter on to the union another 1/4 to 1/2 turn.

Step 3. Connect the Colum n to the P reVent I njector Adapter 1. Place the column on the hanger so that no part of the column touches the bottom or sides of the oven. 2. Insert a 1/16-inch extended nut (P/N 09903392, pkg. of 5) a graphite/ Vespel ferrule (P/N 09920105, pkg. of 10), and a spacer (P/N N6103079) over one end of the column.

CAUTION ATTENTION

Do not use a graphite ferrule since it will extrude up into the body of the PreVent adapter. N'utilisez pas une virole en graphite puisqu'elle sera extrudée dans le corps de l'adaptateur PreVent.

Spacer (P/N N6103079)

1/16-inch Extended Nut (P/N 09903392, pkg. of 5)

1/16-inch x 0.5 mm Graphite/Vespel Ferrule (P/N 09920105, pkg. of 10)

Figure 77. Extended nut, graphite/Vespel ferrule, and spacer. 3. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N9301386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 78. Example of a good tubing cut and bad cuts.

168 . Clarus 690 GC Customer Hardware and Service Guide 4. Insert the column end into the hourglass guide (P/N N6103082). 5. Align the end of the restrictor tubing that protrudes from the PreVent adapter with the open end of the hourglass guide. Then carefully push the hourglass guide, with the attached column, into the PreVent adapter as shown in the following figure, step 2. Notice that the restrictor tubing fits inside the column.

N OTE: Slightly twist the column as you push the hourglass guide into the PreVent adapter. 6. Loosely connect the column nut to the PreVent adapter. 7. Carefully push the column up and into the PreVent adapter until it bottoms in the adapter. Then pull the column back 3 mm. 8. Use two 1/4-inch wrenches to tighten the nut. DO NOT OVERTIGHTEN THE NUT!

PreVent Adapter

Hourglass Guide (P/N N6103082) Spacer Restrictor (P/N N6103079) Tubing

1 Column

2

3

Restrictor Tubing

Column Tubing

Figure 79. Connecting a column to the PreVent adapter in a Clarus SQ 8 system.

Step 4. Connect the Transfer Line to the Clarus SQ 8 1. Install the PreVent device into the Clarus SQ 8 as shown in Figure 81. 2. Take the PreVent ‘injector’ T-piece and screw it onto the 1/16” fitting at the end of the standard transfer line. Be sure the fused silica restrictor is inserted through it, through the heated transfer line and directly into the ion source. 3. Ensure that the end of the restrictor is exposed beyond the T-piece to enable the column to be threaded over it with the aid of an hourglass guide. Figure 82 through Figure 85 show details of the installation.

Installing PreVent on a Clarus SQ 8 MS Detector . 169 4. Figure 86 shows a detail of the column-restrictor interface. Because the restrictor is inserted directly into the end of the capillary column, there are no issues with active sites or dead volumes. Therefore an optimum chromatographic performance is assured.

Figure 80. Diagram showing PreVent device installed into the Clarus SQ 8 detector (The TuboMass is the figure on the left and the Clarus 690 GC is the figure on the right).

Figure 81. Column is detached and PreVent adapter is attached to Clarus SQ 8 transfer line with fused silica restrictor pushed through into ionizer.

170 . Clarus 690 GC Customer Hardware and Service Guide

Figure 82. Restrictor is sealed within adapter with end protruding. Midpoint gas supply is connected to port on adapter.

Figure 83. Hourglass guides restrictor into capillary column

Figure 84. Installation completed.

Installing PreVent on a Clarus SQ 8 MS Detector . 171

Figure 85. Detail of column-restrictor interface at end of Clarus SQ 8 transfer line. Once installed, this PreVent configuration will allow the capillary column to be back-flushed and so offers the following benefits: 

Allows injector maintenance without having to cool or vent the vacuum on the mass spectrometer.



Allows a degraded section of column to be removed from its inlet without the need to cool or pump-down the detector.



Facilitates easier column replacement.



Reduces risk of air entering the detector in the event of a leak or column breakage.



Prevents column bleed from entering the detector while conditioning or when the system is idle.



Allows the injector liner and quartz wool packing to be deactivated in-situ.



Reduces analysis times by eliminating the need for extensive temperature programming to elute unwanted less volatile sample residue from the column.



Increases the life of a column by eliminating temperature programming completely.



Removes the potential of contaminants before they reach the detector.

It will also have the added benefit of maintaining a constant flow of carrier gas into the MS detector in both forward flow and backflush modes during temperature programming.

Step 5. Leak -Check the System 1. Turn on the Clarus 690 GC. When the Clarus 690 GC is turned on, it is set to the initial default settings. 2. From the System Status screen touch the injector to get to the following screen. Touch the numeric field until it turns black and use the plus minus buttons to get the proper split flow of 100 and a ratio of 100.

172 . Clarus 690 GC Customer Hardware and Service Guide

3. Select Valve 3 (or valve 4 if PreVent is installed in channel 2) from the Event drop down menu.

4. Allow the system to equilibrate for several minutes, and then view the Aux 1 pressure, view the Aux Gas button and view the current psig.

Installing PreVent on a Clarus SQ 8 MS Detector . 173

5. Return to the inject screen to view the carrier gas pressure read out which should be at 0 psig.

6. Set the injector pressure to 75 psig. 7. After approximately 5 minutes, if no leaks exist, the Aux 1 reading should be stable at approximately 75 psi. If it is low, then leak-check the fitting connections. If you suspect a leak, test the connections for leaks using a 50/50 mixture of isopropanol/water or an electronic leak detector. To prevent contaminating the system, DO NOT use a soap solution for leak testing. Tighten all leaking connections. If you detect a leak, check and tighten the fittings on the mid-point pressure tee and/or the connections on the rear of the Clarus 690 GC.

CAUTION ATTENTION

Do not overtighten the column nut. If this fitting has a persistent leak, replace the graphite/Vespel ferrule. Ne pas trop serrer l'écrou de la colonne. Si ce raccord présente une fuite persistante, remplacez la ferrule graphite / Vespel.

174 . Clarus 690 GC Customer Hardware and Service Guide

Step 6. Connect the Colum n to the I njector CAUTION ATTENTION

Make certain that no part of the column touches the walls or bottom of the oven. Assurez-vous qu'aucune partie de la colonne ne touche les parois ou le fond du four.

1. Insert a 1/16-inch extended nut (P/N 09903392, pkg of 5) and graphite/ Vespel ferrule (P/N 09920105, pkg of 10) over one end of the column.

1/16-inch Extended Nut (P/N 09903392, pkg of 5)

1/16-inch Graphite/Vespel Fer (P/N 09920105, pkg of 10) Figure 86. Extended nut and graphite/Vespel ferrule. 2. Cut about 1 cm (3/8 inch) from the end of the column tubing using a wafer scribe (P/N N9301386, package of 10 scribes) or other column cutting tool. Break off the tubing at the score mark making sure that the break is clean and square. Examine the cut with a magnifying glass and compare it to the following figure:

Figure 87. Example of a good column cut and bad cuts. 3. Position the column nut on the column so that the back of the nut is 4.4 cm to 5.1 cm (1 3/4 inches to 2 inches) from the end of the column for the CAP injector. or Position the column nut on the column so that the back of the nut is 1 1/2 inches to 1 3/4 inches from the end of the column for the PSS injector. 4. Using typewriter "white out" or a felt-tipped pen, make a mark on the column just beyond the back edge of the column nut.

Installing PreVent on a Clarus SQ 8 MS Detector . 175

CAUTION ATTENTION

To avoid contaminating the system, make certain that the nut and ferrule do not contact the mark on the column. Pour éviter de contaminer le système, assurez-vous que l'écrou et la ferrule ne contactent pas la marque sur la colonne.

5. Locate the capillary injector fitting inside the oven. Refer to the following figures to connect the column to the PSS injector.

CAUTION ATTENTION

CAUTION ATTENTION

The injector terminates in a 1/16-inch fitting. This fitting is fragile. To preserve the integrity of the fitting, carefully connect the nut to prevent cross-threading the fitting and/or overtightening the nut on the fitting. You can also preserve the integrity of the fitting by allowing the injector to cool before connecting a nut. L'injecteur se termine par un raccord de 1/16 de pouce. Ce raccord est fragile. Pour préserver l'intégrité du raccord, connectez soigneusement l'écrou pour empêcher le filetage croisé du raccord et / ou le serrage excessif de l'écrou sur le raccord. Vous pouvez également préserver l'intégrité du raccord en permettant à l'injecteur de refroidir avant de brancher un écrou.

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne.

Figure 88. Capillary column attached to PSS injector fitting.

Step 7. Set the I nitial P ressures 1. From the System Status screen select the oven icon.

176 . Clarus 690 GC Customer Hardware and Service Guide

2. Set the oven to the lowest programmed temperature.

3. Set the transfer line to its operating temperature. 4. From the Events tab turn off the mid-point carrier gas supply by closing Valve 3.

Installing PreVent on a Clarus SQ 8 MS Detector . 177

5. Start the vacuum 6. Set the gas pressure to 30 psi.

N OTE: If the aux 1 pressure is very low, then the graphite ferrule that holds the restrictor inside the PreVent adapter may be leaking. Remedy this by tightening the adapter on the injector and observe if there is an improvement. 7. Adjust the pressure until the Aux pressure is maintained at about 1.5 psig below the required column inlet pressure. Wait a few minutes for the system to stabilize to ensure that no further changes are required. 8. Open Valve 3 and set the Aux 1 pressure to the required column inlet pressure; for example, 1.5 psig higher than the Aux 1 reading established in step 9 above. This ensures a positive gas flow through the Aux 1 supply to the mid-point tee. CAUTION ATTENTION

Valve 3 MUST remain open (ON) during PreVent operation. La vanne 3 DOIT rester ouverte (ON) pendant l'opération PreVent.

9. Condition the system to ready it for an analysis.

178 . Clarus 690 GC Customer Hardware and Service Guide

Step 8. Condition the P SS System 1. Write down the pressures established for Press 1 and Aux 1 in step I. 2. Set Press 1 to 2 psig, Split 1 to 100 mL/min, and Aux 1 to 80 psig. 3. Set oven and PSS injector to the maximum programmed temperature. 4. Allow the system to bake for several hours  overnight if possible. Shorter conditioning times may be possible, but that depends on the condition of the PreVent components and condition of the column. 5. Cool the oven and PSS injector. 6. Reset the initial settings of Press 1 and Aux 1. The system is now ready to use for an analysis.

Step 9. Replacing the Restrictor

To replace a restrictor in the PreVent injector or detector adapter: 1. Turn off the Clarus GC oven and detector and wait until they are cool to the touch. 2. Turn off all gases. 3. Using a 1/4-inch wrench, remove the 1/16-inch column nut and the column from the PreVent adapter. 4. Using two 1/4-inch wrenches, remove the nut and mid-point pressure line from the side arm of the PreVent adapter. 5. Using the appropriate wrenches, remove the PreVent adapter from either the injector or detector. •

Use two 1/4-inch wrenches to remove the PreVent adapter from the injector fitting.



Use two 7/16-inch wrenches to remove the PreVent adapter from the detector fitting.

6. Insert a 1/16-inch rod (P/N N610T100) in the column connection end of the PreVent adapter and push the restrictor tubing and ferrule of the injector end. 7. Install a new restrictor by referring to the appropriate procedure in this chapter.

Installing PreVent on a Clarus SQ 8 MS Detector . 179 Restrictor Tubing

Injector End of Adaptor Mid-Point Pressure Supply

1/16-inch x 0.5 mm Graphite Ferrule (P/N 09903700)

PreVent Adapter (P/N N610-0328)

Restrictor Tubing

Column Connectio

1/16-inch Rod (P/N N610-T100)

Figure 89. Removing a restrictor from a PreVent adapter.

180 . Clarus 690 GC Customer Hardware and Service Guide

P reVent Operating Techniques Sum m ary of the P reVent Techniques Table 7-3 summarizes the various PreVent techniques and the hardware configurations that support them. •

Column Isolation



Solvent Purge



Large Volume Injection



Sample Residue Purge



Time Saver

Table 7-3. Summary of PreVent Techniques and Supporting Configurations Technique

PSS

Restrictor in Injector

Restrictor in Detector

Column Isolation

Yes

Yes

Yes*

Solvent Purge

Yes

Yes

Yes

Large Volume Injection

Yes

Yes

Yes

Sample Residue Purge

Yes

Yes**

Yes

Time Saver

Yes

No

Yes

* Cannot be used while chromatography is in progress ** Not recommended

Table 7-3 demonstrates the advantage of the PSS injector over the conventional Split/Splitless injector in that all the PreVent techniques are supported by this one injector. Several of these techniques are possible with a single restrictor configuration. If a different configuration is required, changing the restrictor type is a simple operation requiring only a few minutes to perform.

Colum n I solation Technique This technique uses the Injector Restrictor Configuration to fully isolate the chromatographic column from actions occurring within the injector. Figure 91 and Figure 92 illustrate the principle. For sample introduction, the inlet pressure (P1) is set up to be higher than the mid-point pressure (P2). P1 and P2 are set as described in “Installing PreVent on the Injector” in this chapter. P2 represents the pressure at the column inlet. Carrier gas will now flow from the injector, through the restrictor and into the column, carrying the vaporized sample with it. The sample may be injected manually using a standard syringe or by autosampler. The injection temperature and split vent operation are set as for normal split or splitless injection techniques. The pressures are not adjusted until the injector needs to be isolated. To isolate the column, Press 1 (P1) is simply reduced to below that of Aux 1 (P2) causing the carrier gas to flow backwards from the mid-point ‘T’, through the restrictor, into the injector and out through the split vent. It is important to ensure that the split vent is open (for example, 50 mL/min) during this step or the technique will not work.

PreVent Operating Techniques . 181 Split Vent Control (Open or Closed)

F Injector

Detector

Restrictor

FLOW

Chromatographic Column P1

P2 Aux 1

Press 1 (20 psig)

(14 psig)

Figure 90. Normal operation. Split Vent Control (50 ml/min)

F Injector

Detector Restrictor

FLOW

Chromatographic Column P1 Press 1 (1 psig)

P2 Aux 1 (14 psig)

Figure 91. Column Isolation Mode. P1 should not be turned off completely, but to a level (for example, 1.0 psig) that maintains forward flow of carrier gas through the liner to prevent the build-up of contamination within the liner while the restrictor is being backflushed and also to maintain a purge flow out of the injector inlet while it is being serviced with the septum cap removed. The reduction in P1 is not normally included as a timed event in the Clarus 690 GC method as this will not be routinely adjusted during each run. The Column Isolation Mode can be used to support the following actions: 

Septum exchange and conditioning even during chromatography



Liner exchange and conditioning even during chromatography



In situ chemical deactivation of the liner even during chromatography



Column and detector protection while the instrument is idle



Column Isolation Mode is also possible with the restrictor fitted to the detector as shown in Figure 93 and Figure 94. In this instance, however, chromatography is not possible

182 . Clarus 690 GC Customer Hardware and Service Guide during column isolation although it now would be possible to condition a column without the effluent reaching the detector. Split Vent Control (Open or Closed) F Chromatographic Column

Injector

Detector

Restrictor

FLOW

P1

P2 Aux 1

Press 1 (20 psig)

(6 psig)

Figure 92. Normal operation. Split Vent Control (50mL/min)

F

Chromatographic Column Detector

Injector

Restrictor

FLOW

P1

P2

Press 1

Aux 1

(1 psig)

(6 psig)

Figure 93. Alternative Column Isolation Mode.

Solvent P urge Technique This technique uses the ability of the PSS injector to rapidly change its liner temperature according to a user-defined program. Figure 95 shows how this ability can be used to remove the solvent from the liner and selectively transfer the less-volatile sample analytes to the column. This technique is best suited to splitless injections of low volatility analytes. While the solvent is being purged from the liner, there is still forward flow into the column and so some of the solvent will still enter the column (and hence the detector). The PreVent system, configured as for the Column Isolation Mode may be used to prevent all traces of the solvent entering the column during the purge process by reversing the flow of carrier gas at the injector column port.

PreVent Operating Techniques . 183

3

2

1

Step 1 : Sample injected into cool liner

Step 2 : Solvent and other volatile components swept from cool liner and out of split vent

Step 3 : Liner heated and the vaporized residue swept into column

Figure 94. Using the PSS to eliminate solvent from the column. Table 7-4 shows typical entries that would be made in a method to enable a 1-minute solvent purge at 100 mL/min and 50 °C. Actual settings will depend upon the type and volume of solvent involved. The initial settings for P1 and P2 are established as described in this chapter. The column oven is programmed as for normal splitless injection except that the temperature is not raised until after the solvent purge step is complete. Table 7-4. Typical Method for Solvent Purge Technique Time

Event

Comment

INITIAL

SPLIT 1 (F) = 100 mL/min

Purge flow rate.

INITIAL

PSS = 50 °C

Purge temperature.

INITIAL

AUX 1 (P2) = 14 psig

Mid-point pressure at column inlet.

INITIAL

PRESS 1 (P1) = 2 psig

Pressure in injector lower than midpoint to keep solvent out of column.

1.00

SPLIT 1 (F) = OFF

Close split vent to transfer everything to column.

1.01

PRESS 1 (P1) = 20 psig

Raise injector pressure so carrier gas flows from liner to column.

1.02

PSS = 350 °C

Heat liner to vaporize sample residue.

2.00

SPLIT = 50 mL/min

Finally open split vent to clean liner for next run.

The Solvent Purge Technique may be used to support the following actions: 

Elimination of peak distortion caused by solvent flooding effects as a result of excess liquid solvent entering the column and becoming fractionated by the carrier gas flow



Elimination of hostile solvents from sensitive columns and detectors



Large volume injections - (See Large Volume Injections below.)

184 . Clarus 690 GC Customer Hardware and Service Guide

Large Volum e I njections (LVI ) The Solvent Purge Technique can also be used to support LVI applications of the PSS injector. The PSS must be used with a 2-mm i.d. liner which is firmly packed with glass wool. A large volume of liquid sample (up to 100 µl - a 50-µl syringe is available for the autosampler) may be injected into the liner which must be kept at a low temperature to prevent (very) explosive vaporization. The solvent is purged out of the split vent as for the Solvent Purge Technique. The purge time and flow rate need to be sufficient to remove all of the solvent (for example, about 100-200 mL of purge gas is needed to remove 50 µl of n-hexane at 50 °C). Extended purge times may result in the loss of some of the analytes and so this technique is only suitable for low volatility compounds (for example, lower than n-decane). Typical method entries are given in Table 7-5. Table 7-5. Typical Method for LVI Technique Time

Event

Comment

INITIAL

SPLIT 1 (F) = 100 mL/min

Purge flow rate.

INITIAL

PSS = 50 °C

Purge temperature.

INITIAL

AUX 1 (P2) = 14 psig

Mid-point pressure at column inlet.

INITIAL

PRESS 1 (P1) = 2 psig

Pressure in injector lower than mid-point to keep solvent out of column.

2.00

SPLIT 1 (F) = OFF

Close split vent to transfer everything to column.

2.01

PRESS 1 (P1) = 20 psig

Raise injector pressure so carrier gas flows from liner to column.

2.02

PSS = 350 °C

Heat liner to vaporize sample residue.

3.00

SPLIT = 50 mL/min

Finally open split vent to clean liner for next run.

Sam ple Residue P urge Technique The separation described under the Solvent Purge Mode can be reversed so that the volatile sample content enters the column and heavy (unwanted) low volatility sample residue left in the liner can be purged out through the split vent. This technique works best with the detector restrictor PreVent configuration so that the traces of low volatility material that do leave the liner can be backflushed from the column (see next Section). Use of the injector restrictor configuration will not prevent some of the low-volatility material from entering the column during the purge process. The figure below illustrates the separation process inside the PSS liner.

PreVent Operating Techniques . 185

3

2

1

Step 1 : Sample injected into cool liner

Step 2 : Volatile components swept from cool liner into column

Step 3 : Liner heated and most of the residue swept out of split vent

Figure 95. Using the PSS to eliminate low volatility sample material from the column. The figures below illustrate how the PreVent system backflushes the heavier sample from the column. For further details on column backflushing refer to the “Time Saver Technique” Table 76 lists some typical entries in a method for the Sample Residue Purge technique. The initial settings for P1 and P2 are established as described in this chapter. Split or splitless injection with a PSS injector is possible and the oven temperature program is set as for normal chromatography. This technique is useful for the determination of volatile components in heavy oils, natural extracts, etc. Split Vent Control (Open or Closed)

F

Chromatographic Column Detector

Injector

Restrictor

FLOW

P1

P2

Press 1

Aux 1

(20 psig)

(6 psig)

Figure 96. Injection and chromatography.

186 . Clarus 690 GC Customer Hardware and Service Guide Split Vent Control (100 ml/min)

F

Chromatographic Column Detector

Injector

Restrictor

FLOW

P1

P2

Press 1

Aux 1

(1 psig)

(80 psig)

Figure 97. Column backflush. Table 7-6. Typical Method for Sample Residue Purge Technique Time

Event

Comment

INITIAL

SPLIT 1 (F) = 100 mL/min

Injection split flow (can use splitless).

INITIAL

PSS = 200 °C

Injection temperature.

INITIAL

AUX 1 (P2) = 14 psig

Mid-point pressure at column inlet.

INITIAL

PRESS 1 (P1) = 20 psig

Pressure in injector higher than mid-point to allow sample into column.

2.00

SPLIT 1 (F) = 100 mL/min

Open split vent (if not already) to reduce the amount of further sample entering column.

20.00

PRESS 1 (P1) = 5 psig

When last peak of interest has eluted, reduce inlet pressure...

20.01

AUX 1 (P2) = 80 psig

and increase mid-point pressure to backflush any sample material left in column.

20.02

PSS = 450 °C

Heat liner to bake out any sample residue and purge it out of the split vent.

Tim e Saver Technique The PreVent system is able to support the classical, single-column, backflush technique. The use of PPC pneumatics enables the backflush conditions to be optimized for rapid backflush of unwanted sample material after the last peak of interest has eluted. The initial settings for P1 and P2 are established as described in this chapter. Split or splitless injection with a PSS injector is possible and the oven temperature program is set as for normal chromatography. When the last peak of interest has eluted, the split vent is opened, P1 is reduced and P2 is increased. Any sample material left in the column is rapidly backflushed even at low oven temperatures eliminating the need, in many instances, for temperature programming. This technique is suitable for the determination of components in any sample where there is unwanted low-volatility material entering the column. By reducing the need for temperature

PreVent Operating Techniques . 187 programming to remove such material, it reduces the analysis time and protects the column by not exposing it to high temperatures. Table 7-7. Typical Method for Column Backflush Technique Event

Comment

INITIAL

SPLIT 1 (F) = 100 mL/min

Injection split flow (can use splitless).

INITIAL

AUX 1 (P2) = 14 psig

Mid-point pressure at column inlet.

INITIAL

PRESS 1 (P1) = 20 psig

Pressure in injector higher than mid-point to allow sample into column.

1.00

SPLIT 1 (F) = 100 mL/min

Open split vent (if not already) to allow backflushed components to escape to vent and help clean liner.

20.00

PRESS 1 (P1) = 5 psig

When last peak of interest has eluted, reduce inlet pressure...

20.01

AUX 1 (P2) = 80 psig

... and increase mid-point pressure to backflush any sample material left in column.

188 . Clarus 690 GC Customer Hardware and Service Guide

PreVent Operating Techniques . 189

8 P P C Fundam entals

190 . Clarus 690 GC Customer Hardware and Service Guide

Introduction . 191

I ntroduction Programmed Pneumatic Control (PPC) is the electronic control of pressures and flows for inlet, detector, and auxiliary gases. The PPC control modules regulate pressures and flows using electronically driven variable flow restrictors. The control modules also contain pressure and flow transducers to provide feedback for complete monitoring. A PPC controller board drives the variable restrictors on the control modules by comparing actual pressures and flows with setpoints determined from user-entered values. This chapter presents an overview of the concepts and use of Programmable Pneumatic Control (PPC) of the Clarus 690 GC. In particular, it describes the following microprocessor-controlled pneumatic functions that comprise PPC: 

Carrier-gas flow, pressure, and linear-velocity programming



Split-flow and split-ratio control



Detector-gas flow control



Auxiliary-gas pressure and flow control

The following unique pneumatic devices make these control functions possible: 

Carrier-gas mass-flow controller



Split-pneumatic controller



Detector-gas flow controller



Auxiliary pressure controller



Pressure-readout module

You can combine one or more of these modules to create the control functions required for a specific pneumatic configuration in the GC. A dedicated microprocessor control system reads all operating parameters from — and sends control signals to — the pneumatic devices. This PPC controller receives pneumatic setpoints, the ambient temperature, pressure, the oven temperature, and GC status information from the main Clarus 690 GC processor. The PPC controller sends back the pneumatic operating parameters and PPC status information to the main GC processor. The PPC controller accommodates up to eight flow controllers and/or twelve pressure controllers, for a maximum of twelve installed PPC devices. The main Clarus 690 GC processor accepts and stores method setpoints and timed events for the configured PPC devices. In addition, the main Clarus 690 GC processor contains PPC functions that include calibration and setup utilities, diagnostics, and fault-condition monitoring.

192 . Clarus 690 GC Customer Hardware and Service Guide

Carrier Gas Control The type of primary PPC carrier gas modules and their operation depends on the inlet (injector) option to which they are connected. For example, the packed-column inlet and the programmedtemperature on-column inlet (POC) use a single carrier gas mass-flow controller plus a pressurereadout module. The split/splitless (capillary) and programmed-temperature split/splitless (PSS) inlets use a combination of a carrier gas mass-flow controller and a split-pneumatic pressure controller. Other carrier-gas controllers may use auxiliary PPC zones (see Auxiliary Carrier Gas Control).

P acked Colum n and P rogram m ed-Tem perature On-Colum n I nlets The PPC carrier gas mass-flow controller delivers a constant flow rate into a column. The controller operates by increasing or decreasing the column head pressure to maintain a constant mass-flow as the column temperature changes, or if you use different columns. Its internal operation is similar to a conventional mass-flow controller, but instead of utilizing a spring-anddiaphragm mechanism for control, the PPC controller reads and sets the flow rate electronically. Unlike a conventional controller, it constantly monitors the ambient room pressure, the tank pressure, and the temperature of each flow module, and it compensates as required to maintain a constant mass-flow output. The schematic in the following figure illustrates the PPC mass-flow controller configuration for packed column and POC inlets.

Figure 98.

Schematic of packed column and POC pneumatics

Carrier Gas M ass-Flow Controller Operating R ange The carrier gas mass-flow controller for packed-column inlets has a nominal flow range of 0 – 30 mL/min for helium carrier gas at 90 psig tank pressure. The controller for the POC inlet has a range of 0 – 10 mL/min helium for capillary column applications. The two controllers differ only in the installed flow-range element (restrictor). By changing the flow-range element, you can use the same controller in different Clarus 690 GC configurations. The adapter kit for packed-column installation of 0.53-mm i.d. capillary columns, for example, contains the appropriate 0 – 10 mL/min element. After changing the flow-range element, you must always recalibrate the flow controller.

M ass-Flow Controller Setup The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery backed-up random access memory (BRAM). This information is written into the

Carrier Gas Control . 193 BRAM during manufacturing. Normally, there is no need to access or modify the type or location of the PPC controllers. If the PPC hardware configuration is changed — by installing or removing PPC devices — the operator must access the PPC connection menus on the Clarus 690 GC touch screen and modify the type and location of the PPC devices. If the installed hardware does not match the stored configuration, the Clarus 690 GC issues a warning message.

CAUTION ATTENTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed. Ne modifiez pas les informations de connexion PPC si la configuration matérielle PPC n'a pas changé.

M ass-Flow Controller Calibration The carrier gas mass-flow controller is factory calibrated to a National Institute of Standards and Technology (NIST) traceable standard with helium carrier gas; however, it requires periodic calibration. Recalibration is required after changing the flow-range element; otherwise the calibration frequency is up to the individual operator. The controller accommodates changing the carrier gas (to other than helium) by approximating a new set of calibration parameters based on the new gas. The approximation is good but controller recalibration is recommended for the best accuracy. It is good laboratory practice to recalibrate all measuring and control devices periodically. The mass-flow controller calibration procedure consists of specifying the reference temperature and pressure; zeroing the internal sensors in the flow controller with the tank pressure off; specifying the correct carrier gas type and installing the correct flow-range element; measuring the background flow rate at zero-flow with the tank pressure on; and measuring the flow rate at 60% of full-scale flow. This procedure is built into the Clarus 690 GC firmware. Once the PPC is set up and calibrated you must configure it.

Carrier Gas M ass-Flow Controller Configuration Once the carrier gas mass-flow controller has been set up and calibrated for the PPC system, you must configure its operating mode on the Clarus 690 GC keypad. For packed-column and POC inlets the selections are Zero, Flow, or Pressure. 1. From the System Status Screen touch the Tools button. 2. From the tools drop down menu select Configuration.

194 . Clarus 690 GC Customer Hardware and Service Guide

3. In the Configuration touch screen touch the PPC icon

4. In the PPC touch screen touch PPC Configure bar.

.

Carrier Gas Control . 195

5. A popup warning appears, select OK.

6. On the Configure PPC Devices screen touch the Zero bar then touch OK to set the gas flow to zero.

196 . Clarus 690 GC Customer Hardware and Service Guide

Headspace Analysis Headspace analysis is the analysis of the vapor lying in equilibrium over a solid/liquid sample in a sealed vial. For practical headspace analysis the sample is sealed in a vapor tight vial, placed in a thermostatted oven and heated to a pre-determined temperature. The sample vial contains the volatile material in equilibrium between the solid/liquid sample and the vapor lying over it. After equilibrium is reached between solid/liquid phase and the vapor phase, a defined amount of the vapor is taken and carried to the column in the gas chromatograph for analysis. With this technique only highly volatile substances reach the column, the non-volatile substances remain in the sample vial. Using this technique, samples containing constituents which are unsuitable for injection with a syringe can be analyzed (e.g. polymers, highly viscous liquids). Suitable fields of application are in the analysis of polymers, certifying of the volatile components in drinks and foodstuffs, blood alcohol levels, water and environmental analysis. For information on vacuum see the section on Vacuum Compensation later in this chapter.

Capillary and P rogram m ed Split/ Splitless I nlets A packed-column or POC inlet requires only a source of flow-controlled carrier gas; however, split-type inlet systems are more complex. The capillary and programmed-temperature split/splitless (PSS) inlets have identical split pneumatics. The split pneumatics occupy two PPC control zones, one for column inlet pressure control and one for split flow rate control.

Split P neum atic Setup The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing. Normally, there is no need to access or modify the type or location of the PPC controllers. If the PPC hardware configuration is changed — by installing or removing PPC devices — you must access the PPC connection menus on the Clarus 690 GC keypad and modify the type and location of the PPC devices. If the installed hardware does not match the stored configuration, the Clarus 690 GC issues a warning message.

Carrier Gas Control . 197

Do not modify the PPC setup information if the PPC hardware configuration has not changed. Ne modifiez pas les informations de configuration PPC si la configuration matérielle PPC n'a pas changé.

CAUTION ATTENTION

Split P neum atic Control A split system such as the capillary or PSS inlet pneumatics controls three parameters: the split flow, the inlet pressure, and the septum purge flow. See the following figure that shows two diagrams showing the forward pressure mode (splitless control) and the backpressure mode (split control) of the PPC split pneumatics in the Clarus 690 GC. Septum Purge

∆p

Controller Temperature

Pressure Regulator

∆p

Restrictor

Charcoal Trap

Flow Monitor ∆p

Solenoid Valve (Closed)

Pressure Control ∆p

Ambient Pressure

Forward Pressure Mode (Splitless Control) Septum Purge

∆p

Controller Temperature

Pressure Regulator

∆p

Restrictor

Charcoal Trap

Total Split Flow Control ∆p

Pressure Control ∆p

Ambient Pressure

Solenoid Valve (Open)

198 . Clarus 690 GC Customer Hardware and Service Guide Backpressure Mode (Split Control) Figure 99. Schematic of the split PPC pneumatics

Split Flow ─ A mass-flow controller supplies carrier gas to the split or PSS inlet system. This is the same mass-flow controller used for packed or POC inlets, but here it is used with a 0 – 300 mL/min flow-range element (fixed restrictor). Carrier gas passes through the flow controller and into the inlet (injector) at the connection labeled Carrier I n . Some of the gas flows across the septum and out the septum purge, while the rest flows down through the inlet liner. At the bottom of the liner, a fraction of the carrier gas enters the column, and the balance flows back up the outside of the liner and out the split vent. An electrical solenoid valve provides positive split flow shutoff when required. I nlet P ressure ─ A pressure transducer connected at the split vent (before the charcoal trap) measures the inlet pressure at the head of the column. The inlet pressure may range from 0 – 100 psig. Septum Purge ─ A constant-flow device (fixed pressure regulator) regulates the septum purge flow at approximately 3 mL/min, independently of the inlet pressure. The regulator is preset in the factory; it is not adjustable.

Split P neum atic Calibration It is recommended that the PPC split modules be recalibrated at regular intervals and this should always be done following a period of inactivity of the GC and if the type of carrier gas is changed (e.g. from helium to hydrogen). It is good laboratory practice to recalibrate all measuring and control devices periodically. The PPC controllers accommodate changing the carrier gas to other than helium by approximating a new set of calibration parameters based on the new gas. The approximation is good but controller recalibration is recommended for the best accuracy. There are four aspects to split injector PPC calibration: 1. Enter the pressure and temperature reference settings to match the flowmeter calibration conditions a. Select Configuration, then Pneumatics, then Reference Settings. b. Enter values for Ref. Pressure and Ref. Temperature to match conditions under which the flowmeter was calibrated. For a bubble flowmeter, enter the current ambient temperature and pressure.

N OTE: The procedure for setting ambient Actual P ressure control as applied to PPC column flow and velocity control is not currently supported due to an open issue in the firmware. Investigation has shown the impact to retention time variability is minor but the issue will be addressed in the future. Please contact your technical support representative for more details or to obtain the future fix. 2. Enter the electronic zero for each of the pressure transducers used within the PPC modules. Variability in transducers signal offsets will directly affect the pressure readings and could introduce errors if not calibrated. Also, calibrate the split zero flow point so that the total

Carrier Gas Control . 199 flow rate would deliver that flow rate to the split. Both “zero’ing” activities are done in the same location.

I M PORTANT: Confirm the injector, oven and detector are cool and the carrier gas supply is disconnected. First turn off carrier supply. Then vent the internal pressure by loosening the septum nut and frit to the PPC module. a. Select Configuration, then Pneumatics, then PPC Configure. b. Choose the channel to calibration (Channel A or Channel B). Select the Zero button for the Carr Flow.

c.

d. Confirm that the Offset1 and Offset2 Actual values are stable and then select the Auto button to enter (or manually enter) the displayed values in the two SetPoint entry fields. e. Press OK to return to the PPC Configure screen. Select the Zero button for the Split Ctrl.

f.

g. Zero the Offset in the same way as in Step d above. h. To confirm the correct entries, check that the carrier gas is displaying 0.0 for all flows and pressures on the instrument injector status page, while the carrier gas supply is still disconnected. i.

Once confirmed, re-tighten any fittings loosened during venting and re-apply the carrier gas to the GC.

3. Perform a split flow calibration so that the split flow set in the method is accurately applied. Errors in the calibration will produce errors in the applied split flow rate producing errors in the split ratio and subsequent quantitation. To perform this calibration: •

A reliable and accurate flowmeter should be connected to the split vent.



Ensure that there’s been enough carrier gas passing through the system before measuring any flow rates to completely flush air from the PPC controllers and injector. a. Select Configuration PPC page, then PPC Configure button. b. Choose the channel to calibration (Channel A or Channel B). c.

Select the correct Gas.

d. Select the Calibrate button. e. Confirm the correct flow Restrictor insert/frit is configured when prompted. • f.

For the split injectors it’s normally a #6 frit which is colored light blue.

Select Next to continue the calibration.

g. The system will now perform a two-point flow calibration. Default points are provided or specific values could be entered, based on intended operating ranges. 4. Set the split flow offset

I M PORTANT: Leaks must be eliminated before proceeding with the calibration.

200 . Clarus 690 GC Customer Hardware and Service Guide I M PORTANT: Split flow offset calibration must always be updated after the carrier gas flow rate through the column has been changed (e.g. following a change to the carrier gas set pressure, flow or velocity; a column exchange; changing to a different carrier gas and rezeroing the sensors). a. Ensure that carrier gas has had sufficient time to fully purge the system before performing this step. To do this, set the split flow to a high value (>200mL/min) for several minutes to help flush the system. b. Set up the GC with the analytical method to be used c.

Allow the GC to equilibrate to the method starting conditions.

d. Go to the main status page and set the split flow rate to zero. e. Let the carrier gas stabilize and note the Total Flow reading. f.

Go to the Configuration screen and select the Injector being used.

g. Enter the Total Flow value recorded above into the Flow Offset entry field.

N OTE: The “Auto” option for this offset can be used. When activated, however, this calibration occurs before every run, increases run time, and could impact results (normally for the better) by changing operating conditions. The “Auto” mode is also used in the carrier gas leak checking feature.

Split P neum atic Carrier Gas Configuration Once the split pneumatic system has been set up and calibrated on the PPC system you must configure its operating mode by using the Clarus 690 GC touch screen. There are configuration choices for column carrier-gas control and split flow or ratio control. The PPC controller includes a mathematical model of a capillary column. This model is enabled by setting Capillary Control Mode ON (see Capillary Control Mode, below). The PPC controller uses this model to calculate the column exit flow at ambient pressure and temperature, the column average linear gas velocity, and the inlet split ratio in real time. You may choose to control the column flow or velocity, in which case the PPC controller calculates the pressure required to maintain the flow or velocity setpoint. You may also choose to control the split ratio; in this case the PPC controller calculates the split flow rate required to obtain the setpoint split ratio. The PPC controller microprocessor carries out the method setpoint program for both carrier-gas channels by taking the setpoint pressure, flow, or velocity value; the column dimensions; the carrier-gas type; the reference temperature and pressure; the oven actual temperature; and the ambient pressure. It calculates the other two parameters in real time. For pressure control, it calculates the flow and velocity; for flow control, it calculates the pressure and velocity; for velocity control, it calculates the pressure and flow. These calculations track ambient pressure changes, changing oven temperatures, and the execution of the multi-ramp carrier-gas program. In each case, the actual pressure in the inlet is controlled by the calculated or method pressure setpoint. For more information on these internal calculations, see the section on “Theory of Capillary Column Control,” in this chapter.

Carrier Gas Control . 201 Capillary Control Mode The first choice for split carrier-gas control is the Capillary Control Mode. This parameter may be set On or Off. Each carrier gas channel may be setup with or without capillary control.

Off ─ uses direct pressure control of the column through a multi-step pressure program and carrier gas timed events. You can directly control the split flow control through the initial setpoint and timed-event control of the split flow controller. Split ratio control, capillary column flow, and capillary column velocity control and information are not available. This mode of operation does not require knowledge of the capillary column dimensions. On ─ enables indirect control of capillary column flow or velocity in addition to direct pressure control, each through a multi-step program and timed events. This mode also enables direct split flow or indirect split ratio control of the split flow controller. The display reports actual pressure, flow, velocity, split flow, and split ratio. In this mode you must enter accurate column length and diameter information (except for pressure-controlled operation with oven tracking). Carrier Gas Configuration without Capillary Column Control When capillary control is turned off for a split PPC controller, the choices for carrier-gas control are None and Pressure.

None ─ removes the split pneumatics from the GC method and turns off the split flow and pressure. P ressure ─ enables direct pressure control of the split inlet. The PPC controller uses a 5 psig initial setpoint with an infinite hold time when Pressure is first selected. The user may add up to three additional setpoints, ramps, and hold times. This mode also enables direct split flow control (see below). Carrier Gas Configuration with Capillary Column Control ON In this mode the choices for carrier-gas control are None, Pressure, Flow, and Velocity. With pressure control you may select either Pressure: Program or Pressure: Oventrack.

None ─ removes the split pneumatics from the GC method and turns off the split flow and

pressure.

P ressure: P rogram ─ enables direct pressure control of the column pressure. The split carriergas pressure program mode gives a multi-ramp program with real-time calculation and display of flow and velocity, plus flow control or ratio control of the split flow. The PPC controller uses a 5 psig initial setpoint with an infinite hold time when P ressure: P rogram is first selected. The pressure range is 0 – 100 psig. You may add up to three additional setpoints, ramps, and hold times. P ressure: Oventrack ─ also enables direct carrier-gas pressure control. The PPC controller automatically varies the initial inlet pressure so that the capillary column flow rate remains constant as the oven temperature changes away from its initial setpoint value. The PPC controller uses a 5 psig initial setpoint when P ressure: Oventrack is first selected. Multi-ramp programming is not available. This mode allows only direct flow control of split flow; however, this mode can be used for constant column flow control when you do not know your column dimensions. Flow ─ enables indirect capillary column mass-flow control. The PPC controller calculates the inlet pressure required to maintain the method flow program setpoints from the column dimensions, carrier-gas type, oven temperature, and ambient pressure. The display shows the flow, pressure, and calculated average linear gas velocity. The PPC controller uses a 2 mL/min initial setpoint with an infinite hold time when Flow is first selected. The column flow has a total flow range of 0 – 500 mL/min (including the split flow). You may add up to three additional

202 . Clarus 690 GC Customer Hardware and Service Guide setpoints, ramps, and hold times. This mode enables flow control or split ratio control of the split flow.

Velocity ─ enables indirect capillary column average linear gas velocity control. The PPC controller calculates the inlet pressure required to maintain the method velocity program setpoints from the column dimensions, carrier-gas type, oven temperature, and ambient pressure. The display shows the velocity, pressure, and calculated flow rate. The PPC controller uses a 30 cm/s initial setpoint with an infinite hold time when Velocity is first selected. You may add up to three additional setpoints, ramps, and hold times. The carrier-gas velocity has a range of 0–200 cm/s. This mode enables flow control or split ratio control of the split flow.

Colum n Dim ensions After selecting any of the capillary control modes (except Pressure: Oventrack) the Clarus 690 GC prompts you for the column dimensions. These parameters control the relationships between the pressure, flow, and velocity as calculated by the PPC controller. If the column dimensions are inaccurate, the PPC controller will produce incorrect readings and will incorrectly control the column flow, velocity, and split ratio. Column Length The column length defaults to 25.0 meters. It has a range from 1.0 – 200.0 meters. Small errors in the column length affect the average linear velocity and the flow rate only slightly. See the section “PPC Tips and Techniques” at the end of this chapter for procedures to correct errors in column length. Column Inner Diameter The column inner diameter defaults to 250 µm. It has a range from 50 –1000 µm. Small errors in the column inner diameter strongly affect the average linear velocity and the flow rate.

Vacuum Com pensation The next Clarus 690 GC split pneumatic configuration parameter in the capillary control mode is vacuum compensation. The PPC controller uses a mathematical model of a capillary column to determine pressure – flow – velocity relationships. The column outlet pressure plays an important role in these calculations. The PPC controller continually monitors the barometric pressure because normal daily atmospheric pressure fluctuations affect the column average linear gas velocity—and thus retention times—if the controlling system does not compensate. Vacuum compensation controls the role of the outlet pressure in the PPC controller. Vacuum compensation may be set On or Off .

Off ─ causes the PPC controller to assume that the column outlet is at atmospheric pressure. On ─ causes the PPC controller to assume that the column outlet is at vacuum. This setting is appropriate for GC-MS systems with a direct interface where the column enters the massspectrometer source. Vacuum compensation should be turned off when an open-split interface is used since it keeps the GC column outlet close to atmospheric pressure.

Split Control M ode Configuration The last Clarus 690 GC carrier-gas configuration parameter is the split control mode. The ratio of the total flow through the inlet liner to the column flow rate — the fraction of sample vapor entering the column, or the split ratio — controls the quantitative transfer of sample into the column. For split ratio control the PPC system must know both the column flow rate and the total gas flow through the inlet liner.

Carrier Gas Control . 203 When capillary control is off, the PPC system does not compute the capillary column flow rate or the split ratio; only split flow control is possible. With capillary control on (except for P ressure: Oventrack mode) the PPC controller calculates the column flow rate and the split ratio. The choices for split-flow control are Flow or Ratio . There is no choice for N one , since the split pneumatics will not operate without incoming carrier-gas flow. Setting the carrier-gas configuration to None automatically sets the split-flow controller configuration to none.

Flow ─ enables direct split flow control of the carrier gas mass-flow controller. You enter a split vent flow rate in the method. The PPC controller adds an optional offset to obtain the total split flow (see “Split Flow Offset” below) and sets that amount at the flow controller. The Clarus 690 GC also displays the calculated split ratio if capillary control is on. Ratio ─ enables indirect split ratio control of the carrier gas mass-flow controller. The user enters a split ratio in the method. The PPC controller calculates the required split vent flow from the column flow rate and adds an optional offset to obtain the total split flow (see “Split Flow Offset” below) and sets that amount at the flow controller. The Clarus 690 GC also displays the calculated split flow rate defined by the following equation: total split flow = split flow + septum purge flow + column flow

N OTE: Any carrier system leaks will add to the observed total split flow.

Split Flow Offset The PPC controller sets and monitors the total split flow through the inlet system. The split ratio is based on the flow out the split vent. Since the septum purge flow is taken from the total split flow rate before passing through the inlet liner, across the column entrance, and out the split vent, split ratio calculations must account for the flow reduction. The Clarus 690 GC includes a split flow offset parameter that is accessed from the CONFIG CARRIER menus when a split controller is setup. This parameter may be set to Fixed or Auto .

Fix ed ─ prompts the user to enter the septum purge offset value. This offset is the septum purge flow plus the column flow. The initial column flow is displayed on the [Carrier Prog] key when the GC is READY at its initial setpoints with capillary control ON. The PPC controller adds this offset to the method setpoint split flow rate to obtain the actual flow delivered to the split inlet. The septum purge flow is factory set at 3 mL/min. To measure the actual septum purge offset, set the carrier pressure to 5 psig or greater and turn off the split flow (set to 0). The remaining split vent flow displayed is the septum purge flow, plus the column flow. The actual septum purge flow can be measured from the split vent bulkhead fitting. Auto ─ measures the offset flow rate before each run by momentarily setting the split vent flow to 0, waiting for the remaining flow to stabilize, and then recording its magnitude. During this time the GC status is PRE-RUN. This operation occurs after equilibration, and before executing any Pre-Run timed events (events with negative times). The Auto measurement includes both column and septum purge flows. N OTE: The split flow screen (shown below) displays the flow from the injector vent as the setpoint and the total split system flow as the actual values. The value in the upper right carrier is the total split flow. The value in the lower right corner is the split vent setpoint. The flow measured at the split vent fitting on top of the GC is the vent flow plus the septum purge flow.

204 . Clarus 690 GC Customer Hardware and Service Guide

Split P neum atic Operation Once the split system has been configured, you may modify the default setpoints in the active GC method, edit the setpoints in stored methods, or generate a new method and enter split pneumatic setpoints. The following parameters control split pneumatic operation. Initial Carrier Setpoint The PPC controller takes setpoints of flow, velocity or pressure depending on the configured split pneumatic carrier gas mode. It receives its initial setpoint from the active GC method. The initial setpoint is established upon setting up a method, ending a run, or otherwise resetting the instrument. The PPC controller holds the setpoint constant at the initial value if no additional carrier-gas entries are made in the GC method. Carrier Gas Program The GC method accommodates up to three additional carrier-gas program steps consisting of a hold time, ramp rate, and plateau setpoint. Sequential setpoints may be greater than or less than the preceding setpoint; the PPC controller will program the split pneumatics with a positive or negative ramp as required. The carrier gas resets to the initial setpoint value in the method at the end of the GC oven program, even if a carrier-gas program is in progress. The carrier gas also resets if the end of the carrier-gas program occurs before the end of the oven program. Carrier Gas Timed Events You may also program carrier-gas timed events to produce step changes in the setpoint. Carriergas timed events can occur before or after a run starts. A carrier-gas timed event supersedes the method setpoint value at the time the event occurs. However, if the carrier-gas program in the method advances to a new step after a carrier-gas event, the new program step or ramp takes control. Split Flow or Ratio Setpoint The GC method includes an initial split flow or ratio setpoint. This setpoint is established upon setting up a method, ending a run, or otherwise resetting the instrument. The PPC controller holds the setpoint constant at the initial value if no additional carrier-gas entries are made in the GC method. Split Flow or Ratio Timed Events The split flow or ratio is changed to a new setpoint by a split timed event. Split flow or ratio events can occur before or after a run starts. A split flow or ratio event supersedes the method setpoint value at the time the event occurs.

Theory of Capillary Colum n Control This section gives an overview of the theory behind PPC. For more detailed information please refer to Reference [1]. The Clarus 690 GC Programmable Pneumatic Control system can program capillary column pressure, flow, or velocity. It accomplishes this by calculating and maintaining the column pressure drop required by the carrier-gas program and column temperature. The PPC controller then reads the actual pressure and uses that value to calculate the actual column flow rate and average linear gas velocity. The column dimensions, ambient pressure, and carrier-gas type also enter into the equations.

Carrier Gas Control . 205

Colum n Tem perature Effects

Figure 100. Carrier-gas viscosity–temperature relationships 8 The viscosity of the carrier gas changes with the column temperature. As the column temperature increases, so does the carrier-gas viscosity. The increasing viscosity raises the resistance of the carrier gas to being pushed through the column. Thus, higher column pressure drops are required to maintain column flow or average linear velocity at higher temperatures. Conversely, keeping the pressure drop across the column constant while increasing the column temperature causes the flow rate and linear velocity to drop off. Figure 101 illustrates the effect of column temperature on carrier-gas viscosity. Notice that the viscosity is different for each carrier gas, and that the “slopes” of the curves are also different in each case. For each carrier gas, the Clarus 690 GC includes an accurate model of this viscositytemperature relationship as represented by the lines in Figure 101. 3.5

(a)

Flow (mL/min)

3.0 2.5 2.0 1.5 1.0 0.5 0.0 -100 -50

0

50

100

150

200

250

300

350

400

450

Column Temperature (°C)

8

J. V. Hinshaw and L.S. Ettre, “Introduction to Open-Tubular Column Gas Chromatography,” (Advanstar Communications, 1994), p. 25. (P/N N930-6007).

206 . Clarus 690 GC Customer Hardware and Service Guide

50 45

(b)

Velocity (cm/s)

40 35 30 25 20 15 10 5 0 -100 -50

0

50

100

150

200

250

300

350

400

450

Column Temperature (°C)

Figure 101. Effect of column temperature on (a) flow and (b) velocity Column: 25-m x 250-µm i.d. Helium carrier at 12 psig constant pressure drop. The viscosity changes caused by increasing temperature affect the flow and velocity through a column with a constant pressure drop. As shown in Figure 102, both flow and velocity decrease across the column temperature range. This decreasing flow and velocity can cause loss of column efficiency as well as changes in flow-sensitive detector operation during programmedtemperature analyses. In this case, the column pressure drop was selected so that the average linear velocity was close to 30 cm/s at 50 °C; at 350 °C the velocity drops to around 20 cm/s. The effect on flow is more severe, dropping from 1.19 mL/min at 50 °C to 0.4 mL/min at 350 °C. Without PPC, compensation for this effect is difficult and limited. Although it is possible to program the PPC controller pressure to crudely compensate, it is much easier and more accurate to use either the flow-programmed or velocity-programmed modes instead.

Flow P rogram m ed Operation The PPC controller operates in a pressure-, flow-, or velocity-programmed mode. The flowprogrammed mode is useful in applications where the column flow rate must be held constant, or where the operator wishes to express the column carrier program in flow terms. Figure 103 shows the pressure program and velocity that result when a 50-m x 530 µm i.d. column operates with nitrogen carrier at a constant 4 mL/min flow across a range of temperatures. 14

Pressure drop (psig)

12

(a)

10 8 6 4 2 0 -100 -50

0

50

100

150

200

250

300

Column Temperature (°C)

350

400

450

Carrier Gas Control . 207

50 45

(b)

Velocity (cm/s)

40 35 30 25 20 15 10 5 0

-100 -50

0

50

100

150

200

250

300

350

400

450

Column Temperature (°C)

Figure 102. Effect of column temperature on (a) pressure and (b) velocity. Column: 50-m x 530-µm i.d. Nitrogen carrier at 4 mL/min constant mass flow

Velocity Program m ed Operation Programming the PPC controller in average linear gas velocity mode adjusts the column pressure drop to maintain the setpoint velocity as the oven temperature and ambient pressure change. This helps ensure better repeatability of chromatograms from one instrument to another — assuming the ovens are calibrated to a common temperature reference standard—by eliminating effects due to differences in altitude or weather. Differences in the columns, however, are not accounted for. 40

Pressure drop (psig)

35 30 25

(b)

(a)

20 15 10 5 0 -100

-50

0

50

100

150

200

250

300

350

400

450

Column Temperature (°C)

Figure 103. Effect of altitude on pressure programs. Column: 15-m x 100-µm i.d. Hydrogen carrier at 30 cm/s constant average linear gas velocity. (a): Sea level; (b): 3000 m altitude Figure 104 shows two pressure-temperature curves, one for a constant velocity of 30 cm/s on a 15-m x 100 µm i.d column with hydrogen carrier with an atmospheric pressure equivalent to sea level, and the other for the same column at the equivalent of 3 000 m altitude. When operating the oven at a constant temperature with identical columns, all solute retention times will be the same in both cases because the velocities are unchanged.

208 . Clarus 690 GC Customer Hardware and Service Guide With oven temperature programming, the retention times will be very similar but not quite the same due to slight differences in the velocity profiles across the two columns at different altitudes.

Pressure / Oventrack M ode

Pressure drop (psig)

60 50 40 30 20 10 0 -100 -50

0

50

100

150

200

250

300

350

400

450

Column Temperature (°C)

Figure 104. Pressure program in oventrack mode. Helium carrier at 25 psig, 50 °C initial conditions The P ressure: Oventrack operating mode maintains a constant column mass flow as the column temperature changes by adjusting the inlet pressure as required. No knowledge of column dimensions is required in this mode: only the initial pressure and temperature are specified. Figure 105 shows the pressure-temperature profile for Pressure: Oventrack mode with an initial pressure of 25 psig at 50 °C with helium carrier gas. This operating mode is convenient for mass-spectrometric or thermal-conductivity detectors where column dimensions are not important.

Vacuum Com pensation 45

Pressure drop (psig)

40 35 30 25 20

(b)

(a)

15 10 5 0 -100 -50

0

50

100

150

200

250

300

350

400

450

Column Temperature (°C)

Figure 105. Pressure program (a) without and (b) with vacuum compensation. Column: 60-m x 320 µm i.d. Helium carrier at 2 mL/min constant flow

Carrier Gas Control . 209 When using a mass-spectrometric detector with the column exit under vacuum, it is necessary to reduce the column pressure drop to account for the reduced exit pressure. The effect of putting the column exit under vacuum is shown in Figure 106 for a 60-m x 320 µm i.d. column with helium carrier at a 2 mL/min constant flow rate.

210 . Clarus 690 GC Customer Hardware and Service Guide

Detector Gas Flow Control The Clarus 690 GC can control up to two detectors, each with up to two primary detector gas flow controllers. The GC determines the type of controllers required for each detector by sensing the presence of the detector amplifier board. Certain detectors can use additional auxiliary detector gas control zones (see “Auxiliary Detector Gas Control,” below). The Clarus 690 GC PPC detector gas flow controller delivers a set flow rate of gas into a device (detector) at atmospheric pressure. Unlike the carrier gas controller, the detector gas controller does not compensate for significant changes in its outlet pressure. It is not suitable for providing column flow. The detector gas controller is not affected by normal changes in the incoming gas supply pressure. The PPC controller compensates for changes in each detector gas controller’s internal temperature. Figure 107 illustrates the detector flow control module.

Figure 106. Schematic of a detector gas flow control module

Detector Gas Flow Controller Operating Range The detector gas flow controller operating range depends on the detector in use. Each flow controller uses a flow-range element (fixed restrictor) that determines the available flow range for a particular gas. The nominal detector gas flow lies within 20 – 50% of the full-scale controller range for the factory-installed flow-range elements. The flow-range elements should not be changed unless the controller is to be used with a different detector. Please refer to Chapter 9, “Maintenance,” the Practical Hints section for information on individual flow-range elements for specific detectors.

N OTE: Only the fixed restrictor changes for various requirements.

Detector Gas Flow Controller Setup The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing; normally there is no need to access or modify the type or location of the PPC controllers. If the PPC hardware configuration is changed — by installing or removing PPC devices — you must access the PPC connection menus on the Clarus 690 GC keypad and modify the type and location of the PPC devices. Refer to the Clarus 690 User’s Guide (P/N 09931374), “Connecting the PPC Modules” in Chapter 6, “Using the Tools Menu,” for detailed instructions.

Detector Gas Flow Control . 211

CAUTION ATTENTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed. Ne modifiez pas les informations de connexion PPC si la configuration matérielle PPC n'a pas changé.

Detector Gas Flow Controller Calibration The detector gas mass-flow controller requires periodic calibration and the calibration is gas specific . The controller is factory calibrated to a National Institutes of Standards Technology (NIST) traceable standard with air, hydrogen, or helium gas, depending on the detector type. Recalibration is required upon changing the gas or the flow-range element; otherwise, the calibration frequency is up to the individual operator. It is good laboratory practice to recalibrate all measuring and control devices periodically. The detector gas flow controller calibration procedure consists of specifying the reference temperature and pressure; zeroing the internal sensors in the flow controller with the tank pressure off; specifying the flow-range element; selecting the gas type (for reference or makeup only); then measuring the flow rate at 20% and 40% of full-scale flow. This procedure is built into the Clarus 690 GC firmware.

Detector Gas M ass-Flow Controller Operation For flame-type detectors, setting up the detector gas flow controllers sets a default flow rate equal to the nominal flow specified for the detectors. For detectors with makeup or reference gases the flow rate defaults to 30 mL/min. Once the detector gas flow controllers have been set up and calibrated on the PPC system, the operator must enter their flow rates on the Clarus 690 GC touch screen. Unlike carrier gas flows, detector gas flow rates do not appear in the GC method. Instead, they are set in the GC configuration. There are no timed events that modify detector flows. Detector gas flow controllers may be set up as Auxiliary gas flow zones if it is necessary to incorporate them into the GC method or to modify them with timed events. See “Auxiliary Detector Gas Control” below for further information.

Flam e Gases The Clarus 690 GC automatically determines gas types for flame detectors, including the flameionization (FID), flame-photometric (FPD), and nitrogen-phosphorous (NPD). Once the flame gas controllers are set up, you may modify the default flow rates as required by entering new values under the in the GC configuration.

Flam e I gnition The PPC version of the FID has flam eout detection and an auto-ignite feature. Flameout detection occurs 0.5 minutes into the PRE-RUN time and it measures the user-entered baseline threshold level in mV. If the threshold level is exceeded, the flame is considered lit; otherwise, the Clarus 690 GC detects that the flame is out, and auto-ignite automatically lights the flame before the start of a run. You can also manually light the flame at any time by pressing the ignite button on the touch screen.

212 . Clarus 690 GC Customer Hardware and Service Guide

M akeup and Reference Gases The Clarus 690 GC automatically determines the presence of reference and makeup gases for detectors. The thermal-conductivity detector (TCD) uses reference gas. The electron-capture detector (ECD), photoionization detector (PID), and, optionally, the thermal-conductivity detector (TCD) use makeup gas. Makeup and reference gases operate in a fashion similar to flame gases, except that you may change the gas type in the PPC controller setup. Once the makeup or reference gas controllers are set up, you may modify the default flow rates as required by entering new values in the active. Detector makeup gas flow controllers may be set up as Auxiliary gas flow zones if it is necessary to incorporate them into the GC method or to modify them with timed events. See “Auxiliary Detector Gas Control” below for further information.

Auxiliary Pressure and Flow Control . 213

Aux iliary Pressure and Flow Control In addition to carrier and detector gas control, the Clarus 690 GC accommodates up to four auxiliary PPC controllers. The auxiliary zones can control carrier gas mass-flow controllers, pressure controllers, pressure readouts, and detector gas flow controllers. These zones all have an initial setpoint, and incorporate timed-event changes.

Aux iliary Carrier Gas Control The Clarus 690 GC controls up to two primary carrier-gas zones and up to four primary detector gas zones. In addition, up to four carrier gas mass flow controllers, pressure controllers, or pressure readouts may be set up as auxiliary PPC carrier zones, subject to a maximum of eight flow controllers (carrier plus detector flow) and twelve zones total in the PPC system. The auxiliary flow control zones do not accommodate split pneumatics, which must be configured as the primary carrier control zones on the GC.

Aux iliary Carrier Gas Setup The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing; normally there is no need to access or modify the type or location of the PPC controllers. Some factory installed options use auxiliary carrier-gas zones. In order to use other auxiliary zones the operator must access the PPC connection menus on the Clarus 690 GC touch screen and modify the type and location of the PPC devices. For auxiliary carrier-gas control you may choose either a carrier gas mass-flow controller, a pressure controller, or a pressure readout.

CAUTION ATTENTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed. Ne modifiez pas les informations de connexion PPC si la configuration matérielle PPC n'a pas changé.

Aux iliary Carrier Gas M ass-Flow Controller Calibration The calibration procedure for mass-flow controllers designated as auxiliary carrier gas zones is the same as for controllers designated as primary carrier zones. Please refer to the section “Mass-Flow Controller Calibration” in this chapter.

Aux iliary Carrier Gas Zone Operation Once the auxiliary carrier gas zone has been set up, you may modify the default setpoints in the active GC method, edit the carrier-gas setpoints in stored methods, or generate a new method and enter carrier-gas setpoints. The following parameters control the auxiliary carrier gas. Initial Auxiliary Carrier Gas Setpoint The auxiliary carrier gas zone takes setpoints of flow or pressure depending on the configured controller. It receives its initial setpoint from the active GC method. The initial setpoint is established upon setting up a method, ending a run, or resetting the instrument. The setpoint is held constant at the initial value if no additional auxiliary carrier entries are made in the GC method. The initial auxiliary carrier gas setpoint appears after touching Carrier Program button in the GC method.

214 . Clarus 690 GC Customer Hardware and Service Guide Auxiliary Carrier Gas Timed Events You may program auxiliary carrier-gas timed events to produce step changes in the setpoint. Auxiliary carrier-gas events can occur before or after a run starts. An auxiliary carrier-gas event supersedes the method setpoint value at the time the event occurs.

Aux iliary Detector Gas Control The Clarus 690 GC controls up to four primary detector gas zones and up to two primary carriergas zones. In addition, up to four detector gas flow controllers may be set up as auxiliary PPC detector zones, subject to a maximum of eight flow controllers (carrier plus detector flow) and twelve zones total in the PPC system.

Aux iliary Detector Gas Setup The PPC system stores information about the type, location, and calibration of its pneumatic devices in battery-backed-up random access memory (BRAM). This information is written into the BRAM during manufacturing: normally there is no need to access or modify the type or location of the PPC controllers. Some factory installed detectors use auxiliary detector gas zones. In order to use other auxiliary zones the operator must access the PPC CONNECTION menus on the Clarus 690 GC touch screen and modify the type and location of the PPC devices.

CAUTION ATTENTION

Do not modify the PPC connection information if the PPC hardware configuration has not changed. Ne modifiez pas les informations de connexion PPC si la configuration matérielle PPC n'a pas changé.

Aux iliary Detector Gas Flow Controller Calibration The calibration procedure for detector gas flow controllers designated as auxiliary detector gas zones is the same as for controllers designated as primary detector zones. Please refer to the section “Detector Gas Flow Controller Calibration” in this chapter.

Aux iliary Detector Gas Zone Operation Once the auxiliary detector gas zone has been set up, you may modify the default setpoints in the active GC method, edit the detector gas setpoints in stored methods, or generate a new method and enter detector gas setpoints. The following parameters control the auxiliary detector gas. Initial Auxiliary Detector Gas Setpoint The auxiliary detector gas zone takes flow setpoints. It receives its initial setpoint from the active GC method. The initial setpoint is established upon setting up a method, ending a run, or resetting the instrument. The setpoint is held constant at the initial value if no additional auxiliary detector gas entries are made in the GC method. Auxiliary Detector Gas Timed Events You may program auxiliary detector gas timed events to produce step changes in the setpoint. Auxiliary detector gas events can occur before or after a run starts. An auxiliary detector gas event supersedes the method setpoint value at the time the event occurs.

PPC Tips and Techniques . 215

P PC Tips and Techniques This section contains tips and techniques that you will find useful when using PPC.

Effect of Flow on P ressurization R ate When first turned on, when a higher setpoint pressure is programmed, the PPC carrier-gas zones require some time to come up to pressure. The pressurization rate is a function of the type of inlet, the incoming carrier-gas flow rate and the outgoing column and septum purge flow rate (if applicable). For typical packed columns the inlet will pressurize within one minute. For a POC inlet where the column flow rate is low it may take longer to reach a steady-state inlet pressure. The PPC gasleak monitor will not trigger until five minutes have elapsed, which should be enough time in all cases. For capillary or PSS inlets, setting a high inlet pressure and a low split flow rate (but not a zero flow) can result in the inlet never coming to pressure because the incoming split flow is less than the column plus septum purge rates. A zero split-flow rate is treated as a special condition for splitless injection and will not cause this problem. In general, use a minimum split flow rate that is the greater of 20 mL/min or four times the sum of the column and septum-purge flows.

N OTE: If the column pressurization rate (psig/min) is too low, increase the split flow rate during this time. If you want to quickly pressurize the column in the split mode, temporarily increase the split vent flow while pressurizing the column. You can set the split vent flow up to a maximum of 500 mL/min.

Correcting the Colum n Dim ensions The PPC controller relies on operator input of column dimensions to calculate pressure – flow and pressure – velocity relationships. If the column dimensions are incorrect, then the calculations will be inaccurate. In general, actual column diameters are within ±2 µm of the nominal specified value. Column lengths may vary for used columns because the ends may have been cut off. Alternatively, the column dimensions may be unknown. In such cases set the Carrier mode to pressure and oventrack, if you need to maintain a constant column flow rate.

Correcting the Colum n Length The average linear gas velocity is the most accurate measure of capillary column conditions, because the low flow rates are difficult to measure. By measuring the average velocity you can compare it to the displayed velocity on the GC, and then make a correction to the column length. This procedure assumes that the column inner diameter entered into the Clarus 690 GC is correct. Set Up the Clarus 690 GC Install a column and turn on the GC. In Configuration set the CAPILLARY CONTROL MODE ON, and choose the VELOCITY mode. Enter the nominal column length and exact diameter. Select the split FLOW mode. In the Method, accept the default velocity of 30 cm/s, or choose another value. Set the oven temperature to at least 100 °C, isothermal. Determine the Unretained Peak Time Inject 1–5 µL of methane (natural gas) and time how long it takes for the methane peak to elute from the column. This is best done by recording the methane chromatogram with a data handling

216 . Clarus 690 GC Customer Hardware and Service Guide system and reading the methane peak time off the chromatogram or report. You may use methylene chloride (CH2Cl2) vapor for an ECD. Calculate the Actual Average Linear Gas Velocity Calculate the average linear gas velocity ( u ) from this formula:

u=

LPPC X 10 tM

where LPPC is the column length (in meters) entered in the Clarus 690 GC and tM is the measured methane peak time in seconds. Compute the Correction Factor Compare the measured velocity ( u ) to the displayed velocity ( u PPC ) and compute the column length correction factor (f):

u f = PPC u If the correction factor is less than 0.8 or greater than 1.2, then you should double check the column diameter and nominal length for errors. Enter the Corrected Column Length Now, go back to the GC configuration and enter the corrected column length (Lnew):

Lnew = f ⋅ LPPC Check the New Length Upon entering the new length the PPC controller will change the inlet pressure up or down to reflect the new conditions. If the measured velocity was less than the displayed value, then the PPC controller will reduce the pressure so that the measured velocity is now equal to the displayed value. If the measured velocity was higher, then the PPC controller will increase the inlet pressure. Check the new value by measuring the methane peak time again. Your calculated linear velocity should now be very close to the displayed value.

Splitless I njection Splitless injection refers to a technique that has a zero split flow rate during injection with an increased split flow rate after injection is complete. There are two ways to accomplish splitless injection on the Clarus 690 GC.

Zero I nitial Split Flow For this method, you set the initial split flow to zero and enter a timed event after injection (typically at 1–2 minutes) that sets the split vent flow to 50–100 mL/min. The Clarus 690 GC will automatically turn off the split vent flow at the end of the run.

Zero Split Flow by P re-R un Event In some cases — with a PSS inlet in particular — it may not be desirable to have zero split vent flow while the instrument is cooling and recycling for another run. Set the initial split vent flow to 50–100 mL/min. Then enter a timed event at -2 minutes to set the split vent flow to 0, and enter

PPC Tips and Techniques . 217 another event after injection (typically at 1–2 minutes) that sets the split vent flow to 50–100 mL/min.

P ressure-P ulse I njection Pressure-pulse injection refers to a technique that facilitates sample transfer from inlet to column by raising the inlet pressure (column flow and velocity) above the normal level for injection, and then lowering it back to a normal setting for peak elution. To perform pressure-pulse injection, enter a Pre-Run timed event at −1 minute that raises the pressure, flow, or velocity — depending on the operating mode — to a level 2–20 times the method’s initial setpoint value. Then enter another timed event 1–2 minutes into the run that reestablishes the initial method setpoint. If you are using a carrier-gas program, then set a 1–2 minute step 1 hold time in the program. The carrier-gas program will “take over” the PPC system at the end of the hold period. To discharge the pressure-pulse, the split flow must have a nonzero volume at the time of discharge. To rapidly change the pressure pulse, the split flow should be zero or as large as possible during pressurization.

Reducing Carrier-Gas Consum ption You can significantly reduce carrier-gas consumption with a split or PSS inlet by setting the initial split flow to 20 mL/min, using a Pre-Run timed event at −2 minutes to increase the flow for split injection or to set the flow to 0 for splitless injection, and then reducing the split flow back to 20 mL/min after completion of the injection. You should delay the flow reduction until at least 5 minutes or longer after injection, or after a PSS inlet has reached its final temperature. Don’t forget to include a timed event at 1–2 minutes that sets the flow to 50–100 mL/min for splitless injection.

218 . Clarus 690 GC Customer Hardware and Service Guide

PPC Tips and Techniques . 219

9 M aintenance

220 . Clarus 690 GC Customer Hardware and Service Guide

Overview . 221

Overview This chapter contains procedures for: •

Autosampler Maintenance — changing a syringe, replacing a vial-locator mechanism and cleaning the autosampler.



Syringe Maintenance — cleaning the 5-µL and 50-µL syringe plungers.



Injector Maintenance — changing septa, replacing a gold seal, changing and repacking packed-injector liners, changing and repacking injector liners on the capillary (CAP), programmed split/splitless (PSS), and programmed on-column (POC) injectors, and changing the charcoal trap on the split/splitless injector.



FID Maintenance — replacing the FID jet, cleaning the FID jet, replacing an O-ring in the FID collector, and cleaning the FID collector and cap.



ECD Maintenance — baking out ECD cells, cleaning the ECD anode, and wipe testing an ECD cell.



PID Maintenance — changing PID lamps, cleaning PID lamp windows, changing PID lamp window seals, and positioning disks.



NPD Maintenance — changing and conditioning the NPD bead and replacing an NPD jet.



FPD Maintenance — cleaning or replacing an FPD optical filter assembly, cleaning the detector liner, cleaning/replacing the detector window, replacing the photomultiplier tube, and cleaning and replacing the FPD jet.



PPC Maintenance — replacing PPC module restrictors.

222 . Clarus 690 GC Customer Hardware and Service Guide

Autosam pler M aintenance Autosampler maintenance consists of changing a syringe and replacing a vial locator mechanism.

Changing a Syringe

1. From the System Status screen select the Run icon.

2. On the Run Type page select the Autosampler radio button and the Task tab. The following page will appear.

3. Select the Park button and the autosampler tower moves to the park position (facing the front of the Clarus 690 GC). 4. Open the tower door on the autosampler tower cover.

Autosampler Maintenance . 223

Figure 107. Autosampler tower in the park position

R em oving a Syringe 1. Locate the plunger assembly shown in the previous illustration. Then, refer to the following figure, as you lift up the plunger cap handle and rotate it until it rests on the collar. Then release the plunger cap handle.

Figure 108. Plunger assembly 2. Hold the syringe by the barrel or syringe nut (see the following figure) and turn the carriage thumbscrew clockwise until the syringe is free. 3. Gently pull the top of the syringe forward until it just clears the carriage assembly. 4. Gently lift the syringe out of the carriage assembly.

224 . Clarus 690 GC Customer Hardware and Service Guide

Figure 109. Removing a syringe

I nstalling a Syringe Please refer to Figure 109 through Figure 111 as you follow these steps. 1. Guide the needle through the hole in the carriage thumbscrew, and then thread the needle through the needle guide. Use your fingers as a guide. 2. Rest the top of the plunger on the plunger cap slide, which is a shelf located on the underside of the plunger assembly. 3. While holding the syringe nut, engage the carriage thumbscrew on the threaded part of the syringe by turning the carriage thumbscrew counterclockwise. 4. Continue turning the thumbscrew counterclockwise. This slowly lowers the needle. Carefully guide the needle through the needle guide into the vial locator. 5. Tighten the carriage thumbscrew.

Figure 110. Installing a syringe

Autosampler Maintenance . 225

Replacing the Vial-Locator M echanism The vial-locator mechanism will wear out with extended use and require replacement. If the autosampler begins missing vials, or if the hole for the syringe needle begins to plug, it is an indication that you should replace the vial-locator mechanism To replace a vial-locator mechanism: 1. Remove the two shoulder screws that secure the locator to the autosampler tower frame. Remove the two springs, then remove the vial locator. Discard the vial locator. 2. Mount the new vial locator (P/N N6101182) on the autosampler tower frame. 3. Install the two shoulder screws through the two springs and into the vial locator. This secures the vial locator to the autosampler tower frame.

WARNING AVERTISSEMENT

When securing the vial-locator molding, be sure that the flag is centered (not touching either side) in the sensor. If it touches a side, adjust the flag by loosening and then tightening the screws. DO N OT ADJUST THE SENSOR. Lors de la fixation du moulage du flacon-localisateur, assurezvous que le drapeau est centré (ne touchez pas les deux côtés) dans le capteur. Si elle touche un côté, ajustez le drapeau en desserrant puis en serrant les vis. NE PAS RÉGLER LE CAPTEUR.

Autosampler Tower Frame

Shoulder Screw (1 of 2) (P/N 09907998) Spring (1 of 2) (P/N N6101183)

Flag

Vial Locator (P/N N6101182)

Figure 111. Exploded view of the vial locator

226 . Clarus 690 GC Customer Hardware and Service Guide

Cleaning the Autosam pler Tray If you have a spill on the autosampler tray clean with a damp cloth. Do not use any abrasive cleaners on the tray. Remove the tray from the autosampler, clean and return to the autosampler.

Figure 112. Clarus 690 Tray (108 positions)

Syringe Maintenance . 227

Syringe M aintenance Syringe maintenance consists of cleaning the 5-µL and 50-µL syringe plungers and servicing idle syringes.

Cleaning the 5- µL and 50- µL Syringe P lungers The 5-µL and 50-µL syringe plungers should be cleaned regularly, after approximately 500 injections, since insolubles can build up and cause friction. To clean the syringe plunger: 1. Remove the syringe using the procedure described in the preceding section. 2. Remove the plunger from the syringe barrel. 3. Wipe the plunger with a tissue soaked in an appropriate solvent. 4. Replace the plunger. 5. Pull and expel the same solvent through the barrel several times. 6. Replace the syringe using the procedure descried in the preceding section.

N OTE: Only syringes distributed by PerkinElmer should be used with the Clarus 690 GC. Plungers are not interchangeable from syringe to syringe.

Servicing I dle Syringes Syringes that are not used for several hours could "freeze," i.e., the syringe plunger will not move. To avoid this condition, PARK the tower, then remove and clean the syringe plunger as described above.

N OTE: If you notice the Clarus 690 GC precision degrading, replace the syringe. The Clarus 690 GC syringe is a consumable part. After extended use, you will need to replace it.

228 . Clarus 690 GC Customer Hardware and Service Guide

I njector M aintenance

CAUTION ATTENTION

If you are analyzing reactive compounds, you should use deactivated liners and wool which are appropriate for your sample type. Si vous analysez des composés réactifs, vous devez utiliser des doublures et de la laine désactivées qui conviennent à votre type d'échantillon.

Injector maintenance consists of changing septa, changing and repacking injector liners, changing the hourglass needle guide on the POC injector, changing and repacking CAP and PSS injector liners, removing a broken liner from the PSS injector body, changing the charcoal trap or replacing charcoal on the split/splitless CAP and PSS injectors.

Changing Septa Septa should be replaced on a regular basis. How often depends on the type of septa used, the temperature of the injection port, and the number of injections made. The septum shipped with your instrument is a Thermogreen LB-2 Septa (P/N N6621028, package of 50). This septum can handle over 200 injections at moderate temperatures. To change a septum: 1. Turn off the injector heater and allow the injector to cool. 2. Remove the injector covers to access the Septum cap

Figure 113.

Removing the injector covers.

3. Remove the septum cap. 4. Remove the old septum from the septum cap or from the top of the injector. Sometimes the septum may get stuck in the septum cap.

Injector Maintenance . 229

Figure 114.

Removing the septum cap.

5. Insert a new septum in the recess in the top of the injector. 6. Replace the septum cap.

N OTE: To minimize the possibility of contamination, avoid unnecessary handling of septa.

Replacing a CAP I njector Gold Seal CAUTION ATTENTION

A gold seal should never be reused. Once the gold seal is removed, it should be replaced with a new gold seal. Reusing a gold seal could result in damage to the injector base. Un sceau d'or ne doit jamais être réutilisé. Une fois le joint d'étanchéité or retiré, il devrait être remplacé par un nouveau sceau d'or. Réutiliser un joint d'étanchéité d'or pourrait endommager la base de l'injecteur.

To replace a gold seal in the CAP injector: 1. Locate the injector connector in the column oven.

Figure 115. Capillary column attached to capillary injector fitting.

230 . Clarus 690 GC Customer Hardware and Service Guide 2. Unscrew and remove the column nut and column inlet.

Figure 116. Gold seal nut inside the oven. 3. Locate the gold seal nut inside the oven. 4. Using a 9/16-in. wrench, loosen and remove the gold seal nut from the CAP injector.

Figure 117. Gold seal nut inside the oven. 5. Remove the gold seal from the gold seal nut and replace it with a new gold seal (P/N 09200623).

Injector Maintenance . 231

Figure 118. Gold seal nut, gold seal, and washer. 6. Install the gold seal nut back into the CAP injector by turning it until it is finger tight. Then using a 9/16-in wrench tighten it another 1/8-turn to make a leak-free seal. DO NOT OVERTIGHTEN! 7. Carefully insert the column and column nut into the gold seal nut; then hand-tighten the column nut onto the capillary injector fitting. 8. Carefully push/slide the column into the capillary injector fitting until the septum (or mark) is aligned with the back of the column nut.

Figure 119. Column, gold seal nut, and capillary column nut. 9. Using one 1/4-inch wrench, tighten the column nut only until the column cannot be pulled out of the fitting. CAUTION ATTENTION

TIP

Do not overtighten column nuts. Overtightening can cause damage to the ferrule and/or column. Ne pas trop serrer les écrous de colonne. Le serrage excessif peut endommager la ferrule et / ou la colonne. To ensure a leak-free connection, cycle the oven temperature to your maximum temperature set in the method then cool. Once cool recheck the tightness of the column nut then retighten if necessary.

232 . Clarus 690 GC Customer Hardware and Service Guide

Changing and R epacking P acked Colum n I njector Liners To improve the performance of the injector used with packed columns, insert a small amount of quartz wool (P/N N6102354) into the top portion of the injector liner (P/N N6101048). The quartz wool accomplishes two things: (1) it wipes the end of the syringe needle to insure that reproducible sample volumes are injected, and (2) it retains any nonvolatile components present in your sample, making cleaning the liner easier. The injector liner should be removed and the wool packing replaced on a regular basis, particularly if your samples contain nonvolatile components that could build up on the wool. This could cause adsorption of peaks of interest, tailing, and loss of sensitivity. You can remove the wool with a small hook on the end of a thin wire, or blow it out using compressed air. To remove a packed injector liner and install new wool: 1. Turn off the injector heater. Allow the injector to cool until it is slightly warm to the touch. Cooling the injector to a temperature that is too low (